KERN ILT 12K0.1DM, ILT 60K0.5DM, ILT 30K0.2DM, ILT 120K1DM Operating Instructions Manual

KERN & Sohn GmbH
Ziegelei 1 D-72336 Balingen E-Mail: info@kern-sohn.com
Tel: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com
Operating instruction Platform scale
KERN ILT
Version 1.2 04/2006 GB
ILT-BA-e-0612
ILT-BA-e-0612
2
GB
KERN ILT
Version 1.2 04/2006
Operating instruction Platform scale
Contents
1 Technical data__________________________________________________ 4 2 Declaration of conformity ________________________________________
6 3 Appliance overview _____________________________________________ 8 4 Basic Information (General)______________________________________ 10
4.1 Proper use______________________________________________________ 10
4.2 Improper Use ___________________________________________________ 10
4.3 Warranty _______________________________________________________ 10
4.4 Monitoring of Test Resources ______________________________________ 10
5 Basic Safety Precautions________________________________________ 11
5.1 Pay attention to the instructions in the Operation Manual_______________ 11
5.2 Personnel training _______________________________________________ 11
6 Transport and storage __________________________________________ 11
6.1 Testing upon acceptance _________________________________________ 11
6.2 Packaging ______________________________________________________ 11
7 Unpacking, Setup and Commissioning ____________________________
11
7.1 Installation Site, Location of Use ___________________________________ 11
7.2 Unpacking ______________________________________________________ 12
7.2.1 Setup ______________________________________________________________ 13
7.2.2 Scope of delivery _____________________________________________________ 13
7.3 Mains connection ________________________________________________ 13
7.4 Connection of peripheral devices___________________________________ 13
7.5 Initial Commissioning ____________________________________________ 13
ILT-BA-e-0612
3
8 User menu ____________________________________________________ 14
8.1 User principle of the menu control __________________________________ 17
8.2 Operation of the balance with PS/2-keyboard _________________________ 18
8.3 Menu selection __________________________________________________ 19
8.3.1 P1 Calibration (adjustment) _____________________________________________ 19
8.3.2 P2 GLP (Good laboratory practice) _______________________________________ 25
8.3.3 P3 Date/time_________________________________________________________ 27
8.3.4 P4 Readout__________________________________________________________ 29
8.3.5 P5 RS-232 __________________________________________________________ 32
8.3.6 P6 Printout __________________________________________________________ 34
8.3.7 P7 Units ____________________________________________________________ 40
8.3.8 P8 Operating modes___________________________________________________ 40
8.3.9 P9 Globals __________________________________________________________ 41
9 Operation_____________________________________________________ 42
9.1 Operating modes (P8) ____________________________________________ 44
9.1.1 Simple weighing ______________________________________________________ 46
9.1.2 Counting ____________________________________________________________ 47
9.1.3 Checkweighing (weighing with tolerance range) _____________________________ 55
9.1.4 Filling ______________________________________________________________ 57
9.1.5 Percent weighing _____________________________________________________ 58
9.1.6 Animal weighing (dynamic weighing) ______________________________________ 60
9.1.7 Formulation__________________________________________________________ 63
9.1.8 Statistics: ___________________________________________________________ 67
10 Data output _________________________________________________ 70
10.1 Technical data___________________________________________________ 70
10.2 Pin allocation of the balance output plug (front view) __________________ 70
10.3 Remote control instructions _______________________________________ 71
11 Service, maintenance, disposal_________________________________ 72
11.1 Cleaning _______________________________________________________ 72
11.2 Service, maintenance_____________________________________________ 72
11.3 Disposal________________________________________________________ 72
