Operating instruction Platform scale
Tabel of contents:
1 Technical data__________________________________________________ 4 2 Declaration of conformty _________________________________________ 5 3 Fundamental information (general)_________________________________ 6
3.1 Intended use _____________________________________________________ 6
3.2 Inappropriate use _________________________________________________ 6
3.3 Guarantee _______________________________________________________ 6
3.4 Monitoring the test substances______________________________________ 7
4 Fundamental safety information ___________________________________ 7
4.1 Observe the information in the operating instructions___________________ 7
4.2 Staff training _____________________________________________________ 7
5 Transport and storage ___________________________________________ 7
5.1 Acceptance check ________________________________________________ 7
5.2 Packaging _______________________________________________________ 7
6 Unpacking, installation and commissioning _________________________ 8
6.1 Place of installation, place of use____________________________________ 8
6.2 Unpacking the Balance ____________________________________________ 8
6.2.1 Installation ___________________________________________________________ 8
6.2.2 Package volume_______________________________________________________ 9
6.3 Mains supply/ Battery operation_____________________________________ 9
6.3.1 Mains supply__________________________________________________________ 9
6.3.2 Battery operation ______________________________________________________ 9
6.4 Initial start-up ___________________________________________________ 11
6.5 Adjustment _____________________________________________________ 11
6.6 Adjusting_______________________________________________________ 11
ECE-BA-e-0711
3
7 Operation_____________________________________________________ 12
7.1 Weighing _______________________________________________________ 12
7.2 Taring (TARE key)________________________________________________ 12
8 Maintenance, upkeep, disposal___________________________________ 13
8.1 Cleaning _______________________________________________________ 13
8.2 Maintenance, upkeep _____________________________________________ 13
8.3 Disposal________________________________________________________ 13
9 Troubleshooting _______________________________________________ 14
ECE-BA-e-0711
4
1 Technical data
KERN ECE10K10 ECE20K20 ECE50K50
Readout 10 g 20 g 50 g Weighing range 10 kg 20 kg 50 kg Taring range (subtractive) 10 kg 20 kg 50 kg
Reproducibility
10 g 20 g 50 g
Linearity
± 20 g ± 40 g ± 100 g
Adjusting weight (not included), adjustment/ tolerance as per M3
5 kg 10 kg 20 kg
Stabilisation time (typical) 2 - 3 sec.
Optimal temp. of operation + 0° C .... + 40° C
Humidity max. 80 % (non-condensing) Size (W x D x H) mm 320 x 300 x 60 Weighing plate mm 320 x 260 Weight kg (net) 1,7
ECE-BA-e-0711
5
2 Declaration of conformty
KERN & Sohn GmbH
D-72322 Balingen-Frommern Postfach 4052 E-Mail: info@kern-sohn.de
Tel: 0049-[0]7433- 9933-0 Fax: 0049-[0]7433-9933-149 Internet: www.kern-sohn.de
Konformitätserklärung
Declaration of conformity for apparatus with CE mark
Konformitätserklärung für Geräte mit CE-Zeichen
Déclaration de conformité pour appareils portant la marque CE
Declaración de conformidad para aparatos con disitintivo CE
Dichiarazione di cofnromitá per apparecchi contrassegnati con la marcatura CE
English
We hereby declare that the product to which this declaration refers conforms with the fol­lowing standards.
Deutsch
Wir erklären hiermit, daß das Produkt, auf das sich diese Erklärung bezieht, mit den nach­stehenden Normen übereinstimmt.
Français
Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente déclaration, est conforme aux normes citées ci-après.
Español
Manifestamos en la presente que el producto al que se refiere esta declaración est´´a de acuerdo con las normas siguientes
Italiano
Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce è confor­me alle norme di seguito citate.
Electronic Parcel Scale: KERN ECE
Mark applied EU Directive Standards
89/336EEC EMC EN 61000-3 : 2001
EN 61000-3 : 2000 EN 61000-3-3 : 1995+A1 : 2001 EN 61000-6-1 :2001
Date: 15.11.2005 Signature:
Gottl. KERN & Sohn GmbH Management
Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0,Fax +49-[0]7433/9933-149
ECE-BA-e-0711
6
3 Fundamental information (general)
3.1 Intended use
The balance you have acquired serves to determine the weighing value of the mate­rial to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing plate. The weighing value can be read off after a stable weighing value has been ob­tained.
3.2 Inappropriate use
Do not use the balance for dynamic weighing. In the event that small quantities are removed or added to the material to be weighed, incorrect weighing results can be displayed due to the “stability compensation“ in the balance. (Example: Slowly drain­ing fluids from a container on the balance.) Do not leave a permanent load on the weighing plate. This can damage the measur­ing equipment. Be sure to avoid impact shock and overloading the balance in excess of the pre­scribed maximum load rating (max.), minus any possible tare weight that is already present. This could cause damage to the balance. Never operate the balance in hazardous locations. The series design is not explo­sion-proof. Structural alterations may not be made to the balance. This can lead to incorrect weighing results, faults concerning safety regulations as well as to destruction of the balance. The balance may only be used in compliance with the described guidelines. Varying areas of application/planned use must be approved by KERN in writing.
