Keysight 1146B User Manual

Keysight 1146B AC/DC Current Probe
User’s Guide
Notices
CAUTION
WARNING
© Keysight Technologies, Inc. 1996 - 2014
No part of this manual may be reprod uced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) with­out prior agreement and written consent from Key­sight Technologies, Inc. as governed by United States and international copyright laws.
Manual Part Number
01146-92005
Second Edition, October 2014
Printed in Malaysia
Published by: Keysight Technologies, Inc. 1400 Fountaingrove Parkway Santa Rosa, CA, 95403
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Keysight dis­claims all warranties, either express or implied, with regard to this manual and any information contained herein, includ ing but not limited to the impl ied war­ranties of merchantability and fitness for a particular purpose. Keysight shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this doc­ument or of any information contained herein. Should Keysight and the user have a separate writ­ten agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
defined in FAR 52.227-19 (June 1987) or any equiva­lent agency regulation or contract clause. Use, dupli­cation or disclosure of Software is subject to Keysight Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June
1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November
1995), as applicable in any technical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the prod­uct or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could resul t in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
Technology Licenses
The hardware and/or software described in this docu­ment are furnished under a license and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a U.S. Gov­ernment prime contract or subcontract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June
1995), or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer software” as
2

Contents

Operation / 5
Before Using the Probe / 5 To Install the Battery / 7 To Perform a Current Measurement / 8 To Clean the Probe / 8
Product Markings / 9
Safety / 10
Concerning the Oscilloscope or Voltage Measuring
Instrument to Which the Probe is Connected / 11
Specifications / 13
Characteristics / 14
Safety Standards / 16 Typical Response Curves / 16
Performance Verification / 18
If the Probe Fails a Test / 19 Test 1. DC Current Measurement Accuracy (10 mV/A
range) / 20
Test 2. AC Current Measurement Accuracy (10 mV/A
range) / 21 Test 3. Bandwidth (10 mV/A range) / 22 Test 4. DC Current Measurement Accuracy (100 mV/A
range) / 23 Test 5. AC Current Measurement Accuracy (100 mV/A
range) / 24 Test 6. Bandwidth Check (100 mV/A range) / 25
3
4
Keysight 1146B AC/DC Current Probe
User’s Guide

