Keysight 1168B User Manual

Keysight 1168/9B-Series Differential and Single-Ended Probes
User’s Guide
Notices
CAUTION
WARNING
© Keysight Technologies, 2005 - 2019
No part of this manual may be reproduced in any form or by any means (including elec­tronic storage and retrieval or translation into a foreign language) without prior agree­ment and written consent from Keysight Technologies, Inc. as governed by United States and international copyright laws.
Manual Part Number
01169-97027
Edition
Twelfth Edition, July 2019
Published by: Keysight Technologies, Inc.
1900 Garden of the Gods Road Colorado Springs, CO 80907 USA
Warranty
The material contained in this docu­ment is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maxi­mum extent permitted by applicable law, Keysight disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par­ticular purpose. Keysight shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or perfor­mance of this document or of any infor­mation contained herein. Should Keysight and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the war­ranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor­dance with the terms of such license.
U.S. Government Rights
The Software is "commercial computer soft­ware," as defined by Federal Acquisition Regulation ("FAR") 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of Defense FAR Supplement ("DFARS")
227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is cus­tomarily provided to the public. Accordingly, Keysight provides the Software to U.S. gov­ernment customers under its standard com­mercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at http://www.key-
sight.com/find/sweula. The license set forth
in the EULA represents the exclusive author­ity by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial computer software documentation that is not customarily pro­vided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose com­mercial computer software or commercial computer software documentation. No addi­tional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obliga­tion to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR 12.211 and 27.404.2 and DFARS
227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR 227.7103-5 (c), as applicable in any technical data. 52.227-14
(June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any tech­nical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAU- TION notice until the indicated con­ditions are fully understood and met.
A WARNING notice denotes a haz­ard. It calls attention to an operat­ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.

Contents

1 Getting Started
2Using Probe Heads
Introduction 8 Probe Handling 12 Using Offset With InfiniiMax Active Probes 16 Slew Rate Requirements for Different Technologies 19 Available Accessories 20 Safety Information 31 Troubleshooting 33
Recommended Configurations 38
1. MX0100A InfiniiMax Micro Probe Head 41
2. N5381B Differential Solder-In 50
3. N2839A Differential Browser 52
4. N5382A Differential Browser 57
5. N5380B SMA 60
6. N5425B with N5426A ZIF Tip 63
7. N5425B with N2884A Fine Wire ZIF Tip 64
8. N2851A QuickTip Probe Head 67
9. N5425B with N5451A Long-Wire ZIF Tip (7 mm) 71
10. N5425B with N5451A Long-Wire ZIF Tip (11 mm) 73
11. E2677B Differential Solder-In Probe Head 75
12. E2678B Differential Socketed Probe Head 77
13. E2675B Differential Browser 79
14. E2679B Single-Ended Solder-in Probe Head 81
15. E2676B Single-Ended Browser 82
16. E2678B Differential Socketed & Damped Wire Accessory 84 Soldering a ZIF Tip to a DUT 86 Using N2884A Fine-Wire ZIF tips 90 N2887A/8A Soft Touch Probe Heads 98
3 Maintaining Probe Heads
MX0100A Micro Probe Heads 106 N5381B/N5382A Probe Heads 109
3
N5451A Long-Wired ZIF Tips 114 E2677B/9A Solder-In Probe Heads 118
4 Calibrating Probes
DC Gain and Offset Calibration 124 Calibration for Solder-In and Socketed Probe Heads 125 Calibration for Hand-Held Browser Probe Heads 131 Calibration for N5380A/B SMA Probe Heads 133 N2887A & N2888A Calibration and Deskew Procedure 134
5 Characteristics and Specifications
General 138 InfiniiMax II Series with N5380B SMA Probe Head 141 Environmental 143 Probe Amplifier Dimensions 144 Probe Heads Dimensions 145
6Performance Plots
MX0100A Micro Probe Head (Full BW) 152 N2839A Differential Browser Head (Full BW) 157 N5380B SMA Probe Head (Full BW) 162 N5380B SMA Probe Head with the 1134B InfiniiMax Probe 165 N5381B & N5382A Differential Probe Heads (Full BW) 166 N5381B Differential Solder-in Probe Head with 2x Longer Wires 171 N5425B ZIF Probe Head (Full BW) 172 N5425B ZIF Probe Head with N5451A Long-Wired ZIF Tip 176 N5426A ZIF Probe Tip Impedance 188 N2851A QuickTip Head with N2849A QuickTip 189 E2675B Differential Browser 190 E2676B Single-Ended Browser 192 E2677B Differential Solder-in Probe Head (High BW) 195 E2678B Differential Socketed Probe Head (High BW) 197 E2678B Differential Socketed Probe Head w/ Damped Wire 199 E2679B Single-Ended Solder-In Probe Head (High BW) 201 N2887A/N2888A Soft Touch Probe Heads 203
7 Performance Verification
To Test Bandwidth 206 To Test Input Resistance 215 Performance Test Record 219
4 1168/9B Probes User’s Guide
8SPICE Models
MX0100A Micro Probe Head 222 N5381B and N5382A Heads 224 N2839A Head 227 N5425B ZIF Head with N5426A ZIF Tip Attached 230 N5426A ZIF Tip 233 N5425B ZIF Head with N5451A Long-Wire ZIF Tip 236 N2851A QuickTip Head with N2849A QuickTip Tip 244 N2887A/N2888A Heads 245 InfiniiMax I Heads 247
9Replacement Parts
N5381B and N5382A Probe Heads 258 N2839A Browser Head 258 E2679B Single-Ended Solder-in Probe Head 258 E2678B Differential Socketed Probe Head 259 E2677B Differential Solder-In Probe Head 260 E2675B Differential Browser Probe Head 260 Other Accessories 261
Index
1168/9B Probes User’s Guide 5
6 1168/9B Probes User’s Guide
Keysight InfiniiMax II Series Probes
CAUTION
WARNING
User’s Guide

