Kewtech KEW2200 User Manual

Page 1
Data Hold Key
Mode Key
KEW 2200 Instruction Manual
Current Sensor
Trigger
CAT. III
600V
V
OFF
1000A
V
Function Switch
LED
HOLD
Lanyard
KEW 2200
digital clamp meter
V/
COM
CAT. II 600V CAT. III 300V
kewtechcorp.com
kewtechcorp.com
V/ΩTerminal
COM Terminal
Page 2
Barrier of Instrument body
p
Ca
Barrier of Test Leads
WARNING
Read through and understand the
instructions contained in this manual
before using the instrument.
1. Specification
Accuracy guaranteed 100% or less of range Temperature 23 ± 5°C
Humidity 45 - 75%
AC A Auto Range
Range Accuracy
40A 0.00, 0.03-41.99A ±1.4 %rdg±6dgt
400A 32.0 -419.9A
1000A 320-1049A
Input protective current AC 1200A
AC V Auto Range
Range Accuracy
4V 40V 3.20-41.99V 400V 32.0 -419.9 V 600V 320-629V
0.000, 0.005- 4.199V
DC V Auto Range
Range Accuracy
400mV ±0.0 -± 419.9mV *1 4V ±0.320 - ±4.199V 40V ±3.20-±41.99V 400V ±32.0 - ±419.9V 600V ±320-±629V
*1 : Accuracy is not guaranteed
AC V/DC V Input impedance >100MΩ (400mV Range) 11MΩ (4V Range) 10MΩ (40/400/600V Range)
(50/60Hz)
±1.6 %rdg±6dgt
(45- 65Hz)
±1.8 %rdg±7dgt
(45- 65Hz)
±2.3 %rdg±8dg t
(65-500Hz)
±1.0%rdg±3dgt
kewtechcorp.com
Page 3
Resistance Auto Range / Continuity
Range Accuracy
400Ω 0. 0-419.9Ω 4kΩ 0.320 -4.199 kΩ 40kΩ 3.20 - 41.9 9 k Ω 400kΩ 32.0 -419.9 kΩ 4MΩ 0.320-4.199 MΩ ±4.0%rdg±4dgt 40MΩ 3.20-41.99 MΩ ±8.0%rdg±4dg t
Cont. 0.0-419.9Ω
±2.0%rdg±4dgt
Buzzer threshold value
50±30Ω
Open-loop voltage
<3.4V (400Ω / Cont Range)
0.7V typ (4kΩ Range)
0.47V typ (40k - 40MΩ Range)
Input protective voltage
AC/DC 600V 10 sec
Measuring method
Dual integration
Over-range indication
OL
Measurement cycle
2.5 times per second
Applicable Standards
• IEC/EN 61010-1/ 61010-2-032/ 61010-031 /
61010-2-030
Pollution degree 2
Indoor use
Altitude up to 2000m
Current measurement section CAT.III 600V
Voltage measurement section CAT.II 600V /
CAT.III 300V
EN61326 (EMC)
In the radio-frequency electromagnetic field of
3V/m, accuracy is within five times the rated accuracy.
EN50581 (RoHS)
kewtechcorp.com
Page 4
Withstand voltage
AC 5320Vrms 5sec between current sensor and enclosure
AC 3540Vrms 5sec between circuit and enclosure
Insulation resistance
>100MΩ /1000V between enclosure and electrical circuit
Operating temperature and humidity range 0 to 40°C 85%RH or less (no condensation)
Storage temperature and humidity range
-20 to 60°C 85%RH or less (no condensation)
Power source
DC3V×2 / R03/LR03 (AAA) ×2
Current consumption < 3mA
Battery life Approx. 350 hours (AC A, continuous, no load)
Dimension, Weight 190(L)×68(W)×20(D)mm
approx. 120g (including batteries)
Accessories
Test leads Model 7107A 1set Battery R03(AAA) 2pcs Instruction manual 1pce Carrying case Model 9160
1pce
2. Other Function
Data Hold
Press the Data Hold Key to freeze the reading. Press the Data Hold Key again to release the display.
