Kern PLJ 360-3NM, PLJ 600-3NM, PLJ 3500-2NM, PLJ 4500-2NM, PLJ 750-3N User guide

KERN & Sohn GmbH
Ziegelei 1 D-72336 Balingen email: info@kern-sohn.com
KERN PLJ_N
Version 3.0 12/2009 GB
Phone: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com
PLJ_N-BA-e-0930
GB
KERN PLJ_N
Version 3.0 12/2009
Operating Manual
Table of Contents
1 Technical Data ................................................................................................... 4
2 Declaration of -Conformity ............................................................................... 7
3 Basic Information (General) ............................................................................. 9
3.1 Proper use................................................................................................................................. 9
3.2 Improper Use ............................................................................................................................ 9
3.3 Warranty .................................................................................................................................... 9
3.4 Monitoring of Test Resources................................................................................................. 9
4 Basic Safety Precautions................................................................................ 10
4.1 Pay attention to the instructions in the Operation Manual ................................................ 10
4.2 Personnel training.................................................................................................................. 10
5 Transportation & Storage ............................................................................... 10
5.1 Testing upon acceptance ...................................................................................................... 10
5.2 Packaging / return transport ................................................................................................. 10
6 Unpacking, Setup and Commissioning ......................................................... 11
6.1 Installation Site, Location of Use.......................................................................................... 11
6.2 Unpacking ............................................................................................................................... 11
6.2.1 Placing...................................................................................................................................... 12
6.2.2 Scope of delivery ...................................................................................................................... 12
6.3 Mains connection ................................................................................................................... 13
6.4 Connection of peripheral devices......................................................................................... 13
6.5 Initial Commissioning ............................................................................................................ 13
6.5.1 Stability display......................................................................................................................... 13
6.5.2 Balance zero display ................................................................................................................ 13
Electronic Precision balance
7 Adjustment....................................................................................................... 13
7.1 Automatic adjustment using internal weight ...................................................................... 15
7.2 Adjustment functions “P1 CAL“......................................................................................... 16
7.2.1 Internal adjustment - „iCAL“ ..................................................................................................... 17
7.2.2 Adjustment test – „tCAL“ .......................................................................................................... 18
7.2.3 Criteria for starting automatic internal adjustment – „ACAL“.................................................... 19
7.2.4 Time setting to start automatic internal adjustment – „CAL t“ ................................................ 20
7.2.5 Printout adjustment protocol – „CAL r“..................................................................................... 21
7.3 Verification .............................................................................................................................. 22
8 Operating elements......................................................................................... 24
8.1 Backlit display ........................................................................................................................ 24
8.2 Keyboard overview in weighing mode................................................................................. 24
9 User menu........................................................................................................ 25
9.1 Navigation in the menu.......................................................................................................... 27
2
PLJ_N-BA-e-0930
10 Basic Operation ........................................................................................... 28
10.1 Simple weighing ..................................................................................................................... 28
10.2 Weighing with taring .............................................................................................................. 28
10.2.1 Taring ................................................................................................................................... 28
10.2.2 Delete tare............................................................................................................................ 28
10.3 Standard weighing unit..........................................................................................................29
10.4 Weighing units switch-over................................................................................................... 30
10.5 Underfloor weighing .............................................................................................................. 30
11 Menu function “P6 Func” - Operating modes .......................................... 31
11.1 Settings for Function P6.1 FFun "ALL" ............................................................................... 33
11.2 Parts counting ........................................................................................................................ 34
11.3 Weighing with tolerance range ............................................................................................. 37
11.4 Percent determination ...........................................................................................................40
11.4.1 Determining the reference weight by weighing (function F4 PrcA) ..................................... 40
11.4.2 Determining the reference weight by entering numeric value (function F4 Prcb) ............. 41
11.5 Density determination – functions “d_Co“ and “d_Li“ ...................................................... 42
12 Menu function “P2 GLP“ - GLP/ISO protocol ............................................ 43
13 Menu function “P3 rEAd” basic settings .................................................. 46
13.1 Filter settings .......................................................................................................................... 46
13.2 Rest position display ............................................................................................................. 47
13.3 Auto Zero................................................................................................................................. 48
13.4 Changing readability – decimal place .................................................................................. 49
14 Menu function “P7 othEr”- additional useful functions .......................... 50
14.1 Display background illumination.......................................................................................... 50
14.2 Acoustic signal for key operation......................................................................................... 51
14.3 Printing balance parameters ................................................................................................. 52
15 Data output RS 232C “P4 Print“ ................................................................. 53
15.1 Technical Data ........................................................................................................................ 53
15.2 Pin allocation of the balance output plug (front view) ....................................................... 53
15.3 Interface cable ........................................................................................................................ 54
15.4 Menu function “ P4 Print ” - RS 232C parameter ............................................................... 55
15.4.1 Setting Baudrate „P4.1 bAud” .............................................................................................. 55
15.4.2 Continuous output in standard weighing unit “P4.2 CntA”................................................... 56
15.4.3 Continuous output in current weighing unit „P4.3 Cntb” ..................................................... 57
15.4.4 Setting data output type: manual/automatic „P4.4 rEPL”.................................................... 58
15.4.5 Setting data output for stable/instable weighing value......................................................... 59
15.4.6 Entering minimum weight for automatic output "P4.6 Lo”................................................... 60
15.5 Communication protocol / remote control commands ...................................................... 61
15.5.1 Manual output ...................................................................................................................... 61
15.5.2 PC controlled output............................................................................................................. 62
15.5.3 Output of date/time............................................................................................................... 62
16 Error messages............................................................................................ 63
17 Service, maintenance, disposal.................................................................. 63
17.1 Cleaning .................................................................................................................................. 63
17.2 Service, maintenance............................................................................................................. 63
17.3 Disposal................................................................................................................................... 63
18 Instant help................................................................................................... 64
PLJ_N-BA-e-0930
3

