Keysight 1000B Series User Manual

Page 1
Keysight 1000B Series Oscilloscopes
User’s Guide
Page 2
Notices
CAUTION
WARNING
© Keysight Technologies, Inc. 2008-2009, 2012
Manual Part Number
54139-97013
Edition
Second Edition, May 2012
Printed in Malaysia
Keysight Technologies, Inc. 1900 Garden of the Gods Road Colorado Springs, CO 80907 USA
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Keysight disclaims all war­ranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fit­ness for a particular purpose. Keysight shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Keysight and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall con­trol.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor­dance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a U.S. Government prime contract or subcon­tract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or d isclosure of Software is subject to Keysight Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Govern­ment users will receive no greater than Lim­ited Rights as defined in FAR 52.227-14
(June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any tech­nical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a haz­ard. It calls attention to an operat­ing procedure, practice, or the like that, if not correctly performed or adhered to, could resul t in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
See also Appendix A, “Safety Notices,” starting on page 143.
2 Keysight 1000B Series Oscilloscopes User’s Guide
Page 3
Keysight 1000B Series Oscilloscopes—At a Glance
The Keysight 1000B Series oscilloscopes are low-cost portable digital storage oscilloscopes (DSOs) that deliver these powerful features:
Channel counts, input bandwidths, sample rates, and memory depths shown in the following table:
Table 1 Keysight 1000B Series Oscilloscope Models
Model Channels Input Bandwid th Sample Rate
(2 channels on-1 channel on)
DSO1052B 2 50 MHz 500 MSa/s-1 GSa/s 8-16 kpts
DSO1072B 2 70 MHz 500 MSa/s-1 GSa/s 8-16 kpts
DSO1102B 2 100 MHz 500 MSa/s-1 GSa/s 8-16 kpts
DSO1152B 2 150 MHz 500 MSa/s-1 GSa/s 8-16 kpts
Bright 5.7 inch QVGA (320 x 240) TFT color LED display and small footprint (to save bench space).
Up to 400 wfms/s refresh rate.
Automatic voltage and time measurements (22) and cursor measurements.
Powerful triggering (edge, pulse width, video, and alternate modes) with adjustable sensitivity (to filter noise and avoid false triggers).
Math function waveforms: add, subtract, multiply, FFT.
USB ports (host and device) for easy printing, saving, and sharing of waveforms, setups, screen BMP files, and CSV data files.
Internal storage for 10 waveforms and 10 setups.
Special digital filter and waveform recorder.
Built-in 6-digit hardware frequency counter.
Multi-language (11) user interface menus and built-in help.
Memory (2 channels on-1 channel on)
Keysight 1000B Series Oscilloscopes User’s Guide 3
Page 4
In This Book
1 Getting Started
2 Displaying Data
3 Capturing Data
4 Making Measurements
This guide shows how to use the Keysight 1000B Series oscilloscopes.
Describes the basic steps to take when first using the oscilloscope.
Describes how to use the horizontal and vertical controls, channel settings, math waveforms, reference waveforms, and display settings.
Describes acquisition modes and how to set up triggers.
Describes voltage, time, and cursor measurements.
5 Saving, Recalling, and Printing Data
Describes how to save, recall, and print data.
6 Oscilloscope Utility Settings
Describes other oscilloscope settings found in the Utility menu.
7 Reference
Contains reference information for the 1000B Series oscilloscopes.
4 Keysight 1000B Series Oscilloscopes User’s Guide
Page 5

Contents

Keysight 1000B Series Oscilloscopes—At a Glance 3
In This Book 4
Figures 13
Tab les 15
1 Getting Started 17
Step 1. Inspect the package contents 18
Step 2. Turn on the oscilloscope 19
Step 3. Load the default oscilloscope setup 22
Step 4. Input a waveform 23
23
Step 5. Use Auto Scale 24
Step 6. Compensate probes 26
Low Frequency Compensation 26 High Frequency Compensation 26
Step 7. Become familiar with the Front Panel Controls 28
Front Panel Overlays for Different Languages 29
Step 8. Become familiar with the oscilloscope display 30
Using the Oscilloscope Softkey Menus 31
Step 9. Use the Run Control keys 33
Step 10. Access the built-in help 34
Keysight 1000B Series Oscilloscopes User’s Guide 5
Page 6
Contents
Securing the Oscilloscope 35
2 Displaying Data 37
Using the Horizontal Controls 38
To adjust the horizontal scale 39 To adjust the horizontal position 40 To display the zoomed time base 41 To change the horizontal time base (Y-T, X-Y, or Roll) 42 To view the sample rate 44
Using the Vertical Controls 45
To turn waveforms on or off (channel, math, or reference) 46 To adjust the vertical scale 46 To adjust the vertical position 46 To specify channel coupling 47 To specify a bandwidth limit 49 To specify the probe attenuation 50 To use a digital filter 50 To change the Volts/Div control sensitivity 51 To invert a waveform 52
Using Math Function Waveforms 54
To add, subtract, or multiply waveforms 54 To display the frequency domain using FFT 55
Using Reference Waveforms 58
To save a reference waveform 58 To export or import reference waveforms 58 To return the reference waveform to its default scale 59
Changing the Display Settings 60
To display waveforms as vectors or dots 60 To clear the display 61 To set waveform persistence 61 To adjust waveform intensity 61
6 Keysight 1000B Series Oscilloscopes User’s Guide
Page 7
To change the grid 61 To adjust the grid brightness 62 To invert screen colors 62 To change the menu display time 62
3 Capturing Data 63
Overview of Sampling 64
Sampling Theory 64 Aliasing 64 Oscilloscope Bandwidth and Sample Rate 65 Oscilloscope Rise Time 67 Oscilloscope Bandwidth Required 68 Memory Depth and Sample Rate 69
Choosing the Sampling Mode 70
To select the real-time sampling mode 70 To select the equivalent-time sampling mode 71
Choosing the Acquisition Mode 73
To select the Normal acquisition mode 74 To select the Average acquisition mode 74 To select the Peak Detect acquisition mode 75
To turn OFF/ON sine(x)/x interpolation 77
Contents
Recording/Playing-back Waveforms 78
To record waveforms 78 To play-back waveforms 79 To store recorded waveforms 80
Adjusting the Trigger Level 82
To adjust the trigger level 82 To force a trigger 82
Choosing the Trigger Mode 84
To set up edge triggers 84 To set up pulse width triggers 85
Keysight 1000B Series Oscilloscopes User’s Guide 7
Page 8
Contents
To set up video triggers 86 To set up alternate triggers 88
Setting Other Trigger Parameters 90
To set the trigger sweep 90 To set the trigger coupling 90 To set the trigger high-frequency reject coupling 91 To change the trigger sensitivity 92 To specify a trigger holdoff 93
Using the External Trigger Input 94
4 Making Measurements 95
Displaying Automatic Measurements 96
To display an automatic measurement 97 To clear automatic measurements from the display 97 To display or hide all automatic measurements 97
8 Keysight 1000B Series Oscilloscopes User’s Guide
Voltage Measurements 98
Vmax (Maximum Voltage) 98 Vmin (Minimum Voltage) 99 Vpp (Peak-to-Peak Voltage) 99 Vtop (Top Voltage) 99 Vbase (Base Voltage) 99 Vamp (Amplitude Voltage = Vtop - Vbase) 99 Vavg (Average Voltage) 99 Vrms (Root-Mean-Square Voltage) 99 Overshoot 100 Preshoot 100
Time Measurements 101
Period 101 Frequency 102 Rise Time 102 Fall Time 102
Page 9
Positive Pulse Wid th 103 Negative Pulse Width 103 Positive Duty Cycle 103 Negative Duty Cycle 103 Delay Between Rising Edges 104 Delay Between Falling Edges 104 Phase Between Rising Edges 105 Phase Between Falling Edges 105
Counter (Frequency) 106
Making Cursor Measurements 107
To use manually adjustable cursors 108 To use tracking cross-hair cursors 109 To display cursors for automatic measurements 110
5 Saving, Recalling, and Printing Data 111
Contents
Saving and Recalling Data 112
To save and recall waveforms 112 To save and recall oscilloscope setups 113 To save screens to BMP or PNG format files 114 To save data to CSV format files 115
Using the Disk Manager 116
To switch between files, path, and directory panes 117 To navigate the directory hierarchy 117 To create new folders 117 To edit folder/file names 118 To delete folders 118 To rename folders 119 To delete files 119 To recall files 119 To rename files 119 To display disk information 120
Keysight 1000B Series Oscilloscopes User’s Guide 9
Page 10
Contents
Printing Screens 121
To choose a PictBridge printer 122 To print with inverted screen colors 122 To choose color or grayscale printing 123 To copy a screen to the printer 123
6 Oscilloscope Utility Settings 125
Displaying System Information 127
Turning Sound ON or OFF 127
Setting the Language (Menu and Help) 128
Performing Mask Tests 129
To enable/disable mask tests 129 To select the source channel for mask tests 129 To run/stop a mask test 130 To turn on/off the mask test message display 130 To set the mask test output condition 130 To stop a mask test on the output condition 132 To set up masks 132
Setting Preferences 135
To set up the screen saver 135 To select the vertical scale reference level 135 To select the USB device port function 135
Running Self-Calibration 137
7 Reference 139
Environmental Conditions 140
Overvoltage Category 140 Pollution Degree 140 Pollution Degree Definitions 140
Measurement Category 141
10 Keysight 1000B Series Oscilloscopes User’s Guide
Page 11
Index 145
Contents
Measurement Category Definitions 141 Transient Withstand Capability 141
141
Specifications and Characteristics 142
Cleaning the Oscilloscope 142
Contacting Keysight 142
ASafety Notices143
Warnings 143
Safety Symbols 144
Keysight 1000B Series Oscilloscopes User’s Guide 11
Page 12
Contents
12 Keysight 1000B Series Oscilloscopes User’s Guide
Page 13

