Kern EMB200-2, EMB600-2, EMB220-1, EMB500-1, EMB1200-1 User guide

...
GB
KERN EMB
Version 2.2 11/2005
Operating Instructions Electronic Balance
1 TECHNICAL DATA ................................................................................................................................ 15
2 DECLARATION OF CONFORMTY..................................................................................................... 16
3 FUNDAMENTAL INFORMATION (GENERAL) ............................................................................... 17
3.1
I
NTENDED USE
3.2
I
NAPPROPRIATE USE
3.3
G
UARANTEE
3.4
M
ONITORING THE TEST SUBSTANCES
4 FUNDAMENTAL SAFETY INFORMATION...................................................................................... 18
4.1
O
BSERVE THE INFORMATION IN THE OPERATING INSTRUCTIONS
4.2
S
TAFF TRAINING
5 TRANSPORT AND STORAGE.............................................................................................................. 18
5.1
A
CCEPTANCE CHECK
5.2
P
ACKAGING
6 UNPACKING, INSTALLATION AND COMMISSIONING............................................................... 19
.................................................................................................................................... 17
............................................................................................................................ 17
....................................................................................................................................... 17
.................................................................................................. 17
......................................................... 18
................................................................................................................................. 18
.......................................................................................................................... 18
........................................................................................................................................ 18
6.1
P
LACE OF INSTALLATION, PLACE OF USE
6.2
U
NPACKING
6.2.1 Installation.................................................................................................................................... 20
6.2.2 List of items supplied..................................................................................................................... 20
6.3
B
ATTERY OPERATION
6.4
M
AINS SUPPLY
6.5
I
NITIAL START-UP
6.6
A
DJUSTMENT
6.6.1 Adjusting ....................................................................................................................................... 21
6.7
U
NDERFLOOR WEIGHING
7 OPERATION............................................................................................................................................. 23
7.1
O
VERVIEW OF THE KEYPAD
7.2
O
PERATING INSTRUCTIONS
7.2.1 Weighing ....................................................................................................................................... 23
7.2.2 Taring............................................................................................................................................ 23
7.2.3 Units.............................................................................................................................................. 24
8 MAINTENANCE, UPKEEP, DISPOSAL .............................................................................................. 24
8.1
C
LEANING
8.2
M
AINTENANCE, UPKEEP
8.3
D
ISPOSAL
9 TROUBLESHOOTING............................................................................................................................ 25
........................................................................................................................................ 20
......................................................................................................................... 20
.................................................................................................................................... 20
............................................................................................................................... 20
...................................................................................................................................... 21
..................................................................................................................... 22
................................................................................................................ 23
................................................................................................................. 23
........................................................................................................................................... 24
...................................................................................................................... 24
............................................................................................................................................ 24
............................................................................................. 19
1 Technical data
KERN EMB200-2 EMB600-2 EMB220-1 EMB500-1
Readout 0,01 g 0,01 g 0,1 g 0,1 g Weighing range 200 g 600 g 220 g 500 g Reproducibility 0,01 g 0,01 g 0,1 g 0,1 g Linearity ± 0,02 g ± 0,03 g ± 0,2 g ± 0,3 g
Adjusting weight (not included)
Stabilization time 2-3 sec. AUTO-OFF-Function
(in battery mode) Units g / oz / ozt / dwt
Operating temperature 0° C .... + 40° C
Air humidity max. 80 % (non-condensing) Case (W x D x H) mm 170x240x39 Weighing plate mm Ø 105 Ø 150 Battery supply (standard) 1 x 9 V Mains adapter (optional) 300 mA / 9V Underfloor weighing Hanging loop below the cover plate as standard
KERN EMB1200-1 EMB2200-0 EMB5.2K1 EMB5.2K5
Readout 0,1 g 1 g 1 g 5 g
200g (F2) 500g (F2)
after 3 min.
