Keysight 1000B Series User Manual

Keysight 1000B Series Oscilloscopes
User’s Guide
Notices
CAUTION
WARNING
© Keysight Technologies, Inc. 2008-2009, 2012
Manual Part Number
54139-97013
Edition
Second Edition, May 2012
Printed in Malaysia
Keysight Technologies, Inc. 1900 Garden of the Gods Road Colorado Springs, CO 80907 USA
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Keysight disclaims all war­ranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fit­ness for a particular purpose. Keysight shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Keysight and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall con­trol.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor­dance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a U.S. Government prime contract or subcon­tract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or d isclosure of Software is subject to Keysight Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Govern­ment users will receive no greater than Lim­ited Rights as defined in FAR 52.227-14
(June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any tech­nical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a haz­ard. It calls attention to an operat­ing procedure, practice, or the like that, if not correctly performed or adhered to, could resul t in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
See also Appendix A, “Safety Notices,” starting on page 143.
2 Keysight 1000B Series Oscilloscopes User’s Guide
Keysight 1000B Series Oscilloscopes—At a Glance
The Keysight 1000B Series oscilloscopes are low-cost portable digital storage oscilloscopes (DSOs) that deliver these powerful features:
Channel counts, input bandwidths, sample rates, and memory depths shown in the following table:
Table 1 Keysight 1000B Series Oscilloscope Models
Model Channels Input Bandwid th Sample Rate
(2 channels on-1 channel on)
DSO1052B 2 50 MHz 500 MSa/s-1 GSa/s 8-16 kpts
DSO1072B 2 70 MHz 500 MSa/s-1 GSa/s 8-16 kpts
DSO1102B 2 100 MHz 500 MSa/s-1 GSa/s 8-16 kpts
DSO1152B 2 150 MHz 500 MSa/s-1 GSa/s 8-16 kpts
Bright 5.7 inch QVGA (320 x 240) TFT color LED display and small footprint (to save bench space).
Up to 400 wfms/s refresh rate.
Automatic voltage and time measurements (22) and cursor measurements.
Powerful triggering (edge, pulse width, video, and alternate modes) with adjustable sensitivity (to filter noise and avoid false triggers).
Math function waveforms: add, subtract, multiply, FFT.
USB ports (host and device) for easy printing, saving, and sharing of waveforms, setups, screen BMP files, and CSV data files.
Internal storage for 10 waveforms and 10 setups.
Special digital filter and waveform recorder.
Built-in 6-digit hardware frequency counter.
Multi-language (11) user interface menus and built-in help.
Memory (2 channels on-1 channel on)
Keysight 1000B Series Oscilloscopes User’s Guide 3
In This Book
1 Getting Started
2 Displaying Data
3 Capturing Data
4 Making Measurements
This guide shows how to use the Keysight 1000B Series oscilloscopes.
Describes the basic steps to take when first using the oscilloscope.
Describes how to use the horizontal and vertical controls, channel settings, math waveforms, reference waveforms, and display settings.
Describes acquisition modes and how to set up triggers.
Describes voltage, time, and cursor measurements.
5 Saving, Recalling, and Printing Data
Describes how to save, recall, and print data.
6 Oscilloscope Utility Settings
Describes other oscilloscope settings found in the Utility menu.
7 Reference
Contains reference information for the 1000B Series oscilloscopes.
4 Keysight 1000B Series Oscilloscopes User’s Guide