1 Getting Started

Introduction 8 Probe Handling 12 Using Offset With InfiniiMax Active Probes 16 Slew Rate Requirements for Different Technologies 19 Available Accessories 20 E2669B Differential Connectivity Kit and Accessories 20 N2833A InfiniiMax II Differential Connectivity Kit and Accessories 21 N5450B Extreme Temp Cable Extension Kit 22 N2880A InfiniiMax In-Line Attenuator Kit 24 N2881A InfiniiMax DC Blocking Capacitors 28 MX0102A Soldering Toolkit 29 Safety Information 31 Troubleshooting 33
With the 1168B (10 GHz) and 1169B (12 GHz) probes, you can probe differential and single-ended signals. The probes provide a large common mode range for measuring differential signals and a large offset range for measuring single-ended signals. These probes are used at extreme frequencies where off-board lead resistors cause undesirable response variation. As a result, the 1168B and 1169B probes were designed using resistor-at-the-tip technology where resistors are located onto the very edge of the probe tip board. The wires or probe tips in front of the resistors are long enough to allow easy connection but are short enough that any resonances caused by them are out of band and don't impact the input impedance.
Before using the probes, refer to “Probe Handling" on page 12.
Before using the probe, refer to “Safety Information" on page 31.
7
1 Getting Started

