Kewaunee INT-1100A2, INT-1100A2-230, INT-1400A2, INT-1400A2-230, INT-1700A2 User guide

...
Page 1
®
Interceptor
Biological Safety Cabinet
Class II A2
Maintenance and Technical Manual
INT-1100A2 & INT-1100A2-230 INT-1400A2 & INT-1400A2-230 INT-1700A2 & INT-1700A2-230 INT-2000A2 & INT-2000A2-230
INT-1100A2-1 & INT-1100A2-1-230 INT-1400A2-1 & INT-1400A2-1-230 INT-1700A2-1 & INT-1700A2-1-230 INT-2000A2-1 & INT-2000A2-1-230
KEWAUNEE
...encouraging new discovery
®
BSCMT-A2-0717
Page 2
Page 3
Table of Contents
CHAPTER 1
Technical Specications 1.1
CHAPTER 2
Replacement & Repair of Key Parts 2.1 Air Chamber Maintenance Item Locations 2.2 Filter Loading 2.3 Fan Failure 2.4 Fan Removal and Replacement 2.5 Pressure Transducer Failure and Replacement 2.6
CHAPTER 3
Decontamination and Repairs 3.1 Filter Replacement Instructions 3.2 Face Velocity Sensor Replacement 3.4 Thimble Connection 3.5 Parts List 3.6
CHAPTER 4
Cabinet Operation 4.1 Startup Procedure 4.3 Air Flow Patterns 4.4 Recommended Workow 4.5 NightSetback/Low Flow Mode 4.5 Control Monitor 4.6 UV Light 4.6 Control Monitor Messages and Meanings 4.7 Control Monitor Screens & Menu Diagrams 4.8
CHAPTER 5
Troubleshooting 5.1 Cabinet Calibration Instructions 5.2 Wiring Schematics 5.4
Page 4
NOTICE:
The Interceptor® Biological Safety Cabinet offers many features to enhance performance, safety, and operator comfort.
Due to the nature of work performed in a Biological Safety Cabinet, it is very important to read the User and Operation Manual and Maintenance and Technical Manual and follow standard operating procedures to avoid infection and other potential injuries.
If this equipment is used in a manner not specied by the manufacturer in this manual, the protection provided by the equipment may be impaired.
®
Also, any maintenance or service to an Interceptor to the instructions contained herein. Maintenance of this product shall be carried out by technicians trained in the mechanical details of this unit.
! WARNING
ê
Class II A2 Biological Safety cabinets must be certified before initial use, after being moved, and after any service, including required annual recertification. Service must be performed by an NSF accredited certifier using NSF/ ANSI 49 criteria and should include, at minimum, the following test:
• Downflow Velocity Profile Test
• Inflow Velocity Test
• Airflow Smoke Pattern Test
• HEPA Filter Leak Test
product must be done according
Page 5
Maintenance and Technical Manual Chapter 1
CHAPTER 1
Technical Specications for Interceptor
Product Number
INT-1100A2/INT-1100A2-1
120 VAC
INT-1100A2-230/ INT-1100A2-1-230
230 VAC
INT-1400A2/INT-1400A2-1
120 VAC
INT-1400A2-230/ INT-1400A2-1-230
230 VAC
INT-1700A2/INT-1700A2-1
120 VAC
INT-1700A2-230/ INT-1700A2-1-230
230 VAC
INT-2000A2/INT-2000A2-1
120 VAC
INT-2000A2-230/ INT-2000A2-1-230
230 VAC
Width of
Cabinet
1100mm
43.3"
1400mm
55.1"
1700mm
66.9"
2000mm
78.7"
Height of
Cabinet
1610mm
63.4"
1610mm
63.4"
1610mm
63.4"
1610mm
63.4"
Depth of
Cabinet
813mm
32"
813mm
32"
813mm
32"
813mm
32"
Fan
Max
HP
0.9
0.9
0.9
0.9
Fan & Light
Max Current/
Power
10.6 Amps 1275 VA
5.2 Amps 1275 VA
10.6 Amps 1275 VA
5.2 Amps 1275 VA
10.6 Amps 1275 VA
5.2 Amps 1275 VA
10.6 Amps 1275 VA
5.2 Amps 1275 VA
®
Outlet
Max Amps
/HZ
9 Amps
60hz
4.5 Amps 50/60hz
9 Amps
60hz
4.5 Amps 50/60hz
9 Amps
60hz
4.5 Amps 50/60hz
9 Amps
60hz
4.5 Amps 50/60hz
Exhaust
CFM
272
358
444
530
Table 1.1: Product Descriptions
NOTES:
All power values measured at 10 inch operating sash height.
The powered base stand, if ordered, will take an additional 8.6 amps at 120 VAC; 60 Hz or 4.3 amps at 230 VAC; 50 Hz.
The electrical outlets inside the Interceptor® are grounded. This is particularly important since all internal surfaces are stainless steel and conduct electricity. Under NO CIRCUMSTANCES use ungrounded plugs in these outlets. It is not recommended to use in excess of 1500 watts of power. Exterior power plugs must not be removed until unit fan and lights are turned off. The unit is to be disconnected from the main voltage by unplugging both plugs to remove power. For electrically-powered base stands, power for disconnect is also accomplished by plug removal from a waste-high or lower plug outlet.
The Interceptor® Cabinet has one power cord. If the Power Base is employed, a total of two power cords are used. When positioning the BSC, always connect BSC and stand power plugs in waste-high or oor-positioned outlets to facilitate disconnection in an emergency. Never block these outlets.
