KERN & SOHN ABJ 320-4NM User Manual

KERN & Sohn GmbH
Ziegelei 1
email: info@kern-sohn.com
Internet: www.kern-sohn.com
Operating instructions Analytical balance
KERN ABS-N_ABJ-NM
Version 1.2
03/2013
GB
D-72336 Balingen
Fax: +49-[0]7433-9933-149
ABS-N_ABJ-NM-BA -e-1312
KERN ABS-N_ABJ-NM
Analytical balance
GB
Version 1.2 03/2013
Operating instructions
Table of Contents
1 Technical Specifications ............................................................................... 4
2 Declaration of conformity ............................................................................. 7
3 Appliance overview ....................................................................................... 8
3.1 Keyboard overview ................................................................................................................... 9
3.1.1 Numeric entry ........................................................................................................................... 10
3.2 Overview of displays .............................................................................................................. 11
4 Basic Information (General) ........................................................................ 12
4.1 Proper use ............................................................................................................................... 12
4.2 Improper Use........................................................................................................................... 12
4.3 Warranty .................................................................................................................................. 12
4.4 Monitoring of Test Resources ............................................................................................... 13
5 Basic Safety Precautions ............................................................................ 13
5.1 Pay attention to the instructions in the Operatio n Manual ................................................ 13
5.2 Personnel training .................................................................................................................. 13
6 Transportation & Storage ............................................................................ 13
6.1 Testing upon acceptance ...................................................................................................... 13
6.2 Packaging / retur n transport ................................................................................................. 14
7 Unpacking, Setup and Commissioning ..................................................... 16
7.1 Installation Site, Lo cation of Use .......................................................................................... 16
7.2 Unpacking and checking ....................................................................................................... 16
7.2.1 Scope of delivery / serial accessories....................................................................................... 17
7.2.2 Placing ...................................................................................................................................... 18
7.3 Mains connection ................................................................................................................... 19
7.3.1 Turning On the Power............................................................................................................... 19
7.4 Initial Commissioning ............................................................................................................ 19
7.5 Connection of periphera l devices ......................................................................................... 19
8 Adjustment ................................................................................................... 20
8.1 Automatic adjustment by PSC (Pe rfect Self Calibration), only models ABJ .................... 20
8.2 Menu settings „I.CAL“ / „E.CAL“ .......................................................................................... 23
8.2.1 Adjustment with internal weight (KERN ABJ) ........................................................................... 24
8.2.2 Adjustment with external weight (KERN ABS).......................................................................... 25
8.3 Adjustment lo g ........................................................................................................................ 26
8.4 Balance identification no. ...................................................................................................... 27
9 Verification ................................................................................................... 28
10 Weighing ....................................................................................................... 30
10.1 Switch-on balance / call-up weighing mode ........................................................................ 30
10.2 Switch off the balance ............................................................................................................ 30
10.3 Set the start display................................................................................................................ 31
10.4 Auto Power-Off function (auto matic switch-off function) .................................................. 33
10.5 Simple weighing ..................................................................................................................... 34
10.6 Taring ....................................................................................................................................... 35
10.7 Switch-over weighing unit ..................................................................................................... 36
2 ABS-N_ABJ-NM-BA -e-1312
10.7.1 Freely programmable weighing unit ..................................................................................... 37
10.7.2 Change position of decimal dot for freely programmable weighing unit ............................... 38
10.7.3 Input minimum weight for freely programmable weighing unit ............................................. 38
10.8 Change readability (1D/10D) .................................................................................................. 39
10.9 Display decimal dot as point or comma ............................................................................... 40
11 Menu ............................................................................................................. 41
11.1 Menu symbol ........................................................................................................................... 41
11.2 Navigation in the menu .......................................................................................................... 42
11.3 Menu overview ........................................................................................................................ 43
11.4 Resetting the menu ................................................................................................................ 44
11.5 Menu Lock ............................................................................................................................... 45
11.5.1 Change password ................................................................................................................ 46
11.6 Log menu settings .................................................................................................................. 47
12 Zero / tare me nu (zeroing and taring functions) ....................................... 48
12.1 Zero tracking function ............................................................................................................ 49
12.2 Auto zero functi o n .................................................................................................................. 50
12.3 Auto Tare function .................................................................................................................. 51
12.4 Zero / tare timing cha n g e function........................................................................................ 52
13 Settings for Stability and Response ........................................................... 53
13.1 Stability and reaction settings via „Easy Setting display“ (without invoking menu) ...... 53
13.2 Selection weighing / dosing mo d e ....................................................................................... 54
13.3 Standstill width ....................................................................................................................... 55
14 Application Functions ................................................................................. 57
14.1 Parts counting ........................................................................................................................ 57
14.2 Percent determination ............................................................................................................ 60
14.2.1 Reference weight = 100 % ................................................................................................... 61
14.2.2 User-defined reference ........................................................................................................ 62
14.2.3 Percent determination .......................................................................................................... 63
14.3 Formula mode ......................................................................................................................... 64
14.3.1 Enable edition „component no.“ ........................................................................................... 66
14.3.2 Enable issue „TOTAL“ .......................................................................................................... 67
15 Check weighing and targe t mode (check and target weighing) .............. 68
15.1 Check weighing ...................................................................................................................... 68
15.2 Target mode ............................................................................................................................ 72
16 Data output ................................................................................................... 75
16.1 Pin connection ........................................................................................................................ 75
16.2 Issue functions ....................................................................................................................... 76
16.2.1 Automatic data output / Auto Print function .......................................................................... 76
16.2.2 Continuous Output function .................................................................................................. 78
16.2.3 Output Timing Change function ........................................................................................... 80
16.2.4 GLP Output function / balance identification number ........................................................... 81
16.3 Communication parameters .................................................................................................. 82
16.3.1 Selection of a standard setting „MODE 1 - MODE 5“ ........................................................... 83
16.3.2 User-defined settings „MODE U“ / setting for KERN –YKB-01N ......................................... 84
16.4 Data Formats ........................................................................................................................... 88
16.5 Remote control instructions .................................................................................................. 92
17 Service, maintenance, disposal .................................................................. 93
17.1 Cleaning................................................................................................................................... 93
17.2 Service, maintenance ............................................................................................................. 95
17.3 Disposal ................................................................................................................................... 95
18 Instant help ................................................................................................... 95
18.1 Error messages ....................................................................................................................... 96
ABS-N_ABJ-NM-BA-e-1312 3

1 Technical Specifications

KERN ABJ
80-4NM
Readability (d) 0.1 mg 0.1 mg 0.1 mg 0.1 mg
Weighing range (max) 80 g 120 g 220 g 320 g
Minimum load (Min) 10 mg 10 mg 10 mg 10 mg
Verification value (e) 1 mg 1 mg 1 mg 1 mg
Verification class I I I I
Reproducibility 0.2 mg 0.2 mg 0.2 mg 0.2 mg
Linearity ± 0.3 mg ± 0.3 mg ± 0.3 mg ± 0.3 mg
Stabilization time 3 sec.
Adjustment weight internal
Warm-up time 4 h 8 h 8 h 8 h
Weighing Units mg, g mg, g, ct
Smallest part weight for piece counting
ABJ
120-4NM
1 mg
ABJ
220-4NM
ABJ
320-4NM
Reference quantities at piece counting
Weighing plate, stainless steel
Dimensions of the housing (B x D x H) [mm]
Dimensions Glass wind screen [mm]
Net weight (kg) 6
Permissible ambient condition
Humidity of air 20 ~ 85 % relative (not condensing)
AC adapter (Primary) AC 100 -240 V, 400 mA 50/60Hz
Rated electric power supply DC 12 V, 1.25 A
Pollution Degree 2
Overvoltage Category Category II
174 x 162 x 227 (weighing space)
5, 10, 20, 50,100
ø 91mm
210 x 340 x 325
+10° C bis +30° C
Altitude Up to 2000 m
Installation Site device may only be used indoors
4 ABS-N_ABJ-NM-BA -e-1312
KERN ABS 80-4N ABS 120-4N
Readability (d) 0.1 mg 0.1 mg
Weighing range (max) 80 g 120 g
Reproducibility 0.2 mg 0.2 mg
Linearity ± 0.3 mg ± 0.3 mg
Recommended adjusting weight not supplied (class)
80 g (E2) 100 g (E2)
Warm-up time 4 h 8 h
Stabilization time 3 sec.
Weighing Units
Smallest part weight for piece counting
Reference quantities at piece counting
Weighing plate, stainless steel
Dimensions of the housing (B x D x H) [mm]
Dimensions Glass wind screen [mm]
mg, g mg, g, ct
1 mg
5, 10, 20, 50,100
ø 91mm
210 x 340 x 325
174 x 162 x 227 (weighing space)
Net weight (kg) 6
Permissible ambient condition
+5° C to +40° C
Humidity of air 20 ~ 85 % relative (not condensing)
AC adapter (Primary) AC 100 -240 V, 400 mA 50/60Hz
Rated electric power supply DC 12 V, 1.25 A
Pollution Degree 2
Overvoltage Category Category II
Altitude Up to 2000 m
Installation Site device may only be used indoors
ABS-N_ABJ-NM-BA-e-1312 5
Smallest part weight for piece
KERN ABS 220-4N ABS 320-4N
Readability (d) 0.1 mg 0.1 mg
Weighing range (max) 220 g 320 g
Reproducibility 0.2 mg 0.2 mg
Linearity ± 0.3 mg ± 0.3 mg
Recommended adjusting weight not supplied (class)
200 g (E2) 300 g (E2)
Stabilization time 3 sec.
Warm-up time 8 h
Weighing Units
counting
Reference quantities at piece counting
Weighing plate, stainless steel
Dimensions of the housing (B x D x H) [mm]
Dimensions Glass wind screen [mm]
174 x 162 x 227 (weighing space)
mg, g, ct
1 mg
5, 10, 20, 50,100
ø 91mm
210 x 340 x 325
Net weight (kg) 6
Permissible ambient condition
+5° C to +40° C
Humidity of air 20 ~ 85 % relative (not condensing)
AC adapter (Primary) AC 100 -240 V, 400 mA 50/60Hz
Rated electric power supply DC 12 V, 1.25 A
Pollution Degree 2
Overvoltage Category Category II
Altitude Up to 2000 m
Installation Site device may only be used indoors
6 ABS-N_ABJ-NM-BA -e-1312

