12.1 Adjustment key and seals .....................................................................................52
12.1.1 Menu overview in the service mode (adjustment switch in the adjustment position) .......... 54
12.2 Menu overview: .....................................................................................................56
MPS / MTS / MXS / MWS-BAPS-e-1710
4
1 Technical data
MPS
200K100NM/PNM
MTS
300K100NM
MXS
300K100NM
MPS
200K100M/PM
MTS
300K100M
MXS
300K100M
: 14 h;
: 14 h;
: 14 h;
KERN (Type)
Trademark
Display 6-position one
Weighing range (max.) 200 kg 300 kg 300kg
Display LCD with 25 mm high digits
Recommended calibration
weight, (class)
Signal rise time
(typical)
200 kg
(M1)
300 kg
(M1)
2–3 s
300 kg
(M1)
Warm-up time 10 min
Operating temperature +5°C …. +35°C
Storage temperature -20°C … +60°C
Air humidity max. 80% (non-condensing)
mains adapter 15 V / 300 mA (EN60601-1)
operation with 6 x 1.5 V battery supply,
Power supply
AA type batteries
operation time 50 h
Auto-Off function after 3 min without load change (possibility of setting)
Terminal (S x G x W) mm 210 x 110 x 50
Scale ready for operation
(W x D x H) mm
275 x 295 x 58
with column:
275 x 460 x 1010
550x550x1060 550x550x61
Scale plate mm
Total weight kg (net) 4.1 6.6
Rechargeable battery
operation (optional)
275 x 295 x 60 550x550 550x550
21.8 15.0
Loading time
operating time: 35 h;
7.2 V / 2000 mA
Loading time
operating time: 45h;
7.2 V / 2000 mA
Loading time
operating time: 50h;
7.2 V / 2000 mA
MPS / MTS / MXS / MWS-BAPS-e-1710
5
KERN (Type)
MWS
300K1LNM
MWS
300K100NM
MWS
400K100DNM
MWS
300K1LM
MWS
400K100DM
Trademark
MWS 300K100M
Display 6-position one
Weighing range (max.) 300 kg 300 kg 300kg; 400kg
Display LCD with 25 mm high digits
Recommended calibration
weight, (class)
Signal rise time
(typical)
300 kg
(M1)
300 kg
(M1)
2 – 3 sec.
400 kg
(M1)
Warm-up time 10 min; 10 min
Operating temperature + 5° C …. + 35° C
Storage temperature - 20°C … + 60°C
Air humidity max. 80 % (non-condensing)
mains adapter 15V / 300 mA ( EN60601-1)
Power supply
operation with 6 x 1.5 V battery supply,
AA type batteries
operation time 50 h
Auto-Off function after 3 min without load change (possibility of setting)
Terminal (B x T x H) mm
Scale ready for operation
(W x D x H) mm
1500x860x68 1155x830x65 1255x1060x69
210 x 110 x 45
Scale plate mm 800x1200 910x740 1000x1000
Total weight kg (net) 42 28,6 42.2
Rechargeable battery
operation (optional)
Loading time:14 h;
operating time: 45 h;
7.2 V / 2000 mA
Loading time:14 h;
operating time: 45h;
7.2 V / 2000 mA
Loading time:14 h;
operating time: 45h;
7.2 V / 2000 mA
MPS / MTS / MXS / MWS-BAPS-e-1710
6
KERN (Type)
MPS
200K100M/PM
MTS
300K100M
MXS
300K100M
: 14 h;
: 14 h;
: 14 h;
Display 6-position one
Weighing range (max.) 200 kg 300 kg 300kg
Display LCD with 25 mm high digits
Recommended calibration
weight, (class)
Signal rise time
(typical)
200 kg
(M1)
300 kg
(M1)
2–3 s
300 kg
(M1)
Warm-up time 10 min
Operating temperature +5°C …. +35°C
Storage temperature -20°C … +60°C
Air humidity max. 80% (non-condensing)
mains adapter 15 V / 300 mA (EN60601-1)
operation with 6 x 1.5 V battery supply,
Power supply
AA type batteries
operation time 50 h
Auto-Off function after 3 min without load change (possibility of setting)
Terminal (S x G x W) mm 210 x 110 x 50
Scale ready for operation
(W x D x H) mm
275 x 295 x 58
with column:
275 x 460 x 1010
550x550x1060 550x550x61
Scale plate mm
275 x 295 x 60 550x550 550x550
Total weight kg (net) 4.1 6.6
Rechargeable battery
operation (optional)
Loading time
operating time: 35 h;
7.2 V / 2000 mA
21.8 15.0
Loading time
operating time: 45h;
7.2 V / 2000 mA
Loading time
operating time: 50h;
7.2 V / 2000 mA
MPS / MTS / MXS / MWS-BAPS-e-1710
7
KERN (Type)
MWS
300K1LM
MWS
300K100M
MWS
400K100DM
Display 6-position one
Weighing range (max.) 300 kg 300 kg 300kg; 400kg
Display LCD with 25 mm high digits
Recommended calibration
weight, (class)
Signal rise time
(typical)
300 kg
(M1)
300 kg
(M1)
2 – 3 sec.