If UV option is on your BSC, be sure safety overrides are never immobilized! UV lamp should NEVER be on while sash is open.
continued on page 1.2
Interceptor Biological Safety Cabinet Class II A2
July 2017
1.1
Page 6
Chapter 1 Maintenance and Technical Manual
Notes: (continued)
Based on the following UL denitions, the Interceptor® may be used in a room with pollution degree 1 or pollution degree 2 conditions:
Pollution Degree 1 No pollution or only dry, non-conductive pollution occurs. The pollution has no inuence.
Pollution Degree 2 Normally only non-conductive pollution occurs. Occasionally, a temporary conductivity
caused by condensation must be expected.
Pollution Degree 3 Conductive pollution occurs, or dry, non-conductive pollution occurs which becomes
conductive, which is expected.
Be aware that the dimensions of a Biological Safety Cabinet may exceed doorway limitations.
Sash Face Opening Chart
Model Height Width Area
INT-1100
INT-1400
INT-1700
INT-2000
Table 1.2
10" 37.3" 2.59 sq ft
10" 49.1" 3.41 sq ft
10" 60.9" 4.23 sq ft
10" 72.7" 5.47 sq ft
1.2
Interceptor Biological Safety Cabinet Class II A2
July 2017
Page 7
Maintenance and Technical Manual Chapter 2
CHAPTER 2
Replacement and Repair of Key Parts
Serial Number
Screw Hole
(both screws must
be removed to open
Top Front Panel)
Air
Chamber
Panel
Figure 2.1 BSC with Top Front Panel open
Conditions 1, 2 and 3 mentioned in this section are all behind the Air Chamber Panel, which is accessible by lifting the Top Front Panel. Before opening the Air Chamber Panel, a complete decontamination of the upper plenum area is required. This decontamination step is essential since the lters and surfaces of the upper chamber will be contaminated.
Factory Seal
Screw Hole
Top Front Panel
(shown in open position)
! WARNING
ê
Opening the Air Chamber Access Panel requires a complete decontamination of the upper plenum area. Decontamination may only be performed by an NSF Accredited Technician.
Interceptor Biological Safety Cabinet Class II A2
July 2017
2.1
Page 8
Chapter 2 Maintenance and Technical Manual
Air Chamber Maintenance Item Locations
Control Board
(on back of Top Front
Exhaust
Filter
Plenum
Access
Panel
(removal required for fan replacement only)
Panel)
Fan
Supply
Filter
Perforated
Panel
Figure 2.2 BSC with Air Chamber Access Panel removed
Pressure Transducer
Conditions Requiring Removal of the Air Chamber Access Panel
1) Filter Replacement
Loading of the lters to a level where an inow face velocity of at least 100 FPM cannot be attained.
2) Fan Failure
Fan fails to operate.
3) Pressure Transducer Failure
The control panel displays Sensor Error and/or erratic velocity readings.
! WARNING
ê
Opening the Air Chamber Access Panel requires a complete decontamination of the upper plenum area. Decontamination may only be performed by an NSF Accredited Technician.
2.2
Interceptor Biological Safety Cabinet Class II A2
July 2017
Page 9
Maintenance and Technical Manual Chapter 2
Filter Loading
Over time, as the Interceptor’s lters load up, inow velocity shown on the control screen will begin to drop. NSF 49 recommends that motor RPM be increased when this occurs until the control unit again shows 105 FPM.
Before considering lter replacement, the motor speed control should be adjusted to increase the air ow. Before considering lter replacement, the fan speed and top damper should be adjusted to increase the air ow. The Fan Speed adjustment is located under Settings|Cabinet Setup on the Control Monitor. Select Cabinet Setup from the menu. Fan Speed is a password protected feature. Additionally you can adjust the top damper and increase the fan speed to attain 105 FPM inow velocity. Check the exhaust/downow balance after such an adjustment is made. The steps must be performed by an NSF Certied Technician.
Figure 2.3a
Menu Screen
Figure 2.3b
Cabinet Setup Screen
Figure 2.3c
Fan Speed Screen
Figure 2.3d
Set Normal Fan Speed Screen
Filters should be changed when they load up to the point that further adjustment of the motor RPM will not increase face velocity to 100 FPM. NSF accepts a range of inow between 100-110 FPM at operating height (10" for the Interceptor).
When the static pressure difference between the negative and positive pressure plenums reaches a certain level (generally between 1.4" and 1.8" wg depending on BSC size and calibration), the net inow will continue to drop over time and no further adjustment of the motor voltage will bring the face velocity back to 105 FPM.
The Filters should be replaced at this time. An inability to maintain 100 FPM mandates lter change-out according to NSF 49.
See Chapter 3 for Filter Replacement procedures.
Interceptor Biological Safety Cabinet Class II A2
July 2017
2.3
Page 10
Chapter 2 Maintenance and Technical Manual
Fan Failure
Fan failure is uncommon. First check the Blower Speed Control and the Power Supplies which are located in the top junction box and can be replaced by a qualied electrician without opening the Air Chamber Access Panel.
Blower Speed Control
24 VDC
Power Supply
24 VAC
Power Supply
Circuit Breakers Power Cord
Figure 2.5 Electrical Junction Box – Blower Speed Control & Power Supplies
Once the Blower Speed Control and Power Supplies are veried and found normal, the Air Chamber Access Panel must be removed to inspect/replace the Fan.
! WARNING
ê
Opening the Air Chamber Access Panel requires a complete decontamination of the upper plenum area. Decontamination may only be performed by an NSF Accredited Technician.
