KERN MPS 200K100NM, MPS 200K100PM, MTS 300K100M, MPS 200K100PNM, MTS 300K100NM Operating Instructions Manual

...
KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
E-Mail: info@kern-sohn.com
Internet: www.kern-sohn.com
Operating Instructions
Wheelchair scale, Transportation stretcher balance
KERN
MPS / MTS / MXS / MWS
Type MPS 200 K1 00M
Type MPS 200 K1 00NM
Type MPS 200 K1 00PM
Type MPS 200 K1 00PNM
Type MTS 300K100M
Type MTS 300K100NM
Type MXS 300 K1 00M
Type MXS 300 K1 00NM
Type MWS 300K100M
Type MWS 300K100NM
Type MWS 400K100DM
Type MWS 400K100DNM
Type MWS 300K1LM
Type MWS 300K1LNM
Version 1.0
Fax: +49-[0]7433-9933-149
Personal weighing scale, Handrail scale, Bariatric scale,
2017-02 GB
MPS / MTS / MXS / MWS-BAPS-e-1710
2
GB
KERN MPS/MTS/MXS/MWS
Transportation stret c he r balance
Version 1.0 2017-02
Operating Instructions Personal weighing sc a le without column / with column, Handrail scal e, Bar iatric scale, Wheelchair scale/
Contents
1 Technical data ............................................................................................ 4
1.1 Tolerances altimeter .............................................................................................. 8
2 Declaration of conformity .......................................................................... 9
3 Basic directions (general information) ..................................................... 9
3.1 Intended use .......................................................................................................... 9
3.2 Inappropriate use ..................................................................................................10
3.3 Guarantee .............................................................................................................10
3.4 Monitoring the test substances ..............................................................................11
4 Basic safety directio ns ............................................................................ 11
4.1 Observing the directions included in the Operating Instructions ............................11
5 Transport and storage ............................................................................. 12
5.1 Check upon delivery .............................................................................................12
5.2 Packaging / return transport ..................................................................................12
6 Unpacking, installation and starting ....................................................... 13
6.1 Place of installation, place of use ..........................................................................13
6.2 Unpacking .............................................................................................................13
6.3 Installation and setting of scale .............................................................................14
6.3.1 Scope of delivery .................................................................................................................. 30
6.3.2 Installation direction for a model with wall bracket ............................................................... 30
6.4 Magnets display unit MWS ....................................................................................31
6.4.1 Transportation of balance .................................................................................................... 31
6.5 Mains socket .........................................................................................................32
6.6 Battery operation/ Rechargeable battery operation (optional)................................32
6.6.1 Battery operation .................................................................................................................. 33
6.6.2 Rechargeable battery operation (optional) ........................................................................... 35
6.7 Initial start-up ........................................................................................................37
6.8 Menu overview ......................................................................................................37
7 Operation .................................................................................................. 38
7.1 Display ..................................................................................................................38
7.2 Display view ..........................................................................................................39
7.3 Overview of keyboard ...........................................................................................40
MPS / MTS / MXS / MWS-BAPS-e-1710
3
8 Using scale ............................................................................................... 41
8.1 Weighing ...............................................................................................................41
8.1.1 Weighing with MWS ............................................................................................................. 41
8.2 Taring ...................................................................................................................42
8.3 HOLD function ......................................................................................................42
8.4 Mother/Child Function ...........................................................................................43
8.5 Determination of Body Mass Index........................................................................43
8.5.1 Classification of BMI values ................................................................................................. 44
8.6 PRE-TARE function ..............................................................................................44
8.6.1 PRE-TARE function with 5 memories .................................................................................. 45
8.7 Printing function ....................................................................................................46
8.7.1 Parameters of RS232 interface ............................................................................................ 46
9 Error messages ........................................................................................ 47
10 Service, maintenance, disposal .............................................................. 47
10.1 Cleaning ...............................................................................................................47
10.2 Cle a nin g / Disinfecting ..........................................................................................47
10.3 Sterilisation ...........................................................................................................47
10.4 Ser vice, m aint enance ...........................................................................................47
10.5
Disposal ................................................................................................................48
11 Troubleshooting ....................................................................................... 49
12 Adjusting ................................................................................................... 50
12.1 Adjustment key and seals .....................................................................................52
12.1.1 Menu overview in the service mode (adjustment switch in the adjustment position) .......... 54
12.2 Menu overview: .....................................................................................................56
MPS / MTS / MXS / MWS-BAPS-e-1710
4