Introduction

Before you can use the probe, you must connect one of the available probe heads to an 1168/9B probe amplifier.
Figure 1 Probe Amplifier with Attached Head
Probe Heads
Figure 2 on page 9 shows the available probe heads and accessories. Six different
InfiniiMax II probe heads can be used including a Zero Insertion Force (ZIF) probe head that uses a ZIF tip that can be installed at many locations on your DUT. The ZIF tip’s small size is critical in probing tight locations and the ZIF feature allows connection without compressing the delicate wires which cannot support this compression. You can also use the probe amplifiers with the InfiniiMax I probe heads (with some limitations).
The differential probe heads offer easy measurement of differential signals and greatly improve the measurement of single-ended signals.
Each available probe head is documented in Chapter 2, “Using Probe Heads".
Compatible Oscilloscopes
Table 1 on page 10 lists the oscilloscopes that are compatible with the 1168/9B
probes. Is Your Oscilloscope Software Up-to-Date? Keysight periodically releases Oscilloscope software updates to support your probe, fix known defects, and incorporate product enhancements. To download the latest firmware, go to www.keysight.com and search for your oscilloscope’s topic. Click on the “Drivers, Firmware & Software” tab.
8 1168/9B Probes User’s Guide
Getting Started 1
NOTE
Figure 2 Available Probe Heads and Accessories
N2849A QuickTips are also compatible with the N2848A InfiniiMode probe heads which are designed for N2830/1/2A and N7000/1/2/3A InfiniiMax III+ and N2800/1/2/3A InfiniiMax III probes.
These probes can also be used with other test instruments that have 50 ohm input such as a spectrum analyzer or 86100D DCA. For more information, check out the Keysight application note, 5989-1869EN.
1168/9B Probes User’s Guide 9
1 Getting Started
NOTE
Table 1 Compatible Infiniium Oscilloscopes
Oscilloscope Model
S Series all
V, 90000 X-, Q-, and Z-Series
90000A Series all
86100C/D Series
80000B Series all
a N5442A adapter required.
b N1022A/B adapter required.
b
a
all
all
The 1168/69A InfiniiMax probes are not compatible with Keysight's InfiniiVision Series oscilloscopes.
Cleaning the probe
If the probe requires cleaning, disconnect it from the oscilloscope and clean it with a soft cloth dampened with a mild soap and water solution. Make sure the probe is completely dry before reconnecting it to the oscilloscope.
Channel Identification Rings
When multiple probes are connected to the oscilloscope, use the channel identification rings to associate the channel inputs with each probe. Place one colored ring near the probe’s channel connector and place an identical color ring near the probe head.
Inspecting the Probe
Inspect the shipping container for damage.
Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the probe has been checked mechanically and electrically.
Check the accessories.
If the contents are incomplete or damaged, notify your Keysight Technologies
Sales Office.
Inspect the probe. If there is mechanical damage or defect, or if the probe does not operate properly or pass calibration tests, notify your Keysight Technologies Sales Office.
10 1168/9B Probes User’s Guide
Getting Started 1
If the shipping container is damaged, or the cushioning materials show signs of stress, notify the carrier as well as your Keysight Technologies Sales Office. Keep the shipping materials for the carrier’s inspection. The Keysight Technologies office will arrange for repair or replacement at Keysight Technologies’ option without waiting for claim settlement.
Figure 3 shows the accessories that are shipped with the 1168/9B probe
amplifiers. The probe amplifiers do not come with a probe head unless selected at the time of order. Any head shown in Figure 2 on page 9 can be ordered at any time for the probes.
Figure 3 Accessories Supplied With the Probe Amplifier
1168/9B Probes User’s Guide 11
1 Getting Started
CAUTION

Probe Handling

This probe has been designed to withstand a moderate amount of physical and electrical stress. However, with an active probe, the technologies necessary to achieve high performance do not allow the probe to be unbreakable. Treat the probe with care. It can be damaged if excessive force is applied to the probe tip. This damage is considered to be abuse and will void the warranty when verified by Keysight Technologies service professionals.
Exercise care to prevent the probe end from receiving mechanical shock.
Store the probe in a shock-resistant case such as the foam-lined shipping case
which came with the probe.
Connecting and Disconnecting Probe Heads
When disconnecting a probe head from an amplifier, pull the probe head connectors straight out of the sockets as shown in Figure 4. When connecting a probe head to an amplifier, push straight in. Always grasp the indentations located on the sides of the amplifier as shown in Figure 4. There are also indentations on many of the probe head sockets so you have a convenient place to grasp there as well.
Figure 4 Properly Pulling the Probe Head Straight Out
Avoid damaging the connection pins. Never bend the probe head in order to “pop” it loose from the amplifier. Do not wiggle the probe head up and down or twist it to remove the connectors from the sockets.
12 1168/9B Probes User’s Guide
Figure 5 Improperly Disconnecting a Probe Head From an Amplifier
CAUTION
CAUTION
Handling the Probe Cable
Getting Started 1
Avoid degrading the probe’s performance. Do not twist, kink, or tightly bend the probe’s cable.
When the probe is attached to an oscilloscope, avoid letting object hit the probe cable where the cable exits the probe amplifier and bend it well beyond its limit.
When storing the probe, coil the cable in a large loops and avoid twisting the cable. Coil the cable in a similar manner to how garden hoses or extension cords are typically coiled. You can start by wrapping the cable around your thumb as shown in Figure 6. Then continue to circle your thumb, but provide a slight twist with each rotation. This allows the cable rotations to lie flat against each other and will eliminate the net twisting of the cable in the end.
Figure 6 Recommended Coil for Storage
1168/9B Probes User’s Guide 13
1 Getting Started
CAUTION
NOTE
Handling the Probe Amplifier
Connecting the Probe to an Oscilloscope
Make the coil’s radius fairly large so it does not induce kinking or bending.
The probe amplifier contains a delicate circuit board. Treat it carefully and take standard precautions (for example, not dropping it repeatedly or from large heights, not getting it wet, not smashing it with heavy objects, etc.). These probes are sensitive ESD devices so standard precautions need to be used to not ruin the probe from the build-up of static charges.
The probes are only meant to be plugged into gold plated BNCs (like those on Infiniium oscilloscopes). To connect the probe to the oscilloscope, do the following steps:
1 As shown in Figure 7, with the lever in the relaxed position push the probe
onto the BNC. The lever moves towards the R (release) and returns to the symbol.
2 Move the lever towards the symbol until snug.
How far the locking mechanism can be pushed to the right varies and will not be the same for every user. Therefore, do not try to force it further to the right because you believe it is unlocked. Instead, gently push it until it is snug.
To disconnect the probe, move and hold the lever at R (release) and pull the probe from the BNC.
Figure 7 Properly Connecting a Probe to the Oscilloscope
Securing Probe Heads and Amplifiers to Your DUTs
When soldering a probe head to a circuit, first provide strain relief by using low temperature hot glue (use as little as possible) or non-conductive double-sided tape. Do not use super glue and do not get the low temperature hot glue on the actual probe head tip as this can damage the precision components of your probing system (only use the low temperature hot glue on the probe head cables). The provided velcro pads can be used to secure your probe amplifier casing to the board.
Once strain relief has been provided, solder the probe tip to the circuit board and then plug the probe head into the probe amplifier.
14 1168/9B Probes User’s Guide
Figure 8 Correct Securing Methods
Getting Started 1
Figure 9 Incorrect Securing Method Because Glue is Placed on the Probe Head Tip
The velcro dots can be used to secure the probe amplifier to a circuit board removing the weight of the probe from the circuit connection. Attach a Velcro dots to both the probe amplifier and the circuit board as shown in Figure 10.
Figure 10 Using the Velcro Dots
1168/9B Probes User’s Guide 15
1 Getting Started