2 Declaration of c onf ormity

KERN & Sohn GmbH
D-72322 Balingen-Frommern
EN
Konformitäts­erklärung
Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den nachstehenden Normen übereinstimmt.
GB
Declaration of conformity
We hereby declare that the product to which this declaration refers conforms to the following standards.
CZ
Prohlášení o shode
Tímto prohlašujeme, že výrobek, kterého se toto prohlášení týká, je v souladu s níže uvedenými normami.
E
Declaración de conformidad
Manifestamos en la presente que el producto al que se refiere esta declaración está de acuerdo con las normas siguientes
F
Déclaration de conformité
Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente déclaration, est conforme aux normes citées ci-après.
I
Dichiarazione di conformitá
Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce è conforme alle norme di seguito citate.
NL
Conformiteit­verklaring
Wij verklaren hiermede dat het product, waarop deze verklaring betrekking heeft, met de hierna vermelde normen overeenstemt.
P
Declaração de conformidade
Declaramos por meio da presente que o produto no qual se refere esta declaração, corresponde às normas seguintes.
PL
Deklaracja zgodności
Niniejszym oświadczamy, że produkt, którego niniejsze oświadczenie dotyczy, jest zgodny z poniższymi normami.
RUS
Заявление о соответствии
Мы заявляем, что продукт, к которому относится данная декларация, соответствует перечисленным ниже нормам.
Signature
Datum
27.03.2013
Ort der Ausstellung
Place of issue
72336 Balingen
Albert Sauter KERN & Sohn GmbH
Geschäftsführer
Managing director
KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0 Fax +49-[0]7433/9933-149, E-Mail: info@kern-sohn.com, Internet: www.kern-sohn.com
Postbox 4052 email: info@kern-sohn.de
Fax: 0049-[0]7433-9933-149 Internet: www.kern-sohn.de
Declaration of conformity
EC-Konformitätserklärung EC-Declaration of -Conformity
Déclaration de conformité EC-Declaración de Conformidad
EC­EC-Dichiarazione di conformit à EC-Conformiteitverklaring EC- Declaração de conformidade EC- Prohlášení o shode EC-Deklaracja zgodności EC-Заявление о соответствии

Electronic Balan ce: KERN ABJ-NM, ABS-N

EU Directive Standards
2004/108/EC EN 61326-1:2006
2006/95/EC EN 61010-1: 2010
Date
Signatur
ABS-N_ABJ-NM-BA-e-1312 7

3 Appliance overview

1. Weighing plate
2. Display
3. Keyboard
4. Levelling screw
5. Bubble level
6. Name plate
7. Wind protection
8. Appliance interface
9. Mains adapter connection
8 ABS-N_ABJ-NM-BA -e-1312

3.1 Keyboard overview

Pressed once and
released
Keep pressed for
about 3 seconds
the operation and
Cancel procedure
(PCS, %).
Switch over
menu
In menu:
Key Designation:
Function
In menu
Menu item back.
Switches between
ON/OFF
CAL
TARE
UNIT
Navigation key
MENU
Navigation key
PRINT
Navigation key
standby modes.
Start adjustment
Tare or set weight display to zero
Switch-over weighing unit. Display stored reference weight
Calculate weighing data via interface
Invoke calibration menu
Invoke zero / tare menu
Invoke unit setting menu.
weighing mode / application mode
Invoke Data Output menu
-
Return to weighing mode:
Press ON/OFF
repeatedly or for 3 s.
-
Confirm entry
Scroll forward in menu
Invoke main menu (press twice) Scroll backwards in
Select next menu item.
ABS-N_ABJ-NM-BA-e-1312 9

3.1.1 Numeric entry

Key Designation: Function
Navigation key
Navigation key
Navigation key Digit selection to the right
Navigation key Confirm entry
ESC Cancel input
With numerical input the indicator [#] appears.
Increase flashing digit
Shift flashing decimal dot to the left
Decrease flashing digit
Shift flashing decimal dot to the right
10 ABS-N_ABJ-NM-BA -e-1312

3.2 Overview of displays

Chpt. 14.3
Units
Symbols
Numerical display/menu
Display Description see chpt.
Battery symbol
Zero tracking function active
Displays adjustment
Flashes prior to start of automatic adjustment, only models ABJ
Stability and reaction settings via „Easy Setting display“ +
Balance is in dosing mode (pouring mode) +
Balance is in formula mode +
Menu lock active +
Menu symbol +
Auto Print function active +
Not documented
Log menu settings
Tolerance marks in check weighing and target mode + Chpt. 15
The stability display indicates that the weighing value is stable. Highlights current setting in the menu.
Negative weighed value
Stand-by mode Ready for start formulation Ready for start continuous data output („MANU ON“)
Chpt.12.1
+
+
Chpt. 8.2.1
+
Chpt. 8.1
Chpt.13.1
Chpt. 13.2
Chpt. 14.3
Chpt. 11.5
Chpt. 11.2
Chpt. 16.2.1
+
Chpt. 10.2
+ Chpt. 14.3 + Chpt. 16.5.2
ABS-N_ABJ-NM-BA-e-1312 11
Indicates numeric value entry. +
Shows the stored reference in piece number or percent determination mode.
Chpt. 3.1.1
+
Chpt.14.1
+ Chpt.14.2.3
Net weight in formula mode +
Total weight (TOTAL) of all components in formula mode +
Memory space piece weight +
Change position decimal dot for freely programmable weighing unit +
Balance is in piece counting mode +
Balance is in percent determination mode with user-defined reference weight + Chpt.14.2.1
Balance is in percent determination mode, reference weight =100% +
Chpt. 14.3
Chpt.14.1
Chpt.10.7.2
Chpt.14.1
Chpt.14.2.2

4 Basic Information (General)

4.1 Proper use

The balance you purchased is intended to determine the weighing value of material to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing plate. As soon as a stable weighing value is reached the weighing value can be read.

4.2 Improper Use

Do not use balance for dynamic add-on weighing procedures, if small amounts of goods to be weighed are removed or added. The “stability compensation“ installed in the balance may result in displaying an incorrect measuring value! (Example: Slowly draining fluids from a container on the balance.) Do not leave permanent load on the weighing plate. This may damage the measuring system. Impacts and overloading exceeding the stated maximum load (max) of the balance, minus a possibly existing tare load, must be strictly avoided. Balance may be damage by this. Never operate balance in explosive environment. The serial version is not explosion protected. The structure of the balance may not be modified. This may lead to incorrect weighing results, safety-related faults and destruction of the balance. The balance may only be used according to the described conditions. Other areas of use must be released by KERN in writing.

4.3 Warranty

Warranty claims shall be voided in case
Our conditions in the operation manual are ignored
The appliance is used outside the described use
The appliance is modified or opened
Mechanical damage or damage by media, liquids, natural wear and tear
The appliance is improperly set up or incorrectly electrically connected
The measuring system is overloaded
12 ABS-N_ABJ-NM-BA -e-1312

4.4 Monitoring of Test Resources

In the framework of quality assurance the measuring-related properties of the balance and, if applicable, the testing weight, must be checked regularly. The responsible user must define a suitable interval as well as type and scope of this test. Information is available on KERN’s home page (www.kern-sohn.com with regard to the monitoring of balance test substances and the test weights required for this. In KERN’s accredited DKD calibration laboratory test weights and balances may be calibrated (return to the national standard) fast and at moderate cost.

5 Basic Safety Precautions

5.1 Pay attention to the instructions in the Operation Manual

Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances.

5.2 Personnel training

The appliance may only be operated and maintained by trained personnel.

6 Transportation & Storage

6.1 Testing upon acceptance

When receiving the appliance, please check packaging immediately, and the appliance itself when unpacking for possible visible damage.
ABS-N_ABJ-NM-BA-e-1312 13

6.2 Packaging / return transport

Keep all parts of the original packaging for a possibly
required return.
Only use original packaging for returning.
Prior to dispatch disconnect all cables and remove
loose/mobile parts.
Reattach possibly supplied transport securing devices.
Secure all parts such as the glass wind screen, the weighing
platform, power unit etc. against shifting and damage.
14 ABS-N_ABJ-NM-BA -e-1312
ABS-N_ABJ-NM-BA-e-1312 15

7 Unpacking, Setup and Commissioning

7.1 Installation Site, Location of Use

The balances are designed in a way that reliable weighing results are achieved in common conditions of use. You will work accurately and fast, if you select the right location for your balance.
Therefore, observe the following for the installation site:
Operate the device only indoors.
Place the balance on a firm, level surface;
Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
Protect the balance against direct draughts due to open windows and doors;
Avoid jarring during weighing;
Protect the balance against high humidity, vapours and dust;
Do not expose the device to extreme dampness for longer periods of time.
Non-permitted condensation (condensation of air humidity on the appliance) may occur if a cold appliance is taken to a considerably warmer environment. In this case, acclimatize the disconnected appliance for ca. 2 hours at room temperature.
Avoid static charge of goods to be weighed or weighing container.
If electro-magnetic fields or static charge occur, or if the power supply is unstable major deviations on the display (incorrect weighing results) are possible. In that case, the location must be changed.

7.2 Unpacking and checking

Open package, take out the appliance and accessories. Verify that there has been no damage and that all packing items are present.
16 ABS-N_ABJ-NM-BA -e-1312

7.2.1 Scope of delivery / serial accessories

1. Balance
2. Weighing plate
3. Carrier weighing plate
4. Screening ring
5. Mains adapter
6. Operating instructions
7. Menu overview
ABS-N_ABJ-NM-BA-e-1312 17

7.2.2 Placing

Attach circular screen, carrier of weighing plate and weighing plate in order.
Level balance with foot screws until the air bubble of the water balance is in the
prescribed circle.
Check levelling regularly
18 ABS-N_ABJ-NM-BA -e-1312

7.3 Mains connection

Supply power to balance via mains adapter. The display lights up and the
The selftest is completed when „OFF“ appears on the display.
Power is supplied via the external mains adapter. The stated voltage value must be the same as the local voltage. Only use original KERN mains adapters. Using other makes requires consent by KERN.

7.3.1 Turning On the Powe r

balance carries out a selftest. In the ABJ models an automatic adjustment takes place.

7.4 Initial Commissioning

In order to obtain exact results with the electronic balances, your balance must have reached the operating temperature (see warming up time chpt. 1). During this warming up time the balance must be connected to the power supply (mains, accumulator or battery).
The accuracy of the balance depends on the local acceleration of gravity. Strictly observe hints in chapter Adjustment.

7.5 Connection of peripheral devices

Before connecting or disconnecting of additional devices (printer, PC) to the data interface, always disconnect the balance from the power supply. With your balance, only use accessories and peripheral devices by KERN, as they are ideally tuned to your balance.
ABS-N_ABJ-NM-BA-e-1312 19

8 Adjustment

As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated - in compliance with the underlying physical weighing principle - to the existing acceleration due to gravity at its place of location (only if the balance has not already been adjusted to the location in the factory). This adjustment process must be carried out for the first commissioning, after each change of location as well as in case of fluctuating environment temperature. To receive accurate measuring values it is also recommended to adjust the balance periodically in weighing operation. Observe stable environmental conditions. A warming up time (see chapter 1) is required for stabilization. Ensure that there are no objects on the weighing plate.

