Software calibration function
Software to connect with PC and USB port
Compact metal case, suitable for use under poor
working conditions
Continuous working period of not less than 50
hours with two Alkaline batteries (AA size);
Auto Power Off to save energy
Dimensions: 132 mm x 76.2 mm
Weight: 345 g
1.2 Main Application & Testing Range
1.2.1 Main Application
Die cavity of molds
Bearings and other parts
Failure analysis of pressure vessels, steam
generators and other equipment
Heavy work piece
Testing the surface of a small hollow space
Material identification in the range of metallic
materials
1.2.2 Testing Range
Please refer to Table 1 and Table 2 in the APPENDIX.
1.3 Configuration
Table 1-1
2. Structure Feature & Testing Principle
2.1 Structure Feature
2.1.1 Hardness Tester Appearance
1. Main body 2.Impact device
2.1.2 Parts of the Main body
1.4 Working Conditions
Working Temperature: 0°C up to +40°C
Storage Temperature: -30°C up to +60°C
Relative humidity: ≤ 90%
Vibrations, strong magnetic field, corrosive medium and
heavy dust should be avoided in the surroundings.
HMR-BA-e-1212 2
1 Socket of USB
2 Aluminium case
3 Belt hole
4 Battery cover
5 Keypad
6 LCD Display
7 Socket of Impact Device
8 Aluminium case, (backside)
9 Label (backside)
The basic testing principle is to use an impact body of a
certain weight which impacts against the testing surface
under certain test force. The impact velocity and the
rebound velocity of the impact body are measured
respectively when the spherical test tip is located 1mm
above the surface being tested.
The calculation formula is as follows:
HL= 1000 x VB/VA
Where HL= Leeb Hardness Value
VB= Rebounding velocity of the impact body
VA= Impact velocity of the impact body
Display: dot matrix LCD,128 x 64 dots
Data memory: max. 500 groups
Power supply: 3V, (2AA size Alkaline batteries)
Continuous working period: about 50 hours
(with backlight off)
Communication interface: USB1.1
4. Preparation & Testing
4.1 Preparation and Inspection before Testing
4.1.1 Preparation of Sample Surface
Preparation of the sample surface should conform to the
relative requirements in APPENDIX Table 3.
During the preparation of the sample surface,
any heating or cold processing on this surface
should be avoided.
Too big roughness of the surface being
measured could cause errors. The surface of the
sample should show metallic luster, be smooth
and polished, without any stains of oil etc.
Support of the sample is not necessary for
heavy objects. Medium-weight samples have to
be placed onto a smooth and stable plane. It
must be placed in absolute equability and
without any wobble.
Curved surface: The best testing surface, of
course, is flat. When the curvature radius R of
the surface to be tested is smaller than 30 mm
(D, DC, D+15, C, E and DL type of impact
device) and smaller than 50 mm (G type of
impact device), the small support rings should
be chosen.
The sample should have enough thickness, the
minimum thickness of it should conform to
Table 3.
*For a sample with a hardened layer on its surface,
the depth of the hardened layer should also conform
to Table 3.
surface. If the sample is a big plate, long rod or
bending piece, it can be deformed and become
unstable, even though its weight and thickness
is big enough; accordingly, the test value might
not be accurate. So the sample should be
reinforced or supported at its back.
Magnetism of the sample itself should be
avoided.
4.1.2 System Setting
See 6.9 for details.
4.1.3 Presetting Testing conditions
See 6.5 for details.
4.2 Testing Program
The hardness tester has to be verificated by using the
hardness block. The error and repeatability of the
displayed value should be within the regulation of
APPENDIX Table 2.
4.2.1 Start-up
1) The plug of the impact device has to be inserted into the
socket of the impact device of the instrument.
2) The key has to be pressed to power on. Now the
instrument is in testing condition.
4.2.2 Testing
- The release button on the upside of the impact device
has to be pressed to start testing. The sample and the
impact device as well as the operator are all required to be
stable now. The action direction should pass the axis of
the impact device.
- Each measurement area of the sample usually needs 3
to 5 times of testing operations. The result data dispersion
should not be more than the mean value ± 15 HL.
- The distance between any two impact points or from the
centre of any impact point to the edge of the sample
should conform to the regulation of Table 4-1.
- If an accurate conversion from the Leeb hardness value
to another hardness value is requested, a contrastive test
is needed to get conversion relations for the special
material. An inspected qualified Leeb hardness tester and
a corresponding hardness tester have to be used to test at
the same sample respectively. For each hardness value,
each homogeneous measurement, 5 points of Leeb
hardness value are needed. They are in the surroundings
of more than three indentations which need conversion of
hardness. The Leeb hardness arithmetic average value
and the corresponding hardness average value as
correlative value respectively are used to establish an
individual hardness contrastive curve. The contrastive
curve at least should include three groups of correlative
data.
Table 4-1
4.2.3 Reading the Measured Value
4.2.4 Power Off
The key has to be pressed to power off.
5. Advice
- The impact device must be replaced during the
instrument is powered off. Otherwise the type of impact
device cannot be identified by the main body. The circuit
board of the main body may get damaged.
- If the testing times are less than the preset times value,
the current test value can not be saved. 【 AVG 】 can be
pressed to end the testing process in advance if the values
shall be saved.
- When pressing 【 AVG 】 to end testing in advance, the
【 Auto save】 , 【 Auto transfer】 settings will not work.
- Only D and DC type of impact device have got the
function of strength measurement. Using other types of
impact device, the 【 Set hardness orбb 】 can not be
modified. The 【 Set hardness orбb 】 setting will be set
to【 Hardness 】 automatically after replacing the impact
device, whether the setting is【 Hardness 】 or not before.
- Not all materials can be converted to all hardness style
values. The hardness style is reset to HL automatically
after changing material. So, material has to be selected
first before changing the hardness style.
6. Operation in Details
6.1 Power On
The key has to be pressed to power on the instrument.
Following is shown on the screen:
The type of impact device will automatically be detected
during powering on. This information will be displayed on
the screen (Probe Type). After several seconds, the
HMR-BA-e-1212 4
Loading...
+ 8 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.