Operation

The 1146B AC/DC Current Oscilloscope Probe expands oscilloscope applications in industrial, automotive or power environments, and is ideal for analysis and measurement of distorted current waveforms and harmonics. The probe permits accurate display and measurement of currents from 100 mA to 100A rms, DC to 100 kHz without breaking into the circuit. The probe uses Hall effect technology to measure AC and DC signals. The probe connects directly to an oscilloscope through a 2 meter coaxial cable with an insulated BNC.
The 1146B probe is compatible with any analog or digital oscilloscope or other voltage measuring instrument which has the following features:
- BNC input connector.
- Range capable of displaying 0.2V to 0.5V per division.
- Minimum input impedance of 1 M
Ω.
Before Using the Probe
- Install the battery that was shipped with the probe. Refer to
“To Install the Battery" on page 7.
- Read the warnings listed under “Safety" on page 10.
- Confirm that the probe has not been damaged in transit. If the unit is damaged, or fails to operate according to the specifications, contact your dealer or Keysight representative.
- Check that the contents of the shipment agree with the packing slip. If anything is missing, contact your nearest Keysight Technologies, Inc. Sales Office. If the shipment was damaged, contact the carrier, then contact the nearest Keysight Technologies, Inc. Sales Office.
5
Figure 1 Control and Connector Identification
Zero-Adjust Knob
Operation
Use the zero adjust knob to eliminate the probe’s offset before performing a measurement.
Battery Power Indicator
To ensure proper readings with your current probe, be sure that the green battery indicator is lit during measurement. If not, replace the 9V battery as described in “To Install the Battery" on page 7.
Peak Overload (OL) Indicator
If the red Peak Overload indicator illuminates during measurement, this indicates that the peak value exceeds the instrument response level and that the output is distorted. Use the range selection switch to change the probe to a higher range, if possible.
Range Selection Switch
Use the range selection switch to turn the probe OFF and to select from two ranges: 10 mV/A and 100 mV/A.
6
WARNING
It is possible to change the current probe ‘s range without removing the
WARNING
probe from the current carrying conductor, but never exceed the permissible peak ratings of 1000 mV peak or 2000 mV peak-to-peak maximum. The peak ratings by range are: 10A peak on the 100 mV/A range, 100A peak on the 10 mV/A range.
To Install the Battery
Before using the probe for the first time, you must install the battery. When the probe is turned on, the green battery indication LED should light up. If not, replace the battery.
1 Disconnect the probe from the circuit and the oscilloscope.
Do not replace the battery while probe is in use.
2 Set the probe’s switch to the OFF position.
3 Unscrew the battery compartment screw (shown on Figure 1 on
page 6) and pull out the battery compartment cover.
4 Install the new 9V battery and put the cover back on.
Operation
7
To Perform a Current Measurement
WARNING
WARNING
WARNING
1 Read the warnings in “Safety" on page 10.
2 Connect the current probe to an input channel on the
oscilloscope.
Connect the probe to the oscilloscope or voltage measuring instrument before clamping the probe around a conductor.
3 Set the current probe to its least sensitive range (10 mV/A).
4 Select the 0.5 V/division range on the oscilloscope.
5 Clamp the probe on the conductor to be measured and read the
current flowing directly on your oscilloscope.
You may also use your oscilloscope to amplify the signal while using the 100 mV/A probe range (which offers the best accuracy and least phase shift).
It is possible to change the range on the current probe without removing the probe from the current carrying conductor, but never exceed the permissible peak ratings of 1000 mV peak or 2000 mV peak-to-peak maximum. The peak ratings by range are: 10A peak on the 100 mV/A range, 100A peak on the 10 mV/A range.
Operation
To Clean the Probe
1 Remove the probe from any circuit under test and from the
oscilloscope.
2 Clean the external surfaces of the probe with a soft cloth
dampened with a mixture of mild detergent and water.
3 Clean the mating surfaces of the probe’s jaws so that they are free