1 Technical data

MPS
200K100NM/PNM
MTS
300K100NM
MXS
300K100NM
MPS
200K100M/PM
MTS
300K100M
MXS
300K100M
: 14 h;
: 14 h;
: 14 h;
KERN (Type)
Trademark Display 6-position one
Weighing range (max.) 200 kg 300 kg 300kg Display LCD with 25 mm high digits Recommended calibration
weight, (class) Signal rise time
(typical)
200 kg
(M1)
300 kg
(M1)
2–3 s
300 kg
(M1)
Warm-up time 10 min Operating temperature +5°C …. +35°C Storage temperature -20°C … +60°C Air humidity max. 80% (non-condensing)
mains adapter 15 V / 300 mA (EN60601-1)
operation with 6 x 1.5 V battery supply,
Power supply
AA type batteries
operation time 50 h
Auto-Off function after 3 min without load change (possibility of setting) Terminal (S x G x W) mm 210 x 110 x 50
Scale ready for operation (W x D x H) mm
275 x 295 x 58
with column:
275 x 460 x 1010
550x550x1060 550x550x61
Scale plate mm
Total weight kg (net) 4.1 6.6 Rechargeable battery
operation (optional)
275 x 295 x 60 550x550 550x550
21.8 15.0
Loading time
operating time: 35 h;
7.2 V / 2000 mA
Loading time
operating time: 45h;
7.2 V / 2000 mA
Loading time
operating time: 50h;
7.2 V / 2000 mA
MPS / MTS / MXS / MWS-BAPS-e-1710
5
KERN (Type)
MWS
300K1LNM
MWS
300K100NM
MWS
400K100DNM
MWS
300K1LM
MWS
400K100DM
Trademark
MWS 300K100M
Display 6-position one Weighing range (max.) 300 kg 300 kg 300kg; 400kg Display LCD with 25 mm high digits Recommended calibration
weight, (class) Signal rise time
(typical)
300 kg
(M1)
300 kg
(M1)
2 – 3 sec.
400 kg
(M1)
Warm-up time 10 min; 10 min Operating temperature + 5° C …. + 35° C Storage temperature - 20°C … + 60°C Air humidity max. 80 % (non-condensing)
mains adapter 15V / 300 mA ( EN60601-1)
Power supply
operation with 6 x 1.5 V battery supply,
AA type batteries
operation time 50 h
Auto-Off function after 3 min without load change (possibility of setting) Terminal (B x T x H) mm
Scale ready for operation (W x D x H) mm
1500x860x68 1155x830x65 1255x1060x69
210 x 110 x 45
Scale plate mm 800x1200 910x740 1000x1000 Total weight kg (net) 42 28,6 42.2
Rechargeable battery operation (optional)
Loading time:14 h;
operating time: 45 h;
7.2 V / 2000 mA
Loading time:14 h;
operating time: 45h;
7.2 V / 2000 mA
Loading time:14 h;
operating time: 45h;
7.2 V / 2000 mA
MPS / MTS / MXS / MWS-BAPS-e-1710
6
KERN (Type)
MPS
200K100M/PM
MTS
300K100M
MXS
300K100M
: 14 h;
: 14 h;
: 14 h;
Display 6-position one Weighing range (max.) 200 kg 300 kg 300kg Display LCD with 25 mm high digits Recommended calibration
weight, (class) Signal rise time
(typical)
200 kg
(M1)
300 kg
(M1)
2–3 s
300 kg
(M1)
Warm-up time 10 min Operating temperature +5°C …. +35°C Storage temperature -20°C … +60°C Air humidity max. 80% (non-condensing)
mains adapter 15 V / 300 mA (EN60601-1)
operation with 6 x 1.5 V battery supply,
Power supply
AA type batteries
operation time 50 h
Auto-Off function after 3 min without load change (possibility of setting) Terminal (S x G x W) mm 210 x 110 x 50
Scale ready for operation (W x D x H) mm
275 x 295 x 58
with column:
275 x 460 x 1010
550x550x1060 550x550x61
Scale plate mm
275 x 295 x 60 550x550 550x550 Total weight kg (net) 4.1 6.6 Rechargeable battery
operation (optional)
Loading time
operating time: 35 h;
7.2 V / 2000 mA
21.8 15.0
Loading time
operating time: 45h;
7.2 V / 2000 mA
Loading time
operating time: 50h;
7.2 V / 2000 mA
MPS / MTS / MXS / MWS-BAPS-e-1710
7
KERN (Type)
MWS
300K1LM
MWS
300K100M
MWS
400K100DM
Display 6-position one Weighing range (max.) 300 kg 300 kg 300kg; 400kg Display LCD with 25 mm high digits Recommended calibration
weight, (class) Signal rise time
(typical)
300 kg
(M1)
300 kg
(M1)
2 – 3 sec.
400 kg
(M1)
Warm-up time 10 min; 10 min Operating temperature + 5° C …. + 35° C Storage temperature - 20°C … + 60°C Air humidity max. 80 % (non-condensing)
mains adapter 15V / 300 mA ( EN60601-1)
Power supply
operation with 6 x 1.5 V battery supply,
AA type batteries
operation time 50 h
Auto-Off function after 3 min without load change (possibility of setting) Terminal (B x T x H) mm
Scale ready for operation (W x D x H) mm
1500x860x68 1155x830x65 1255x1060x69
210 x 110 x 45
Scale plate mm 800x1200 910x740 1000x1000 Total weight kg (net) 42 28,6 42.2
Rechargeable battery operation (optional)
Loading time:14 h;
operating time: 45 h;
7.2 V / 2000 mA
Loading time:14 h;
operating time: 45h;
7.2 V / 2000 mA
Loading time:14 h;
operating time: 45h;
7.2 V / 2000 mA
MPS / MTS / MXS / MWS-BAPS-e-1710
8

1.1 Tolerances altimeter Measured value (cm)

90 ± 0.5 100 ± 1.0 150 ± 1.0 200 ± 1.0
Tolerance (cm)
MPS / MTS / MXS / MWS-BAPS-e-1710
9

2 Decl a ration of conformity

Only those people who can stand on the scales platform with their
To view the current EC/EU Declaration of Conformity go to:
www.kern-sohn.com/ce

