No part of this manual may be reproduced in
any form or by any means (including electronic storage and retrieval or translation into
a foreign language) without prior agreement
and written consent from Keysight Technologies, Inc. as governed by United States and
international copyright laws.
Manual Part Number
54139-97013
Edition
Second Edition, May 2012
Printed in Malaysia
Keysight Technologies, Inc.
1900 Garden of the Gods Road
Colorado Springs, CO 80907 USA
Warranty
The material contained in this document is
provided “as is,” and is subject to being
changed, without notice, in future editions.
Further, to the maximum extent permitted by
applicable law, Keysight disclaims all warranties, either express or implied, with regard
to this manual and any information contained
herein, including but not limited to the
implied warranties of merchantability and fitness for a particular purpose. Keysight shall
not be liable for errors or for incidental or
consequential damages in connection with
the furnishing, use, or performance of this
document or of any information contained
herein. Should Keysight and the user have a
separate written agreement with warranty
terms covering the material in this document
that conflict with these terms, the warranty
terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a
U.S. Government prime contract or subcontract, Software is delivered and licensed as
“Commercial computer software” as defined
in DFAR 252.227-7014 (June 1995), or as a
“commercial item” as defined in FAR 2.101(a)
or as “Restricted computer software” as
defined in FAR 52.227-19 (June 1987) or any
equivalent agency regulation or contract
clause. Use, duplication or d isclosure of
Software is subject to Keysight Technologies’
standard commercial license terms, and
non-DOD Departments and Agencies of the
U.S. Government will receive no greater than
Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14
(June 1987) or DFAR 252.227-7015 (b)(2)
(November 1995), as applicable in any technical data.
Safety Notices
A CAUTION notice denotes a hazard.
It calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like
that, if not correctly performed or
adhered to, could resul t in personal
injury or death. Do not proceed
beyond a WARNING notice until the
indicated conditions are fully
understood and met.
See also Appendix A, “Safety Notices,”
starting on page 143.
2Keysight 1000B Series Oscilloscopes User’s Guide
Page 3
Keysight 1000B Series Oscilloscopes—At a Glance
The Keysight 1000B Series oscilloscopes are low-cost portable digital storage
oscilloscopes (DSOs) that deliver these powerful features:
•Channel counts, input bandwidths, sample rates, and memory depths shown
in the following table:
Table 1Keysight 1000B Series Oscilloscope Models
ModelChannelsInput Bandwid thSample Rate
(2 channels on-1 channel on)
DSO1052B250 MHz500 MSa/s-1 GSa/s8-16 kpts
DSO1072B270 MHz500 MSa/s-1 GSa/s8-16 kpts
DSO1102B2100 MHz500 MSa/s-1 GSa/s8-16 kpts
DSO1152B2150 MHz500 MSa/s-1 GSa/s8-16 kpts
•Bright 5.7 inch QVGA (320 x 240) TFT color LED display and small footprint (to
save bench space).
•Up to 400 wfms/s refresh rate.
•Automatic voltage and time measurements (22) and cursor measurements.
•Powerful triggering (edge, pulse width, video, and alternate modes) with
adjustable sensitivity (to filter noise and avoid false triggers).
•Math function waveforms: add, subtract, multiply, FFT.
•USB ports (host and device) for easy printing, saving, and sharing of
waveforms, setups, screen BMP files, and CSV data files.
•Internal storage for 10 waveforms and 10 setups.
•Special digital filter and waveform recorder.
•Built-in 6-digit hardware frequency counter.
•Multi-language (11) user interface menus and built-in help.
Memory
(2 channels on-1 channel on)
Keysight 1000B Series Oscilloscopes User’s Guide3
Page 4
In This Book
1Getting Started
2Displaying Data
3Capturing Data
4Making Measurements
This guide shows how to use the Keysight 1000B Series oscilloscopes.
Describes the basic steps to take when first using the oscilloscope.
Describes how to use the horizontal and vertical controls, channel settings, math
waveforms, reference waveforms, and display settings.
Describes acquisition modes and how to set up triggers.
Describes voltage, time, and cursor measurements.
5Saving, Recalling, and Printing Data
Describes how to save, recall, and print data.
6Oscilloscope Utility Settings
Describes other oscilloscope settings found in the Utility menu.
7Reference
Contains reference information for the 1000B Series oscilloscopes.
4Keysight 1000B Series Oscilloscopes User’s Guide
Page 5
Contents
Keysight 1000B Series Oscilloscopes—At a Glance3
In This Book4
Figures13
Tab les15
1Getting Started17
Step 1. Inspect the package contents18
Step 2. Turn on the oscilloscope19
Step 3. Load the default oscilloscope setup22
Step 4. Input a waveform23
23
Step 5. Use Auto Scale24
Step 6. Compensate probes26
Low Frequency Compensation26
High Frequency Compensation26
Step 7. Become familiar with the Front Panel Controls28
Front Panel Overlays for Different Languages29
Step 8. Become familiar with the oscilloscope display30
Using the Oscilloscope Softkey Menus31
Step 9. Use the Run Control keys33
Step 10. Access the built-in help34
Keysight 1000B Series Oscilloscopes User’s Guide5
Page 6
Contents
Securing the Oscilloscope35
2Displaying Data37
Using the Horizontal Controls38
To adjust the horizontal scale39
To adjust the horizontal position40
To display the zoomed time base41
To change the horizontal time base (Y-T, X-Y, or Roll)42
To view the sample rate44
Using the Vertical Controls45
To turn waveforms on or off (channel, math, or reference)46
To adjust the vertical scale46
To adjust the vertical position46
To specify channel coupling47
To specify a bandwidth limit49
To specify the probe attenuation50
To use a digital filter50
To change the Volts/Div control sensitivity51
To invert a waveform52
Using Math Function Waveforms54
To add, subtract, or multiply waveforms54
To display the frequency domain using FFT55
Using Reference Waveforms58
To save a reference waveform58
To export or import reference waveforms58
To return the reference waveform to its default scale59
Changing the Display Settings60
To display waveforms as vectors or dots60
To clear the display61
To set waveform persistence61
To adjust waveform intensity61
6Keysight 1000B Series Oscilloscopes User’s Guide
Page 7
To change the grid61
To adjust the grid brightness62
To invert screen colors62
To change the menu display time62
3Capturing Data63
Overview of Sampling64
Sampling Theory64
Aliasing64
Oscilloscope Bandwidth and Sample Rate65
Oscilloscope Rise Time67
Oscilloscope Bandwidth Required68
Memory Depth and Sample Rate69
Choosing the Sampling Mode70
To select the real-time sampling mode70
To select the equivalent-time sampling mode71
Choosing the Acquisition Mode73
To select the Normal acquisition mode74
To select the Average acquisition mode74
To select the Peak Detect acquisition mode75
To turn OFF/ON sine(x)/x interpolation77
Contents
Recording/Playing-back Waveforms78
To record waveforms78
To play-back waveforms79
To store recorded waveforms80
Adjusting the Trigger Level82
To adjust the trigger level82
To force a trigger82
Choosing the Trigger Mode84
To set up edge triggers84
To set up pulse width triggers85
Keysight 1000B Series Oscilloscopes User’s Guide7
Page 8
Contents
To set up video triggers86
To set up alternate triggers88
Setting Other Trigger Parameters90
To set the trigger sweep90
To set the trigger coupling90
To set the trigger high-frequency reject coupling91
To change the trigger sensitivity92
To specify a trigger holdoff93
Using the External Trigger Input94
4Making Measurements95
Displaying Automatic Measurements96
To display an automatic measurement97
To clear automatic measurements from the display97
To display or hide all automatic measurements97
Positive Pulse Wid th103
Negative Pulse Width103
Positive Duty Cycle103
Negative Duty Cycle103
Delay Between Rising Edges104
Delay Between Falling Edges104
Phase Between Rising Edges105
Phase Between Falling Edges105
Counter (Frequency)106
Making Cursor Measurements107
To use manually adjustable cursors108
To use tracking cross-hair cursors109
To display cursors for automatic measurements110
5Saving, Recalling, and Printing Data111
Contents
Saving and Recalling Data112
To save and recall waveforms112
To save and recall oscilloscope setups113
To save screens to BMP or PNG format files114
To save data to CSV format files115
Using the Disk Manager116
To switch between files, path, and directory panes117
To navigate the directory hierarchy117
To create new folders117
To edit folder/file names118
To delete folders118
To rename folders119
To delete files119
To recall files119
To rename files119
To display disk information120
Keysight 1000B Series Oscilloscopes User’s Guide9
Page 10
Contents
Printing Screens121
To choose a PictBridge printer122
To print with inverted screen colors122
To choose color or grayscale printing123
To copy a screen to the printer123
6Oscilloscope Utility Settings125
Displaying System Information127
Turning Sound ON or OFF127
Setting the Language (Menu and Help)128
Performing Mask Tests129
To enable/disable mask tests129
To select the source channel for mask tests129