Figures

Figure 1. Power Switch 21 Figure 2. [Default Setup] Key 22 Figure 3. [Auto Scale] Key 24 Figure 4. Low Frequency Probe Compensation 26 Figure 5. High Frequency Probe Compensation 27 Figure 6. Front Panel 28 Figure 7. Oscilloscope Display 30 Figure 8. Softkey Menus 31 Figure 9. Run Control Keys 33 Figure 10. Securing the Instrument 35 Figure 11. Horizontal Controls 38 Figure 12. Status Bar, Trigger Position, and Horizontal Scale Control Indicators 39 Figure 13. Zoomed Time Base Window 41 Figure 14. X-Y Display Format Showing Out-of-Phase Waveforms 43 Figure 15. Vertical Controls 45 Figure 16. DC Coupling Control 48 Figure 17. AC Coupling Control 48 Figure 18. BW Limit Control OFF 49 Figure 19. BW Limit Control ON 50 Figure 20. Waveform Before Inversion 52 Figure 21. Waveform After Inversion 53 Figure 22. Math Scale Setting Value 54 Figure 23. FFT Waveform 57 Figure 24. [Display] Key 60 Figure 25. Alaising 65 Figure 26. Theoretical Brick-Wall Frequency Response 66 Figure 27. Sample Rate and Oscilloscope Bandwidth 67 Figure 28. Real-Time Sampling Mode 70
Keysight 1000B Series Oscilloscopes User’s Guide 13
Page 14
Figures
Figure 29. Equivalent-Time (Repetitive) Sampling Mode 71 Figure 30. [Acquire] Key 73 Figure 31. Noisy Waveform Without Averaging 74 Figure 32. Noisy Waveform With Averaging 75 Figure 33. Peak Detect Waveform 76 Figure 34. Trigger Controls 82 Figure 35. Line Synchronization 87 Figure 36. Field Synchronization 88 Figure 37. Alternate Triggers 89 Figure 38. Trigger Holdoff 93 Figure 39. [Meas] Key 96 Figure 40. Voltage Measurement Points 98 Figure 41. Period and Frequency Measurements 101 Figure 42. Rise Time and Fall Time Measurements 102 Figure 43. Positive Pulse Width and Negative Pulse Width Measurements 103 Figure 44. Delay Measurements 104 Figure 45. Phase Measurements 105 Figure 46. [Cursors] Key 108 Figure 47. USB Host Port on Front Panel 111 Figure 48. [Save/Recall] Key 112 Figure 49. Disk Manager 116 Figure 50. Editing Folder/File Names in Disk Manager 118 Figure 51. USB Device Port 121 Figure 52. [Print] Key 122 Figure 53. [Utility] Key 126 Figure 54. Mask Test Display 130 Figure 55. Schematic Diagram of Mask Pass/Fail out 131 Figure 56. Mask Pass/Fail out 132 Figure 57. Mask Test Mask Setting 133 Figure 58. Calibration Screen 137
14 Keysight 1000B Series Oscilloscopes User’s Guide
Page 15

Tables

Table 1. Keysight 1000B Series Oscilloscope Models 3 Table 2. Power Requirements 19 Table 3. Environmental Characteristics 20 Table 4. Auto Scale Default Settings 25 Table 5. Front Panel Controls 29 Table 6. FFT Window Characteristics 56
Keysight 1000B Series Oscilloscopes User’s Guide 15
Page 16
Tables
16 Keysight 1000B Series Oscilloscopes User’s Guide
Page 17
Keysight 1000B Series Oscilloscopes User’s Guide

1 Getting Started

Step 1. Inspect the package contents 18 Step 2. Turn on the oscilloscope 19 Step 3. Load the default oscilloscope setup 22 Step 4. Input a waveform 23 Step 5. Use Auto Scale 24 Step 6. Compensate probes 26 Step 7. Become familiar with the Front Panel Controls 28 Step 8. Become familiar with the oscilloscope display 30 Step 9. Use the Run Control keys 33 Step 10. Access the built-in help 34 Securing the Oscilloscope 35
This chapter describes the basic steps to take when first using the oscilloscope.
17
Page 18
1 Getting Started

Step 1. Inspect the package contents

1 Inspect the shipping container for damage.
Keep a damaged shipping container or cushioning material until you have inspected the contents of the shipment for completeness and have checked the oscilloscope mechanically and electrically.
2 Verify that you received the following items in the oscilloscope packaging:
• Oscilloscope.
• Power cord.
• N2862A 10:1 10 MΩ passive probes, quantity = 2.
• Documentation CD.
• Front panel overlay (if language option other than English is chosen).
If anything is missing, or if you need to order additional probes, power cords, etc., contact your nearest Keysight Technologies sales office.
3 Inspect the oscilloscope.
• If there is mechanical damage or a defect, or if the oscilloscope does not operate properly or does not pass performance tests, notify your Keysight Technologies sales office.
• If the shipping container is damaged, or the cushioning materials show signs of stress, notify the carrier; then, contact your nearest Keysight Technologies sales office.
Keep the shipping materials for the carrier’s inspection.
The Keysight Technologies sales office will arrange for repair or replacement at Keysight’s option without waiting for claim settlement.
18 Keysight 1000B Series Oscilloscopes User’s Guide
Page 19

Step 2. Turn on the oscilloscope

WARNING
The next few steps (turning on the oscilloscope, loading the default setup, and inputting a waveform) will provide a quick functional check to verify the oscilloscope is operating correctly.
1 Connect the power cord to a power source.
Use only power cords designed for your oscilloscope.
Use a power source that delivers the required power.
Table 2 Power Requirements
Name Typical Value
Line rating: ~Line 50 W max
Getting Started 1
100-120 V/50/60/400 Hz, ±10% 100-240 V/50/60 Hz, ±10%
To avoid electric shock, be sure the oscilloscope is properly grounded.
Keysight 1000B Series Oscilloscopes User’s Guide 19
Page 20
1 Getting Started
Table 3 Environmental Characteristics
Name Typical Value
Ambient temperature: Operating 0 °C to +50 °C
Non-operating –20 °C to +60 °C
Humidity: Operating 80% RH (non-condensing) at +40 °C for 24 hr
Non-operating 60% RH (non-condensing) at +60 °C for 24 hr
Altitude: Operating to 3,000 m (9,842 ft)
Non-operating to 15,000 m (49,213 ft)
Vibration: Keysight class GP and MIL-PRF-28800F; Class 3 random
Shock: Keysight class GP and MIL-PRF-28800F; (operating 30 g, 1/2
sine, 11-ms duration, 3 shocks/axis along major axis. Total of 18 shocks)
Pollution degree 2: Normally only dry non-conductive pollution occurs.
Occasionally a temporary conductivity caused by condensation must be expected.
Indoor use: Rated for indoor use only.
20 Keysight 1000B Series Oscilloscopes User’s Guide
Page 21
2 Turn on the oscilloscope.
Getting Started 1
Power switch

Figure 1 Power Switch

Keysight 1000B Series Oscilloscopes User’s Guide 21
Page 22
1 Getting Started

Step 3. Load the default oscilloscope setup

You can recall the factory default setup any time you want to return the oscilloscope to its original setup.
1 Press the front panel [Default Setup] key.

Figure 2 [Default Setup] Key

2 When the Default menu appears, press [Menu On/Off] to turn off the menu.
(The Undo softkey in the Default menu lets you cancel the default setup and go back to the previous setup.)
22 Keysight 1000B Series Oscilloscopes User’s Guide
Page 23

Step 4. Input a waveform

CAUTION
1 Input a waveform to a channel of the oscilloscope.
Use one of the supplied passive probes to input the Probe Comp signal from the front panel of the oscilloscope.
To avoid damage to the oscilloscope, make sure that the input voltage at the BNC connector does not exceed the maximum voltage (300 Vrms, CAT I).
Getting Started 1
Keysight 1000B Series Oscilloscopes User’s Guide 23
Page 24
1 Getting Started

Step 5. Use Auto Scale

The oscilloscope has an auto scale feature that automatically sets the oscilloscope controls for the input waveforms present.
24 Keysight 1000B Series Oscilloscopes User’s Guide

Figure 3 [Auto Scale] Key

Auto scale requires waveforms with a frequency greater than or equal to 50 Hz and a duty cycle greater than 1%.
1 Press the front panel [Auto Scale] key. 2 When the AUTO menu appears, press [Menu On/Off] to turn off the menu.
The oscilloscope turns on all channels that have waveforms applied and sets the vertical and horizontal scales appropriately. It also selects a time base range based on the trigger source. The trigger source selected is the highest-numbered channel that has a waveform applied.
(The Undo softkey in the AUTO menu lets you cancel the auto scale and go back to the previous setup.)
Page 25
Getting Started 1
The oscilloscope is configured to the following default control settings:
Table 4 Auto Scale Default Settings
Menu Setting
Horizontal time base Y-T (amplitude vs. time)
Acquisition mode Normal
Vertical coupling Adjusted to AC or DC according
to the waveform.
Vertical "V/div" Adjusted
Volts/Div Coarse
Bandwidth limit OFF
Waveform invert OFF
Horizontal position Center
Horizontal "s/div" Adjusted
Trigger type Edge
Trigger source Measure the channel with input
waveform automatically.
Trigger coupling DC
Trigger voltage Midpoint setting
Trigger sweep Auto
Keysight 1000B Series Oscilloscopes User’s Guide 25
Page 26
1 Getting Started
Low frequency compensation adjustment
Over Compensated
Under Compensated
Correctly Compensated

Step 6. Compensate probes

Compensate probes to match your probe to the input channel. You should compensate a probe whenever you attach it for the first time to any input channel.