12 Instant help _________________________________________________
73
ILT-BA-e-0612
4
1 Technical data
KERN ILT 12K0.1DM ILT 30K0.2DM
Readability (d) 0.1/0.2 g 0.2/0.5 g Weighing range (max) 6/12 kg 12/30 kg Minimum load 5/10 g 10/25 g Verification value 1/2 g 2/5 g Verification class II II Reproducibility 0.2/0.4 g 0.4/1 g Linearity ± 0.2/0.4 g ± 0.4/1 g Recommended adjusting
weight (not supplied)
10 kg (F2) 20 kg (F2)
Minimum piece weight for counting function
> 0.1 g > 0.2 g
Stabilization time 4 sec. Weighing units
(calibrated appliances)
g, kg
Weighing units (non-calibrated appliances)
g, kg, ct, lb, oz, ozt, dwt, t {h}, t {S}, t {t}, mom, GN, N, t,
freely programmable
Reference quantities at piece counting
10, 20, 50,100, freely selectable
Weighing plate, stainless steel
300 x 300 mm
Dimensions of the housing (B x D x H)
300 x 300 x 75 (platform)
197 x 110 x 35 ( display) Net weight (kg) 7,7 Permitted environmental
condition
-15° C to 30° C
Humidity of air max. 80 % relative (not condensing) Voltage 230 V/50Hz
ILT-BA-e-0612
5
KERN ILT 60K0.5DM ILT 120K1DM
Readability (d) 0.5/1 g 1/2 g Weighing range (max) 30/60 kg 60/120 kg Minimum load 25/50 g 50/100 g Verification value 5/10 g 10/20 g Verification class II II Reproducibility 1/2 g 2/4 g Linearity ± 1/2 g ± 2/4 g Recommended adjusting
weight (not supplied)
50 kg (F2) 100 kg (F2)
Minimum piece weight for counting function
> 1 g > 1 g
Stabilization time 4 sec. Weighing units
(calibrated appliances)
g, kg
Weighing units (non-calibrated appliances)
g, kg, ct, lb, oz, ozt, dwt, t {h}, t {S}, t {t}, mom, GN, N, t,
freely programmable
Reference quantities at piece counting
10, 20, 50,100, freely selectable
Weighing plate, stainless steel
410 x 410 mm
Dimensions of the housing (B x D x H)
410 x 410 x 90 (platform)
197 x 110 x 35 ( display) Net weight (kg) 13 Permitted environmental
condition
-15° C to 30° C
Humidity of air max. 80 % relative (not condensing) Voltage 230 V/50Hz
ILT-BA-e-0612
6
2 Declaration of conformity
KERN & Sohn GmbH
D-72322 Balingen-Frommern Postfach 4052 E-Mail: info@kern-sohn.de
Tel: 0049-[0]7433- 9933-0 Fax: 0049-[0]7433-9933-149 Internet: www.kern-sohn.de
Konformitätserklärung
Declaration of conformity for apparatus with CE mark
Konformitätserklärung für Geräte mit CE-Zeichen
Déclaration de conformité pour appareils portant la marque CE
Declaración de conformidad para aparatos con marca CE
Dichiarazione di conformitá per apparecchi contrassegnati con la marcatura CE
English
We hereby declare that the product to which this declaration refers conforms with the following standards.
Deutsch
Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den nachstehenden Normen übereinstimmt.
Français
Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente déclaration, est conforme aux normes citées ci-après.
Español
Manifestamos en la presente que el producto al que se refiere esta declaración est´´a de acuerdo con las normas siguientes
Italiano
Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce è conforme alle norme di seguito citate.
Electronic Balance: KERN ILT
Mark applied EU Directive Standards
89/336/EEC EMC
EN45501 EN55022
73/23/EEC Low Voltage EN60950
Date: 30.01.2006 Signature:
Gottl. KERN & Sohn GmbH Management
Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0,Fax +49-[0]7433/9933-149
ILT-BA-e-0612
7
KERN & Sohn GmbH
D-72322 Balingen-Frommern Postfach 4052 E-Mail: info@kern-sohn.de
Tel: 0049-[0]7433- 9933-0 Fax: 0049-[0]7433-9933-149 Internet: www.kern-sohn.de
Konformitätserklärung
Declaration of conformity for apparatus with CE mark
Konformitätserklärung für Geräte mit CE-Zeichen
Déclaration de conformité pour appareils portant la marque CE
Declaración de conformidad para aparatos con marca CE
Dichiarazione di conformitá per apparecchi contrassegnati con la marcatura CE
English
We hereby declare that the product to which this declaration refers conforms with the following standards.
This declaration is only valid with the certificate of conformity by a notified body.
Deutsch
Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den nachstehenden Normen übereinstimmt.
Diese Erklärung gilt nur in Verbindung mit der Konformitätsbescheinigung einer benannten Stelle.
Français
Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente déclaration, est conforme aux normes citées ci-après.
Cette déclaration est valide seulement avec un certificat de conformité d’un organisme notifié.