Contents

Keysight 1000B Series Oscilloscopes—At a Glance 3
In This Book 4
Figures 13
Tab les 15
1 Getting Started 17
Step 1. Inspect the package contents 18
Step 2. Turn on the oscilloscope 19
Step 3. Load the default oscilloscope setup 22
Step 4. Input a waveform 23
23
Step 5. Use Auto Scale 24
Step 6. Compensate probes 26
Low Frequency Compensation 26 High Frequency Compensation 26
Step 7. Become familiar with the Front Panel Controls 28
Front Panel Overlays for Different Languages 29
Step 8. Become familiar with the oscilloscope display 30
Using the Oscilloscope Softkey Menus 31
Step 9. Use the Run Control keys 33
Step 10. Access the built-in help 34
Keysight 1000B Series Oscilloscopes User’s Guide 5
Contents
Securing the Oscilloscope 35
2 Displaying Data 37
Using the Horizontal Controls 38
To adjust the horizontal scale 39 To adjust the horizontal position 40 To display the zoomed time base 41 To change the horizontal time base (Y-T, X-Y, or Roll) 42 To view the sample rate 44
Using the Vertical Controls 45
To turn waveforms on or off (channel, math, or reference) 46 To adjust the vertical scale 46 To adjust the vertical position 46 To specify channel coupling 47 To specify a bandwidth limit 49 To specify the probe attenuation 50 To use a digital filter 50 To change the Volts/Div control sensitivity 51 To invert a waveform 52
Using Math Function Waveforms 54
To add, subtract, or multiply waveforms 54 To display the frequency domain using FFT 55
Using Reference Waveforms 58
To save a reference waveform 58 To export or import reference waveforms 58 To return the reference waveform to its default scale 59
Changing the Display Settings 60
To display waveforms as vectors or dots 60 To clear the display 61 To set waveform persistence 61 To adjust waveform intensity 61
6 Keysight 1000B Series Oscilloscopes User’s Guide
To change the grid 61 To adjust the grid brightness 62 To invert screen colors 62 To change the menu display time 62
3 Capturing Data 63
Overview of Sampling 64
Sampling Theory 64 Aliasing 64 Oscilloscope Bandwidth and Sample Rate 65 Oscilloscope Rise Time 67 Oscilloscope Bandwidth Required 68 Memory Depth and Sample Rate 69
Choosing the Sampling Mode 70
To select the real-time sampling mode 70 To select the equivalent-time sampling mode 71
Choosing the Acquisition Mode 73
To select the Normal acquisition mode 74 To select the Average acquisition mode 74 To select the Peak Detect acquisition mode 75
To turn OFF/ON sine(x)/x interpolation 77
Contents
Recording/Playing-back Waveforms 78
To record waveforms 78 To play-back waveforms 79 To store recorded waveforms 80
Adjusting the Trigger Level 82
To adjust the trigger level 82 To force a trigger 82
Choosing the Trigger Mode 84
To set up edge triggers 84 To set up pulse width triggers 85
Keysight 1000B Series Oscilloscopes User’s Guide 7
Contents
To set up video triggers 86 To set up alternate triggers 88
Setting Other Trigger Parameters 90
To set the trigger sweep 90 To set the trigger coupling 90 To set the trigger high-frequency reject coupling 91 To change the trigger sensitivity 92 To specify a trigger holdoff 93
Using the External Trigger Input 94
4 Making Measurements 95
Displaying Automatic Measurements 96
To display an automatic measurement 97 To clear automatic measurements from the display 97 To display or hide all automatic measurements 97
8 Keysight 1000B Series Oscilloscopes User’s Guide
Voltage Measurements 98
Vmax (Maximum Voltage) 98 Vmin (Minimum Voltage) 99 Vpp (Peak-to-Peak Voltage) 99 Vtop (Top Voltage) 99 Vbase (Base Voltage) 99 Vamp (Amplitude Voltage = Vtop - Vbase) 99 Vavg (Average Voltage) 99 Vrms (Root-Mean-Square Voltage) 99 Overshoot 100 Preshoot 100
Time Measurements 101
Period 101 Frequency 102 Rise Time 102 Fall Time 102
Positive Pulse Wid th 103 Negative Pulse Width 103 Positive Duty Cycle 103 Negative Duty Cycle 103 Delay Between Rising Edges 104 Delay Between Falling Edges 104 Phase Between Rising Edges 105 Phase Between Falling Edges 105
Counter (Frequency) 106
Making Cursor Measurements 107
To use manually adjustable cursors 108 To use tracking cross-hair cursors 109 To display cursors for automatic measurements 110
5 Saving, Recalling, and Printing Data 111
Contents
Saving and Recalling Data 112
To save and recall waveforms 112 To save and recall oscilloscope setups 113 To save screens to BMP or PNG format files 114 To save data to CSV format files 115
Using the Disk Manager 116
To switch between files, path, and directory panes 117 To navigate the directory hierarchy 117 To create new folders 117 To edit folder/file names 118 To delete folders 118 To rename folders 119 To delete files 119 To recall files 119 To rename files 119 To display disk information 120
Keysight 1000B Series Oscilloscopes User’s Guide 9
Contents
Printing Screens 121
To choose a PictBridge printer 122 To print with inverted screen colors 122 To choose color or grayscale printing 123 To copy a screen to the printer 123
6 Oscilloscope Utility Settings 125
Displaying System Information 127
Turning Sound ON or OFF 127
Setting the Language (Menu and Help) 128
Performing Mask Tests 129
To enable/disable mask tests 129 To select the source channel for mask tests 129 To run/stop a mask test 130 To turn on/off the mask test message display 130 To set the mask test output condition 130 To stop a mask test on the output condition 132 To set up masks 132
Setting Preferences 135
To set up the screen saver 135 To select the vertical scale reference level 135 To select the USB device port function 135
Running Self-Calibration 137
7 Reference 139
Environmental Conditions 140
Overvoltage Category 140 Pollution Degree 140 Pollution Degree Definitions 140
Measurement Category 141
10 Keysight 1000B Series Oscilloscopes User’s Guide
Index 145
Contents
Measurement Category Definitions 141 Transient Withstand Capability 141
141
Specifications and Characteristics 142
Cleaning the Oscilloscope 142
Contacting Keysight 142
ASafety Notices143
Warnings 143
Safety Symbols 144
Keysight 1000B Series Oscilloscopes User’s Guide 11
Contents
12 Keysight 1000B Series Oscilloscopes User’s Guide

