No part of this manual may be reproduced
in any form or by any means (including
electronic storage and retrieval or translation into a foreign language) without prior
agreement and written consent from
Keysight Technologies as governed by
United States and international copyright
laws.
Manual Part Number
16860-97003
Edition
Second Edition, July 2017
Printed in Malaysia
Published By
Keysight Technologies
Plot No. CP-11, Sector-8
IMT Manesar, Gurgaon - 122051, India
Technical Assistance
ccr.keysight.com@keysight.com
www.keysight.com/find
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS SUBJECT TO BEING CHANGED, WITHOUT
NOTICE, IN FUTURE EDITIONS. FURTHER,
TO THE MAXIMUM EXTENT PERMITTED BY
APPLICABLE LAW, KEYSIGHT DISCLAIMS
ALL WARRANTIES, EITHER EXPRESS OR
IMPLIED WITH REGARD TO THIS MANUAL
AND ANY INFORMATION CONTAINED
HEREIN, INCLUDING BUT NOT LIMITED TO
THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT
BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN
CONNECTION WITH THE FURNISHING,
USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED
HEREIN. SHOULD KEYSIGHT AND THE
USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT
THAT CONFLICT WITH THESE TERMS, THE
WARRANTY TERMS IN THE SEPARATE
AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software described in
this document are furnished under a
license and may be used or copied only in
accordance with the terms of such license.
U.S. Government Rights
The Software is "commercial computer
software," as defined by Federal Acquisition
Regulation ("FAR") 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of
Defense FAR Supplement ("DFARS")
227.7202, the U.S. government acquires
commercial computer software under the
same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to
U.S. government customers under its standard commercial license, which is embodied in its End User License Agreement
(EULA), a copy of which can be found at
http://www.keysight.com/find/sweula. The license set
forth in the EULA represents the exclusive
authority by which the U.S. government
may use, modify, distribute, or disclose the
Software. The EULA and the license set
forth therein, does not require or permit,
among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial
computer software documentation that is
not customarily provided to the public; or
(2) Relinquish to, or otherwise provide, the
government rights in excess of these rights
customarily provided to the public to use,
modify, reproduce, release, perform, display, or disclose commercial computer
software or commercial computer software
documentation. No additional government
requirements beyond those set forth in the
EULA shall apply, except to the extent that
those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR
and the DFARS and are set forth specifically
in writing elsewhere in the EULA. Keysight
shall be under no obligation to update,
revise or otherwise modify the Software.
With respect to any technical data as
defined by FAR 2.101, pursuant to FAR
12.211 and 27.404.2 and DFARS 227.7102,
the U.S. government acquires no greater
than Limited Rights as defined in FAR
27.401 or DFAR 227.7103-5 (c), as applicable in any technical data. 52.227-14
(June 1987) or DFAR 252.227-7015 (b)(2)
(November 1995), as applicable in any
technical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls
attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in
damage to the product or loss of important
data. Do not proceed beyond a CAUTION
notice until the indicated conditions are
fully understood and met.
A WARNING notice denotes a hazard. It
calls attention to an operating procedure,
practice, or the like that, if not correctly
performed or adhered to, could result in
personal injury or death. Do not proceed
beyond a WARNING notice until the indicated conditions are fully understood and
met.
2Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 3
16860 Series Logic Analyzers—At a Glance
Pods for Probe
Inputs
Run/Stop Keys
USB 2.0 Ports (2)
Scroll Knob
Standby (Power) Button
Touch Screen
On/Off Button
The Keysight Technologies 16860 series logic analyzers are portable logic analyzers that
range from 34 to 136 logic acquisition channels, depending on the model.
Table 1 Model comparisons
Keysight model number16861A16862A16863A16864A
Logic acquisition channels3468102136
Features, Logic Acquisition
•2 M to 400 M samples of memory depth per channel (depending on memory option),
software upgradeable.
•350 MHz or 700 MHz maximum state clock rate (depending on state speed option),
software upgradeable.
•Full Channel Timing Mode at 2.5 GHz sampling with 12.5 GHz Timing Zoom.
•Half Channel Timing Mode at 5.0 GHz sampling with 12.5 GHz Timing Zoom
•Quarter Channel Timing Mode at 10 GHz sampling with 12.5 GHz Timing Zoom
Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide3
•Eye scan (automatic threshold and sample position setup) feature.
•Single-ended and differential probing support
Page 4
Features, Mainframe
•Built-in 15 inch color touch screen with 1024 x 768 resolution
•Removable hard drive
•10/100/1000 Base-T LAN port
•USB 2.0 ports (six total, two on front, four on back)
•One PCI expansion slot
•One PCI Express x1 expansion slot
•Windows 10 installation
Windows 10 support has been added to the 16860-series logic analyzers and these
are now shipped from the factory with Windows 10 operating system preinstalled.