Español
Manifestamos en la presente que el producto al que se refiere esta declaración está de acuerdo con las normas siguientes.
Esta declaración solo será válida acompañada del certificado de conformidad de conformidad de la parte nominal.
Italiano
Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce è conforme alle norme di seguito citate.
Questa dichiarazione sarà valida solo se accompagnata dal certificato di conformità della parte nominale.
Model: KERN ILT
EU Directive Standards
EC-type-approval certificate no. Issued by
90/384/EEC EN 45501 PL 06 001
GUM
Date: 30.01.2006 Signature:
Gottl. KERN & Sohn GmbH Management
Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-07433/9933-0,Fax +49-074433/9933-149
ILT-BA-e-0612
8
3 Appliance overview
Models: ILT 12K0.1DM, ILT 30K0.2M
1. Mains connection
2. PS/2 keyboard
3. RS 232 interface
Air bubble
ON/OFF switch
ILT-BA-e-0612
9
Models: ILT 60k0.50M, ILT 120K1DM
1. ON/OFF switch
2. Mains connection
3. PS/2 keyboard
4. RS 232 interface
Air bubble (underneath weighing plate)
ILT-BA-e-0612
10
4 Basic Information (General)
4.1 Proper use
The balance you purchased is intended to determine the weighing value of material to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing plate. As soon as a stable weighing value is reached the weighing value can be read.
4.2 Improper Use
Do not use balance for dynamic add-on weighing procedures, if small amounts of goods to be weighed are removed or added. The “stability compensation“ installed in the balance may result in displaying an incorrect measuring value! (Example: Slowly draining fluids from a container on the balance.) Do not leave permanent load on the weighing plate. This may damage the measuring system. Impacts and overloading exceeding the stated maximum load (max) of the balance, minus a possibly existing tare load, must be strictly avoided. Balance may be damage by this. Never operate balance in explosive environment. The serial version is not explosion protected. The structure of the balance may not be modified. This may lead to incorrect weighing results, safety-related faults and destruction of the balance. The balance may only be used according to the described conditions. Other areas of use must be released by KERN in writing.
4.3 Warranty
Warranty claims shall be voided in case
Our conditions in the operation manual are ignored
The appliance is used outside the described uses
The appliance is modified or opened
Mechanical dammage or dammage by media, liquids, natural wear and tear
The appliance is improperly set up or incorrectly electrically connected
The measuring system is overloaded
4.4 Monitoring of Test Resources
In the framework of quality assurance the measuring-related properties of the balance and, if applicable, the testing weight, must be checked regularly. The responsible user must define a suitable interval as well as type and scope of this test. Information is available on KERN’s home page (www.kern-sohn.com with regard to the monitoring of balance test substances and the test weights required for this. In KERN’s accredited DKD calibration laboratory test weights and balances may be calibrated (return to the national standard) fast and at moderate cost.
ILT-BA-e-0612
11
5 Basic Safety Precautions
5.1 Pay attention to the instructions in the Operation Manual
Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances.
5.2 Personnel training
The appliance may only be operated and maintained by trained personnel.
6 Transport and storage
6.1 Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the appliance itself when unpacking for possible visible damage.
6.2 Packaging
Keep all parts of the original packaging in case you need to return the appliance. Only use original packaging for returning. Before sending, disconnect all connected cables and loose/movable parts. Attach possibly existing transport safeguards. Secure all parts, e.g. weighing plate, power unit etc., to prevent slipping and damage.
7 Unpacking, Setup and Commissioning
7.1 Installation Site, Location of Use
The balances are designed in a way that reliable weighing results are achieved in common conditions of use. You will work accurately and fast, if you select the right location for your balance.
Therefore, observe the following for the installation site:
Place the balance on a firm, level surface;
Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
Protect the balance against direct draughts due to open windows and doors;
Avoid jarring during weighing;
Protect the balance against high humidity, vapours and dust;
Do not expose the device to extreme dampness for longer periods of time.
Non-permitted condensation (condensation of air humidity on the appliance) may occur if a cold appliance is taken to a considerably warmer environment. In this case, acclimatize the disconnected appliance for ca. 2 hours at room temperature.
Avoid static charge of goods to be weighed or weighing container.