To run/stop a mask test130
To turn on/off the mask test message display130
To set the mask test output condition130
To stop a mask test on the output condition132
To set up masks132
Setting Preferences135
To set up the screen saver135
To select the vertical scale reference level135
To select the USB device port function135
Keysight 1000B Series Oscilloscopes User’s Guide11
Page 12
Contents
12Keysight 1000B Series Oscilloscopes User’s Guide
Page 13
Figures
Figure 1. Power Switch21
Figure 2. [Default Setup] Key22
Figure 3. [Auto Scale] Key24
Figure 4. Low Frequency Probe Compensation26
Figure 5. High Frequency Probe Compensation27
Figure 6. Front Panel28
Figure 7. Oscilloscope Display30
Figure 8. Softkey Menus31
Figure 9. Run Control Keys33
Figure 10. Securing the Instrument35
Figure 11. Horizontal Controls38
Figure 12. Status Bar, Trigger Position, and Horizontal Scale Control Indicators39
Figure 13. Zoomed Time Base Window41
Figure 14. X-Y Display Format Showing Out-of-Phase Waveforms43
Figure 15. Vertical Controls45
Figure 16. DC Coupling Control48
Figure 17. AC Coupling Control48
Figure 18. BW Limit Control OFF49
Figure 19. BW Limit Control ON50
Figure 20. Waveform Before Inversion52
Figure 21. Waveform After Inversion53
Figure 22. Math Scale Setting Value54
Figure 23. FFT Waveform57
Figure 24. [Display] Key60
Figure 25. Alaising65
Figure 26. Theoretical Brick-Wall Frequency Response66
Figure 27. Sample Rate and Oscilloscope Bandwidth67
Figure 28. Real-Time Sampling Mode70
Keysight 1000B Series Oscilloscopes User’s Guide13
Page 14
Figures
Figure 29. Equivalent-Time (Repetitive) Sampling Mode71
Figure 30. [Acquire] Key73
Figure 31. Noisy Waveform Without Averaging74
Figure 32. Noisy Waveform With Averaging75
Figure 33. Peak Detect Waveform76
Figure 34. Trigger Controls82
Figure 35. Line Synchronization87
Figure 36. Field Synchronization88
Figure 37. Alternate Triggers89
Figure 38. Trigger Holdoff93
Figure 39. [Meas] Key96
Figure 40. Voltage Measurement Points98
Figure 41. Period and Frequency Measurements101
Figure 42. Rise Time and Fall Time Measurements102
Figure 43. Positive Pulse Width and Negative Pulse Width Measurements103
Figure 44. Delay Measurements104
Figure 45. Phase Measurements105
Figure 46. [Cursors] Key108
Figure 47. USB Host Port on Front Panel111
Figure 48. [Save/Recall] Key112
Figure 49. Disk Manager116
Figure 50. Editing Folder/File Names in Disk Manager118
Figure 51. USB Device Port121
Figure 52. [Print] Key122
Figure 53. [Utility] Key126
Figure 54. Mask Test Display130
Figure 55. Schematic Diagram of Mask Pass/Fail out131
Figure 56. Mask Pass/Fail out132
Figure 57. Mask Test Mask Setting133
Figure 58. Calibration Screen137
14Keysight 1000B Series Oscilloscopes User’s Guide
Page 15
Tables
Table 1. Keysight 1000B Series Oscilloscope Models3
Table 2. Power Requirements19
Table 3. Environmental Characteristics20
Table 4. Auto Scale Default Settings25
Table 5. Front Panel Controls29
Table 6. FFT Window Characteristics56
Keysight 1000B Series Oscilloscopes User’s Guide15
Page 16
Tables
16Keysight 1000B Series Oscilloscopes User’s Guide
Page 17
Keysight 1000B Series Oscilloscopes
User’s Guide
1Getting Started
Step 1. Inspect the package contents 18
Step 2. Turn on the oscilloscope 19
Step 3. Load the default oscilloscope setup 22
Step 4. Input a waveform 23
Step 5. Use Auto Scale 24
Step 6. Compensate probes 26
Step 7. Become familiar with the Front Panel Controls 28
Step 8. Become familiar with the oscilloscope display 30
Step 9. Use the Run Control keys 33
Step 10. Access the built-in help 34
Securing the Oscilloscope 35
This chapter describes the basic steps to take when first using the oscilloscope.
17
Page 18
1Getting Started
Step 1. Inspect the package contents
1 Inspect the shipping container for damage.
Keep a damaged shipping container or cushioning material until you have
inspected the contents of the shipment for completeness and have checked the
oscilloscope mechanically and electrically.
2 Verify that you received the following items in the oscilloscope packaging:
• Oscilloscope.
• Power cord.
• N2862A 10:1 10 MΩ passive probes, quantity = 2.
• Documentation CD.
• Front panel overlay (if language option other than English is chosen).
If anything is missing, or if you need to order additional probes, power cords,
etc., contact your nearest Keysight Technologies sales office.
3 Inspect the oscilloscope.
• If there is mechanical damage or a defect, or if the oscilloscope does not
operate properly or does not pass performance tests, notify your Keysight
Technologies sales office.
• If the shipping container is damaged, or the cushioning materials show signs
of stress, notify the carrier; then, contact your nearest Keysight Technologies
sales office.
Keep the shipping materials for the carrier’s inspection.
The Keysight Technologies sales office will arrange for repair or replacement
at Keysight’s option without waiting for claim settlement.
18Keysight 1000B Series Oscilloscopes User’s Guide
Page 19
Step 2. Turn on the oscilloscope
WARNING
The next few steps (turning on the oscilloscope, loading the default setup, and
inputting a waveform) will provide a quick functional check to verify the
oscilloscope is operating correctly.
1 Connect the power cord to a power source.
Use only power cords designed for your oscilloscope.
Use a power source that delivers the required power.
To avoid electric shock, be sure the oscilloscope is properly grounded.
Keysight 1000B Series Oscilloscopes User’s Guide19
Page 20
1Getting Started
Table 3Environmental Characteristics
NameTypical Value
Ambient temperature:Operating 0 °C to +50 °C
Non-operating –20 °C to +60 °C
Humidity:Operating 80% RH (non-condensing) at +40 °C for 24 hr
Non-operating 60% RH (non-condensing) at +60 °C for 24 hr
Altitude:Operating to 3,000 m (9,842 ft)
Non-operating to 15,000 m (49,213 ft)
Vibration:Keysight class GP and MIL-PRF-28800F; Class 3 random
Shock:Keysight class GP and MIL-PRF-28800F; (operating 30 g, 1/2
sine, 11-ms duration, 3 shocks/axis along major axis. Total of 18
shocks)
Pollution degree 2:Normally only dry non-conductive pollution occurs.
Occasionally a temporary conductivity caused by condensation
must be expected.
Indoor use:Rated for indoor use only.
20Keysight 1000B Series Oscilloscopes User’s Guide
Page 21
2 Turn on the oscilloscope.
Getting Started1
Power switch
Figure 1Power Switch
Keysight 1000B Series Oscilloscopes User’s Guide21
Page 22
1Getting Started
Step 3. Load the default oscilloscope setup
You can recall the factory default setup any time you want to return the
oscilloscope to its original setup.
1 Press the front panel [Default Setup] key.
Figure 2[Default Setup] Key
2 When the Default menu appears, press [Menu On/Off] to turn off the menu.
(The Undo softkey in the Default menu lets you cancel the default setup and go
back to the previous setup.)
22Keysight 1000B Series Oscilloscopes User’s Guide
Page 23
Step 4. Input a waveform
CAUTION
1 Input a waveform to a channel of the oscilloscope.
Use one of the supplied passive probes to input the Probe Comp signal from the
front panel of the oscilloscope.
To avoid damage to the oscilloscope, make sure that the input voltage at the BNC
connector does not exceed the maximum voltage (300 Vrms, CAT I).
Getting Started1
Keysight 1000B Series Oscilloscopes User’s Guide23
Page 24
1Getting Started
Step 5. Use Auto Scale
The oscilloscope has an auto scale feature that automatically sets the oscilloscope
controls for the input waveforms present.
24Keysight 1000B Series Oscilloscopes User’s Guide
Figure 3[Auto Scale] Key
Auto scale requires waveforms with a frequency greater than or equal to 50 Hz
and a duty cycle greater than 1%.
1 Press the front panel [Auto Scale] key.
2 When the AUTO menu appears, press [Menu On/Off] to turn off the menu.
The oscilloscope turns on all channels that have waveforms applied and sets
the vertical and horizontal scales appropriately. It also selects a time base
range based on the trigger source. The trigger source selected is the
highest-numbered channel that has a waveform applied.
(The Undo softkey in the AUTO menu lets you cancel the auto scale and go back
to the previous setup.)
Page 25
Getting Started1
The oscilloscope is configured to the following default control settings:
Table 4Auto Scale Default Settings
MenuSetting
Horizontal time baseY-T (amplitude vs. time)
Acquisition modeNormal
Vertical couplingAdjusted to AC or DC according
to the waveform.
Vertical "V/div"Adjusted
Volts/DivCoarse
Bandwidth limitOFF
Waveform invertOFF
Horizontal positionCenter
Horizontal "s/div"Adjusted
Trigger typeEdge
Trigger sourceMeasure the channel with input
waveform automatically.
Trigger couplingDC
Trigger voltageMidpoint setting
Trigger sweepAuto
Keysight 1000B Series Oscilloscopes User’s Guide25
Page 26
1Getting Started
Low frequency compensation adjustment
Over Compensated
Under Compensated
Correctly Compensated
Step 6. Compensate probes
Compensate probes to match your probe to the input channel. You should
compensate a probe whenever you attach it for the first time to any input channel.
Low Frequency Compensation
For the supplied passive probes:
1 Set the Probe menu attenuation to 10X. If you use the probe hooktip, ensure a
proper connection by firmly inserting the tip onto the probe.