Earlier, Windows 7 operating system was preinstalled on these logic analyzers.
•Keysight Logic and Protocol Analyzer application which takes the complexity out of
making logic analyzer measurements. You can perform all operations directly from
one window.
Supplied Accessories
•USB mouse
•USB keyboard
•Accessory pouch and power cord
Snap the accessories pouch to the top of the 16860 series logic analyzer. Use it to
store probe leads, accessories, or manuals. Use the tie-down straps under the flap to
conveniently hold pod cables not in use or during transport.
Optional Accessories
•Probes
Front Panel
Refer to the online help in the logic and protocol analyzer software for information on
using the analyzer.
Rear Panel
Before you turn on the system, connect your mouse and keyboard to the 16860 Series
logic analyzer rear panel.
16860-Series Logic Analyzer Dimensions
All dimensions given below are in Millimeters (mm).
4Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 5
Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide5
Page 6
In This Guide…
This guide describes how to set up, use, and update your 16860 series logic analyzer. It
also contains information on solving problems and additional safety information.
1First-Time Set Up
This chapter describes the steps in setting up the logic analyzer for the first time.
2Getting Started with the Logic Analyzer
This chapter describes some basic features of the logic analyzer and tells you where to
find more information on using the logic analyzer.
3Updating the Logic Analyzer
This chapter describes how to change your logic analyzer’s network and Windows
Firewall settings. It also describes how to get and install software updates.
4Solving Problems
This chapter describes the logic analyzer’s built-in self tests and recovering system
software.
5Safety Notices
This chapter contains additional safety information.
6Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 7
Contents
1 First-Time Set Up
2 Getting Started with the Logic Analyzer
Proper Cooling12
First-Time Set Up Considerations13
Determine Whether the Instrument will be Connected to the
Network
Get Network Setup Information13
First-Time Set Up On 16860 Series Logic Analyzers14
Additional First-Time Setup Steps16
Establishing an Administrator Account (recommended for standalone and
networked instruments)
Protecting the Instrument’s Operating System16
13
16
Contents
Probing18
Making a Measurement19
Snap to Edge Markers21
Left-Click Menus22
Tree Structure Bus/Signal Names22
Offline Analysis23
Marker Overview23
Tool Tips24
Trigger History25
Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide7
Page 8
Contents
To close the Windows OS and place the system in a standby
state26
Using the Windows Shutdown26
Using a Short Press of the Power Button26
Using a Long Press of the Power Button27
Unplugging the Power Cord or Power Loss27
Specifications and Characteristics28
3 Updating the Logic Analyzer
Changing Network Settings32
Changing Windows Firewall Settings33
To give other applications/ports access through the firewall33
To restore the logic analyzer firewall defaults33
Software Installation and Upgrades35
Keysight Logic and Protocol Analyzer Software35
Optional Products35
4 Solving Problems
Running Self-Tests38
If the logic analyzer powers itself down40
If the Power LEDs do not come on40
If you get a registry error and the system won’t boot40
If there are network connection problems41
If there are touch screen problems42
Using the System Recovery Software45
Contacting Keysight Service/Support49
5 Safety Notices
Warnings52
8Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 9
Index
Contents
To clean the instrument54
Safety Symbols55
Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide9
Page 10
Contents
10Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 11
Keysight 16860 Series Portable Logic Analyzers
Installation/Quick Start Guide
1First-Time Set Up
Proper Cooling / 12
First-Time Set Up Considerations / 13
First-Time Set Up On 16860 Series Logic Analyzers / 14
Additional First-Time Setup Steps / 16
Page 12
1First-Time Set Up
NOTE
Proper Cooling
Allow at least 5 cm (2 inches) of space between instruments for proper
cooling.
This instrument requires 100-120VAC 50/60/400Hz & 100-240VAC 50/60Hz mains
capable of supplying 325 watts. Do not apply more than 25 volts to any Probe
Adapter signal line. Do not apply more than 5.5 volts to the BNC Trigger In or Trigger
Out.
This instrument is intended for office and lab use only in:
- Temperature range of 5-40°C
- Altitude upto 4000 m
- Maximum Humidity 90% non-condensing at +40°C.
12Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 13
First-Time Set Up Considerations
NOTE
Determine Whether the Instrument will be Connected to the Network
If you plan to use COM automation to programmatically control a logic analyzer from a
remote computer, be aware that the remote computer and the logic analyzer must both
have the same kind of network membership; that is, they must both be members of a
"Workgroup", or they must both be members of a "Domain". This is due to the security
requirements of Distributed COM.