1 Technical Data

KERN PLJ 360-3NM PLJ 600-3NM
Weighing range (max) 360 g 600 g
Readability (d) 1 mg 1 mg
Minimum load (Min) 20 mg 20 mg
Verification value (e) 10 mg 10 mg
Verification class II II
Reproducibility 1 mg 15 mg
Linearity ± 2 mg ± 3 mg
Stabilization time (typical) 4 sec 4 sec
Minimum unit weight at piece counting
Warm-up time 2 hours 4 hours
Adjustment weight internal
Reference quantities at piece counting
Weighing Units g, ct, mg
Electric Supply 230 V / 50 Hz, 11V AC
Operating temperature + 15° C …. + 30° C
Humidity of air max. 80 % (not condensing)
B x D x H [mm]
shield [mm]
1 mg 5 mg
10, 20, 50, freely selectable
with windscreen 206 x 335 x 157 Dimensions housing
without windscreen 206 x 335 x 85
exterior 150 x 150 x 80 Dimensions glass draught
interior 140 x 140 x 65
Weighing plate (stainless steel) [mm]
Weight kg (net) 4 kg
Underfloor weighing Standard hooks
Interface RS 232C
4
128 x 128
PLJ_N-BA-e-0930
KERN PLJ 3500-2NM PLJ 4500-2NM
Weighing range (max) 3500 g 4500 g
Readability (d) 10 mg 10 mg
Minimum load (Min) 500 mg 500 mg
Verification value (e) 100 mg 100 mg
Verification class II II
Reproducibility 10 mg 10 mg
Linearity ± 20 mg ± 20 mg
Stabilization time (typical) 4 sec 4 sec
Minimum unit weight at piece counting
Warm-up time 2 hours 4 hours
Adjustment weight internal
Reference quantities at piece counting
Weighing Units g, ct, mg
Electric Supply 230 V / 50 Hz, 11V AC
Operating temperature + 15° C …. + 30° C
Humidity of air max. 80 % (not condensing)
Dimensions housing
B x D x H [mm]
Weighing plate (stainless steel) [mm]
10 mg 50 mg
10, 20, 50, freely selectable
206 x 335 x 85
195 x 195
Weight kg (net) 4.5 kg
Underfloor weighing standard hooks
Interface RS 232C
PLJ_N-BA-e-0930
5
KERN PLJ 750-3N
Weighing range (max) 750 g
Readability (d) 1 mg
Reproducibility 15 mg
Linearity ± 3 mg
Stabilization time (typical) 4 sec
Minimum unit weight at piece counting
Warm-up time 4 hours
Adjustment weight internal
Reference quantities at piece counting
Weighing Units g, ct, mg
Electric Supply 230 V / 50 Hz, 11V AC
Operating temperature + 15° C …. + 30° C
Humidity of air max. 80 % (not condensing)
B x D x H [mm]
shield [mm]
10, 20, 50, freely selectable
with windscreen 206 x 335 x 157 Dimensions housing
without windscreen 206 x 335 x 85
exterior 150 x 150 x 80 Dimensions glass draught
interior 140 x 140 x 65
5 mg
Weighing plate (stainless steel) [mm]
Weight kg (net) 4 kg
Underfloor weighing Standard hooks
Interface RS 232C
6
128 x 128
PLJ_N-BA-e-0930

2 Declaration of -Conformity

KERN & Sohn GmbH
D-72322 Balingen-Frommern Postfach 4052 E-Mail: info@kern-sohn.de
Tel: 0049-[0]7433- 9933-0 Fax: 0049-[0]7433-9933-149 Internet: www.kern-sohn.de
Konformitätserklärung
EC-Konformitätserklärung EC-Declaration of -Conformity
Déclaration de conformité EC-Declaración de Conformidad
EC­EC-Dichiarazione di conformità EC-Conformiteitverklaring EC- Declaração de conformidade EC- Prohlášení o shode EC-Deklaracja zgodności EC-Заявление о соответствии
D
GB
CZ
E
F
I
NL
P
PL
RUS
Konformitäts­erklärung Declaration of conformity Prohlášení o shode Declaración de conformidad Déclaration de conformité Dichiarazione di conformitá Conformiteit­verklaring Declaração de conformidade Deklaracja zgodności
Заявление о соответствии
Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den nachstehenden Normen übereinstimmt. We hereby declare that the product to which this declaration refers conforms with the following standards. Tímto prohlašujeme, že výrobek, kterého se toto prohlášení týká, je v souladu s níže uvedenými normami. Manifestamos en la presente que el producto al que se refiere esta declaración está de acuerdo con las normas siguientes Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente déclaration, est conforme aux normes citées ci-après. Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce è conforme alle norme di seguito citate. Wij verklaren hiermede dat het product, waarop deze verklaring betrekking heeft, met de hierna vermelde normen overeenstemt. Declaramos por meio da presente que o produto no qual se refere esta declaração, corresponde às normas seguintes. Niniejszym oświadczamy, że produkt, którego niniejsze oświadczenie dotyczy, jest zgodny z poniższymi normami.
Мы заявляем, что продукт, к которому относится данная декларация, соответствует перечисленным ниже нормам.
Electronic Balance: KERN PLJ_N
Mark applied EU Directive Standards
2004/108/EC EMC
2006/95/EC Low Voltage
Date: 27.01.2010 Signature:
Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0,Fax +49-[0]7433/9933-149
PLJ_N-BA-e-0930
EN 61326 : 2006
EN 61010-1 : 2004
Gottl. KERN & Sohn GmbH Management
7
KERN & Sohn GmbH
D-72322 Balingen-Frommern Postfach 4052 E-Mail: info@kern-sohn.de
Tel: 0049-[0]7433- 9933-0 Fax: 0049-[0]7433-9933-149 Internet: www.kern-sohn.de
Konformitätserklärung
EC-Konformitätserklärung EC-Declaration of -Conformity
Déclaration de conformité EC-Declaración de Conformidad
EC­EC-Dichiarazione di conformità EC-Conformiteitverklaring EC- Declaração de conformidade EC- Prohlášení o shode EC-Deklaracja zgodności EC-Заявление о соответствии
D
GB
CZ
E
F
I
NL
P
PL
RUS
Konformitäts­erklärung
Declaration of conformity
Prohlášení o shode
Declaración de conformidad
Déclaration de conformité
Dichiarazione di conformitá
Conformiteit­verklaring
Declaração de conformidade
Deklaracja zgodności
Заявление о соответствии
Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den nachstehenden Normen übereinstimmt. Diese Erklärung gilt nur in Verbindung mit der Konformitätsbescheinigung einer benannten Stelle. We hereby declare that the product to which this declaration refers conforms with the following standards. This declaration is only valid with the certificate of conformity by a notified body. Tímto prohlašujeme, že výrobek, kterého se toto prohlášení týká, je v souladu s níže uvedenými normami. Toto prohlášení platí pouze ve spojitosti s deklarací o souladu uvedeného pracoviště se směrnicemi EU. Manifestamos en la presente que el producto al que se refiere esta declaración está de acuerdo con las normas siguientes. Esta declaración solo será válida acompañada del certificado de conformidad de una institución renombrada. Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente déclaration, est conforme aux normes citées ci-après. Cette déclaration est valide seulement avec un certificat de conformité d’un organisme notifié. Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce è conforme alle norme di seguito citate. Questa dichiarazione sarà valida solo se accompagnata dal certificato di conformità della parte nominale. Wij verklaren hiermede dat het product, waarop deze verklaring betrekking heeft, met de hierna vermelde normen overeenstemt. Deze verklaring geldt uitsluitend in
verbinding met het certificaat van overeenstemming vanwege een daarmee belaste instantie.
Declaramos por meio da presente que o produto no qual se refere esta declaração, corresponde às normas seguintes. Esta declaração vale só em combinação com um certificado de conformidade duma instituição nomeada. Niniejszym oświadczamy, że produkt, którego niniejsze oświadczenie dotyczy, jest zgodny z poniższymi normami. Niniejsze oświadczenie obowiązuje wyłącznie w połączeniu z oświadczeniem o zgodności danego miejsca.
Мы заявляем, что продукт, к которому относится данная декларация, соответствует перечисленным ниже нормам. Эта декларация действует
совместно с удостоверением соответствия названной лаборатории.
Electronic Balance: KERN PLJ_NM
EU Directive Standards
90/384/EEC EN 45501 TCM 128/08-4606 CMI
EC-type-approval certificate no.
Issued by
Date: 27.01.2010 Signature:
Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0,Fax +49-[0]7433/9933-149
8
Gottl. KERN & Sohn GmbH Management
PLJ_N-BA-e-0930

3 Basic Information (General)

3.1 Proper use

The balance you purchased is intended to determine the weighing value of material to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing plate. As soon as a stable weighing value is reached the weighing value can be read.