Figures

Figure 1. Power Switch 21 Figure 2. [Default Setup] Key 22 Figure 3. [Auto Scale] Key 24 Figure 4. Low Frequency Probe Compensation 26 Figure 5. High Frequency Probe Compensation 27 Figure 6. Front Panel 28 Figure 7. Oscilloscope Display 30 Figure 8. Softkey Menus 31 Figure 9. Run Control Keys 33 Figure 10. Securing the Instrument 35 Figure 11. Horizontal Controls 38 Figure 12. Status Bar, Trigger Position, and Horizontal Scale Control Indicators 39 Figure 13. Zoomed Time Base Window 41 Figure 14. X-Y Display Format Showing Out-of-Phase Waveforms 43 Figure 15. Vertical Controls 45 Figure 16. DC Coupling Control 48 Figure 17. AC Coupling Control 48 Figure 18. BW Limit Control OFF 49 Figure 19. BW Limit Control ON 50 Figure 20. Waveform Before Inversion 52 Figure 21. Waveform After Inversion 53 Figure 22. Math Scale Setting Value 54 Figure 23. FFT Waveform 57 Figure 24. [Display] Key 60 Figure 25. Alaising 65 Figure 26. Theoretical Brick-Wall Frequency Response 66 Figure 27. Sample Rate and Oscilloscope Bandwidth 67 Figure 28. Real-Time Sampling Mode 70
Keysight 1000B Series Oscilloscopes User’s Guide 13
Figures
Figure 29. Equivalent-Time (Repetitive) Sampling Mode 71 Figure 30. [Acquire] Key 73 Figure 31. Noisy Waveform Without Averaging 74 Figure 32. Noisy Waveform With Averaging 75 Figure 33. Peak Detect Waveform 76 Figure 34. Trigger Controls 82 Figure 35. Line Synchronization 87 Figure 36. Field Synchronization 88 Figure 37. Alternate Triggers 89 Figure 38. Trigger Holdoff 93 Figure 39. [Meas] Key 96 Figure 40. Voltage Measurement Points 98 Figure 41. Period and Frequency Measurements 101 Figure 42. Rise Time and Fall Time Measurements 102 Figure 43. Positive Pulse Width and Negative Pulse Width Measurements 103 Figure 44. Delay Measurements 104 Figure 45. Phase Measurements 105 Figure 46. [Cursors] Key 108 Figure 47. USB Host Port on Front Panel 111 Figure 48. [Save/Recall] Key 112 Figure 49. Disk Manager 116 Figure 50. Editing Folder/File Names in Disk Manager 118 Figure 51. USB Device Port 121 Figure 52. [Print] Key 122 Figure 53. [Utility] Key 126 Figure 54. Mask Test Display 130 Figure 55. Schematic Diagram of Mask Pass/Fail out 131 Figure 56. Mask Pass/Fail out 132 Figure 57. Mask Test Mask Setting 133 Figure 58. Calibration Screen 137
14 Keysight 1000B Series Oscilloscopes User’s Guide