2 Attach the probe tip to the probe compensation connector and the ground lead
to the probe compensator ground connector.
3 Press the [Auto Scale] front panel key.
Figure 4Low Frequency Probe Compensation
4 If waveform does not appear like the Correctly Compensated waveform shown
in Figure 4, then use a nonmetallic tool to adjust the low frequency
compensation adjustment on the probe for the flattest square wave possible.
High Frequency Compensation
For the supplied passive probes:
26Keysight 1000B Series Oscilloscopes User’s Guide
Page 27
Getting Started1
High frequency compensation adjustments
Over Compensated
Under Compensated
Correctly Compensated
1 Using the BNC adapter, connect the probe to a square wave generator.
2 Set the square wave generator to a frequency of 1 MHz, an amplitude of
3 Vp-p, and an output termination of 50Ω.
3 Press the [Auto Scale] front panel key.
Figure 5High Frequency Probe Compensation
4 If waveform does not appear like the Correctly Compensated waveform shown
in Figure 5, then use a nonmetallic tool to adjust the 2 high frequency
compensation adjustments on the probe for the flattest square wave possible.
Keysight 1000B Series Oscilloscopes User’s Guide27
Page 28
1Getting Started
Entry knobHorizontal controlsRun controls
Menu controlsVertical controlsTrigger controls
Softkeys
Setup and File controls
Math functionsReference waveforms
Step 7. Become familiar with the Front Panel Controls
Before using the oscilloscope, familiarize yourself with the front panel controls.
The front panel has knobs, keys, and softkeys. Knobs are used most often to make
adjustments. Keys are used for run controls and to change other oscilloscope
settings via menus and softkeys.
28Keysight 1000B Series Oscilloscopes User’s Guide
Figure 6Front Panel
Page 29
Getting Started1
The definitions of the front panel knobs, keys, and softkeys are as follows:
Table 5Front Panel Controls
ControlsConsists of these knobs and keys
Entry knobFor the adjustment of defined controls.
Setup controls[Auto Scale] and [Default Setup] front panel keys.
File controls[Save/Recall] and [Print] front panel keys.
Horizontal controlsPosition knob, [Horiz] front panel key, and scale knob.
Run controls[Run/Stop] and [Single] front panel keys.
Menu controls[Cursors], [Meas], [Acquire], [Display] , and [Utility] front panel keys.
Trigger controlsTrigger [Level] knob, [Menu], and [Force Trigger] front panel keys.
SoftkeysFive gray keys from top to bottom on the right-hand side of the screen,
which select the adjacent menu items in the currently displayed
menu.
Front Panel Overlays for Different Languages
If you choose a language option other than English, you get a front panel overlay
for your language option.
To install a front panel overlay:
1 Insert the tabs on the left side of the overlay into the appropriate slots on the
front panel.
2 Gently press the overlay over the knobs and buttons.
3 When the overlay is against the front panel, insert the tabs on the right side of
the overlay into the slots on the front panel.
4 Let the overlay flatten out. It should remain secure on the front panel.
Keysight 1000B Series Oscilloscopes User’s Guide29
Page 30
1Getting Started
Channel 1 Status,
DC coupling, 500 mV per division
Timebase
Status,
500 us per
division
Channel 1
Reference
Acquisition
Status
Waveform
Trigger,
rising edge,
channel 1,
1.56 V
Waveform window
position in memory
Trigger position
in memory
Trigger position in
waveform window
Step 8. Become familiar with the oscilloscope display
30Keysight 1000B Series Oscilloscopes User’s Guide
Figure 7Oscilloscope Display
Page 31
Using the Oscilloscope Softkey Menus
Getting Started1
Figure 8Softkey Menus
When one of the oscilloscope front panel keys turns on a menu, you can use the
five softkeys to choose items from the menu.
Some common menu item choices are:
Accesses the next page of items in the menu.
Accesses the previous page of items in the menu.
Keysight 1000B Series Oscilloscopes User’s Guide31
Page 32
1Getting Started
Returns to the previous menu in the hierarchy.
The [Menu On/Off] front panel key turns off the menu or turns on the last accessed
menu on again. The Menu Display item in the Display menu lets you select the
amount of time menus are displayed (see “To change the menu display time” on
page 62).
32Keysight 1000B Series Oscilloscopes User’s Guide
Page 33
Step 9. Use the Run Control keys
There are two front panel keys for starting and stopping the oscilloscope’s
acquisition system: [Run/Stop] and [Single].
Getting Started1
Figure 9Run Control Keys
• When the [Run/Stop] key is green, the oscilloscope is acquiring data. To stop
acquiring data, press [Run/Stop]. When stopped, the last acquired waveform is
displayed.
• When the [Run/Stop] key is red, data acquisition is stopped. To start acquiring
data, press [Run/Stop].
• To capture and display a single acquisition (whether the oscilloscope is running
or stopped), press [Single]. After capturing and displaying a single acquisition,
the [Run/Stop] key is red.
Keysight 1000B Series Oscilloscopes User’s Guide33
Page 34
1Getting Started
Step 10. Access the built-in help
The oscilloscope has built-in quick help information. To access the built-in help:
1 Press and hold the front panel key, softkey, or pushable knob on which you
would like quick help information.
The built-in help is available in 11 different languages (see “Setting the Language
(Menu and Help)” on page 128).
34Keysight 1000B Series Oscilloscopes User’s Guide
Page 35
Securing the Oscilloscope
To secure a 1000B Series oscilloscope to its location, you can use a Kensington
lock or the security loop.
Getting Started1
Hole for Kensington lock
Security loop
Figure 10Securing the Instrument
Keysight 1000B Series Oscilloscopes User’s Guide35
Page 36
1Getting Started
36Keysight 1000B Series Oscilloscopes User’s Guide
Page 37
Keysight 1000B Series Oscilloscopes
User’s Guide
2Displaying Data
Using the Horizontal Controls 38
Using the Vertical Controls 45
Using Math Function Waveforms 54
Using Reference Waveforms 58
Changing the Display Settings 60
This chapter describes how to use the horizontal and vertical controls, channel
settings, math waveforms, reference waveforms, and display settings.
37
Page 38
2Displaying Data
Using the Horizontal Controls
The horizontal controls consist of:
• The horizontal scale knob — changes the oscilloscope’s time per division
setting using the center of the screen as a reference.
• The horizontal position knob — changes the position of the trigger point relative
to the center of the screen.
•The [Horiz] key — displays the Horizontal menu which lets you display the
zoomed (delayed) time base, change the time base mode, and display the
sample rate.
Figure 11Horizontal Controls
Figure 12 shows the screen icon descriptions and control indicators.
38Keysight 1000B Series Oscilloscopes User’s Guide
Page 39
Displaying Data2
Horizontal
scale
setting
Sample rate
Displayed waveform window ([ ])
position in memory
Trigger position
in memory
Trigger position in
waveform window
To adjust the horizontal scale
Keysight 1000B Series Oscilloscopes User’s Guide39
Figure 12Status Bar, Trigger Position, and Horizontal Scale Control Indicators
• Turn the horizontal scale knob to change the horizontal time per division
(time/div) setting (and the oscilloscope’s sample rate — see “Memory Depth
and Sample Rate” on page 69).
The time/div setting changes in a 1-2-5 step sequence.
The time/div setting is also known as the sweep speed.
Page 40
2Displaying Data
When the time/div is set to 50 ms/div or slower, the oscilloscope enters Slow
Scan mode (see “Slow Scan Mode” below).
When the horizontal scale is set to 20 ns or faster, the oscilloscope uses
sine(x)/x interpolation to expand the horizontal time base.
• Push the horizontal scale knob to toggle between the zoomed time base and
the normal time base display (see “To display the zoomed time base” on
page 41).
The time/div setting is displayed in the status bar at the top left of the screen.
Because all channels are displayed in the same time base (except in the Alternate
trigger mode), the oscilloscope displays one time/div setting for all channels.
Slow Scan Mode
When the horizontal scale is set to 50 ms/div or slower, the oscilloscope enters
Slow Scan mode.
In the Slow Scan mode, peak detect acquisition is used so that no data is missed
(even the though the Acquire menu may show a different acquisition mode
setting). The oscilloscope acquires sufficient data for the pre-trigger part of the
display, then waits for the trigger. When the trigger occurs, the oscilloscope
continues to capture data for the post-trigger part of the display.
When using the Slow Scan mode to view low frequency signals, the channel
coupling should be set to “DC”.
The Slow Scan mode lets you see dynamic changes (like the adjustment of a
potentiometer) on low frequency waveforms. For example, Slow Scan mode is
often used in applications like transducer monitoring and power supply testing.
To adjust the horizontal position
• Turn the horizontal position knob to change the position of the trigger point
relative to the center of the screen.
The position knob adjusts the horizontal position of all channels, math
functions, and reference waveforms.
• Push the horizontal position knob to “zero” the trigger point (in other words,
move it to the center of the screen).
40Keysight 1000B Series Oscilloscopes User’s Guide
Page 41
To display the zoomed time base
Zoomed window
Zoomed window view
The zoomed time base (also known as delayed sweep time base), magnifies a
portion of the original waveform display (now on the top half of the screen) and
displays it in a zoomed time base on the bottom half of the screen.
1 To toggle the zoomed time base “ON” or “OFF”, either push the horizontal scale
knob or press [Horiz] key followed by the Zoom softkey in the Horizontal menu.