Using Offset With InfiniiMax Active Probes

It is important to understand how the 1168/9B InfiniiMax probes behave with respect to offset when different probe head / signal combinations are used.
The purpose of offset in active probes or oscilloscope front ends is to allow the subtraction of most or all of the dc component of the input signal so the signal can better utilize the dynamic range of the input. When using an InfiniiMax probe with an Infiniium oscilloscope, you can select the case (see the three cases described below) that applies for your measurement by selecting the Probes button under the channel setup menu. This allows you to select which type of probe head is being used and, if it is a differential probe head, allows you to select whether you are probing a differential or single-ended signal. With these inputs, the oscilloscope will use the proper type of offset for your measurement case. The specifics for each case are discussed below.
When adjusting the offset for a particular probe head, make sure to have a triggered signal.
Case 1. A single-ended probe head probing a single-ended signal
For this case, the offset control on the oscilloscope controls the probe offset and the channel offset is set to zero. This allows the offset voltage to be subtracted from the input signal before the signal gets to the differential amplifier. Since this subtraction is done before any active circuits, the offset range is large (±16V). Note that the minus probe tip is not present when using a single-ended probe head which means nothing is plugged into the "–" input of the probe amp. This is normal and causes no problems.
Case 2. A differential probe head probing a single-ended signal
For this case, the offset control on the oscilloscope controls the probe offset and the channel offset is set to zero. This allows the offset voltage to be subtracted from the input signal before the signal gets to the differential amplifier. Since this subtraction is done before any active circuits, the offset range is large (±16V). A differential probe can make higher bandwidth and more accurate measurements on single-ended signals than a single-ended probe and this method of applying offset to only the plus side of a differential probe means there is no sacrificing of offset range.
When Probe is selected in the Probe Offset dialog box as shown in Figure 11 on page 17, the InfiniiMax probe provides a very large offset range (up to ±16V) for probing single-ended signals and a large common-mode range for probing differential signals. For information on properly using probe offset to ensure that you can get the maximum performance and dynamic range from the InfiniiMax probe, refer to Keysight application note 5988-9264EN.
16 1168/9B Probes User’s Guide
Getting Started 1
Figure 11 Probe Offset Dialog Box (Probe Setting)
Case 3. A differential probe head probing a differential signal
For this case, the offset control on the oscilloscope controls the oscilloscope channel offset. The probe offset is not used and set to zero. Since the plus and minus sides of differential signals have the same dc component, it will be subtracted out and the output of the probe will by definition be centered around ground.
The channel offset allows the waveform seen on screen to be moved as desired. The allowable dc component in the plus and minus signals is determined by the common mode range of the probe.
Figure 12 shows Normal selected in the Probe Offset dialog box. When probing
differential signals Normal allows you to apply probe offset using the oscilloscope’s front-panel vertical offset controls.
1168/9B Probes User’s Guide 17
1 Getting Started
Figure 12 Probe Offset Dialog Box (Normal Setting)
18 1168/9B Probes User’s Guide