3.2 Improper Use

Do not use balance for dynamic weighings. In the event that small quantities are removed or added to the material to be weighed, incorrect weighing results can be displayed due to the “stability compensation“ in the balance. (Example: Slowly draining fluids from a container on the balance.) Do not leave permanent load on the weighing plate. This may damage the measuring system.
-­Impacts and overloading exceeding the stated maximum load (max) of the balance, minus a possibly existing tare load, must be strictly avoided. Balance may be damage by this. Never operate balance in explosive environment. The serial version is not explosion protected. The structure of the balance may not be modified. This may lead to incorrect weighing results, safety-related faults and destruction of the balance. The balance may only be used according to the described conditions. Other areas of use must be released by KERN in writing.

3.3 Warranty

Warranty claims shall be voided in case
Our conditions in the operation manual are ignored
The appliance is used outside the described uses
The appliance is modified or opened
Mechanical damage or damage by media, liquids, natural wear and tear
The appliance is improperly set up or incorrectly electrically connected
The measuring system is overloaded

3.4 Monitoring of Test Resources

In the framework of quality assurance the measuring-related properties of the balance and, if applicable, the testing weight, must be checked regularly. The responsible user must define a suitable interval as well as type and scope of this test. Information is available on KERN’s home page (www.kern-sohn.com with regard to the monitoring of balance test substances and the test weights required for this. In KERN’s accredited DKD calibration laboratory test weights and balances may be calibrated (return to the national standard) fast and at moderate cost.
PLJ_N-BA-e-0930
9

4 Basic Safety Precautions

4.1 Pay attention to the instructions in the Operation Manual

Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances.

4.2 Personnel training

The appliance may only be operated and maintained by trained personnel.

5 Transportation & Storage

5.1 Testing upon acceptance

When receiving the appliance, please check packaging immediately, and the appliance itself when unpacking for possible visible damage.

5.2 Packaging / return transport

Keep all parts of the original packaging for a possibly required return.
Only use original packaging for returning.
Prior to dispatch disconnect all cables and remove loose/mobile
parts.
Reattach possibly supplied transport securing devices.
Secure all parts such as glass wind screen, weighing platform, power
unit etc. against shifting and damage.
10
PLJ_N-BA-e-0930

6 Unpacking, Setup and Commissioning

6.1 Installation Site, Location of Use

The balances are designed in a way that reliable weighing results are achieved in common conditions of use. You will work accurately and fast, if you select the right location for your balance.
Therefore, observe the following for the installation site:
Place the balance on a firm, level surface;
Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
Protect the balance against direct draughts due to open windows and doors;
Avoid jarring during weighing;
Protect the balance against high humidity, vapours and dust;
Do not expose the device to extreme dampness for longer periods of time.
Non-permitted condensation (condensation of air humidity on the appliance) may occur if a cold appliance is taken to a considerably warmer environment. In this case, acclimatize the disconnected appliance for ca. 2 hours at room temperature.
Avoid static charge of goods to be weighed or weighing container.
Major display deviations (incorrect weighing results) may be experienced should electromagnetic fields (e.g. due to mobile phones or radio equipment), static electricity accumulations or instable power supply occur. Change location or remove source of interference.

6.2 Unpacking

Carefully remove the balance from the packaging, remove plastic cover and setup balance at the intended workstation.
PLJ_N-BA-e-0930
11

6.2.1 Placing

Design layout of weighing balance:
Assemble all parts as per illustration. To ensure earthing of weighing plate, Place carrier (2) right at rear.
Models readability d = 1 mg: Models readability d = 10 mg:
2
Level balance with foot screws until the air bubble of the water balance is in the prescribed circle

6.2.2 Scope of delivery

Serial accessories:
Balance
Mains power supply
Operating Manual
Draught shield (models providing readability of d =
1mg only)
Hook for underfloor weighing
12
PLJ_N-BA-e-0930

6.3 Mains connection

Power is supplied via the external mains adapter. The stated voltage value must be the same as the local voltage. Only use original KERN mains adapters. Using other makes requires consent by KERN. After connection to the power supply the balance will carry out a self-test. The balance is ready for weighing when the weight display appears. If the display diverts from zero, press the TARE-key.

6.4 Connection of peripheral devices

Before connecting or disconnecting of additional devices (printer, PC) to the data interface, always disconnect the balance from the power supply.
With your balance, only use accessories and peripheral devices by KERN, as they are ideally tuned to your balance.

6.5 Initial Commissioning

In order to obtain exact results with the electronic balances, your balance must have reached the operating temperature (see warming up time chap. 1). During this warming up time the balance must be connected to the power supply. The accuracy of the balance depends on the local acceleration of gravity. Strictly observe hints in chapter Adjustment.

6.5.1 Stability display

The appearance of the stability symbol [ ] on the display indicates that the balance is in a stable state. If the status is instable the [ ] display disappears.

6.5.2 Balance zero display

If an exact zero reading is not displayed on the balance in spite of the weighing dish being empty, press the TARE key and the balance will start resetting to zero [ a ]
.

7 Adjustment

As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated - in compliance with the underlying physical weighing principle - to the existing acceleration due to gravity at its place of location (only if the balance has not already been adjusted to the location in the factory). This adjustment process must be carried out for the first commissioning, after each change of location as well as in case of fluctuating environment temperature. To receive accurate measuring values it is also recommended to adjust the balance periodically in weighing operation.
PLJ_N-BA-e-0930
13
Adjustment with internal weight
Automatic start (chpt. 7.1)
1. Temperature controlled (chpt.
7.1)
A change in temperature of 3 °C triggers the weighing balance to send a message announcing internal adjustment:
2. Time controlled (chpt.7.1)
After a set time interval (1 – 12 h) in the menu (P1.5 CALt) has passed the weighing balance will send a message about internal adjustment:
14
PLJ_N-BA-e-0930

7.1 Automatic adjustment using internal weight

With the internal adjustment weight, the weighing accuracy can be checked and re­adjusted at any time.
The automatic adjustment function is always enabled. You can start adjustment at any time by pressing the CAL-key manually.
Automatic adjustment is started each time
after the weighing balance was disconnected from the mains
when temperature changes take place
at the end of a time interval
Temperature / time controlled adjustment is taking place:
5 minutes before the start of automatic adjustment, this will be announced by a "º C“, (change of temperature) or a ““ (after a certain time interval ends) symbol on the display.
or
The user must complete his/her weighing process within this time. After 5 minutes [ CAL 30 ] appears on the display.
This starts a “count down” of 30 seconds [ CAL 30 ] [ CAL 0 ]. During these 30 seconds it is possible to cancel the adjustment by pressing the TARE key. This makes the balance return to weighing mode in order to e. g. complete an unfinished measurement.
After a further 5 minutes the automatic adjustment will restart and [ CAL 30 ] appears.
PLJ_N-BA-e-0930
15