Tables

Table 1. Keysight 1000B Series Oscilloscope Models 3 Table 2. Power Requirements 19 Table 3. Environmental Characteristics 20 Table 4. Auto Scale Default Settings 25 Table 5. Front Panel Controls 29 Table 6. FFT Window Characteristics 56
Keysight 1000B Series Oscilloscopes User’s Guide 15
Tables
16 Keysight 1000B Series Oscilloscopes User’s Guide
Keysight 1000B Series Oscilloscopes User’s Guide

1 Getting Started

Step 1. Inspect the package contents 18 Step 2. Turn on the oscilloscope 19 Step 3. Load the default oscilloscope setup 22 Step 4. Input a waveform 23 Step 5. Use Auto Scale 24 Step 6. Compensate probes 26 Step 7. Become familiar with the Front Panel Controls 28 Step 8. Become familiar with the oscilloscope display 30 Step 9. Use the Run Control keys 33 Step 10. Access the built-in help 34 Securing the Oscilloscope 35
This chapter describes the basic steps to take when first using the oscilloscope.
17
1 Getting Started

Step 1. Inspect the package contents

1 Inspect the shipping container for damage.
Keep a damaged shipping container or cushioning material until you have inspected the contents of the shipment for completeness and have checked the oscilloscope mechanically and electrically.
2 Verify that you received the following items in the oscilloscope packaging:
• Oscilloscope.
• Power cord.
• N2862A 10:1 10 MΩ passive probes, quantity = 2.
• Documentation CD.
• Front panel overlay (if language option other than English is chosen).
If anything is missing, or if you need to order additional probes, power cords, etc., contact your nearest Keysight Technologies sales office.
3 Inspect the oscilloscope.
• If there is mechanical damage or a defect, or if the oscilloscope does not operate properly or does not pass performance tests, notify your Keysight Technologies sales office.
• If the shipping container is damaged, or the cushioning materials show signs of stress, notify the carrier; then, contact your nearest Keysight Technologies sales office.
Keep the shipping materials for the carrier’s inspection.
The Keysight Technologies sales office will arrange for repair or replacement at Keysight’s option without waiting for claim settlement.
18 Keysight 1000B Series Oscilloscopes User’s Guide