Slew Rate Requirements for Different Technologies

The following table shows the slew rates for several different technologies. The maximum allowed input slew rate is 25 V/ns for single-ended signals and 40 V/ns for differential signals. Table 2 shows that the maximum required slew rate for the different technologies is much less that of the probe.
Table 2 Slew Rate Requirements
Getting Started 1
Max Single-Ended
Differential
Name of Technology
PCI Express (3GIO) YES 9.6 19.2 50 1.6
RapidIO Serial 3.125Gb YES 8.0 16.0 60 1.6
10GbE XAUI (4x3.125Gb) YES 8.0 16.0 60 1.6
1394b YES 8.0 16.0 60 1.6
Fibre Channel 2125 YES 8.0 16.0 75 1
Gigabit Ethernet 1000Base-CX YES 7.8 15.5 85 2.2
RapidIO 8/16 2Gb YES 7.2 14.4 50 1.2
Infiniband 2.5Gb YES 4.8 9.6 100 1.6
HyperTransport 1.6Gb YES 4.0 8.0 113 1.5
SATA (1.5Gb) YES 1.3 2.7 134 0.6
USB 2.0 YES 0.9 1.8 375 1.1
DDR 200/266/333 NO 7.2 n/a 300 3.6
PCI NO 4.3 n/a 500 3.6
Signal
Slew Rate (V/ns)
a
Max Differential Slew Rate (V/ns)
Driver Min Edge
b
Rate (20%-80% ps)
Max Transmitter Level (Diff V)
AGP-8X NO 3.1 n/a 137 0.7
a The probe specification is 25 V/ns
b The probe specification is 40 V/ns
1168/9B Probes User’s Guide 19
1 Getting Started

