Kett LE-330J, LH-330J, LZ-330J Operating Manual

330J Series Coating Thickness Testers
LE-330J/LH-330J/LZ-330J
Operating Manual
Coating Thickness Tester Safety Precautions
Be sure to carefully follow all safety precautions.
Carefully read operating manual.
Immediately contact our service representative if the unit malfunctions or does not operate properly.
Measuring of Warning Indications.
The symbols indicated below are used in the operating manual and on the unit itself in order to prevent accidents due to misuse of the product. These symbols have the following meanings:
This symbol indicates information important for the understand in order to safety operate the product.
Caution
1. Measuring Principles and Characteristics ........................................................................... 4
2. Instrument view ................................................................................................................... 6
3. Display................................................................................................................................. 8
4. Keypad Fanctions................................................................................................................ 9
5. Specifications ...................................................................................................................... 12
6. Preparations for Operation .................................................................................................. 13
7. Measuring Procedure .......................................................................................................... 22
8. Functions ............................................................................................................................. 24
9. Easy Calibration (For Eddy Current method only)............................................................... 32
10.Notes for Measuring and Handling ...................................................................................... 33
11.Data output format ............................................................................................................... 35
Contents
4
• The model LE-330J electromagnetic coating thickness tester is designed to measure the thickness of non-magnetic coat­ings such as paint or plating on magnetic metal substrates (iron or steel).
• The model LH-330J Eddy Current coating thickness tester is designed to measure the thickness of insulating coatings such as alumite or paint on non-magnetic metal substrates (such as aluminum or copper, etc.).
• The model LZ-330J dual-type coating thickness tester is designed to measure the thickness of either non-magnetic or insulating coatings on either magnetic or non-magnetic metal substrates.
Applicable Coatings
• A Beryllium-copper calibration foil is required in order to measure the thickness of plating coatings.
Model
LZ-330J
LE-330J LH-330J
Applicable Coatings
Substrate
Iron Steel Aluminum, Copper, Brass, etc.
Paint Plastic Lacquer Resin Rubber Enamel Lining Zinc Chrome Tin Copper Aluminum Other
Paint Alumite Rubber Plastic Enamel Lacquer Resin Other
1. Measuring Principles and Characteristics
5
• LH-330J
(Eddy Current measurement method: For measuring the thickness of insulating coatings on non-magnetic metal substrates)
An eddy current is produced in the surface of a metal when a coil through which a current of fixed frequency is brought near the metal. This eddy current and the voltage at the ends of the coil change in proportion to the distance be­tween the coil and the metal surface. This change can be determined from the current value and this is used to cal­culate the thickness of the coating.
• LE-330J
(Electromagnetic measurement method : For measur­ing the thickness of non-magnetic coatings on mag­netic metal substrates)
When an alternating current electromagnet is brought near iron (or other magnetic metal) the number of magnetic flux lines passing through the coil changes in proportion to the distance, thereby causing a change in the voltage at the ends of the coil. This change in voltage is deter­mined from the current value and this is used to compute the thickness of the coating.
• LZ-330J(Dual Electromagnetic and Eddy Current Type)
The model LZ-330J is a dual type coating thickness tester which features both the electromagnetic and Eddy Current measurement methods.
Low-Frequency Power Supply
Non-Magnetic Coating Iron • Steel
Thickness D
Electromagnetic coil
Current Change i
Induction coil
Insulating Coating Non-Magnetic Metal
Thickness D
High-Frequency Power Supply
Current Change i
i
6
2. Instrument view
Front Panel Rear Panel
Main part Probe
Top View
Fe Probe (Black)
Electromagnetic (LEP-J)
NFe Probe (Gray) Eddy Current (LHP-J)
Probe Connector Dust Cap printer & Computer Out put Connector
Display
Keyboard
Battery Box
Connector
Connector
Measurement Part
Measurement Part
7
Accessories
Standard Calibration Foils Probe Adaptor
Carrying Case Batteries size AA (4pcs) Calibration Foilcase
Operating Manual printer VZ-330
Iron Substrate (FE-J) Aluminum Substrate (NFE-J)
(For LE-330J,LZ-330J) (For LH-330J,LZ-330J)
(LE-330J,LZ-330J :
6pcs /
LH-330J :
5pcs
)
Optional Accessories
8
Foil Calibrations ( P. 17)
Measurment Mode
Calibration Procedure ( P. 16,17)
Limit Mode ( P. 25)
Battery Check ( P. 34)
Upper Limit ( P. 25)
Lower Limit ( P. 25) Unit of Measurement ( P. 23)
3. Display
• Please refer to the page shown by " P. " about the function and meaning of each pattern.