Standalone
You are going to use the logic analyzer as a standalone instrument (no
LAN connection). Printing requires a printer attached to the logic analyzer,
sharing files is done via USB flash drive or floppy disk.
Connected to a Network
If you require the ability to print, share files, remotely control the logic
analyzer, etc. over the network, you need to either join a workgroup or
domain.
First-Time Set Up1
Get Network Setup Information
For networking, find out whether the network that your instrument will be
connected to uses the Microsoft Work group model or Microsoft Domain
model. In order for your instrument to perform tasks over the network
(print, transfer files, etc.) your instrument must be configured
appropriately for the networking model in use at your site.
Information you may require to complete your network configuration may
include, but is not limited to:
• The Workgroup or Domain name.
• Methodology for assigning IP addresses at your site (for example,
DHCP).
• Domain Name Service (DNS).
Other network information that may be readily available from your network
administrator.
Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide13
Page 14
1First-Time Set Up
First-Time Set Up On 16860 Series Logic Analyzers
The first time you turn on a 16860 series logic analyzer (shipped with
Windows 10 installation), you are automatically logged on to the system
using the Administrator account and the Windows 10 desktop is displayed.
Note: Windows 10 support has been added to the 16860-series logic
analyzers and these are now shipped from the factory with Windows 10
operating system preinstalled. Earlier, Windows 7 operating system
was preinstalled on these logic analyzers. I
Once you are logged on to the system, you must set the following logic
analyzer configurations based on how you want to use your logic analyzer.
You use the standard Windows 10 procedures to set these configurations.
Table 2 Logic Analyzer Configurations on a 16860 Series Logic Analyzer with Windows 10
Configuration TaskLogic Analyzer Usage
Standalone (No LAN)WorkgroupDomain
Protect your instrument’s PCN/ATurn on Automatic Updates Turn on Automatic Updates
Specify computer nameIf your organization has a standard naming scheme, use that scheme. Otherwise, choose a descriptive
Specify the administrator passwordKeysight recommends entering an administrator password for administrator accounts. See page 16 for
Specify the domain name (if applicable)No
Enter your domain settings.N/AN/AEnter the name/password of the
Configure connectivity to the InternetN/AN/AChoose the appropriate Internet
Set up a high-speed connectionN/AN/AEnter either the fixed IP/DNS
name (less than 15 characters)
actions that require administrator accounts.
(Is given WORKGROUP
membership)
Installation
No
(Is given WORKGROUP
membership)
Yes
Enter the domain name the logic
analyzer will log onto.
user who can add the logic analyzer
to the domain.
connectivity.
assigned by your network
administrator or select “Obtain
Automatically”.
14Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 15
Configuration TaskLogic Analyzer Usage
Standalone (No LAN)WorkgroupDomain
First-Time Set Up1
Determine if you want to connect the
instrument’s PC to the Internet directly
or through a network.
Register with MicrosoftThe 16860 series of Logic Analyzers have the Windows 10 operating system already enabled to run.
Specify users for the instrument’s PCYou can add multiple users with administrator permissions to allow these users to use the logic analyzer.
N/AKeysight does not support
connecting to the Internet
through a modem directly
connected to the logic
analyzer.
This registration is to allow Microsoft to contact you.
N/A
Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide15
Page 16
1First-Time Set Up
Additional First-Time Setup Steps
Establishing an Administrator Account (recommended for standalone and networked instruments)
Protecting the Instrument’s Operating System
Keysight recommends that you set up an Administrator account for your
logic analyzer. You have to be logged in to the Administrator account to do
the following:
• Install most software (applies whether the instrument is networked or
standalone).
• Change the workgroup.
• Perform "Windows Updates".
• Change firewall settings.
Microsoft recommends the following three steps to ensure the
instrument's PC is protected.
1 Use an Internet firewall.
2 Get operating system updates.
3 Use up-to-date antivirus software.
16Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 17
Keysight 16860 Series Portable Logic Analyzers
Installation/Quick Start Guide
2Getting Started with
the Logic Analyzer
Probing / 18
Making a Measurement / 19
Snap to Edge Markers / 21
Left-Click Menus / 22
Tree Structure Bus/Signal Names / 22
Offline Analysis / 23
Marker Overview / 23
Tool Tips / 24
Trigger History / 25
Specifications and Characteristics / 28
Page 18
2Getting Started with the Logic Analyzer
Probing
Probing is the key to effective and efficient use of logic analyzers. Keysight
Technologies offers a wide variety of probing accessories that support
general-purpose and application-specific measurement needs. We
provide reliable, electrically and mechanically unobtrusive probes that
make it easy to connect your Keysight logic analyzer to your system under
test. For more information on Keysight probes:
• See the quick reference card that came with your logic analyzer for an
overview of probing solutions.