7.2 Adjustment functions “P1 CAL“

Models PLJ:
P1.1 iCAL | [internal adjustment]
P1.2 ECAL | [external adjustment] - not documented
P1.3 tCAL | [adjustment test]
P1.4 ACAL | both/nonE/tenno/tinnE [criteria for starting automatic adjustment]
P1.5 CALt | 1 h ÷ 12h [time setting for starting automatic adjustment]
P1.6 CALr | YES/no [printout adjustment log]
16
PLJ_N-BA-e-0930

7.2.1 Internal adjustment - „iCAL“

The automatic adjustment function with internal adjusting weight is permanently enabled. It is also possible to start adjusting at any time by pressing the CAL-key or by invoking the menu item “P1.1 iCAL“.
Operation: Display:
Start balance by pressing
Press .
Press . Ensure that there are no objects on the weighing plate.
Confirm by pressing ; automatic adjustment will follow. After successful adjustment the balance automatically returns to weighing mode.
Information: Adjustment errors are indicated by an acoustic signal and an error message will appear on the display.
Press and repeat adjustment process.
It is possible to cancel adjustment by pressing
0.0000 g
P1 CAL
P1.1 iCAL
CAL
- - - - - - - -
0.0000 g
PLJ_N-BA-e-0930
17

7.2.2 Adjustment test – „tCAL“

Here, deviation from the last adjustment is determined. This is only a check, i.e. no values are changed.
Operation: Display:
Start balance by pressing
Press .
Press .
Press repeatedly button. Ensure that there are no objects on the weighing plate.
Press , adjustment test is carried out automatically. The result is displayed
Press , balance returns to menu
Press repeatedly until “SAVE“? appears
Press
Press balance automatically jumps back to weighing mode.
to save the changes you made.
, to cancel the changes you made. Afterwards the
0.0000 g
P1 CAL
P1.1 iCAL
P1.3 tCAL
CAL
diff
d 0,0042 g
P1.3 tCAL
SAVE?
0.0000 g
18
PLJ_N-BA-e-0930

7.2.3 Criteria for starting automatic internal adjustment – „ACAL“

Note:
This adjustment function is locked [Er 9 lock] in verifiable devices.
Operation: Display:
Start balance by pressing
Press .
Press .
Press repeatedly until function “ACAL“ appears
Press , current setting will be flashing
Press the arrow keys ( ), to select among the settings below:
nonE Automatic adjustment disabled
tEmp Start automatic adjustment after change in temperature
timE Start automatic Adjustment dependent on setting
of function P1.5 CALt
both Start automatic Adjustment dependent on temperature and time
0.0000 g
P1 CAL
P1.1 iCAL
P1.4 ACAL
nonE
timE
Confirm setting by pressing . Weighing balance returns to menu.
Press repeatedly until “SAVE“? appears.
The changes you made will be saved by pressing .
To cancel changes . Afterwards the balance automatically jumps back to weighing mode.
PLJ_N-BA-e-0930
P1.4 ACAL
SAVE ?
0,0000 g
19

7.2.4 Time setting to start automatic internal adjustment – „CAL t“

Note:
This adjustment function is locked [Er 9 lock] in verifiable devices.
Set the time interval after which automatic adjustment is to be started in this menu item as follows:
Operation: Display:
Start balance by pressing
Press .
Press .
Press repeatedly until function "P1.5 CALt“ appears
Press , current setting will be flashing
Use the arrow keys ( ) to select a time interval 1 – 12 h
Confirm setting by pressing . Weighing balance returns to menu.
Press repeatedly until “SAVE“? appears.
0.0000 g
P1 CAL
P1.1 iCAL
P1.5 CALt
1 h
12 h
P1.5 CALt
SAVE ?
The changes you made will be saved by pressing .
To cancel changes, press the automatically jumps back to weighing mode.
20
key. Afterwards the balance
0.0000 g
PLJ_N-BA-e-0930

7.2.5 Printout adjustment protocol – „CAL r“

Go to this menu item to enable the function used to produce printouts for adjustment data.
Enable/disable function:
Operation: Display:
Start balance by pressing
Press .
Press .
Press repeatedly until function “CALr“ appears
Press , current setting will be flashing
Press the arrow keys ( ), to select among the settings below:
no Data output disabled
yes Data output enabled
0.0000 g
P1 CAL
P1.1 ECAL
P1.6 CALr
Models PLJ
P1.3 CALr
Models PLS
no
YES
Confirm setting by pressing . Weighing balance returns to menu.
Press repeatedly until “SAVE“? appears.
The changes you made will be saved by pressing
Press to cancel the changes you made. Afterwards the balance automatically jumps back to weighing mode.
P1.6 CALr
Models PLJ
P1.3 CALr
Models PLS
SAVE ?
0.0000 g
PLJ_N-BA-e-0930
21
Data output adjustment log:
The content of the data output is determined in menu P2 GLP. All parameters set to YES“ will be issued.
Example:
P2 GLP
P2.1 uSr |
P2.2 PrJ |
P2.3 Ptin | YES
P2.4 PdAt | YES
P2.5 PuS | YES
P2.6 PPrJ | YES
P2.7 PId | YES
P2.8 PFrn | YES

7.3 Verification

General introduction:
According to EU directive 90/384/EEC balances must be officially verified if they are used as follows (legally controlled area):
a) For commercial transactions if the price of goods is determined by weighing. b) For the production of medicines in pharmacies as well as for analyses in the
medical and pharmaceutical laboratory.
c) For official purposes. d) For manufacturing final packages.
In cases of doubt, please contact your local trade in standard.
Verification instructions
An EU type approval exists for balances described in their technical data as verifiable. If a balance is used where obligation to verify exists as described above, it must officially verified and re-verified in regular intervals. Re-verification of a balance is carried out according to the respective national regulations. The validity for verification of balances in Germany is e.g. 2 years. The legal regulation of the country where the balance is used must be observed! After verification the balance is sealed at the indicated positions.
Verification of the balance is invalid without the "seal".
22
PLJ_N-BA-e-0930
Position of the “official seals“:
Position of the official seals
Balances with obligation to verify must be taken out of operation if:
- The weighing result of the balance is outside the error limit. Therefore,
in regular intervals load balance with known test weight (ca. 1/3 of the max. load) and compare with displayed value.
- The reverification deadline has been exceeded.
PLJ_N-BA-e-0930
23

8 Operating elements

8.1 Backlit display

Very contrastful display which can also be red in the darkness.

