The device must be repaired only by trained specialist staff or personnel with professional
formation (such as a repair-specialist accredited by law concerning verification). The service
manual is obligatory for repair work. After repair, original conditions of the device have to be
restored. Only original spare parts should be used.
Instructions about conformity-evaluated scales:
Repair must be carried only at 100% compliance with the type approval. A violation of this
specification will result in a loss of the type approval! After successful repair the balance will
have to be reverified before it can be used again in a statutorily regulated field.
Detailed instructions about conformity-evaluated scales:
Repair must be carried only at 100% compliance with the type approval. A violation of this
specification will result in a loss of the type approval!
After successful repair the balance will have to be reverified before it can be used again in a
statutorily regulated field.
2 Introdution
This service manual covers the MWB series and is edited for the authorized servicing
personnel.
Note all rights are reserved. Copying any part of this manual is prohibited without our
permission.
In this lineup it is an eco product, Therefore, it is not intended to represent the repair manual
in detail, since the construction of the balance is very simple. It is therefore only referring to
the list of related to disposal spare parts.
4 MWB-SH-e-1810.docx
3 Appliance overview
Bubble level and handle (2x)
Conveyor roller (2x)
1. Cover of weighing cell feet
2. Weighing surface
3. Bubble level
4. Hand grip
5. Connection cable
6. Display Unit
MWB-SH-e-1810.docx 5
3.1 Details
Display unit rear side
1
Battery compartment
2
Connecting cable
display unit - platform
6 MWB-SH-e-1810.docx
3.2 Overview display
Display
Designation
Description
GROSS
Gross weight display
Lights up during indication of the gross
weight
NET
Net weight display
Lights up during indication of the net weight
Illuminated after weighing scale was tared
ZERO
Zeroing display
Should the balance not display exactly zero
despite empty scale pan, press the button.
Your balance will be set to zero after a short
standby time.
STABLE
Stability display
Scales are in a steady state
BMI
Body Mass Index
Is actively displayed during active BMI
function
HOLD
HOLD function
Is displayed with active hold function
Battery symbol
Lights when the voltage drops below the
prescribed minimum.
Lights when the battery capacity is almost
exhausted.
Lights when the battery is fully charged.
MWB-SH-e-1810.docx 7
3.3 Overview Keyboard
Button
Designation
Function
ON/OFF button
Turn on/off
HOLD button
Hold function
BMI key
Calculation of the Body Mass Index
In menu:
Confirm selection
For numeric entry:
Confirm numerical value
Function key
In menu:
Call up menu
Select menu items
For numeric entry:
Increase numerical value
Zero setting key
Weighing scale will be reset to „0.0“
For numeric entry:
Change decimal place
TARE key
Tare balance
8 MWB-SH-e-1810.docx
4 Basic Information (General)
4.1 Proper use
This weighing scale is designed for determining the weight of a person whilst standing or
sitting.
On multifunctional weighing scales, the weighed person should carefully step onto the center
of the weighing platform and remain standing without moving. If weighed with a wheelchair,
the wheelchair should remain in the center of the weighing plate.
As soon as a stable weighing value is reached the weighing value can be read.
The weighing scale is designed for continuous duty.
4.2 Improper Use
Do not use these scales for dynamic weighing processes.
Do not leave permanent load on the weighing plate. This may damage the measuring
system.
Impacts and overloading exceeding the stated maximum load (max) of the weighing plate,
minus a possibly existing tare load, must be strictly avoided. This could cause damage to the
balance.
Never operate balance in explosive environment. The serial version is not explosion
protected. It should be noted that a flammable mixture of anesthetics and oxygen or laughing
gas may occur.
The structure of the balance may not be modified. This may lead to incorrect weighing
results, safety-related faults and destruction of the balance.
The balance may only be used according to the described conditions. Other areas of use
must be released by KERN in writing.
The balance cannot be used to determine a body weight in practice of medicine.
4.3 Warranty
Warranty claims shall be voided in case
Our conditions in the operation manual are ignored
The appliance is used outside the described uses
Modification or opening of appliances
Mechanical damage and damage caused by media, liquids,
Natural wear and tear
The appliance is improperly set up or incorrectly electrically connected
The measuring system is overloaded
Dropping of scales
4.4 Monitoring of Test Resources
In the framework of quality assurance the measuring-related weighing properties of the
balances and, if applicable, the testing weight, must be checked regularly. The responsible
user must define a suitable interval as well as type and scope of this test. Information is
available on KERN’s home page (www.kern-sohn.com) with regard to the monitoring of
balance test substances and the test weights required for this. In KERN’s accredited DKD
calibration laboratory test weights and balances may be calibrated (return to the national
standard) fast and at moderate cost.