•Go to www.keysight.com/find/logic_analyzer_probes for the most
up-to-date information on probe compatibility with a given logic
analyzer and purchasing information.
•Go to www.keysight.com and search for Probing Solutions for Logic
Analysis Systems (publication 5968-4632E) for a catalog of detailed
probing information.
To connect and set up a probe, follow the instructions that came with the
probe.
18Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 19
Making a Measurement
Getting Started with the Logic Analyzer2
1 Start up the logic analyzer, connect to your target system, and select
Setup>Bus/Signal.
2 Assign buses and signals and click OK.
Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide19
Page 20
2Getting Started with the Logic Analyzer
3 Enter a trigger value in the Waveform View.
4 Acquire and view data.
20Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 21
Snap to Edge Markers
Getting Started with the Logic Analyzer2
You can set the markers to jump to the nearest edge of a waveform. A
yellow target will show which edge the marker will jump to.
Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide21
Page 22
2Getting Started with the Logic Analyzer
Left-Click Menus
Drag a box around parts of the data. When you release the mouse button,
a menu appears.
Tree Structure Bus/Signal Names
Show individual signals in buses by clicking on the + to expand.
22Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 23
Offline Analysis
Marker Overview
Getting Started with the Logic Analyzer2
Install the Keysight Logic and Protocol Analyzer application on another
Windows 7/10 or Windows 2000 computer and you can analyze data
without the 16860 Series logic analyzer.
1 Follow the instructions in “Software Installation and Upgrades" on
page 35 to load the 16860 Series logic analyzer software on your PC.
2 Acquire a trace with the logic analyzer.
3 Save a configuration file with data.
4 Copy the file to the PC being used for offline analysis.
5From the Keysight Logic and Protocol Analyzer application software on
your PC, open the configuration file.
The marker overview display shows marker locations within the data and
the part of data currently being displayed.
Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide23
Page 24
2Getting Started with the Logic Analyzer
Tool Tips
Pause the cursor over a tool Icon and a “Tool Tip Window” will appear.
24Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 25
Trigger History
Getting Started with the Logic Analyzer2
Save and recall trigger settings.
Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide25
Page 26
2Getting Started with the Logic Analyzer
NOTE
To close the Windows OS and place the system in a standby state
You can place the logic analyzer in a standby state by using either
Windows Shutdown or a short press of the power button on the front
panel of the logic analyzer.
When powering off the logic analyzer, wait until the fans stop turning (about 15 seconds)
before turning the logic analyzer back on. This ensures that internal circuitry restarts in a
known state.
Using the Windows Shutdown
On the logic analyzer desktop click Start>Shut Down or if you are running
remote desktop, click Start>Windows Security>Shut Down. This:
• Closes all programs that are running.
• Writes all data to the disk.
• Turns off the power supply except for the motherboard standby power.
• If the system is unplugged while it is off and then plugged back in, the
system will not power on until the power button is pressed.
Using a Short Press of the Power Button
Pressing the power button on the logic analyzer for a short time (less than
2 seconds):
• Closes all programs that are running.
• Writes all data to the disk.
• Turns off the power supply except for the motherboard standby power.
If the power button on the Remote Desktop is depressed for a short time
(less than 2 seconds or so), a message on the logic analyzer screen (not on
the remote desktop) will pop up asking if you really want to power down.
Clicking the 'yes' button results in the above events. Clicking 'no' will
prevent the shutdown. Not answering the dialog box (clicking neither "Yes"
nor "No") will have no effect. Further short presses on the power button
will have no effect.
26Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 27
Using a Long Press of the Power Button
CAUTION
CAUTION
NOTE
Only use a long press of the power button when the logic analyzer is unresponsive to
Windows Shutdown or a short press of the power button.
Pressing the power button for more than 4 seconds will place the system
in a standby state abruptly:
• Programs that are running will not be properly shut down. Any data that
has not been written to the disk will be lost.
• The power supply will be turned off except for the motherboard standby
power.
• If the system is unplugged while it is off and then plugged back in, the
system will not power on until the power button is pressed
Unplugging the Power Cord or Power Loss
Only unplug the power cord when the logic analyzer is unresponsive to all other methods
for powering off the system.
Getting Started with the Logic Analyzer2
Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide27
The main power cord is to be used as the system disconnecting device.It
disconnects the mains circuit from the mains supply. When the system is
plugged back in, it will power up and boot into Windows.
Position the logic analyzer such that it is not difficult to unplug the main power cord.
Page 28
2Getting Started with the Logic Analyzer
CAUTION
CAUTION
CAUTION
Specifications and Characteristics
This product is designed for use in Installation Category II and Pollution Degree 2
environment.