Low Frequency Compensation

For the supplied passive probes: 1 Set the Probe menu attenuation to 10X. If you use the probe hooktip, ensure a
proper connection by firmly inserting the tip onto the probe.
2 Attach the probe tip to the probe compensation connector and the ground lead
to the probe compensator ground connector.
3 Press the [Auto Scale] front panel key.
Figure 4 Low Frequency Probe Compensation
4 If waveform does not appear like the Correctly Compensated waveform shown
in Figure 4, then use a nonmetallic tool to adjust the low frequency compensation adjustment on the probe for the flattest square wave possible.

High Frequency Compensation

For the supplied passive probes:
26 Keysight 1000B Series Oscilloscopes User’s Guide
Page 27
Getting Started 1
High frequency compensation adjustments
Over Compensated
Under Compensated
Correctly Compensated
1 Using the BNC adapter, connect the probe to a square wave generator. 2 Set the square wave generator to a frequency of 1 MHz, an amplitude of
3 Vp-p, and an output termination of 50Ω.
3 Press the [Auto Scale] front panel key.
Figure 5 High Frequency Probe Compensation
4 If waveform does not appear like the Correctly Compensated waveform shown
in Figure 5, then use a nonmetallic tool to adjust the 2 high frequency compensation adjustments on the probe for the flattest square wave possible.
Keysight 1000B Series Oscilloscopes User’s Guide 27
Page 28
1 Getting Started
Entry knob Horizontal controls Run controls
Menu controlsVertical controls Trigger controls
Softkeys
Setup and File controls
Math functions Reference waveforms

Step 7. Become familiar with the Front Panel Controls

Before using the oscilloscope, familiarize yourself with the front panel controls.
The front panel has knobs, keys, and softkeys. Knobs are used most often to make adjustments. Keys are used for run controls and to change other oscilloscope settings via menus and softkeys.
28 Keysight 1000B Series Oscilloscopes User’s Guide

Figure 6 Front Panel

Page 29
Getting Started 1
The definitions of the front panel knobs, keys, and softkeys are as follows:
Table 5 Front Panel Controls
Controls Consists of these knobs and keys
Entry knob For the adjustment of defined controls.
Setup controls [Auto Scale] and [Default Setup] front panel keys.
File controls [Save/Recall] and [Print] front panel keys.
Horizontal controls Position knob, [Horiz] front panel key, and scale knob.
Run controls [Run/Stop] and [Single] front panel keys.
Menu controls [Cursors], [Meas], [Acquire], [Display] , and [Utility] front panel keys.
Trigger controls Trigger [Level] knob, [Menu], and [Force Trigger] front panel keys.
Vertical controls Vertical position knobs, vertical scale knobs, channel ([1], [2], etc.)
[Math], and [Ref] front panel keys.
Softkeys Five gray keys from top to bottom on the right-hand side of the screen,
which select the adjacent menu items in the currently displayed menu.

Front Panel Overlays for Different Languages

If you choose a language option other than English, you get a front panel overlay for your language option.
To install a front panel overlay: 1 Insert the tabs on the left side of the overlay into the appropriate slots on the
front panel.
2 Gently press the overlay over the knobs and buttons. 3 When the overlay is against the front panel, insert the tabs on the right side of
the overlay into the slots on the front panel.
4 Let the overlay flatten out. It should remain secure on the front panel.
Keysight 1000B Series Oscilloscopes User’s Guide 29
Page 30
1 Getting Started
Channel 1 Status, DC coupling, 500 mV per division
Timebase Status, 500 us per division
Channel 1 Reference
Acquisition Status
Waveform
Trigger, rising edge, channel 1,
1.56 V
Waveform window position in memory
Trigger position in memory
Trigger position in waveform window

Step 8. Become familiar with the oscilloscope display

30 Keysight 1000B Series Oscilloscopes User’s Guide

Figure 7 Oscilloscope Display

Page 31

Using the Oscilloscope Softkey Menus

Getting Started 1
Figure 8 Softkey Menus
When one of the oscilloscope front panel keys turns on a menu, you can use the five softkeys to choose items from the menu.
Some common menu item choices are:
Accesses the next page of items in the menu.
Accesses the previous page of items in the menu.
Keysight 1000B Series Oscilloscopes User’s Guide 31
Page 32
1 Getting Started
Returns to the previous menu in the hierarchy.
The [Menu On/Off] front panel key turns off the menu or turns on the last accessed menu on again. The Menu Display item in the Display menu lets you select the amount of time menus are displayed (see “To change the menu display time” on page 62).
32 Keysight 1000B Series Oscilloscopes User’s Guide
Page 33

Step 9. Use the Run Control keys

There are two front panel keys for starting and stopping the oscilloscope’s acquisition system: [Run/Stop] and [Single].
Getting Started 1

Figure 9 Run Control Keys

• When the [Run/Stop] key is green, the oscilloscope is acquiring data. To stop acquiring data, press [Run/Stop]. When stopped, the last acquired waveform is displayed.
• When the [Run/Stop] key is red, data acquisition is stopped. To start acquiring data, press [Run/Stop].
• To capture and display a single acquisition (whether the oscilloscope is running or stopped), press [Single]. After capturing and displaying a single acquisition, the [Run/Stop] key is red.
Keysight 1000B Series Oscilloscopes User’s Guide 33
Page 34
1 Getting Started

Step 10. Access the built-in help

The oscilloscope has built-in quick help information. To access the built-in help: 1 Press and hold the front panel key, softkey, or pushable knob on which you
would like quick help information.
The built-in help is available in 11 different languages (see “Setting the Language
(Menu and Help)” on page 128).
34 Keysight 1000B Series Oscilloscopes User’s Guide
Page 35

Securing the Oscilloscope

To secure a 1000B Series oscilloscope to its location, you can use a Kensington lock or the security loop.
Getting Started 1
Hole for Kensington lock
Security loop

Figure 10 Securing the Instrument

Keysight 1000B Series Oscilloscopes User’s Guide 35
Page 36
1 Getting Started
36 Keysight 1000B Series Oscilloscopes User’s Guide
Page 37
Keysight 1000B Series Oscilloscopes User’s Guide

2 Displaying Data

Using the Horizontal Controls 38 Using the Vertical Controls 45 Using Math Function Waveforms 54 Using Reference Waveforms 58 Changing the Display Settings 60
This chapter describes how to use the horizontal and vertical controls, channel settings, math waveforms, reference waveforms, and display settings.
37
Page 38
2 Displaying Data

Using the Horizontal Controls

The horizontal controls consist of:
• The horizontal scale knob — changes the oscilloscope’s time per division setting using the center of the screen as a reference.
• The horizontal position knob — changes the position of the trigger point relative to the center of the screen.
•The [Horiz] key — displays the Horizontal menu which lets you display the zoomed (delayed) time base, change the time base mode, and display the sample rate.

Figure 11 Horizontal Controls

Figure 12 shows the screen icon descriptions and control indicators.
38 Keysight 1000B Series Oscilloscopes User’s Guide
Page 39
Displaying Data 2
Horizontal scale setting
Sample rate
Displayed waveform window ([ ]) position in memory
Trigger position in memory
Trigger position in waveform window

To adjust the horizontal scale

Keysight 1000B Series Oscilloscopes User’s Guide 39

Figure 12 Status Bar, Trigger Position, and Horizontal Scale Control Indicators

• Turn the horizontal scale knob to change the horizontal time per division (time/div) setting (and the oscilloscope’s sample rate — see “Memory Depth
and Sample Rate” on page 69).
The time/div setting changes in a 1-2-5 step sequence.
The time/div setting is also known as the sweep speed.
Page 40
2 Displaying Data
When the time/div is set to 50 ms/div or slower, the oscilloscope enters Slow Scan mode (see “Slow Scan Mode” below).
When the horizontal scale is set to 20 ns or faster, the oscilloscope uses sine(x)/x interpolation to expand the horizontal time base.
• Push the horizontal scale knob to toggle between the zoomed time base and the normal time base display (see “To display the zoomed time base” on page 41).
The time/div setting is displayed in the status bar at the top left of the screen. Because all channels are displayed in the same time base (except in the Alternate trigger mode), the oscilloscope displays one time/div setting for all channels.
Slow Scan Mode
When the horizontal scale is set to 50 ms/div or slower, the oscilloscope enters Slow Scan mode.
In the Slow Scan mode, peak detect acquisition is used so that no data is missed (even the though the Acquire menu may show a different acquisition mode setting). The oscilloscope acquires sufficient data for the pre-trigger part of the display, then waits for the trigger. When the trigger occurs, the oscilloscope continues to capture data for the post-trigger part of the display.
When using the Slow Scan mode to view low frequency signals, the channel coupling should be set to “DC”.
The Slow Scan mode lets you see dynamic changes (like the adjustment of a potentiometer) on low frequency waveforms. For example, Slow Scan mode is often used in applications like transducer monitoring and power supply testing.