of dirt or foreign matter. If rust is present, gently clean the mating surfaces with a soft cloth that has been lightly oiled with standard machine oil. Do not leave excessive oil residue.
Make sure that the probe is completely dry before reconnecting it to a power source.
8

Product Markings

Product Markings
This symbol indicates the Environmental Protection Use Period (EPUP) for the product’s toxic substances for the China RoHS requirements.
The CE symbol indicates the European Community.
CAUTION. Risk of Danger. Refer to the manual for more information.
CAUTION. Risk of Electric Shock. Refer to the manual for more information.
This symbol signifies that the probe is protected by double or reinforced insulation. Only use specified replacement parts when servicing the probe.
Earth terminal symbol: Used to indicate a circuit common connected to grounded chassis.
WEEE Directive notice. This marking is on a notice sticker and is not actually fixed to the product.
9

Safety

WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
Safety
This manual provides information and warnings essential for operating this probe in a safe manner and for maintaining it in safe operating condition. Before using this equipment and to ensure safe operation and to obtain maximum performance from the probe, carefully read and observe the following warnings, cautions, and notes.
This apparatus has been designed and tested in accordance with IEC61010, Safety Requirements for Measuring Apparatus, and has been supplied in a safe condition. Note the external markings on the instrument that are described in this document.
Connect the probe to the oscilloscope or voltage measuring instrument before clamping the probe around a conductor.
Never use the probe on circuits rated higher than 600 Vac RMS CAT II or 300 Vac RMS CAT III or with float voltage greater than 600 V.
Never leave the probe clamped around a cond uctor while it is not connected to an oscilloscope or voltage measuring instrument.
Carefully center the conductor inside the probe jaws and ascertain that the probe is perpendicular to the cond uctor before closing the jaws.
Check the magnetic mating surfaces of the probe jaws; these should be free of dirt, rust, or other foreign matter.
Do not use a probe which is cracked, damaged or has defective leads.
10
Safety
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
Avoid, if possible, the proximity of other cond uctors which may create noise.
Do not install substitute parts or perform any unauthorized modification to the probe.
Do not clamp the probe on a bare/un-insulated conductor.
Do not operate the probe or oscilloscope in the presence of flammable gasses or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard.
Do not use the probe or oscilloscope in a manner not specified by the manufacturer.
Service instructions are for trained service personnel. To avoid dangerous electric shock, do not perform any service unless qualified to do so. Do not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present.
Concerning the Oscilloscope or Voltage Measuring Instrument to Which the Probe is Connected
Whenever it is likely that the ground protection is impaired, you must make the instrument inoperative and secure it against any unintended operation.
If you energize the instrument by an auto transformer (for voltage reduction or mains isolation), the ground pin of the input connector terminal must be connected to the earth terminal of the power source.
11
Safety
WARNING
WARNING
WARNING
Before turning on the instrument, you must connect the protective earth terminal of the instrument to the protective conductor of the (mains) power cord. The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. You must not negate the protective action by using an extension cord (power cable) without a protective conductor (grounding). Ground ing one conductor of a two-conductor outlet is not sufficient protection.
Only fuses with the required rated current, vol tage, and specified type (normal blow, time delay, etc.) should be used. Do not use repaired fuses or short-circuited fuseholders. To do so could cause a shock or fire hazard.
Capacitors inside the instrument may retain a charge even if the instrument is disconnected from its source of supply.
12