400 kg
(M1)
Warm-up time 10 min; 10 min
Operating temperature + 5° C …. + 35° C
Storage temperature - 20°C … + 60°C
Air humidity max. 80 % (non-condensing)
mains adapter 15V / 300 mA ( EN60601-1)
Power supply
operation with 6 x 1.5 V battery supply,
AA type batteries
operation time 50 h
Auto-Off function after 3 min without load change (possibility of setting)
Terminal (B x T x H) mm
Scale ready for operation
(W x D x H) mm
1500x860x68 1155x830x65 1255x1060x69
210 x 110 x 45
Scale plate mm 800x1200 910x740 1000x1000
Total weight kg (net) 42 28,6 42.2
Rechargeable battery
operation (optional)
Loading time:14 h;
operating time: 45 h;
7.2 V / 2000 mA
Loading time:14 h;
operating time: 45h;
7.2 V / 2000 mA
Loading time:14 h;
operating time: 45h;
7.2 V / 2000 mA
MPS / MTS / MXS / MWS-BAPS-e-1710
8
1.1 Tolerances altimeter
Measured value (cm)
90 ± 0.5
100 ± 1.0
150 ± 1.0
200 ± 1.0
Tolerance (cm)
MPS / MTS / MXS / MWS-BAPS-e-1710
9
2 Decl a ration of conformity
Only those people who can stand on the scales platform with their
To view the current EC/EU Declaration of Conformity go to:
www.kern-sohn.com/ce
3 Basic directions (general information)
3.1 Intended use
The scales are used for determining the body weight of people in the standing position.
The person being weighed shall stand in the middle of the weighing surface and remain calm.
The value may be read after the scales stabilizes.
The scales are designed for continuous use.
both legs may enter it.
The scales platform is equipped in the anti-slippery surface which shall not be taken
off or covered during the time the people are being weighed.
MPS / MTS / MXS / MWS-BAPS-e-1710
10
3.2 Inappropriate use
Do not use the scales for dynamic weighing.
Do not leave a permanent load on the weighing plate. This can damage the measuring equipment.
Be sure to avoid impact shock and overloading the scale in excess of the prescribed
maximum load rating (max.), minus any possible tare weight that is already present.
This could result in damage of the scale.
Never operate the scale in hazardous locations. The series design is not explosionproof. Attention should be paid that flammable mixture may also be for me d fro m
anaesthesiological means that contain oxygen or laughing gas (nitrous oxide).
Construction alterations may not be made to the scale. This can lead to incorrect
weighing results, faults concerning safety regulations as well as to destruction of the
scale.
The scale may only be used in compliance with the described guidelines. Other areas of application/planned use must be approved by KERN in writing.
Do not use the scales for determining body weight in medicine.
3.3 Guarantee
The guarantee shall become void in the event of the following:
• non-observation of our guidelines in the Operating Instructions,
• use outside the described applications,
• alteration to or opening the device,
• mechanical damage or damage caused by media, liquids,
• usual wear and tear,
• inappropriate erection or electric instal lation,
• overloading of the measuring equipment,
• scale falling down.
MPS / MTS / MXS / MWS-BAPS-e-1710
11
3.4 Monitoring the test substances
The metrology features of the balance and any possible available adjusting weight
must be checked at regular intervals within the scope of quality assurance. For this
purpose, the responsible user must define a suitable interval as well as the nature
and scope of this check. Information is available on KERN’s home page (www.kern-
sohn.com) with regard to the monitoring of balance test substances and the test
weights required for this. Test weights and balances can be adjusted quickly and at a
reasonable price at KERN’s accredited DKD (Deutscher Kalibrierdienst) calibration
laboratory (return to national standard).
In the case of the scales for weighing people provided with the scale to determine a
body size, it is recommended to carry out the check of its measuring accuracy because determination of the human body size is always connected with a very large
inaccuracy.
4 Basic safety directions
4.1 Observing the directions included in the Operating Instructions
Carefully read this oper ati on ma nual before
setup and commissioning, even if you are already familiar with KERN balances.
MPS / MTS / MXS / MWS-BAPS-e-1710
12
5 Transport and storage
Secure all parts such as the weighing platform, power unit etc.
5.1 Check upon delivery
Please check the packaging immediately upon delivery and the device during unpacking for any visible signs of external damage.
5.2 Packaging / return transport
Keep all parts of the original packaging for a possibly required
return.
Only use original packaging for returning.
Prior to dispatch disconnect all cables and remove
loose/mobile parts.
Reattach possibly supplied transport securing devices.
against shifting and damage.