Step 2. Turn on the oscilloscope

WARNING
The next few steps (turning on the oscilloscope, loading the default setup, and inputting a waveform) will provide a quick functional check to verify the oscilloscope is operating correctly.
1 Connect the power cord to a power source.
Use only power cords designed for your oscilloscope.
Use a power source that delivers the required power.
Table 2 Power Requirements
Name Typical Value
Line rating: ~Line 50 W max
Getting Started 1
100-120 V/50/60/400 Hz, ±10% 100-240 V/50/60 Hz, ±10%
To avoid electric shock, be sure the oscilloscope is properly grounded.
Keysight 1000B Series Oscilloscopes User’s Guide 19
1 Getting Started
Table 3 Environmental Characteristics
Name Typical Value
Ambient temperature: Operating 0 °C to +50 °C
Non-operating –20 °C to +60 °C
Humidity: Operating 80% RH (non-condensing) at +40 °C for 24 hr
Non-operating 60% RH (non-condensing) at +60 °C for 24 hr
Altitude: Operating to 3,000 m (9,842 ft)
Non-operating to 15,000 m (49,213 ft)
Vibration: Keysight class GP and MIL-PRF-28800F; Class 3 random
Shock: Keysight class GP and MIL-PRF-28800F; (operating 30 g, 1/2
sine, 11-ms duration, 3 shocks/axis along major axis. Total of 18 shocks)
Pollution degree 2: Normally only dry non-conductive pollution occurs.
Occasionally a temporary conductivity caused by condensation must be expected.
Indoor use: Rated for indoor use only.
20 Keysight 1000B Series Oscilloscopes User’s Guide
2 Turn on the oscilloscope.
Getting Started 1
Power switch

Figure 1 Power Switch

Keysight 1000B Series Oscilloscopes User’s Guide 21
1 Getting Started

Step 3. Load the default oscilloscope setup

You can recall the factory default setup any time you want to return the oscilloscope to its original setup.
1 Press the front panel [Default Setup] key.

Figure 2 [Default Setup] Key

2 When the Default menu appears, press [Menu On/Off] to turn off the menu.
(The Undo softkey in the Default menu lets you cancel the default setup and go back to the previous setup.)
22 Keysight 1000B Series Oscilloscopes User’s Guide

Step 4. Input a waveform

CAUTION
1 Input a waveform to a channel of the oscilloscope.
Use one of the supplied passive probes to input the Probe Comp signal from the front panel of the oscilloscope.
To avoid damage to the oscilloscope, make sure that the input voltage at the BNC connector does not exceed the maximum voltage (300 Vrms, CAT I).
Getting Started 1
Keysight 1000B Series Oscilloscopes User’s Guide 23
1 Getting Started

Step 5. Use Auto Scale

The oscilloscope has an auto scale feature that automatically sets the oscilloscope controls for the input waveforms present.
24 Keysight 1000B Series Oscilloscopes User’s Guide

Figure 3 [Auto Scale] Key

Auto scale requires waveforms with a frequency greater than or equal to 50 Hz and a duty cycle greater than 1%.
1 Press the front panel [Auto Scale] key. 2 When the AUTO menu appears, press [Menu On/Off] to turn off the menu.
The oscilloscope turns on all channels that have waveforms applied and sets the vertical and horizontal scales appropriately. It also selects a time base range based on the trigger source. The trigger source selected is the highest-numbered channel that has a waveform applied.
(The Undo softkey in the AUTO menu lets you cancel the auto scale and go back to the previous setup.)
Getting Started 1
The oscilloscope is configured to the following default control settings:
Table 4 Auto Scale Default Settings
Menu Setting
Horizontal time base Y-T (amplitude vs. time)
Acquisition mode Normal
Vertical coupling Adjusted to AC or DC according
to the waveform.
Vertical "V/div" Adjusted
Volts/Div Coarse
Bandwidth limit OFF
Waveform invert OFF
Horizontal position Center
Horizontal "s/div" Adjusted
Trigger type Edge
Trigger source Measure the channel with input
waveform automatically.
Trigger coupling DC
Trigger voltage Midpoint setting
Trigger sweep Auto
Keysight 1000B Series Oscilloscopes User’s Guide 25
1 Getting Started
Low frequency compensation adjustment
Over Compensated
Under Compensated
Correctly Compensated

Step 6. Compensate probes

Compensate probes to match your probe to the input channel. You should compensate a probe whenever you attach it for the first time to any input channel.

