1. Technical Data
6 V 1.2 A
Battery Size 23A (1 x 1.5V)
KERN HFB 150K50 HFB 300K100 HFB 600K200
Readability (d) 50 g 100 g 200 g
Weighing range (max) 150 kg 300 kg 600 kg
Taring range (subtractive) 150 kg 300 kg 600 kg
Reproducibility 50 g 100 g 200 g
Linearity ±100 g ±200 g ±400 g
Recommended adjustment
weight, not added (class)
Stabilization time 2 s
Precision 0.2 % of max.
Warm-up time 10 min
Units kg, lb
Allowable ambient
temperature
Relative humidity 10 to 80 %, non-condensing
Supply voltage 220V - 240V AC 50 Hz
Secondary voltage
powerpack
Rechargeable battery
(standard equipment)
150 kg (M1) 300 kg (M1) 500 kg (M1)
0…+40 °C
9V, 800 mA
Service life – background illumination ON 30 h
Service life – background illumination OFF 40 h
Loading time 12 h
Display Digit height 25 mm
Housing size W x D x H 163 mm x 126 mm x 159 mm
Housing material Synthetic material
Material hook and shackle Nickel plated steel
Net weight 2 kg
Remote control (standard
equipment)
HFB-BA-e-1212 3
W x D x H 48 x 16 x 95 mm
4
1.1 Dimensions
HFB-BA-e-1212
1.2 EC-Declaration of -Conformity
KERN & Sohn GmbH
D-72322 Balingen-Frommern
E-Mail: info@kern-sohn.de
Tel: 0049-[0]7433- 9933-0
Internet: www.kern-sohn.de
D
Konformitätserklärung
Wir erklären hierm it, dass das Produkt, auf das sich diese Erk lärung bezieht,
mit den nachstehenden Normen übereinstimmt.
GB
Declaration of
conformity
We hereby declare that the product to which this dec laration refers c onforms
with the following standards.
CZ
Prohlášení o
shode
Tímto prohlašujem e, že v ýrobek , k terého se toto proh lášení týká , je v souladu
s níže uvedenými normami.
E
Declaración de
conformidad
Manifestamos en la presente que el producto al que se refiere esta
declaración está de acuerdo con las normas siguientes
F
Déclaration de
conformité
Nous déclarons avec cela res ponsabilité que le produit, auquel se r apporte la
présente déclaration, est conforme aux normes citées ci-après.
I
Dichiarazione di
conformitá
Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si
riferisce è conforme alle norme di seguito citate.
NL
Conformiteitverklaring
Wij verklaren hiermede dat het product, waarop deze verklaring betrekking
heeft, met de hierna vermelde normen overeenstemt.
P
Declaração de
conformidade
Declaramos por meio da presente que o produto no qual se refere esta
declaração, corresponde às normas seguintes.
PL
Deklaracja
zgodności
Niniejszym oświadczamy, że produkt, którego niniejsze oświadczenie dotyczy,
jest zgodny z poniższymi normami.
RUS
Заявление о
соответствии
Мы заявляем, что продукт, к которому относится данная декларация,
EG-Konformitätserklärung EC-Declaration of -Conformity
EC-
Déclaration de conformité EC-Declaración de Conformidad
EC-Dichiarazione di conformità EC-Conformiteitverklaring
EC- Declaração de conformidade EC- Prohlášení o shode
EC-Deklaracja zgodnościEC-Заявление о соответствии
Electronic Balan ce: KERN HFB
EN61000-3-3:1995+A1:2001+A2:2005
HFB-BA-e-1212 5
6
2. General Safety Instructions
Duties of the owner-operator
Follow the national accident prevention regulations as well as the working, operating
and safety regulations of the owner-operator.
Observe all safety regulations of the crane manufacturer.
The balance may only be used for the proposed purpose. Any type of use which
is not specified in these operating instructions, will be considered as improper
use. The customer is solely responsible for material damage and injury of persons
resulting from an improper use, Messrs. KERN & Sohn will not be liable under
any circumstance.
Messrs. KERN & Sohn cannot be held liable, if the crane scales are modified or
used improperly and if damage is resulting from such use.
Carry out regular maintenance and care of the crane scales, the crane and the
load suspension devices (see chapter 9).
Log the test result and keep it in the logbook.
Organizational measures
Only trained and instructed staff may operate the balance.
Make sure that the operating instructions are kept nearby the operation site of the
crane scales.
Assembly, commissioning and maintenance should only be carried out by trained
specialists.
Repair of safety-relevant pieces may only be carried out by KERN or by service
partners authorized by Messrs. KERN. (competence certificate or training).
Use original spare parts only.
All repairs and spare parts must be documented by the service partner (see list,
chap. 10.3).
All maintenance must be documented (see checklist chap. 9.3).
Load suspending components may only be exchanged as a complete spare parts
set. The dimensions of the new components must be noted (see checklist chapter
9.3).
Ambient conditions
Never operate the crane scales in explosive environment. The serial version is
not explosion protect e d.
Operate the crane scales only under environmental conditions as specified in
these operating instructions (especially in chapter 1 „Technical data“).