8.1 Automatic adjustment by PSC (Perfect Self Calibration), only models ABJ

The default setting for balances of the series ABJ prompts automatic adjustment via the PSC function (cannot be switched off). The moment a change in temperature is detected this function uses the internal tare weight to carry out fully-automatic adjustment with the help of a temperature sensor. Adjustment in weighing mode is carried out automatically under the following conditions:
(1) When there is a change in the surrounding temperature (t 2°C)
(2) When about four hours has passed since the previous calibration.
(3) When the balance is switched from standby status to weighing mode
and condition (1) or (2) has been met.
If one of the above conditions was met in weighing mode, the weight symbol flashes for about two minutes in order to notify the pending adjustment;
20 ABS-N_ABJ-NM-BA -e-1312
Case 1: The load on the weighing plate is near zero.
The weight symbol flashes for approx. two minutes, followed by „PSC.RUN“.
After that the internal adjustment is started automatically. In order to ensure proper PSC operation, prevent vibrations and air flow.
As soon as the display in grams reappears after completing adjustment via PSC, the balance returns to weighing mode.
ABS-N_ABJ-NM-BA-e-1312 21
To prevent a start-up of adjustment during a measuring process, actuate the
again for two minutes, followed by „PLS.CAL“.
Case 2: Weighing plate loaded
The gram display flashes for two minutes, followed by „PLS.CAL“.
Unload weighing plate. The gram display flashes again for 2 minutes; after that the internal adjustment is started automatically. In order to ensure proper PSC operation, prevent vibrations and air flow.
As soon as the display in grams reappears after completing adjustment via PSC, the balance returns to weighing mode.
ON/OFF button as soon as „PLS.CAL“ is displayed. The gram display flashes
22 ABS-N_ABJ-NM-BA -e-1312

8.2 Menu settings „I.CAL“ / „E.CAL“

I.CAL:
Adjustment with internal weight (see chpt.
8.1)
E.CAL:
Adjustment with external weight (see chpt.
8.2)
The saved adjustment (I.CAL or E.CAL) can now directly be invoked via the
CAL button.
To invoke the adjustment function in weighing mode,
press CAL for 3 sec.
Acknowledge using PRINT, the current setting is
displayed.
Select the desired adjustment using the navigation keys
( )
Confirm with TARE
Press ON/OFF repeatedly or 3 sec., the balance will
return into weighing mode
ABS-N_ABJ-NM-BA-e-1312 23

8.2.1 Adjustment with internal weight (KERN ABJ)

When an optional printer is connected and the GLP function activated,
fter the
With the internal adjustment weight, the weighing accuracy can be checked and re­adjusted at any time.
Condition: Menu setting „I.CAL“.
while starting the adjustment log „WAIT“ will be displayed. A
printout the adjustment will be continued automatically.
Adjustment may be interrupted with ON/OFF, „ABORT“ is displayed.
Press the CAL-button and adjustment will take place
automatically.
The indicator will be shown.
After successful adjustment the balance automatically
returns to weighing mode. In case of an adjustment error (e.g. objects on the weighing plate) the display will show an error message, repeat adjustment. When an optional printer is connected and the GLP function is connected, the adjustment log will be edited, see chpt. 8.3.
24 ABS-N_ABJ-NM-BA -e-1312

8.2.2 Adjustment with external we ight (KERN ABS)

When an optional printer is connected and the GLP function activated,
After the
g the adjustment process within 60 s no operation is carried out,
Condition: Menu setting „E.CAL“.
The weight to be used depends on the capacity of the scale. Carry out
adjustment as near as possible to the balance’s maximum weight (recommended adjustment weight see chpt. 1). Weights of different nominal values or tolerance classes may be used for adjustment but are not optimal for technical measuring. The accuracy of the adjustment weight must correspond approximately to or, if possible, be better than, the readability of the balance. Minimum weight „Adjustment weight“:
ABS 80-4N / ABS 120-4N : 50 g ABS 220-4N / ABS 320-4N:100 g
Information about test weights you will find in the internet under:
http://www.kern-sohn.com
while starting the adjustment log „WAIT“ will be displayed.
printout the adjustment process will be continued automatically.
If durin
ERR C“ will be displayed. Press ON/OFF a restart.
In weighing mode press CAL. The weight value of the
recommended adjustment weight (see chpt. 1) appears
flashing. The indicator will be shown.
If the value shall be changed, press MENU, the active
digit flashes. Carry out the desired setting using navigation buttons (see chapter 3.1.1 „Numeric input“).
With flashing adjustment weight display, place the
adjustment weight carefully on the center of the weighing plate within 60 s. Close wind screen doors completely.
Wait until the zero display flashes.
Take away the adjustment weight and close the wind
screen doors. After successful adjustment the balance automatically returns to weighing mode. In case of an adjustment error (e.g. objects on the weighing plate) the display will show an error message, repeat adjustment. When an optional printer is connected and the GLP function is connected, the adjustment log will be edited, see chpt. 8.3.
ABS-N_ABJ-NM-BA-e-1312 25

8.3 Adjustment log

- - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - -
KERN & Sohn GmbH
Company
Balance identification no. (see chpt. 8.4)
-COMPLETE
-SIGNATURE-
prepared by
- - - - - - - - - - - - - - - - -
Make sure that the communication parameter of balance and printer are the
Call function
This function enables automatic log issue after each adjustment. These logs may be issued by using an optional printer.
Printout example (KERN YKB-01N):
CAL –EXTERNAL
TYPE ABJ 220-4NM
SN WBIIAB000I
ID 1234
REF 200.0000g
BFR 200.0001g
AFT 200.0000g
Mode of adjustment
Model
Serial no.
Used adjustment weight
Before adjustment
After adjustment
same.
In weighing mode press the CAL key for 3 sec. Use the navigation buttons ( ) to select „GLP.OUT“.
The current setting is marked by the stability display ( )
With stability display ( ) Function activated
Without stability display ( ) Function disabled
26 ABS-N_ABJ-NM-BA -e-1312
How to change settings Select menu item
Enter balance identification no. (max. 4 characters)
Press TARE
Press ON/OFF repeatedly or for 3 sec., the balance will
return into weighing mode

8.4 Balance identification no.

This setting is for the balance ID number that is output along with the adjustment report.
Call up menu
In weighing mode press the MENU button twice
Press the navigation buttons ( ) repeatedly until
TOOLS“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly
until „SYSTEM“ is displayed.
Confirm with PRINT
Press the navigation buttons ( ) repeatedly until
BAL.ID“ is displayed.
Press TARE, the currently set ID no. is displayed
(factory setting 0000).
Enter the desired ID no. using navigation buttons (see
chapter 3.1.1 „Numeric input“).
Press ON/OFF repeatedly or 3 sec., the balance will
return into weighing mode
ABS-N_ABJ-NM-BA-e-1312 27

9 Verification

General introduction:
According to EU directive 90/384/EEC or 2009/23EG balances must be officially verified if they are used as follows (legally controlled area):
a) For commercial transactions if the price of goods is determined by weighing. b) For the production of medicines in pharmacies as well as for analyses in the
medical and pharmaceutical laboratory. c) For official purposes d) For manufacturing final packages
In cases of doubt, please contact your local trade in standard.
Verification notes:
An EU type approval exists for balances described in their technical data as verifiable. If a balance is used where obligation to verify exists as described above, it must be verified and re-verified at regular intervals. Re-verification of a balance is carried out according to the respective national regulations. The validity for verification of balances in Germany is e.g. 2 years. The legal regulation of the country where the balance is used must be observed!
Verification of the balance is invalid without the seal.
The seal marks attached on verified balances point out that the balance may only be opened and serviced by trained and authorised specialist staff. If the seal mark is destroyed, verification looses its validity. Please observe all national laws and legal regulations. In Germany a re-verification will be necessary.
28 ABS-N_ABJ-NM-BA -e-1312
Position of seals
Seal
Data plate
Seal
ABS-N_ABJ-NM-BA-e-1312 29

10 Weighing

10.1 Switch-on balance / call-up weighing mode

Status balance Call-up weighing mode
Display switched off
Display
Display
All segments light up
Balance is in menu
After numeric input

10.2 Switch off the balance

Press ON/OFF. The balance is in standby mode, that
means that the balance is now in state ready-for­operation. Immediately after switching-on it is ready for operation (press any key) without warm-up time.
Press ON/OFF.
After display press any key
Press any key
Press ON/OFF repeatedly or for 3 sec.
To switch-off the balance completely, separate balance
from power supply.
If [WAIT] or [SET] are displayed, do not separate the balance from power
supply.
30 ABS-N_ABJ-NM-BA -e-1312

10.3 Set the start display

For the start display you can choose between three types.
Start display Explanation Menu setting
1. Weighing mode
2. Display
3. All segments
After connection to the power supply the balance will start in weighing mode.
After connection to the power supply the balance displays . After pressing a certain key the balance will carry out a segment test and start in weighing mode automatically.
After connection to the power supply the balance displays . After pressing any key the balance will carry out a segment test. Only after having pressed
TARE the weighing mode will
start.
ABS-N_ABJ-NM-BA-e-1312 31
Select menu item
Set display type
Call up menu
In weighing mode press the MENU button twice
Press the navigation buttons ( ) repeatedly until
TOOLS“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly
until „SYSTEM“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly
until „START“ is displayed. The current setting is marked by the stability display ( ).
Acknowledge using PRINT, the current setting is
displayed.
Use the navigation buttons () to select the desired
setting, e.g. „SEM.AUTO“.
Weighing mode
All segments
Confirm with TARE. The current setting is marked by the
stability display ( ).
Return to weighing mode
Press ON/OFF repeatedly or for 3 sec.
display
32 ABS-N_ABJ-NM-BA -e-1312

10.4 Auto Power-Off function (automatic switch-off function)

Stability display ( )
yes
no
To save battery when function is enabled backlight of display will automatically switch off after a defined time during which there was no change of load or activity.
1. Call up menu In weighing mode press the MENU button twice
2. Select function
Press the navigation buttons () repeatedly until
TOOLS“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly
until „SYSTEM“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly
until „AUTO.OFF“ is displayed. The current setting is marked by the stability display ( ).
Further steps depend on the desired setting:
Function set/update cancel
ON
off
 
Press PRINT and
continue with step 3.
Press TARE and
continue with step 3.
Press TARE and
continue with step 4.
continue with step 4.
3. Set time, after which the display has to switch off
Use the navigation key ( ) to enter the desired time in
Confirm with TARE.
4. Return to weighing mode
Press ON/OFF repeatedly or for 3 sec.
ABS-N_ABJ-NM-BA-e-1312 33
minutes (max. 99 min.), „Numeric input“. see chpt. 3.1.1.