This instrument has auto-ranging line voltage input. Ensure that the supply voltage
is within the specified range and voltage fluctuations do not to exceed 10 percent of
the nominal supply voltage.
Do not apply more than 5.5 volts to Trigger In BNC or Trigger Out BNC.
The following electrical and operating characteristics are not
specifications, but are typical operating characteristics for the Keysight
16860 series logic analyzers.
28Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Electrical Characteristics
Power Requirements16861A, 16862A, 16863A, and 16864A: 00-120VAC 50/60/400Hz &100-240VAC
50/60Hz mains capable of supplying 325 watts.
CAT II (Line voltage in appliance and to wall outlet).
Installation category II, Pollution degree 2.
More specifications and characteristics for your instrument and
measurement modules are in the on-line help. To find these, go to:
1Click Help>Help Topics.
Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide29
Page 30
2Getting Started with the Logic Analyzer
2Click the Logic Analyzer - Specifications and Characteristics help book.
30Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 31
Keysight 16860 Series Portable Logic Analyzers
Installation/Quick Start Guide
3Updating the Logic
Analyzer
Changing Network Settings / 32
Changing Windows Firewall Settings / 33
Software Installation and Upgrades / 35
Page 32
3Updating the Logic Analyzer
Changing Network Settings
On Windows 7
To make network changes after the first-time power up:
The following steps are applicable for a logic analysis system with
Windows 7 operating system:
• To change from a workgroup to a domain or vice versa, click Computer
and select Properties. Click the Change Settings link in the Computer name, domain and workgroup settings section. Then select the Computer Name tab
and click Change.
• To change network connection properties, click Start>Control
Panel>Network and Sharing Center. Select the connection you want to
change and then set the properties of this connection.
For more information on changing network settings, refer to the Windows
7 online help.
On Windows 10
To make network changes after the first-time power up:
The following steps are applicable for a logic analysis system with
Windows 10 operating system:
• To change from a workgroup to a domain or vice versa, click Control
Panel from the Startup menu, and then click System. Click the Change
Settings link in the Computer name, domain and workgroup settings section.
Then select the Computer Name tab and click Change.
• To change network connection properties, click Start>Control
Panel>Network and Sharing Center. Select the connection you want to
change and then set the properties of this connection.
For more information on changing network settings, refer to the Windows
10 online help.
32Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 33
Changing Windows Firewall Settings
NOTE
When 16860 series logic analyzers are shipped from the factory, the
Windows Firewall is enabled and set up with the exceptions required by
the logic analyzer.
To give other applications/ports access through the firewall
For example, you may need to change firewall settings in order to:
• Use NetOp to remotely control the logic analyzer.
• Use RealVNC to remotely control the logic analyzer.
To change firewall settings on Windows OS to give other
applications/ports access:
1 From the Windows Start menu, choose Start > Control Panel > System and
Security > Windows Firewall.
2Click the Allow an app or a feature through Windows firewall link.
3 The list of allowed programs and features is displayed. If the required
program or service is listed, check its box to enable it; otherwise, click
Allow another program... to give unlisted programs permission to
penetrate the firewall.
Refer to the application’s documentation for more information on port
numbers or other firewall setup information.
Updating the Logic Analyzer3
Note that there are separate Windows Firewall profiles: Domain for when the computer
has domain membership, and Standard for when the computer has workgroup
membership. If you change the type of membership, any changes you made to the
Windows Firewall settings will have to be made again.
To restore the logic analyzer firewall defaults
1 From the Windows Start menu, choose Start > Control Panel > System
Security.
2Click the Windows Firewall link.
3Click Restore defaults in the left pane of the window.
4 Again click Restore defaults.
5 Uninstall and then reinstall the Logic Analyzer software from
www.keysight.com/find/lpa-sw-download.
The logic analyzer firewall defaults enable the following exceptions:
Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide33
Page 34
3Updating the Logic Analyzer
NOTE
Note that there are separate Windows Firewall profiles: Domain for when the computer
has domain membership, and Standard for when the computer has workgroup
membership. When you restore the logic analyzer firewall defaults, the defaults for both
profiles are restored.
• Keysight Logic Analysis Application program.
• Keysight Logic Analysis Service program.
• File and Printer Sharing.
• Microsoft RPC Endport Mapper TCP port.
• Remote Desktop service.
• Web Server (HTTP) port.
The logic analyzer firewall defaults also enable the following ICMP
(Internet Control Message Protocol) settings:
• Allow incoming echo request.
• Allow outgoing destination unreachable.
• Allow outgoing time exceeded.
34Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 35
Software Installation and Upgrades
Keysight Logic and Protocol Analyzer Software
The Keysight 16860 Series logic analyzers are shipped with software
already installed, including any licensed optional products that you may
have purchased. The licensing process (described below) has been
performed and the (*.lic) license file is in place.
To upgrade/reinstall the logic and protocol analyzer software or to install
this software onto your PC for offline processing:
• Download the Logic and Protocol Analyzer software from
http://www.keysight.com/find/lpa-sw-download and then install the
software using the downloaded file. These systems are not shipped with
an application CD. You need the Keysight Logic and Protocol Analyzer
software version 6.30 or later for these systems.
To be notified when software upgrades are available for downloading from
the web, please sign up for e-mail notifications at:
www.keysight.com/find/emailupdates
Optional Products
If applicable, follow the instructions on the entitlement certificate that
came with your purchase to enable the optional tool.
Updating the Logic Analyzer3
Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide35
Page 36
3Updating the Logic Analyzer
36Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 37
Keysight 16860 Series Portable Logic Analyzers
Installation/Quick Start Guide
4Solving Problems
Running Self-Tests / 38
If the logic analyzer powers itself down / 40
If the Power LEDs do not come on / 40
If you get a registry error and the system won’t boot / 40
If there are network connection problems / 41
If there are touch screen problems / 42
Using the System Recovery Software / 45
Contacting Keysight Service/Support / 49
Page 38
4Solving Problems
Running Self-Tests
1 In the Keysight Logic and Protocol Analyzer application, click Help>Self
Tes t.
2In the Analysis System Self Test dialog, double-click the test you want to
run.
38Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 39
Solving Problems4
Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide39
Page 40
4Solving Problems
If the logic analyzer powers itself down
If the logic analyzer warns you it is powering down before it powers down,
the dialog will describe the problem (such as a fan/overtemp problem). If
the logic analyzer just powers down, it is likely a power supply problem.
If the Power LEDs do not come on
If the lights do not come on and if the system powers up momentarily
when you plug it in, make sure the power button hasn't become jammed or
stuck in the pushed in position.
If you get a registry error and the system won’t boot
When powered off unexpectedly (for example, with a long press of the
power button or a power loss), Windows may not flush all system
information to disk and is vulnerable to problems. When you power back
on you may get a "Registry Error" and the system won't boot. If this
happens you will need to restore the system software using the recovery
process. This process overwrites all the user data and restores the system
back to the original condition in which it was shipped. Refer to the topic
“Using the System Recovery Software" on page 45 to know more about
this process.
It is best to keep regular backups of your data and use proper power down
procedures. If you have data that was not backed up that you need to try
to save:
1 Physically remove the disk drive.
2 Mount it under another Windows 10 system.
3 Move the data to the boot disk until you can restore the operating
system.
40Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 41
If there are network connection problems
See the “Network Troubleshooting Guide” topic in the Keysight Logic Analyzer application’s online help. Briefly:
• The Windows Firewall must be set to allow access to the “Keysight
Logic Analysis” service in order to use remote connections (see “To
restore the logic analyzer firewall defaults" on page 33).
• The Windows Firewall must be set to respond to the following ICMP
(Internet Control Message Protocol) requests in order to use remote
connections:
• Allow incoming echo requests.
• Allow outgoing destination unreachable.
• Allow outgoing time exceeded.
(See “To restore the logic analyzer firewall defaults" on page 33.)
• The link activity light must be on; If these LEDs are not on, the LAN
segment may be dead.
• If connecting two logic analysis systems directly together (that is, no
hub or switch between them), make sure you are using a special
cross-over cable.
If your IP address is in the same subnet as a working computer’s (use
“ipconfig /all | more” to view IP addresses on machine in question), it is
possible that your network administrators have denied access to any
resources if the computer is not in the domain. In this case, you need to
add the computer (logic analysis system) to the domain.
Solving Problems4
Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide41
Page 42
4Solving Problems
If there are touch screen problems
If you observe the following touch screen problems:
• If the Touch Off button on the instrument’s front panel doesn’t work.
• If there are problems with the right-click functionality on the touch
screen.
• If the accuracy of touch locations is off.
You may be able to calibrate the touch screen to fix the problem by
following the procedures described below for 16860 series of logic
analyzers.
1 From the Windows Start menu, choose Start>All Programs>Keysight Logic
Analyzer>Utilities>Touch Screen>Touch Screen Calibration and Settings.
2 In the Microchip TSHARC Control Panel dialog, select the Calibration
tab.
42Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 43
Solving Problems4
3 By default, the number of points is set to 4 and the Inset is set at
20.0%.
4Click Begin Calibration.