2.4
Interceptor Biological Safety Cabinet Class II A2
July 2017
Page 11
Maintenance and Technical Manual Chapter 2
Fan Removal and Replacement
Once the BSC is decontaminated and the Air Chamber Access Panel is removed, removal and replacement of the fan requires the unbolting of the fan from inside the Plenum Chamber as outlined in the steps below:
Plenum Access
Cover Panel
(removal required for fan replacement only)
Fan Bolts
(access from inside Plenum)
(nuts are attached to fan frame)
Fan Bolts
Color-coded Wires
FanFan Gasket
Figure 2.6 Attachment of Fan to Inner Plenum Chamber – (Model INT-1400 shown in photo)
1) Remove the exhaust lter (see lter replacement instruction)
2) Remove the Color-coded Wires from the Fan.
3) Remove the Plenum Access Cover Panel.
4) Remove the Bolts holding Fan in place. The two lower Bolts are accessible from inside the Plenum Chamber.
5) Carefully dislodge the Fan from its cushioned Gasket and remove from the Plenum.
6) Inspect the Fan mounting surface on the Plenum and remove any Gasket material that may have stuck to the surface.
6) Install replacement Fan by reversing the steps above. (Make sure new Fan has been tted with attachment nuts and new Gasket).
Replacement fans are available from Kewaunee Scientic Corporation. See
page 3.7 for part numbers.
Interceptor Biological Safety Cabinet Class II A2
July 2017
2.5
Page 12
Chapter 2 Maintenance and Technical Manual
Pressure Transducer Failure
Check the security of the connections at indicated locations on the BSC Control Board. If the connections are secure, proceed with replacing the transducer.
Pressure
Transducer
Connection
Figure 2.7 BSC Control Board with cover removed – located on back of Front Panel
Pressure Transducer Replacement
After removing the Air Chamber Access Panel, the Pressure Transducer is located in the far right hand side of the Air Chamber. Replacement requires disconnecting, then reconnecting all wires and the pressure hose.
! WARNING
ê
Opening the Air Chamber Access Panel requires a complete decontamination of the upper plenum area. Decontamination may only be performed by an NSF Accredited Technician.
Pressure Transducer Location
Figure 2.8 Pressure Transducer Location Figure 2.9 Pressure Transducer
2.6
Interceptor Biological Safety Cabinet Class II A2
July 2017
Page 13
Maintenance and Technical Manual Chapter 3
CHAPTER 3
Decontamination / Repairs
Decontamination
Upon leaving the factory, the BSC Control Monitor was calibrated and fan voltage adjusted to the proper values:
Figure 3.1 Control Monitor – Info Screen Figure 3.2 Control Monitor – Face Velocity 105 FPM
When new, the Interceptor® Biological Safety Cabinet Control Panel shows key values on the display. Remaining Filter Capacity should be 100%; Face Velocity should read 105 FPM (+- 2 FPM).
After much use, these values (remaining Filter Capacity and Face Velocity) will both decrease. When this inevitable change occurs, some adjustment of motor RPM can restore face velocity, but will not restore the lter capacity readout. (see Chapter 2)
Below the 50% lter capacity level, it may no longer be possible to achieve 105 FPM by simply raising the motor RPM. Since NSF 49 requires the minimum operating average face velocity be between 100 FPM and 110 FPM, velocity readings below 100 FPM require lter replacement.
Since both lters lie inside the contaminated Air Chamber, the BSC must be decontaminated before opening up the panel to replace these lters.
! WARNING
ê
Opening the Air Chamber Access Panel requires a complete decontamination of the upper plenum area. Decontamination may only be performed by an NSF Accredited Technician.
Interceptor Biological Safety Cabinet Class II A2
July 2017
3.1
Page 14
Chapter 3 Maintenance and Technical Manual
Filter Replacement
Exhaust Filter
Exhaust
Filter
Scissor
Clamp
Supply Filter Scissor Clamp
Figure 3.3 Filter Replacement Front View
Supply Filter
Scissor Clamp
Supply Filter
Exhaust Filter Scissor Clamp
Exhaust Filter
Supply Filter
3.2
Figure 3.4 Filter Replacement End View
Interceptor Biological Safety Cabinet Class II A2
July 2017
Page 15
Maintenance and Technical Manual Chapter 3
Supply and Exhaust Filter Removal Instructions
1. Lower sash to lowest position
2. Using a 5/16" hex driver, loosen and remove all hex head bolts and then remove access panel.
Exhaust Filter Removal/Replacement
1. Using a 1/2" socket, alternately loosen the two Exhaust Filter Scissor Clamps until Exhaust Filter can be easily removed by sliding forward.
2. Insert new Exhaust Filter and insure proper alignment.
3. Using the 1/2" socket, alternately tighten the two Exhaust Filter Scissor Clamps until the Exhaust Filter is securely contained.
Supply Filter Removal/Replacement
1. Using a 1/2" socket, alternately loosen the two Exhaust Filter Scissor Clamps allowing the Exhaust Filter to drop down 1/2 inch.
2. Using the 1/2" socket, alternately loosen the two Supply Filter Scissor Clamps until the Supply Filter can be easily removed by sliding forward.
3. Insert a new Supply Filter and insure proper alignment.
4. Using the 1/2" socket, alternately tighten the two Supply Filter Scissor Clamps until the Exhaust Filter is securely contained.
5. Using the 1/2" socket, alternately tighten the two Exhaust Filter Scissor Clamps until the Exhaust Filter is securely contained.
Interceptor Biological Safety Cabinet Class II A2
July 2017
3.3
Page 16
Chapter 3 Maintenance and Technical Manual
Replacement of Face Velocity Sensor
The thermal velocity sensor, located in the exhaust output box at the BSC top left, is designed to measure exhaust velocity and translate it into average face velocity at a 10" opening. If the velocity readout starts being irregular, the thermal sensor may need replacement.
Figure 3.5 Face Velocity Sensor
Be sure the sensor window openings are perpendicular to the exhaust plane. Repair and/or replacement of the thermal sensor can be done without decontamination by replacing the sensor with a new unit. Once this has been done, velocity must be recalibrated using the controller calibration program outlined in Chapter 5.