3 Basic directions (general information)

3.1 Intended use

The scales are used for determining the body weight of people in the standing posi­tion. The person being weighed shall stand in the middle of the weighing surface and re­main calm.
The value may be read after the scales stabilizes. The scales are designed for continuous use.
both legs may enter it.
The scales platform is equipped in the anti-slippery surface which shall not be taken off or covered during the time the people are being weighed.
MPS / MTS / MXS / MWS-BAPS-e-1710
10

3.2 Inappropriate use

Do not use the scales for dynamic weighing. Do not leave a permanent load on the weighing plate. This can damage the measur­ing equipment.
Be sure to avoid impact shock and overloading the scale in excess of the prescribed maximum load rating (max.), minus any possible tare weight that is already present. This could result in damage of the scale. Never operate the scale in hazardous locations. The series design is not explosion­proof. Attention should be paid that flammable mixture may also be for me d fro m anaesthesiological means that contain oxygen or laughing gas (nitrous oxide). Construction alterations may not be made to the scale. This can lead to incorrect weighing results, faults concerning safety regulations as well as to destruction of the scale. The scale may only be used in compliance with the described guidelines. Other are­as of application/planned use must be approved by KERN in writing. Do not use the scales for determining body weight in medicine.

3.3 Guarantee

The guarantee shall become void in the event of the following:
non-observation of our guidelines in the Operating Instructions,
use outside the described applications,
alteration to or opening the device,
mechanical damage or damage caused by media, liquids,
usual wear and tear,
inappropriate erection or electric instal lation,
overloading of the measuring equipment,
scale falling down.
MPS / MTS / MXS / MWS-BAPS-e-1710
11

3.4 Monitoring the test substances

The metrology features of the balance and any possible available adjusting weight must be checked at regular intervals within the scope of quality assurance. For this purpose, the responsible user must define a suitable interval as well as the nature and scope of this check. Information is available on KERN’s home page (www.kern-
sohn.com) with regard to the monitoring of balance test substances and the test
weights required for this. Test weights and balances can be adjusted quickly and at a reasonable price at KERN’s accredited DKD (Deutscher Kalibrierdienst) calibration laboratory (return to national standard).
In the case of the scales for weighing people provided with the scale to determine a body size, it is recommended to carry out the check of its measuring accuracy be­cause determination of the human body size is always connected with a very large inaccuracy.

4 Basic safety directions

4.1 Observing the directions included in the Operating Instructions

Carefully read this oper ati on ma nual before
setup and commissioning, even if you are al­ready familiar with KERN balances.
MPS / MTS / MXS / MWS-BAPS-e-1710
12

5 Transport and storage

Secure all parts such as the weighing platform, power unit etc.

5.1 Check upon delivery

Please check the packaging immediately upon delivery and the device during un­packing for any visible signs of external damage.

5.2 Packaging / return transport

Keep all parts of the original packaging for a possibly required
return.
Only use original packaging for returning. Prior to dispatch disconnect all cables and remove
loose/mobile parts.
Reattach possibly supplied transport securing devices.
against shifting and damage.
MPS / MTS / MXS / MWS-BAPS-e-1710
13

6 Unpacking, installation and star ti ng

6.1 Place of installation, place of use

The scale is designed in such a way that reliable weighing results can be achieved under normal application conditions. By selecting the correct location for your scale, you will be able to work quickly and precisely.
Therefore, please observe the following when choosing a place of installation:
- Place the scale on a firm, level surface;
- Avoid extreme heat as well as temperature fluctuation caused by installing the
scale next to a radiator or in the direct sunlight;
- Protect the scale against direct draughts due to open windows and doors;
- Avoid shaking during weighing;
- Protect the scale against high humidity, vapours and dust;
- Do not expose the device to extreme dampness for longer periods of time. In-
admissible bedewing (condensation of air moisture on the device) can occur if a cold device is taken into a significantly warmer environment. In this case, please keep the device for approx. 2 ho ur s at room temp er a ture after it has been disconnected fro m mains s uppl y ;
- Avoid static charge build-up on the scale and people to be weighed;
- Avoid contact with water.
Major dis play deviati on s (i ncorrect weighing results ) are possibl e if electrom ag n e t ic fields occur (e.g. coming from mobile phones or radio equipment) as well as due to static charging and instable power supply. It is necessary then to change the scale locatio n or remove disturbance source.

6.2 Unpacking

Carefull y r emove indivi d ua l scale parts or the whole scal e fr om its packagi ng and po­sition the scale in its intended working location. When the mains adapter is used , be careful not to cause the danger of falling over the power cable.
MPS / MTS / MXS / MWS-BAPS-e-1710
14