2 When the zoomed time base is “ON”:
• The top half of the display shows the original waveform and the portion
being magnified.
• The horizontal scale knob changes the magnification (widens or narrows the
area of magnification).
• The horizontal position knob moves the area of magnification forward and
backward on the original waveform.
• The bottom half of the display shows the magnified data in the zoomed time
base.
Displaying Data2
Figure 13Zoomed Time Base Window
Keysight 1000B Series Oscilloscopes User’s Guide41
Page 42
2Displaying Data
To change the horizontal time base (Y-T, X-Y, or Roll)
1 Press [Horiz].
2 In the Horizontal menu, press Time Base.
3 Continue pressing the Time Base softkey or turn the entry knob to select
between:
Y-TAmplitude vs. time. This is the typical horizontal time base
setting.
X-YChannel 2 (X-axis) vs. Channel 1 (Y-axis), see “X-Y Format” on
page 42.
RollIn Roll mode, the waveform display rolls from right to left, and
the minimum horizontal scale setting is 500 ms/div. No trigger
or horizontal position control is available. Roll mode is used in
applications similar to the ones for which Slow Scan mode is
used (see “Slow Scan Mode” on page 40).
42Keysight 1000B Series Oscilloscopes User’s Guide
X-Y Format
This format compares the voltage level of two waveforms point by point. It is
useful for studying phase relationships between two waveforms. This format only
applies to channels 1 and 2. Choosing the X-Y display format displays channel 1
on the horizontal axis and channel 2 on the vertical axis.
The oscilloscope uses the untriggered sample acquisition mode and waveform
data is displayed as dots. The sampling rate can vary from 4 kSa/s to 100 MSa/s,
and the default sampling rate is 1 MSa/s.
Page 43
Figure 14X-Y Display Format Showing Out-of-Phase Waveforms
Displaying Data2
Keysight 1000B Series Oscilloscopes User’s Guide43
Page 44
2Displaying Data
To view the sample rate
The following modes or functions are not available in X-Y format:
• Automatic voltage or time measurements.
• Cursor measurements.
•Mask testing.
• Math function waveforms.
• Reference waveforms.
• Zoomed time base display.
• Displaying waveforms as vectors.
• Horizontal position knob.
• Trigger controls.
1 Press [Horiz].
2 In the Horizontal menu, the Sa Rate menu item d isplays the sample rate used for
the current horizontal scale setting.
44Keysight 1000B Series Oscilloscopes User’s Guide
See Also“Memory Depth and Sample Rate” on page 69.
Page 45
Using the Vertical Controls
Scale knobs
Position knobs
The vertical controls consist of:
• The channel ([1], [2], [3], and [4]), [Math], and [Ref] front panel keys — turn
waveforms on or off (and display or hide their menus).
• The vertical scale knobs — change the amplitude per division setting for a
waveform, using either ground or the center of the screen as a reference
(depending on a preference setting).
• The vertical position knobs — change the vertical position of the waveform on
the screen.
Displaying Data2
Figure 15Vertical Controls
Keysight 1000B Series Oscilloscopes User’s Guide45
Page 46
2Displaying Data
To turn waveforms on or off (channel, math, or reference)
To adjust the vertical scale
Pressing the channel ([1], [2], [3], and [4]), [Math], and [Ref] front panel keys have
the following effect:
• If the waveform is off, the waveform is turned on and its menu is displayed.
• If the waveform is on and its menu is not displayed, its menu will be displayed.
• If the waveform is on and its menu is displayed, the waveform is turned off and
its menu goes away.
When an input channel waveform is on:
• Turn its vertical scale knob to change the amplitude per division setting.
The amplitude/div setting changes in a 1-2-5 step sequence from 2 mV/div to
10 V/div (with “1X” probe attenuation).
Either ground or the center of the screen is used as a reference, depending on
the “Expand Reference” preference setting (see “To select the vertical scale
reference level” on page 135). The “center of screen” reference is not available
for math function or reference waveforms.
• Push its vertical scale knob to toggle between vernier (fine scale) adjustment
and normal adjustment.
With vernier adjustment, the amplitude/div setting changes in small steps
between the normal (coarse scale) settings.
The Volts/Div item in a channel’s menu also toggles between vernier and
normal adjustment (see “To change the Volts/Div control sensitivity” on
page 51).
Vernier adjustment is not available for math function or reference waveforms.
The amplitude/div setting is displayed in the status bar at the bottom of the
screen.
To adjust the vertical position
Adjusting their vertical position lets you compare waveforms by aligning them
above one another or on top of each other.
When an input channel waveform is on:
46Keysight 1000B Series Oscilloscopes User’s Guide
Page 47
• Turn the vertical position knob to change the vertical position of the waveform
on the screen.
Notice that the ground reference symbol on the left side of the display moves
with the waveform.
• Push the vertical position knob to “zero” the ground reference (in other words,
move it to the center of the screen).
Notice that, as you adjust the vertical position, a message showing the position of
the ground reference relative to the center of the screen is temporarily displayed in
the lower left-hand corner of the screen.
To specify channel coupling
1 If the channel’s menu is not currently displayed, press the channel key ([1], [2],
[3], or [4]).
2 In the Channel menu, press Coupling.
3 Continue pressing the Coupling softkey or turn the entry knob to select
between:
Displaying Data2
DCPasses both DC and AC components of the input waveform to
the oscilloscope. See Figure 16.
You can quickly measure the DC component of the waveform by
simply noting its distance from the ground symbol.
ACBlocks the DC component of the input waveform and passes
the AC component. See Figure 17.
This lets you use greater sensitivity (amplitude/div settings) to
display the AC component of the waveform.
GNDThe waveform is disconnected from the oscilloscope input.
Keysight 1000B Series Oscilloscopes User’s Guide47
Page 48
2Displaying Data
DC Coupling
DC Coupling Status
Figure 16DC Coupling Control
AC Coupling
AC Coupling Status
Figure 17AC Coupling Control
48Keysight 1000B Series Oscilloscopes User’s Guide
Page 49
To specify a bandwidth limit
When high frequency components of a waveform are not important to its analysis,
the bandwidth limit control can be used to reject frequencies above 20 MHz. See
Figure 19 and Figure 18.
1 If the channel’s menu is not currently displayed, press the channel key ([1], [2],
[3], or [4]).
2 In the Channel menu, press BW Limit to toggle the bandwidth limit setting “ON”
and “OFF”.
Displaying Data2
Bandwidth limit OFF
Figure 18BW Limit Control OFF
Keysight 1000B Series Oscilloscopes User’s Guide49
Page 50
2Displaying Data
To specify the probe attenuation
20 MHz Bandwidth
Bandwidth ON Status
Figure 19BW Limit Control ON
For correct measurements, you must match the oscilloscope’s probe attenuation
factor settings with the attenuation factors of the probes being used.
The probe attenuation factor setting changes the vertical scaling of the
oscilloscope so that the measurement results reflect the actual voltage levels at
the probe tip.
1 If the channel’s menu is not currently displayed, press the channel key ([1], [2],
[3], or [4]).
2 In the Channel menu, press Probe.
3 Continue pressing the Probe softkey or turn the entry knob to select the
appropriate attenuation factor.
To use a digital filter
You can apply a digital filter to the sampled waveform data.
1 If the channel’s menu is not currently displayed, press the channel key ([1], [2],
[3], or [4]).
50Keysight 1000B Series Oscilloscopes User’s Guide
Page 51
Displaying Data2
2 In the Channel menu, press Digital Filter.
3 In the Filter menu, press Filter Type, and continue pressing the Filter Type softkey
or turn the entry knob to select between:
LPF (Low Pass Filter).
HPF (High Pass Filter).
BPF (Band Pass Filter).
BRF (Band Reject Filter).
4 Depending on the type of filter selected, press Upper Limit and/or Lower Limit,
and turn the entry knob to adjust the limit.
The horizontal scale control sets the maximum value for the upper and lower
limits.
Digital filters are not available when:
• The horizontal scale is 20 ns/div or lower.
• The horizontal scale is 50 ms/div or higher.
To change the Volts/Div control sensitivity
When you need to adjust the amplitude/div setting in smaller increments, you can
change the sensitivity ofthe vertical scale control.
1 If the channel’s menu is not currently displayed, press the channel key ([1], [2],
[3], or [4]).
2 In the Channel menu, press Volts/Div to toggle between:
CoarseThe verical scale knob changes the amplitude/div setting in a
1-2-5 step sequence from 2 mV/div to 10 V/div (with “1X”
probe attenuation).
FineAlso known as vernier, the vertical scale knob changes the
amplitude/div setting in small steps between the normal
(coarse scale) settings.
Keysight 1000B Series Oscilloscopes User’s Guide51
Page 52
2Displaying Data
To invert a waveform
You can also toggle between coarse and fine settings by pushing the vertical scale
knob (see “To adjust the vertical scale” on page 46).
You can invert a waveform with respect to the ground level.
1 If the channel’s menu is not currently displayed, press the channel key ([1], [2],
[3], or [4]).
2 In the Channel menu, press Invert to toggle between “ON” and “OFF”.
Figure 20 and Figure 21 show the changes before and after inversion.
Figure 20Waveform Before Inversion
52Keysight 1000B Series Oscilloscopes User’s Guide
Page 53
Figure 21Waveform After Inversion
Displaying Data2
Keysight 1000B Series Oscilloscopes User’s Guide53
Page 54
2Displaying Data
Using Math Function Waveforms
The math functions control allows the selection of the math functions:
•Add.