Measured value etc.
Electromagnetic
( P. 6,13)
Eddy Current
( P. 6,13)
Measurement Number Symbol
Lot Number ( P. 24) Data ( P. 24)
Date ( P. 24) or measurement number
Minimum value ( P. 23,28,29,36) Maximum value ( P. 23,28,29,36) Standard deviation ( P. 23,28,29,36) Average value ( P. 23,28,29,36)
9
Key Name Function
ON key, OFF key This key is a power supply switch.
1. Number keys 1. This key is used for a number input.
2. Dual-function keys 2. These keys have two functions.
1. CLEAR key
1. This key is used to correct mistakes made when inputting
2. DATA TRANSFER Key
numeric data.
2. This key is also used to transmit measure ment data and other information to a printer or computer.
( press
)
ENTER key
Use to enter the numeric value indicated in the display. Also used to proceed to the next operation.
FUNCTION key
This key is used to execute the Function Mode functions of the numeric keys. (Date, Lot Number, Statistical calculations, etc.)
4. Keypad Fanctions
10
Dual-function Keys : The following keys also have a thing with two functions.
These are called function mode key. These ke ys are pressed immediately after presetting a function k e y. A result -- as a function mode key -- functioning.
Function mode
Numeric key
Key Name Function
DA TE Key Used to print the date.
( press )
LOT Number Key Used to input lot numbers.
( press )
DELETE Key Used to delete mistaken measurement data from the measurement
value memory. ( press
)
LOT RESULT Key Used to perform statistical calculations by lot.
( press )
11
Key Name Function
SUBSTRATE CALIBRATION key
Substrate ‘0’ calibration. ( press
)
FOIL CALIBRATION key For making foil calibrations.
( press )
CONTINUOUS key
1. Hold measurement (for auto result entry)
2.
Continuous measurement (requires
key
for result entry)
(Each press of this key switches the mode.)
( press )
LIMIT SET key For presetting the upper and lower limits.
FINAL RESULT key Used to perform overall statistical calculations.
( press )
ALL CLEAR key Used to delete all data from the measurement value memory.
(Measured values, various comments, etc.) ( press )
APPLICATION NUMBER key Used to select and confirm the calibration memory location at
which calibration data is stored. ( press
)
12
5. Specifications
LE-330J LH-330J LZ-330J
Electromagnetic Eddy Current Electromagnetic,Eddy Current
Non-magnetic coatings on magnetic metal substrates Insulating coatings on non-magnetic metal substrates Non-magnetic coatings on magnetic metal substrates
Insulating coatings on non-magnetic metal substrates
0 ~1500µm or 60.0 mils 0 ~800µm or 32.0 mils
0 ~1500µm or 60.0 mils,0 ~800µm or 32.0 mils
LEP-J (Fe) LHP-J (NFe) LEP-J (Fe),LHP-J (NFe)
±1µm under 50µm ±2% at 50µm or greater
0.1µm under 100µm 1µm at 100µm or greater µm or mils (selection)
5 x 5mm
2180 coating thickness measurements maximum
44
Electromagnetic x
4,
Eddy Current x
4
No. of measurements, number of block numbers, average value, standard deviation, maximum value, minimum value
Digital (LCD, minimum displayed digit is 0.1µm)
RS-232C
1.5V alkaline batteries (size AA) x 4 60 hours (continuous use)
0 ~ 40˚C
Main Unit: 75(W) x 140(D) x 31(H) mm, 0.5kg Probe:
Ø13
x
94 mm
Iron substrate
(FE-J)
Aluminum substrate
(NFE-J)
Iron substrate (FE-J), Aluminum substrate (NFE-J)
* Standard Calibration Foils x 6 or 5, Probe Adaptor, Carrying Case, Batteries (size AA) x 4, Calibration Foil Case, Operating Manual
Beryllium Copper Calibration Foil, Printer VZ-330
* The accessory plastic calibration foils supplied have had their actual values measured. For this reason the values indicated above are
only approximate and are not exactlythe same actual thicknesses of the supplied foils.