To adjust the horizontal position

• Turn the horizontal position knob to change the position of the trigger point relative to the center of the screen.
The position knob adjusts the horizontal position of all channels, math functions, and reference waveforms.
• Push the horizontal position knob to “zero” the trigger point (in other words, move it to the center of the screen).
40 Keysight 1000B Series Oscilloscopes User’s Guide
Page 41

To display the zoomed time base

Zoomed window
Zoomed window view
The zoomed time base (also known as delayed sweep time base), magnifies a portion of the original waveform display (now on the top half of the screen) and displays it in a zoomed time base on the bottom half of the screen.
1 To toggle the zoomed time base “ON” or “OFF”, either push the horizontal scale
knob or press [Horiz] key followed by the Zoom softkey in the Horizontal menu.
2 When the zoomed time base is “ON”:
The top half of the display shows the original waveform and the portion
being magnified.
The horizontal scale knob changes the magnification (widens or narrows the
area of magnification).
The horizontal position knob moves the area of magnification forward and
backward on the original waveform.
The bottom half of the display shows the magnified data in the zoomed time
base.
Displaying Data 2
Figure 13 Zoomed Time Base Window
Keysight 1000B Series Oscilloscopes User’s Guide 41
Page 42
2 Displaying Data

To change the horizontal time base (Y-T, X-Y, or Roll)

1 Press [Horiz]. 2 In the Horizontal menu, press Time Base. 3 Continue pressing the Time Base softkey or turn the entry knob to select
between:
Y-T Amplitude vs. time. This is the typical horizontal time base
setting.
X-Y Channel 2 (X-axis) vs. Channel 1 (Y-axis), see “X-Y Format” on
page 42.
Roll In Roll mode, the waveform display rolls from right to left, and
the minimum horizontal scale setting is 500 ms/div. No trigger or horizontal position control is available. Roll mode is used in applications similar to the ones for which Slow Scan mode is used (see “Slow Scan Mode” on page 40).
42 Keysight 1000B Series Oscilloscopes User’s Guide
X-Y Format
This format compares the voltage level of two waveforms point by point. It is useful for studying phase relationships between two waveforms. This format only applies to channels 1 and 2. Choosing the X-Y display format displays channel 1 on the horizontal axis and channel 2 on the vertical axis.
The oscilloscope uses the untriggered sample acquisition mode and waveform data is displayed as dots. The sampling rate can vary from 4 kSa/s to 100 MSa/s, and the default sampling rate is 1 MSa/s.
Page 43
Figure 14 X-Y Display Format Showing Out-of-Phase Waveforms
Displaying Data 2
Keysight 1000B Series Oscilloscopes User’s Guide 43
Page 44
2 Displaying Data

To view the sample rate

The following modes or functions are not available in X-Y format:
• Automatic voltage or time measurements.
• Cursor measurements.
•Mask testing.
• Math function waveforms.
• Reference waveforms.
• Zoomed time base display.
• Displaying waveforms as vectors.
• Horizontal position knob.
• Trigger controls.
1 Press [Horiz]. 2 In the Horizontal menu, the Sa Rate menu item d isplays the sample rate used for
the current horizontal scale setting.
44 Keysight 1000B Series Oscilloscopes User’s Guide
See Also “Memory Depth and Sample Rate” on page 69.
Page 45

Using the Vertical Controls

Scale knobs
Position knobs
The vertical controls consist of:
• The channel ([1], [2], [3], and [4]), [Math], and [Ref] front panel keys — turn waveforms on or off (and display or hide their menus).
• The vertical scale knobs — change the amplitude per division setting for a waveform, using either ground or the center of the screen as a reference (depending on a preference setting).
• The vertical position knobs — change the vertical position of the waveform on the screen.
Displaying Data 2

Figure 15 Vertical Controls

Keysight 1000B Series Oscilloscopes User’s Guide 45
Page 46
2 Displaying Data

To turn waveforms on or off (channel, math, or reference)

To adjust the vertical scale

Pressing the channel ([1], [2], [3], and [4]), [Math], and [Ref] front panel keys have the following effect:
• If the waveform is off, the waveform is turned on and its menu is displayed.
• If the waveform is on and its menu is not displayed, its menu will be displayed.
• If the waveform is on and its menu is displayed, the waveform is turned off and its menu goes away.
When an input channel waveform is on:
• Turn its vertical scale knob to change the amplitude per division setting.
The amplitude/div setting changes in a 1-2-5 step sequence from 2 mV/div to 10 V/div (with “1X” probe attenuation).
Either ground or the center of the screen is used as a reference, depending on the “Expand Reference” preference setting (see “To select the vertical scale
reference level” on page 135). The “center of screen” reference is not available
for math function or reference waveforms.
• Push its vertical scale knob to toggle between vernier (fine scale) adjustment and normal adjustment.
With vernier adjustment, the amplitude/div setting changes in small steps between the normal (coarse scale) settings.
The Volts/Div item in a channel’s menu also toggles between vernier and normal adjustment (see “To change the Volts/Div control sensitivity” on page 51).
Vernier adjustment is not available for math function or reference waveforms.
The amplitude/div setting is displayed in the status bar at the bottom of the screen.

To adjust the vertical position

Adjusting their vertical position lets you compare waveforms by aligning them above one another or on top of each other.
When an input channel waveform is on:
46 Keysight 1000B Series Oscilloscopes User’s Guide
Page 47
• Turn the vertical position knob to change the vertical position of the waveform on the screen.
Notice that the ground reference symbol on the left side of the display moves with the waveform.
• Push the vertical position knob to “zero” the ground reference (in other words, move it to the center of the screen).
Notice that, as you adjust the vertical position, a message showing the position of the ground reference relative to the center of the screen is temporarily displayed in the lower left-hand corner of the screen.

To specify channel coupling

1 If the channel’s menu is not currently displayed, press the channel key ([1], [2],
[3], or [4]).
2 In the Channel menu, press Coupling. 3 Continue pressing the Coupling softkey or turn the entry knob to select
between:
Displaying Data 2
DC Passes both DC and AC components of the input waveform to
the oscilloscope. See Figure 16. You can quickly measure the DC component of the waveform by
simply noting its distance from the ground symbol.
AC Blocks the DC component of the input waveform and passes
the AC component. See Figure 17. This lets you use greater sensitivity (amplitude/div settings) to
display the AC component of the waveform.
GND The waveform is disconnected from the oscilloscope input.
Keysight 1000B Series Oscilloscopes User’s Guide 47
Page 48
2 Displaying Data
DC Coupling
DC Coupling Status
Figure 16 DC Coupling Control
AC Coupling
AC Coupling Status
Figure 17 AC Coupling Control
48 Keysight 1000B Series Oscilloscopes User’s Guide
Page 49

To specify a bandwidth limit

When high frequency components of a waveform are not important to its analysis, the bandwidth limit control can be used to reject frequencies above 20 MHz. See
Figure 19 and Figure 18.
1 If the channel’s menu is not currently displayed, press the channel key ([1], [2],
[3], or [4]).
2 In the Channel menu, press BW Limit to toggle the bandwidth limit setting “ON”
and “OFF”.
Displaying Data 2
Bandwidth limit OFF
Figure 18 BW Limit Control OFF
Keysight 1000B Series Oscilloscopes User’s Guide 49
Page 50
2 Displaying Data

To specify the probe attenuation

20 MHz Bandwidth
Bandwidth ON Status
Figure 19 BW Limit Control ON
For correct measurements, you must match the oscilloscope’s probe attenuation factor settings with the attenuation factors of the probes being used.
The probe attenuation factor setting changes the vertical scaling of the oscilloscope so that the measurement results reflect the actual voltage levels at the probe tip.
1 If the channel’s menu is not currently displayed, press the channel key ([1], [2],
[3], or [4]).
2 In the Channel menu, press Probe. 3 Continue pressing the Probe softkey or turn the entry knob to select the
appropriate attenuation factor.

To use a digital filter

You can apply a digital filter to the sampled waveform data.
1 If the channel’s menu is not currently displayed, press the channel key ([1], [2],
[3], or [4]).
50 Keysight 1000B Series Oscilloscopes User’s Guide
Page 51
Displaying Data 2
2 In the Channel menu, press Digital Filter. 3 In the Filter menu, press Filter Type, and continue pressing the Filter Type softkey
or turn the entry knob to select between:
LPF (Low Pass Filter).
HPF (High Pass Filter).
BPF (Band Pass Filter).
BRF (Band Reject Filter).
4 Depending on the type of filter selected, press Upper Limit and/or Lower Limit,
and turn the entry knob to adjust the limit.
The horizontal scale control sets the maximum value for the upper and lower limits.
Digital filters are not available when:
• The horizontal scale is 20 ns/div or lower.
• The horizontal scale is 50 ms/div or higher.