MPS / MTS / MXS / MWS-BAPS-e-1710
13
6 Unpacking, installation and star ti ng
6.1 Place of installation, place of use
The scale is designed in such a way that reliable weighing results can be achieved
under normal application conditions.
By selecting the correct location for your scale, you will be able to work quickly and
precisely.
Therefore, please observe the following when choosing a place of installation:
- Place the scale on a firm, level surface;
- Avoid extreme heat as well as temperature fluctuation caused by installing the
scale next to a radiator or in the direct sunlight;
- Protect the scale against direct draughts due to open windows and doors;
- Avoid shaking during weighing;
- Protect the scale against high humidity, vapours and dust;
- Do not expose the device to extreme dampness for longer periods of time. In-
admissible bedewing (condensation of air moisture on the device) can occur if
a cold device is taken into a significantly warmer environment. In this case,
please keep the device for approx. 2 ho ur s at room temp er a ture after it has
been disconnected fro m mains s uppl y ;
- Avoid static charge build-up on the scale and people to be weighed;
- Avoid contact with water.
Major dis play deviati on s (i ncorrect weighing results ) are possibl e if electrom ag n e t ic
fields occur (e.g. coming from mobile phones or radio equipment) as well as due to
static charging and instable power supply. It is necessary then to change the scale
locatio n or remove disturbance source.
6.2 Unpacking
Carefull y r emove indivi d ua l scale parts or the whole scal e fr om its packagi ng and position the scale in its intended working location. When the mains adapter is used , be
careful not to cause the danger of falling over the power cable.
MPS / MTS / MXS / MWS-BAPS-e-1710
14
6.3 Installation and setting of scale
Personal weighing scale MPS with wall bracket:
Scope of delivery:
MPS / MTS / MXS / MWS-BAPS-e-1710
15
Personal weighing scale MPS-PM with column:
Scope of delivery :
• Balance with display unit and tripod
• Mains adapter
• 4 screws
MPS / MTS / MXS / MWS-BAPS-e-1710
16
Assembly:
Remove cap (1)
Pull cable with Plug-in con-
Place supporting foot next to
Unscrew the screw (2)
nection (3) through the supporting foot (4) and pull it out
at the end (5)
the balance
MPS / MTS / MXS / MWS-BAPS-e-1710
17
Introduce cable completely
Reinsert cap (1)
Use the 4 screws to attach
into the tripod tube (6)
Screw-in screw (2)
When screwing the screw ensure that the plug-in connector inside the tripod foot is not
squeezed.
the tripod to the bottom of
the balance
Level balance with foot screws until the air
bubble of the water balance is in the prescribed circle.
Check levelling regularly.
MPS / MTS / MXS / MWS-BAPS-e-1710
18
Adjust foot screw of the tripod in a
way that the tripod has a safe base
and is not loose.
MPS / MTS / MXS / MWS-BAPS-e-1710
19
Mounting height measuring rod MSF 200:
How to attach to
KERN scale
Use the 2 screws for screwing the bracket into the supplied
insert nuts on the stand of the scales.
Extend the height measuring rod and tighten it in the lower
hole of the bracket with the help of the appertaining screw.
The height measuring rod can be mounted in the same way at
the back of the support stand.
MPS / MTS / MXS / MWS-BAPS-e-1710
20
Scale MTS with handrail:
Scope of delivery:
MPS / MTS / MXS / MWS-BAPS-e-1710
21
Assembly:
Fasten 3 corner elements to the platform, using 4 screws each time.
Place the handrail on 3 corner elements and screw it.
Fasten the terminal holder to the handrail with 3 screws.
Remove the side rubber plugs at both sides of the display.
Fasten the display to the holder with both handwheels.
Adjust the display position with handwheels.
MPS / MTS / MXS / MWS-BAPS-e-1710
22
Bariatric scale MXS:
Scope of delivery:
MPS / MTS / MXS / MWS-BAPS-e-1710
23
Wheelchair scale MWS and Transportation stretcher balance MWS-L:
MWS
MWS-L
Scope of delivery:
MPS / MTS / MXS / MWS-BAPS-e-1710
24
Direction concerning installation of external column on MPS model without
column, MXS and MWS models
Fasten the round plate to the aluminium profile with screws.
Fasten the wall bracket to the top of aluminium profile top with screws.
Remove the side rubber plugs at both sides of the display.
Fasten the display to the bracket with both handwheels.
Adjust the display position with handwheels.
Fasten the cable with cable clips.
MPS / MTS / MXS / MWS-BAPS-e-1710
25
Assembly of holding bracket set MWS-A02 at the MWS models
1
Railing
2
Handrail butts
3
Cross strut
6
Screw
7
Hexagonal socket
wrench
8
Screw
(for installation cross strut)
4
Support
5
Tapped bush
9
Screw
(for installation display
unit)
MPS / MTS / MXS / MWS-BAPS-e-1710
26
We recommend engaging a second person to assist you during installation.
Carefully remove the plastic cover from
the weighing scale, ensuring that the
weighing scale does not get scratched in
the process.