Low Frequency Compensation

For the supplied passive probes: 1 Set the Probe menu attenuation to 10X. If you use the probe hooktip, ensure a
proper connection by firmly inserting the tip onto the probe.
2 Attach the probe tip to the probe compensation connector and the ground lead
to the probe compensator ground connector.
3 Press the [Auto Scale] front panel key.
Figure 4 Low Frequency Probe Compensation
4 If waveform does not appear like the Correctly Compensated waveform shown
in Figure 4, then use a nonmetallic tool to adjust the low frequency compensation adjustment on the probe for the flattest square wave possible.

High Frequency Compensation

For the supplied passive probes:
26 Keysight 1000B Series Oscilloscopes User’s Guide
Getting Started 1
High frequency compensation adjustments
Over Compensated
Under Compensated
Correctly Compensated
1 Using the BNC adapter, connect the probe to a square wave generator. 2 Set the square wave generator to a frequency of 1 MHz, an amplitude of
3 Vp-p, and an output termination of 50Ω.
3 Press the [Auto Scale] front panel key.
Figure 5 High Frequency Probe Compensation
4 If waveform does not appear like the Correctly Compensated waveform shown
in Figure 5, then use a nonmetallic tool to adjust the 2 high frequency compensation adjustments on the probe for the flattest square wave possible.
Keysight 1000B Series Oscilloscopes User’s Guide 27
1 Getting Started
Entry knob Horizontal controls Run controls
Menu controlsVertical controls Trigger controls
Softkeys
Setup and File controls
Math functions Reference waveforms

Step 7. Become familiar with the Front Panel Controls

Before using the oscilloscope, familiarize yourself with the front panel controls.
The front panel has knobs, keys, and softkeys. Knobs are used most often to make adjustments. Keys are used for run controls and to change other oscilloscope settings via menus and softkeys.
28 Keysight 1000B Series Oscilloscopes User’s Guide

Figure 6 Front Panel

Getting Started 1
The definitions of the front panel knobs, keys, and softkeys are as follows:
Table 5 Front Panel Controls
Controls Consists of these knobs and keys
Entry knob For the adjustment of defined controls.
Setup controls [Auto Scale] and [Default Setup] front panel keys.
File controls [Save/Recall] and [Print] front panel keys.
Horizontal controls Position knob, [Horiz] front panel key, and scale knob.
Run controls [Run/Stop] and [Single] front panel keys.
Menu controls [Cursors], [Meas], [Acquire], [Display] , and [Utility] front panel keys.
Trigger controls Trigger [Level] knob, [Menu], and [Force Trigger] front panel keys.
Vertical controls Vertical position knobs, vertical scale knobs, channel ([1], [2], etc.)
[Math], and [Ref] front panel keys.
Softkeys Five gray keys from top to bottom on the right-hand side of the screen,
which select the adjacent menu items in the currently displayed menu.

Front Panel Overlays for Different Languages

If you choose a language option other than English, you get a front panel overlay for your language option.
To install a front panel overlay: 1 Insert the tabs on the left side of the overlay into the appropriate slots on the
front panel.
2 Gently press the overlay over the knobs and buttons. 3 When the overlay is against the front panel, insert the tabs on the right side of
the overlay into the slots on the front panel.
4 Let the overlay flatten out. It should remain secure on the front panel.
Keysight 1000B Series Oscilloscopes User’s Guide 29
1 Getting Started
Channel 1 Status, DC coupling, 500 mV per division
Timebase Status, 500 us per division
Channel 1 Reference
Acquisition Status
Waveform
Trigger, rising edge, channel 1,
1.56 V
Waveform window position in memory
Trigger position in memory
Trigger position in waveform window

Step 8. Become familiar with the oscilloscope display

30 Keysight 1000B Series Oscilloscopes User’s Guide

Figure 7 Oscilloscope Display

Loading...
+ 118 hidden pages