KERN MBB-M series Service Manual

KERN & Sohn GmbH
Ziegelei 1 D-72336 Balingen E-Mail: info@kern-sohn.com
Tel: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com
Service manual
Baby weighing scale
KERN MBB-M (approval series)
Version 1.0 3/2010 GB
MBB-M-SH-e-1010
MBB-M-SH-e-1010
2
GB
KERN MBB-M
Version 1.0 3/2010
Service manual
Baby weighing scale
Table of Contents
1 Basic Information .............................................................................................. 3
2 Technical Data ................................................................................................... 4
3 Overview of Display / Keyboard ....................................................................... 5
4 Error Message Indication .................................................................................. 6
5 Trouble Shooting ............................................................................................... 7
6 Calibration .......................................................................................................... 8
6.1 Calibration switch and seals ................................................................................. 8
6.2 Calibration procedure .......................................................................................... 10
7 Service Mode ................................................................................................... 12
7.1 Navigation in the menu ........................................................................................ 12
7.2 Description ............................................................................................................ 13
7.3 Setting of maximum capacity and resolution .................................................... 14
8 Approval Mode – Menu Overview .................................................................. 15
9 Schematics ...................................................................................................... 16
10 System Block Diagram .................................................................................... 20
11 Display Format ................................................................................................ 21
12 Layout ............................................................................................................... 22
13 Assembly Drawing .......................................................................................... 25
MBB-M-SH-e-1010
3
1 Basic Information
Grundlegende Hinweise
The device must be repaired only by trained specialist staff or personnel with professional formation (such as a repair-specialist accredited by law concerning verification).
The service manual is obligatory for repair work. After repair, original conditions of the device have to be restored. Only original spare parts should be used.
Instructions about conformity-evaluated scales: Repair must be carried only at 100% compliance with the type approval.
A violation of this specification will result in a loss of the type approval! After successful repair the balance will have to be reverified before it can be
used again in a statutorily regulated field.
Das Gerät darf nur von geschultem oder beruflich ausgebildetem Fachpersonal (z. B. eichrechtlich anerkannter Instandsetzer) repariert werden.
Die Serviceanleitung ist bindend für Reparaturen. Das Gerät muss nach erfolgter Reparatur wieder in den Originalzustand
zurückversetzt werden. Es dürfen nur Originalersatzteile verwendet werden.
Hinweis zu konformitätsbewerteten Waagen: Reparatur darf nur in 100% -iger Übereinstimmung mit der
Bauartzulassung erfolgen. Ein Verstoß gegen diese Vorgabe führt zum Erlöschen der Bauartzulassung!
Nach erfolgreicher Reparatur muss eine Nacheichung erfolgen, um die Waage wieder im gesetzlich geregelten Bereich verwenden zu können.
MBB-M-SH-e-1010
4
2 Technical Data
KERN MBB 15K5M MBB 15K2DM
Weighing range (max) 15 kg 6 kg; 15 kg Minimum load (Min) 100 g 40 g Verification value (e) 5 g 2 g; 5 g Display LCD 24mm digit height
1.0 inch LCD with 5 and ½ digits
Recommended calibra­tion weight, (class)
12 kg
(M1) Warm-up time 10 min Operating temperature + 5° C …. + 35° C Storage temperature - 20°C … + 60°C Humidity of air max. 80 % (not condensing)
Power Supply
Line adapter 15V / 300 mA ( EN60601-1)
Battery operation 6 x 1.5V, size AA Auto Off After 3 min without load change, adjustable Housing display unit
(B x D x H) mm
210 x 110 x 45
Weighing pan (WxD) mm 550 x 240 Calibrated in accordance
with 90/384/EEC
Medical grade III
Medical product in accor­dance with 93/42/EEC
Category I with measuring function
MBB-M-SH-e-1010
5
3 Overview of Display / Keyboard
Display view
Display Designation Description
1
0
Zeroing display Should the balance not display exactly zero
despite empty scale pan, press the
but­ton. Your balance will be set to zero after a short standby time.