* Standard Accessories Plastic Calibration Foils LE-330J, LZ-330J : Approx. 10,50,100,350,800,1000 (µm)
LH-330J : Approx. 10,50,100,350,800, (µm)
Model Measuring Method
Applications Measuring Range
Probe type Measuring Accuracy Resolution Display Unit
Min. Measurement Surface Area Data Memory Capacity NO. of Calibration Parameter Memories Statistical Fanctions
Display Interface Power Supply
Battery Life Temperature Dimensions & Weight Accessories
Accessories (Common use) Optional Accessories
13
6. Preparations for Operation
1. Insert the batteries
The battery cover is opened by sliding the battery cover on the rear of the case in the direction indicated by the arrowhead mark ( " " ). Insert 4 size AA alkaline batteries into the batter y compartment as indicated by the picture in the compartment. Take care to align the positive and negative ends of the batteries correctly.
2. Selecting and Connecting the probe (LZ-330J)
Select a probe appropriate to the type of sample to be measured with switching the main unit off.
(LE-330J • LZ-330J) Non-magnetic coating on magnetic metal substrates: Electromagnetic type probe Fe probe (Black)<LEP-J>
(LH-330J • LZ-330J) Insulating coatings on non-magnetic metal substrates: Eddy current type probe NFe probe (Gray)<LHP-J>
• The probe can be easily connected to the main unit by turning a little while softly pressing it into the probe socket on clockwise of the main unit.
• When removing the probe from the main unit, always be sure to grip, the metal fitting at the end of the probe cord connected to the main unit. Note that the probe can not be disconnected by pulling on the blac k portion of the cord.
Slide to open
Load baatteries into the main unit
Connecting the probe
14
Using probe Using probe adapter
for carved probe
Measuring Range Adjustment point ( Electromagnetic )
0~50µm
Iron substrate
10µm25µm50µm
50~500µm
Iron substrate
50µm 100µm 500µm
500~1500µm
Iron substrate
500µm 1000µm 1500µm
Measuring Range Adjustment point ( Eddy Current )
0~50µm
Aluminum substrate
10µm25µm50µm
50~300µm
Aluminum substrate
50µm 100µm 300µm
300~800µm
Aluminum substrate
350µm 500µm 800µm
4. Using the probe
The probe features a 1-point contact, fixed pressure design which ensures fixed load applied the chip on the tip the probe. As shown in the diagram, grip the probe near the tip and press it against the measurement surface so that it is perpendicular to the surface. To make the next measurement, lift the probe tip at least 10mm from the measurement surface and then press it down again.
•The probe adapter can be used to obtain stab le, reliab le measurements when measuring coatings on curved surfaces such as pipes , etc., or when making continuous measurements on a flat surface.
3. Adjustment preparation
Adjustments must be made attempting to make measurements. However, measurement calibrations are stored in the unit’s computer memory, therefore new adjustments are not required if the sample to be measured has been measured previously. In this case, the previous adjustments can be recalled from memory.
1A standard "substrate" of the same material,shape and
thickness as the sample to be measured is used to make calibration adjustments. Hav e the standard substrate ready before beginning the adjustment procedure.
•Use bare standard substrates which are free of coatings such as plating or paint. Standard substrates of this type are ref erred to as "z ero f oils" in this manual.
2In order to obtain the most accurate precise measurement
results possible, adjustments are perf ormed with zero foils and standard foils (samples of known thickness) using the Standard Foil 4-point Adjustment Method.
•Ref er to the chart to select a set of 3 standard foils appropriate for making adjustments for the thickness of the coating to be measured.
<Standard Foil Combination Example>
•The supplied standard f oil values will not necessarily be e xactly the same as those indicated in the chart. Instead, standard plates with actually measured values close to these values are supplied.
•An attached standard board is used for an easy accuracy check.
It attaches in the point of a probe with an adapter. Furthermore, a screw is tightened.
15
5. Calibration
Specify calibration memory address at which to store calibration settings.
The LZ-330J has the capacity to store up to a maximum of 4 calibration settings for both electromagnetic and eddy current measurement modes, and the LE-330J has the capacity to store up calibration settings for 4 electromagnetic mode and LH-330J also has the capacity to store up calibration setting for 4 eddy current mode. The calibrations used for a particular measurement are selected depending upon the type of sample being measured. The calibration memory address (APL#) at which a particular set of calibration settings is stored is selected at the beginning of the adjustment (calibration) procedure.