10.5 Simple weighing

- - - - - - - - - - - - - - - - - KERN & Sohn GmbH
Balance identification no. (see chpt. 8.4)
- - - - - - - - - - - - - - - - -
50.0010 g
Measuring Value
-SIGNATURE-
prepared by
- - - - - - - - - - - - - - - - -
50.0010 g
Measuring Value
A warm-up time is required for stabilisation (see chpt. 1).
Wait for zero display, reset to zero using TARE.
Place the goods to be weighed and close the wind screen doors
Wait until the stability display appears [ ].
Read weighing result.
When an optional printer is connected, the weighing value can be edited.
Print-out example with enabled GLP function (see chpt. 8.3):
Company
TYPE ABJ 220-4NM
SN WBIIAB000I
Model
Serial no.
ID 1234
Print-out example with disabled GLP function (see chpt. 8.3):
34 ABS-N_ABJ-NM-BA -e-1312

10.6 Taring

When the balance is unloaded the saved taring value is displayed with
To delete the stored tare value, remove load from weighing pan and
The dead weight of any weighing container may be tared away by pressing a button, so that the following weighings show the net weight of the goods to be weighed.
Put vessel of goods to be weighed on weighing plate and close the wind
screen doors.
Wait until the stability display appears ( ), then press TARE. The weight of
the container is now internally saved.
Weigh the goods to be weighed and close the wind screen doors.
Wait until the stability display appears ( ).
Read net weight.
Note:
The balance is able to only store one taring value at a time.
negative sign.
press TARE.
The taring process can be repeated any number of times. The limit is reached when the whole weighing range is exhausted.
ABS-N_ABJ-NM-BA-e-1312 35

10.7 Switch-over w eighing unit

Enable/disable units
on the balance, the unit in which the balance has been
By pressing the UNIT key, the display can be switched over to the units enabled
before in the menu.
Call up menu
In weighing mode press the UNIT key for 3 sec. The
current setting is marked by the stability display ( ).
With stability display ( ) Unit enabled
Without stability display ( ) Unit disabled
Press TARE
Use the navigation buttons ( ) to select more units and to enable/disable them as described before.
Press ON/OFF repeatedly or for 3 sec., the balance will
return into weighing mode
How to change weighing units
To return to the previously enabled weighing units go to
weighing mode by UNIT.
When switching­switched off, will be displayed.
36 ABS-N_ABJ-NM-BA -e-1312

10.7.1 Freely programmable weighing unit

if necessary enable units
With a numeric input of the conversion factor, the position of the decimal
For the freely programmable weighing unit, no unit symbol will appear in the display.
Call up menu
In weighing mode press the UNIT key for 3 sec. Use the
navigation buttons ( ) to select „UNIT.U“. The
current setting is marked by the stability display ( ).
With stability display ( ) Unit enabled
Without stability display ( ) Unit disabled
Press TARE
Enter conversion factor
1. Press TARE and current setting will be displayed.
2. Use the navigation buttons to enter the desired
conversion factor (see chapter 3.1.1 „Numeric input“).
3. Press ON/OFF repeatedly or for 3 sec., the balance will
return into weighing mode
How to change weighing units
To return to the previously enabled weighing units go to weighing mode by
UNIT.
ABS-N_ABJ-NM-BA-e-1312 37
dot can be changed, see chpt. 10.7.2

10.7.2 Change pos ition of decimal dot for freely programmable weighing unit

The position of the decimal dot can only be changed by a numeric input of the conversion factor (see chpt. 9.7.1, step 2).
When the first cipher flashes, press PRINT repeatedly
until the decimal dot flashes.
Select the desired position using the navigation keys
( ). If no decimal dot shall be set, press MENU repeatedly until the display symbol appears.
Confirm with TARE.
Press ON/OFF repeatedly or for 3 sec., the balance will
return into weighing mode

10.7.3 Input minimum weight for freely programmable weighing unit

Call up menu
In the menu item „CONV.K“ (see chpt. 9.7.1) use the
navigation buttons to select ( ) „MIN.D“.
Press TARE and current setting will be displayed.
Use the navigation buttons to enter the desired
minimum weight, see chapter 3.1.1 „Numeric input“.
Press ON/OFF repeatedly or for 3 sec., the balance will
return into weighing mode
38 ABS-N_ABJ-NM-BA -e-1312

10.8 Change readability (1D/10D)

Select menu item
Change readability from 1D to 10 D
Call up menu
In weighing mode press the MENU button twice
Press the navigation buttons ( ) repeatedly until
TOOLS“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly
until „TARGET“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly
until „PARAM.W“ is displayed.
 Confirm with PRINT  Press the navigation buttons ( ) repeatedly
until „CHG.MIN“ is displayed.
Confirm with PRINT
1. Use the navigation buttons ( ) to select „10 D“.
2. Confirm with TARE. The current setting is marked by the
stability display ( ).
3. Press ON/OFF repeatedly or for 3 sec., the balance will
return into weighing mode
In order to reset readability to 1D, repeat step 1 – 3 accordingly for 1D.
ABS-N_ABJ-NM-BA-e-1312 39

10.9 Display decimal dot as point or comma

Select menu item
Selection point/comma
Call up menu
In weighing mode press the MENU button twice
Press the navigation buttons ( ) repeatedly until
TOOLS“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly
until „SYSTEM“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly
until „POINT“ is displayed.
Acknowledge using PRINT, the current setting is
displayed.
Select the desired adjustment using the navigation keys
( )
PERIOD:
COMMA
Confirm with TARE. The current setting is marked by the
stability display ( ).
Decimal dot is displayed as a point
Decimal dot is displayed as a comma
Return to weighing mode
Press ON/OFF repeatedly or for 3 sec.
40 ABS-N_ABJ-NM-BA -e-1312

11 Menu

To adapt the behaviour of the balance to your requirements, go to the menu. Usually the default setting of the menu is such that you do not need to make any changes. If you encounter special conditions of use, go to the menu in order to set your balance according to your individual requirements.
Menu structure:
Menu designation Invoke menu Explanation
Main Menu
Calibration Menu
Zero / tare Menu
Data Output Menu
Unit setting Menu
Main menu
Adjustment
Set to zero/taring
Data output
Weighing Units

11.1 Menu symbol

After invoking the menu, the menu symbol [ ] will appear. The arrangement depends on the navigation in the menu.
Symbol arrangement Explanation
Arrangement filled inside Shows the current setting
Circular arch left/right Higher or lower menu level selectable
Circular arch top / bottom Further selectable menu settings
ABS-N_ABJ-NM-BA-e-1312 41

11.2 Na vigation in the m enu

After selecting the function in the display via the navigation
the present setting for the function.
Selecting a menu item to the left.
Back to weighing mode.
How to invoke a menu, see chpt. 11
• Menu structure
Select and pass through menu items to bottom ().
Select setting within one function
Select and pass through menu items to top ().
Select setting within one function
buttons ( ), the change is invoked by pressing the TARE key.
Confirm and store the setting momentarily appearing on the display by actuating the TARE key. The stability display shows
Selecting a menu item to the right ().
Leaving the function
Press shortly the ON/OFF key: Back to previous menu.
Long actuation of ON/OFF key:
42 ABS-N_ABJ-NM-BA -e-1312

11.3 Menu overview

+ See also the attached menu map
ABS-N_ABJ-NM-BA-e-1312 43

11.4 Resetting the menu

Select menu item
Resetting the menu
This will return all the settings to factory setting. The reference value stored in previous use of piece counting or percentage conversion will also be cleared. Factory settings are marked by a “*” in the menu oversight.
Call up menu
In weighing mode press the MENU button twice
Press the navigation buttons ( ) repeatedly until
TOOLS“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly
until „SYSTEM“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly
until „RESET“ is displayed.
Confirm with TARE. Confirm OK inquiry by pressing TARE, the password
inquiry will be shown.
Use the navigation keys () to enter the password
see chpt. 3.1.1 „Numeric input“.
+ Standard password (factory setting): „9999“.
+
Change password see chpt. 11.5.1
Confirm with TARE. The balance will be reset to factory
setting and automatically returns into the weighing mode.
44 ABS-N_ABJ-NM-BA -e-1312

11.5 Menu Lock

Either
The menu setting operations can be locked so that the settings cannot be inadvertently changed. This menu lock is set with the following procedure.
Switch-on power supply of the balance and wait until
OFF“ is displayed
or
Set balance into stand-by mode, see chpt. 10.2
Select menu item
1. Press MENU for 3 sec., the password inquiry appears.
2. Use the navigation keys ( ) to enter the password
see chpt. 3.1.1 „Numeric input“.
+
Standard password (factory setting): „9999“.
+ Change password, see following chapter + If a wrong password is entered, „ERR N“
will appear. Restart with step 1.
3. Confirm with TARE. The menu lock is enabled, the
symbol is activated. This is followed by a reappearance of "oFF" or .
or
In weighing mode the symbol shows the menu lock.
If a menu item selection is attempted in locked status,
the message “LOCKED” appears and the menu
selection is not allowed. To deactivate menu disabled, follow the sequence of operations below:
Remove the menu lock
With indication “oFF” or „ “, repeat step 1 – 3
ABS-N_ABJ-NM-BA-e-1312 45

11.5.1 Change password

Select menu item Change password
Standard password (factory setting): „9999“
Call up menu
In weighing mode press the MENU button twice
Press the navigation buttons ( ) repeatedly until
TOOLS“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly
until „SYSTEM“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly
until „PASS.WRD“ is displayed.
Confirm with TARE. Use the navigation keys ( ) to enter the currently set
password see chpt. 3.1.1 „Numeric input“.
Confirm with TARE. When „OK“ appears, the password
input has been correct, if „ERR N“ appears, it was
wrong. In this case repeat the input with the correct password.
Use the navigation keys () to enter the new
password
„Numeric input“ see chpt. 3.1.1.
Confirm with TARE.
Confirm again with TARE (or discard by ON/OFF).
Return to weighing mode
Press ON/OFF repeatedly or for 3 sec.
46 ABS-N_ABJ-NM-BA -e-1312

11.6 Log menu settings

Select menu item
Activate menu item
When an optional printer is connected, a list of the current menu settings can be printed out.
+ Data output, see chpt. 16
Call up menu
In weighing mode press the MENU button twice
Press the navigation buttons ( ) repeatedly until
TOOLS“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly
until „SYSTEM“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly
until „COND.OUT“ is displayed.
Confirm with TARE. Confirm OK inquiry using TARE.
Confirm with TARE. The printout is started, the symbol
is displayed.
The balance returns automatically into weighing mode.
ABS-N_ABJ-NM-BA-e-1312 47