To change the Volts/Div control sensitivity

When you need to adjust the amplitude/div setting in smaller increments, you can change the sensitivity ofthe vertical scale control.
1 If the channel’s menu is not currently displayed, press the channel key ([1], [2],
[3], or [4]).
2 In the Channel menu, press Volts/Div to toggle between:
Coarse The verical scale knob changes the amplitude/div setting in a
1-2-5 step sequence from 2 mV/div to 10 V/div (with “1X” probe attenuation).
Fine Also known as vernier, the vertical scale knob changes the
amplitude/div setting in small steps between the normal (coarse scale) settings.
Keysight 1000B Series Oscilloscopes User’s Guide 51
Page 52
2 Displaying Data

To invert a waveform

You can also toggle between coarse and fine settings by pushing the vertical scale knob (see “To adjust the vertical scale” on page 46).
You can invert a waveform with respect to the ground level.
1 If the channel’s menu is not currently displayed, press the channel key ([1], [2],
[3], or [4]).
2 In the Channel menu, press Invert to toggle between “ON” and “OFF”.
Figure 20 and Figure 21 show the changes before and after inversion.
Figure 20 Waveform Before Inversion
52 Keysight 1000B Series Oscilloscopes User’s Guide
Page 53
Figure 21 Waveform After Inversion
Displaying Data 2
Keysight 1000B Series Oscilloscopes User’s Guide 53
Page 54
2 Displaying Data

Using Math Function Waveforms

The math functions control allows the selection of the math functions:
•Add.
• Subtract.
• Multiply.
• FFT (Fast Fourier Transform).
The mathematical result can be measured using the grid and cursor controls.
The amplitude of the math waveform can be adjusted using a menu item selection in the Math menu and the entry knob. The adjustment range is in a 1-2-5 step from 0.1% to 1000%.
The math scale setting is shown at the bottom of the display.
Math Scale

Figure 22 Math Scale Setting Value

To add, subtract, or multiply waveforms

1 Press [Math]. 2 In the Math menu, press Operate.
54 Keysight 1000B Series Oscilloscopes User’s Guide
Page 55
3 Continue pressing the Operate softkey or turn the entry knob to select
NOTE
“A + B”, “A - B”, or “A x B”.
4 Press Source A, and continue pressing the softkey to select the desired input
channel.
5 Press Source B, and continue pressing the softkey to select the desired input
channel.
6 To invert the result of the addition, subtraction, or multiplication (with respect
to the reference level), select Invert to toggle between “ON” and “OFF”.

To display the frequency domain using FFT

The FFT math function mathematically converts a time-domain waveform into its frequency components. FFT waveforms are useful for finding the harmonic content and distortion in systems, for characterizing noise in DC power supplies, and for analyzing vibration.
To display a waveform’s FFT:
1 Press [Math]. 2 In the Math menu, press Operate. 3 Continue pressing the Operate softkey or turn the entry knob to select “FFT”. 4 In the FFT menu, press Source, and continue pressing the softkey to select the
desired input channel.
Displaying Data 2
The FFT of a waveform that has a DC component or offset can cause incorrect FFT waveform magnitude values. To minimize the DC component, choose AC Coupling on the source waveform.
To reduce random noise and aliasing components (in repetitive or single-shot waveforms), set the oscilloscope acquisition mode to averaging.
5 Press Window, and continue pressing the softkey or turn the entry knob to
select the desired window:
There are four FFT windows. Each window has trade-offs between frequency resolution and amplitude accuracy. What you want to measure and your source waveform characteristics help determine which window to use. Use the guidelines in Table 6 to select the best window.
Keysight 1000B Series Oscilloscopes User’s Guide 55
Page 56
2 Displaying Data
NOTE
Table 6 FFT Window Characteristics
Window Characteristics Best for measuring
Rectangle Best frequency resolution, worst
magnitude resolution. This is essentially the same as no window.
Hanning, Hamming
Blackman Best magnitude, worst frequency
Better frequency, poorer magnitude accuracy than Rectangular. Hamming has slightly better frequency resolution than Hanning.
resolution.
Transients or bursts, the waveform levels before and after the event are nearly equal. Equal-amplitude sine waves with fixed frequencies. Broadband random noise with a relatively slow varying spectrum.
Sine, periodic, and narrow-band random noise. Transients or bursts where the waveform levels before and after the events are significantly different.
Single frequency waveforms, to find higher order harmonics.
6 Press Display to toggle between a “Split” screen display and a “Full Screen”
display.
7 Press and turn the entry knob to adjust the vertical position of the
FFT waveform.
8 Press and turn the entry knob to adjust the vertical scale of the
FFT waveform.
9 Press Scale to toggle between “V
” and “dBV
RMS
RMS
” units.
To display FFT waveforms with a large dynamic range, use the dBVrms scale. The dBVrms scale displays component magnitudes using a log scale.
10 Use the horizontal position knob to adjust the frequency per division.
The frequency scale is displayed on the screen. Use this to display the frequencies associated with the peaks in the FFT waveform.
56 Keysight 1000B Series Oscilloscopes User’s Guide
Page 57
Figure 23 FFT Waveform
NOTE
NOTE
Displaying Data 2
FFT Resolution
The FFT resolution is the quotient of the sampling rate and the number of FFT points (fS/N). With a fixed number of FFT points (1024), the lower the sampling rate, the better the resolution.
Nyquist Frequency and Aliasing in the Frequency Domain
The Nyquist frequency is the highest frequency that any real-time digitizing oscilloscope can acquire without aliasing. This frequency is half of the sample rate. Frequencies above the Nyquist frequency will be under sampled, which causes aliasing. The Nyquist frequency is also called the folding frequency because aliased frequency components fold back from that frequency when viewing the frequency domain.
Keysight 1000B Series Oscilloscopes User’s Guide 57
Page 58
2 Displaying Data
NOTE

Using Reference Waveforms

You can save a reference waveform to an internal, nonvolatile memory location and then display it on the oscilloscope along with other captured waveforms.
You can also export/import reference waveforms to/from an external USB drive when it is connected to the front panel USB host port.
Reference waveforms are displayed (that is, turned on/off) just like other waveforms (see page 46).
The reference waveform function is not available in X-Y mode.

To save a reference waveform

1 Before saving a waveform as a reference, set the waveform’s scale and position
as desired.
These settings will become the reference waveform’s defaults.
2 Press [Ref]. 3 In the REF menu, press Source, and continue pressing the softkey or turn the
entry knob to select the waveform you want to save.
4 Press Location to choose “Internal”. 5 Press Save.

To export or import reference waveforms

To export or import from external storage (when a USB drive is connected to the front panel USB host port):
1 Press [Ref]. 2 If exporting a waveform, in the REF menu, press Source, and continue pressing
the softkey or turn the entry knob to select the waveform you want to export.
3 Presh Location to choose “External”. 4 Press Save or Import.
58 Keysight 1000B Series Oscilloscopes User’s Guide
Page 59
5 Use the disk manager dialog to navigate to the folder where you want to export
the file or to select the file you want to import (see “To navigate the d irectory
hierarchy” on page 117).
6 In the Save or Import menu:
To export the waveform, press New File, enter the filename (see “To edit
folder/file names” on page 118), and press Save.
To load the selected waveform (.wfm file), press Import.

To return the reference waveform to its default scale

1 Press [Ref]. 2 In the REF menu, press Reset.
The scale and position of the waveform as originally saved are restored.
Displaying Data 2
Keysight 1000B Series Oscilloscopes User’s Guide 59
Page 60
2 Displaying Data

Changing the Display Settings

Figure 24 [Display] Key

To display waveforms as vectors or dots

1 Press [Display].
60 Keysight 1000B Series Oscilloscopes User’s Guide
Page 61
2 In the Display menu, press Type to toggle the waveform display between:
Vectors The oscilloscope connects the sample points by using digital
interpolation. Digital interpolation maintains linearity by using a sin(x)/x
digital filter. The digital interpolation is suitable for real time sampling and is most effective at 20 ns or faster horizontal scale settings.
Dots The sample points are displayed.

To clear the display

1 Press [Display]. 2 In the Display menu, press Clear.

To set waveform persistence

1 Press [Display]. 2 In the Display menu, press Persist to toggle the waveform display between:
Displaying Data 2
Infinite Sample points remain displayed until the display is cleared or
persistence is set to “OFF”.
OFF

To adjust waveform intensity

1 Press [Display]. 2 In the Display menu, press Intensity and turn the entry knob to adjust the
waveform intensity.

To change the grid

1 Press [Display]. 2 In the Display menu, press Grid, and continue pressing the softkey or turn the
entry knob to select between:
Keysight 1000B Series Oscilloscopes User’s Guide 61
Page 62
2 Displaying Data

To adjust the grid brightness

To invert screen colors

Display grid and coordinates on the axes.
Displays coordinates on the axes.
Turns off the grid and coordinates.
1 Press [Display]. 2 In the Display menu, press GridBright and turn the entry knob to adjust the
grid brightness.
1 Press [Display] > Preference. 2 In the Display menu, press Screen to toggle the screen between “Normal” or
“Inverted” colors.
Inverted screen colors are sometimes useful when printing or saving screens.

To change the menu display time

The menu display time is how long menus remain on the screen after a front panel key or softkey has been pressed.
1 Press [Display]. 2 In the Display menu, press Menu Display, and continue pressing the softkey or
turn the entry knob to select “1 s”, “2 s”, “5 s”, “10 s”, “20 s”, or “Infinite” menu display time.
62 Keysight 1000B Series Oscilloscopes User’s Guide
Page 63
Keysight 1000B Series Oscilloscopes User’s Guide

3 Capturing Data

Overview of Sampling 64 Choosing the Acquisition Mode 73 Choosing the Acquisition Mode 73 Recording/Playing-back Waveforms 78 Adjusting the Trigger Level 82 Choosing the Trigger Mode 84 Setting Other Trigger Parameters 90 Using the External Trigger Input 94
This chapter describes sampling and acquisition modes and how to set up triggers.
63
Page 64
3 Capturing Data

Overview of Sampling

To understand the oscilloscope’s sampling and acquisition modes, it is helpful to understand sampling theory, aliasing, oscilloscope bandwidth and sample rate, oscilloscope rise time, oscilloscope bandwidth required, and how memory depth affects sample rate.