Available Accessories

This section lists the kits and accessories that are available in addition to the individual probe heads described in Chapter 2, “Using Probe Heads".
E2669B Differential Connectivity Kit and Accessories
The optional E2669B differential connectivity kit provides multiple quantities of the three InfiniiMax I probe heads as shown in Figure 13. These probe heads allow full bandwidth probing of differential and single-ended signals. The kit can be ordered at the same time as 1168/9B probe amplifiers.
Figure 13 E2669B Differential Connectivity Kit (not to scale)
Table 3 Supplied Accessories (Sheet 1 of 2)
Qty
Description
E2678B Differential Socketed Head 2
E2677B Differential Solder-In Head 4
E2675B Differential Browser 1
160W damped wire accessory 12
82W resistor for full bandwidth 96
Socket for 25 mil (25/1000 inch) square pins, female on both ends
25 mil female socket with 20 mil round male pin on other end
Heat shrink socket accessory 8
Header adapter, 91W 4
Supplied
8
8
Used With
E2678B E2677B E2675B
Part Number
01130-21303
01130-81506
01131-85201
01131-85202
01130-41101
01130-63201
a
82W resistor template 1
20 1168/9B Probes User’s Guide
01131-94309
Table 3 Supplied Accessories (Sheet 2 of 2)
Getting Started 1
Qty
Description
91W resistor for full bandwidth 80
150W resistor for medium bandwidth 40
91W resistor template 1
150W resistor template 1
Resistive tip (blue), 91W 20
Ergonomic handle 1
Supplied
a Not orderable.
N2833A InfiniiMax II Differential Connectivity Kit and Accessories
The optional N2833A differential connectivity kit provides multiple quantities of the four InfiniiMax II probe heads as shown in Figure 14. These probe heads allow full bandwidth probing of differential and single-ended signals. You can order this kit either at the same time as 1168/9B probe amplifiers or separately later.
Used With
E2678B E2677B E2675B
Part Number
0700-2353
0700-2350
01131-94311
01131-94308
01131-62107
01131-43201
a
Figure 14 Probe Heads Included in the N2833A Differential Connectivity Kit (not to scale)
1168/9B Probes User’s Guide 21
1 Getting Started
Table 4 Supplied Accessories
Qty
Description
N5381B InfiniiMax II 12 GHz Differential Solder-In Probe Head 2 N5381B
0.007 inch tin-plated nickel wire
0.005 inch tin-plated nickel wire Trim Gauge
N5425B InfiniiMax II 12 GHz Differential ZIF Solder-in Probe Head
N5426A InfiniiMax 12 GHz ZIF Tip Kit 2 kits (10 tips in
N2851A InfiniiMax II QuickTip Probe Head 2 N2851A
N2849A InfiniiMax QuickTip Tips Kit 2 kits (4 tips in
N2839A InfiniiMax II Browser Probe Head 1 N2839A
Spring-loaded tips Tweezer for replacing tips Protective end cap
* - Indicates the part number of an accessory not orderable separately as an individual product.
Supplied
1 1 1
2 N5425B
each kit)
each kit)
20 1 1
Part Number
01169-81301 * 01169-21306 *
-
N5426A
N2849A
-
-
-
N5450B Extreme Temp Cable Extension Kit
The extreme temperature cable extension kit is an accessory that allows an oscilloscope probe to be used to monitor a device in a temperature chamber. Keysight’s Infiniimax probe amplifiers have a specified operating temperature range from 5 can be operated over a much larger range of temperatures. Use the extension cables to physically separate the amplifier from the probe head which allows you to operate the probe head inside a temperature chamber while the probe amplifier remains outside the chamber. To ensure a high-quality measurement, the N5450B cable set have been phase-matched at the factory. A coupling tag is included with the cables to ensure the cables stay as a matched pair. To install the coupling tag, slip the small end of each cable through the holes in the tag. The tag can be positioned anywhere along the length of the cable and can withstand the temperature ranges specified.
o
C to 40o C, but the probe heads
22 1168/9B Probes User’s Guide
Table 5 Probing Temperature Ranges
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
NOTE
NOTE
Getting Started 1
Probe Head
Configuration
N5381B –40° to +85° > 250
E2677B –25° to +80° > 1000
E2678B –25° to +80° > 1000
N5425B + N5426A -40° to +85° > 500
N5451A –25° to +80° > 1000
MX0100A -55° C dwell, 1000 hours minimum
a Refers to the probe head or tip that is attached to the cable extension kit.
a
Operating Temperature Range (oC)
+150° C dwell, 1000 hours minimum
-55° C to 150° C cycles, 1000 cycles minimum (as per JEDEC JESD22-A104 revision E)
Expected Lifetime of
the Probe Head (cycles)
> 1000
Avoid rapid changes in temperature that can lead to moisture accumulating in the form of condensation on the probe components, as well as the DUT. If this occurs, wait until the moisture has evaporated before making any measurements.
Additional care must be taken when handling probe heads used during extreme temperature cycling because this process makes the probe heads less robust.
Secure the ends of the extension cable near the probe head in the temperature chamber such that the probe head legs are not tugged or moved around significantly.