12 Zero / tare menu (ze roing and taring functions)

In the event that small quantities are removed or added
is advisable to switch off this function.
This function is used to correct automatically small
Selectable functions:
1. Zero tracking function
+ see chpt 12.1
2. Auto zero function
+ see chpt 12.2
3. Auto tare function
+
see chpt 12.3
This function is used to correct automatically small weight variations which appear directly after switching-on.
to the material to be weighed, incorrect weighing results can be displayed due to the “stability compensation“ in the balance. (e.g. slow flow of liquids from a container placed on the balance, evaporating processes). When apportioning involves small variations of weight, it
weight variations which appear after a measurement (e.g. soiled weighing plate) after the stability display.
After data output an automatic taring is carried out
4. Zero / tare timing change function
+
see chpt 12.4
Selectable if the balance tares or gets set to zero before or after indication of the stability mark.
48 ABS-N_ABJ-NM-BA -e-1312

12.1 Zero tracking function

The zero tracking function „A.ZERO“ is turned-on by factory.
1. Check menu setting
Zero tracking symbol Zero tracking function
is displayed
is not displayed
2. Call function
In weighing mode press TARE for 3 sec. and if
necessary press repeatedly the navigation buttons ( ). until „Z.TRC“ is displayed. The current setting is marked by the stability display ( ).
With stability display ( ) Function turned on
Without stability display ( ) Function switched off
3. Enable/dis able function
Press TARE
switched on
switched off
4. Return to weighing mode
Press ON/OFF repeatedly or for 3 sec.
ABS-N_ABJ-NM-BA-e-1312 49

12.2 Auto zero function

The auto zero function is not available when formula mode is enabled (see chapter 14.3).
Stability display ( )
yes
no
When entering the zero range observe the currently set weighing unit.
Example for a balance d = 0.0001 g
Unit
Lower limit
Upper limit
g
0.0001 g
0.0099 g
ct
0.001 ct
0.099 ct
1. Call up menu
Press TARE for 3 sec. in weighing mode.
2. Select function
Press the navigation buttons () repeatedly until
A.ZERO“ is displayed. The current setting is marked by
the stability display ( ).
Further steps depend on the desired setting:
Function set/update cancel
3. Define zero range
Press TARE
ON
off
 
Press PRINT and
continue with step 3.
Press TARE and
continue with step 3.
Press TARE and
continue with step 4.
continue with step 4.
Use the navigation keys () to enter the zero range
see chpt. 3.1.1 „Numeric input“.
4. Return to weighing mode
Press ON/OFF repeatedly or for 3 sec.
If the weighing unit is changed at a later moment, the zero range must be adapted to the new weighing unit (step 3.).
Upper limit zero range: 99 d (in the displayed weighing unit) Lower limit zero range: 1d (in the displayed weighing unit)
50 ABS-N_ABJ-NM-BA -e-1312

12.3 Auto Tare function

1. Call up menu
Press TARE for 3 sec. in weighing mode.
2. Select function
Press the navigation buttons () repeatedly
until „A.TARE“ is displayed. The current setting is marked
by the stability display ( ).
With stability display ( ) Function turned on
Without stability display ( ) Function switched off
3. Enable/dis able function
Press TARE
4. Return to weighing mode
Press ON/OFF repeatedly or for 3 sec.
ABS-N_ABJ-NM-BA-e-1312 51

12.4 Zero / tare timing change function

The zero / tare timing change function can be switched-on, when the „Auto zero“ and the „Auto tare“ function are enabled.
1. Call up menu
Press TARE for 3 sec. in weighing mode.
2. Select function
Press the navigation buttons () repeatedly
until „TARE.F“ is displayed.
Confirm with PRINT. The current setting is marked by the
stability display ( ).
With stability display ( ) Function turned on
Without stability display ( ) Function switched off
3. Enable/dis able function
Press TARE
4. Return to weighing mode
Press ON/OFF repeatedly or for 3 sec.
52 ABS-N_ABJ-NM-BA -e-1312

13 Settings for Stability and Response

The easy setting display flashes for a short time. Inputs are only possible
switched off.
Exists the possibility to tune the stability of the display and the degree of reaction of the balance to the requirements of certain applications or the environmental conditions. Most measurements can be carried out using default settings, that is, standard mode. In standard weighing mode, stability and reaction have the same priority. For certain applications such as e.g. dosage do use the dosing mode (pouring mode). In dosing mode the reaction degree has the higher priority. Beside the selection standard / dosing mode the stability of the display and the reaction degree of the balance can additionally adapted in the menu. Please note that data processing for greater stability generally slows the response and processing for higher response reduces stability.
13.1 Stability and reaction settings via „Easy Setting display“
(without invoking menu)
In weighing mode press MENU shortly. The easy setting
display [ ] flashes.
When the display is flashing, set stability or reaction via
UNIT or PRINT as described in the following.
Priority on reaction
Easy setting display
Operation
Every time the button is hit, reaction will increase
Priority on stability
Every time the button is hit, stability will increase
within this time. Using ON/OFF the flashing Easy setting display can be
ABS-N_ABJ-NM-BA-e-1312 53

13.2 S election weighing / dosi ng m ode

Select menu item
Select menu item
Call-up weighing mode:
This is the default setting. Use this mode, when neither stability nor reaction time must be increased or reduced.
Call up menu
In weighing mode press the MENU button twice
Press the navigation buttons ( ) repeatedly until
STAND“ is displayed.
Confirm with TARE. The balance returns automatically
into weighing mode.
Call-up dosing mode:
Use this function if you wish to increase display speed, e.g. during apportioning. However, please note that the balance is very susceptible to ambience conditions.
Call up menu
In weighing mode press the MENU button twice
Press the navigation buttons ( ) repeatedly
until „POURING“ is displayed.
Confirm with TARE. From here on the balance is in
dosing mode (pouring mode), symbolised by the indicator .
54 ABS-N_ABJ-NM-BA -e-1312

13.3 Standstill width

Select menu item
Set range
If the stability display lights up ( ), the weighing result will be stable within the range indicated by the standstill width.
Set range for stability determination:
Call up menu
In weighing mode press the MENU button twice
Press the navigation buttons ( ) repeatedly until
TOOLS“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly
until „PARAMW“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly
until „BAND“ is displayed.
Acknowledge using PRINT, the current setting is
displayed.
Use the navigation buttons ( ) to select the desired
setting (0.5d, 1d, 10d, 50d, 100d, 1000d can be selected).
0.5D
1000d
Stability display ( ) very quiet environment
Stability display ( ) busy environment
Confirm with TARE. The current setting is marked by the
stability display ( ).
Return to weighing mode
Press ON/OFF repeatedly or for 3 sec.
ABS-N_ABJ-NM-BA-e-1312 55
Set reaction time:
Select menu item
Call up menu
In weighing mode press the MENU button twice
Press the navigation buttons ( ) repeatedly until
TOOLS“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly
until „PARAMW“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly
until „STB.Mk“ is displayed.
Set reaction time
Confirm with PRINT. The current setting is marked by
the stability display ( ).
with ( )
without ( ) Default
Confirm with TARE. The current setting is marked by the
stability display ( ).
Return to weighing mode
Stability display appears faster, but with less exactness.
Press ON/OFF repeatedly or for 3 sec.
56 ABS-N_ABJ-NM-BA -e-1312

14 Application Functions

The application functions can be combined with the check weighing or

14.1 P arts counting

With parts counting you can either count parts into a container or remove parts from a container. To count a greater number of parts the average weight per part has to be determined with a small quantity (reference quantity). The larger the reference quantity, the higher the counting exactness. High reference must be selected for small parts or parts with considerably different sizes.
1. Enable function and set reference
Call up menu
In weighing mode press the MENU button twice
target function (see chpt. 14).
The balance starts in the mode, in which it has been switched off.
For switching over between application and weighing mode, press MENU
for 3 sec.
Select piece counting mode Press the navigation buttons () repeatedly until
APL.FUNC“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly
until „PCS“ is displayed.
Press TARE. The display shows the currently set
memory location.
Enter memory location for reference
The balance offers the possibility to store five different
piece weights at the same time.
Press the navigation buttons () repeatedly until the
desired memory location is displayed, then
press TARE.
Display, if no piece weight is stored
Display, if a piece weight is stored
ABS-N_ABJ-NM-BA-e-1312 57
Set reference
If under the indicated memory location no reference is stored, [- - -] will be displayed.
Place an empty container on the balance and tare it by
pressing TARE.
 Press MENU twice.  Press the navigation buttons ( ) repeatedly until the
desired reference quantity is displayed (possible reference quantities 5, 10, 20, 50, 100).
Use MENU to scroll forward. Use UNIT to scroll backward.
Fill number of pieces into the container according to the
selected reference quantity.
Wait for the stability display ( ), then confirm by
pressing TARE. The balance determines the average
piece weight. From here on the appliance is in piece counting mode
and counts all pieces which are on the weighing plate.
2. Switching over betw een piece counting and weighing mode
3. Count the items
Call-up the piece weight stored in piece counting mode
Every time UNIT is pressed (approx. for 3 sec.), the next
Place an empty container on the balance and tare it by
Fill weighing goods into the container and read the piece
memory location [ ] is called up.
pressing TARE.
quantity.
58 ABS-N_ABJ-NM-BA -e-1312
Change or add piece w eight
Or count more pieces
or
The menu item to change the stored reference is called up, the currently set reference quantity is displayed.
+ see „4. Change or add piece weight“
The stored piece weight is displayed in grams, marked by the
symbol . Using PRINT the reference weight (UW= Unit weight)
can be edited when a printer is connected.
e.g.: UW = 1.0001
To go back to piece counting display press UNIT anew.
Switching over between piece counting and weighing mode
4. Change or add piece weight
In piece counting mode call-up the desired memory location
Every time UNIT is pressed (approx. for 3 sec.), the next
memory location [ ] is called up.
Place an empty container on the balance and tare it by
pressing TARE.
Press MENU twice, the currently set reference quantity
is displayed.
Press the navigation buttons () repeatedly until the
desired reference quantity is displayed (possible reference quantities 5, 10, 20, 50, 100).
Use MENU to scroll forward. Use UNIT to scroll backward.
Fill number of pieces into the container according to the
selected reference quantity.
Wait for the stability display ( ), then confirm by
pressing TARE.
The balance is now in parts counting mode and counts
all units on the weighing plate.
ABS-N_ABJ-NM-BA-e-1312 59

14.2 Percent determination

Percent weighing allows to display weight in percent, in relation to a reference weight.
The balance offers two possibilities:
1. Reference = 100 %
2. Reference = defined by user
Call up menu In weighing mode press the MENU button twice
Select percent weighing mode
Press the navigation buttons () repeatedly until
APL.FUNC“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly
until „PERCENT“ is displayed.
Further steps:
+ Reference weight = 100 %, see chpt. 14.2.1 + Reference weight = XX %, see chpt. 14.2.2
60 ABS-N_ABJ-NM-BA -e-1312