2
Stability display Scales are in a steady state
3
Power supply connected Illuminates in the event of power supply via
mains adaptor
4
HOLD function active Hold/Save function active
5
Net weight display Net weight will be displayed
6
Weight value display Current weight value will be displayed
MBB-M-SH-e-1010
6
Keyboard
Button Designation Function
ON/OFF-switch Turn on/off
Zero setting key Balance will be reset to 0.0 kg.
Possible up to max. 2% of maximum load for verified scales or 2 % or 100% of maximum load for all other scales (selectable via menu)
HOLD button Hold function/Calculation of a stable weight value
TARE button Tare balance
4 Error Message Indication
A. ErrE : EEprom Error B. ErrL : Low zero count C. Err : High zero count / Overload D. : Low battery indication
E.
ooooo
: The scale plate was loaded during switching on the scale
MBB-M-SH-e-1010
7
5 Trouble Shooting
MBB-M-SH-e-1010
8
6 Calibration
6.1 Calibration switch and seals
After verification the display device is sealed at the indicated positions.
Verification of the balance is invalid without the "seal". Position of the official seals:
1. Rear of display device 2. Battery compartment
3. Bottom side of platform
MBB-M-SH-e-1010
9
Position of calibration switch:
Position of calibration switch Status
1. to the left
Test mode
Shows internal count, e.g. 100052 Zero count: 5000 ~ 200000 Span count: 10000 ~ 250000
Note: If the value of internal count is outside the reference value, please change the load cell or the main board.
2. concentric
Service mode
Calibration position – enables calibration
3. to the right
Approval mode
Verification position – calibration locked
1 2 3
Switch key
MBB-M-SH-e-1010
10
6.2 Calibration procedure
Observe stable environmental conditions. A warming up time of approx. 10 minutes is required for stabilization.
Attention: In calibrated balances the calibration function is switch locked. To carry out calibra­tion, set the switch to calibration position (centre position). (See chpt. 6.1).
Display Operation
a
z000
Ö Start balance by pressing
OFF
ON
SETUP
Ø
UNKT
Ö Press for approx. 3 sec, in the display appears „SETUP“, followed by „UNIT“
CAL
K
B
Ö Press
TARE
repeatedly until “CALIB“ appears
CAL U
Ö Press
HOLD
, „CAL U“ appears
Ö Press
TARE
, the appeared triangle W must be located in the
upper left part of the display. If it is not, press again
TARE
.
MBB-M-SH-e-1010
11
CAL U
Ø
CAL O
Ö Press
HOLD
repeatedly until “CAL 0“ appears.
30770
Ø
CAL O
Ö Press
TARE
,
the display will show a numeric value.
Ö Then press ,
“CAL 0“ appears.
CAL 5
Ö Press
HOLD
,
“CAL 5“ appears.
12.0
Ö Press
TARE
Enter the required size of the calibration weight. For this select the digit to be changed
with
HOLD
and the numeric value using
TARE
.
0
Ö Acknowledge by .
82077
Ö Place the calibration weight carefully in the centre of the
weighing pan and a numerical value will appear in the dis­play.
Ö Press
HOLD
, the calibration process is started.
12.0
The balance will automatically return to weighing mode upon successful completion of the calibration and will display the weight of the calibration weight.
Take away calibration weight.
For verified scales, turn off the scales and move the calibration switch back to verification position.
MBB-M-SH-e-1010
12
7 Service Mode
The balance must be switched over to the service mode. To achieve this, set the calibration switch to calibration position (See chpt. 6.1).
In the service mode all parameters of the balance can be modified.
7.1 Navigation in the menu
When the scale is switched on, press and hold the [0]-key for approx. 3
seconds until “SETUP“, followed by ”UNIT“ appears.
Press the [ TARE ]-key repeatedly until the desired function appears. Confirm the selected function by pressing the [HOLD]-key. The first parameter
appears. To select the desired parameter, use the [HOLD]-key and confirm by pressing the [TARE]-key.
To exit the menu and to store, press the [ TARE ]-key repeatedly until “END“ ap­pears, then confirm by pressing the [HOLD]-key. The balance returns automatically into weighing mode.