3.4
Interceptor Biological Safety Cabinet Class II A2
July 2017
Page 17
Maintenance and Technical Manual Chapter 3
Thimble Connection (Canopy)
An Interceptor® Class II, A2 Biological Safety Cabinet can be used with minute quantities of volatile toxic chemicals or trace amounts of radionuclides when properly ducted through a Thimble Connection with Airow Sensor. The Thimble mounts to the top of the exhaust lter guard with four (4) screws and connects to either a 8" or 10" exhaust duct. The Airow Sensor connects to the cabinet’s monitor to warn when the exhaust ventilation is not operating.
Figure 3.6 Thimble Connection with Airow Sensor (Canopy)
Thimble Exhaust requirements
The building ventilation system must be veried that it can handle the required exhaust volumes of the cabinet before connection. Thimble Exhaust CFM should be
1.3 times (130%) of the Biological Safety Cabinet exhaust. (See instructions shipped
with Thimble Connection for proper installation and connection procedures.)
Table 3.3
! WARNING
ê
A hood connected to a building ventilation system with a Thimble Connection may become dysfunctional if the ventilation system fails. Check building ventilation system before using cabinet!
Size of Biological
Safety Cabinet
1100mm 272 354
1400mm 358 465
1700mm 444 577
2000mm 530 689
Biological Safety
Cabinet CFM
Required Thimble
Exhaust CFM
Interceptor Biological Safety Cabinet Class II A2
July 2017
3.5
Page 18
Chapter 3 Maintenance and Technical Manual
Replaceable Parts List
Part Numbers
Parts INT-1100 INT-1400 INT-1700 INT-2000
24 VAC Transformer F-7424-00 F-7424-00 F-7424-00 F-7424-00
24 VDC Power Supply F-7241-00 F-7241-00 F-7241-00 F-7241-00
Motor Controller F-7421-00 F-7421-00 F-7421-00 F-7421-00
Micro Switch F-7242-00 F-7242-00 F-7242-00 F-7242-00
Fluorescent / UV Relay F-7232-00 F-7232-00 F-7232-00 F-7232-00
20 Amp Circuit Breaker F-7240-00 F-7240-00 F-7240-00 F-7240-00
10 Amp Circuit Breaker F-7240-04 F-7240-04 F-7240-04 F-7240-04
0.3 Amp Circuit Breaker F-7430-00 F-7430-00 F-7430-00 F-7430-00
UV Lamp F-7249-18 F-7249-00 F-7249-00 F-7249-00
T5 Lamp F-6347-34 F-6347-46 F-7248-00 F-7248-00
Removable Work Tray 050007-36 050007-48 050007-60 050007-72
Work Tray Chrome Knob F-1145-00 F-1145-00 F-1145-00 F-1145-00
Bottom Intake Grille 052008-36 052008-48 052008-60 052008-72
Sash Glass with Handles Assembly 050009-36-FIN 050009-48-FIN 050009-60-FIN 050009-72-FIN
Exhaust HEPA filter F-7275-36-FIN F-7275-48-FIN F-7275-60-FIN F-7275-72-FIN
Supply HEPA filter F-7276-36-FIN F-7276-48-FIN F-7276-60-FIN F-7276-72-FIN
Removable Work Tray Support 052063-00 052063-00 052063-00 052063-00
Sash Sweeper Plate 052076-36 052076-48 052076-60 052076-72
Top Exhaust Balance Damper 052077-00 052077-00 052077-00 052077-00
Damper Telescoping Section 050166-00 050166-00 050166-00 050166-00
Downflow Perforated Grille 050079-36 050079-48 050079-60 050079-72
Arm Rest Bracket 050195-0A 050195-0A 050195-0A 050195-0A
Arm Rest Plastic End Cap F-7279-00 F-7279-00 F-7279-00 F-7279-00
Aluminum Arm Rest F-7280-36 F-7280-48 F-7280-60 F-7280-72
Drain Valve Plate 050110-00 050110-00 050110-00 050110-00
316 SS Pipe Fitting F-7211-11 F-7211-11 F-7211-11 F-7211-11
316 SS Ball Valve F-7211-00 F-7211-00 F-7211-00 F-7211-00
Safety Labels 120 VAC F-8200-01 F-8200-01 F-8200-01 F-8200-01
Pressure Transducer F-7433-00 F-7433-00 F-7433-00 F-7433-00
Velocity Sensor F-7425-00 F-7425-00 F-7425-00 F-7425-00
Fan Assembly F-7420-00 F-7420-00 F-7420-00 F-7420-00
Gas Spring Assembly F-7224-00 F-7224-00 F-7223-00 F-7223-00
Belt Drive to Counterweight F-6343-00 F-6343-00 F-6343-00 F-6343-00
Table 3.2
3.6
Interceptor Biological Safety Cabinet Class II A2
July 2017
Page 19
Maintenance and Technical Manual Chapter 4
CHAPTER 4
Cabinet Operation
Theory of Operation
The Interceptor® is a Class II, A2 Biological Safety Cabinet. Its function is to isolate bacterial samples from cross-contamination and protect the user and the environment around the cabinet from being contaminated by biological or particulate material inside the cabinet.
It accomplishes this objective by bathing biological samples in HEPA ltered clean air while keeping contaminants contained by having a continuous substantial inow of air through the 10” cabinet sash opening.
The BSC will safely operate at a temperature range of 55ºF to 85ºF (13ºC - 29ºC) and a relative humidity of 10% to 70%.
Class II, A2 Biological Safety Cabinets are made to safely contain bacterial samples that fall into the Biological Safety levels 1, 2, 3, and 4 if accompanied by level appropriate protection garments. The following is a description of each level as published in the CDC BMBL 5th edition:
“Biological Safety level 1 (BSL-1) is the basic level of protection and is appropriate for agents that are not known to cause disease in normal, healthy humans. Biological Safety level 2 (BSL-2) is appropriate for handling moderate-risk agents that cause human disease of varying severity by ingestion or through percutaneous or mucous membrane exposure. Biological Safety level 3 (BSL-3) is appropriate for agents with a known potential for aerosol transmission, for agents that may cause serious and potentially lethal infections and that are indigenous or exotic in origin. Exotic agents that pose a high individual risk of life­threatening disease by infectious aerosols and for which no treatment is available are restricted to high containment laboratories that meet biosaftey level 4 (BSL-4) standards.”
! WARNING
ê
Class II A2 Biological Safety cabinets must be certified before initial use, after being moved, and after any service, including required annual recertification. Service must be performed by an NSF accredited certifier using NSF/ ANSI 49 criteria and should include, at minimum, the following test:
• Downflow Velocity Profile Test
• Inflow Velocity Test
• Airflow Smoke Pattern Test
• HEPA Filter Leak Test
Interceptor Biological Safety Cabinet Class II A2
July 2017
4.1
Page 20
Chapter 4 Maintenance and Technical Manual
Touch Pad Contol Monitor
Home/Face Velocity Screen
Asset Tag
Face Velocity
Warnings & Alarms
Latch Banner
Light ON/OFF
Button
Fan ON/OFF
Button
Settings
Button
Alarm Mute
Button
Local Time
12h /24hr
Unit of Measure
FPM or M/S
UV Light ON/OFF Button
Information
Button
Alarm LED
Home
Button
Figure 4.1 Touch Pad Control Monitor
4.2
Interceptor Biological Safety Cabinet Class II A2
July 2017
Page 21
Maintenance and Technical Manual Chapter 4
Startup Procedure
Plug in the Power Cord to a properly sized electrical receptacle. Raise the vertical sash to the 10" operating line. The Control Monitor will show the Kewaunee Contact Screen, Figure 4.2, for several seconds and will then display Home/Face Velocity Screen. Clear the the Power Failure/Latched Banner, Figure 4.3, by touching the banner once.
The Latched Banner is provided to inform the user that an unsave condition has taken place within the BSC. The Latch Banner will display until user clears it from the screen.
Figure 4.2 Kewaunee Contact Screen
Figure 4.3 Home/Face Velocity Screen - Latched/Power Failure
Important: Cabinet must be certified before being put into service.
Press the Fan button once and the fan motor will start. As the fan spins up, the face velocity will increase until the operating face velocity is reached and the cabinet is safe to use Figure 4.4 - 4.6. The Interceptor and viral matter and delivering HEPA ltered, clean air to the Biological Safety Cabinet. The air ow pattern is shown in Figure 4.7 on following page.
®
is now intercepting all dust, bacteria,
Figure 4.4 Home/Face Velocity Screen - Fan Start
Figure 4.5 Home/Face Velocity Screen - Fan Warm-Up
Interceptor Biological Safety Cabinet Class II A2
July 2017
Figure 4.6 Home/Face Velocity Screen - Safe Operating Condition
4.3
Page 22
Chapter 4 Maintenance and Technical Manual
Interceptor Class II, A2 Air Flow Patterns
It is recommended that the cabinet fan be turned on and allowed to operate for several minutes removing any suspended particulates. Cabinet interior should then be wiped down with 70% ethanol (EtOH), or other approved disinfectant.
HEPA
Filtered Air
Contaminated
Air
HEPA
Filtered Air
Room Air
Figure 4.7 Interceptor Biological Safety Cabinet Air Flow Patterns
4.4
Interceptor Biological Safety Cabinet Class II A2
July 2017
Page 23
Maintenance and Technical Manual Chapter 4
Suggested or Recommended Workow
Once skilled in your lab’s Standard Operating Procedure (SOP), don appropriate protective garments and gear suitable to the level of safety required. Place items for use in the procedure in the left section of the work area. Place equipment in the center of the work area. Designate the right section of the BSC for waste to be disposed of at the procedure’s end.
As you perform the procedure, waste materials will accumulate on the right side of the cabinet.
After the procedure is nished, remove waste materials, disinfect as is customary at your facility, and place materials away. Each time the unit is used, it should be left clean. Follow your SOP for shut down or night setback procedures.
Night Setback/Low Flow Mode
The Interceptor® Biological Safety Cabinet is equipped with a two-speed motor to save energy when it is not being used for active procedures. The cabinet is placed into Night Setback/Low Flow Mode by closing the sash to a 1" opening. At 1", the cabinet exhausts about 50% of the air the unit processes at a 10" opening. Do not
run active experiments at this 1" sash height.
When the sash is moved to a 1" opening, Night Setback/Flow Flow Mode, the Control Monitor will look like Figure 4.11 below.
NOTE: Low ow is set to 50% of run speed. It is not necessary to calibrate the BSC for this position.
Figure 4.11 Home/Face Velocity Screen­Night Set-back/Low Flow Mode
Interceptor Biological Safety Cabinet Class II A2
July 2017
4.5
Page 24
Chapter 4 Maintenance and Technical Manual
Control Monitor
The Control Monitor Home/Face Velocity Screen can alert you to several possible problems as you proceed:
1) High inow velocity (face velocity).
2) Low inow velocity (face velocity).
3) Sash Height Alarm – Always set the sash at the 10" line when conducting experiments. This is the setting that has received NSF certication.
4) UV Light on and Fan on.