5 If you want to enable the right-click functionality on the touch screen,
select the Click Settings tab and then select the Enable Right-Click
checkbox. You can also set the area on the touch screen on which you
want to enable the right-click functionality.
Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide43
Page 44
4Solving Problems
If calibrating the touch screen doesn’t fix the problem, see “Contacting
Keysight Service/Support" on page 49.
44Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 45
Using the System Recovery Software
NOTE
Restoring your system software might be necessary for the following
reasons:
• Hard drive (SSD) failure.
• Virus in the system or unstable system.
• Intentional disk clean - for example if you are passing the system to
another team or returning it to a rental company and you do not want
any data left on it.
The 16860 series of Logic Analyzers are now shipped with the Windows 10
operating system preinstalled. Earlier, Windows 7 was preinstalled on
these logic analyzers.
In addition, 16860 series of Logic Analyzers also have the Keysight Logic
Analyzer software version 6.30 or later preinstalled.
These systems do not have any recovery CD or DVD in their shipment. If
you need to restore the logic analysis system software, you run the
recovery process on the hard drive (SSD) of the system. This recovery
process uses the hidden partition on the hard drive to restore the hard disk
drive back to its original state in which it was shipped. When you run this
process, the recovered hard disk drive contains:
• the Windows 10 operating system
• the version of the Logic and Protocol Analyzer software which was
installed when the system was originally shipped and not the latest or
upgraded version of the software that might be available at the time of
system recovery.
• the license files of any licensed optional products that you had
purchased with the Logic Analyzer and that were installed when the
system was originally shipped.
Solving Problems4
When you run the recovery process, the software licensing Host ID of your logic analyzer
may change. If this happens, the restored license files will not work with the new host ID
and you will need rehosted license files. To get these files, you can contact your nearest
Keysight sales/service office. To locate a sales or service office, go to
www.keysight.com/find/contactus.
Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide45
Page 46
4Solving Problems
CAUTION
Running the recovery process reformats your logic analyzer’s hard disk drive to the state
in which it was originally shipped to you. All user data files and programs are overwritten
when the recovery process runs. Therefore, save your data to a CD or to another machine
before you start recovering the system software.
To run the recovery process
1 Shut down and then restart your Logic Analyzer.
While booting up, the Logic Analyzer displays the Windows Boot
Manager screen. This screen provides you the following options:
• Microsoft Windows 7 or 10 (as per the installed operating system) -
This is the default selected option. This option starts the Logic
Analyzer with the installed Windows operating system. This is the
normal startup of the system.
• Keysight Recovery system - You should select this option when you
want to restore/repair your Logic Analyzer software by running the
Keysight recovery process.
2 After a few seconds, Logic Analyzer automatically starts with the option
selected in the Windows Boot Manager. Select the Keysight Recovery
46Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 47
Solving Problems4
system option in the Windows Boot Manager screen using the arrow
keys and then press Enter.
The recovery process starts preparing the system for recovery and
displays the following screen with options to choose. You can enter:
• 1 to run Check Disk on the Logic Analyzer’s hard disk drive. If the
recovery process encounters any problems while running Check
Disk, it reports these problems else it returns to the Keysight
Recovery system prompt on completion of Check Disk.
• 2 to restore your Logic Analyzer software back to its original state in
which it was shipped.
• 3 to view a document that provides information on the recovery
process.
• 4 to repair the Logic Analysis system hard disk drive.
• 5 to exit the recovery process and restart the Logic Analyzer in the
normal mode.
3 Select the second option in the above screen by entering 2 in the text
box and clicking OK.
4 A warning message is displayed stating that the recovery process
overwrites the data on C: drive. If you want to save your data to a CD or
to another machine before you proceed further, then exit the recovery
process and save your data. Else, click OK to proceed with the recovery
process.
Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide47
Page 48
4Solving Problems
The recovery process starts.
If the recovery process is able to restore the system software
successfully, then the following screen is displayed.
5Click OK to proceed. You Logic Analysis system now reboots to
Windows 10.
6 The Windows Boot Manager screen is displayed with Microsoft Windows
10 as the default selection. The system automatically starts with this
48Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 49
Solving Problems4
NOTE
default selection. Alternatively, you can press Enter to proceed with
the default selection.
There are situations when you are not able to run the recovery process, (for instance,
when the hard disk drive of your system fails) or when running the recovery process does
not recover your system software. In such situations, you can send your Logic Analysis
system for hard disk repair/replacement to Keysight. Alternatively, you can contact your
nearest Keysight sales/service office. To locate a sales or service office, go to
www.keysight.com/find/contactus.