ILT-BA-e-0612
12
If electro-magnetic fields or static charge occur, or if the power supply is unstable major deviations on the display (incorrect weighing results) are possible. In that case, the location must be changed.
7.2 Unpacking
Carefully remove the balance from the packaging and setup balance at the intended workstation.
Remove the transport arrestment (models ILT 60k0.50M, ILT 120K1DM):
Screw in the 4 foot screws
Remove the 4 plastic inserts (2)
Place the weighing pan (3) on the balance
ILT-BA-e-0612
13
7.2.1 Setup
Level balance with foot screws until the air bubble of the water balance is in the prescribed circle.
7.2.2 Scope of delivery
Serial accessories:
Balance
Weighing plate
Mains power supply
Operating Manual
Safety hood above control panel
7.3 Mains connection
Power is supplied via the external mains power supply. The stated voltage value must be the same as the local voltage. Only use original KERN mains power supplies. Using other makes requires consent by KERN.
7.4 Connection of peripheral devices
Before connecting or disconnecting of additional devices (printer, PC) to the data interface, always disconnect the balance from the power supply. With your balance, only use accessories and peripheral devices by KERN, as they are ideally tuned to your balance.
7.5 Initial Commissioning
A warming up time of 15 minutes after switching on stabilizes the measuring values. The accuracy of the balance depends on the local acceleration of gravity. Strictly observe hints in chapter Adjustment.
ILT-BA-e-0612
14
8 User menu
The user menu has nine main menus (P1 - P9), arranged in the following sub-menus:
P1 Calibration
01 External calibration * * * * * * * * Function 02 User calibration * * * * * * * * Function 03 Calibration test * * * * * * * * Function 04 Print report 1 on
P2 GLP
01 User 02 Project 03 Time print 1 on 04 Date print 1 on 05 User print 1 on 06 Project print 1 on 07 Id printout 1 on 08 Last Cal. print 1 on
P3 Date/time
01 Date format 0 DA/MO/YR 02 Time format 0 24 hour 03 Time * * * * * * * * Function 04 Date * * * * * * * * Function 05 Display time 1 on 06 Display date 1 on
P4 Readout
01 Filter 3 normal 02 Med. filter 1 active 03 Disp.refresh 1 04 Auto zero 1 on 05 Last digit 1 always 06 Negative 0 disabled
ILT-BA-e-0612
15
P5 RS-232
01 Baud rate 1 4800 02 Parity 0 None 03 Data bits 2 8 bit 04 Stop bits 1 1 Stop bit 05 Handshake 0 none 06 Auto. Print 0 None 07 Interval 1 * 0.04 s 08 Min. mass 4 10 d 09 Print on stab 1 enabled 10 Pause 1 * 0.1 s
P6 Printouts
01 Prontout men. * * * * * * * * Function 02 1. Printout dest. 0 RS+Mem 03 Printout No. 0 Standard 04 Printout Mem. No. 0 Standard 05 1. Printout start 1 06 1. Printout stop 1 07 2. Printout start 1 08 2. Printout stop 1
.. ………………….
13 1. String 14 2. String
.. ………………….
92 80. String
P7 Units 01 Gram 1 enabled 02 Kilogram 0 disabled 03 Pound 0 disabled 04 Ounce 0 disabled 05 Troy ounce 0 disabled 06 Carat 0 disabled 07 Dwt 0 disabled 08 Teal Hk. 0 disabled 09 Teal S. 0 disabled 10 T.eal T 0 disabled 11 Momme 0 disabled 12 Grain 0 disabled 13 Newton 0 disabled 14 Tical 0 disabled 15 User defined 0 disabled 16 User factor 0.1
P8 Operating modes
01 Parts Count 1 enabled 02 Checkweighing 1 enabled 03 Filling 1 enabled 04 Percent 1 enabled 05 Animal 1 enabled 06 Density 1 enabled 07 Formulation 1 enabled 08 Statistics 1 enabled
ILT-BA-e-0612
16
P9 Globals
01 ID setting * * * * * * * * Function 02 ID autoprint 0 no 03 Beep 1 on 04 Language * * * * * * * * Function 05 Backlight 1 on 06 Contrast * * * * * * * * Function 07 Screensaver 0 disabled 08 Balance Id WL141475 09 Software rev. MBT 1b24 10 Parameter printout * * * * * * * * Function 11 Parameter receive. * * * * * * * * Function
Main menu:
1. Menu number (P1 - P9)
2. Menu designation
3. Cursor ( ► ) indicates the current menu item
Submenu:
1. Main menu number (P1 – P9)
2. Sub-menu number and designation
3. Parameter value or ****** i.e. this
menu item is stored as function
4. Description of function
5. Cursor ( ► ) shows the current
menu item (e.g. P1).