• Subtract.
• Multiply.
• FFT (Fast Fourier Transform).
The mathematical result can be measured using the grid and cursor controls.
The amplitude of the math waveform can be adjusted using a menu item selection
in the Math menu and the entry knob. The adjustment range is in a 1-2-5 step
from 0.1% to 1000%.
The math scale setting is shown at the bottom of the display.
Math Scale
Figure 22Math Scale Setting Value
To add, subtract, or multiply waveforms
1 Press [Math].
2 In the Math menu, press Operate.
54Keysight 1000B Series Oscilloscopes User’s Guide
Page 55
3 Continue pressing the Operate softkey or turn the entry knob to select
NOTE
“A + B”, “A - B”, or “A x B”.
4 Press Source A, and continue pressing the softkey to select the desired input
channel.
5 Press Source B, and continue pressing the softkey to select the desired input
channel.
6 To invert the result of the addition, subtraction, or multiplication (with respect
to the reference level), select Invert to toggle between “ON” and “OFF”.
To display the frequency domain using FFT
The FFT math function mathematically converts a time-domain waveform into its
frequency components. FFT waveforms are useful for finding the harmonic content
and distortion in systems, for characterizing noise in DC power supplies, and for
analyzing vibration.
To display a waveform’s FFT:
1 Press [Math].
2 In the Math menu, press Operate.
3 Continue pressing the Operate softkey or turn the entry knob to select “FFT”.
4 In the FFT menu, press Source, and continue pressing the softkey to select the
desired input channel.
Displaying Data2
The FFT of a waveform that has a DC component or offset can cause incorrect FFT waveform
magnitude values. To minimize the DC component, choose AC Coupling on the source
waveform.
To reduce random noise and aliasing components (in repetitive or single-shot waveforms), set
the oscilloscope acquisition mode to averaging.
5 Press Window, and continue pressing the softkey or turn the entry knob to
select the desired window:
There are four FFT windows. Each window has trade-offs between frequency
resolution and amplitude accuracy. What you want to measure and your source
waveform characteristics help determine which window to use. Use the
guidelines in Table 6 to select the best window.
Keysight 1000B Series Oscilloscopes User’s Guide55
Page 56
2Displaying Data
NOTE
Table 6FFT Window Characteristics
WindowCharacteristicsBest for measuring
RectangleBest frequency resolution, worst
magnitude resolution. This is essentially
the same as no window.
Hanning,
Hamming
BlackmanBest magnitude, worst frequency
Better frequency, poorer magnitude
accuracy than Rectangular. Hamming
has slightly better frequency resolution
than Hanning.
resolution.
Transients or bursts, the waveform
levels before and after the event are
nearly equal. Equal-amplitude sine
waves with fixed frequencies.
Broadband random noise with a
relatively slow varying spectrum.
Sine, periodic, and narrow-band
random noise. Transients or bursts
where the waveform levels before and
after the events are significantly
different.
Single frequency waveforms, to find
higher order harmonics.
6 Press Display to toggle between a “Split” screen display and a “Full Screen”
display.
7 Press and turn the entry knob to adjust the vertical position of the
FFT waveform.
8 Press and turn the entry knob to adjust the vertical scale of the
FFT waveform.
9 Press Scale to toggle between “V
” and “dBV
RMS
RMS
” units.
To display FFT waveforms with a large dynamic range, use the dBVrms scale. The dBVrms
scale displays component magnitudes using a log scale.
10 Use the horizontal position knob to adjust the frequency per division.
The frequency scale is displayed on the screen. Use this to display the
frequencies associated with the peaks in the FFT waveform.
56Keysight 1000B Series Oscilloscopes User’s Guide
Page 57
Figure 23FFT Waveform
NOTE
NOTE
Displaying Data2
FFT Resolution
The FFT resolution is the quotient of the sampling rate and the number of FFT points (fS/N).
With a fixed number of FFT points (1024), the lower the sampling rate, the better the
resolution.
Nyquist Frequency and Aliasing in the Frequency Domain
The Nyquist frequency is the highest frequency that any real-time digitizing oscilloscope can
acquire without aliasing. This frequency is half of the sample rate. Frequencies above the
Nyquist frequency will be under sampled, which causes aliasing. The Nyquist frequency is also
called the folding frequency because aliased frequency components fold back from that
frequency when viewing the frequency domain.
Keysight 1000B Series Oscilloscopes User’s Guide57
Page 58
2Displaying Data
NOTE
Using Reference Waveforms
You can save a reference waveform to an internal, nonvolatile memory location
and then display it on the oscilloscope along with other captured waveforms.
You can also export/import reference waveforms to/from an external USB drive
when it is connected to the front panel USB host port.
Reference waveforms are displayed (that is, turned on/off) just like other
waveforms (see page 46).
The reference waveform function is not available in X-Y mode.
To save a reference waveform
1 Before saving a waveform as a reference, set the waveform’s scale and position
as desired.
These settings will become the reference waveform’s defaults.
2 Press [Ref].
3 In the REF menu, press Source, and continue pressing the softkey or turn the
entry knob to select the waveform you want to save.
4 Press Location to choose “Internal”.
5 Press Save.
To export or import reference waveforms
To export or import from external storage (when a USB drive is connected to
the front panel USB host port):
1 Press [Ref].
2 If exporting a waveform, in the REF menu, press Source, and continue pressing
the softkey or turn the entry knob to select the waveform you want to
export.
3 Presh Location to choose “External”.
4 Press Save or Import.
58Keysight 1000B Series Oscilloscopes User’s Guide
Page 59
5 Use the disk manager dialog to navigate to the folder where you want to export
the file or to select the file you want to import (see “To navigate the d irectory
hierarchy” on page 117).
6 In the Save or Import menu:
• To export the waveform, press New File, enter the filename (see “To edit
folder/file names” on page 118), and press Save.
• To load the selected waveform (.wfm file), press Import.
To return the reference waveform to its default scale
1 Press [Ref].
2 In the REF menu, press Reset.
The scale and position of the waveform as originally saved are restored.
Displaying Data2
Keysight 1000B Series Oscilloscopes User’s Guide59
Page 60
2Displaying Data
Changing the Display Settings
Figure 24[Display] Key
To display waveforms as vectors or dots
1 Press [Display].
60Keysight 1000B Series Oscilloscopes User’s Guide
Page 61
2 In the Display menu, press Type to toggle the waveform display between:
VectorsThe oscilloscope connects the sample points by using digital
interpolation.
Digital interpolation maintains linearity by using a sin(x)/x
digital filter. The digital interpolation is suitable for real time
sampling and is most effective at 20 ns or faster horizontal
scale settings.
DotsThe sample points are displayed.
To clear the display
1 Press [Display].
2 In the Display menu, press Clear.
To set waveform persistence
1 Press [Display].
2 In the Display menu, press Persist to toggle the waveform display between:
Displaying Data2
InfiniteSample points remain displayed until the display is cleared or
persistence is set to “OFF”.
OFF
To adjust waveform intensity
1 Press [Display].
2 In the Display menu, press Intensity and turn the entry knob to adjust the
waveform intensity.
To change the grid
1 Press [Display].
2 In the Display menu, press Grid, and continue pressing the softkey or turn the
entry knob to select between:
Keysight 1000B Series Oscilloscopes User’s Guide61
Page 62
2Displaying Data
To adjust the grid brightness
To invert screen colors
Display grid and coordinates on the axes.
Displays coordinates on the axes.
Turns off the grid and coordinates.
1 Press [Display].
2 In the Display menu, press GridBright and turn the entry knob to adjust the
grid brightness.
1 Press [Display] > Preference.
2 In the Display menu, press Screen to toggle the screen between “Normal” or
“Inverted” colors.
Inverted screen colors are sometimes useful when printing or saving screens.
To change the menu display time
The menu display time is how long menus remain on the screen after a front panel
key or softkey has been pressed.
1 Press [Display].
2 In the Display menu, press Menu Display, and continue pressing the softkey or
turn the entry knob to select “1 s”, “2 s”, “5 s”, “10 s”, “20 s”, or “Infinite”
menu display time.
62Keysight 1000B Series Oscilloscopes User’s Guide
Page 63
Keysight 1000B Series Oscilloscopes
User’s Guide
3Capturing Data
Overview of Sampling 64
Choosing the Acquisition Mode 73
Choosing the Acquisition Mode 73
Recording/Playing-back Waveforms 78
Adjusting the Trigger Level 82
Choosing the Trigger Mode 84
Setting Other Trigger Parameters 90
Using the External Trigger Input 94
This chapter describes sampling and acquisition modes and how to set up
triggers.
63
Page 64
3Capturing Data
Overview of Sampling
To understand the oscilloscope’s sampling and acquisition modes, it is helpful to
understand sampling theory, aliasing, oscilloscope bandwidth and sample rate,
oscilloscope rise time, oscilloscope bandwidth required, and how memory depth
affects sample rate.
Sampling Theory
The Nyquist sampling theorem states that for a limited bandwidth (band-limited)
signal with maximum frequency f
must be greater than twice the maximum frequency f
signal be uniquely reconstructed without aliasing.
f
= fS/2 = Nyquist frequency (fN) = folding frequency
MAX
Aliasing
, the equally spaced sampling frequency fS
MAX
, in order to have the
MAX
Aliasing occurs when signals are under-sampled (fS < 2f
distortion caused by low frequencies falsely reconstructed from an insufficient
number of sample points.