The memory addresses that can be used for electromagnetic and eddy current calibration settings are as follows:
(LE-330J • LZ-330J)
Electromagnetic Mode (Fe Probe) A1, A2, A3, A4
(LH-330J • LZ-330J)
Eddy Current Mode (NFe Probe) A5, A6, A7, A8
•After po wering on and after switching between electromagnetic and eddy current mode, the unit is automatically set to the previous APL#. If it is
not necessary to change this, press key.
1) Calibration Memory Address Selection Procedure
Example: Using an electromagnetic probe
Step
Operation Display
1
2
3
4
5
16
Press the ENT key 10 times.
Step
Operation Display Explanation
Press the key .
Press the key .
Press the key 10 times while pressing the probe against the substrate.
•The substrate should be an uncoated (no plating, paint, etc.) sample of the same shape and material as that to be measured.
After pressing the key 10 times the buzzer will sound intermittently and "IF" will appear in the display.
Press the key while holding the probe toward the air.
•The probe must be in the air, don’t touch any surface or others.
1
2
3
4
○○○○○○
○○○○
2) Substrate Calibration Procedure
Once substrate calibration settings have been registered it is not necessary to perform this operation each time a measurement is made. However this procedure must be performed again when changes such as replacement of the probe are made.
Example: Using an electromagnetic probe
17
3) Standard Foil Calibration Procedure
Turn on the probe and the main unit and hold the tip of the probe toward the air for 2 ~ 3 seconds.
Example: Performing calibration using the zero foil and standard foils (40, 100 and 400µm foils).
1
2
5
5
6
7
Press the key.
Press the key .
Measure the (substrate) 5 times. The buzzer will sound and the measured value will be displayed each time.
•Substrate: An uncoated (no plating, paint, etc.) sample of the same shape and material as that to be measured.
•Depending upon the type of material, the value displa yed ma y diff er greatly from that indicated in the example to the left (a value close to zero). However, the procedure can be continued by setting it close to the setting value by performing steps 9 and 0.
○○○
Step
Operation Display Explanation
(Measure 5 times)
<Calibration using the zero foil>
18
Press the key while holding the probe toward the air. "---" is shown in the display.
Input the coating thickness of the substrate (0.0µm). Press the key. The number displayed disappears and S.
NO increments from 0 to 1 to indicate that the unit is entering the standard foil calibration mode.
8
9
0
<Calibration using the standard foil (40µm)>
!
1
2
5
Place the 40µm calibration foil on the substrate and repeat the measurement procedure 5 times.
•Depending upon the type of material, the v alue display ed may diff er greatly from that indicated in the example to the left. However, you can continue the procedure by setting it near the setting value by performing steps # and $.
○○○
Step
Operation Display Explanation
(Measure 5 times)
19
Press the key while holding the probe toward the air. "---" is shown in the display.
Input the coating thickness of the calibration foil (40.0µm).
Press the key. The number displayed disappears and S. NO increments from 1 to 2 to indicate that the unit is ready
for calibration using the second standard foil.
@
#
$
<Calibration using the calibraiton foil (100µm)>
%
Place the 100µm standard foil on the zero foil (substrate) and repeat the measurement procedure 5 times.
•Depending upon the type of material, the value displa yed ma y diff er greatly from that indicated in the example to the left. However, you can continue the procedure by setting it near the setting value by performing steps & and * .
1
2
5
Step
Operation Display Explanation
(Measure 5 times)
○○○
20
Place the 400µm standard foil on the substrate and repeat the measurement procedure 5 times.
•Depending upon the type of material, the v alue display ed may diff er greatly from that indicated in the example to the left. However, you can continue the procedure by setting it near the setting value by performing steps ­and = .
Press the key while holding the probe toward the air. "---" is shown in the display.
Input the coating thickness of the calibraiton foil (100.0µm).
Press the key. The number displayed disappears and S. NO increments from 2 to 3 to indicate that the unit is ready
for calibration using the third standard foil.
^
&
*
<Calibration using the standard foil (400µm)>
(
Step
Operation Display Explanation
(Measure 5 times)
1
2
5
○○○
21
Press the key while holding the probe toward the air. "--­" is shown in the display.