Do not expose the crane scales to strong humidity. Non-per mitt e d condensation
(condensation of air humidity on the appliance) may occur if a cold appliance is
taken to a considerably warmer environment. In this case, acclimatize the
disconnected appliance for ca. 2 hours at room temperature.
Do not operate the crane scales in corrosive environment.
Protect the crane scales against high humidity, vapours and dust.
HFB-BA-e-1212
Major display deviations (incorrect weighing results) may be experienced should
electromagnetic fields (e.g. due to mobile phones or radio equipment), static
electricity accumulations or instable power supply occur. Change locat ion or
remove source of interference.
Proper use
The balance you purchased is intended to determine the weighing value of material
to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the
material to be weighed is suspended on the crane hook only vertically, manually,
carefully and without jerks. As soon as a stable weighing value is reached the
weighing value can be read.
Use the crane scales only for lifting and weighing of freely movable loads.
Danger of injury due to improper use. Not allowed are e.g.:
- Exceeding the allowed nominal load of crane, crane scales or any type of
load attachment devic es
- Conveying persons,
- Pulling loads over an inclined surface,
- Tearing-off, pulling or towing loads.
Modifications or reconstructions of the crane scales or of the crane are not
allowed.
Improper Use
Do not use balance for dynamic weighing. In the event that small quantities are
removed or added to the material to be weighed, incorrect weighing results can be
displayed due to the “stability compensation“ in the balance. (Example: Slowly
draining fluids from a container suspended on the balance.) Do not leave permanent
load suspended on the balance. This may damage the measuring system as well as
safety-relevant parts.
The balance may only be used according to the described conditions. Other areas of
use must be released by KERN in writing.
Warranty
Warranty claims shall be voided in case
Our conditions in the operation manual are ignored
The appliance is used outside the described uses
The appliance is modified or opened
Mechanical damage and damage caused by media, liquids,
Natural wear and tear
The appliance is improperly set up or incorrectly electrically connected
The meas ur i ng sy s tem is overloaded
Safe working
Do not stand underneath suspended loads!
Position the crane in a way that the load is lifted vertically.
When working with the crane and crane scales wear personal safety equipment
(helmet, safety shoes etc.).
HFB-BA-e-1212 7
8
Monitoring of Test Resources
In the framework of quality assurance the measuring-related properties of the
balance and, if applicable, the testing weight, must be checked regularly. The
responsible user must define a suitable interval as well as type and scope of this test.
Information is available on KERN’s home page (www.kern-sohn.com) with regard to
the monitoring of balance test substances and the test weights required for this. In
KERN’s accredited DKD calibration laboratory test weights and balances may be
calibrated (return to the national standard) fast and at moderate cost.
Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the
appliance itself when unpacking for possible visible damage.
Initial Commissioning
In order to obtain exact results with the electronic balances, your balance must have
reached the operating temperature (see warming up time chap. 1).
During this warming up time the balance must be connected to the power supply
(mains, accumulator or battery).
The accuracy of the balance depends on the local acceleration of gravity.
Strictly observe hints in chapter Adjustment.
Check the original dimensions, see chap. 4.2.
Shutdown and storage
Take off the crane scales from the crane and dismantle all load attachment
devices from the crane scales.
Do not store the crane scales at open air
HFB-BA-e-1212
3. The crane scales at a glance
4
Safety bracket
The crane scales are a multi-purpose and cost-saving solution for overhead weighing
applications such as e.g. recycling, metal processing, machine engineering, transport
and logistics.
With the radio remote control operation will be more comfortable yet.
3.1 Overview
Fig. 1: Scales fr ont an d rear vi ew Fig. 2: Radio remote control
the capacity of the rechargeable battery will soon be
exhausted. The balance will be ready to operate for
about another 30 minutes, then it will switch off
automatically.
a the weight is in the area around the zero point
TARE the scales have been tared
HOLD the data-hold function is active
kg the current weighing unit is „kg“
lb the current weighing unit is „lb“
The LED display above is illuminated when the battery is being
recharged.
HFB-BA-e-1212
3.3 Keyboard
Button Designation Description of function
Switch-over weighing unit.
UNIT button
Exit menu, back to weighing mode.
HOLD button
TARE button Taring
ZERO button
ON/OFF key Turn on or off the balanc e
3.4 Numerica l input via the navigation buttons
Button Navigation button Function
ESC Delete
The weight display is „frozen“ via the HOLD button
until the HOLD-button is pressed anew.
Corrects the zero point of the balance. The display is
set to zero.
Digit selection
Increase flashi ng digit
HFB-BA-e-1212 11
Terminate input
12
3.5 Radio remote control
The balance can be operated by the radio remote control like by a keyboard. All
functions (excepted ON/OFF) can be selected.
The red LED must light up when any button is pressed. If it does not light up, the
batteries in the remote control must be exchanged.
Range on free surface (free of buildings) approx. 20 m.
3.6 Sticker
1
2
3
4
Do not stand or go under suspended loads.
Taring
Zeroing
Switch-over weighing unit
The weight display is „frozen“ via the
HOLD button until the HOLD-button is
pressed anew.
(Example)
Do not use on building sites.
Suspended loads have to be observed constantly.
Do not exceed the nominal load of crane scales.
The product meets the requirements of the German
Equipment and Product Safety Act.
HFB-BA-e-1212