Indicated“ ”on LCD
Low battery indication
Indicated “
Sleep Function
Automatically powers off after about 10min of inactivity. To disable the sleep function, power the instrument on with the Data Hold Key pressed.
(indicated " " for about 2 seconds on LCD)
” on LCD at 2.3±0.15V or less
kewtechcorp.com
Page 5
Correct
Incorrect
3. AC A Measurement
DANGER
Never measure current while the test leads
are inserted into the V/ and/or COM
Terminals.
Press the trigger to open the Current Sensor and clamp the single conductor (Dia. 33mm max.) under test.
Correct Incorrect
KEW 2200
KEW 2200
digital clamp meter
NOTE
Measurement accuracy is guaranteed when the
measured object is placed at the centre (zone A)
of the Current Sensor. In zone B, 4% of tolerance
should be added to the specified accuracy. In
zone C, measured values should be considered as
reference values (Accuracy is not guaranteed).
digital clamp meter
C
B
B
A
C
kewtechcorp.com
Center (zone A)
Centre (zone A)
Page 6
4. AC V/DC V Measurement
Black test lead
ACV
Red test lead
DC
V
Black(-)
Red(+)
digital clamp meter
KEW 2200
digital clamp meter
KEW 2200
DANGER
600V
Never make measurement on a circuit in
which voltage over 600V exists.
AC V DC V
kewtechcorp.com
Page 7
5.
Resistance (Continuity) Measurement
WARNING
Never use the instrument on an
energized circuit.
Test lead
Resistance
KEW 2200
digital clamp meter
Press
/ Press
Test lead
Continuity
LCD
KEW 2200
digital clamp meter
Beeps when less than 50
kewtechcorp.com
±30.
Page 8
6. Safety Warnings
This instrument has been designed, manufactured and tested according to IEC 61010: Safety requirements for Electronic Measuring apparatus, and delivered in the best condition after having passed final inspection. This instruction manual contains warnings and safety rules which must be observed by the user to ensure safe operation of the instrument and retain it in a safe condition. Therefore, read through these operating instructions before using the instrument.
WARNING
• Read through and understand the
instructions contained in this manual before using the instrument.
• Keep the manual at hand to enable quick
reference whenever necessary.
• The instrument is to be used only in its
intended applications.
• Understand and follow all the safety
instructions contained in the manual.
• It is essential that the above instructions
are adhered to.
• Failure to follow the above instructions
may impair the protection provided by the instrument and test leads, and may cause injury, instrument damage and/or damage to equipment under test.
The symbol indicated on the instrument means that the user must refer to the related parts in the manual for safe operation of the instrument. It is essential to read the instructions wherever the symbol appears in the manual.
DANGER is reserved for conditions and actions that are likely to cause serious or fatal injury.
WARNING is reserved for conditions and actions that can cause serious or fatal injury.
CAUTION is reserved for conditions and actions that can cause injury or instrument damage.
kewtechcorp.com
Page 9
Marks listed below are used on this instrument.
User mus t refer to the manual.
Instrument with double or reinforced insulation
Indicates that this instrument can clamp on bare conductors when measuring a voltage corresponding to the applicable measurement category, which is marked next to this symbol.
AC
DC
Ground (Earth)
This instrument is subject to WEEE Directive (2002/96/EC). Please contact our dealer near you at disposal.
Measurement Category
CAT.II
Primary electrical circuits of equipment connected to an AC electrical outlet by a power cord.
CAT.III
Primary electrical circuits of the equipment connected directly to the distribution panel, and feeders from the distribution panel to outlets.
CAT.IV
The circuit from the service drop to the service entrance, and to the power meter and primary over current protection device (distribution panel).
Current measurement section of this instrument is designed for CAT.III 600V and Voltage measurement section is for CAT.III 300V / CAT.II 600V respectively.
Test leads 7107A with the Cap is designed for CAT.IV 600V / CAT.III 1000V and without the Cap is for CAT.II 1000V.
DANGER
• Never make measurement on a circuit in
which voltage over AC/DC600V exists.