Input the coating thickness of the calibraiton foil (400.0µm).
Press the key. The display returns to its original state. This completes the calibration procedure.
•Calibrations remain stored in the unit's non-v olatile computer memory even when the unit's power is turned off.
)
-
=
Step
Operation Display Explanation
•5 measurements are made at each stage of the zero and calibration foil adjustment procedure in order to obtain the average measurement value.
• If you make errors while entering numeric values during the calibration procedure , press the key to delete the value and then input the correct value. However, please note that corrections cannot be made after the key has been
pressed. If y ou press the key after entering incorrect values, turn the power off once and then back on and resume the procedure from step 5.
22
7. Measuring Procedure
(1)Select & Connect the probe
In the LZ-330J, select the appropriate type of probe for the type of material to be measured. (
P. 6,13)
(2) Turn on the main unit's power.
Press the ON key.
(3) Adjustment
Before beginning a measurement, check to see if the unit has been calibration for measurement of that type of material. If calibration has not been performed, perform the calibration operation and register the calibration settings. (
P. 15)
(4) Delete contents of the measurement value memory
Except in cases in which you wish to continue making measurements with data acquired previ­ously, you must delete the entire contents of the measurement value memory. (
P. 26)
(5) Connect the printer
Connect the printer to the unit if you wish to print out results while performing measurement and other operations. (It is also possible to connect the unit to a printer after completing measure­ments and print out the measurement values and the results of statistical calculations. (The printer is an optional accessory.)
(6) Specify calibration settings
Specify the calibration memory address of cali­bration settings registered for the type of material being measured. (
P. 15)
(Example: Specifying the calibration settings stored at calibration memory address 2.)
•When the display appears as indicated belo w:
• When the display indicates a calibration memory address:
23
(9) Data Processing (Statistical Processing)
• Example: Lot Statistical Calculations
AV: Average value SD: Standard deviation MAX: Maximum value MIN: Minimum value
1.Performing calculations without using a printer
2.Printing out the results when connected to a printer
3.Batch transmission of data to the printer
•Data can also be transmitted to a computer in the same way.
(10) Converting Measurement Value Display Units
You can convert between µm and mils as the measurement value units displayed. To do so, first turn off the units power, then press the key to turn the unit on again while holding down the key.
while holding down
(7) Enter comments such as the date and lot number, etc.
Commentary information which can be entered includes the date, lot num­ber, and upper and lower measurement value limits. This information is input as necessary. (
P. 24,25)
•Example: Inputting lot number 125
(8) Measurement
Quickly press the probe against the surface to be measured in such a wa y that the probe is perpendicular to the surface. The next measurement can be performed after first removing the tip of the probe at least 10mm from the surface being measured.
•"FFF" displayed in the display during measurement indicates that all measurement value memory locations are in use. If this occurs , perform the procedure described in (4) "Delete contents of the
measurement value memory" (
P. 26) above and then make the
measurement again.
•"HH" or "LL" will appear in the display if the measured coating thickness values exceed the upper or lower bounds of the unit's measurement range.
24
Step
Operation Display
1
2
3
4
5
Step
Operation Display
1
2
3
4
8. Functions
(1) Entering the Date
Example: Entering the date as May 6, 2003
(2) Entering the Lot Number
Example:
ntering 651 as the lot number
•Up to three digits can be input for the lot number.
•Lot number input can be input as many times as necessary during measurement.
25
Step
Operation Display
1
2
3
4
5
6
(3) Setting Limit Values
Example: Setting the upper limit at 100µm and the lower limit at 50µm.
•Upper and lower limits are set when it is necessary to check that coating thickness values are within a certain range.
•A buzzer sounds and the " " indicator appears in the display if measured values fall outside of this range.
•Keys are pressed in the following sequence in order to delete the upper and lower limit settings.
26
Step
Operation Display
1
2
3
(4) AC
(All Clear: delete the contents of measurement value memory)
This operation deletes the measurement data, statistical results and com­ments from the unit's memory.
•"FFF" display in the display during measurement indicates that all measure ment value memory locations are in use. If this occurs, perform the procedure to delete and then measure again.
•"HH" or "LL" will appear in the display if the measured coating thickness values exceed the upper or lower bounds of the unitis measurement range.