Mount all 4 handrail butts (2) on the
weighing scale frame.
Ensure that the handrail butt and cable conduit are positioned to the right of
the power unit supply point. (s. fig.)
Take the two hexagonal socket wrenches (7) and fix all handrail butts with the
help of the screws 6 (3x) and the tapped
bushes 5 (2x) as shown on the illustration.
Make sure that all screws are tightened properly.
MPS / MTS / MXS / MWS-BAPS-e-1710
27
Ensuring perfect fit, put the handrail containing three holes for the display unit on
the handrail butt (1) and cable conduit.
(see fig.)
Using the holders 4 (2x) fasten the hand-
rail to the handrail butts.
Once again, use screws 6 (3x) and
tapped bushes 5 (3x).
Carry out same sequence of operations
for second handrail.
Use the two screws (8) to fasten the
cross strut (3).
Use the three screw s to fasten the
mounting plate to the handrail.
MPS / MTS / MXS / MWS-BAPS-e-1710
28
Take a screw driver and remove the
The reading direction of the display unit is optional and can be adapted during
the installation of the application.
plastic covers from both sides of the display unit.
Screw the display unit to the handle, using the synthetic screws included in the
delivery,
Display unit facing inward
Display unit facing outward
MPS / MTS / MXS / MWS-BAPS-e-1710
29
Using the synthetic covers, seal both
holes in the two handrail butts without
cross struts.
Once installation is complete, check all screws for tight fit. Otherwise the
person to be weighed may suffer an i njur y.
General direction concerning setting up the previously mentioned scales
Place a personal weighing scale in the intended location and level it with the adjustable rubber feet until the air bubble in the spirit level (located in the centre of the scale
plate) is in the centre.
When scales with large and heavy platforms are installed and transported (a scale
plate folded upwards), take care not to drop a scale as this could cause its damage.
MPS / MTS / MXS / MWS-BAPS-e-1710
30
6.3.1 Scope of delivery
Standard accessories:
• Mains adapter
• Operating Instructions
6.3.2 Installation direction for a model with wall bracket
The display unit of the MWS has two magnets on the rear side, use these magnets to
fix the display unit on metalli c surfaces.
6.4.1 Transportation of balance
There exists the possibility to fix the display unit at the platform using the two magnets on the rear side, whereby the balance and the display unit can be transported
together without problems (see fig. below).
MPS / MTS / MXS / MWS-BAPS-e-1710
32
Connection CN 3 for
Connection CN 4 for
6.5 Mains socket
Power supply is carried out by means of the external mains adapter which also provides separation between mains and a scale. The printed voltage value must be
compliant with local voltage.
It is allowed to use only admitted, original power supply adaptors of KERN company.
6.6 Battery operation/ Rechargeable battery operation (optional)
(only models with battery operation and rechargeable battery operation)
MPS / MTS / MXS / MWS-BAPS-e-1710
batteries (AA x 6)
rechargeable battery
33
6.6.1 Battery operation
off the battery cover on the lower
On models where the back of the display unit is not directly accessible, remove the
two black rotary knobs from both sides of the display unit in order to open the battery
compartment and remove the display unit from the holder.
Lift-
side of the balance
Carefully take out the battery holder
(1)
Insert 6 batteries (AA).
Ensure that the batteries are inserted in the correct direction
MPS / MTS / MXS / MWS-BAPS-e-1710
34
Insert battery holder with the inserted
batteries into the display unit
Ensure that the cables are not
squeezed
Close the battery cover
If the batteries are run down, "LO" appears in the display. To turn
ON
off, press the button and immediately change the batteries.
OFF
If the balance is not used for a longer time, take out the batteries
and store them separately. Leaking battery liquid could damage
the balance.
MPS / MTS / MXS / MWS-BAPS-e-1710
35
6.6.2 Rechargeable battery operation (optional)
n the lower
1
When an optional rechargeable battery is used, proceed as follows:
On models where the back of the display unit is not directly accessible, remove the
two black rotary knobs from both sides of the display unit in order to open the battery
compartment and remove the display unit from the holder.
Lift-off the battery cover o
side of the balance
Carefully take out the battery holder
(1)
Carefully pull-out plug (2) from the
connection CN 4 (3)
MPS / MTS / MXS / MWS-BAPS-e-1710
36
Carefully insert the rechargeable bat-
tery block and insert plug into connection CN 3
Ensure that the cables are not
squeezed
Close the battery cover
If the rechargeable battery is exhausted, „LO“ is displayed. The
rechargeable battery is loaded via the provided plug-in power
supply unit (loading time 14 h for a complete loading).
If the balance is not used for a longer time, take out the rechargeable battery and store it separately. Leaking liquid could damage
the balance.