). Aliasing is the signal
MAX
64Keysight 1000B Series Oscilloscopes User’s Guide
Page 65
Figure 25Alaising
Capturing Data3
Oscilloscope Bandwidth and Sample Rate
An oscilloscope’s bandwidth is typically described as the lowest frequency at
which input signal sine waves are attenuated by 3 dB (-30% amplitude error).
At the oscilloscope bandwidth, sampling theory says the required sample rate is
f
=2fBW. However, the theory assumes there are no frequency components above
S
f
(fBW in this case) and it requires a system with an ideal brick-wall frequency
MAX
response.
Keysight 1000B Series Oscilloscopes User’s Guide65
Page 66
3Capturing Data
f
S
f
N
-3dB
Attenuation
Frequency
0dB
Figure 26Theoretical Brick-Wall Frequency Response
66Keysight 1000B Series Oscilloscopes User’s Guide
However, digital signals have frequency components above the fundamental
frequency (square waves are made up of sine waves at the fundamental frequency
and an infinite number of odd harmonics), and typically, for 1 Ghz bandwidths and
below, oscilloscopes have a Gaussian frequency response.
Page 67
Figure 27Sample Rate and Oscilloscope Bandwidth
f
S
f
N
fS/4
-3dB
Attenuation
Aliased frequency
components
Frequency
Limiting oscilloscope bandwidth (fBW) to 1/4 the sample rate (fS/4) reduces
frequency components above the Nyquist frequency (f
N
).
0dB
Capturing Data3
So, in practice, an oscilloscope’s sample rate should be four or more times its
bandwidth: f
components have a greater amount of attenuation.
See AlsoEvaluating Oscilloscope Sample Rates vs. Sampling Fidelity: How to Make the
Most Accurate Digital Measurements, Keysight Application Note 1587
=4fBW. This way, there is less aliasing, and aliased frequency
S
Oscilloscope Rise Time
Closely related to an oscilloscope’s bandwidth specification is its rise time
specification. Oscilloscopes with a Gaussian-type frequency response have an
Keysight 1000B Series Oscilloscopes User’s Guide67
approximate rise time of 0.35/f
An oscilloscope’s rise time is not the fastest edge speed that the oscilloscope can
accurately measure. It is the fastest edge speed the oscilloscope can possibly
produce.
based on a 10% to 90% criterion.
BW
Page 68
3Capturing Data
Oscilloscope Bandwidth Required
The oscilloscope bandwidth required to accurately measure a signal is primarily
determined by the signal’s rise time, not the signal's frequency. You can use these
steps to calculate the oscilloscope bandwidth required:
1 Determine the fastest edge speeds.
You can usually obtain rise time information from published specifications for
devices used in your designs.
2 Compute the maximum “practical” frequency component.
From Dr. Howard W. Johnson’s book, High-Speed Digital Design – A Handbook of Black Magic, all fast edges have an infinite spectrum of frequency
components. However, there is an inflection (or “knee”) in the frequency
spectrum of fast edges where frequency components higher than f
knee
are
insignificant in determining the shape of the signal.
f
= 0.5 / signal rise time (based on 10% - 90% thresholds)
knee
f
= 0.4 / signal rise time (based on 20% - 80% thresholds)
knee
3 Use a multiplication factor for the required accuracy to determine the
oscilloscope bandwidth required.
Required
accuracy
20%f
10%fBW = 1.3 x f
3%fBW = 1.9 x f
Oscilloscope bandwidth
required
= 1.0 x f
BW
knee
knee
knee
See AlsoChoosing an Oscilloscope with the Right Bandwidth for your Application, Keysight
68Keysight 1000B Series Oscilloscopes User’s Guide
Page 69
Memory Depth and Sample Rate
The number of points of oscilloscope memory is fixed (except when divided
between channel pairs), and there is a maximum sample rate associated with
oscilloscope’s analog-to-digital converter; however, the actual sample rate is
determined by the time of the acquisition (which is set according to the
oscilloscope’s horizontal time/div scale).
sample rate = number of samples / time of acquisition
For example, when storing 10 us of data in 10,000 points of memory, the actual
sample rate is 1 GSa/s.
Likewise, when storing 1 s of data in 10,000 points of memory, the actual sample
rate is 10 kSa/s.
The actual sample rate, is displayed in the Horizontal menu (see “To view the
sample rate” on page 44).
The oscilloscope achieves the actual sample rate by throwing away (decimating)
unneeded samples.
Capturing Data3
Keysight 1000B Series Oscilloscopes User’s Guide69
Page 70
3Capturing Data
1
1
1
1
N
= Waveform Sample is From
Trigger Time
Choosing the Sampling Mode
The osilloscope can operate in real-time or equivalent-time sampling modes.
You can choose the oscilloscope’s sampling mode in the Acquire menu (accessed
by pressing the [Acquire] front panel button).
To select the real-time sampling mode
In the real-time sampling mode, single waveforms are sampled at uniformly
spaced intervals. See Figure 28.
70Keysight 1000B Series Oscilloscopes User’s Guide
Figure 28Real-Time Sampling Mode
Use the real-time sampling mode when capturing (non-repetitive) single-shot or
pulse waveforms.
The 1000B Series oscilloscopes provide real-time sampling rates up to 500 MSa/s
(when two channels are on) or 1 GSa/s (when one channel is on).
To select the Real-Time sampling mode:
1 Press [Acquire].
2 In the Acquire menu, select Sampling to choose the “Real Time” sampling mode.
In the real-time sampling mode, when the horizontal scale is set to 20 ns or faster,
the oscilloscope uses sine(x)/x interpolation to expand the horizontal time base.
Page 71
To select the equivalent-time sampling mode
1
1
1
1
2
3
4
5
4
4
4
3
3
3
5
5
5
2
2
2
N
= Waveform Sample is From
Trigger Time
D1
D2
D3
D4
In the equivalent-time sampling mode (also known as repetitive sampling),
multiple waveforms are sampled using randomly differing delays from the trigger
to yield higher effective sampling rates.
Capturing Data3
Keysight 1000B Series Oscilloscopes User’s Guide71
In the equivalent-time sampling mode, the effective sample rate can be greater
because the time between samples in the acquisition is shorter.
Equivalent-time sampling mode requires a repetitive waveform with a stable
trigger.
Use the equivalent-time sampling mode to capture repetitive high-frequency
signals at greater sample rates than are available in the real-time sampling mode.
Do not use the equivalent-time mode for single-shot events or pulse waveforms.
The benefits of the equivalent-time sampling mode are negligible when sample
rates are the same as available in the real-time sampling mode.
In the 1000B Series oscilloscopes, the equivalent time sampling mode can achieve
up to 40 ps of horizontal resolution (equivalent to 25 GSa/s).
To select the Equivalent-Time sampling mode:
1 Press [Acquire].
Page 72
3Capturing Data
2 In the Acquire menu, select Sampling to choose the “Equ-Time” sampling mode.
72Keysight 1000B Series Oscilloscopes User’s Guide
Page 73
Choosing the Acquisition Mode
The osilloscope can operate in normal, average, or peak detect acquisition modes.
You can choose the oscilloscope’s acquisition mode in the Acquire menu
(accessed by pressing the [Acquire] front panel key).
Capturing Data3
Figure 30[Acquire] Key
Keysight 1000B Series Oscilloscopes User’s Guide73
Page 74
3Capturing Data
To select the Normal acquisition mode
To select the Average acquisition mode
In the Normal acquisition mode, acquisitions are made and displayed one after the
other.
To select the Normal acquisition mode:
1 Press [Acquire].
2 In the Acquire menu, press Acquisition.
3 Continue pressing the Acquisition softkey or turn the entry knob to select
“Normal”.
In the Average acquisition mode, acquisitions are made, and the running average
over the specified number of acquisitions is displayed.
Use the Average acquisition mode to remove random noise from the waveform
and to improve measurement accuracy.
Figure 31Noisy Waveform Without Averaging
74Keysight 1000B Series Oscilloscopes User’s Guide
Page 75
Figure 32Noisy Waveform With Averaging
Capturing Data3
The Average acquisition mode decreases the screen refresh rate.
To select the Average acquisition mode:
1 Press [Acquire].
2 In the Acquire menu, press Acquisition.
3 Continue pressing the Acquisition softkey or turn the entry knob to select
“Averag e”.
4 Press Averages and turn the entry knob to select the desired number (2, 4, 8,
16, 32, 64, 128, or 256).
To select the Peak Detect acquisition mode
In Normal or Average acquisition modes, at longer horizontal time/div settings,
the oscilloscope’s analog-to-digital converter samples at a rate that yields more
samples than can be stored in a limited amount of oscilloscope memory.
Consequently, samples are thrown away (decimated), and you can miss narrow
excursions on a signal.
Keysight 1000B Series Oscilloscopes User’s Guide75
Page 76
3Capturing Data
However, in the Peak Detect acquisition mode, acquisitions are made at the
fastest sample rate, and the minimum and maximum values for the period
associated with the actual sample rate are stored. This way, you can capture
narrow excursions on a signal at longer horizontal time/div settings.
Figure 33Peak Detect Waveform
Because minimum and maximum values for a sample period are stored, you can
use the Peak Detect acquisition mode to avoid waveform aliasing.