6.3 Installation and setting of scale

Personal weighing scale MPS with wall bracket:
Scope of delivery:
MPS / MTS / MXS / MWS-BAPS-e-1710
15
Personal weighing scale MPS-PM with column:
Scope of delivery :
Balance with display unit and tripod
Mains adapter
4 screws
MPS / MTS / MXS / MWS-BAPS-e-1710
16
Assembly:
Remove cap (1) Pull cable with Plug-in con-
Place supporting foot next to
Unscrew the screw (2)
nection (3) through the sup­porting foot (4) and pull it out at the end (5)
the balance
MPS / MTS / MXS / MWS-BAPS-e-1710
17
Introduce cable completely
Reinsert cap (1)
Use the 4 screws to attach
into the tripod tube (6)
Screw-in screw (2)
When screwing the screw en­sure that the plug-in connect­or inside the tripod foot is not squeezed.
the tripod to the bottom of the balance
Level balance with foot screws until the air
bubble of the water balance is in the pre­scribed circle.
Check levelling regularly.
MPS / MTS / MXS / MWS-BAPS-e-1710
18
Adjust foot screw of the tripod in a
way that the tripod has a safe base and is not loose.
MPS / MTS / MXS / MWS-BAPS-e-1710
19
Mounting height measuring rod MSF 200:
How to attach to
KERN scale
Use the 2 screws for screwing the bracket into the supplied insert nuts on the stand of the scales.
Extend the height measuring rod and tighten it in the lower hole of the bracket with the help of the appertaining screw.
The height measuring rod can be mounted in the same way at the back of the support stand.
MPS / MTS / MXS / MWS-BAPS-e-1710
20
Scale MTS with handrail:
Scope of delivery:
MPS / MTS / MXS / MWS-BAPS-e-1710
21
Assembly: Fasten 3 corner elements to the platform, using 4 screws each time.
Place the handrail on 3 corner elements and screw it.
Fasten the terminal holder to the handrail with 3 screws.
Remove the side rubber plugs at both sides of the display. Fasten the display to the holder with both handwheels. Adjust the display position with handwheels.
MPS / MTS / MXS / MWS-BAPS-e-1710
22
Bariatric scale MXS:
Scope of delivery:
MPS / MTS / MXS / MWS-BAPS-e-1710
23
Wheelchair scale MWS and Transportation stretcher balance MWS-L:
MWS
MWS-L
Scope of delivery:
MPS / MTS / MXS / MWS-BAPS-e-1710
24
Direction concerning installation of external column on MPS model without column, MXS and MWS models
Fasten the round plate to the aluminium profile with screws.
Fasten the wall bracket to the top of aluminium profile top with screws.
Remove the side rubber plugs at both sides of the display. Fasten the display to the bracket with both handwheels. Adjust the display position with handwheels. Fasten the cable with cable clips.
MPS / MTS / MXS / MWS-BAPS-e-1710
25
Assembly of holding bracket set MWS-A02 at the MWS models
1 Railing
2 Handrail butts
3 Cross strut
6 Screw
7 Hexagonal socket
wrench
8 Screw
(for installation cross strut)
4 Support
5 Tapped bush
9 Screw
(for installation display unit)
MPS / MTS / MXS / MWS-BAPS-e-1710
26
We recommend engaging a second person to assist you during installa­tion.
Carefully remove the plastic cover from the weighing scale, ensuring that the weighing scale does not get scratched in the process.
Mount all 4 handrail butts (2) on the weighing scale frame.
Ensure that the handrail butt and ca­ble conduit are positioned to the right of the power unit supply point. (s. fig.)
Take the two hexagonal socket wrench­es (7) and fix all handrail butts with the help of the screws 6 (3x) and the tapped bushes 5 (2x) as shown on the illustra­tion.
Make sure that all screws are tight­ened properly.
MPS / MTS / MXS / MWS-BAPS-e-1710
27
Ensuring perfect fit, put the handrail con­taining three holes for the display unit on the handrail butt (1) and cable conduit. (see fig.)
Using the holders 4 (2x) fasten the hand- rail to the handrail butts. Once again, use screws 6 (3x) and tapped bushes 5 (3x).
Carry out same sequence of operations for second handrail.
Use the two screws (8) to fasten the cross strut (3).
Use the three screw s to fasten the mounting plate to the handrail.
MPS / MTS / MXS / MWS-BAPS-e-1710
28
Take a screw driver and remove the
The reading direction of the display unit is optional and can be adapted during the installation of the application.
plastic covers from both sides of the dis­play unit.
Screw the display unit to the handle, us­ing the synthetic screws included in the delivery,
Display unit facing inward
Display unit facing outward
MPS / MTS / MXS / MWS-BAPS-e-1710
29
Using the synthetic covers, seal both holes in the two handrail butts without cross struts.
Once installation is complete, check all screws for tight fit. Otherwise the person to be weighed may suffer an i njur y.
General direction concerning setting up the previously mentioned scales
Place a personal weighing scale in the intended location and level it with the adjusta­ble rubber feet until the air bubble in the spirit level (located in the centre of the scale plate) is in the centre. When scales with large and heavy platforms are installed and transported (a scale plate folded upwards), take care not to drop a scale as this could cause its damage.
MPS / MTS / MXS / MWS-BAPS-e-1710
30

6.3.1 Scope of delivery

Standard accessories:
Mains adapter
Operating Instructions

6.3.2 Installation direction for a model with wall bracket

(personal weighing scale, bariatric scale, wheelchair scale, transportation stretcher scales)
MPS / MTS / MXS / MWS-BAPS-e-1710
31

6.4 Magnets display unit MWS

The display unit of the MWS has two magnets on the rear side, use these magnets to fix the display unit on metalli c surfaces.

6.4.1 Transportation of balance

There exists the possibility to fix the display unit at the platform using the two mag­nets on the rear side, whereby the balance and the display unit can be transported together without problems (see fig. below).
MPS / MTS / MXS / MWS-BAPS-e-1710
32
Connection CN 3 for
Connection CN 4 for

6.5 Mains socket

Power supply is carried out by means of the external mains adapter which also pro­vides separation between mains and a scale. The printed voltage value must be compliant with local voltage. It is allowed to use only admitted, original power supply adaptors of KERN company.