MPS / MTS / MXS / MWS-BAPS-e-1710
37
6.7 Initial start-up
Function
Settings
Description
SEtuP
A. oFF
Automatic turning
180 s
Automatic turning of f after 3 m inutes
240 s
Automatic turning of f after 4 m inutes
300 s
Automatic turning of f after 5 m inutes
oFF
Automatic turning off turned off
120 s
Automatic turning of f after 2 m inutes
burr
on
Acoustic signal turned on
oFF
Acoustic signal turned off
End
Leaving menu after pressing button
To obtain accurate weighing results with electronic scales, the appropriate operating
temperature is to be provided for them (refer to „Warm-up time”, section 1).
During warm-up the scale must be connected to power supply and switched on
(mains socket or batteries).
The accuracy of the scale depends on the local acceleration of gravity.
The value of acceleration of gravity is given on the rating plate.
6.8 Menu overview
When the scale is switched on, hold the [→0←] key pressed for about 3 seconds
until the display shows successively the „SETUP” and „A.OFF” symbols.
Selection is carried out with the [TARE] and [HOLD] keys.
off, Auto Off function
Acoustic signal
HOLD
MPS / MTS / MXS / MWS-BAPS-e-1710
38
7 Operation
7.1 Display
Type MPS 200K100M
Type MPS 200K100PM
Type MPS 200K100NM
Type MPS 200K100PNM
MPS / MTS / MXS / MWS-BAPS-e-1710
39
Scale zero display: If the scale does not show exactly
zeroed again.
Stabilisation display:
unstable condition, the stabilisation display [ο] dis appears.
It is illuminated when mains supply is via the mains
adaptor.
7.2 Display view
No. Display Description
1
2 [ο]
3
4 BMI
5 HOLD
6 PRE-TARE
7 NET
[→0←]
zero value although the scale pan is unloaded, press the
[→0←] key. After a short waiting time, the scale will be
If the display shows the stabilisation display [ο], the
scale is in the stable condition. When the scale is in the
Calculated value of the body mass index (BMI).
Hold function / saving function is active.
Initially set tare value is active.
The net weight is displayed.
8 WEIGHT
The present weight value is displayed.
MPS / MTS / MXS / MWS-BAPS-e-1710
40
7.3 Overview of keyboard
kg display.
Key Description
ON/OFF
PRINT
BMI
HOLD
0
M 1-5
PRE-TARE
TARE
CLEAR
0..9
Switching on/off the scale
Data transmission via interface
Determination of Body Mass Index
Hold function / determination of stable weighing value
The scale is reset to 0,0
It is possible to set max. up to 2% of maximal load for
verified scales, and 2% or 100% of maximal load for
common scales (possibility of selection in the menu)
Memories 1–5 were called
Calling the tare function with set values
Taring the scale
Clearing the digits entered manually
Entering digits
ENTER
Using the entered digits
MPS / MTS / MXS / MWS-BAPS-e-1710
41
8 Using s c a le
8.1 Weighing
Switch on the scale with the [ON/OFF] key. The diagnostic scale self-check is
performed and then the software version is displayed. The scale is ready for
weighing when the „0,00 kg” weight display is shown.
Direction: The [→0←] key makes it possible to zero the scale if necessary and
at any time.
Place a person in the middle of the scale. Wait until the stability display (ο) is
shown and then read the weighing result.
Direction:
If a person is heavier than the weighing range, the display will show the „Err” symbol
(= overload).
8.1.1 Weighing with MWS
Due to the great dimensions and the big weighing range especially suitable for
weighing of immobile people on transportation stretchers, wheelchairs or overweight
people in the adiposity range
8.1.1.1 Weighing with transportation stretcher or wheelchair
Place transportation stretcher/wheelchair with the person on the centre of the
scales
Arrest the brakes of the transportation stretcher/of the wheelchair
Do not leave the person unattended
When the person is lying/sitting quietly, read weighing value 1
Loosen the brakes and carefully pull off the transportation stretcher/wheelchair
with the person
After that weigh the transportation stretcher/wheelchair without the person and
subduct this weight from weighing value 1, from there results the person’s
weight.
MPS / MTS / MXS / MWS-BAPS-e-1710
42
8.2 Taring
The dead weight of any initial load used for weighing may be tared away by pressing
the key, so that the following weighing shows the real weight of a person to be
weighed.
E.g. when a rubber mat is put on the scale plate, the scale does not
show 0 value.
To start the taring process, press the [TARE] key. Now internal weight
saving is performed and value of 0.0 kg is displayed.
Place a person in the middle of the scale plate.
Then read the weight on the display.
Direction:
The scale can store only one tare value.
When the scale is unloaded, the saved tare value is displayed with „negative” sign.
To delete the saved tare value, unload the scale plate and then press the [TARE]
key.
8.3 HOLD function
The scale is provided with the integrated hold function (determination of average value). It enables people to be weighed accurately although they are not still on the
scale plate.
Note: Determination of average value is not possible when a person moves too
much.
Switch on the scale with the [ON/OFF] key. The diagnostic scale self-check is
performed. The scale is ready for weighing when the „0.00 kg” weight display
is shown.