6. Cursor ( ► ) shows the current
setting (e.g. 06).
7. Status
8. Use the arrow key to modify the
parameter value, the active parameter blinks.
ILT-BA-e-0612
17
8.1 User principle of the menu control
In the menu the settings of the balance can be modified and functions can be activated. This way, the balance can be adjusted to individual weighing requirements. Proceed as follows:
Ö Switch on the balance using the -key
Ö After actuating the key the display shows the main menus that can be set
P1 Calibration P2 GLP P3 Date/Time P4 Readout P5 RS-232 P6 Printouts P7 Units P8 Modes P9 Global
s
Setup2.02.05
Ö Use the cursor ( ► ) to select your setting
Use the key to move the cursor ( ) down, the key to move it up
Ö The key confirms your selected setting, the display shows the sub-menu or the
active menu item blinks:
Setup
P1
01 Int. calibr. 02 Ext. calibr. 03 User calibr. 04 Calib r. test 05 Weig ht. co rr. 06 Aut calibr. 07
Print report
o
Auto cal. time
08
******** ******** ******** ********
0.0 0
1
function
function function
of f
on
function
31.01.05
1
1 hour
Ö Use the arrow keys to move within and enter into the menu:
Cursor control up
Parameter selection within function
Increasing the value for selected numeral or number
Access to main menu
Access to sub-menu
Move to next menu item
on the right inside menu
Selection of the number to be changed to the right (the respective active position flashes)
Return to main menu
Selection of the number
to be changed to the left (the respective active position flashes)
Cursor control down
Parameter selection within function
Decreasing the value for selected numeral or number
ILT-BA-e-0612
18
Ö Actuating the -key will confirm your setting; to interrupt use the -key Ö Press -key to return to menu Ö Actuate -key again, a message will inquire whether you want to save the
changes you made.
Ö Actuating the key will save the selected setting. Ö If you do not want to save your selected setting, press the key
Ö The balance returns automatically into weighing mode.
Weighing
0.0000
g
0% 100%
0
27.01.05
8.2 Operation of the balance with PS/2-keyboard
Using a PS/2 keyboard will help you to enter numbers and texts faster and easier.
Key allocation:
Balance keyboard
PS- keyboard
ILT-BA-e-0612
19
8.3 Menu selection
8.3.1 P1 Calibration (adjustment)
As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated - in compliance with the underlying physical weighing principle - to the existing acceleration due to gravity at its place of location (only if the balance has not already been adjusted to the location in the factory). This adjustment process must be carried out for the first commissioning, after each change of location as well as in case of fluctuating environment temperature. To receive accurate measuring values it is also recommended to adjust the balance periodically in weighing operation.
Procedure when adjusting:
Observe stable environmental conditions. A warming up time of ca. 1 hour is required for stabilization. Note! During the adjustment process no objects should be on the weighing plate. In calibrated balances the adjustment function is switch locked. To carry out an adjustment, the switch must be unlocked, see chpt. 8.3.1.2
Operation Display
Ö Call up menu item "P1 Calibration" (see chapt. 8.1)
Ö Press -button
P1 Calibration P2 GLP P3 Date/Time P4 Readout P5 RS-232 P6 Printouts P7 Units P8 Modes P9 Globals
Setup2.02.05
01 Ext. Cal - adjustment with external weight (locked for verifiable devices)
Ö Use the cursor ( ) to select "01 Ext. cal."
Einstellungen
P1
01 02 03 04
Ext.
test
Pr
User calibr.
C
int report
calibr.
alibr.
******** ******** ********
1
func tion
func tion on
func tion
29.12.04
Ö Press -button
Ensure that there are no objects on the weighing plate.
Ö Press -key, on the display appears the weight
value of the adjusting weight.
ILT-BA-e-0612
20
Ö Now move the adjusting weight to the centre of
the weighing plate, press -key. If the adjustment was successful, this will be followed by automatic reset to the menu.