6.6 Battery operation/ Rechargeable battery operation (optional) (only models with battery operation and rechargeable battery operation)

MPS / MTS / MXS / MWS-BAPS-e-1710
batteries (AA x 6)
rechargeable battery
33

6.6.1 Battery operation

off the battery cover on the lower
On models where the back of the display unit is not directly accessible, remove the two black rotary knobs from both sides of the display unit in order to open the battery compartment and remove the display unit from the holder.
Lift-
side of the balance
Carefully take out the battery holder
(1)
Insert 6 batteries (AA).
Ensure that the batteries are in­serted in the correct direction
MPS / MTS / MXS / MWS-BAPS-e-1710
34
Insert battery holder with the inserted
batteries into the display unit
Ensure that the cables are not squeezed
Close the battery cover
If the batteries are run down, "LO" appears in the display. To turn
ON
off, press the button and immediately change the batteries.
OFF
If the balance is not used for a longer time, take out the batteries and store them separately. Leaking battery liquid could damage the balance.
MPS / MTS / MXS / MWS-BAPS-e-1710
35

6.6.2 Rechargeable battery operation (optional)

n the lower
1
When an optional rechargeable battery is used, proceed as follows:
On models where the back of the display unit is not directly accessible, remove the two black rotary knobs from both sides of the display unit in order to open the battery compartment and remove the display unit from the holder.
Lift-off the battery cover o
side of the balance
Carefully take out the battery holder
(1)
Carefully pull-out plug (2) from the
connection CN 4 (3)
MPS / MTS / MXS / MWS-BAPS-e-1710
36
Carefully insert the rechargeable bat-
tery block and insert plug into connec­tion CN 3
Ensure that the cables are not squeezed
Close the battery cover
If the rechargeable battery is exhausted, „LO“ is displayed. The rechargeable battery is loaded via the provided plug-in power supply unit (loading time 14 h for a complete loading). If the balance is not used for a longer time, take out the recharge­able battery and store it separately. Leaking liquid could damage the balance.
MPS / MTS / MXS / MWS-BAPS-e-1710
37

6.7 Initial start-up

Function
Settings
Description
SEtuP
A. oFF
Automatic turning
180 s
Automatic turning of f after 3 m inutes
240 s
Automatic turning of f after 4 m inutes
300 s
Automatic turning of f after 5 m inutes
oFF
Automatic turning off turned off
120 s
Automatic turning of f after 2 m inutes
burr
on
Acoustic signal turned on
oFF
Acoustic signal turned off
End
Leaving menu after pressing button
To obtain accurate weighing results with electronic scales, the appropriate operating temperature is to be provided for them (refer to „Warm-up time”, section 1). During warm-up the scale must be connected to power supply and switched on (mains socket or batteries).
The accuracy of the scale depends on the local acceleration of gravity. The value of acceleration of gravity is given on the rating plate.

6.8 Menu overview

When the scale is switched on, hold the [0] key pressed for about 3 seconds until the display shows successively the „SETUP” and „A.OFF” symbols. Selection is carried out with the [TARE] and [HOLD] keys.
off, Auto Off func­tion
Acoustic signal
HOLD
MPS / MTS / MXS / MWS-BAPS-e-1710
38

7 Operation

7.1 Display

Type MPS 200K100M Type MPS 200K100PM
Type MPS 200K100NM Type MPS 200K100PNM
MPS / MTS / MXS / MWS-BAPS-e-1710
39
Scale zero display: If the scale does not show exactly
zeroed again.
Stabilisation display:
unstable condition, the stabilisation display [ο] dis appears.
It is illuminated when mains supply is via the mains adaptor.

7.2 Display view No. Display Description

1
2 [ο]
3 4 BMI
5 HOLD
6 PRE-TARE
7 NET
[0]
zero value although the scale pan is unloaded, press the [0] key. After a short waiting time, the scale will be
If the display shows the stabilisation display [ο], the scale is in the stable condition. When the scale is in the
Calculated value of the body mass index (BMI).
Hold function / saving function is active.
Initially set tare value is active.
The net weight is displayed.
8 WEIGHT
The present weight value is displayed.
MPS / MTS / MXS / MWS-BAPS-e-1710
40

7.3 Overview of keyboard

kg display.
Key Description
ON/OFF
PRINT
BMI
HOLD
0
M 1-5
PRE-TARE
TARE
CLEAR
0..9
Switching on/off the scale
Data transmission via interface
Determination of Body Mass Index
Hold function / determination of stable weighing value
The scale is reset to 0,0
It is possible to set max. up to 2% of maximal load for
verified scales, and 2% or 100% of maximal load for
common scales (possibility of selection in the menu)
Memories 1–5 were called
Calling the tare function with set values
Taring the scale
Clearing the digits entered manually
Entering digits
ENTER
Using the entered digits
MPS / MTS / MXS / MWS-BAPS-e-1710
41

8 Using s c a le

8.1 Weighing

Switch on the scale with the [ON/OFF] key. The diagnostic scale self-check is
performed and then the software version is displayed. The scale is ready for weighing when the „0,00 kg” weight display is shown. Direction: The [0] key makes it possible to zero the scale if necessary and at any time.
Place a person in the middle of the scale. Wait until the stability display (ο) is
shown and then read the weighing result. Direction: If a person is heavier than the weighing range, the display will show the „Err” symbol (= overload).