8.2 Keyboard overview in weighing mode

Key Function
Turn on/off
Invoke adjustment function
Function key (changing parameters after
selecting a function)
Switch-over key into operating modes
Weighing units switch-over
Invoke menu
Arrow keys for navigation in menu
24
Data export to external device
(printer or PC)
Confirm/save settings
Taring
Leave menu
Set weight display at zero
PLJ_N-BA-e-0930

9 User menu

The user menu has seven main menus (P1 – P7), arranged in the following sub­menus:
P1 CAL [adjustment]
P1.1 iCAL | [Internal adjustment] PLJ only
P1.2 ECAL | [external adjustment] – not documented
P1.3 tCAL | [adjustment test]
P1.4 ACAL | both/nonE/tenno/tinnE [criteria for starting automatic adjustment]
P1.5 CALt | 1 h ÷ 12h [time setting for starting automatic adjustment]
P1.6 CALr | YES/no [printout adjustment log]
P2 GLP [good lab practice]
P2.1 USr | _ [user]
P2.2 PrJ | _ [project]
P2.3 Ptin | YES/no [printout time]
P2.4 PdAt | YES/no [printout date]
P2.5 PUSr | YES/no [printout user]
P2.6 PPrJ | YES/no [printout project]
P2.7 PId | YES/no [printout serial number weighing balance]
P2.8 PFr | YES/no [frame printout]
P3 rEAd [basic settings]
P3.1 AuE | Stand/Slouu/FASt [filter settings]
P3.2 ConF | FASt_rEL/Fast/rEL [rest control display]
P3.3 Auto | On/OFF [Auto zero]
P3.4 Ldi9 | ALuuAYS/never/uu_StAb [delete last decimal place]
P4 Print [parameter for serial interface RS 232]
P4.1 bAud | 2400/4800/9600/19200 [baud rate]
P4.2 CntA | YES/no [continuous output standard weighing unit]
P4.3 Cntb | YES/no [continuous output currently set weighing unit]
P4.4 rEPL | YES/no [manual (press key) or automatic output]
P4.5 PStb | YES/no [output stable/instable weighing values]
P4.6 Lo | 000.0000 [input minimal weight for automatic output]
PLJ_N-BA-e-0930
25
P5 unit [weighing unit]
P5.1 StUn | g/mg/ct/oz/ozt/dwt/t/mom/G [standard weighing unit, e.g. “g”]
P5.2 mg | YES/no [mg - milligram]
P5.3 Ct | YES/no [ct carat]
P5.4 oZ | YES/no [oz ounce]
P5.5 oZt | YES/no [ozt – Troy ounce]
P5.6 dwt | YES/no [dwt pennyweight]
P5.7 t | YES/no [t Tael]
P5.8 nno | YES/no [mom - Momme]
P5.9 Gr | YES/no [gr Grain]
P6 Func [operating modes]
P6.1 FFun | ALL/PcS/HiLo/PrcA/Prcb/d_Co/d_Li [selection enabled operating modes]
P6.2 PcS | YES/no [piece counting]
P6.3 HiLo | YES/no [weighing with tolerance control]
P6.4 PrcA | YES/no [percentage weighing, ref. weight determination "weighing"]
P6.5 Prcb | YES/no [percentage weighing, ref. weight determination “numeric“]
P6.6 d_Co | YES/no [determination density “solids”]
P6.7 d_Li | YES/no [determination density “liquids”]
P7 othEr [additional useful functions]
P7.1 bL | On/Aut/OFF [display background illumination]
P7.2 bLbA | no/20/30/40/50/60/70/80/90/100 [Background illumination of display auto-off after x
sec.]
P7.3 bEEP | [
P7.4 PrnS | [“Balance parameters” printout]
Key sound]
26
PLJ_N-BA-e-0930

9.1 Navigation in the menu

Keyboard overview in menu:
Key Function
Call main menu
Menu selection
Confirm menu selection
Invoke sub-menu
To select parameters in a function or to increase/decrease
the value of a digit or number press the - and -key.
To shift to the right or left by one menu item press the - and -key.
Confirm/save settings
Leave menu
Storing / jumping back to weighing mode
Any changes made in the balance memory will only be saved when the storing process is complete.
To achieve this, press the TARE-key several times until “SAVE“? appears.
Any changes carried out are stored by pressing the PRINT key. To cancel changes, press the TARE key. Afterwards the balance automatically jumps back to weighing mode.
PLJ_N-BA-e-0930
27

10 Basic Operation

10.1 Simple weighing

Start balance by pressing the ON/OFF-key and wait for zero display [ a ].
If an exact zero reading is not displayed on the balance in spite of the weighing dish being empty, press the TARE-key and the balance will start resetting to zero.
To change to different weighing units, press the UNIT-key [g], [mg], [ct].
Place goods to be weighed on balance
Wait until the stability display appears [ ]
Read weighing result.
Turn off balance by pressing the ON/OFF key.

10.2 Weighing with taring

The dead weight of any weighing container may be tared away by pressing a button or by numerical input, so that the following weighings show the net weight of the goods to be weighed.

10.2.1 Taring

Place the weighing box and press the TARE-key. The zero display appears.
The tare weight is saved until it is deleted. Information: The tare procedure can be repeated as many times as necessary, for example with initial weighing of several components for a mix (add-on weighing). The limit is reached when the total weighing range capacity is full. After removing the taring container the total weight is displayed as negative display.

10.2.2 Delete tare

Unload the balance and press the TARE –key; the zero display appears.
28
PLJ_N-BA-e-0930

10.3 Standard weighing unit

Selected weighing unit will be retained even after disconnection from the mains.
Invoke menu P5 Unit:
Press repeatedly
button
The currently set weighing unit is flashing
[g] [mg] [ct]
Press the -key repeatedly until the desired weighing unit (See chpt. 1 “Technical Specifications “) appears
The balance returns to menu.
Return to weighing mode:
Press the TARE -button repeatedly until the "SAVE"? inquiry appears. Confirm
query by pressing the PRINT-key or cancel by pressing the TARE-key. The balance returns to weighing mode; the display show the set weighing unit. The set weighing unit remains even after disconnection from the mains.
PLJ_N-BA-e-0930
29

10.4 Weighing units switch-over

A weighing unit may be changed by pressing the UNIT-key several times:
[g] [mg] [ct]
Information:
In calibrated models selection is restricted to options [g] and [mg] [ct]. The different weighing models have integrated different foreign weighing units. For details please refer to chpt. 1, “Technical specifications”.

10.5 Underfloor weighing

Objects unsuitable for placing on the weighing scale due to size or shape may be weighed with the help of the flush-mounted platform. Proceed as follows:
Switch off balance.
Open the closing lid on the bottom of your balance.
Suspend hook for underfloor weighing carefully and completely.
Place your balance over an opening.
Suspend the goods to be weighed from the hook and carry out the weighing.
Fig. 1: Setup of balance for underfloor weighing
!
CAUTION
Always ensure that all suspended objects are stable enough to hold the desired
goods to be weighed safely (danger of breaking).
Never suspend loads that exceed the stated maximum load (max) (danger of
breaking)
Always ensure that there are no persons, animals or objects that might be damaged underneath the load.
NOTICE
After completing the underfloor weighing the opening on the bottom of the balance must always be closed (dust protection).
30
PLJ_N-BA-e-0930

11 Menu function “P6 Func” - Operating modes

Functions can be selected in Menu "6.1.FFun", which are then made available to the operator without having to access the menu every time. All activated operating
modes can be called directly by pressing the key.
Menu activation:
Press repeatedly button
A function can only be activated directly in Menu "6.1.FFun".
If only one operating mode is callable via the key, simply select the desired
function by pressing the arrow keys ( ) and confirm by pressing the PRINT- key.
If several operating modes are callable via the key, select function “ALL“ by
pressing the arrow keys ( ) and confirm by pressing the PRINT-key. The selection as to which Operating modes are deposited in "ALL" is made in the next chapter 11.1.
PLJ_N-BA-e-0930
31
Confirm settings by pressing the PRINT key. Balance jumps back to sub-
menu 6.1.FFun.
Return to weighing mode:
Press the TARE-button repeatedly until the "SAVE"? inquiry appears.
Confirm inquiry by pressing the PRINT-button to save changes and to return to Weighing mode.
-or­press the TARE-button to return to Weighing mode without saving.
Back to parameter selection using the F-key:
The F-key allows you to change set function parameters after having selected a function without having to return to the menu first.
Example: You are in piece counting mode with set reference piece number 20 pcs. You wish to change these to 10 pcs.
Press the F-key; the set reference piece number 20 pcs appears. Press arrow keys to change to 10 pcs and confirm by pressing the PRINT-key. "load" appears on the display. Place the reference weight on balance and confirm
by pressing the PRINT-key.
Remove the reference weight and continue with piece counting.
32
PLJ_N-BA-e-0930