Place a person in the centre of the scale plate.
Press the [HOLD] key. When the triangle is flashing on the display, the scale
takes some measuring values and then the calculated average value is dis-
played.
Press the [HOLD] key again to return the scale to the normal weighing mode.
Pressing the [HOLD] key makes it possible to activate the function at any
time.
MPS / MTS / MXS / MWS-BAPS-e-1710
43
8.4 Mother/Child Function
The “Mother/Child” function makes it possible to determine the body weight of infants
and small children held in an adult’s arms.
Turn on the scales using the [ON/OFF] switch. The scales enters the self-
diagnostic mode. The device is ready for operation as soon as “0.00 kg” is
shown on the weight display.
Position the adult person in the middle of the scales platform. After showing
the stability indicator, the scales displays the person’s weight. A triangle s ym-
bol is shown under the “WEIGHT” symbol.
Press the [TARE] button and the readout changes to “0.00 kg”.
Place the child in the arms of the adult person. After showing the stability indi-
cator, the scales displays the child’s weight. The triangle symbol is now dis-
played under the “NET” symbol.
Press the [TARE] button again and the readout again changes to “0.00 kg”.
After stepping off the scales, the total weight of the adult person and the child
is displayed as a negative value.
Press the [TARE] button again. The saved tare value is cancelled and the
scales can be used for weighing again.
8.5 Determination of Body Mass Index
When you obtain a stable weight and display shows 0.0 kg, place a person in the
middle of the scale plate. Wait until the weighing value is stable. Then press the BMI
key. Now enter a body height.
Please take note that reliable determination of BMI index is only possible for body
height from 100 cm to 250 cm and weight > 10 kg.
A body height entered as the last one is flashing on the display. Now you can enter a
different value with the numerical keypad. Confirm the entered value with the
ENTER key, and then a person’s BMI index will be displayed.
When the BMI index value is displayed, it is presented with the arrow pointing the
BMI symbol. To return to the weighing mode, press the BMI key once again and the
arrow at the BMI symbol will disappear.
MPS / MTS / MXS / MWS-BAPS-e-1710
44
8.5.1 Classification of BMI values
overweight
Classification of weight for adults over 18 years on the basis of Body Mass Index
according to WHO, 2000 EK IV and WHO 2004 (WHO - World Health Organization).
Category BMI (kg/m2) Risk of diseases accompanying
Underweight < 18.5 low
Normal weight 18.5 – 24.9 average
Overweight
Preobesity
I degree of obesity
II degree of obesity
III degree of obesity
8.6 PRE-TARE function
When a tare weight (rubber mat, clothes, ...) is known, this value can be entered
manually.
If the PRE-TARE key is pressed shortly, the flashing display will be shown.
The PRE-TARE function is active as long as the small arrow is pointing the ”PRETARE” symbol on the display.
The value used as the last one will be displayed. If a different value is required, a
new weight value can be entered with the numerical keypad. By pressing the ENTER
key, the new value is confirmed and used. Then the entered value with minus sign
will be shown on the display.
When a person is placed on the scale plate, the display will show a weight value less
the value entered previ ousl y.
Pressing the PRE-TARE key again will return the scale to the normal weighing mode.
> 25.0
25.0 – 29.9
30.0 – 34.9
35.0 – 39.9
> 40
slightly increased
increased
high
very hig h
MPS / MTS / MXS / MWS-BAPS-e-1710
45
8.6.1 PRE-TARE function with 5 memories
Owing to this function it is possible to store 5 Pre-Tare values (e.g. for different
wheelchairs), and then call up them if necessary.
Saving PRE-Tare values:
To enable a later calling up values from the memory, they are to be entered into the
memory first. It is carried out in the following way:
The scale plate is unloaded, and the scale display is showing 0,0 kg.
Put a weight, whose value is to be saved (e.g. an empty wheelchair), on the scale
plate and wait until a stable weight display is shown.
Press the M1-5 key repeatedly until the display will show the „ni” (M) symbol.
Press the key with digit (1..5) shortly to indicate which number a value is to be
saved under. The previously displayed weight value is flashing for 3 seconds.
When the flashing is finished, press again the key with digit previously pressed and
the weighing value is saved in the memory (short audible signal).
By pressing the CLEAR key, the scale will return to the weighing mode without
saving the value.
The display will show the real value of the weight placed on the scale plate. When
the weight is removed, the display will show 0,0 kg.
Calling up the PRE-Tare value from memory:
Press the PRE-Tare key repeatedly until the display will show the „ni” (M) symbol.
Pressing the key with digit (1..5) will display the flashing weight value saved there.
The small arrow, additionally shown on the display, is pointing to the „PRE-TARE”
symbol. By pressing another key with digit (1..5), the appropriate also flashing
weight value will be displayed. Press the ENTER key and the value will be accepted
and shown on the display as the PRE-Tare value with minus sign.