8.1.1 Weighing with MWS

Due to the great dimensions and the big weighing range especially suitable for weighing of immobile people on transportation stretchers, wheelchairs or overweight people in the adiposity range
8.1.1.1 Weighing with transportation stretcher or wheelchair
Place transportation stretcher/wheelchair with the person on the centre of the
scales
Arrest the brakes of the transportation stretcher/of the wheelchair
Do not leave the person unattended
When the person is lying/sitting quietly, read weighing value 1 Loosen the brakes and carefully pull off the transportation stretcher/wheelchair
with the person
After that weigh the transportation stretcher/wheelchair without the person and
subduct this weight from weighing value 1, from there results the person’s
weight.
MPS / MTS / MXS / MWS-BAPS-e-1710
42

8.2 Taring

The dead weight of any initial load used for weighing may be tared away by pressing the key, so that the following weighing shows the real weight of a person to be weighed.
E.g. when a rubber mat is put on the scale plate, the scale does not
show 0 value.
To start the taring process, press the [TARE] key. Now internal weight
saving is performed and value of 0.0 kg is displayed.
Place a person in the middle of the scale plate. Then read the weight on the display.
Direction:
The scale can store only one tare value. When the scale is unloaded, the saved tare value is displayed with „negative” sign. To delete the saved tare value, unload the scale plate and then press the [TARE] key.

8.3 HOLD function

The scale is provided with the integrated hold function (determination of average val­ue). It enables people to be weighed accurately although they are not still on the scale plate.
Note: Determination of average value is not possible when a person moves too much.
Switch on the scale with the [ON/OFF] key. The diagnostic scale self-check is
performed. The scale is ready for weighing when the „0.00 kg” weight display
is shown.
Place a person in the centre of the scale plate. Press the [HOLD] key. When the triangle is flashing on the display, the scale
takes some measuring values and then the calculated average value is dis-
played.
Press the [HOLD] key again to return the scale to the normal weighing mode. Pressing the [HOLD] key makes it possible to activate the function at any
time.
MPS / MTS / MXS / MWS-BAPS-e-1710
43

8.4 Mother/Child Function

The “Mother/Child” function makes it possible to determine the body weight of infants and small children held in an adult’s arms.
Turn on the scales using the [ON/OFF] switch. The scales enters the self-
diagnostic mode. The device is ready for operation as soon as “0.00 kg” is
shown on the weight display.
Position the adult person in the middle of the scales platform. After showing
the stability indicator, the scales displays the person’s weight. A triangle s ym-
bol is shown under the “WEIGHT” symbol.
Press the [TARE] button and the readout changes to “0.00 kg”. Place the child in the arms of the adult person. After showing the stability indi-
cator, the scales displays the child’s weight. The triangle symbol is now dis-
played under the “NET” symbol.
Press the [TARE] button again and the readout again changes to “0.00 kg”. After stepping off the scales, the total weight of the adult person and the child
is displayed as a negative value.
Press the [TARE] button again. The saved tare value is cancelled and the
scales can be used for weighing again.

8.5 Determination of Body Mass Index

When you obtain a stable weight and display shows 0.0 kg, place a person in the middle of the scale plate. Wait until the weighing value is stable. Then press the BMI key. Now enter a body height. Please take note that reliable determination of BMI index is only possible for body height from 100 cm to 250 cm and weight > 10 kg.
A body height entered as the last one is flashing on the display. Now you can enter a different value with the numerical keypad. Confirm the entered value with the ENTER key, and then a person’s BMI index will be displayed.
When the BMI index value is displayed, it is presented with the arrow pointing the BMI symbol. To return to the weighing mode, press the BMI key once again and the arrow at the BMI symbol will disappear.
MPS / MTS / MXS / MWS-BAPS-e-1710
44

8.5.1 Classification of BMI values

overweight
Classification of weight for adults over 18 years on the basis of Body Mass Index according to WHO, 2000 EK IV and WHO 2004 (WHO - World Health Organization).
Category BMI (kg/m2) Risk of diseases accompanying
Underweight < 18.5 low Normal weight 18.5 – 24.9 average Overweight
Preobesity I degree of obesity II degree of obesity III degree of obesity

8.6 PRE-TARE function

When a tare weight (rubber mat, clothes, ...) is known, this value can be entered manually. If the PRE-TARE key is pressed shortly, the flashing display will be shown. The PRE-TARE function is active as long as the small arrow is pointing the ”PRE­TARE” symbol on the display. The value used as the last one will be displayed. If a different value is required, a new weight value can be entered with the numerical keypad. By pressing the ENTER key, the new value is confirmed and used. Then the entered value with minus sign will be shown on the display. When a person is placed on the scale plate, the display will show a weight value less the value entered previ ousl y. Pressing the PRE-TARE key again will return the scale to the normal weighing mode.
> 25.0
25.0 – 29.9
30.0 – 34.9
35.0 – 39.9 > 40
slightly increased increased high very hig h
MPS / MTS / MXS / MWS-BAPS-e-1710
45