Specifications

All probe specifications are warranted based on the following conditions:
- Within one year of calibration
- 23° C ±5° C
- 20% to 75% relative humidity
-Probe zeroed
- 1 minute warm up
- Battery at 9V ± 0.1V
- External magnetic field < 40 A/m
- No adjacent current carrying conductor
-1 M
Ω /100 pF probe termination
Table 1 Specifications
Item Specification
Specifications
Measurement Accuracy (Input Current: dc to 1 kHz)
50 mA to 10A < ±3% of reading ±50 mA
50 mA to 40A n/a < ±4% of reading ±50 mA
40A to 100A n/a < ±15% of reading ±100 mA
Bandwid th (–3 dB) dc to 100 kHz
100 mV/A Sensitivity Setting
(with derating, see “Typical Response Curves" on page 16)
13
10 mV/A Sensitivity Setting

Characteristics

All probe characteristics are the typical performance values and are not warranted. Characteristics are based on these conditions:
- Within one year of calibration
- 23° C ±5° C
- 20% to 75% relative humitidy
-Probe zeroed
- 1 minute warm up
- Battery at 9V ± 0.1V
- External magnetic field < 40 A/m
- No adjacent current carrying conductor
-1 M
- Conductor centered in jaw
Table 2 Characteristics
Characteristics
Ω /100 pF probe termination
100 mV/A
Item
Current Measurement Range 0A to 10A dc or peak ac 0A to 100A dc or peak ac
Phase Shift (DC to 1 kHz, no DC component)
Noise 3 mV 480 µV
Slew Rate 0.3 V/µs 20 mV/µs
Insertion Impedance (50 Hz/60 Hz) 0.01Ω 0.01Ω
Rise or Fall Time 3 µs 4 µs
Influence of Adjacent Conductor < 0.2 mA/A ac < 0.2 mA/A ac
Influence of Conductor Position in Jaw
Overload Indication Red LED (OL) indicates input is out of range
Maximum Working Voltage (refer to safety warnings and standards)
Maximum Floating Voltage (refer to safety warnings and standards)
Sensitivity Setting
< 1.5° dc to 65 Hz < 1.0° dc to 65 Hz
0.5% of reading in jaw 0.5% of reading in jaw
600Vac RMS CAT II 300Vac RMS CAT III
600Vac RMS CAT II 300Vac RMS CAT III
10 mV/A Sensitivity Setting
600Vac RMS CAT II 300Vac RMS CAT III
600Vac RMS CAT II 300Vac RMS CAT III
14
Table 3 Environmental and Mechanical Characteristics
Item Characteristic
Environmental Conditions
Characteristics
Operating Temperature Range
Storage Temperature Range
Temperature Influence < 0.2% per ° C
Max Operating Humidity 10° C to 30° C: 85% RH (without condensation)
Altitude Operating: 0 to 2000m
Indoor Use This probe is rated for indoor use only
Mechanical
Zero Adjustment 20 turn potentiometer
Maximum cable diameter 11.8 mm
Case Protection IP20 per IEC 529
Drop Test 1.0 m on 38 mm of oak on concrete, tested according to IEC
Mechanical Shock 100 G; test per IEC 68-2-27
Vibration Tested per IEC 68-2-6, frequency range 10 Hz to 55 Hz, amplitude
0° C to 50° C
–30° C to 80° C
40° C to 50° C: 45% RH (without condensation)
Non operating: 0 to 12,000m
61010
0.15 mm
Handle Lexan 920A, UL 94 V2
Dimensions 231 mm x 36 mm x 67 mm
Weight 330 g (11.6 oz) with battery
Color Light gray
Output cable Insulated coaxial cable with insulated BNC connector
Output cable length 2m
15
Table 4 Battery Characteristics
Battery Characteristic
Battery Type 9V Alkaline (NEDA 1604A, IEC 6LR61)
Low Battery Indication Green LED (ON) when battery voltage 6.5V
Typical Battery Consumption 8.6 mA
Typical Battery Life 55 hours
Safety Standards
- Double insulation or reinforced insulation between primary or secondary and outer case of the handle, per IEC61010.
- 600V Category III, Pollution degree 2.
- 300V Category IV, Pollution degree 2.
- 5550V 50/60 Hz between primary or secondary and the outer case of the handle.
- 3250V 50/60 Hz between primary and secondary
Typical Response Curves
Characteristics
Figure 2 Typical Frequency Response Curve
16
Figure 3 Typical Phase Shift Response Curve
Characteristics
Accuracy
Figure 4 Linearity for a DC Signal Range 10 mV/A
17