A selection is made by pressing the [HOLD]-key and the [ TARE]-key
*
*
*
*
*
*
*
*
* *
*
* default setting
MBB-M-SH-e-1010
13
7.2 Description
Weighing unit: Kg Dividing steps (resolution) and weighing range (max) Selection multi-range scales / single range scales Single-range scales Multi-range scales Filter: Fast / normal / slow Auto Zero Tracking: 0.25d / 0.5d / 1d / 3d / OFF Stabilisation range: 0.25d / 0.5d / 1d / 3d / OFF Zero range: 2% / 100% Overload range: 9d / 2% Calibration Auto off: 120 sec / 180 sec / 240 sec / 300 sec / OFF Audible signal: ON/OFF Reset to default settings (default setup) Exit menu
MBB-M-SH-e-1010
14
7.3 Setting of maximum capacity and resolution
A selection is made by pressing the [HOLD]-key and the [ TARE]-key
Table 1- Dividing steps
NO. 500d 1000d 1500d 2000d 2500d 3000d 6000d
1 1.002 kg 1.001 kg 1.501 kg 1.0005 kg 1.2505 kg 1.5005 kg 1.2002 kg 2 2.505 kg 2.002 kg 3.002 kg 2.001 kg 2.501 kg 3.001 kg 3.0005 kg 3 5.01 kg 5.005 kg 7.505 kg 4.002 kg 5.002 kg 6.002 kg 6.001 kg 4 10.02 kg 10.01 kg 15.01 kg 10.005 kg 12.505 kg 15.005 kg 12.002 kg 5 25.05 kg 20.02 kg 30.02 kg 20.01 kg 25.01 kg 30.01 kg 30.005 kg 6 50.1 kg 50.05 kg 75.05 kg 40.02 kg 50.02 kg 60.02 kg 60.01 kg 7 100.2 kg 100.1 kg 150.1 kg 100.05 kg 125.05 kg 150.05 kg 120.02 kg 8 250.5 kg 200.2 kg 300.2 kg 200.1 kg 250.1 kg 300.1 kg 300.05 kg
9 501 kg 500.5 kg 750.5 kg 400.2 kg 500.2 kg 600.2 kg 600.1 kg 10 1002 kg 1001 kg 1501 kg 1000.5 kg 1250.5 kg 1500.5 kg 1200.2 kg 11 2505 kg 2002 kg 3002 kg 2001 kg 2501 kg 3001 kg 3000.5 kg 12 5010 kg 5005 kg 7505 kg 4002 kg 5002 kg 6002 kg 6001 kg 13 10020 kg 10010 kg 15010 kg 10005 kg 12505 kg 15005 kg 12002 kg
*
6.002
15.005
30.01
Maximum capacity
Dividing steps (resolution)
- standard 3000d -
MBB-M-SH-e-1010
15
8 Approval Mode – Menu Overview
Menu overview for calibrated scales (calibration switch in verification position)
When the scale is switched on, press and hold the [0]-key for about 3 seconds until “SETUP“ followed by “A.OFF“ appears.
A parameter selection is made by pressing the [ TARE] -key and the [HOLD] - key.
A.OFF = Auto off: 120 s / 180 s / 240 s / 300 s / OFF bUrr = Audible signal: ON/OFF End = Exiting the menu
When “End” is selected, press the [HOLD] -key to return to normal weighing mode.
SETUP
A.OFF
120
180
OFF
240
300
OFF
BURR
0n
END
MBB-M-SH-e-1010
16
9 Schematics
Main Board CH-0651
MBB-M-SH-e-1010
17
MBB-M-SH-e-1010
18
MBB-M-SH-e-1010
19
Power Board CH-0723
Junction Board CH-0725
MBB-M-SH-e-1010
20
10 System Block Diagram
MBB-M-SH-e-1010
21
11 Display Format
MBB-M-SH-e-1010
22
12 Layout
Primary Side of Main Board CH-0651
MBB-M-SH-e-1010
23
Secondary Side of Main Board CH-0651
MBB-M-SH-e-1010
24
Primary Side of Power Board CH-0723
Secondary Side of Power Board CH-0723
Primary Side of Junction Board CH-0725
Secondary Side of Junction Board CH-0725
MBB-M-SH-e-1010
25
13 Assembly Drawing