MWB-SH-e-1810.docx 9
5 Basic Safety Precautions
5.1 Pay attention to the instructions in the Operation Manual
Carefully read this operation manual before setup
and commissioning, even if you are already familiar
with KERN balances.
6 Transport and storage
6.1 Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the appliance itself
when unpacking for possible visible damage.
6.2 Packaging / return transport
Keep all parts of the original packaging for a possibly required return.
Only use original packaging for returning.
Prior to dispatch disconnect all cables and remove loose/mobile parts.
Reattach possibly supplied transport securing devices.
Secure all parts against shifting and damage.
10 MWB-SH-e-1810.docx
7 Unpacking, Setup and Commissioning
7.1 Place of installation / place of operation
The balances are designed in a way that reliable weighing results are achieved in common
conditions of use. You will work accurately and fast, if you select the right location for your
balance.
On the installation site observe the following:
Place scales on a stable, even surface
Avoid extreme heat as well as temperature fluctuation caused by installing next to a
radiator or in the direct sunlight;
Protect the balance against direct draughts due to open windows and doors;
Avoid jarring during weighing;
Protect the balance against high humidity, vapours and dust;
Do not expose the device to extreme dampness for longer periods of time. Non-permitted
condensation (condensation of air humidity on the appliance) may occur if a cold
appliance is taken to a considerably warmer environment. In this case, acclimatize the
disconnected appliance for ca. 2 hours at room temperature.
Avoid static charge of the balance and of the person to be weighed.
Avoid contact with water.
Major display deviations (incorrect weighing results) may be experienced should
electromagnetic fields (e.g. due to mobile phones or radio equipment), static electricity
accumulations or instable power supply occur. In that case, the location must be changed.
7.2 Unpacking
Remove the individual components of the balance or the complete balance from the
packaging with care and install at the intended location. When using the power pack, ensure
that the power cable does not produce a risk of stumbling.
Level balance with foot screws until the air bubble of the water
balance is in the prescribed circle.
Check levelling regularly.
7.3 Scope of delivery
7.4 Levelling
MWB-SH-e-1810.docx 11
8 Menu
Call up menu
In weighing mode, press and the first function [F1
oFF] will be displayed.
Select function
With help of , the individual functions can be
selected one after the other.
Change settings
Confirm selected function by . The current setting
will be displayed.
Select desired setting by and confirm with ,
the balance returns to the menu.
Exit menu/
Return to weighing
mode
Press , the balance will return to weighing mode.
Function
Settings
Description
Adjustment
Adjustment
d 0, d 0.0, d 0.00,
d 0.000, d 0.0000
Capacity
Automatic cutout
Auto Off
oFF 3
Automatic shutdown after 3 min.
oFF 5
Automatic shutdown after 5 min.
oFF 15
Automatic shutdown after 15 min.
8.1 Navigation in the menu
8.2 Menu overview
12 MWB-SH-e-1810.docx
Gravity
Gravity
not documented
MWB-SH-e-1810.docx 13
9 Error Messages
Display
Description
Zero range exceeded
(on start-up or when pressing the key)
Load on weighing pan
Excess load, during zero setting of weighing scale
Incorrect adjusting process
Fault on load cell
Value outside the A/D converter range
Damaged weighing cell
Damaged electronics
Unable to initialize zero point
Measuring cell defective / overloaded
Object on weighing pan / contact
Transport safety device has not been removed
Main board defective
OL or --------
Overload
Unload, switch off and adjust the balance
------or „0“
Underload
Unload, switch off and adjust the balance
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
14 MWB-SH-e-1810.docx
10 Servicing, maintenance, disposal
Before any maintenance, cleaning and repair work disconnect the appliance
from the operating voltage.
In case of a fault in the program sequence, the balance should be shortly switched off.
The weighing process must then be restarted from the beginning.
Failure:
Possible causes:
The displayed weight does
not glow.
The balance is not switched on.
The mains supply connection has been interrupted
(mains cable not plugged in/faulty).
Power supply interrupted.
Battery inserted incorrectly or empty.
No battery inserted.
The displayed weight is
permanently changing
Draught/air movement
Table/floor vibrations
The seating surface is in contact with foreign bodies or
is not correctly positioned.
Electromagnetic fields / static charging (choose different
location/switch off interfering device if possible)
10.1 Cleaning
10.2 Cleaning / disinfecting
Clean weighing platform (such as seat pan) as well as casing with household detergents or
commercially available disinfectants, e.g. 70% isopropanol. We recommend a disinfectant
suitable for wiping disinfection. Please follow manufacturer’s instructions.
Do not use abrasive or aggressive cleaners such as spirits or alcohol or similar as they might
damage the high-quality surface.
10.3 Servicing, maintenance
The appliance may only be opened by trained service technicians who are authorized by
KERN.
Disconnect the scales before opening.
10.4 Disposal
Disposal of packaging and appliance must be carried out by operator according to valid
national or regional law of the location where the appliance is used.