Prevent abrasion and tears in the cable’s jacket, do not rest the extension cables on any metal objects or objects with sharp edges.
Do not kink the cables. The cables are designed to be flexible, but are not designed to be bent sharply.
Keep your extreme temperature testing probes separate from the probes they use under milder conditions. This is because cycling probe heads through extreme temperature ranges has a marked affect on their lifetimes as listed in Table 5. Only the lifetime of the probe head is affected by temperature cycling. The extension cables and probe amplifier should not need to be replaced with extended temperature cycling.
Discoloration or texture changes are possible with the extension cables. These changes do not, however, affect the performance or the quality of a measurement.
1168/9B Probes User’s Guide 23
1 Getting Started
N2880A InfiniiMax In-Line Attenuator Kit
The in-line attenuators are an accessory for probes. The maximum input range of the 1168B/9B probes are 3.3 V
. If you need to measure larger signals, the
p-p
probe’s design allows you to add the N2880A InfiniiMax in-line attenuators between the probe head and probe amplifier to increase the maximum input range (as listed in Table 6 on page 25). Additionally, these attenuators enable you to increase the offset range of the probe as specified in Table 6 on page 25. When using the N2880A In-Line Attenuators, the bandwidth and rise time of your probing system is not affected. There is, however, a trade-off in noise (refer to
Table 6) and in the accuracy of DC offset relative to the input.
The N2880A provides a pair of 6 dB, 12 dB, and 20 dB attenuators. The attenuators come as matched pairs and should only be used with each other. Each attenuator has a serial number. The pair of matching attenuators in each set will have the same four digit numeric prefix and will differ by the last letter (one attenuator in the matched pair will be labeled A and the other will be labeled B).
Compatibility with probe heads and amplifiers
All InfiniiMax I and II probe heads and amplifiers are compatible with the N2880A In-line attenuators. However, the following two limitations should be considered when planning to use N2880A attenuators.
The maximum input voltage of the InfiniiMax I probe heads is ±30 Vdc and so they should not be used to measure signals that exceed this range. This places a practical limit of 20 dB on the attenuators used with the InfiniiMax I probing system. Larger attenuation ratios will only degrade the noise performance and gain of the system.
Due to the N5380B dual-SMA probe head’s maximum input voltage specification of 2.28 V large enough to require an added attenuator.
, the N5380B is not suitable for measuring signals
RMS
24 1168/9B Probes User’s Guide
.
NOTE
Red = dB(Vout/Vin) + 10.8 dB of probe
Blue = dB (Vout/Vin) + 12 dB attenuator + 10.8 dB of probe
Black = dB(Vout/Vin) + 6dB attenuator + 10.8 dB
Green = dB(Vout/Vin) + 20 dB attenuator + 10.8 dB of probe
Table 6 N2880A With 1168B/9B Probe Amplifiers
Getting Started 1
Added Attenuator
Maximum Input Range
(mains isolated circuits only)
Offset Range
Typical Noise Referred to
Maximum Allowed Input Slew Rate
a
(se = single-ended) (diff = differential)
Nominal DC Attenuation of Probe System
None 3.3 Vp-p ±16V 2.2 mV RMS se: 25 V/ns, diff: 40 V/ns 3.45:1
6 dB (2:1) 6.6 Vp-p ±30 V
12 dB (4:1) 13.2 Vp-p ±30 V
20 dB (10:1) 33.3 Vp-p ±30 V
a These slew rate do not apply when the N5380B SMA probe head is used with the InfiniiMax amplifiers.
b The actual range of DC voltage for N2880A is > ±30 V, but the usable range of DC voltage at the probe input is limited to ±30 Vdc.
b
b
b
6.3 mV RMS se: 50 V/ns, diff: 80 V/ns 6.9:1
13.2 mV RMS se: 100 V/ns, diff: 160 V/ns 13.8:1
33.4 mV RMS se: 250 V/ns, diff: 400 V/ns 34.5:1
The values shown above do not apply to the N5380B dual-SMA probe head. Due to the maximum input voltage specification of 2.28 VRMS (mains isolated circuits only) for the N5380B, it is not suitable for measuring signals large enough to require an added attenuator.
Frequency Response Plots
Below are the frequency response plots for four setups: the probe without any attenuators, the probe with the 6 dB attenuators, the probe with the 12 dB attenuators, and the probe with the 20 dB attenuators.
Figure 15 Frequency Response
1168/9B Probes User’s Guide 25
1 Getting Started
Extended Offset Range with N2880A Attenuators
The use of N2880A attenuators allows an increased offset range (±30 Vdc) of the probe as listed in Table 6 on page 25.
However, if you are using the12 dB (4:1) or 20 dB (10:1) N2880A attenuators with the N5381A/B probe head and InfiniiMax II probe amplifier, an extended offset range (±60 Vdc) is allowed to enable you to measure higher input voltages.
When you connect the appropriate attenuators, probe head, and probe amplifier to the oscilloscope, the extended offset range is available using the Offset field of the External Scaling dialog box in the Infiniium software GUI. You access this dialog
box by clicking Setup > Probe Configuration.... and then enabling External Scaling.