The -key may be used to cancel the adjustment. Errors during adjustment or the use of an incorrect adjusting weight will result in an error message. Repeat adjustment.
Einstellungen
P1
01 02 03 04
Ext.
test
Pr
User calib r.
C
int repo rt
calibr.
alibr.
******** ******** ********
1
func tion
func tion on
func tion
29.12.04
Ö Press -key until you enter the menu/weighing
mode (see chapt. 8.1)
02 User cal. (locked in calibrated appliances)
Ö Use the cursor ( ) to select „02 User cal.“
Einstellungen
P1
01 02 03 04
Ext.
test
Pr
calibr.
alibr.
User calib r.
C
int repo rt
******** ******** ********
1
func tion
func tion on
func tion
29.12.04
Ö Press -key, on the display appears the weight
value of the adjusting weight. The first digit of the displayed value is flashing. Now you can change the weight value by pressing the arrow keys. After adjusting, confirm your new adjusting weight by pressing the -key.
Einstellungen
Benutzerk ierungalibr Masse eingeben g :
12000.00
[]
Ö Ensure that there are no objects on the weighing
plate. Press -button
Ö Press -key, the display shows the weight
value of the adjusting value set by you
ILT-BA-e-0612
21
Ö Now move the adjusting weight to the centre of
the weighing plate, press -key. If the adjustment was successful, this will be followed by automatic reset to the menu.
The -key may be used to cancel the adjustment. Errors during adjustment or the use of an incorrect adjusting weight will result in an error message. Repeat adjustment.
Einstellungen
P1
01 02 03 04
Ext.
test
Pr
calibr.
alibr.
User calib r.
C
int repo rt
******** ******** ********
1
func tion
func tion on
func tion
29.12.04
We recommend:
Carry out all adjustments as close as possible to the maximum load of the balance. Info about test weights can be found on the Internet at: http://www.kern­sohn.com
03 Cal. test
Here, deviation from the last adjustment is determined. This is only a check, i.e. no values are changed.
Ö Use the cursor ( ) to select "03 Cal. test"
Einstellungen
P1
01 02 03 04
Ext.
test
Pr
calibr.
alibr.
User ca libr.
C
int rep ort
******** ******** ********
1
func tion
func tion on
func tion
29.12.04
Ö Press -button
Calibration
Calibration check Clear the p an [Enter]
Ö Press -key, on the display appears the weight
value of the adjusting weight.
Calibration
Calibration check weight measuring
Einstellungen
Calibration check Load weight []
12000.0g
Enter
Ö Now set the adjusting weight in the centre of the
weighing plate. Press -key; calibration test will be executed automatically
Kalibrierung
Kalibrierungstest Bitte warten
Ö The result is displayed
Calibration
Check results Cal.: 180.1354 Act.: 180.1438 Diff.: - 0.0084
ILT-BA-e-0612
22
04 Print report – printout of adjusting data
Ö Use the cursor ( ) to select "04 Print report"
Ö Press -key, the current menu items starts
flashing Highlight your settings by using the arrow keys
0 Print report deactivated 1 Print report activated
Einstellungen
P1
01 02 03 04
Ext.
test
Pr
calibr.
alibr.
User calib r.
C
int repo rt
******** ******** ********
1
func tion
func tion on
func tion
29.12.04
Ö Confirm with the -key
Ö With "Print report" activated, you will receive a printout of your adjustment data
after each adjustment (example):
8.3.1.1 Verification
General introduction: According to EU directive 90/384/EEC balances must be officially verified if they are used as follows (legally controlled area):
a) For commercial transactions if the price of goods is determined by weighing b) For the production of medicines in pharmacies as well as for analyses in the
medical and pharmaceutical laboratory c) For official purposes. d) For manufacturing final packages.
In cases of doubt, please contact your local trade in standard.
Verification instructions An EU type approval exists for balances described in their technical data as verifyable. If a balance is used where obligation to verify exists as described above, it must officially verified and re-verified in regular intervals. Re-verification of a balance is carried out according to the respective national regulations. The validity for verification of balances in Germany is e.g. 2 years. The legal regulation of the country where the balance is used must be observed!
***Report of external calibration***
Calibration : external
Diff.. : 0.0038 g
Signature………………………….