14.2.1 Reference weight = 100 %

Set reference
(Minimum weight: Readability d x 100)
Select percent weighing mode, see chpt. 14.2
Press TARE.
If necessary, press the navigation buttons ( )
repeatedly until „SAMPLE“ is displayed.
 Press TARE.
Display, if no reference weight is stored
Display, if a reference weight is stored
If necessary, place an empty container on the balance
and tare it by pressing TARE.
Press MENU twice.
Put a reference weight which corresponds to 100 %.
Wait for the stability display ( ), then confirm by
pressing TARE.
From here on the weight of the sample is displayed in percentage in terms of the reference weight, see chpt.
13.2.3.
ABS-N_ABJ-NM-BA-e-1312 61

14.2.2 User-defined reference

Set reference
Select percent weighing mode, see chpt. 13.2
Press TARE.
If necessary, press the navigation buttons ( )
repeatedly until „OPTION“ is displayed.
 Press TARE.
Display, if no reference is stored
Display, if a reference is stored
If necessary, place an empty container on the balance
and tare it by pressing TARE.
Press MENU twice.
Press PRINT and current setting will be displayed.
Enter a percentage of your choice via the navigation
buttons, see chpt. 3.1.1 „Numeric input“
Put a reference weight which corresponds to the entered
percent value.
Wait for the stability display ( ), then confirm by
pressing TARE.
The display symbol [ ] characterises the percentage determination with user-defined reference value.
From here on the weight of the sample is displayed in percentage in terms of the reference weight, see chpt.
14.2.3.
62 ABS-N_ABJ-NM-BA -e-1312

14.2.3 Percent determination

In weighing mode press MENU for 3 sec., the currently
set percentage determination mode is displayed.
Place an empty container on the balance and tare it by
pressing TARE.
Place goods to be weighed on balance.
The weight of the sample is displayed in percentage in terms of the reference weight.
Either carry out more percent weighing
or
The menu item to change the stored reference appears.
+ see chpt. 14.2.2 / 14.2.3 „Set reference“
The stored reference weight is displayed in grams, marked by the
symbol . Use PRINT to edit the weight value of the reference
value when a printer is connected.
or
Back to percent display, press UNIT anew.
Switching over between percentage determination and weighing mode
ABS-N_ABJ-NM-BA-e-1312 63

14.3 Formula mode

The formula function allows to add on various components of a mixture. For check purpose, the weight of all components (CMP001, (CMP002 etc.), as well as the total weight (TOTAL) can be printed out. The balance works with a separated memory for the weight of the weighing container and of the formula components.
In formula mode the „Auto zero“ function is not active (see chpt. 12.2).
1. Connect printer (see chpt. 15 „Data output“)
2. Call up menu
In weighing mode press the MENU button twice
3. Select formula mode
Press the navigation buttons () repeatedly until
APL.FUNC“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly
until „FORMULA“ is displayed. The formula symbol [ ] appears.
Press TARE. The indicator is displayed, from
here on the balance is in formula mode.
Enable the edition „component no.“(see chpt. 14.3.1) and „Total“ (see chpt. 14.3.2).
4. Weighing components
If necessary, place an empty container on the balance
and tare it by pressing TARE.
Press PRINT , the symbol goes out.
When the GLP function (see chpt. 8.3) is activated, the headline will be printed.
Weigh first component. Press PRINT.
When standstill control ( ) is completed, the weighing value of the first component (CMP001) is issued to the optional printer. The displayed value is added into the total adding memory. This is followed by the automatic taring, the symbol [ ] is displayed.
64 ABS-N_ABJ-NM-BA -e-1312
Company
Weigh further components in the same way
During formulation weighing can be displayed at any time by pressing
MENU (for 3 sec.).
5. Complete formulation process
Press ON/OFF. When the symbol [G] is displayed, the
total weight (TOTAL) of all components is displayed and
edited to the printer.
When the symbol is displayed, the balance is ready
for further measurements.
6. Return to weighing mode
Press ON/OFF for 3 sec.
Printout example „GLP ON“ (KERN YKB-01N):
Model Serial no. ID no.
prepared by
ABS-N_ABJ-NM-BA-e-1312 65

14.3.1 Enable edition „component no.“

Function ELM.NUM enabled
Function ELM.NUM disabled
FORMULATION MODE
FORMULATION MODE
CMP001=
0.5361 g
0.5361 g
CMP002=
0.5422 g
0.5422 g
CMP003=
0.4488 g
0.4488 g
TOTAL=
1.5271 g
TOTAL=
1.5271 g
Select formula mode, see chpt. 14.3 When the symbol appears, press the MENU
button twice.
Press the navigation buttons () repeatedly until
ELM.NUM“ is displayed.
Confirm with PRINT. The current setting is marked by
the stability display ( ).
with ( )
Edition „component no.“ (e.g. CMP001)
without ( ) No edition „component no.“
Change using TARE.
Back to formula mode
Press ON/OFF repeatedly or for 3 sec.
Printout examples (KERN YKB-01N):
66 ABS-N_ABJ-NM-BA -e-1312

14.3.2 Enable issue „TOTAL“

Function TOTAL enabled
Function TOTAL disabled
FORMULATION MODE
FORMULATION MODE
CMP001=
0.5361 g
CMP001=
0.5361 g
CMP002=
0.5422 g
CMP002=
0.5422 g
CMP003=
0.4488 g
CMP003=
0.4488 g
TOTAL
1.5271 g
Select formula mode, see chpt. 14.3 When the symbol appears, press the MENU
button twice.
Press the navigation buttons () repeatedly
until „TOTAL“ is displayed.
Confirm with PRINT. The current setting is marked by
the stability display ( ).
with ( ) Issue „TOTAL“
without ( ) No issue „TOTAL“
Change using TARE.
Back to formula mode
Press ON/OFF repeatedly or for 3 sec.
Printout examples (KERN YKB-01N):
ABS-N_ABJ-NM-BA-e-1312 67

15 Check weighi ng a nd target m ode ( che c k a nd target weighing)

The check weighing or target mode can be applied for the application

15.1 Check weighing

In many cases not the nominal value of the weighed goods is the decisive parameter, but the deviation from this nominal value. Such applications are for example the weight check of equivalent packages or the process check of parts in a fabrication process.
functions (see chpt. 14).
The balance starts in the mode, in which it has been switched off.
The indicators
, or
in the display show, where the weighed goods are
located within the tolerance range. These indicators are only functioning in check weighing or in target mode, otherwise they cannot be seen.
The indicators provide the following information
Condition Grading Indicator
OVR.RNG < sample weight
HI.LIM < sample weight OVR.RNG
LO.LIM sample weight HI.LM
UND.RG sample weight < LO.LIM
Sample weight < UND.RG
Beyond tolerance limit
Upper tolerance limit
Inside tolerance range
Lower tolerance limit
Beyond tolerance limit
No
indicator
No
indicator
68 ABS-N_ABJ-NM-BA -e-1312
Stability display ( )
yes
no
0. Call up menu
In weighing mode press the MENU button twice
1. Select function
Press the navigation buttons () repeatedly until
TOOLS“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly
until „CHECK.W“ is displayed. The current setting is marked by the stability display ( ).
Further steps depend on the desired setting:
Function set/update cancel
Press PRINT and
continue with step 3.
Press TARE and
continue with step 3.
2. Set limit values
ON
off
 
Press TARE and
continue with step 4.
continue with step 4.
When entering the limit values ensure that the values match logically one with another, i.e. the lower limit value must not be greater than the upper one.
If not considered, the balance will adapt the limit values automatically.
Press the navigation buttons () repeatedly until
HI.LIM“ is displayed.
Press TARE. The current setting will be displayed. Use the navigation buttons ( ) to enter the desired
value, see chpt. 3.1.1 „Numeric input“.
ABS-N_ABJ-NM-BA-e-1312 69
Confirm with TARE.
Press the navigation buttons ( ) repeatedly
until „LO.LIM“ is displayed.
Press TARE. The current setting will be displayed. Use the navigation buttons ( ) to enter the desired
value, see chpt. 3.1.1 „Numeric input“.
Confirm with TARE.
Press the navigation buttons ( ) repeatedly
until „UND.RNG“ is displayed.
Press TARE. The current setting will be displayed. Use the navigation buttons ( ) to enter the desired
value, see chpt. 3.1.1 „Numeric input“.
Confirm with TARE.
Press the navigation buttons ( ) repeatedly
until „OVR.RNG“ is displayed.
Press TARE. The current setting will be displayed. Use the navigation buttons ( ) to enter the desired
value, see chpt. 3.1.1 „Numeric input“.
Confirm with TARE.
70 ABS-N_ABJ-NM-BA -e-1312
3. Return to weighing mode
Press ON/OFF repeatedly or for 3 sec.
4. Start tolerance control
If necessary, place an empty container on the balance and tare it by pressing
TARE.
Place weighed goods and wait until the indicator
, or
appears. With the help of the indicator check if the weighed goods are under, inside or over the default tolerance
Example of entry:
HI.LIM 7.0000 g LO.LIM 6.0000 g UND.RNG 5.0000 g OVR.RNG 8.0000 g
Sample weight < UND.RG
(sample weight < 5 .0000g)
No indicator is
displayed
UND.RG sample weight < LO.LIM
(sample weight 5.0000g - 5.9999g)
LO.LIM sample weight HI.LM
(sample weight 6.0000g - 7.000g)
HI.LIM < sample weight OVR.RNG
(sample weight 7.0001g - 8.0000g)
Sample weight > OVR.RNG
(sample weight > 8.0000 g)
is displayed
is displayed
is displayed
No indicator is
displayed
ABS-N_ABJ-NM-BA-e-1312 71

15.2 Target mode

This mode e.g. is used for weighing constant liquid quantities or for assessment of missing quantities or excess quantities.
The target value is the numeric value which corresponds to the nominal quantity of the used unit. Beside the target value a tolerance value is entered. This is a numerical value which is plus/minus over or under the acceptable target value.
The indicators
, or
advise, when the target value is reached. These indicators are only functioning in check weighing or in target mode, otherwise they cannot be seen.
The indicators provide the following information:
Condition Grading Indicator
Weight more than the nominal
Great difference to target value
flashes slowly
weight and above the upper tolerance
Weight within tolerance (target value ± tolerance)
Weight less than nominal
Small difference to target value
flashes fast
Target value accepted
Small difference to target value
flashes fast
weight and below the lower tolerance
Great difference to target value
flashes fast
72 ABS-N_ABJ-NM-BA -e-1312
Stability display ( )
yes
no
1. Call up menu In weighing mode press the MENU button twice
2. Select function
Press the navigation buttons () repeatedly until
TOOLS“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly
until „TARGT“ is displayed. The current setting is marked
by the stability display ( ).
Further steps depend on the desired setting:
Function set/update cancel
ON
off
3. Set target value and tolerance
Press the navigation buttons () repeatedly until the
display to enter the target value „TG.VAL“ appears.
Press TARE. The current setting will be displayed.
 
