1 Technical data
KERN EW 150-3M EW 600-2M EW 1500-2M
Readout (d) 0,001 g 0,01 g 0,01 g
Verification value (e) 0,01 g 0,1 g 0,1 g
Min. load (Min) 0,02 g 0,5 g 0,5 g
Taring range 150 g 600 g 1.500 g
Reproducibility 0,002 g 0,01 g 0,01 g
Linearity ± 0,003 g ± 0,02 g ± 0,02 g
Recommended adjusting weight, not included (class)
1 x 100g
+ 1 x 20g (F1)
500 g (F2)
1 x 1 kg
+ 1 x 200 g (F1)
Stabilisation time 2 sec.
Weighing plate, stainless steel Ø 110 mm Ø 140 mm Ø 140 mm
Balance dimensions 180 x 230 x 85 mm
Mains adapter 230 V, 50/60 Hz, balance 9 V DC, 400 mA
Net weight (kg) 2,0
Air humidity max. 80 % relative (non-condensing)
Ambient temperature 0° C to 40° C
Authorised surroundings conditions for verifiable applications
10° C to 30° C
Units g, ct
Vibration filter yes
Interface RS 232 C interface (as option)
Accu as option
EW-BA-defsi-0620
18
KERN EW 3000-2M EW 6000-1M
Readout (d)
0,01 g 0,1 g
Verification value (e)
0,1 g 1 g
Weighing range (Max)
3.000 g 6.000 g
Min. load (Min)
0,5 g 5 g
Taring range
3.000 g 6.000 g
Reproducibility
0,01 g 0,1 g
Linearity
± 0,02 g ± 0,2 g
Recommended adjusting weight, not included (class)
1 x 2 kg
+
1 x 1kg (E2)
5 kg (F2)
Stabilisation time
2 sec.
Weighing plate, stainless steel
Ø 140 mm
Balance dimensions
180 x 230 x 85 mm
Mains adapter
230 V, 50/60 Hz, balance 9 V DC, 400 mA
Net weight (kg)
2,0
Air humidity max. 80 % relative (non-condensing)
Ambient temperature 0° C to 40° C
Authorised surroundings conditions for verifiable applications
10° C to 30 ° C
Units g, ct
Vibration filter yes
Interface RS 232 C interface (as option)
Accu as option
EW-BA-defsi-0620
19
2 Unpacking and standard accessories
Take the balance carefully out of the packing and store the plastic wrap for a possible future transport.
The standard accessories consist of:
Balance Weighing pan Weighing pan support External mains adapter Operating instructions
3 Setting up the balance
Place the pan base packed with the weighing pan on the balance. Fix it on the
shaft by driving the knurled nut in the centre. Place the weighing pan properly on the pan base.
Place the balance on a firm surface. See also chapter 11. Drive the adjusting legs to centre the bubble in the circle of the level (front left in
the housing).
EW-BA-defsi-0620
20
4 External view
Key functions
ON/OFF ON/OFF key for lighting display
Print Print key or data transmission key
F
Function key, to call various functions
TARE Zeroing / Taring key – parameter selection key
Loading...