Step
Operation Display
1
2
3
4
(5) DEL (Partial deletion of measurement value data)
You can delete specific measurement values such as erroneous measurement data which you do not wish to be used in statistical calculations.
Example: Delete data for the forth measurement made.
•After statistical processing has been done it is not possible to delete a portion of the measurement data used for processing in order to change the processed results.
27
Step
Operation Display
1
2
To Continuous Mode To Hold Mode
(6) CONT (Changing the Measurement Value Display Mode)
You can switch the measurement value display mode between the Hold (fixed value display) and Continuance (continuous monitoring) modes.
•The Continuance Mode is used in cases such as the measurement of coatings on complexly shaped objects in which measurement values have a tendency to be unstable. This allows you to obtain stable measurement values and makes statistical processing easier and more reliable.
<Making Measurements in the Continuance Mode>
Press the tip of the probe against the surface to be measured and wait for the value sho wn in the displa y to stabiliz e, then press the key. The value displayed at that point will be stored in memory as the
measured value.
<Making Calibrations in the Continuance Mode>
Use the key as described above when performing steps 7, ! , % and ( of the standard foil calibration procedure.
•You can switch back and forth between the Hold and Continuance modes by repeating this key operation procedure .
28
(7) LOT RES (Lot statistical calculations)
The 330J series are capable of providing statistical results by lot. However, lot statistics can not be obtained without at least 2 sets of mea­surement data for a particular lot.
•Please note that if a printer is not connected to the unit you will not be able to redisplay the results of the statistical calculations after you
press
key at the end of this procedure.
•Perform the following k ey operations to calculate statistical inf ormation when a printer is connected to the unit. (The key operation procedure is simpler in this case.)
•The formulas used to obtain the various statistical values and the symbols used in the display are as follows:
N : Number of measurements performed.
Average Value AV:
Standard Deviation SD:
Maximum Value Max
Minimum Value Min
Step
Operation
Display
1
2
3
4
5
6
Example: Measurement data for the lot is 162µm x 2 and 163µm x 2 and a printer is not connected.
(Average value)
(Standard deviation)
(Maximum value)
(Minimum value)
29
Step
Operation
Display Explanation of Calculated Results
1
2
3
4
5
(8) RESULT (Overall statistical results)
1. When not using a printer
Example: Setting to lot 1 and making 3 measurements (X
1
~ X3), and then setting to lot 2 and making 2 measurements (X4 ~ X5). Calculate statistical
results based on all of the measurement data thus collected. No printer connected. Y1 in the explanation column represents the average value of measurements X
1
, X2 and X3. Y2 represents the average value of measurements X4 and X5.
Press the key.
Average value of all data acquired (X1 ~ X5)
•To reconfirm the statistical results, press the 1 key before pressing the key at the end of the procedure.
Standard deviation value for all data acquired (X1 ~ X5)
Maximum value among all measured values (X1 ~ X5)
Minimum value among all measured v alues (X1 ~ X5)
30
Step
Operation
Display Explanation of Calculated Results
6
7
8
9
0
Average of Y1 and Y
2
Standard deviation of Y1 and Y
2
Maximum value of Y1 and Y
2
Minimum value of Y1 and Y
2
Returns to the measurement display.
•You can reconfirm any of the statistical values by pressing the corresponding ke y indicated above before pressing the
key .
31
2. When connected to a printer
Step
Operation
Display
1
2
(9) DATA TRANS (Data transmission)
You can send measurement data or various statistical data obtained pre vi­ously (overall or by lot) to the printer all at once. (This data can also be transmitted to a computer in the same way.)
Step
Operation
Display
1
2
(transmitting)
(transmission
completed)
32
9. Easy Calibration (For Eddy Current method only)
Simplified "easy" adjustment can be made in order to make Eddy Current measurement mode measurements. (Easy adjustment is not possible for the electromagnetic measurement mode.) If the coating thickness does not exceed 500µm, measurement results are as precise as those obtained using standard foil calibration adjustments. In the easy adjustment procedure, only the substance calibrations are perf ormed. Once substance calibrations have been registered it is not necessary to perform this procedure each time measurements are made. However, the substance calibration procedure must be redone in cases such as if the probe is replaced with a new one, etc.
<Easy Adjustment Procedure>
This type of calibration adjustment is only valid for the eddy current measurement mode.