Sampling Theory

The Nyquist sampling theorem states that for a limited bandwidth (band-limited) signal with maximum frequency f must be greater than twice the maximum frequency f signal be uniquely reconstructed without aliasing.
f
= fS/2 = Nyquist frequency (fN) = folding frequency
MAX

Aliasing

, the equally spaced sampling frequency fS
MAX
, in order to have the
MAX
Aliasing occurs when signals are under-sampled (fS < 2f distortion caused by low frequencies falsely reconstructed from an insufficient number of sample points.
). Aliasing is the signal
MAX
64 Keysight 1000B Series Oscilloscopes User’s Guide
Page 65
Figure 25 Alaising
Capturing Data 3

Oscilloscope Bandwidth and Sample Rate

An oscilloscope’s bandwidth is typically described as the lowest frequency at which input signal sine waves are attenuated by 3 dB (-30% amplitude error).
At the oscilloscope bandwidth, sampling theory says the required sample rate is f
=2fBW. However, the theory assumes there are no frequency components above
S
f
(fBW in this case) and it requires a system with an ideal brick-wall frequency
MAX
response.
Keysight 1000B Series Oscilloscopes User’s Guide 65
Page 66
3 Capturing Data
f
S
f
N
-3dB
Attenuation
Frequency
0dB
Figure 26 Theoretical Brick-Wall Frequency Response
66 Keysight 1000B Series Oscilloscopes User’s Guide
However, digital signals have frequency components above the fundamental frequency (square waves are made up of sine waves at the fundamental frequency and an infinite number of odd harmonics), and typically, for 1 Ghz bandwidths and below, oscilloscopes have a Gaussian frequency response.
Page 67
Figure 27 Sample Rate and Oscilloscope Bandwidth
f
S
f
N
fS/4
-3dB
Attenuation
Aliased frequency
components
Frequency
Limiting oscilloscope bandwidth (fBW) to 1/4 the sample rate (fS/4) reduces frequency components above the Nyquist frequency (f
N
).
0dB
Capturing Data 3
So, in practice, an oscilloscope’s sample rate should be four or more times its bandwidth: f components have a greater amount of attenuation.
See Also Evaluating Oscilloscope Sample Rates vs. Sampling Fidelity: How to Make the
Most Accurate Digital Measurements, Keysight Application Note 1587
(http://literature.cdn.keysight.com/litweb/pdf/5989-5732EN.pdf)
=4fBW. This way, there is less aliasing, and aliased frequency
S

Oscilloscope Rise Time

Closely related to an oscilloscope’s bandwidth specification is its rise time specification. Oscilloscopes with a Gaussian-type frequency response have an
Keysight 1000B Series Oscilloscopes User’s Guide 67
approximate rise time of 0.35/f
An oscilloscope’s rise time is not the fastest edge speed that the oscilloscope can accurately measure. It is the fastest edge speed the oscilloscope can possibly produce.
based on a 10% to 90% criterion.
BW
Page 68
3 Capturing Data

Oscilloscope Bandwidth Required

The oscilloscope bandwidth required to accurately measure a signal is primarily determined by the signal’s rise time, not the signal's frequency. You can use these steps to calculate the oscilloscope bandwidth required:
1 Determine the fastest edge speeds.
You can usually obtain rise time information from published specifications for devices used in your designs.
2 Compute the maximum “practical” frequency component.
From Dr. Howard W. Johnson’s book, High-Speed Digital Design – A Handbook of Black Magic, all fast edges have an infinite spectrum of frequency components. However, there is an inflection (or “knee”) in the frequency spectrum of fast edges where frequency components higher than f
knee
are
insignificant in determining the shape of the signal.
f
= 0.5 / signal rise time (based on 10% - 90% thresholds)
knee
f
= 0.4 / signal rise time (based on 20% - 80% thresholds)
knee
3 Use a multiplication factor for the required accuracy to determine the
oscilloscope bandwidth required.
Required accuracy
20% f
10% fBW = 1.3 x f
3% fBW = 1.9 x f
Oscilloscope bandwidth required
= 1.0 x f
BW
knee
knee
knee
See Also Choosing an Oscilloscope with the Right Bandwidth for your Application, Keysight
Application Note 1588 (http://literature.cdn.keysight.com/litweb/pdf/5989-5733EN.pdf)
68 Keysight 1000B Series Oscilloscopes User’s Guide
Page 69

Memory Depth and Sample Rate

The number of points of oscilloscope memory is fixed (except when divided between channel pairs), and there is a maximum sample rate associated with oscilloscope’s analog-to-digital converter; however, the actual sample rate is determined by the time of the acquisition (which is set according to the oscilloscope’s horizontal time/div scale).
sample rate = number of samples / time of acquisition
For example, when storing 10 us of data in 10,000 points of memory, the actual sample rate is 1 GSa/s.
Likewise, when storing 1 s of data in 10,000 points of memory, the actual sample rate is 10 kSa/s.
The actual sample rate, is displayed in the Horizontal menu (see “To view the
sample rate” on page 44).
The oscilloscope achieves the actual sample rate by throwing away (decimating) unneeded samples.
Capturing Data 3
Keysight 1000B Series Oscilloscopes User’s Guide 69
Page 70
3 Capturing Data
1
1
1
1
N
= Waveform Sample is From
Trigger Time

Choosing the Sampling Mode

The osilloscope can operate in real-time or equivalent-time sampling modes.
You can choose the oscilloscope’s sampling mode in the Acquire menu (accessed by pressing the [Acquire] front panel button).

To select the real-time sampling mode

In the real-time sampling mode, single waveforms are sampled at uniformly spaced intervals. See Figure 28.
70 Keysight 1000B Series Oscilloscopes User’s Guide
Figure 28 Real-Time Sampling Mode
Use the real-time sampling mode when capturing (non-repetitive) single-shot or pulse waveforms.
The 1000B Series oscilloscopes provide real-time sampling rates up to 500 MSa/s (when two channels are on) or 1 GSa/s (when one channel is on).
To select the Real-Time sampling mode:
1 Press [Acquire]. 2 In the Acquire menu, select Sampling to choose the “Real Time” sampling mode.
In the real-time sampling mode, when the horizontal scale is set to 20 ns or faster, the oscilloscope uses sine(x)/x interpolation to expand the horizontal time base.
Page 71

To select the equivalent-time sampling mode

1
1
1
1
2
3
4
5
4
4
4
3
3
3
5
5
5
2
2
2
N
= Waveform Sample is From
Trigger Time
D1
D2
D3
D4
In the equivalent-time sampling mode (also known as repetitive sampling), multiple waveforms are sampled using randomly differing delays from the trigger to yield higher effective sampling rates.
Capturing Data 3
Keysight 1000B Series Oscilloscopes User’s Guide 71
Figure 29 Equivalent-Time (Repetitive) Sampling Mode
In the equivalent-time sampling mode, the effective sample rate can be greater because the time between samples in the acquisition is shorter.
Equivalent-time sampling mode requires a repetitive waveform with a stable trigger.
Use the equivalent-time sampling mode to capture repetitive high-frequency signals at greater sample rates than are available in the real-time sampling mode.
Do not use the equivalent-time mode for single-shot events or pulse waveforms.
The benefits of the equivalent-time sampling mode are negligible when sample rates are the same as available in the real-time sampling mode.
In the 1000B Series oscilloscopes, the equivalent time sampling mode can achieve up to 40 ps of horizontal resolution (equivalent to 25 GSa/s).
To select the Equivalent-Time sampling mode:
1 Press [Acquire].
Page 72
3 Capturing Data
2 In the Acquire menu, select Sampling to choose the “Equ-Time” sampling mode.
72 Keysight 1000B Series Oscilloscopes User’s Guide
Page 73

Choosing the Acquisition Mode

The osilloscope can operate in normal, average, or peak detect acquisition modes.
You can choose the oscilloscope’s acquisition mode in the Acquire menu (accessed by pressing the [Acquire] front panel key).
Capturing Data 3

Figure 30 [Acquire] Key

Keysight 1000B Series Oscilloscopes User’s Guide 73
Page 74
3 Capturing Data

To select the Normal acquisition mode

To select the Average acquisition mode

In the Normal acquisition mode, acquisitions are made and displayed one after the other.
To select the Normal acquisition mode:
1 Press [Acquire]. 2 In the Acquire menu, press Acquisition. 3 Continue pressing the Acquisition softkey or turn the entry knob to select
“Normal”.
In the Average acquisition mode, acquisitions are made, and the running average over the specified number of acquisitions is displayed.
Use the Average acquisition mode to remove random noise from the waveform and to improve measurement accuracy.
Figure 31 Noisy Waveform Without Averaging
74 Keysight 1000B Series Oscilloscopes User’s Guide
Page 75
Figure 32 Noisy Waveform With Averaging
Capturing Data 3
The Average acquisition mode decreases the screen refresh rate.
To select the Average acquisition mode:
1 Press [Acquire]. 2 In the Acquire menu, press Acquisition. 3 Continue pressing the Acquisition softkey or turn the entry knob to select
“Averag e”.
4 Press Averages and turn the entry knob to select the desired number (2, 4, 8,
16, 32, 64, 128, or 256).