To select the Peak Detect acquisition mode:
1 Press [Acquire].
2 In the Acquire menu, press Acquisition.
3 Continue pressing the Acquisition softkey or turn the entry knob to select
“Peak Detect”.
76Keysight 1000B Series Oscilloscopes User’s Guide
Page 77
To turn OFF/ON sine(x)/x interpolation
When sample points are displayed as vectors (instead of dots) and sine(x)/x
interpolation is on, curved lines are drawn between sample points. When sine(x)/x
interpolation is off, straight lines are drawn.
The effects of sine(x)/x interpolation are only noticeable when the horizontal scale
is set to 20 ns or faster.
1 Press [Acquire].
2 In the Acquire menu, press Sinx/x to turn sine(x)/x interpolation “OFF” or “ON”.
Capturing Data3
Keysight 1000B Series Oscilloscopes User’s Guide77
Page 78
3Capturing Data
Recording/Playing-back Waveforms
You can record waveforms from input channels or from the mask test output, with
a maximum acquisition depth of 1000 frames.
The ability to record mask test output is especially useful for capturing abnormal
waveforms over a long period of time.
To record waveforms
To record waveforms:
1 Press [Acquire].
2 In the Acquire menu, press Sequence.
3 In the Sequence menu, press Mode.
4 Continue pressing the Mode softkey or turn the entry knob to select Record.
To select the source channel for record ing
1 In the Sequence menu ([Acquire] > Sequence > Mode=Record), press Source.
2 Continue pressing the Source softkey or turn the entry knob to select the
desired input channel or the mask test output.
To specify the mask test output, see “To set the mask test output condition” on
page 130.
To select the number of frames to record
1 In the Sequence menu ([Acquire] > Sequence > Mode=Record), press End Frame.
2 Turn the entry knob to select a number from 1 to 1000.
To start/stop recording
1 In the Sequence menu ([Acquire] > Sequence > Mode=Record), press Operate to
start or stop recording.
Appears on the menu when not recording; press Operate to start
recording.
Appears on the menu when recording; press Operate to stop
recording.
78Keysight 1000B Series Oscilloscopes User’s Guide
Page 79
To select the interval between recorded frames
1 In the Sequence menu ([Acquire] > Sequence > Mode=Record), press Interval.
2 Turn the entry knob to select an interval from 1 ms to 1000 s.
To play-back waveforms
To play-back waveforms:
1 Press [Acquire].
2 In the Acquire menu, press Sequence.
3 In the Sequence menu, press Mode.
4 Continue pressing the Mode softkey or turn the entry knob to select
“Play back”.
To play-back/stop the record ing
1 In the Sequence menu ([Acquire] > Sequence > Mode=Play back), press Operate to
play-back or stop the recording.
Capturing Data3
Appears on the menu when not playing-back; press Operate to
start playing-back the recording.
Appears on the menu when playing-back; press Operate to stop
playing-back.
To select continuous or one-time play-back
1 In the Sequence menu ([Acquire] > Sequence > Mode=Play back), press Play Mode
to toggle between:
Continuous play-back.
One-time play-back.
To select the interval between played-back frames
1 In the Sequence menu ([Acquire] > Sequence > Mode=Play back), press Interval.
2 Turn the entry knob to select an interval from 1 ms to 20 s.
Keysight 1000B Series Oscilloscopes User’s Guide79
Page 80
3Capturing Data
To store recorded waveforms
To select the start frame
1 In the Sequence menu ([Acquire] > Sequence > Mode=Play back), press Start Frame.
2 Turn the entry knob to select a number from l to 1000.
To select the current frame
1 In the Sequence menu ([Acquire] > Sequence > Mode=Play back), press
Current Frame.
2 Turn the entry knob to select a number from l to 1000.
To select the end frame
1 In the Sequence menu ([Acquire] > Sequence > Mode=Play back), press End Frame.
2 Turn the entry knob to select a number from l to 1000.
To store recorded waveforms:
1 Press [Acquire].
2 In the Acquire menu, press Sequence.
3 In the Sequence menu, press Mode.
4 Continue pressing the Mode softkey or turn the entry knob to select
“Storage”.
To select the start frame
1 In the Sequence menu ([Acquire] > Sequence > Mode=Storage), press Start Frame.
2 Turn the entry knob to select a number from l to 1000.
To select the end frame
1 In the Sequence menu ([Acquire] > Sequence > Mode=Storage), press End Frame.
2 Turn the entry knob to select a number from l to 1000.
To select internal/external recording storage location
1 In the Sequence menu ([Acquire] > Sequence > Mode=Storage), press Location to
toggle between Internal and External.
80Keysight 1000B Series Oscilloscopes User’s Guide
Page 81
Capturing Data3
InternalRecordings are saved and loaded from oscilloscope internal
memory.
ExternalRecordings are saved, loaded, exported, and imported from an
external USB drive.
To save a recording
1 In the Sequence menu ([Acquire] > Sequence > Mode=Storage), press Save.
2 If the External location has been selected, use the Disk Manager to name and
save the waveform recording file. See “Using the Disk Manager” on page 116.
To load a recording
1 In the Sequence menu ([Acquire] > Sequence > Mode=Storage), press Load.
2 If the External location has been selected, use the Disk Manager to select and
load the waveform recording file. See “Using the Disk Manager” on page 116.
To import/export recordings
1 Because you can only export and import waveform recordings from an external
drive, select the External location. See “To select internal/external recording
storage location” on page 80.
2 In the Sequence menu ([Acquire] > Sequence > Mode=Storage), press Imp./Exp..
3 Use the Disk Manager to select the file and import or export the waveform
recording. See “Using the Disk Manager” on page 116.
Keysight 1000B Series Oscilloscopes User’s Guide81
Page 82
3Capturing Data
Adjusting the Trigger Level
Figure 34Trigger Controls
To adjust the trigger level
• Turn the trigger [Level] knob.
Trigger level knob
Two things happen:
• The trigger level value is displayed at the lower left-hand corner of the
screen.
• A line is displayed showing the location of the trigger level with respect to
the waveform (except when using AC coupling or LF reject coupling modes).
• Push the trigger [Level] knob to set the level at 50% of the signal’s vertical
amplitude.
To force a trigger
To make an acquisition even if no valid trigger has been found:
1 Press [Force Trigger].
Forcing a trigger is useful, for example, when you want to display the DC voltage
of a level signal.
The [Force Trigger] key has no effect if the acquisition is already stopped.
82Keysight 1000B Series Oscilloscopes User’s Guide
Page 83
Capturing Data3
When the oscilloscope’s front panel is locked by a remote program (shown by a
red “Rmt” on the upper-right part of the display), pressing the [Force Trigger] key
returns the front panel to Local control.
Keysight 1000B Series Oscilloscopes User’s Guide83
Page 84
3Capturing Data
Choosing the Trigger Mode
The trigger determines when captured data should be stored and displayed.
When a trigger is set up properly, it can convert unstable displays or blank screens
into meaningful waveforms.
When the oscilloscope starts to acquire a waveform, it collects enough data so
that it can draw the waveform to the left of the trigger point. The oscilloscope
continues to acquire data while waiting for the trigger condition to occur. After it
detects a trigger, the oscilloscope continues to acquire enough data so that it can
draw the waveform to the right of the trigger point.
The oscilloscope provides these trigger modes:
EdgeCan be used with analog and digital circuits. An edge trigger
occurs when the trigger input passes through a specified
voltage level with the specified slope.
PulseIs used to find pulses with certain widths.
VideoIs used to trigger on fields or lines for standard video
waveforms.
AlternateIs used to trigger on non-synchronized signals.
To set up edge triggers
1 Press [Menu].
2 In the Trigger menu, press Mode.
3 Continue pressing the Mode softkey or turn the entry knob to select “Edge”.
4 Then, either push the entry knob or press Mode again.
5 Press Source and continue pressing the softkey or turn the entry knob to
select the waveform to trigger on:
CH1 - CH2The oscilloscope input channel.
EXTThe external trigger input.
AC LineThe AC power line.
84Keysight 1000B Series Oscilloscopes User’s Guide
Page 85
6 Press Slope and continue pressing the softkey or turn the entry knob to
select the edge to trigger on:
Rising edge.
Falling edge.
Both rising and falling edges.
To set up pulse width triggers
A pulse width trigger occurs when a pulse that matches the pulse definition is
found in a waveform.
The width setting can be adjusted from 20 ns to 10 s.
1 Press [Menu].
2 In the Trigger menu, press Mode.
3 Continue pressing the Mode softkey or turn the entry knob to select “Pulse”.
4 Then, either push the entry knob or press Mode again.
5 Press Source and continue pressing the softkey or turn the entry knob to
select the waveform to trigger on:
Capturing Data3
CH1 - CH2The oscilloscope input channel.
EXTThe external trigger input.
6 Press When and continue pressing the softkey or turn the entry knob to
select the type of pulse to trigger on:
Positive pulse greater than the width setting.
Positive pulse less than the width setting.
Negative pulse greater than the width setting.
Negative pulse less than the width setting.
Keysight 1000B Series Oscilloscopes User’s Guide85
Page 86
3Capturing Data
NOTE
To set up video triggers
7 Press Setting and turn the entry knob to adjust the width setting.
Video triggering is used to trigger on fields or lines of NTSC, PAL, or SECAM
standard video waveforms.
When the video trigger mode is selected, the trigger coupling is set to AC.