8.6.1 PRE-TARE function with 5 memories

Owing to this function it is possible to store 5 Pre-Tare values (e.g. for different wheelchairs), and then call up them if necessary.
Saving PRE-Tare values:
To enable a later calling up values from the memory, they are to be entered into the memory first. It is carried out in the following way: The scale plate is unloaded, and the scale display is showing 0,0 kg. Put a weight, whose value is to be saved (e.g. an empty wheelchair), on the scale plate and wait until a stable weight display is shown. Press the M1-5 key repeatedly until the display will show the „ni” (M) symbol. Press the key with digit (1..5) shortly to indicate which number a value is to be saved under. The previously displayed weight value is flashing for 3 seconds. When the flashing is finished, press again the key with digit previously pressed and the weighing value is saved in the memory (short audible signal). By pressing the CLEAR key, the scale will return to the weighing mode without saving the value. The display will show the real value of the weight placed on the scale plate. When the weight is removed, the display will show 0,0 kg.
Calling up the PRE-Tare value from memory:
Press the PRE-Tare key repeatedly until the display will show the „ni” (M) symbol. Pressing the key with digit (1..5) will display the flashing weight value saved there. The small arrow, additionally shown on the display, is pointing to the „PRE-TARE” symbol. By pressing another key with digit (1..5), the appropriate also flashing weight value will be displayed. Press the ENTER key and the value will be accepted and shown on the display as the PRE-Tare value with minus sign. Now you can place e.g. a person in a wheelchair or on the transportation stretcher on the scale, and only a person‘s weight will be displayed. To return to the normal weighing mode when the scale plate is unloaded, press the PRE-Tare key shortly again. This will also result in disappearing the small arrow pointing the „PRE-TARE” symbol.
MPS / MTS / MXS / MWS-BAPS-e-1710
46
Printing Pre-Tare memory (refer also to section 8.6):
G
88.8
kg
Gross weight
T
2.0
kg
Taring weight
N
86.8
kg
Net weight
180.0
cm
Size of the person
24.4
BMI
BMI value
To do this, press the PRE-Tare key repeatedly until the display will show the „ni” (M) symbol. Short pressing the PRINT key will activate printing of the values saved in 5 memo­ries.
M1 0.0 kg M2 7.0 kg M3 10.0 kg M4 30.0 kg M5 50.0 kg

8.7 Printing function

To use this function, you need the RS232 interface cable (available as accessories) which is connected with the round plug at the terminal back.
Note: In medical applications, only the peripheral equipment meeting EN 60601-1 standard can be connected to the interface.
When a scale is in the weighing mode, pressing the PRINT key will result in output of the particular data, presented below, via the interface. It is the standard way to output data, which cannot be changed.

8.7.1 Parameters of RS232 interface

Set parameters of the scal e int er face on t he connected device. It is not possible to change the scale parameters. Baud rate: 9600 bps Parity check: no Data length: 8 bits Stop bit: 1 bit Handshake: no or Xon/Xoff Data code: ASCII
MPS / MTS / MXS / MWS-BAPS-e-1710
47

9 Error messages

Remove dirt immediately.
The following messages can be shown on the display during switching on or using the scale. ERRL: Too small weight on the scale.
ooooo:
The scale plate was loaded during switching on the scale. Unload the scale
plate.
ERR: Overload, too large weight on the scale plate.

10 Service, maintenance, disposal

10.1 Cleaning

Disconnect the unit from the mains power supply prior to cleaning, maintenance and repair activities.

10.2 Cleaning / Disinfecting

Clean weighing platform (such as seat) as well as casing with household detergents or commercially available disinfectants, e.g. 70% isopropanol. We recommend a dis­infectant suitable for wiping disinfection. Please follow manufacturer’s instructions.
Do not use abrasive or aggressive cleaners such as spirits or alcohol or similar as they might damage the high-quality surface.
To prevent cross-contamination (fungal skin infection) please observe the following time intervals for disinfection:
Weighing plate before and after any measurement with direct skin contact
When required:
o Display o Foil-covered keyboard
Do not spray disinfectants onto appliance. Make sure that disinfectant does not penetrate the interior of
the balance.

10.3 Sterilisation

Sterilisation of the appl i ance is not all owed.

10.4 Service, maintenance

The device may only be operated and maintained by trained service technicians who are authorised by KERN. Disconnect the scale from mains supply before its opening.
MPS / MTS / MXS / MWS-BAPS-e-1710
48

10.5 Disposal

Disposal of packaging and device must be carried out by an operator according to valid national or regional law of the location where the device is used.
MPS / MTS / MXS / MWS-BAPS-e-1710
49

11 Troubleshooting

The scale should be switched off for a short time following an interruption in the pro­gram sequence and disconn ected from mains supply. It is then necessary to repeat the weighing process from the beginning.
Interruption Possible cause
Weight display is not illumi­nated.
Weight display changes continuously
The scale is not switched on.
The mains supply connection has been interrupted
(mains cable not plugged in/ faul ty).
Check the fus e o f the mains ad apter / green LED is illuminated next to the fuse.
Mains fail ur e.
Batteries are incorrectly inserted or discharged.
No batteries.
Draught/air movement
Table/floor vibrations
The weighing plate is in contact with foreign
matters or is installed incorrectly.
Electromagnetic fields/static charging (choose a different location for the scale, switch off an inter­fering device if possible).
The weighing result is obviously incorrect
Should other error messages occur, switch the scale off and then on again. If the er­ror message remains, inform the manufacturer.
MPS / MTS / MXS / MWS-BAPS-e-1710
The scale display is not set to zero
Incorr ect adj us t me nt.
Great fluctuations in temperature.
Warm-up time was ignored.
Electr omagnetic fields/static charging (choose a
different location for the scale, switch off an inter­fering device if possible).
50
Operation
Display
SETUP
UNKT
CAL
��
K
B
CAL U
of the display. If not, press the [TARE] key.
CAL U
CAL O
30770
CAL O