To select the Peak Detect acquisition mode

In Normal or Average acquisition modes, at longer horizontal time/div settings, the oscilloscope’s analog-to-digital converter samples at a rate that yields more samples than can be stored in a limited amount of oscilloscope memory. Consequently, samples are thrown away (decimated), and you can miss narrow excursions on a signal.
Keysight 1000B Series Oscilloscopes User’s Guide 75
Page 76
3 Capturing Data
However, in the Peak Detect acquisition mode, acquisitions are made at the fastest sample rate, and the minimum and maximum values for the period associated with the actual sample rate are stored. This way, you can capture narrow excursions on a signal at longer horizontal time/div settings.
Figure 33 Peak Detect Waveform
Because minimum and maximum values for a sample period are stored, you can use the Peak Detect acquisition mode to avoid waveform aliasing.
To select the Peak Detect acquisition mode:
1 Press [Acquire]. 2 In the Acquire menu, press Acquisition. 3 Continue pressing the Acquisition softkey or turn the entry knob to select
“Peak Detect”.
76 Keysight 1000B Series Oscilloscopes User’s Guide
Page 77

To turn OFF/ON sine(x)/x interpolation

When sample points are displayed as vectors (instead of dots) and sine(x)/x interpolation is on, curved lines are drawn between sample points. When sine(x)/x interpolation is off, straight lines are drawn.
The effects of sine(x)/x interpolation are only noticeable when the horizontal scale is set to 20 ns or faster.
1 Press [Acquire]. 2 In the Acquire menu, press Sinx/x to turn sine(x)/x interpolation “OFF” or “ON”.
Capturing Data 3
Keysight 1000B Series Oscilloscopes User’s Guide 77
Page 78
3 Capturing Data

Recording/Playing-back Waveforms

You can record waveforms from input channels or from the mask test output, with a maximum acquisition depth of 1000 frames.
The ability to record mask test output is especially useful for capturing abnormal waveforms over a long period of time.

To record waveforms

To record waveforms:
1 Press [Acquire]. 2 In the Acquire menu, press Sequence. 3 In the Sequence menu, press Mode. 4 Continue pressing the Mode softkey or turn the entry knob to select Record.
To select the source channel for record ing
1 In the Sequence menu ([Acquire] > Sequence > Mode=Record), press Source. 2 Continue pressing the Source softkey or turn the entry knob to select the
desired input channel or the mask test output.
To specify the mask test output, see “To set the mask test output condition” on page 130.
To select the number of frames to record
1 In the Sequence menu ([Acquire] > Sequence > Mode=Record), press End Frame. 2 Turn the entry knob to select a number from 1 to 1000.
To start/stop recording
1 In the Sequence menu ([Acquire] > Sequence > Mode=Record), press Operate to
start or stop recording.
Appears on the menu when not recording; press Operate to start recording.
Appears on the menu when recording; press Operate to stop recording.
78 Keysight 1000B Series Oscilloscopes User’s Guide
Page 79
To select the interval between recorded frames
1 In the Sequence menu ([Acquire] > Sequence > Mode=Record), press Interval. 2 Turn the entry knob to select an interval from 1 ms to 1000 s.

To play-back waveforms

To play-back waveforms:
1 Press [Acquire]. 2 In the Acquire menu, press Sequence. 3 In the Sequence menu, press Mode. 4 Continue pressing the Mode softkey or turn the entry knob to select
“Play back”.
To play-back/stop the record ing
1 In the Sequence menu ([Acquire] > Sequence > Mode=Play back), press Operate to
play-back or stop the recording.
Capturing Data 3
Appears on the menu when not playing-back; press Operate to start playing-back the recording.
Appears on the menu when playing-back; press Operate to stop playing-back.
To select continuous or one-time play-back
1 In the Sequence menu ([Acquire] > Sequence > Mode=Play back), press Play Mode
to toggle between:
Continuous play-back.
One-time play-back.
To select the interval between played-back frames
1 In the Sequence menu ([Acquire] > Sequence > Mode=Play back), press Interval. 2 Turn the entry knob to select an interval from 1 ms to 20 s.
Keysight 1000B Series Oscilloscopes User’s Guide 79
Page 80
3 Capturing Data

To store recorded waveforms

To select the start frame
1 In the Sequence menu ([Acquire] > Sequence > Mode=Play back), press Start Frame. 2 Turn the entry knob to select a number from l to 1000.
To select the current frame
1 In the Sequence menu ([Acquire] > Sequence > Mode=Play back), press
Current Frame.
2 Turn the entry knob to select a number from l to 1000.
To select the end frame
1 In the Sequence menu ([Acquire] > Sequence > Mode=Play back), press End Frame. 2 Turn the entry knob to select a number from l to 1000.
To store recorded waveforms:
1 Press [Acquire]. 2 In the Acquire menu, press Sequence. 3 In the Sequence menu, press Mode. 4 Continue pressing the Mode softkey or turn the entry knob to select
“Storage”.
To select the start frame
1 In the Sequence menu ([Acquire] > Sequence > Mode=Storage), press Start Frame. 2 Turn the entry knob to select a number from l to 1000.
To select the end frame
1 In the Sequence menu ([Acquire] > Sequence > Mode=Storage), press End Frame. 2 Turn the entry knob to select a number from l to 1000.
To select internal/external recording storage location
1 In the Sequence menu ([Acquire] > Sequence > Mode=Storage), press Location to
toggle between Internal and External.
80 Keysight 1000B Series Oscilloscopes User’s Guide
Page 81
Capturing Data 3
Internal Recordings are saved and loaded from oscilloscope internal
memory.
External Recordings are saved, loaded, exported, and imported from an
external USB drive.
To save a recording
1 In the Sequence menu ([Acquire] > Sequence > Mode=Storage), press Save. 2 If the External location has been selected, use the Disk Manager to name and
save the waveform recording file. See “Using the Disk Manager” on page 116.
To load a recording
1 In the Sequence menu ([Acquire] > Sequence > Mode=Storage), press Load. 2 If the External location has been selected, use the Disk Manager to select and
load the waveform recording file. See “Using the Disk Manager” on page 116.
To import/export recordings
1 Because you can only export and import waveform recordings from an external
drive, select the External location. See “To select internal/external recording
storage location” on page 80.
2 In the Sequence menu ([Acquire] > Sequence > Mode=Storage), press Imp./Exp.. 3 Use the Disk Manager to select the file and import or export the waveform
recording. See “Using the Disk Manager” on page 116.
Keysight 1000B Series Oscilloscopes User’s Guide 81
Page 82
3 Capturing Data

Adjusting the Trigger Level

Figure 34 Trigger Controls

To adjust the trigger level

• Turn the trigger [Level] knob.
Trigger level knob
Two things happen:
The trigger level value is displayed at the lower left-hand corner of the
screen.
A line is displayed showing the location of the trigger level with respect to
the waveform (except when using AC coupling or LF reject coupling modes).
• Push the trigger [Level] knob to set the level at 50% of the signal’s vertical amplitude.

To force a trigger

To make an acquisition even if no valid trigger has been found: 1 Press [Force Trigger].
Forcing a trigger is useful, for example, when you want to display the DC voltage of a level signal.
The [Force Trigger] key has no effect if the acquisition is already stopped.
82 Keysight 1000B Series Oscilloscopes User’s Guide
Page 83
Capturing Data 3
When the oscilloscope’s front panel is locked by a remote program (shown by a red “Rmt” on the upper-right part of the display), pressing the [Force Trigger] key returns the front panel to Local control.
Keysight 1000B Series Oscilloscopes User’s Guide 83
Page 84
3 Capturing Data

Choosing the Trigger Mode

The trigger determines when captured data should be stored and displayed.
When a trigger is set up properly, it can convert unstable displays or blank screens into meaningful waveforms.
When the oscilloscope starts to acquire a waveform, it collects enough data so that it can draw the waveform to the left of the trigger point. The oscilloscope continues to acquire data while waiting for the trigger condition to occur. After it detects a trigger, the oscilloscope continues to acquire enough data so that it can draw the waveform to the right of the trigger point.
The oscilloscope provides these trigger modes:
Edge Can be used with analog and digital circuits. An edge trigger
occurs when the trigger input passes through a specified voltage level with the specified slope.
Pulse Is used to find pulses with certain widths.
Video Is used to trigger on fields or lines for standard video
waveforms.
Alternate Is used to trigger on non-synchronized signals.

To set up edge triggers

1 Press [Menu]. 2 In the Trigger menu, press Mode. 3 Continue pressing the Mode softkey or turn the entry knob to select “Edge”. 4 Then, either push the entry knob or press Mode again. 5 Press Source and continue pressing the softkey or turn the entry knob to
select the waveform to trigger on:
CH1 - CH2 The oscilloscope input channel.
EXT The external trigger input.
AC Line The AC power line.
84 Keysight 1000B Series Oscilloscopes User’s Guide
Page 85
6 Press Slope and continue pressing the softkey or turn the entry knob to
select the edge to trigger on:
Rising edge.
Falling edge.
Both rising and falling edges.