11.1 Settings for Function P6.1 FFun "ALL"

The selection of the menu items, which can then be called up by pressing the button, is made at this point.
Return to weighing mode:
Press the TARE-button repeatedly until the "SAVE"? inquiry appears.
Confirm inquiry by pressing the PRINT-button to save changes and to return to Weighing mode.
-or­press the TARE-button to return to Weighing mode without saving.
All activated operating modes can now be called up in Weighing mode by
pressing the
button.
Press the
button in Weighing mode and the initial activated function
will appear: Select desired function by pressing the arrow keys ( ).
PLJ_N-BA-e-0930
33

11.2 Parts counting

Before the balance can count parts, it must know the average part weight (i.e. reference). Proceed by putting on a certain number of the parts to be counted. The balance determines the total weight and divides it by the number of parts (the so­called reference quantity). Counting is then carried out on the basis of the calculated average piece weight. As a rule:
The higher the reference quantity the higher the counting exactness.
Call add-up function
The current setting is displayed
Press the arrow keys ( ), to select the desired reference piece number.
Reference piece number 10, 20 or 50
Confirm selected reference piece number by pressing the PRINT-key (e.g. 20)
34
PLJ_N-BA-e-0930
Make reference:
Place as many pieces to add-up as required by the set reference piece number.
tare, as required
Remove reference weight. The balance is now in parts counting mode counting all units on the weighing plate.
for selection ”optional reference piece number“ FrEE
Use the arrow keys   to select the digit to be changed
Use the arrow keys   to select the number
PLJ_N-BA-e-0930
35
Confirm the entered reference piece number by pressing the PRINT-key
“LoAd“ appears on the display.
Place as many counting parts on the balance as the set reference quantity
requires, confirm by pressing the PRINT key. Information:
If there is no load on the weighing plate when the PRINT-key is pressed, “Er8 outr“ will appear briefly on the display before the display of the balance returns automatically to weighing mode.
Remove reference weight. The balance is now in parts counting mode counting all units on the weighing plate.
Return to weighing mode
36
PLJ_N-BA-e-0930

11.3 Weighing with tolerance range

For weighing with tolerance ranges you can enter individual upper and lower limits. For tolerance controls such as dosaging, apportioning or sorting the scale will display violated upper or lower limits and show the tolerance tag.
The triangular tolerance marker (S) in the upper part of the display shows whether the goods to be weighed are within the two tolerance limits. The tolerance marker is only in operation during operating mode tolerance weighing; it is otherwise not visible.
The tolerance marker provides the following information:
g
0.000
Lo
ok
Hi
Call function
lo
ok
Hi
Goods to be weighed below tolerance limit Goods to be weighed within tolerance range Goods to be weighed above tolerance limit
PLJ_N-BA-e-0930
37
Setting the lower tolerance limit "Lo“
LQ
0000.000
Use the arrow keys to select the digit to be changed, enabled digit will be flashing
Use the arrow keys   to select the number
Confirm the entered lower tolerance tag by pressing the PRINT-key.
Setting the upper tolerance limit “Hi“.
g
HKK
008.8900
g
38
000.0000
Use the arrow keys to select the digit to be changed, enabled digit will be flashing
Use the arrow keys   to select the number
Confirm the entered lower tolerance tag by pressing the PRINT-key.
g
009.5500
g
PLJ_N-BA-e-0930
The balance is now in checkweighing mode Put on goods to be weighed, tolerance control is started
g
0.000
Information:
If invalid values are entered such as lower tolerance limit greater than upper tolerance limit, the balance will issue the error message and return automatically to weighing mode.
Return to weighing mode
PLJ_N-BA-e-0930
39

11.4 Percent determination

Percent determination allows weight display in percent, in relation to a reference weight.

11.4.1 Determining the reference weight by weighing (function F4 PrcA)

Call function
Make reference
Place the reference weight
Confirm by pressing the PRINT key
This weight is adopted as reference (100%).
Remove reference weight. The balance is now in percent determining mode Place the load on the balance; percentage value in relation to reference body is shown on display:
40
PLJ_N-BA-e-0930

11.4.2 Determining the reference weight by entering numeric value (function F4 Prcb)

Call function
Make reference
You will be asked to enter the reference weight by a flashing message.
Use the arrow keys   to select the digit to be changed, enabled digit
will be flashing.
⇒⇒
Use the arrow keys   to select the number
Confirm the entered reference weight by pressing the PRINT-key.
Now you can place the test objects onto the weighing plate; the percentage to the reference body is displayed
PLJ_N-BA-e-0930
41
Return to weighing mode
Press repeatedly
button

11.5 Density determination – functions “d_Co“ and “d_Li“

For procedure of density determination please see the operating instructions "Set density"
42
PLJ_N-BA-e-0930

12 Menu function “P2 GLP“ - GLP/ISO protocol

2
Quality assurance systems require printouts of weighing results as well as of correct adjustment of the balance stating date and time and balance identification. The easiest way is to have a printer connected.
The content of the data output is determined in menu “P2 GLP“. All parameters set to “YES“ will be issued. Examples:
P2 GLP
P2.1 uSr | max. 8 digits
P2.2 PrJ | max. 8 digits
P2.3 Ptin | YES
P2.4 PdAt | YES
P2.5 PuS | YES
P2.6 PPrJ | YES
P2.7 Pid | YES
P2.8 PFrn | YES
P2.8 PFrn: YES P2.8 PFrn: no
---------------------------------------
Date :20.03.07 Time :11.31.07 UserID :Mustermann Balance ID :180151
Date : Time : UserID : Balance ID :180151
19.3406 g
19.3406 g
----------------------------------------
Definition of a standard log:
Operation: Display:
Call menu point „P2 GLP” :
Press .
Press .
Press . The first menu item „P2.1 Usr“ for entering the user name appears.
0.0000 g
P1 CAL
P2 GLP
P2.1 Usr
PLJ_N-BA-e-0930
43
The first digit is flashing
To change the number value, press the keys.
Press the keys, to select the digit on the
left/right (the relevant current digit flashes).
+ .
Confirm input with
button.
Enter user name or number (max. 8 digits) with the help of the arrow keys
Example: Data input:
HEMN
Data output:
KERN
Overview data input / data output:
After confirming by pressing the balance will return to the menu.
, the next menu item “P2.2 PrJ“ used to enter the project name
appears.
P2.1 Usr
P2.2 PrJ
44
PLJ_N-BA-e-0930
Press , first digit is flashing
Press the arrow keys to enter project name or number (max. 8 digits).
Confirm setting by pressing . Weighing balance returns to menu.
Press until the next menu item “P2.3 Ptin” used to enter the time appears.
Press , current setting will be flashing
Press the arrow keys   to select among the settings below:
no = time not displayed YES = time displayed
Confirm setting by pressing . Weighing balance returns to menu.
_
P2.2 PrJ
P2.3 Ptin
no
YES
P2.3 Ptin
Press until the next menu item “P2.4 PdAt” used to enter the
P2.4 PdAt
date appears.
Use the same steps to enter further GLP-Parameter.
Menu overview
P2.1 Usr | _ [user]
P2.2 PrJ | _ [project]
P2.3 Ptin | YES/no [printout time]
P2.4 PdAt | YES/no [printout date]
P2.5 PUSr | YES/no [printout user]
P2.6 PPrJ | YES/no [printout project]
P2.7 Pid | YES/no [printout serial number weighing balance]
P2.8 PFr | YES/no [frame printout]
Return to weighing mode:
Press the TARE-button repeatedly until the "SAVE"? inquiry appears. Confirm inquiry by pressing the PRINT-button to save changes and to return to Weighing mode.
-or-
press the TARE-button to return to Weighing mode without saving.
PLJ_N-BA-e-0930
45