Contacting Keysight Service/Support
To locate a sales or service office near you, go to:
http://www.keysight.com/find/contactus
Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide49
Page 50
4Solving Problems
50Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 51
Keysight 16860 Series Portable Logic Analyzers
Installation/Quick Start Guide
5Safety Notices
Warnings / 52
To clean the instrument / 54
Safety Symbols / 55
This apparatus has been designed and tested in accordance with IEC
Publication 1010, Safety Requirements for Measuring Apparatus, and has
been supplied in a safe condition. This is a Safety Class I instrument
(provided with terminal for protective earthing). Before applying power,
verify that the correct safety precautions are taken (see the following
warnings). In addition, note the external markings on the instrument that
are described under "Safety Symbols."
Page 52
5Safety Notices
Warnings
• Before turning on the instrument, you must connect the protective
earth terminal of the instrument to the protective conductor of the
(mains) power cord. The mains plug shall only be inserted in a socket
outlet provided with a protective earth contact. You must not negate
the protective action by using an extension cord (power cable) without
a protective conductor (grounding). Grounding one conductor of a
two-conductor outlet is not sufficient protection.
• Only fuses with the required rated current, voltage, and specified type
(normal blow, time delay, etc.) should be used. Do not use repaired
fuses or short-circuited fuse holders. To do so could cause a shock or
fire hazard.
• Do not use this instrument with a power cord other than the one
provided by Keysight. Other power cords may not have appropriate
ratings.
• If you energize this instrument by an auto transformer (for voltage
reduction or mains isolation), the common terminal must be connected
to the earth terminal of the power source.
• Whenever it is likely that the ground protection is impaired, you must
make the instrument inoperative and secure it against any unintended
operation.
• Service instructions are for trained service personnel. To avoid
dangerous electric shock, do not perform any service unless qualified to
do so. Do not attempt internal service or adjustment unless another
person, capable of rendering first aid and resuscitation, is present.
• Do not install substitute parts or perform any unauthorized
modification to the instrument.
• Capacitors inside the instrument may retain a charge even if the
instrument is disconnected from its source of supply.
• Do not operate the instrument in the presence of flammable gasses or
fumes. Operation of any electrical instrument in such an environment
constitutes a definite safety hazard.
• Do not use the instrument in a manner not specified by the
manufacturer.
• To optimize your comfort and productivity, it is important that you set
up your work area correctly and use your equipment properly. Refer to
the Working in Comfort topic in the online help or set-up and use
recommendations
• Position equipment so that it is not difficult to disconnect the power
cord.
52Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 53
Safety Notices5
• If incorporated into a system, the safety of the system is the
responsibility of the system installer. If the equipment is used in a
manner not specified by the manufacturer, the protection provided by
the equipment may be impaired. See "For Your Safety" booklet
publication number 9320-6797. Use only the power cord supplied with
the instrument.
Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide53
Page 54
5Safety Notices
WARNING
To clean the instrument
To prevent an electrical shock, disconnect the logic analyzer from mains before
cleaning. Use a dry cloth or one slightly dampened with water to clean the external
case parts. Do not attempt to clean internally.
If the instrument requires cleaning:
1 Remove power from the instrument.
2 Clean the external surfaces of the instrument with a soft cloth
dampened with water.
3 Make sure that the instrument is completely dry before reconnecting it
to a power source.
54Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
Page 55
Safety Symbols
!
Safety Notices5
Instruction manual symbol: the product is marked with this symbol when it
is necessary for you to refer to the instruction manual in order to protect
against damage to the product.
Indicates hazardous voltages and potential for electrical shock.
Earth terminal symbol: Used to indicate a circuit common connected to
grounded chassis.
The C-tick mark is a registered trademark of the Spectrum Management
Agency of Australia. This signifies compliance with the Australia EMC
Framework regulations under the terms of the Radio Communication Act
of 1992.
CE compliance marking to the EU Safety and EMC Directives.
ISM GRP-1A classification according to the international EMC standard.
ICES/NMB-001 compliance marking to the Canadian EMC standard.
KC certification mark to demonstrate compliance with the South Korean
EMC requirements.
South Korean Class A EMC declaration:
This equipment is Class A suitable for professional use and is for use in
electromagnetic environments outside of the home.
CSA is the Canadian certification mark to demonstrate compliance with
the Safety requirements.
Notice for the European Community: This product complies with the WEEE
Directive (2002/96/EC) marking requirements. The affixed label indicates
that you must not discard this electrical/electronic product in domestic
Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide55
Page 56
5Safety Notices
household waste. Product Category: With reference to the equipment
types in the WEEE Directive Annex I, this product is classed as a
“Monitoring and Control instrumentation” product.
Do not dispose in domestic household waste. To return unwanted
products, contact your local Keysight office or get more information from
www.keysight.com/environment/product.
56Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide