Kett Riceter M Series Instruction Manual

Accessories
DisplayMain Body
0611.TO.0106.001K
Backed by Kett’s advanced technology and 50 year experience as a manufacturer specializing in many types of moisture measuring testers for various purpuses,the Riceter Model m always enables the user to accurately measure the moisture content 3 from 6 types of grain without using any conversion tables.The Model m requires the user only to insert the sampling tray and turn the handle until it stops. All other procedures are automatically conducted with built-in micro-pro­cessed temperature compensator and sensitivity adjuster.
Sampling Tray Brush Spoon with tweezers Battery Carrying Case
Rice Husker (optional)
Grain Crushing Handle
Testing Chamber
series
Grain Moisture Tester
Instruction Manual
"POWER" Button "AVE" Button (Average) "SELECT" Button
Specifications
Measuring Principle :Electric Resistance Indication :Digital display Applications & Measuring Ranges
:Refer to
Riceter m series Application Table
Measuring Accuracy :S.E.C : 0.5% (920%)
(Standard Error of Calibration)
Temperature Compensation
:Dual automatic Battery :4 pcs x 1.5V ("AA"size) Dimensions : 164(W) x 94(D) x 65(H) Weight : approx. 443g Net Accessories
: 2 pcs.Sampling Tray / 1pc.Brush /
1pc.Spoon with tweezers / 4pcs."AA"Battery / Instruction Manual / 1 pc.Rice Husker(optional) / Carrying Case
Moisture Content Number of tests Selection Indicator Battery Indicator
1
First,confirm that the figures "8 TIMES 88.8%"
and several other marks appear when the "POWER" button is pressed with your finger. Should the figures not appear ,or mark appear, the batteries are exhausted. Replace all of the batteries with new ones, according to the diagram illustrated in the battery chamber.
2
Press the
"SELECT"
button and repeat until
the Selection Indicator of the sample you wish to test appears at the bottom of the display.
3
Take one spoonful of the grain to be measured,
and place it evenly in one layer in the Sampling
Tray.
NOTE: Sort out unripe or degenerated grains from the tray in order to avoid error.
4
Rotate the Crushing Handle counterclockwise,
and fully insert the Sampling Tray into the Testing Chamber.
Fully rotate the Crushing
Handle until stop is reached
, so that the sample may be well crushed. Press the "POWER " button, the most reliable moisture content is then displayed to be read directly without using any conversion tables.
5
The following indicators each have important
meanings. The illuminator lights for 2 sec­onds when the "POWER " button is pressed and for 4 seconds when the moisture content is displayed or the "SELECT" button is pressed.
The Battery Indicator indicates the battery power remaining.This indicator is displayed for approxi­mately 2 seconds when the "POWER" button is pressed.
INSTRUCTIONS
Battery Indicator
Full
Half
Empty
Over Indicator
Under Indicator Selection Indicator
The Over Indicator " "is displayed when the moisture content of the measured sample exceeds the upper limit of the measurement range, or in the event of problems with the measurement operation ,or if irregularities are detected. The Under Indicator " " is displayed when the moisture content of the sample is below the lower limit of the measurement range.
NOTE: Over or under indicators may also be displayed in the event that the measurement pro­cedure is perfomed without a sample loaded on the sampling tray, there is condensation within the testing chamber, or insulation is poor due to high humidity or soiling. If this should occur, thor­oughly clean the testing chamber and allow it to dry through natural evaporation if damp.
6
The averaged moisture content of 2-9
sequential tests is indicated when the "AVE" button is pressed with in 2 minutes after the tests. In this case," " or " "marked tests are disregarded. Numbers of tests being performed is shown on the left part of the digital display. But in the following cases, the reading of the number of tests goes back to "1".
1. More than 2 minutes have passed after the tests.
2. After the tests are already averaged.
3. More than 9 tests are performed
.
POWER OFF
1. Continue to press the "POWER" button for 2
seconds, the unit will power off.
2. When inactive, the unit will automatically power
off affer 5 minutes.
After each measurement, clean the Testing Chamber, Crushing Handle and Sampling Tray with the supplied brush.
Too Many Insufficient
Evenly Placed in one layer