Performance Verification

NOTE
Performance Verification
To ensure that the probe complies with the factory specifications, perform the six tests in this section in the order listed. Connect the required equipment listed in Table 5 as shown in Figure 6 on page 19.
Allow 30 minutes for warm up before starting the tests. While the test system is warming up, clean the magnetic contacts on the probe’s jaw.
Table 5 Required Equipment for Performance Verification Tests
Recommended
Description Critical Specifications
Model/Part Numbers
Digital Multimeter (2 required)
AC/DC voltage and current measurement accuracy better than ±0.1% of reading Input resistance in AC/DC voltage mode
Keysight 34401A or Keysight 3458A
1MΩ
Signal Generator DC to 100 kHz sine waves
Able to generate more than 10 mA in the test coil (e.g. an inductive load).
Patch Cable Assembly Banana plug connectors
36 in (91 cm)
Length
Adapter BNC (m) to dual banana Pomona 1296
Adapter BNC (f) to dual banana Pomona 1269
Banana Jack (2 required) Stackable Pomona 1325-0
Coil 10 turns (requires about 4 ft or
1.2 m of transformer wire or wire wrap wire.)
Construct a 10-turn coil with a diameter of about 4 inches (10 cm) as shown in Figure 5 on page 19. Use wire wrap or transformer wire and attach stackable banana plugs to the coil for connection to the instruments during testing.
Keysight 33120A or Fluke 5500A
Pomona 1440-36-0
OK Industries R30B-0100(100 ft roll of 30AWG wire wrap wire)
18
Figure 5 10 Turn Coil
Performance Verification
Figure 6 Equipment Setup
If the Probe Fails a Test
1 Replace the battery. Verify that the battery voltage is 9V ±0.1V.
2 Ensure that:
- The probe's magnetic contact surfaces are clean
- The probe tip closes completely around the coil's conductors
- The coil is centered in the probe tip opening
- The probe tip is perpendicular to the coil's conductors
19
Performance Verification
NOTE
NOTE
- The probe current flow indicator points in the correct direction
3 Demagnetize the probe by opening and closing the clamp 5 to 10
times.
Test 1. DC Current Measurement Accuracy (10 mV/A range)
1 Configure the instruments as follows:
DMM 1: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . DC Amps
DMM 2: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . DC Volts
Probe’s Range Selector Switch . . . . . . . . . . . . . . . . . . . . 10 mV/A
2 Disconnect the probe from the coil and adjust the zero control to
minimize the probe's DC output voltage. Record this zero offset voltage (V
3 Clamp the probe around the coil.
Ensure that the DC current direction in the coil matches the current direction arrow on the probe. Position the probe as perpendicular as possible to the coil.
4 Set up the Keysight 33120A as follows:
Waveform:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sine
Frequency: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 kHz
Amplitude: . . . . . . . . . . . . . . . . . . . . . . . . . . 500 mVpp (minimum)
Offset: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +1.0 Vdc
5 Measure the current in the coil on DMM 1 and record this value (I
in Table 4.
) in Tabl e 6 on page 21.
z
)
c
The current in the coil must be 10 mA. Increase the generator offset if necessary to achieve a current reading of at least 10 mA.
6 Measure the voltage output by the probe on DMM 2 and record
) in Tab le 6.
(V
probe
7 Do the calculations specified in Tab le 6 to determine if the probe
passes the test.
20
Performance Verification
NOTE
Table 6 DC Current Measurement Accuracy Record (10 mV/A range)
Measurement Recorded Value
Probe Zero Offset Voltage (Vz)
Current in Coil (Ic)
Probe Output Voltage (V
probe
)
Calculate Maximum Probe Output Voltage As: V
probe max
= (Ic * 10 * 10 m V/A * 1.04) + V
Calculate Minimum Probe Output Voltage As: V
probe min
= (Ic * 10 * 10 m V/A * 0.96) + V
z
Probe Passes Test if: V
probe min
V
probe
V
probe max
Test 2. AC Current Measurement Accuracy (10 mV/A range)
8 Set the DMM modes to:
DMM 1: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . AC Amps
DMM 2: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . AC Volts
9 Set the Keysight 33120A to:
Waveform: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sine
Frequency:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 kHz
Amplitude: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Vpp
Offset: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 Vdc
10 Measure the current (I
on page 22.
The current in the coil must be 10 mA. Increase the generator offset if necessary to achieve a current reading of at least 10 mA.
Z
) in the coil on DMM 1 and record in Table 7
c
11 Measure the voltage output by the probe (V
) on DMM 2 and
probe
record in Ta ble 7.
12 Do the calculations specified in Tab le 7 to determine if the probe
passes the test.
21
Performance Verification
NOTE
Table 7 AC Current Measurement Accuracy Record (10 mV/A range)
Measurement Recorded Value
Current in Coil (Ic)
Probe Output Voltage (V
probe
)
Calculate Maximum Probe Output Voltage As: V
probe max
= (Ic * 10 * 10 mV/A * 1.04)
Calculate Minimum Probe Output Voltage As: V
probe min
= (Ic * 10 * 10 mV/A * 0.96)
Probe Passes Test if: V
probe min
V
probe
V
probe max
Test 3. Bandwidth (10 mV/A range)
13 Increase the generated frequency to 100 kHz.
14 Measure the current in the coil (I
page 23.
The current in the coil must be 10 mA. Increase the generator offset if necessary to achieve a current reading of at least 10 mA.
15 Measure the voltage output (V