Press the ENT key 10 times.
Step
Operation Display Explanation
1
2
3
4
○○○○○○
○○○○
Press the key.
Press the key.
Press the key 10 times while pressing the probe against the substrate.
•The substrate should be an uncoated (no plating, paint, etc.) sample of the same shape and material as that to be measured.
After pressing the key 10 times the buzzer will sound intermittently and "IF" will appear in the display.
Press the key while holding the probe toward the air.
•The probe must be in the air, don’t touch any surface or others.
33
10. Notes for Measuring and Handling
(1) Confirm the type of material being measured.
Be sure to check the type of material and select the correct probe type before beginning measurements.
(2) Do not damage the probe or get it dirty.
Accurate measurement results cannot be obtained if the chip on the tip of the probe is damaged or dirty. Do not pound the probe against the measurement surface or move the probe laterally while it is pressed down upon the surface. When finished making measurements, use a soft cloth wet with benzine or alcohol to clean the tip of the probe.
(3) Handle the calibration foils with care.
The thickness of the calibration foils has been measured v ery precisely. You will not be ab le to obtain accurate measurement results if you use calibration foils which ha ve been scr atched, bent or otherwise damaged. Be particularly careful not to subject the thinnest foil, the 10µm foil, to wear. If a standard foil becomes damaged while being used, please contact the dealer from whom you purchased the tester and order a replacement of the same thickness. Although the thickness of replacement foils may slightly different than that of the original foils, this does not pose a problem for calibration adjustments.
Caution
Caution
34
(4) Replace the batteries immediately if the low-voltage indicator is displayed.
When the batteries become weak, a "BAT" indicator will appear in the display when the power is first turned on or while the tester is in use. Immediately replace the batteries with 4 new size AA alkaline batteries when this indicator is displayed. Please note that calibration adjustments are maintained in the unit's non-volatile memory even when the batteries are exhausted.
(5) Adjustment & Inspection
In order to maintain precise performance the coating thickness testers should be inspected at least once per year. Please contact the merchant from whom you purchasd your unit regarding inspection.
Measurement of paper or a film
After calibration on Iron or Aluminum Substrate, the thickness of paper or a film can be measured on the Iron or Aluminum Substrate.
Caution
35
11. Data output format
(1) Transmission of data
The 330 series may be connected to either an optipnal printer or PC printer to out measurement data or statistical results.
Digits Output
0 12345678910111213 14 15 16 17 18 19 20 21 22 23 24
*APPLNO=1
LF CR *LOT100 *DATE12:34:56 *U.LIM 50.0 µm *L.LIM 5.0 µm
N=0001 23.4 µm N=0002 65.2 ! µm N=0003
-
1.0 ! µm
N=0004
Example: Data Code
$1B $57 $01 $20 $4E $3D $30 $30 $30 $31 $20 $20 $32 $33 $2E $34 $20 $20 $20 $20 $20 $75 $6D $0A $0D
N=0001 23.4 µm
Notes
*APPL NO=1 *LOT 100 *DATE 12:34 :56 *U.L IM 50 .0 µm *L.LIM 5.0 µm N=0001 23.4 µm N=0002 65.2!µm N=0003
-
1.0!µm
N=0004
<Data communications>
Standard RS-232C
<Interface Specifications>
Baud rate : 2400bps Data bits : 8 bits Stop bits : 1 bit Parity bits : None Sync format : synchronous Signal Level : ±9V
<Output Contents>
Soft Command : 3 digits Data & Comment
: 20 digits LF Code : 1 digit CR Code : 1 digit Total : 25 digits
One digit is consisted of two bytes and contents as per ASCII code table.
36
Operation Print Out Notes
Example: The date as May 6, 2003
Statistical calculations for Lot 100.
Statistical calculations for Lot 200.
(Measurement)
(Measurement)
(2) Example Printout
Performing measurements and statistical calculations with a printer connected to the unit and printing out the results thus obtained.
um µm
37
Operation Print Out Notes
Overall statistical calculations for measurement data from N = 0001 to N = 0006.
Statistical processing of results of statistical calculations for LOT 001 and LOT 002 (AV).
Statistical calculations for Lot 300. (Measurement data prior to N = 0006 is not subject to this statistical processing.)
(Measurement)
MEMO
0601•TO•0102•200
Loading...