200 g (M1)
AA(2x 1,5V) 9V
200g (M1)
Weighing range 1.200 g 2.200 g 5.200 g 5.200 g Reproducibility 0,1 g 1 g 1 g 5 g Linearity ± 0,2 g ± 2 g ± 3 g ± 10 g
Adjusting weight (not included) 1.000 g (M1) 1.000 g (M1) 2.000 g (M1) 2.000g (M1)
Stabilization time 2-3 sec. AUTO-OFF-Function
(in battery mode) Units g / oz / ozt / dwt
Operating temperature 0° C .... + 40° C
Air humidity max. 80 % (non-condensing) Case (W x D x H) mm 170x240x39 Weighing plate mm Ø 150 Battery supply (standard) 9V AA (2x 1,5V) 9V AA (2x 1,5 V) Mains adapter (optional) 300 mA / 9V Underfloor weighing Hanging loop below the cover plate as standard
after 3 min.
EMB-BA-defsi-0522
15
2 Declaration of conformty
KERN & Sohn GmbH
D-72322 Balingen-Frommern Postfach 4052 E-Mail: info@kern-sohn.de
Tel: 0049-[0]7433- 9933-0 Fax: 0049-[0]7433-9933-149 Internet: www.kern-sohn.de
Konformitätserklärung
Declaration of conformity for apparatus with CE mark
Konformitätserklärung für Geräte mit CE-Zeichen
Déclaration de conformité pour appareils portant la marque CE
Declaración de conformidad para aparatos con disitintivo CE
Dichiarazione di cofnromitá per apparecchi contrassegnati con la marcatura CE
English
Deutsch
Français
Español
Italiano
We hereby declare that the product to which this declaration refers conforms with the fol­lowing standards.
Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den nach­stehenden Normen übereinstimmt.
Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente déclaration, est conforme aux normes citées ci-après.
Manifestamos en la presente que el producto al que se refiere esta declaración est´´a de acuerdo con las normas siguientes
Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce è confor­me alle norme di seguito citate.
Electronic Balances: KERN EMB
Mark applied EU Directive Standards
89/336EEC EMC EN 61000-6-3 : 2001
EN 61000-3-2 : 2000 EN 61000-3-3 : 1995+A1 : 2001 EN 61000-6-1 : 2001
Date: 15.11.2005 Signature:
Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0,Fax +49-[0]7433/9933-149
Gottl. KERN & Sohn GmbH Management
16
EMB-BA-defsi-0421
3 Fundamental information (general)
3.1 Intended use
The balance you have acquired serves to determine the weighing value of the mate­rial to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing plate. The weighing value can be read off after a stable weighing value has been ob­tained.
3.2 Inappropriate use
Do not use the balance for dynamic weighing. In the event that small quantities are removed or added to the material to be weighed, incorrect weighing results can be displayed due to the “stability compensation“ in the balance. (Example: Slowly drain­ing fluids from a container on the balance.) Do not leave a permanent load on the weighing plate. This can damage the measur­ing equipment. Be sure to avoid impact shock and overloading the balance in excess of the pre­scribed maximum load rating (max.), minus any possible tare weight that is already present. This could cause damage to the balance. Never operate the balance in hazardous locations. The series design is not explo­sion-proof. Structural alterations may not be made to the balance. This can lead to incorrect weighing results, faults concerning safety regulations as well as to destruction of the balance. The balance may only be used in compliance with the described guidelines. Varying areas of application/planned use must be approved by KERN in writing.
3.3 Guarantee
The guarantee is not valid following
- non-observation of our guidelines in the operating instructions
- use outside the described applications
- alteration to or opening of the device
- mechanical damage and damage caused by media, liquids
- natural wear and tear
- inappropriate erection or electric installation
- overloading of the measuring equipment
3.4 Monitoring the test substances
The metrology features of the balance and any possible available adjusting weight must be checked at regular intervals within the scope of quality assurance. For this purpose, the answerable user must define a suitable interval as well as the nature and scope of this check. Information is available on KERN’s home page (www.kern-
sohn.com) with regard to the monitoring of balance test substances and the test
weights required for this. Test weights and balances can be adjusted quickly and at a reasonable price in KERN’s accredited DKD calibration laboratory (return to na­tional normal).
EMB-BA-defsi-0522
17
Loading...
+ 8 hidden pages