Figure 4.12 Home/Face Velocity Screen Warning – Check Sash Height
UV Light
The UV germicidal disinfecting light may be used if one has been installed; fully close the sash, and turn on the light using the UV Light button. Because UV rays are dangerous, the light will not operate unless the sash is fully closed. Glass stops UV radiation. If the UV light is not turned off before the sash is raised, the safety interlock will disable the UV light.
! WARNING
ê
If the UV option is on your BSC, be sure safety overrides are never immobilized! UV light should NEVER be on while sash is open.
4.6
Interceptor Biological Safety Cabinet Class II A2
July 2017
Page 25
Maintenance and Technical Manual Chapter 4
Control Monitor Messages and Meanings
Variable Measurement Method Importantance Recommended Action
Airow
(optimal value is 105 FPM)
Since the inow at any point of the 10" sash opening can vary widely, it is the average inow (CFM/sq ft) that is actually measured. Since the “air in” through the 10" sash opening exactly equals “air out”, the exhaust port, the Interceptor converts the FPM exhaust port velocity into FPM inow velocity at the sash opening.
Proper inow at 10" assures containment of contaminants within the cabinet. NSF 49 requires a minimum inow velocity of 100 FPM.
®
The Home/Face Velocity Screen shows airow. Any velocity between 100 FPM and 110 FPM is acceptable. If face velocity falls below 100 FPM, discontinue work and contact your Accredited Technician for remediation.
Remaining Filter Life
Sash Alarm Banner
UV Light
(when installed)
The static pressure differential between the negative and positive pressure is proportional to lter loading. It is measured by the Pressure Sensor and converted by the Contol Module to a lter loading percentage.
A micro switch interaction with the counterweight activates this warning.
When the sash is closed and the fan is off, the disinfecting UV Light may be activated using the UV Light button on the Control Panel Touch Pad.
The Interceptor® exceeds NSF requirements for maximum static load.
The cabinet is designed to be used at a 10" sash opening. Any other opening is inappropriate. If the fan is on at any height other than 10", the Yellow
Sash Alarm Banner
displays.
Be sure UV Light Safety Overrides are never immobilized! UV light should NEVER be on while sash is open.
An Accredited Technician may need to adjust fan speed to maintain face velocity or change lter.
Return the sash to 10" or turn the unit off and close the sash.
Close sash and turn fan off to reactivate UV.
Remaining UV Lamp Life
Table 4.1
Internal countdown timer. Setting of 2000 hours can be reset to manufacturers recommendations.
When the hood UV light is on, it is important to know how much life is remaining in the UV lamp before its UV output diminishes. The remaining UV lamp life natation is displayed in the Info/About screen.
Interceptor Biological Safety Cabinet Class II A2
July 2017
Replace UV lamp, even if it still glows, when ‘0’ hours remaining is indicated.
4.7
Page 26
Chapter 4 Maintenance and Technical Manual
Monitor Screen Displays
The Kewaunee Interceptor® Biological Safety Cabinet employs a sophisticated control system using a built in tough screen Control Monitor to display menus and options. The following pages illustrate the different states of the monitor with illustrations of the various screens and schematics of each of the menus. The schematic charts show all options for each menu item as well as the factory setting.
Startup
Figure 4.13
Home/Face Velocity Screen
displays when cabinet is plugged in
Press Power Failure banner to clear Press Settings for Configuration
4.8
Interceptor Biological Safety Cabinet Class II A2
July 2017
Page 27
Maintenance and Technical Manual Chapter 4
Monitor Screen Displays
Home/Face Velocity Screen
Fan OFF
displays when Fan is Off
Press Fan Switch to Start
Figure 4.14
Figure 4.15
Home/Face Velocity Screen
Fan Warmup
displays when Fan is On during warmup period
Home/Face Velocity Screen
Fan On
displays when Fan is On after warmup period
Press Light button to turn on light
Figure 4.16
Interceptor Biological Safety Cabinet Class II A2
July 2017
4.9
Page 28
Chapter 4 Maintenance and Technical Manual
Monitor Screen Displays
Settings Menu
Used for User Settings and Cabinet Setup and Calibration
Press System for User Settings
Press Cabinet Setup for Cabinet Setup and Calibration
Figure 4.17
System Menu
Figure 4.18
Use to set User Settings
Press Alarms/Sound to adjust alarm and volume settings
Press Airflow Settings to adjust Sensor sensitivity and Airflow Resolution
Press Advanced to Configure Asset Tag and Change Passwords
Change Password Menu
Use to change Passwords
Press Admin Password to change
Press Certifier Password to change
Figure 4.19
4.10
Interceptor Biological Safety Cabinet Class II A2
July 2017
Page 29
Maintenance and Technical Manual Chapter 4
Monitor Screen Displays
Cabinet Setup Menu
Use to Setup Cabinet
Press Time to set date/time
Press Fan Speed to adjust airflow
Press Sensor Calibration to calibrate airflow sensor
Press UV Settings to adjust UV Timer
Figure 4.20
Figure 4.21
Calibration Setup Menu
Use to Calibrate Monitor Screen at time of installation and after filter changes
(only requred if actual airflow velocity does not match displayed airflow velocity)
Press Calibrate and follow prompts
Note:
To Be Performed by Certified Technician Only
Interceptor Biological Safety Cabinet Class II A2
July 2017
4.11
Page 30
Chapter 4 Maintenance and Technical Manual
Monitor Menu Schematics
Settings Menu
Menu Item
System to System Menu See Table 4.3
Cabinet Setup to Cabinet Setup Menu See Table 4.10
Back Return to Home Screen
Table 4.2
System Menu
Menu Item
Alarms/Sound to Alarm/Sound Menu See Table 4.4
Airflow Settings to Airflow Settings Menu See Table 4.6
Advanced
password
protected
password
protected
to Advanced Menu
See Table 4.7
Back Return to Home Screen
Table 4.3
Alarms/Sound Menu
Menu Item Input Option Factory Setting
Device Volume Hi, Med, Low
Key Beep On, Off
Set Minimum Volumes Hi, Med, Low
Alarm Controls to Alarm Controls Menu See Table 4.5
password protected
password protected
Back Return to System Menu
Table 4.4
Med
On
Low
4.12
Alarm Controls Menu
(password protected)
Menu Item Input Option Factory Setting
Ringback Off, 10s, 30s, 1min, 5min, 10min, 60min
Alarm Delay 1s, 2s, 5s, 10s
Startup Delay 30s, 1min, 2min, 5min
1min
1s
30s
Back Return to Alarms/Sound Menu
Table 4.5
Interceptor Biological Safety Cabinet Class II A2
July 2017
Page 31
Maintenance and Technical Manual Chapter 4
Monitor Menu Schematics
Airflow Settings Menu (password protected)
Menu Item Input Option Factory Setting
Sensitivity Low, Med, Hi
Airflow Resolution 1 FPM, 2 FPM, 5 FPM
Back Return to System Menu
Table 4.6
Med
1 FPM
Advanced Menu (password protected)
Menu Item Input Option Factory Setting
Configure Asset Tag to Configure Asset Tag Menu See Table 4.8
Change Password to Password Menu See Table 4.9
Back Return to System Menu
Table 4.7
Configure Asset Tag Menu (password protected)
Menu Item Input Option Factory Setting
Change Asset Tag ############
Show Asset On, Off
Back Return to Advanced Menu
Table 4.8
none
ON
Change Password Menu (password protected)
Menu Item Input Option Factory Setting
Enter Old Password - Admin
Admin Password
Certifier Password
Back Return to Advanced Menu
Table 4.9
Enter New Password - Admin Enter Confirm Pass - Admin
Enter Old Password - Certify Enter New Password - Certify Enter Confirm Pass - Certify
9999
1001
Interceptor Biological Safety Cabinet Class II A2
July 2017
4.13
Page 32
Chapter 4 Maintenance and Technical Manual
Monitor Menu Schematics
Cabinet Setup Menu
Menu Item
Time to Time Setting Menu See Table 4.11
password protected
Fan Speed to Fan Speed Menu See Table 4.13
password protected
Sensor Calibration to Sensor CalibrationMenu See Table 4.14
UV Settings to UV Settings Menu See Table 4.15
password protected
Back Return to Home Screen
Table 4.10
Time Setting Menu (password protected)
Menu Item Input Option Factory Setting
Show Time On, Off
On
Date & Time Settings to Date & Time Settings Menu See Table 4.12
Back Return to Cabinet Setup Menu
Table 4.11
Date & Time Setting Menu (password protected)
Menu Item Input Option Factory Setting
Date Enter Current Date Current Date
Time Enter Current Time Current Time
Time Format 12hr, 24hr
12hr
Back Return to Cabinet Setup Menu
Table 4.12
Fan Speed Menu (password protected)
Menu Item Input Option Factory Setting
Normal Fan Speed Set Normal Fan Speed ##%
NSB Fasn Speed Set NSB Fan Speed ##%
Back Return to Cabinet Setup Menu
Table 4.13
4.14
Interceptor Biological Safety Cabinet Class II A2
July 2017
Page 33
Maintenance and Technical Manual Chapter 4
Monitor Menu Schematics
Sensor Calibration Menu
Menu Item
Calibration Info
Calibrate Sensor to Sensor Calibration Instructions See Page 5.2
Back Return to Cabinet Setup Menu
Table 4.14
UV Settings Menu
Menu Item Input Option Factory Setting
UV Timer 30min, 60min, 90min, 2hr, 4hr, 6hr
UV Lamp Life #### (in hours)
Reset UV Lamp Life Okay, Cancel
Back Return to Cabinet Setup Menu
Table 4.15
password protected
(password protected)
Calibration Date:
(Date Last Calibrated)
MMM DD, YYYY
30min
2000hrs
Interceptor Biological Safety Cabinet Class II A2
July 2017
4.15
Page 34
Chapter 4 Maintenance and Technical Manual
4.16
Interceptor Biological Safety Cabinet Class II A2
July 2017
Page 35
Maintenance and Technical Manual Chapter 5
CHAPTER 5
Troubleshooting
Problem Possible Cause Recommended Fix
No lights or fan
No fan
Fluorescent light not working
UV light does not illuminate
Reduced face velocity
Table 5.1
a. Unit unplugged Plug it in b. BSC breaker open Reset breaker c. Building breaker open Check outlet breaker a. Sash not at operating height Move sash to operating height b. Sash microswitch not activated Adjust microswitch position so switch is activated when
sash is at operating height
c. Motor control broken Replace motor control. If condition persists, motor may
need replacement. Contact Kewaunee for assistance.
a. Lamp burnt out (look for dark
rings at opposite ends of glass
uorescent tube) b. Lamp wiring defective Inspect and repair c. Bad lamp ballast (symptom is
intermittent light) a. Sash must be closed for UV light to
work (regular glass blocks UV rays) b. Lamp burnt out Replace lamp
c. Lamp wiring defective Inspect and repair d. Bad lamp ballast (symptom is
intermittent light) a. HEPA lter loaded Increase motor RPM or replace lter if 100 fpm
b. Towels have clogged towel screen
or visible bafe louvers c. Exhaust outlet clogged with debris Clear outlet
Replace lamp
Replace ballast located inside rooftop circuit box
Close sash
Replace ballast located in electrical box
face velicity is no longer achievable
Clean towel screen (see Chapter 5 User & Operation Manual)
Technical Support
For further technical assistance, contact Kewaunee at 704-873-7202 or www.kewaunee.com.