Press PRINT and
continue with step 3.
Press TARE and
continue with step 3.
Press TARE and
continue with step 4.
continue with step 4.
Use the navigation buttons () to enter the desired
value, see chpt. 3.1.1 „Numeric input“.
Confirm with TARE.
Press the navigation buttons ( ) repeatedly until the
display to enter the target value „LM.VAL“ appears.
Press TARE. The current setting will be displayed.
Use the navigation buttons ( ) to enter the desired
value, see chpt. 3.1.1 „Numeric input“.
ABS-N_ABJ-NM-BA-e-1312 73
Confirm with TARE.
4. Return to weighing mode
Press ON/OFF repeatedly or for 3 sec.
5. Start tolerance control
If necessary, place an empty container on the balance and tare it by pressing
TARE.
Place weighed goods and wait until the indicator
, or
appears. With the help of the indicator check if the weighed goods are under, inside or over the default tolerance
Example of entry: TG.VAL LM.VAL
100.0000 g
10.0000 g
Weight less than nominal
flashes slowly
weight and below the lower tolerance
Weight within tolerance [target value ± tolerance]
(90.0000g – 110.000g)
Weight more than the nominal weight and above the upper tolerance
flashes fast
flashes fast
flashes slowly
74 ABS-N_ABJ-NM-BA -e-1312

16 Data output

The appliance interface allows a bi-directional data exchange from the balance to external devices. This data exchange is asynchronous using ASCII - Code. The following conditions must be met to provide successful communication between the weighing balance and the printer.
Use a suitable cable to connect the weighing balance to the interface of the printer/PC.
Faultless operation requires an adequate KERN interface cable (optional).
Communication parameters (baud rate, bits and parity) of balance and printer must match.

16.1 Pin connection

Due to connection of an optional KERN interface cable the balance is equipped with
a RS232C interface.
Balance (RS-232C)
3 TXD
2 RXD
6 DSR
5 SG
4 DTR
7 CTS
8 RTS
ABS-N_ABJ-NM-BA-e-1312 75

16.2 Issue functions

stable / positive
stable / negative
Stabilisation / zero display
Check weighing
1. Invoke Data Output menu
2. Select function
by the stability display ( ).

16.2.1 Automatic data output / Auto Print function

Data output ensues automatically without pressing PRINT, as soon as the respective
output condition is fulfilled. This is defined by the menu setting.
Tab. 1:
Mode 1
LD. .
Mode 2
LD.UL.
Mode 3
LD. .Z
Mode 4
LD.UL.Z
Mode 5
LD.OK .
- - -
-
- - -
Enable Auto Print function:
- -
-
-
Output for stable and positive weighing value
Output for stable and positive or negative weighing value.
Output for stable and positive weighing value. New output only after zero display and stabilisation
Output for stable and positive or negative weighing value. New output only after zero display and stabilisation.
If the Auto Print function is connected to the check weighing function, data of stable weighing values are output with indicator display
.
.
Press PRINT for 3 sec. in weighing mode.
Press the navigation buttons ( ) repeatedly
until „APL.PRN“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly until
AUTO.PRN“ is displayed. The current setting is marked
76 ABS-N_ABJ-NM-BA -e-1312
Stability display ( )
Press TARE and
5.
no
4. If necessary, set condition for zero display
( )
An other output when the display goes back to 50% of the previous weighing value.
5. Return to weighing mode
displayed.
weighing value is issued automatically.
automatically.
Further steps depend on the desired setting:
Function set/update cancel
yes
3. Set output condition
Use the navigation buttons () to select the desired
mode, e.g. Mode 3 (details see Tab. 1)
Confirm with TARE
Press PRINT Select the desired adjustment using the navigation keys
RET.0 RET.50%
ON
off
An other output when the display goes back to zero.
 
Press PRINT and
continue with step 3.
Press TARE and
continue with step 3.
continue with step
continue with step
5.
With selection „RET.0“
Confirm with TARE
Press ON/OFF repeatedly or for 3 s. From here on the
Auto Print function is active, the indicator is
5. Place goods to be weighed on balance
If necessary, place an empty container on the balance and tare it by pressing
TARE.
Place weighed goods and wait until the stability display ( ) appears. The
6. Remove the weighed good.
Wait until the stability ( ) / zero display appears. The weighing value is issued
ABS-N_ABJ-NM-BA-e-1312 77

16.2.2 Continuous Output function

1. Invoke Data Output menu
2. Select function
the stability display ( ).
3. Set start/end of the c onti nuous issue manually
by the stability display ( ).
Press PRINT for 3 sec. in weighing mode.
Press the navigation buttons ( ) repeatedly
until „APL.PRN“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly until
SEQ.PRN“ is displayed. The current setting is marked by
with ( ) on
without ( ) off
Change using TARE.
or automatically
Press PRINT Press the navigation buttons ( ) repeatedly
until „MANU“ is displayed. The current setting is marked
The continuous data output starts after
with ( )
pressing of PRINT and ends by pressing ON-OFF
without ( )
The continuous data output starts automatically
Change using TARE
78 ABS-N_ABJ-NM-BA -e-1312
5. Return to weighing mode
Press ON/OFF repeatedly or for 3 s.
the indicator will appear.
With deactivated menu setting „MANU“ (see step 3,
automatically, step 7 is not applied.
6. Tare weighing container
weighing value is issued automatically.
7. Press PRINT (only with menu setting „MANU ON)
The continuous data output starts, the indicator goes out.
8. Place goods to be weighed on balance (data issue interval approx. 100 msec).
4. Connect filter
Press the navigation buttons () repeatedly
until „NO.FIL“ is displayed. The current setting is marked
by the stability display ( ).
with ( ) Weighing value is filtered
without ( ) Weighing value is not filtered
Change using TARE
With activated menu setting „MANU“ (see step 3, with )
without ) starts the continuous data output
If necessary, place an empty container on the balance and tare it by pressing
TARE.
Place weighed goods and wait until the stability display ( ) appears. The
Every display change is continuously issued,
Interrupt continuous data output ON-OFF = interrupt PRINT = restart
ABS-N_ABJ-NM-BA-e-1312 79

16.2.3 Output Timing Change function

1. Invoke Data Output menu
2. Select function
stability display ( ).
3. Return to weighing mode
Press ON/OFF repeatedly or for 3 s.
This function allows to select if the data output shall ensue with stable or instable
weighing value after pressing PRINT.
Press PRINT for 3 sec. in weighing mode.
Press the navigation buttons ( ) repeatedly
until „APL.PRN“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly until
PRINT.F“ is displayed.
Confirm with PRINT. The current setting is marked by the
with ( )
without ( ) Data output only after stability display
Change using TARE.
Instantaneous data output without waiting until the stability display appears
80 ABS-N_ABJ-NM-BA -e-1312

16.2.4 GLP Output function / balance identification number

KERN & Sohn GmbH
Company
Balance identification no. (see chpt. 8.4)
-SIGNATURE-
prepared by
- - - - - - - - - - - - - - - - -
With the GLP Output function the printouts of weighing results are completed with a bottom row and a head line. Contents of the headline and bottom row see following printout example.
+ Enable GLP Output function, see chpt. 8.3 + Enter balance identification number, see chpt. 8.4
Printout example:
TYPE ABJ 220-4NM
SN WBIIAB000I
Model
Serial no.
ID 1234
200.0000g
Weighing result
In the ABS/ABJ_N series it is not possible to state date and time.
ABS-N_ABJ-NM-BA-e-1312 81

16.3 Communication parameters

Standard
Standard
Standard
Standard
Standard
Setting for
Menu selection
Manufactu rer
Shimadzu (Standard)
Shimadzu *
user­defined
user­defined
user­defined
Hand­shake
user­defined
Data format
Shimadzu Standard
Shimadzu Standard
Mettler Standard
Sartorius Standard
A-D Standard
user­defined
user­defined
By calling-up a standard setting „MODE 1 - MODE 5“ all communication parameters
are already preset (see chpt. 16.3.1). The subsequent standard setting must be selected according to the printer (details see the following table).
In the menu item „MODE U“ all parameters can be set defined by user (see chpt.
16.3.2).
User-defined settings
setting 1
MODE 1 MODE 2 MODE 3 MODE 4 MODE 5 MODE U MODE U
setting 2
setting 3
Mettler Sartorius A-D - -
setting 4
setting 5
KERN –
YKB-01N
Baud Rate 1200 1200 2400 1200 2400
Parity None (8) None (8) Even (7) Odd (7) Even (7)
Stop bit 1 1 2 2 2
Hardware Hardware off Hardware off
Separator C/R C/R C/R + L/F C/R + L/F C/R + L/F
1200
None (8)
1
off
DF.1
C/R
*only if the balance can send a feedback to the PC (without error: OK [C/R], at error NG [C/R].
82 ABS-N_ABJ-NM-BA -e-1312