For this extended offset range, the maximum allowed AC voltage is dependent on the DC voltage on the signal. The graph in Figure 16 depicts the DC voltage and the corresponding allowed maximum AC voltage for this extended range. It is recommended that you use the maximum AC voltage values depicted in this graph to set the extended offset value in the Infiniium GUI.
Figure 16 Extended offset range - Maximum AC voltage recommendations corresponding to
the DC voltage
You can also calculate the maximum allowed AC voltage using the following equations.
26 1168/9B Probes User’s Guide
Getting Started 1
Maximum V
Maximum V
ACRMS
AC_PEAK
= 30 * (60 - VDC) / 60
= 42.4 * (60 - VDC) / 60
Calibrating and Configuring Attenuators on an Infiniium Series Oscilloscope
The software in the Infiniium oscilloscopes will detect a probe when it is connected and by default will assume that no additional attenuators are installed. If you want to scale readings and settings on the oscilloscope so they are correct with the attenuators installed, refer to the procedures below for your specific oscilloscope series.
Calibrating Attenuators on an Infiniium Series Oscilloscope
You cannot calibrate your InfiniiMax probes with the attenuators attached. Calibrate the InfiniiMax probes as you normally would (with no attenuators), configure the attenuators as discussed in the next section, and begin probing.
Configuring Attenuators on an Infiniium Series Oscilloscope
First, plug your InfiniiMax probe amplifier / probe head into one of the oscilloscope channels with the attenuators connected. Then enter the Probe Setup dialog box (can be reached via Setup > Probes on the oscilloscope menu). Press the Configure Probing System button. A pop-up window will appear where you can select External Scaling. Click the Decibel radio button under the External Scaling section and then set the Gain field to either –6 dB, –12 dB, or –20 dB depending on the attenuator you are using (be sure to include the negative sign). Finally, you will need to manually set the Offset field in this dialog box to zero out the signal.
1168/9B Probes User’s Guide 27
1 Getting Started
N2881A InfiniiMax DC Blocking Capacitors
The DC blocking capacitors are an accessory for the probes. The architecture of the InfiniiMax probing system allows you to place the N2881A DC blocking caps in between the probe amplifier and the probe head as shown in Figure 17. The capacitors block out the DC component of the input signal (up to 30 Vdc).
Figure 17 Blocking Caps Between Probe Amplifier and Head
You can use the blocking capacitors with the N2880A In-Line Attenuators. The order of the two products in the probing system (that is, which one is closest to the probe amplifier) does not matter.
Figure 18 on page 28 shows the frequency response plot of the blocking
capacitors (no probe included).
Figure 18 DC Blocking Cap Insertion Loss (S21) versus Frequency (DC Blocking Cap only)
28 1168/9B Probes User’s Guide
MX0102A Soldering Toolkit
Straight
Tweezers
Cutting
Tweezer s
Double-sided Foam Tape
Low Temperature
Solder Wire
Regular Solder Wire
Probe Tip Wire
Kapton
Tap e
The optional MX0102A soldering toolkit provides tools that can make soldering tasks easier. For instance, you can use the tools available in this kit while soldering the lead wires of the MX0100A Micro probe head to a DUT (see page 44).
Getting Started 1
Table 7 Accessories supplied in the soldering toolkit
Description
Straight Tweezers
Qty Supplied
1 8710-2837 (Anti-magnetic straight pointed tip 120mm) For general purpose manipulation / movement of com-
ponents such as probe tip wires and probe head.
Cutting Tweezers
1 8710-2838 (Narrow oblique head 115mm) To cut a probe tip wire to a desired length.
Kapton Tape (36 yards roll)
1 0460-3121 To provide strain-relief to the neck portion of the probe
head by taping it to a flat surface (such as a DUT circuit board).
1168/9B Probes User’s Guide 29
Part Number
a
1 Getting Started
Description
Double-sided Foam Tape
To provide strain-relief to either the neck portion of the probe head or the plastic housings by taping it to a flat surface such as a tabletop or a DUT circuit board.
Regular Solder Wire
Lead free, .009" diameter, 2 feet long To attach the probe tip wires to a DUT using standard lead-free soldering temperatures (330 °C to 350 °C). (NOTE: This alloy melts at 217
o
C.)
Low Temperature Solder Wire
Lead free, .010" diameter, 2 feet long To attach the probe tip wires to a DUT using a low tem-
perature setting on your soldering iron. (NOTE: This alloy melts at 138
o
C.)
Probe Tip Wire
.004" diameter, 2 feet long To add ground wires to your probe tip if InfiniiMode mea-
surements (differential, single ended, and common mode signals with a single probe tip) are desired. Clip as short as possible using the cutting tweezers included in the kit.
Qty
Supplied
Part Number
a
10 0460-3122
1 MX0102-21302
1 MX0102-21303
1 MX0102-21301
a You can reorder these items using the part numbers included in the table above.
30 1168/9B Probes User’s Guide
Loading...
+ 234 hidden pages