Interceptor Biological Safety Cabinet Class II A2
July 2017
5.1
Page 36
Chapter 5 Maintenance and Technical Manual
Cabinet Calibration Instructions
Primary Method
Setup and Measurements
1. Install a Flow Hood on Interceptor® BSC per NSF #49 methodology A9.3.
2. Turn on Interceptor® fan and let cabinet stabilize 30 seconds.
3. Press Settings button on Home Screen.
4. Press Cabinet Setup on Settings Menu.
5. Press Fan Speed on Cabinet Settings Menu.
6. Enter Certifier Password. (default password 1001)
7. Press Normal Fan Speed.
8. Adjust Normal Fan Speed to set Inflow CFM based on data label on left facia.
9. Press Back to return to Fan Speed Menu.
10. Press NSB Fan Speed.
11. Adjust Night Setback Fan Speed to 50% of Normal Fan Speed.
12. Press Back to return to Cabinet Settings Menu.
13. Remove Flow Hood.
Confirm Downflow CFM matches data label. Adjust as required by sliding exhaust damper on top of cabinet.
Calibration
14. Press Sensor Calibration on Cabinet Setup Menu.
15. Press Calibrate.
16. Follow instructions on Control Screen to complete Calibration process.
17. Step 1: Fan will turn off. Press Continue when Airflow Uniformity stabilizes.
18. Enter fan off face velocity. Typically 0. Press Enter.
19.
20. Cabinet control will flash Calibration Success and then display Calibration Settings.
21. Press Back to return to Cabinet Setup Menu or Home Button to return to Home Screen.
22. Calibration is complete.
Step 2: Fan speed will increase to Normal Fan Speed setting (from step 8 above). Press Continue when Airflow Uniformity Statbilizes.
5.2
Interceptor Biological Safety Cabinet Class II A2
July 2017
Page 37
Maintenance and Technical Manual Chapter 5
Cabinet Calibration Instructions
Secondary Method
Setup and Measurements
1. Turn on Interceptor® fan and let cabinet stabilize 30 seconds.
Take inflow readings at 3" sash opening per NSF 49 section A.9.3.4.2. Readings shall be taken using the Secondary Inflow Measurement Method listed on the next page. (Be sure to adjust readings using air density
2.
3. Press Settings button on Home Screen.
4. Press Cabinet Setup on Settings Menu.
5. Press Fan Speed on Cabinet Settings Menu.
6. Enter Certifier Password. (default password 1001)
7. Press Normal Fan Speed.
8. Adjust Normal Fan Speed and then repeat Steps 2 through 8 until a Face Velocity of 105
9. Press Back to return to Fan Speed Menu.
corrections for Temperature, Pressure, and Humidity, or use an instrument that makes the adjustments automatically.)
Average the readings and then calculate the Alternate Face Velocity (V V
= 3/10 x CF x V
alt
(If the Alternate Face Velocity is 105
av.
+
5 FPM skip to step 13.)
-
), using the formula:
alt
+
-
5 FPM is achieved.
10. Press NSB Fan Speed.
11. Adjust Night Setback Fan Speed to 50% of Normal Fan Speed.
12. Press Back to return to Cabinet Setup Menu.
Confirm Downflow CFM matches data label. Adjust as required by sliding exhaust damper on top of cabinet.
Calibration
13. Return Sash to 10" Operating Position
14. Press Sensor Calibration on Cabinet Setup Menu.
15. Press Calibrate.
16. Follow instructions on Control Screen to complete Calibration process.
17. Step 1: Fan will turn off. Press Continue when Airflow Uniformity stabilizes.
18. Enter fan off face velocity. Typically 0. Press Enter.
19.
20. Cabinet control will flash Calibration Success and then display Calibration Settings.
21. Press Back to return to Cabinet Setup Menu or Home Button to return to Home Screen.
22. Calibration is complete.
Step 2: Fan speed will increase to Normal Fan Speed setting (from step 8 above). Press Continue when Airflow Uniformity Statbilizes.
Interceptor Biological Safety Cabinet Class II A2
July 2017
5.3
Page 38
Chapter 5 Maintenance and Technical Manual
Secondary Inow Measurement Method
1. Secure the Anemometer Probe to the outside of the sash glass using tape so that the probe is positioned vertically, facing down. Position the center of the sensing element 1.5” below the top of the 3” opening, directly above the Measurement Point. (as shown below)
2. Record the Velocity Reading.
3. Repeat for all the Measurement Points as shown on the chart below and as marked on the lower face of the Cabinet Airfoil behind the Armrest. Measurement Points are marked with small vertical slots .
Note: Armrest must be in place for testing and daily operation.
Secondary Method Measurements
Model Number
INT-1100A2 4.61” 4.63” 1.31
INT-1400A2 4.44” 4.00” 1.29
INT-1700A2 5.10” 5.05” 1.28
INT-2000A2 4.25” 4.00” 1.16
Table 5.2
Anemometer Probe
Dimension from Sides
Dimension
Between Points
Correction
Factor (CF)
Cabinet
Airfoil
3"
3"
5.4
1½"
1½"
Armrest
Measurement Points
Interceptor Biological Safety Cabinet Class II A2
July 2017
Page 39
Maintenance and Technical Manual Chapter 5
Wiring Schematic
Interceptor Biological Safety Cabinet Class II A2
July 2017
5.5
Page 40
LABORATORY PRODUCTS GROUP
P.O. Box 1842 • Statesville, NC 28687-1842 Phone: 704-873-7202 • Fax: 800-932-3296 Email: kscmarketing@kewaunee.com
© Copyright 2017 Printed in U.S.A. BSCMT-A2-0717
Website: www.kewaunee.com
Loading...