11 Instant help
MWB-SH-e-1810.docx 15
The weighing result is
obviously incorrect
The display of the balance is not at zero.
Adjustment is no longer correct.
Great fluctuations in temperature.
The balance is on an uneven surface.
Electromagnetic fields / static charging (choose different
location/switch off interfering device if possible)
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
16 MWB-SH-e-1810.docx
12 Adjustment
Prepare the required adjustment weight. The adjustment weight to be applied
depends on the capacity of a weighing scale, see chap. 1.
Carry out adjustment as closely as possible to admissible maximum load of
weighing scale. Info about test weights can be found on the Internet at:
http://www.kern-sohn.com.
Observe stable environmental conditions. For warm-up time required for
stabilization
Switch-on balance by and during the selftest press .[F0
CAL] is displayed.
Press , [CAL] will be displayed
Press again, [ULOAD] will be displayed.
Wait for stability display, then press , an adjustment weight
will be displayed. The right digit flashes. Or confirm value with
, or enter a new value with the keys and confirm
by .
Place a corresponding adjustment weight, wait for stability display
and confirm by .
[--------] will be shortly displayed, followed by [F0 CAL]
The adjustment is now finished. Should an error occur, repeat
adjustment.
As the acceleration value due to gravity is not the same at every location on earth, each
display unit with connected weighing plate must be coordinated - in compliance with the
underlying physical weighing principle - to the existing acceleration due to gravity at its place
of location (only if the weighing system has not already been adjusted to the location in the
factory). This adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To receive
accurate measuring values it is also recommended to adjust the display unit periodically in
weighing operation.
Procedure:
MWB-SH-e-1810.docx 17
13 MAINTENENCE
13.1 General
If the scale does not operate properly, find out the problem as possible. Determine
whether the problem is constant or alternate. Be aware that problems can be caused by
mechanical or electrical influences.
Check the following.
Water
Corrosive materials
Vibrations or temperature or wind
Physical damage
Check the scale cables for damage, and check all connections and connecters for any
loose contact or incorrect connection
13.2 Cleaning
Disconnect the power before cleaning.
Use a cloth with mild suds and light cleaning agents.
Make sure that fluid not able to get into the device.
Use a clean and soft cloth for rub off.
13.3 Determine the Problem
Determine whether the problem is in the PCB or the Load Cell
Remove power from the system, and disconnect the load cell connection from
the PCB
Connect the PCB to a load cell simulator
Reapply power and test the PCB
If problem goes away, its source is probably in the Load cell. Check the
wiring, connecter, load cell and mechanical components of the load cell.
If problem persists, its source is probably in the PCB. Check the PCB voltages,
connecters, cables and function programs
13.4 Check the Load cell
Remove power from the system, and disconnect the PCB from the Load cell
Check the moisture, or foreign material inside.
Make sure all leads are connected and correctly.
Check load cell for proper input and output resistances
18 MWB-SH-e-1810.docx
13.5 Load Cell Connections
Measuring Points
Resistance
Red (+ Exc) to White ( –Exc)
420 ±20Ω
Green (+Sig) to Black ( –Sig)
350Ω ±5Ω
Signal + Green
Supply + Red
Signal - Black
Supply Shield
Whit
e
Load cell connection for the 4 load
cell’s.
RED --------- EXE +
WHITE------- SIG –
GREEN----- SIG +
BLACK------ EXE YELLOW--- SHIELD
Connection for the indicator
RED --------- EXE +
Empty------- SEN +
GREEN----- SIG +
BLACK------ SHIELD
BLUE-------- SIG –
Empty------- SEN –
YELLOW--- EXE –
MWB-SH-e-1810.docx 19
13.6 Check PCB Voltages
If the problem is in the PCB, use a multimeter to check the following voltages
13.7 PCB Input Voltage
Check the PCB input power connecter voltage
Voltage must be minimum 9VDC in to the pin AD+
13.8 Check Battery Voltage
Check the Battery Voltage
The battery voltage below the 1.5VDC, replace the battery and install new
1.5V battery.
20 MWB-SH-e-1810.docx
13.9 Replace Main Board
Release 4 pcs screw from bottom of the
scale by using cross screw driver.
Open the top cover, the main board lie on top
cover, use cross screw driver loose 4 screw
for main board, pull out all connector on the
main board, then you can bring out main
board carefully.
Install the new main board, plug all connector at last,After check anything,
try to turn on the power,if anything is OK,close the top cover,replace main
board completely.
13.10 Replace Battery
Open battery compartment cover
Remove the dry battery
Insert new batteries ,Lock the
battery cover again.
MWB-SH-e-1810.docx 21
14 Drawing
22 MWB-SH-e-1810.docx
15 Schematic
MWB-SH-e-1810.docx 23
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.