Now you can place e.g. a person in a wheelchair or on the transportation stretcher on
the scale, and only a person‘s weight will be displayed.
To return to the normal weighing mode when the scale plate is unloaded, press the
PRE-Tare key shortly again. This will also result in disappearing the small arrow
pointing the „PRE-TARE” symbol.
MPS / MTS / MXS / MWS-BAPS-e-1710
46
Printing Pre-Tare memory (refer also to section 8.6):
G
88.8
kg
Gross weight
T
2.0
kg
Taring weight
N
86.8
kg
Net weight
180.0
cm
Size of the person
24.4
BMI
BMI value
To do this, press the PRE-Tare key repeatedly until the display will show the „ni” (M)
symbol.
Short pressing the PRINT key will activate printing of the values saved in 5 memories.
M1 0.0 kg
M2 7.0 kg
M3 10.0 kg
M4 30.0 kg
M5 50.0 kg
8.7 Printing function
To use this function, you need the RS232 interface cable (available as accessories)
which is connected with the round plug at the terminal back.
Note: In medical applications, only the peripheral equipment meeting EN 60601-1
standard can be connected to the interface.
When a scale is in the weighing mode, pressing the PRINT key will result in output of
the particular data, presented below, via the interface. It is the standard way to output
data, which cannot be changed.
8.7.1 Parameters of RS232 interface
Set parameters of the scal e int er face on t he connected device. It is not possible to
change the scale parameters.
Baud rate: 9600 bps
Parity check: no
Data length: 8 bits
Stop bit: 1 bit
Handshake: no or Xon/Xoff
Data code: ASCII
MPS / MTS / MXS / MWS-BAPS-e-1710
47
9 Error messages
Remove dirt immediately.
The following messages can be shown on the display during switching on or using
the scale.
ERRL: Too small weight on the scale.
ooooo:
The scale plate was loaded during switching on the scale. Unload the scale
plate.
ERR: Overload, too large weight on the scale plate.
10 Service, maintenance, disposal
10.1 Cleaning
Disconnect the unit from the mains power supply prior to cleaning,
maintenance and repair activities.
10.2 Cleaning / Disinfecting
Clean weighing platform (such as seat) as well as casing with household detergents
or commercially available disinfectants, e.g. 70% isopropanol. We recommend a disinfectant suitable for wiping disinfection. Please follow manufacturer’s instructions.
Do not use abrasive or aggressive cleaners such as spirits or alcohol or similar as
they might damage the high-quality surface.
To prevent cross-contamination (fungal skin infection) please observe the following
time intervals for disinfection:
• Weighing plate before and after any measurement with direct skin contact
• When required:
o Display
o Foil-covered keyboard
Do not spray disinfectants onto appliance.
Make sure that disinfectant does not penetrate the interior of
the balance.
10.3 Sterilisation
Sterilisation of the appl i ance is not all owed.
10.4 Service, maintenance
The device may only be operated and maintained by trained service technicians who
are authorised by KERN.
Disconnect the scale from mains supply before its opening.
MPS / MTS / MXS / MWS-BAPS-e-1710
48
10.5 Disposal
Disposal of packaging and device must be carried out by an operator according to
valid national or regional law of the location where the device is used.
MPS / MTS / MXS / MWS-BAPS-e-1710
49
11 Troubleshooting
The scale should be switched off for a short time following an interruption in the program sequence and disconn ected from mains supply. It is then necessary to repeat
the weighing process from the beginning.
Interruption Possible cause
Weight display is not illuminated.
Weight display changes
continuously
• The scale is not switched on.
• The mains supply connection has been interrupted
(mains cable not plugged in/ faul ty).
• Check the fus e o f the mains ad apter / green LED
is illuminated next to the fuse.
• Mains fail ur e.
• Batteries are incorrectly inserted or discharged.
• No batteries.
• Draught/air movement
• Table/floor vibrations
• The weighing plate is in contact with foreign
matters or is installed incorrectly.
• Electromagnetic fields/static charging (choose a
different location for the scale, switch off an interfering device if possible).
The weighing result is
obviously incorrect
Should other error messages occur, switch the scale off and then on again. If the error message remains, inform the manufacturer.
MPS / MTS / MXS / MWS-BAPS-e-1710
• The scale display is not set to zero
• Incorr ect adj us t me nt.
• Great fluctuations in temperature.
• Warm-up time was ignored.
• Electr omagnetic fields/static charging (choose a
different location for the scale, switch off an interfering device if possible).
50
Operation
Display
SETUP
��
UNKT
��
CAL
��
K
B
CAL U
��
of the display. If not, press the [TARE] key.
CAL U
��
CAL O ��
30770
CAL O
��
12 Adjusting
Observe stable environmental conditions. The warm-up time (refer to chapter 1) is
required to ensure the scale stabilisation.
Note:
In the case of verified scales, adjusting is locked with the jumper. To carry out adjusting, the jumper is to be set in the adjusting position (central position). (refer to 12.2).