1 Press [Menu].
2 In the Trigger menu, press Mode.
3 Continue pressing the Mode softkey or turn the entry knob to select “Video”.
4 Then, either push the entry knob or press Mode again.
5 Press Polarity to toggle between:
Normal polarity — trigger on the negative edge of the sync
pulse.
Inverted polarity — trigger on the positive edge of the sync
pulse.
Normal Polarity Sync triggers always occur on negative-going horizontal sync pulses. If the
video waveform has positive-going horizontal sync pulses, use the Inverted Polarity selection.
6 Press Sync and continue pressing the softkey or turn the entry knob to select
what to trigger on:
All LinesTrigger on all lines.
Line NumTrigger on a selected line.
If you select “Line Num”, press the following Line Num menu
item and turn the entry knob to select the line number.
Odd FieldTrigger on an odd field.
Even FieldTrigger on an even field.
7 Press Standard to toggle between:
86Keysight 1000B Series Oscilloscopes User’s Guide
Page 87
NTSCTrigger on an NTSC video waveform.
PAL/ SECAMTrigger on a PAL or SECAM video waveform.
Capturing Data3
Figure 35Line Synchronization
Keysight 1000B Series Oscilloscopes User’s Guide87
Page 88
3Capturing Data
Figure 36Field Synchronization
88Keysight 1000B Series Oscilloscopes User’s Guide
To set up alternate triggers
The Alternate trigger mode splits the display horizontally and lets you trigger on
two, non-synchronized signals.
1 Press [Menu].
2 In the Trigger menu, press Mode.
3 Continue pressing the Mode softkey or turn the entry knob to select
“Alterna te”.
4 Then, either push the entry knob or press Mode again.
5 Press Select to select the channel to set up triggering on, either “CH1” or
“CH2”.
At this point, the remaining items in the Trigger menu let you set up
independent triggers for the selected channel.
For each source, you can set up edge, pulse width, or video triggering. You can
also specify other trigger setup options, except trigger sweep.
Page 89
Figure 37Alternate Triggers
Capturing Data3
Keysight 1000B Series Oscilloscopes User’s Guide89
Page 90
3Capturing Data
Setting Other Trigger Parameters
These are trigger system parameters that apply in all trigger modes.
To set the trigger sweep
Trigger sweep specifies whether acquisitions occur without a triggern or only with
a trigger.
1 Press [Menu].
2 In the Trigger menu, press Sweep.
3 Continue pressing the Sweep softkey or turn the entry knob to select one of
these trigger sweep settings:
AutoAcquire waveform even when no trigger occurs.
NormalAcquire waveform when trigger occurs.
90Keysight 1000B Series Oscilloscopes User’s Guide
To set the trigger coupling
Trigger coupling is used to filter low frequency signal components or DC offsets
from the trigger path when they interfere with achieving stable triggers.
Trigger coupling is similar to channel coupling (see page 47), but it only affects
the triggering system and does not change how the signal is displayed.
To set the trigger coupling:
1 Press [Menu].
2 In the Trigger menu, press Set Up.
3 In the Set Up menu, press Coupling.
Page 91
4 Continue pressing the the Coupling softkey or turn the entry knob to select
one of these trigger coupling settings:
DCSets the trigger coupling to DC.
ACSets the trigger coupling to AC — use for waveforms greater
than 50 Hz.
LF RejectSets the trigger coupling to low frequency reject (10 kHz
cutoff).
To set the trigger high-frequency reject coupling
Trigger high-frequency reject coupling (100 kHz cutoff) is used to filter high
frequency signal components from the trigger path when they interfere with
achieving stable triggers.
To set the trigger high-frequency reject coupling:
1 Press [Menu].
2 In the Trigger menu, press Set Up.
3 In the Set Up menu, press HF Reject to toggle between “ON” and “OFF”.
Capturing Data3
Keysight 1000B Series Oscilloscopes User’s Guide91
Page 92
3Capturing Data
To change the trigger sensitivity
Trigger sensitivity specifies the vertical change that must occur in order for a
trigger to be recognized. In the 1000B Series oscilloscopes, you can adjust the
trigger sensitivity.
For example, to reduce the influence of noise, you can lower the trigger sensitivity
(by increasing the vertical change required to trigger).
To change the trigger sensitivity:
1 Press [Menu].
2 In the Trigger menu, press Set Up.
3 In the Set Up menu, press Sensitivity and turn the entry knob to adjust the
sensitivity setting.
The trigger sensitivity can be adjusted from 0.1 div to 1 div.
92Keysight 1000B Series Oscilloscopes User’s Guide
Page 93
To specify a trigger holdoff
Trigger holdoff can be used to stabilize a waveform. The holdoff time is the
oscilloscope's waiting period before starting a new trigger. The oscilloscope will
not trigger until the holdoff time has expired.
Capturing Data3
Figure 38Trigger Holdoff
To specify a trigger holdoff:
1 Press [Menu].
2 In the Trigger menu, press Set Up.
3 In the Set Up menu, press Holdoff and turn the entry knob to adjust the
holdoff setting.
To reset the trigger holdoff
1 In the Set Up menu, select the Holdoff Reset menu item to return the trigger
holdoff setting to the 100 ns minimum value.
Keysight 1000B Series Oscilloscopes User’s Guide93
Page 94
3Capturing Data
Using the External Trigger Input
You can trigger on external inputs by selecting “EXT” as the trigger source in all
trigger modes except Alternate.
94Keysight 1000B Series Oscilloscopes User’s Guide
Page 95
Keysight 1000B Series Oscilloscopes
User’s Guide
4Making Measurements
Displaying Automatic Measurements 96
Voltage Measurements 98
Time Measurements 101
Counter (Frequency) 106
Making Cursor Measurements 107
This chapter shows how to make automatic voltage measurements, automatic
time measurements, and cursor measurements.
95
Page 96
4Making Measurements
Displaying Automatic Measurements
You can use the [Meas] key to display automatic measurements. The oscilloscope
has 22 automatic measurements and a hardware frequency counter (see “Vol tage
Measurements” on page 98 and “Time Measurements” on page 101).
Figure 39[Meas] Key
96Keysight 1000B Series Oscilloscopes User’s Guide
Page 97
Making Measurements4
To display an automatic measurement
1 Press [Meas].
2 In the Measure menu, press Source to select the input channel on which to
make the automatic measurement.
3 Press Voltage (for voltage measurements) or Time (for time measurements) and
turn the entry knob to select the desired measurement.
4 Then, either push the entry knob or press Voltage or Time again to add the
measurement to the bottom of the display.
If the measurement result is displayed as "*****", the measurement cannot be
performed with the current oscilloscope settings.
A maximum of three measurements can be displayed at the bottom of the display.
When three measurements are displayed and you add a new one, the
measurements shift to the left, pushing the first measurement result off screen.
See Also“To display cursors for automatic measurements” on page 110.
To clear automatic measurements from the display
1 Press [Meas].
2 In the Measure menu, press Clear to clear all automatic measurements from the
display.
To display or hide all automatic measurements
1 Press [Meas].
2 In the Measure menu, press Display All to toggle the display of all automatic
measurements “ON” or “OFF”.
Keysight 1000B Series Oscilloscopes User’s Guide97
Page 98
4Making Measurements
Vbase
Vtop
Overshoot
Vmax
Overshoot
Preshoot
Vpp
Vmin
Vamp
Voltage Measurements
There are 10 automatic voltage measurements:
•Vmax (Maximum Voltage).
• Vmin (Minimum Voltage).
• Vpp (Peak-to-Peak Voltage).
• Vtop (Top Voltage).
• Vbase (Base Voltage).
• Vamp (Amplitude Voltage = Vtop - Vbase).
•Vavg (Average Voltage).
• Vrms (Root-Mean-Square Voltage).
• Overshoot.
• Preshoot.
Figure 40Voltage Measurement Points
Vmax (Maximum Voltage)
98Keysight 1000B Series Oscilloscopes User’s Guide
The maximum amplitude. The most positive peak voltage measured over the entire
waveform. See Figure 40 on page 98.
Page 99
Vmin (Minimum Voltage)
RMS
x
i
2
i1=
n
n
--------------- -=
The minimum amplitude. The most negative peak voltage measured over the
entire waveform. See Figure 40 on page 98.
Vpp (Peak-to-Peak Voltage)
Peak-to-peak voltage. See Figure 40 on page 98.
Vtop (Top Voltage)
Voltage of the waveform's flat top, useful for square and pulse waveforms. See
Figure 40 on page 98.
Vbase (Base Voltage)
Voltage of the waveform's flat base, useful for square and pulse waveforms. See
Figure 40 on page 98.
Making Measurements4
Vamp (Amplitude Voltage = Vtop - Vbase)
Voltage between Vtop and Vbase of a waveform. See Figure 40 on page 98.
Vavg (Average Voltage)
The arithmetic mean over the entire waveform.
Vrms (Root-Mean-Square Voltage)
The true root-mean-square voltage over the entire waveform.
Where:
x
= value at ith point.
i
n = number of points.
Keysight 1000B Series Oscilloscopes User’s Guide99
Page 100
4Making Measurements
Overshoot
Preshoot
Defined as (Vmax-Vtop)/Vamp, useful for square and pulse waveforms. See
Figure 40 on page 98.
Defined as (Vmin-Vbase)/Vamp, useful for square and pulse waveforms. See
Figure 40 on page 98.
100Keysight 1000B Series Oscilloscopes User’s Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.