13 Menu function “P3 rEAd” basic settings

Basic settings can be changed and functions activated in Menu "P3 rEAd". It is now possible to change individual weighing requirements.

13.1 Filter settings

This menu item allows the balance to be set according to specific ambient conditions and measuring purposes.
Call up menu:
Press repeatedly button
Press the button as often as
necessary until the required setting* appears:
confirm
* Press the arrow keys ( ), to select among the settings below: FASt Sensitive and fast (very quiet set-up location). StAnd Default Slouu Robust but slow (very busy set-up location)
Return to weighing mode:
Press the TARE-button repeatedly until the "SAVE"? inquiry appears.
Confirm inquiry by pressing the PRINT-button to save changes and to return to Weighing mode.
-or­press the TARE-button to return to Weighing mode without saving.
46
PLJ_N-BA-e-0930

13.2 Rest position display

Call up menu:
Press repeatedly button
Press the button as often as
necessary until the required setting appears:
confirm
* Press the arrow keys ( ), to select among the settings below:
FASt_rEL = rest position control fast
FASt = rest position control fast and accurate
rEL = rest position control accurate
Return to weighing mode:
Press the TARE-button repeatedly until the "SAVE"? inquiry appears.
Confirm inquiry by pressing the PRINT-button to save changes and to return to Weighing mode.
-or­press the TARE-button to return to Weighing mode without saving.
PLJ_N-BA-e-0930
47

13.3 Auto Zero

This function helps to tare automatically deviations from the zero display
Call up menu:
Press repeatedly button
Press repeatedly button
Return to weighing mode:
Press the TARE-button repeatedly until the "SAVE"? inquiry appears.
Confirm inquiry by pressing the PRINT-button to save changes and to return to Weighing mode.
-or­press the TARE-button to return to Weighing mode without saving.
Press the button as often as
necessary until the required setting appears (no = off, yes = on)
confirm
48
PLJ_N-BA-e-0930

13.4 Changing readability – decimal place

Readability may be reduced by 1 digit on the weighing balances, as required. The last decimal place will be rounded and removed from the display. Call up menu:
Press repeatedly button
Press repeatedly button
Press the button as often as
necessary until the required setting appears:
confirm
* Press the arrow keys ( ), to select among the settings below:
AlwAYS =shows last decimal place nEuEr =does not show last decimal place uu_StAb = last decimal place will only be shown for stable weighing values
Return to weighing mode:
Press the TARE-button repeatedly until the "SAVE"? inquiry appears.
Confirm inquiry by pressing the PRINT-button to save changes and to return to Weighing mode. -or­press the TARE-button to return to Weighing mode without saving.
PLJ_N-BA-e-0930
49

14 Menu function “P7 othEr”- additional useful functions

Here, you can set the parameters that influence the operation of the balance, such as background lighting and key sounds.

14.1 Display background illumination

Call up menu:
Press repeatedly button
Press the button as often as
necessary until the required setting appears:
confirm
* Press the arrow keys ( ), to select among the settings below:
ON = Background illumination on OFF = Background illumination off Aut The background illumination will be switched off
automatically 10 sec after having reached a stable weighing value.
Return to weighing mode:
Press the TARE-button repeatedly until the "SAVE"? inquiry appears.
Confirm inquiry by pressing the PRINT-button to save changes and to return to Weighing mode.
-or-
press the TARE-button to return to Weighing mode without saving. Information: Backlighting automatically switched off 10 seconds after achieving a stable weighing value The background light will switch on automatically after the weighing value has changed.
50
PLJ_N-BA-e-0930

14.2 Acoustic signal for key operation

Call up menu:
Return to weighing mode:
Press the TARE-button repeatedly until the "SAVE"? inquiry appears.
Confirm inquiry by pressing the PRINT-button to save changes and to
return to Weighing mode.
-or-
press the TARE-button to return to Weighing mode without saving.
Press repeatedly button
Press the button as often as
necessary until the required setting appears (no = off, yes = on)
confirm
PLJ_N-BA-e-0930
51

14.3 Printing balance parameters

Call up menu:
Press the F-key; output of balance parameter is achieved via RS 232 interface.
Press repeatedly button
Press repeatedly button
52
PLJ_N-BA-e-0930

15 Data output RS 232C “P4 Print“

15.1 Technical Data

8-bit ASCII Code 8 data bits, 1 stop bit, no parity bit Baudrate selectable from 2400 - 19200 Baud (factory setting 4800 Baud.) For operation with interface faultless operation is only ensured with the correct
KERN – interface cable (max. 2m)
Transfer modes:
Manually after pressing the PRINT key
Continuously, according to setting
Automatically according to stability display
On request by external device (For remote control commands, see chpt. 16.5))
Output conditions:
stable – Output for stable weighing value
unstable – sequential output by pressing the PRINT button (Marks in the
printout: <?>)

15.2 Pin allocation of the balance output plug (front view)

5 1
9 6
Pin 2: Receive data Pin 3: Transmit data Pin 5: Signal ground
PLJ_N-BA-e-0930
53

15.3 Interface cable

PC 25-channel
Balance
PC 9-channel
Printer
Balance
Balance
Balance
2 (RxD) 3 (TxD) 4 (DTR) 5 (GND) 6 (DSR)
54
PLJ_N-BA-e-0930

15.4 Menu function “ P4 Print ” - RS 232C parameter

15.4.1 Setting Baudrate „P4.1 bAud”

Call up menu:
Press repeatedly button
Press the button as often as
necessary until the required setting appears:
confirm
* Press the arrow keys ( ), to select among the settings below:
2400 bit/s
4800 bit/s
9600 bit/s
19200 bit/s
Return to weighing mode:
Press the TARE -button repeatedly until the "SAVE"? inquiry appears.
Confirm inquiry by pressing the PRINT-button to save changes and to
return to Weighing mode.
-or-
press the TARE-button to return to Weighing mode without saving.
PLJ_N-BA-e-0930
55