12 Adjusting

Observe stable environmental conditions. The warm-up time (refer to chapter 1) is required to ensure the scale stabilisation.
Note:
In the case of verified scales, adjusting is locked with the jumper. To carry out adjust­ing, the jumper is to be set in the adjusting position (central position). (refer to 12.2).
Turn the scale on using the [ON/OFF] key.
Press and keep the [0] key pressed for about 3 seconds until the display shows successively the „SETUP” and „UNIT” symbol.
Press the [TARE] key until the „CAL ib” symbol is displayed.
Press the [HOLD] key.
Press the [TARE] key. The triangle must be displayed at the top right side
Press the [HOLD] key repeatedly until the „CAL 0” symbol is displayed.
Press the [TARE] key, the display will show the present numerical value.
Then press the [ENTER] key.
MPS / MTS / MXS / MWS-BAPS-e-1710
51
Press the [HOLD] key.
CAL 5
ue with the [TARE] key.
200.0
cess is started.
82077
Press the [TARE] key. Enter the required calibration weight value (refer to
chapter 1, „Technical data”): Select the item to be changed with the HOLD key and set its numerical val-
Confirm by pressing the [ENTER] key.
Place the calibration weight carefully in the centre of the scale plate, and the display will show a numerical value. Press the [ENTER] key. The adjustment pro-
When the adjustment is finished successfully, the scale is automatically switched over to the weighing mode again and the calibration weight value will be displayed.
Remove the calibration weight.
MPS / MTS / MXS / MWS-BAPS-e-1710
52

12.1 Adjustment key and seals

When verification of a scale is finished, the positions indicated on the scale are sealed.
Verification of a scale without a seal is invalid. Position of seals:
1. Backside 2. Battery compartment
MPS / MTS / MXS / MWS-BAPS-e-1710
53
3. MPS
4. MXS, MTS
5. MWS
MPS / MTS / MXS / MWS-BAPS-e-1710
54
Position of adjustment switch:
Position of adjustment switch
Status
A.OFF
BURR
1 2 3
1. left side not documented
2. centre
adjustment position - adjustment is possible
3. right s ide verificati on pos it ion – adjustment locking

12.1.1 Menu overview in the service mode (adjustment switch in the adjustment position)

Overview is only used to check the set parameters by authorised verification offices. Changes may only be introduced in parameters of the automatic switching off function „
” and audible signal „
”.
MPS / MTS / MXS / MWS-BAPS-e-1710
55
Navigation in menu:
END
When the scale is switched on, press and hold the [0] key press ed for
about 3 seconds until the display shows successively the „SETUP” and „UNIT” symbol.
Press the [TARE] key repeatedly until the required function is displayed. Confirm the selected function with the [HOLD] key. The first parameter will be
displayed. Select the required parameter with the [HOLD] key and confirm the selection with the [TARE] key.
To exit the menu and save the settings, press the [TARE] key until the „ symbol is displayed and then confirm with the [HOLD] key. The scale is automatically returned to the weighing mode.
Selection is carried out with the [HOLD] and [TARE] key.
MPS / MTS / MXS / MWS-BAPS-e-1710
56

12.2 Menu o vervi ew:

Function
Settings
Description
SEtuP
Unit
on-off
Weight unit: kg
Grad
3000 d – 6000 d –
2500 d - 2000 d
The values of the scale elementary division, weigh-
Ut.-d
Full – S-Ut
Selection: one-range scales (Full)- / multi-range
FillE
Fast – Nor. - SLo
Filter: fast - normal - slow
Auto 0
0.25 d – 0.5 d – 1 d – 3 d - OFF
Auto-Zero-Tracking
Stab
0.25 d – 0.5 d – 1 d – 3 d - off
Stabilisierungsbereich
Orang
2 Pct – 100 Pct.
Nullbereich: 2 % / 100 %
Ould
9 d – 2 Pct.
Überlastbereich: 9 d / 2 %
CALib
CAL-U – CAL–0 CAL-5
Justierung
A.Off
120s/180s/240s/ 300s/off
Automatische Abschaltfunktion
burr
on/off
Acoustic signal
default
Resetting to factory settings
End
Leaving the menu after pressing button
10000 d – 500 d – 1000 d – 1500 d -
ing range (max) and elementary division (d)
scales (S-Ut)
MPS / MTS / MXS / MWS-BAPS-e-1710
HOLD
57
Description:
Weight unit: kg Scale divisions, weighing range (max.) and read-out (d) Selection of multi-range / single-range scale Single-range scale Multi-range scale Filter: fast / normal / slow Automatic zero tracking: 0,25 d/ 0,5 d/ 1 d/ 3 d/ OFF Stabilisation range: 0,25 d/ 0,5 d/ 1 d/ 3 d/ OFF Zero range: 2% / 100% Overload range: 9 d / 2% Adjusting Auto off function: 120 s / 180 s / 240 s / 300 sec/ OFF Audible signal: ON/OFF Restoring the factory settings (default settings) Exiting the menu
MPS / MTS / MXS / MWS-BAPS-e-1710