record in Ta ble 8.
16 Do the calculations specified in Tab le 8 to determine if the probe
passes the test.
) and record in Table 8 on
c
) by the probe on DMM 2 and
probe
22
Performance Verification
NOTE
NOTE
Table 8 Bandwidth Record (10 mV/A range)
Measurement Recorded Value
Current in Coil (Ic)
Probe Output Voltage (V
probe
)
Calculate Maximum Probe Output Voltage As: V
probe max
= (Ic * 10 * 10 mV/A * 1.04)
Calculate Minimum Probe Output Voltage As: V
probe min
= (Ic * 10 * 10 mV/A * 0.96) * 0.707
Probe Passes Test if: V
probe min
V
probe
V
probe max
Test 4. DC Current Measurement Accuracy (100 mV/A range)
17 Configure the instruments as follows:
DMM 1: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . DC Amps
DMM 2: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . DC Volts
Probe’s Range Selector Switch . . . . . . . . . . . . . . . . . . . .100 mV/A
18 Disconnect the probe from the coil and adjust the zero control to
minimize the probe's DC output voltage. Record this zero offset voltage (V
) in Tabl e 9 on page 24.
z
19 Clamp the probe around the coil.
Ensure that the DC current direction in the coil matches the current direction arrow on the probe. Position the probe as perpendicular as possible to the coil.
20 Set up the Keysight 33120A as follows:
Waveform: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sine
Frequency: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 kHz
Amplitude: . . . . . . . . . . . . . . . . . . . . . . . . . . 500 mVpp (minimum)
Offset: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +1.0 Vdc
21 Measure the current in the coil on DMM 1. Record this value (I
Tabl e 9.
The current in the coil must be 10 mA. Increase the generator offset if necessary to achieve a current reading of at least 10 mA.
23
c
) in
Performance Verification
NOTE
22 Measure the voltage output (V
) by the probe on DMM 2, and
probe
record in Ta ble 9.
23 Do the calculations specified in Tab le 9 to determine if the probe
passes the test.
Table 9 DC Current Measurement Accuracy Record (100 mV/A range)
Measurement Recorded Value
Probe Zero Offset Voltage (Vz)
Current in Coil (Ic)
Probe Output Voltage (V
Calculate Maximum Probe Output Voltage As: V
probe max
= (Ic * 10 * 100 mV/A * 1.03) + V
Calculate Minimum Probe Output Voltage As: V
probe min
= (Ic * 10 * 100 mV/A * 0.97) + V
Probe Passes Test if: V
probe min
V
probe
V
Test 5. AC Current Measurement Accuracy (100 mV/A range)
24 Change the DMM settings to:
25 Set up the Keysight 33120A settings to:
)
probe
z
z
probe max
DMM 1: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . AC Amps
DMM 2: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . AC Volts
Waveform: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sine
Frequency: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 kHz
Amplitude: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Vpp
Offset: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 Vdc
26 Measure the current in the coil (Ic) on DMM 1 and record this in
Tabl e 10 on page 25.
The current in the coil must be 10 mA. Increase the generator offset if necessary to achieve a current reading of at least 10 mA.
27 Measure the voltage output (V
) by the probe on DMM 2, and
probe
record in Ta ble 10.
24
Performance Verification
NOTE
28 Do the calculations specified in Table 10 to determine if the probe
passes the test.
Table 10 AC Current Measurement Accuracy Record (100 mV/A range)
Measurement Recorded Value
Current in Coil (Ic)
Probe Output Voltage (V
probe
)
Calculate Maximum Probe Output Voltage As: V
probe max
= (Ic * 10 * 100 mV/A * 1.03)
Calculate Minimum Probe Output Voltage As: V
probe min
= (Ic * 10 * 100 mV/A * 0.97)
Probe Passes Test if: V
probe min
V
probe
V
probe max
Test 6. Bandwidth Check (100 mV/A range)
29 Increase the generated frequency to 100 kHz.
30 Measure the current in the coil (I
in Tabl e 11 on page 26.
The current in the coil must be 10 mA. Increase the generator offset if necessary to achieve a current reading of at least 10 mA.
31 Measure the voltage output (V
record in Ta ble 11.
32 Do the calculations specified in Table 11 to determine if the probe
passes the test.
) and record the current reading
c
by the probe on DMM 2, and
probe)
25
Performance Verification
Table 11 Bandwidth Record (100 mV/A range)
Measurement Recorded Value
Current in Coil (Ic)
Probe Output Voltage (V
probe
)
Calculate Maximum Probe Output Voltage As: V
probe max
= (Ic * 10 * 100 mV/A * 1.03)
Calculate Minimum Probe Output Voltage As: V
probe min
= (Ic * 10 * 100 mV/A * 0.97) * 0.707
Probe Passes Test if: V
probe min
V
probe
V
probe max
26

Index

Numerics
10 mV/A, 7 100 mV/A, 7 10-turn coil, 18
B
battery
indicator, 6 installation, 7
typical life, 16 battery power indicator, 6 battery, installing, 7
C
characteristics, 14 cleaning, 8 compatibility, 5
G
green indicator light, 6
H
Hall effect, 5
operating temperature range, 15
P
Peak Overload (OL) Indicator, 6 peak ratings, 7 probe, cleaning, 8
R
range selection switch, 6 red indicator light, 6
S
safety, 10 safety warnings, 10 specifications, 13
T
transformer wire, 18 Typical Frequency Response, 16 Typical Phase Shift Response, 17
V
verification, 18
I
indicator lights, 6
K
knob, 6
L
Linearity for a DC signal Range, 17
O
OL, 6 Operating, 15
W
warm up time, 18 WEEE Directive, 9 wire wrap, 18
Z
zero adjustment knob, 6
27
Keysight Technologies, Inc
Printed in Malaysia
Manual Part Number: 01146-92005
*01146-92005*
Loading...