KERN EOA 10K-3, EOA 30K-2 Operating Manual

KERN & Sohn GmbH
Ziegelei 1
Phone: +49-[0]7433- 9933-0 D-72336 Balingen
Fax: +49-[0]7433-9933-149 E-Mail: info@kern-sohn.com
Internet: www.kern-sohn.com
GB
KERN EOA
Version 1.0 10/2013
Operating manual Platform balances
KERN
EOA 10K-3
EOA 30K-2
Readability (d)
0.005 kg
0.01 kg
Weighing range (max)
15 kg
35 kg
Reproducibility
0.005 kg
0.01 kg
Linearity
± 0.025 kg
± 0.05 kg
Warm-up time
10 minutes
Weighing Units
kg
Recommended adjustment weight, not added (class)
10 kg
(M2)
20 kg
( M2 )
Stabilization time (typical)
3 sec.
Operating temperature
+ 10° C ... + 35° C
Humidity of air
max. 80 % (not condensing)
Weighing plate mm
315 x 305
Electric Supply
Input Voltage: 15 V / 600 mA
Battery
9 V block
Operating period: 60 h
Auto-off
3 minutes
Weight kg (net)
3
KERN
EOA 50K-2
EOA 100K-2
Readability (d)
0.02 kg
0.05 kg
Weighing range (max)
60 kg
150 kg
Reproducibility
0.02 kg
0.05 kg
Linearity
± 0.1 kg
± 0.25 kg
Warm-up time
10 minutes
Weighing Units
kg
Recommended adjustment weight, not added (class)
40 kg
(M1)
100 kg
(M1)
Stabilization time (typical)
3 sec.
Operating temperature
+ 10° C ... + 35° C
Humidity of air
max. 80 % (not condensing)
Weighing plate mm
315 x 305
Electric Supply
Input Voltage: 15 V / 600 mA
Battery
9 V block
Operating period: 60 h
Auto-off
3 minutes
Weight kg (net)
3
I
KERN & Sohn GmbH
D-72322 Balingen-Frommern
Phone: 0049-[0]7433- 9933-0 Postbox 4052
Fax: 0049-[0]7433-9933-149 E-Mail: info@kern-sohn.de
Internet: www.kern-sohn.de
Konformitäts-
Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht,
D
erklärung
mit den nachstehenden Normen übereinstimmt.
Declaration of
We hereby declare that the product to which this declaration refers conforms
EN
conformity
to the following standards.
Prohlášení o
Tímto prohlašujeme, že výrobek, kterého se toto prohlášení týká, je v souladu
CZ
shode
s níže uvedenými normami.
Declaración de
Manifestamos en la presente que el producto al que se refiere esta
E
conformidad
declaración está de acuerdo con las normas siguientes
Déclaration de
Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la
F
conformité
présente déclaration, est conforme aux normes citées ci-après.
Dichiarazione di
Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si conformitá
riferisce è conforme alle norme di seguito citate.
Conformiteit-
Wij verklaren hiermede dat het product, waarop deze verklaring betrekking
NL
verklaring
heeft, met de hierna vermelde normen overeenstemt.
Declaração de
Declaramos por meio da presente que o produto no qual se refere esta
P
conformidade
declaração, corresponde às normas seguintes.
Deklaracja
Niniejszym oświadczamy, że produkt, którego niniejsze oświadczenie dotyczy,
PL
zgodności
jest zgodny z poniższymi normami.
Заявление о
Мы заявляем, что продукт, к которому относится данная декларация,
RUS
соответствии
соответствует перечисленным ниже нормам.
EU Directive
Standards
2004/108/EC
EN 61326-1: 2006 EN 61326-2-2: 2006 EN 61000-3-2: 2006 EN 61000-3-3: 2008
2006/95/EC
EN 60950-1: 2006+A11: 2009
Datum
23.09.2013
Date
Signatur
Signature
Ort der Ausstellung
72336 Balingen
Albert Sauter
Place of issue
KERN & Sohn GmbH
Geschäftsführer
Managing director
KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0 Fax +49-[0]7433/9933-149, E-Mail: info@kern-sohn.com, Internet: www.kern-sohn.com
Carefully read this operation manual before setup and
commissioning, even if you are already familiar with KERN balances.
All language versions contain a non-binding translation.
The original German is binding.
Keep all parts of the original packaging for a possibly
required return.
Only use original packaging for returning. Prior to dispatch disconnect all cables and remove
loose/mobile parts.
Reattach possibly supplied transport securing devices. Secure all parts such as the glass wind screen, the
weighing platform, power unit etc. against shifting and damage.
Balance with weighing plate and display unit Operating manual Battery
Operation
Display
.Start balance by pressing
Await zero display
Press and keep it pressed, [CAL] is
displayed
After approx.5 seconds appears automatically the size
of the recommended adjustment weight
(example)
Place adjustment weight in the middle of the weighing
plate, short time later [F] will appear.
Afterwards the balance automatically jumps back to weighing mode. In the display there appears the value of the adjustment weight.
(example)
Button
Description
Function
ON/OFF button
Switch on-off balance
TARE button
short pressing: Tare balance long pressing: Invoke adjustment
Operation
Display
.Start balance by pressing
Await zero display Place goods onto weighing plate.
Now the weight is displayed, after the standstill control appears the weighing unit [kg] right-hand in the display.
(example)
If the material to be weighed is heavier than the weighing range, the display will show [E] (=Overload).
Operation
Display
Turn on weighing scale and wait for zero to
be displayed.
Place the tare vessel (ex. 3kg) on the weighing plate
(example)
and press the button.
The balance display goes to [0].
Put the sample into the tare vessel. Weight of sample is indicated.
(example)
If after finishing the weighing process the
button is pressed again, [0] appears anew in the display.
Operation
Display
Turn on weighing scale and wait for zero to
be displayed.
Put the nominal weight on the weighing plate and tare
(example)
to [0] using the button.
Remove the nominal weight.
The nominal weight appears as negative display.
(example)
Put the test objects subsequently on the weighing
plate, the respective deviation from the nominal weight is displayed with the respective sign to „+“ and „-“.
(example)
Back to weighing mode by pressing the
button and with unloaded weighing plate
Before any maintenance, cleaning and repair work disconnect the appliance from the operating voltage.
Fault
The displayed weight does not glow.
The displayed weight is permanently changing
The weighing value is obviously wrong
Possible cause
The balance is not switched on.
The mains supply connection has been interrupted (mains cable not plugged in/faulty).
Power supply interrupted.
Batteries are inserted incorrectly or empty
No batteries inserted.
Draught/air movement
Table/floor vibrations
The weighing plate is in contact with foreign matter.
Electromagnetic fields / static charging (choose different location/switch off interfering device if possible)
The display of the balance is not at zero
Adjustment is no longer correct.
Great fluctuations in temperature. Electromagnetic fields / static charging
(choose different location/switch off interfering device if possible)