15.4.2 Continuous output in standard weighing unit “P4.2 CntA”

Call up menu:
* Press the arrow keys ( ), to select among the settings below: CntA no Function deactivated CntA yes Function activated
Return to weighing mode:
Press the TARE -button repeatedly until the "SAVE"? inquiry appears.
Confirm inquiry by pressing the PRINT-button to save changes and to
return to Weighing mode.
-or-
press the TARE-button to return to Weighing mode without saving.
Press repeatedly button
Press the button as often as
necessary until the required setting appears:
confirm
56
PLJ_N-BA-e-0930

15.4.3 Continuous output in current weighing unit „P4.3 Cntb”

* Press the arrow keys ( ), to select among the settings below: Cntb no Function deactivated Cntb yes Function activated
Return to weighing mode:
Press the TARE -button repeatedly until the "SAVE"? inquiry appears.
Confirm inquiry by pressing the PRINT-button to save changes and to
return to Weighing mode.
-or-
press the TARE-button to return to Weighing mode without saving.
Press repeatedly button
Press repeatedly button
Press the button as often as
necessary until the required setting appears:
confirm
PLJ_N-BA-e-0930
57

15.4.4 Setting data output type: manual/automatic „P4.4 rEPL”

* Press the arrow keys ( ), to select among the settings below: rEPl no Manual issue after pressing the PRINT key. rEPl yes Automatic issue of first stable weighing value
Press repeatedly button
Press repeatedly button
Press the button as often as
necessary until the required setting appears:
confirm
Sequence of operations:
1. Taring
2. Place weight, issue of first stable weighing value
3. Renewed output only possible after weight was removed. Condition: Display +/- 50 display steps from zero
4. Place next weight.
Return to weighing mode:
Press the TARE -button repeatedly until the "SAVE"? inquiry appears.
Confirm inquiry by pressing the PRINT-button to save changes and to return to Weighing mode.
-or­press the TARE-button to return to Weighing mode without saving.
58
PLJ_N-BA-e-0930

15.4.5 Setting data output for stable/instable weighing value

“P4.5 PStb” (models with non-verifiable setting only)
* Press the arrow keys ( ), to select among the settings below: PStb no Output even for unstable weighing value PStb yes Output for stable weighing value only
Return to weighing mode:
Press the TARE -button repeatedly until the "SAVE"? inquiry appears.
Confirm inquiry by pressing the PRINT-button to save changes and to return to Weighing mode.
-or­press the TARE-button to return to Weighing mode without saving.
Press repeatedly button
Press repeatedly button
Press the
necessary until the required setting appears:
confirm
button as often as
PLJ_N-BA-e-0930
59

15.4.6 Entering minimum weight for automatic output "P4.6 Lo”

Weighing value issued automatically if current weighing value exceeds entered minimum value. The next weighing value will not be issued unless the weighing value has meanwhile dropped below the entered weighing value.
Press repeatedly button
Press repeatedly button
Press the keys, to select the
digit on the left/right (the relevant current digit flashes).
To change the number value, press the   keys.
Return to weighing mode:
Press the TARE -button repeatedly until the "SAVE"? inquiry appears.
Confirm inquiry by pressing the PRINT-button to save changes and to return to Weighing mode.
-or­press the TARE-button to return to Weighing mode without saving.
confirm
60
PLJ_N-BA-e-0930

15.5 Communication protocol / remote control commands

Instruction: Meaning of Instruction:
Z
T
S
SI
SU
SUI
C1
C0
CU1
CU0
Set weight display at zero
Taring
Send weight value immediately
Send stable weight value
Send stable weighing value in current weighing value
Send weighing value immediately in current weighing unit
Turn on continuous transmission in standard weighing unit
Turn off continuous transmission in standard weighing unit
Turn on continuous transmission in current weighing unit
Turn off continuous transmission in current weighing unit
Complete each instruction with CR LF.

15.5.1 Manual output

The user may start output manually by pressing the PRINT-key (For settings see chapter 16.4.5, function “P4.5 PStb”, locked for verifiable devices).
Data record format:
1 2 3 4 - 12 13 14 - 16 17 18
Stability indicator
Blank
Signs
Weight
Blank
Unit
CR
LF
PLJ_N-BA-e-0930
61
Stability indicator Space if stable, ? if not stable ^ if overload v if underload Presign: Space, if positive negative sign, if negative Weight: 9 signs, right justified Unit: 3 signs, left justified

15.5.2 PC controlled output

Response message of balance after remote instruction was sent: XX_ Instruction: XX_A CR LF Instruction accepted; will be executed XX_ I CR LF Instruction received; impossible to carry out
XX_^ CR LF Instruction received but time overflow error occurred
XX_v CR LF Instruction received, but insufficient load XX_E CR LF Error during execution, timeout for stable weighing value
exceeded
Data record format:
1 - 3 4 5 6 7 8 - 16 17 18 - 20 21 22
control
Remote
command
Blank
Stability
indicator
mark
Blank
Weight
Blank
Instruction: 1. up to 3 signs Stability indicator: Space if stable, ? if not stable ^ if overload v if underload Presign: Space, if positive negative sign, if negative Weight: 9 signs, right justified Unit: 3 signs, left justified

15.5.3 Output of date/time

Output of date and time is enabled in menu item ”P2 GLP“:
PdAt – yes
LF
Unit
CR
Ptin - yes
62
PLJ_N-BA-e-0930

16 Error messages

Er1 Hi
Er2 nuLL
Er3 FuL1
Er4 FuL2
Er5 rout
Er7 tout
Er8 outr
Er9 Lock
Er10 cal
Initial weight error
Value below allowed range
Value above allowed range
Weighing range exceeded
Value outside allowed range e.g. tare value <= 0, Reference weight = 0
Zero/Tare not possible as stable weighing value is still pending
Input outside range e.g. for tolerance control: Input upper limit<lower limit
Function blocked
Adjustment error e.g. incorrect adjustment weight}

17 Service, maintenance, disposal

17.1 Cleaning

Before cleaning, please disconnect the appliance from the operating voltage.
Please do not use aggressive cleaning agents (solvents or similar agents), but a cloth dampened with mild soap suds. Ensure that no liquid penetrates into the device and wipe with a dry soft cloth. Loose residue sample/powder can be removed carefully with a brush or manual vacuum cleaner.
Spilled weighing goods must be removed immediately.

17.2 Service, maintenance

The appliance may only be opened by trained service technicians who are authorized by KERN. Before opening, disconnect from power supply.

17.3 Disposal

Disposal of packaging and appliance must be carried out by operator according to valid national or regional law of the location where the appliance is used.
PLJ_N-BA-e-0930
63

18 Instant help

In case of an error in the program process, briefly turn off the balance and disconnect from power supply. The weighing process must then be restarted from the beginning.
Help:
Fault Possible cause
The displayed weight does not glow.
The displayed weight is permanently changing
The balance is not switched on.
The mains supply connection has been interrupted
(mains cable not plugged in/faulty).
Power supply interrupted.
(Rechargeable) batteries are inserted incorrectly or
empty
No (rechargeable) batteries inserted.
Draught/air movement
Table/floor vibrations
Weighing plate has contact with other objects.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if possible)
The weighing result is obviously incorrect
Should other error messages occur, switch balance off and then on again. If the error message remains inform manufacturer.
The display of the balance is not at zero
Adjustment is no longer correct.
Great fluctuations in temperature.
Warm-up time was ignored.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if possible)
64
PLJ_N-BA-e-0930
Loading...