Turn the scale on using the [ON/OFF] key.
Press and keep the [→0←] key pressed for about 3
seconds until the display shows successively the
„SETUP” and „UNIT” symbol.
Press the [TARE] key until the „CAL ib” symbol is
displayed.
Press the [HOLD] key.
Press the [TARE] key.
The triangle must be displayed at the top right side
Press the [HOLD] key repeatedly until the „CAL 0”
symbol is displayed.
Press the [TARE] key, the display will show the
present numerical value.
Then press the [ENTER] key.
MPS / MTS / MXS / MWS-BAPS-e-1710
51
Press the [HOLD] key.
CAL 5
��
ue with the [TARE] key.
200.0
cess is started.
82077
Press the [TARE] key.
Enter the required calibration weight value (refer to
chapter 1, „Technical data”): Select the item to be
changed with the HOLD key and set its numerical val-
Confirm by pressing the [ENTER] key.
Place the calibration weight carefully in the centre of
the scale plate, and the display will show a numerical
value. Press the [ENTER] key. The adjustment pro-
When the adjustment is finished successfully, the
scale is automatically switched over to the weighing
mode again and the calibration weight value will be
displayed.
Remove the calibration weight.
MPS / MTS / MXS / MWS-BAPS-e-1710
52
12.1 Adjustment key and seals
When verification of a scale is finished, the positions indicated on the scale are
sealed.
Verification of a scale without a seal is invalid.
Position of seals:
1. Backside 2. Battery compartment
MPS / MTS / MXS / MWS-BAPS-e-1710
53
3. MPS
4. MXS, MTS
5. MWS
MPS / MTS / MXS / MWS-BAPS-e-1710
54
Position of adjustment switch:
Position of adjustment switch
Status
A.OFF
BURR
1 2 3
1. left side not documented
2. centre
adjustment position - adjustment is
possible
3. right s ide verificati on pos it ion – adjustment locking
12.1.1 Menu overview in the service mode (adjustment switch in the adjustment
position)
Overview is only used to check the set parameters by authorised verification offices.
Changes may only be introduced in parameters of the automatic switching off
function „
” and audible signal „
”.
MPS / MTS / MXS / MWS-BAPS-e-1710
55
Navigation in menu:
END
When the scale is switched on, press and hold the [→0←] key press ed for
about 3 seconds until the display shows successively the „SETUP” and
„UNIT” symbol.
Press the [TARE] key repeatedly until the required function is displayed.
Confirm the selected function with the [HOLD] key. The first parameter will be
displayed. Select the required parameter with the [HOLD] key and confirm the
selection with the [TARE] key.
To exit the menu and save the settings, press the [TARE] key until the „
symbol is displayed and then confirm with the [HOLD] key. The scale is automatically
returned to the weighing mode.
Selection is carried out with the [HOLD] and [TARE] key.
”
MPS / MTS / MXS / MWS-BAPS-e-1710
56
12.2 Menu o vervi ew:
Function
Settings
Description
SEtuP
Unit
on-off
Weight unit: kg
Grad
3000 d – 6000 d –
2500 d - 2000 d
The values of the scale elementary division, weigh-
Ut.-d
Full – S-Ut
Selection: one-range scales (Full)- / multi-range
FillE
Fast – Nor. - SLo
Filter: fast - normal - slow
Auto 0
0.25 d – 0.5 d – 1
d – 3 d - OFF
Auto-Zero-Tracking
Stab
0.25 d – 0.5 d –
1 d – 3 d - off
Stabilisierungsbereich
Orang
2 Pct – 100 Pct.
Nullbereich: 2 % / 100 %
Ould
9 d – 2 Pct.
Überlastbereich: 9 d / 2 %
CALib
CAL-U – CAL–0
CAL-5
Justierung
A.Off
120s/180s/240s/
300s/off
Automatische Abschaltfunktion
burr
on/off
Acoustic signal
default
Resetting to factory settings
End
Leaving the menu after pressing button
10000 d – 500 d –
1000 d – 1500 d -
ing range (max) and elementary division (d)
scales (S-Ut)
MPS / MTS / MXS / MWS-BAPS-e-1710
HOLD
57
Description:
Weight unit: kg
Scale divisions, weighing range (max.) and read-out (d)
Selection of multi-range / single-range scale
Single-range scale
Multi-range scale
Filter: fast / normal / slow
Automatic zero tracking: 0,25 d/ 0,5 d/ 1 d/ 3 d/ OFF
Stabilisation range: 0,25 d/ 0,5 d/ 1 d/ 3 d/ OFF
Zero range: 2% / 100%
Overload range: 9 d / 2%
Adjusting
Auto off function: 120 s / 180 s / 240 s / 300 sec/ OFF
Audible signal: ON/OFF
Restoring the factory settings (default settings)
Exiting the menu
MPS / MTS / MXS / MWS-BAPS-e-1710
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.