To set up pulse width triggers

A pulse width trigger occurs when a pulse that matches the pulse definition is found in a waveform.
The width setting can be adjusted from 20 ns to 10 s.
1 Press [Menu]. 2 In the Trigger menu, press Mode. 3 Continue pressing the Mode softkey or turn the entry knob to select “Pulse”. 4 Then, either push the entry knob or press Mode again. 5 Press Source and continue pressing the softkey or turn the entry knob to
select the waveform to trigger on:
Capturing Data 3
CH1 - CH2 The oscilloscope input channel.
EXT The external trigger input.
6 Press When and continue pressing the softkey or turn the entry knob to
select the type of pulse to trigger on:
Positive pulse greater than the width setting.
Positive pulse less than the width setting.
Negative pulse greater than the width setting.
Negative pulse less than the width setting.
Keysight 1000B Series Oscilloscopes User’s Guide 85
Page 86
3 Capturing Data
NOTE

To set up video triggers

7 Press Setting and turn the entry knob to adjust the width setting.
Video triggering is used to trigger on fields or lines of NTSC, PAL, or SECAM standard video waveforms.
When the video trigger mode is selected, the trigger coupling is set to AC.
1 Press [Menu]. 2 In the Trigger menu, press Mode. 3 Continue pressing the Mode softkey or turn the entry knob to select “Video”. 4 Then, either push the entry knob or press Mode again. 5 Press Polarity to toggle between:
Normal polarity — trigger on the negative edge of the sync pulse.
Inverted polarity — trigger on the positive edge of the sync pulse.
Normal Polarity Sync triggers always occur on negative-going horizontal sync pulses. If the video waveform has positive-going horizontal sync pulses, use the Inverted Polarity selection.
6 Press Sync and continue pressing the softkey or turn the entry knob to select
what to trigger on:
All Lines Trigger on all lines.
Line Num Trigger on a selected line.
If you select “Line Num”, press the following Line Num menu item and turn the entry knob to select the line number.
Odd Field Trigger on an odd field.
Even Field Trigger on an even field.
7 Press Standard to toggle between:
86 Keysight 1000B Series Oscilloscopes User’s Guide
Page 87
NTSC Trigger on an NTSC video waveform.
PAL/ SECAM Trigger on a PAL or SECAM video waveform.
Capturing Data 3
Figure 35 Line Synchronization
Keysight 1000B Series Oscilloscopes User’s Guide 87
Page 88
3 Capturing Data
Figure 36 Field Synchronization
88 Keysight 1000B Series Oscilloscopes User’s Guide

To set up alternate triggers

The Alternate trigger mode splits the display horizontally and lets you trigger on two, non-synchronized signals.
1 Press [Menu]. 2 In the Trigger menu, press Mode. 3 Continue pressing the Mode softkey or turn the entry knob to select
“Alterna te”.
4 Then, either push the entry knob or press Mode again. 5 Press Select to select the channel to set up triggering on, either “CH1” or
“CH2”.
At this point, the remaining items in the Trigger menu let you set up independent triggers for the selected channel.
For each source, you can set up edge, pulse width, or video triggering. You can also specify other trigger setup options, except trigger sweep.
Page 89
Figure 37 Alternate Triggers
Capturing Data 3
Keysight 1000B Series Oscilloscopes User’s Guide 89
Page 90
3 Capturing Data

Setting Other Trigger Parameters

These are trigger system parameters that apply in all trigger modes.

To set the trigger sweep

Trigger sweep specifies whether acquisitions occur without a triggern or only with a trigger.
1 Press [Menu]. 2 In the Trigger menu, press Sweep. 3 Continue pressing the Sweep softkey or turn the entry knob to select one of
these trigger sweep settings:
Auto Acquire waveform even when no trigger occurs.
Normal Acquire waveform when trigger occurs.
90 Keysight 1000B Series Oscilloscopes User’s Guide

To set the trigger coupling

Trigger coupling is used to filter low frequency signal components or DC offsets from the trigger path when they interfere with achieving stable triggers.
Trigger coupling is similar to channel coupling (see page 47), but it only affects the triggering system and does not change how the signal is displayed.
To set the trigger coupling:
1 Press [Menu]. 2 In the Trigger menu, press Set Up. 3 In the Set Up menu, press Coupling.
Page 91
4 Continue pressing the the Coupling softkey or turn the entry knob to select
one of these trigger coupling settings:
DC Sets the trigger coupling to DC.
AC Sets the trigger coupling to AC — use for waveforms greater
than 50 Hz.
LF Reject Sets the trigger coupling to low frequency reject (10 kHz
cutoff).

To set the trigger high-frequency reject coupling

Trigger high-frequency reject coupling (100 kHz cutoff) is used to filter high frequency signal components from the trigger path when they interfere with achieving stable triggers.
To set the trigger high-frequency reject coupling:
1 Press [Menu]. 2 In the Trigger menu, press Set Up. 3 In the Set Up menu, press HF Reject to toggle between “ON” and “OFF”.
Capturing Data 3
Keysight 1000B Series Oscilloscopes User’s Guide 91
Page 92
3 Capturing Data

To change the trigger sensitivity

Trigger sensitivity specifies the vertical change that must occur in order for a trigger to be recognized. In the 1000B Series oscilloscopes, you can adjust the trigger sensitivity.
For example, to reduce the influence of noise, you can lower the trigger sensitivity (by increasing the vertical change required to trigger).
To change the trigger sensitivity:
1 Press [Menu]. 2 In the Trigger menu, press Set Up. 3 In the Set Up menu, press Sensitivity and turn the entry knob to adjust the
sensitivity setting.
The trigger sensitivity can be adjusted from 0.1 div to 1 div.
92 Keysight 1000B Series Oscilloscopes User’s Guide
Page 93

To specify a trigger holdoff

Trigger holdoff can be used to stabilize a waveform. The holdoff time is the oscilloscope's waiting period before starting a new trigger. The oscilloscope will not trigger until the holdoff time has expired.
Capturing Data 3
Figure 38 Trigger Holdoff
To specify a trigger holdoff:
1 Press [Menu]. 2 In the Trigger menu, press Set Up. 3 In the Set Up menu, press Holdoff and turn the entry knob to adjust the
holdoff setting.
To reset the trigger holdoff
1 In the Set Up menu, select the Holdoff Reset menu item to return the trigger
holdoff setting to the 100 ns minimum value.
Keysight 1000B Series Oscilloscopes User’s Guide 93
Page 94
3 Capturing Data

Using the External Trigger Input

You can trigger on external inputs by selecting “EXT” as the trigger source in all trigger modes except Alternate.
94 Keysight 1000B Series Oscilloscopes User’s Guide
Page 95
Keysight 1000B Series Oscilloscopes User’s Guide

4 Making Measurements

Displaying Automatic Measurements 96 Voltage Measurements 98 Time Measurements 101 Counter (Frequency) 106 Making Cursor Measurements 107
This chapter shows how to make automatic voltage measurements, automatic time measurements, and cursor measurements.
95
Page 96
4 Making Measurements

Displaying Automatic Measurements

You can use the [Meas] key to display automatic measurements. The oscilloscope has 22 automatic measurements and a hardware frequency counter (see “Vol tage
Measurements” on page 98 and “Time Measurements” on page 101).

Figure 39 [Meas] Key

96 Keysight 1000B Series Oscilloscopes User’s Guide
Page 97
Making Measurements 4

To display an automatic measurement

1 Press [Meas]. 2 In the Measure menu, press Source to select the input channel on which to
make the automatic measurement.
3 Press Voltage (for voltage measurements) or Time (for time measurements) and
turn the entry knob to select the desired measurement.
4 Then, either push the entry knob or press Voltage or Time again to add the
measurement to the bottom of the display.
If the measurement result is displayed as "*****", the measurement cannot be performed with the current oscilloscope settings.
A maximum of three measurements can be displayed at the bottom of the display. When three measurements are displayed and you add a new one, the measurements shift to the left, pushing the first measurement result off screen.
See Also “To display cursors for automatic measurements” on page 110.

To clear automatic measurements from the display

1 Press [Meas]. 2 In the Measure menu, press Clear to clear all automatic measurements from the
display.

To display or hide all automatic measurements

1 Press [Meas]. 2 In the Measure menu, press Display All to toggle the display of all automatic
measurements “ON” or “OFF”.
Keysight 1000B Series Oscilloscopes User’s Guide 97
Page 98
4 Making Measurements
Vbase
Vtop
Overshoot
Vmax
Overshoot
Preshoot
Vpp
Vmin
Vamp

Voltage Measurements

There are 10 automatic voltage measurements:
•Vmax (Maximum Voltage).
• Vmin (Minimum Voltage).
• Vpp (Peak-to-Peak Voltage).
• Vtop (Top Voltage).
• Vbase (Base Voltage).
• Vamp (Amplitude Voltage = Vtop - Vbase).
•Vavg (Average Voltage).
• Vrms (Root-Mean-Square Voltage).
• Overshoot.
• Preshoot.

Figure 40 Voltage Measurement Points

Vmax (Maximum Voltage)

98 Keysight 1000B Series Oscilloscopes User’s Guide
The maximum amplitude. The most positive peak voltage measured over the entire waveform. See Figure 40 on page 98.
Page 99

Vmin (Minimum Voltage)

RMS
x
i
2
i 1=
n
n
--------------- -=
The minimum amplitude. The most negative peak voltage measured over the entire waveform. See Figure 40 on page 98.

Vpp (Peak-to-Peak Voltage)

Peak-to-peak voltage. See Figure 40 on page 98.

Vtop (Top Voltage)

Voltage of the waveform's flat top, useful for square and pulse waveforms. See
Figure 40 on page 98.

Vbase (Base Voltage)

Voltage of the waveform's flat base, useful for square and pulse waveforms. See
Figure 40 on page 98.
Making Measurements 4

Vamp (Amplitude Voltage = Vtop - Vbase)

Voltage between Vtop and Vbase of a waveform. See Figure 40 on page 98.

Vavg (Average Voltage)

The arithmetic mean over the entire waveform.

Vrms (Root-Mean-Square Voltage)

The true root-mean-square voltage over the entire waveform.
Where:
x
= value at ith point.
i
n = number of points.
Keysight 1000B Series Oscilloscopes User’s Guide 99
Page 100
4 Making Measurements

Overshoot

Preshoot

Defined as (Vmax-Vtop)/Vamp, useful for square and pulse waveforms. See
Figure 40 on page 98.
Defined as (Vmin-Vbase)/Vamp, useful for square and pulse waveforms. See
Figure 40 on page 98.
100 Keysight 1000B Series Oscilloscopes User’s Guide
Loading...