• Do not attempt to make measurement in the
presence of flammable gasses. Otherwise, the use of the instrument may cause sparking, which can lead to an explosion.
• Never attempt to use the instrument if its
surface or your hand is wet.
kewtechcorp.com
Page 10
• Do not exceed the maximum allowable input of
any measuring range.
• Never open the Battery cover during a
measurement.
• To avoid electrical shock by touching the
equipment under test or its surroundings, be sure to wear insulated protective gear.
• Never measure current while the test leads are
inserted into the input terminals.
• Barriers on the instrument body and the test
leads provide protection to keep your fingers and hands from touching an object under test.
• Keep your fingers and hands behind the
barriers during measurement.
WARNING
• Never attempt to make measurement if any
abnormal conditions, such as broken case and exposed metal parts are found on the instrument or test leads.
• Verify proper operation on a known source
before use or taking action as a result of the indication of the instrument.
Firmly attach the Caps to the test leads
when performing measurements at CAT. III or higher test environment.
When the KEW2200 and the test leads
are combined and used together, whichever is the lowest category & voltage to earth specification applies.
• Do not rotate the Function Switch while the
test leads are being connected.
Do not install substitute parts or make any modification to the instrument. For repair or re-calibration, return the instrument to your local distributor from where it was purchased.
CAUTION
• Use of this instrument is limited to domestic,
commercial and light industry applications.
If equipment generating strong electromagnetic Interference or strong magnetic fields due to large currents exist nearby, the instrument may not work correctly.
Set the Function Switch to an appropriate position before starting measurement.
Firmly insert the test leads.
The LCD may show some readings at the
ACV and the DCV ranges even while the test leads are open. And, it may show some digits
kewtechcorp.com
Page 11
instead of 0 when short-circuiting the test leads. However, these phenomena doesn’t affect measurement results.
This instrument isn’t dust & waterproof.
Keep away from dust and water.
Be sure to power off the instrument after use.
When the instrument will not be in use for a long period, place it in storage after removing the batteries.
Do not expose the instrument to the direct sun, high temperature and humidity or dewfall.
• Use a slightly damp cloth or neutral
detergent for cleaning the instrument. Do not use abrasives or solvents.
7. Battery Replacement
WARNING
• Replace the batteries when a Low Battery
Voltage warning " " mark(< 2.3±0.15V)
is indicated on the LCD. Otherwise, precise
measurement cannot be made. Note that
when the battery is completely exhausted, the LCD goes blank without showing " " mark.
• Do not try to replace the batteries if the
surface of the instrument is wet.
• Disconnect the test leads from the object
under test and power off the instrument before opening the Battery Compartment Cover for Battery replacement.
CAUTION
• Do not mix old and new batteries.
• Install batteries in correct polarity as
indicated in the Battery Compartment.
(1) Set the Function Switch to "OFF" position.
(2) Unscrew and remove the Battery
Compartment Cover on the bottom of the instrument.
(3) Replace the batteries observing correct
polarity. Use new two
R03/LR03 (AAA)
1.5V batteries. (4) Install the Battery
Compartment and tighten the screw.
kewtechcorp.com
Page 12
Certificate
of Conformity &
Warranty Registration
This instrument has been calibrated
using equipment which has itself been
calibrated to standards traceable to
International Standards monitored by
BIPM (International Bureau of Weights
and Measures)
This cer tificate guarantees that the
product has been fully inspected and
conforms to all the relevant published
specifications.
Free Two Year Warranty
Kewtech's Two Year Warranty enhances the customers' legal rights/ it covers all
manufacturing defects for a t wo year
period but Kewtech reserves the right to
exclude abuse or accidental damage.
To register your free guarantee simply go
to kewtechcorp.com - the link is on the
home page.
Re-calibration Service
Regular re-calibration is recommended
for this instrument. Kewtech recommends
that with normal use the instrument
is calibrated at least once in every 12
month interval.
When the instrument is due for re-calibration return it to the address below marked for the attention of the
Calibration Department.
Kewtech Corp Ltd
Shaw Wood Business Park
Shaw Wood Way
Doncaster DN2 5TB
t: 01302 761044
kewtechcorp.com