16.3.1 Selection of a standard setting „MODE 1 - MODE 5“

1. Invoke Data Output menu
2. Select function
current setting is marked by the stability display ( ).
3. Return to weighing mode
Press ON/OFF repeatedly or for 3 s.
Press PRINT for 3 sec. in weighing mode.
Press the navigation buttons ( ) repeatedly
until „COMM.SET“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly until the
desired mode „MODE 1 - MODE 5“ is displayed. The
with ( ) on
without ( ) off
Change using TARE.
ABS-N_ABJ-NM-BA-e-1312 83
16.3.2 User-defined settings „MO DE U“ /
1. Invoke Data Output menu
2. Select function
current setting is marked by the stability display ( ).
setting for KERN –YKB-01N
In the menu item „MODE U“ every communication parameter can be set individually.
Press PRINT for 3 sec. in weighing mode.
Press the navigation buttons ( ) repeatedly
until „COMM.SET“ is displayed.
Confirm with PRINT Press the navigation buttons ( ) repeatedly until the
desired mode „MODE 1 - MODE 5“ is displayed. The
with ( ) on
without ( ) off
Change using TARE.
84 ABS-N_ABJ-NM-BA -e-1312
stability display ( ).
Press the navigation buttons ( ) repeatedly until the
B0.300
B0.9600
300bps
9600bps
stability display ( ).
Press the navigation buttons ( ) repeatedly until the
3. Set the communication speed (Baud rate)
 Press PRINT  Press the navigation buttons ( ) repeatedly
until „START“ is displayed.
Press PRINT. The current setting is marked by the
desired setting (e.g. 9600 bps) appears.
Press TARE.
Selectable settings:
Display
Baud rate
4. Set parity
Use ON/OFF to return to menu
B0.600 B0.1200 B0.2400 B0.4800
600bps 1200bps 2400bps 4800bps
B.19.2k B.38.4k
19.2k bps 38.4k bps
Press the navigation buttons () repeatedly
until „PARITY“ is displayed.
Press PRINT. The current setting is marked by the
desired setting (e.g. P.NONE) appears.
Press TARE.
Selectable settings:
Display P.NONE P.ODD P.EVEN
Parity Small parity, 8 bit Odd parity, 7 bit Straight parity, 7 bit
ABS-N_ABJ-NM-BA-e-1312 85
stability display ( ).
Press the navigation buttons ( ) repeatedly until the
stability display ( ).
Press the navigation buttons ( ) repeatedly until the
Hand­shake
Hardware
handshake
5. Stop bit settings
 Use ON/OFF to return to menu  Press the navigation buttons ( ) repeatedly
until „STOP“ is displayed.
Press PRINT. The current setting is marked by the
desired setting (e.g. S. 1) appears.
Press TARE.
Selectable settings:
Display S. 1 S. 2
Stop bit 1 bit 2 bit
6. Set handshake
 Use ON/OFF to return to menu  Press the navigation buttons ( ) repeatedly
until „HAND.SHK“ is displayed.
Press PRINT. The current setting is marked by the
desired setting (e.g. HS.HW) appears.
Press TARE.
Selectable settings:
Display HS.OFF HS.HW HS.SW HS.TiM
No handshake
Software handshake Timer handshake
86 ABS-N_ABJ-NM-BA -e-1312
stability display ( ).
Press the navigation buttons ( ) repeatedly until the
Options:
Data length 8 -2
stability display ( ).
Press the navigation buttons ( ) repeatedly until the
Press ON/OFF repeatedly or for 3 s.
7. Set data format
 Use ON/OFF to return to menu  Press the navigation buttons ( ) repeatedly
until „D.FORM“ is displayed.
Press PRINT. The current setting is marked by the
desired setting (e.g. DF.1) appears.
Press TARE.
Selectable settings (details see chpt. 15.4):
Display DF.1 DF.2 DF.3 DF.4 DF.FREE
Data format
Shimadzu
Standard
Shimadzu
Standard
Mettler
Standard
Sartorius
Standard
Head byte 1 -17,
8. Set separator
 Use ON/OFF to return to menu  Press the navigation buttons ( ) repeatedly
until „DELIM“ is displayed.
Press PRINT. The current setting is marked by the
desired setting (e.g. DF.1) appears.
Press TARE.
Selectable settings:
Display CR LF CR+LF COMMA WINI
Separator CR LF CR+LF COMMA not documented
9. Return to weighing mode
ABS-N_ABJ-NM-BA-e-1312 87

16.4 Data Formats

Position
data
ASCII code
Explanation
Negative weighing value = minus sign 2DH
2 20H
3 20H 4 9
39H 5 .
2EH 6 9
39H 7 9
39H 8 4
34H 9 9
39H
10
g
67H
11
20H
At CR+LF data length will be increased
Printout examples KERN YKB-01N
In the menu item „D.FORM“ four data formats „DF.1-DF.4“ are available.
+ Menu setting, see chpt. 16.3.2, step 7 „Setting data format“.
1. Data format 1 „DF.1“
Example 9.9949 g:
1 20H
12 C/R 0DH
Positive weighing value = space character 20H
Numeric weighing value is displayed in 8 positions.
Not required positions = space character 20H
Where necessary, Overload is displayed in 2 positions with O L
Weighing unit
Separator C/R = 0DH, L/F = 0AH
9.9949g Stable or instable/positive weighing value
-9.9949g Stable or instable/negative weighing value
88 ABS-N_ABJ-NM-BA -e-1312
2. Data format 2 „DF.2“
Position
Data
ASCII code
Explanation
1 20H
2 20H 3 U
55H 4
20H 5
20H 6
20H 7
20H
8 20H 9 9
39H
10
.
2EH
11
9
39H
12
9
39H
13
4
34H
14
9
39H
15
20H
16
g
67H
At CR+LF data length will be increased
Printout examples KERN YKB-01N
Example 9.9949 g:
Positive weighing value = space character 20H
Negative weighing value = minus sign 2DH
Stable weighing value = S (stable) 53H
Unstable weighing value = U (unstable) 55H
Numeric weighing value is displayed in 8 positions.
Not required positions = space character 20H
Where necessary, Overload is displayed in 2 positions with O L
17 C/R 0DH
S 9.9949 g Stable/positive weighing value
U 9.9949 g Instable/positive weighing value
S -9.9949 g Stable/negative weighing value
U -9.9949 g Instable/negative weighing value
Weighing unit
Separator C/R = 0DH, L/F = 0AH
ABS-N_ABJ-NM-BA-e-1312 89
3. Data format 3 „DF.3“
Position
Data
ASCII code
Explanation
1 + 2BH
2 20H 3
20H
4 20H 5 9
39H 6 .
2EH 7 9
39H 8 9
39H 9 4
34H
10
9
39H
11
20H
12
g
67H
13
20H
14
20H
At CR+LF data length will be increased
Printout examples KERN YKB-01N
Example 9.9949 g:
Positive weighing value = plus sign 2BH
Negative weighing value = minus sign 2DH
Numeric weighing value is displayed in 8 positions.
Not required positions = space character 20H
Where necessary, Overload is displayed in 2 positions with O L
15 C/R 0DH
+9.9949 g Stable or instable/positive weighing value
-9.9949 g Stable or instable/negative weighing value
Weighing unit
Separator C/R = 0DH, L/F = 0AH
90 ABS-N_ABJ-NM-BA -e-1312
4. Data format 4 „DF.4“
Position
Data
ASCII code
Explanation
1 S 53H
2 20H 3
20H 4
2BH 5
20H
6 + 2BH 7 9
39H 8 .
2EH 9 9
39H
10
9
39H
11
4
34H
12
9
39H
13
20H
14
20H
15
g
67H
At CR+LF data length will be increased
Printout examples KERN YKB-01N
Example 9.9949 g:
Stable weighing value = S (stable) 53H
Unstable weighing value = U (unstable) 55H
Positive weighing value = plus sign 2BH
Negative weighing value = minus sign 2DH
Numeric weighing value is displayed in 8 positions.
Not required positions = space character 20H
Where necessary, Overload is displayed in 2 positions with O L
16 C/R 0DH
S +9.9949 g Stable/positive weighing value
U +9.9949 g Instable/positive weighing value
S -9.9949 g Stable/negative weighing value
U -9.9949 g Instable/negative weighing value
Weighing unit
Separator C/R = 0DH, L/F = 0AH
ABS-N_ABJ-NM-BA-e-1312 91

16.5 Remote control instructions

S: stable
S: stable
Command Function
D01 Continuous data output
D02 Continuous data output of stable weighing values
D03 Status of stability display is attached to the data in the
continuous output.
U: instable
D05 Single output
D06 Automatic output
D07 Single output. Status of stability display is attached to the
data in the output.
U instable
D08 Single output with stable weighing value
D09 Cancel output
BREAK
Q
CAL
TARE
T
PRINT
Function as , see chpt. 3.1
Function as , see chpt. 3.1
Function as , see chpt. 3.1
Function as , see chpt. 3.1
92 ABS-N_ABJ-NM-BA -e-1312

17 Service, maintenance, disposal

Please do not use aggressive cleaning agents (solvents or

17.1 Cleaning Before cleaning, disconnec t the appliance from the operating voltage.

1. Display
2. Weighing
plate
3. casing,
housing, case, box
4. Glass doors
Fig. 1: Clean the balance
similar agents), but a cloth dampened with mild soap suds.
Remove weighing plate, clean it wet and dry it before installation
Please do not use aggressive cleaning agents (solvents or similar agents), but a cloth dampened with mild soap suds. Take care that the device is not penetrated by fluids and polish it with a dry soft cloth.
Loose residue sample/powder can be removed carefully with a brush or manual vacuum cleaner.
Spilled weighing goods m ust be removed immediately.
These can be removed as described in the following. After that clean with a commercially available glass cleaner.
Handle glass doors with care.
Attention: Risk of breakage
Risk of cuts.
Keep away your hands/fingers from the running rail.
ABS-N_ABJ-NM-BA-e-1312 93
Remove, screening ring, weighing
2. Remove the plastic handle by turning.
Do no touch the support of the
glass door carefully acc. to
4. Re-install the glass door in reverse order.
1.
plate and carrier of weighing plate
weighing plate. This could cause damage to the balance.
3. Remove
fig.
To secure the glass door always reattach the plastic handle.
Fig. 2: Remove the glass doors
94 ABS-N_ABJ-NM-BA -e-1312

17.2 Service, maintenance

The appliance may only be opened by trained service technicians who are authorized by KERN. Before opening, disconnect from power supply.

17.3 Disposal

Disposal of packaging and appliance must be carried out by operator according to valid national or regional law of the location where the appliance is used.

18 Instant help

Possible causes of errors:
In case of an error in the program process, briefly turn off the balance and disconnect from power supply. The weighing process must then be restarted from the beginning.
Fault Possible cause
The displayed weight does not appear
The displayed weight is permanently changing
The weighing result is obviously incorrect
The balance is not switched on.
The mains supply connection has been interrupted
(mains cable not plugged in/faulty).
Power supply interrupted.
Draught/air movement
Glass doors not closed
Table/floor vibrations
Weighing plate has contact with other objects.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if possible)
The display of the balance is not at zero
Adjustment is no longer correct.
The balance is on an uneven surface.
Great fluctuations in temperature.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if possible)
The desired weighing unit
cannot be called by UNIT
key.
Automatic adjustment carried out frequently.
No data transfer between printer and balance.
The menu settings cannot be changed.
ABS-N_ABJ-NM-BA-e-1312 95
Unit was not activated beforehand.
Severe temperature variations in the room or the
instrument
Communication settings are wrong.
The menu is locked Remove the menu lock.

18.1 Error messages

Error message
Explication
Remedy
ERR H Hardware error
High zero point shift during adjustment
ERR C
CAL D Display instable
Objects present on weighing plate.
Missing weighing plate
Switch balance off and on again. If the error message remains inform manufacturer.
Use ON/OFF to return into
weighing mode. Restart adjustment process.
Check ambient conditions (draught, vibrations etc.) Use
ON/OFF to return into weighing
mode. Restart adjustment process.
ERR N
ERR W Wrong application
COM ERR Wrong remote control order.
- OL Missing weighing plate Install weighing plate correctly
OL Overload Reduce load ABORT Process cancelled WAIT Wait for process
BUSY
Error at numerical input (e. g. wrong password)
When adjustment was started objects were present on weighing plate
Correct input
The balance changes to the previous state. Correct application.
The balance changes to the previous state. Correct remote control order.
Remove object and continue adjustment
Should other error messages occur, switch balance off and then on again. If the error message remains inform manufacturer.
96 ABS-N_ABJ-NM-BA -e-1312
Loading...