This probe has a cable with a built-in wear indicator. When the insulation of the cable
deteriorates, the wear indicator becomes visible. Do not use the probe, when the
contrasting color (Black) is visible through the cable jacket. Using a product with a
worn cable may result in electric shock, fire, or equipment failure.
Avoid contact of the probe cable with hazardous live voltages. The rated voltage of the
probe tip is higher than the rated voltage of the probe cable. If the cable contacts
hazardous live voltages, it will not provide adequate protection. Contact of the cable
with hazardous live voltages may result in electric shock, fire, or equipment failure.
10070/3/4D Probes
User’s Guide
The input voltage should not exceed the input voltage rating of the measuring instrument.
An E2697A 1 MΩ input adapter is required to use these probes with a 50Ω input
oscilloscope. The E2697A includes one 10073D general purpose, 500 MHz, 10:1,
passive probe.
Cleaning. To clean the probe, disconnect it from the instrument and clean it with mild detergent and
water. Make sure the probe is completely dry before reconnecting it to the instrument.
If the probe assembly is used in a manner not specified by the manufacturer, the protection
provided by the probe assembly may be impaired.
QtyDescriptionItem
Keysight 5081-7690 Accessory Kit
1 1Retractable Hook Tip (also available as part number 5018-7697)
2 Colored identification tags
3 Ground Bayonet
4 IC Tip
5 Adjustment Tool
6 Ground Lead (black)
7 BNC adapter (not part of kit, order separately: 5081-7705)
8
1
1
1
1
−
1
5
2
3
4
7
6

© Keysight Technologies 2013 - 2014, 2018
10070-92016, February 2018, Printed in Japan
*10070-92016*
Keysight Technologies
1900 Garden of the Gods Road
Colorado Springs, CO 80907
10
100
1000
0.01 0.1 1 10 100 1000
Voltage (V)
Frequency (MHz)
Derating Curve (Mains Isolated)
10073D
10074D
10070D
10
100
1000
0.01 0.1 1 10 100 1000
Voltage (V)
10073D
10074D
10070D
Frequency (MHz)
Derating Curve (CAT II)
10070D 10073D 10074DCharacteristics
Operating Temperature –10º C to 55° C
Humidity 90% RH @ 40° C non-condensing
Shock
400g 1/2 sine wave on tip end
50g 1/2 sine wave on BNC end
Propagation Delay
7 ns (1.5m probe cable length)
Altitude
3000m (indoor use only)
System Bandwidth 20 MHz 500 MHz 150 MHz
150 Vpk CAT II
(post receptacle mains)
400 Vpk CAT II (post receptacle mains)
1 MΩ 2.2 MΩ 10 MΩ
< 17.5 ns < 700 ps < 2.33 ns
1:1 10:1 10:1
none 6 – 15 pF 9 – 17 pF
no yes yes
≈70 pF ≈12 pF ≈15 pF
System Risetime
(calculated)
Attenuation Ratio
Input Resistance (when
terminated into 1 MΩ)
Input Capacitance
Maximum Input
(dc + peak ac)
Comensation Range
Probe Readout
500 Vpk (mains isolated)300 Vpk (mains isolated)