3.3 Guarantee
The guarantee is not valid following
- non-observation of our guidelines in the operating instructions
- use outside the described applications
- alteration to or opening of the device
- mechanical damage and damage caused by media, liquids
- natural wear and tear
- inappropriate erection or electric installation
- overloading of the measuring equipment
ECE-BA-e-0711
7
3.4 Monitoring the test substances
The metrology features of the balance and any possible available adjusting weight must be checked at regular intervals within the scope of quality assurance. For this purpose, the answerable user must define a suitable interval as well as the nature and scope of this check. Information is available on KERN’s home page (
www.kern­sohn.com) with regard to the monitoring of balance test substances and the test
weights required for this. Test weights and balances can be adjusted quickly and at a reasonable price in KERN’s accredited DKD calibration laboratory (return to na­tional normal).
4 Fundamental safety information
4.1 Observe the information in the operating instructions
Please read the operating instructions carefully before erecting and commissioning, even if you already have experience with KERN balances.
4.2 Staff training
The device may only be operated and looked after by trained members of staff.
5 Transport and storage
5.1 Acceptance check
Please check the packaging immediately upon delivery and the device during un­packing for any visible signs of external damage.
5.2 Packaging
Please retain all parts of the original packaging in case it should be necessary to re­turn items at any time. Only the original packaging should be used for return consignments. Before despatch, disconnect all attached cables and loose/movable parts. Secure all parts, e.g. power unit etc., to prevent slipping and damage.
ECE-BA-e-0711
8
6 Unpacking, installation and commissioning
6.1 Place of installation, place of use
The balance is constructed in such a way that reliable weighing results can be achie­ved under normal application conditions. By selecting the correct location for your balance, you will be able to work quickly and precisely.
Therefore please observe the following at the place of installation:
- Place the balance on a firm, level surface;
- Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
- Protect the balance against direct draughts due to open windows and doors;
- Avoid jarring during weighing;
- Protect the balance against high humidity, vapours and dust;
- Do not expose the device to extreme dampness for longer periods of time. In-
admissible bedewing (condensation of air moisture on the device) can occur if a cold device is taken into a significantly warmer environment. In this case, please acclimatise the device for approx. 2 hours at room temperature after it has been disconnected from the mains.
- Avoid static charging of the material to be weighed, weighing container and
windshield.
Major display deviations (incorrect weighing results) are possible if electromagnetic fields occur as well as due to static charging and instable power supply. It is then ne­cessary to change the location.
6.2 Unpacking the Balance
Carefully remove the balance from the packaging and out of the plastic covering, then place the balance on the assigned position.
6.2.1 Installation
Install the balance in such a fashion that the weighing plate is absolutely horizontal.
ECE-BA-e-0711
9
6.2.2 Package volume
Serial fittings:
KERN ECB
Balance
Operating manual
Power adaptor
6.3 Mains supply/ Battery operation
6.3.1 Mains supply
Electric power supply is by means of the external mains supply circuit. The printed voltage level must comply with the local voltage. Only use original KERN mains supply circuits. The use of other makes is subject to approval by Kern.
Mains adapter output voltage: 9V DC (polarity: inner terminal is negative)
6.3.2 Battery operation
Open the battery cover on the underside of the balance. Insert 6 x 1.5volt (AA) bat­teries. Replace the battery cover. To preserve battery life, the balance switches off automatically after 3 minutes if no weighing has taken place.
For battery operation the balance has an automatic switchoff function which can be activated and deactivated in the menu (chap. 7.5). When the battery power is used up the display will show “LO“. Press the ON/OFF key and change the batteries at once. When the balance is not in use for a longer period of time remove batteries and keep them separately. Leakage of battery liquid might damage the balance.
The optionally supplied battery is charged with the supplied power supply.
The battery has a service life of c. 100 hours. The charging period for total charge is c. 10 hours.
ECE-BA-e-0711
10
Rechargeable Battery Installation:
1. Remove battery cover
2. Slide spacer to left or right, lever down on corner, then remove the spacer
3. Insert and connect rechargeable battery
ECE-BA-e-0711
11
6.4 Initial start-up
A warm-up time of 5 minutes stabilises the measured values after switching on. The accuracy of the balance depends on the local acceleration of the fall.
Please be sure to observe the information in the chapter on adjusting.
6.5 Adjustment
As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated – in compliance with the underlying physical weighing principle - to the existing acceleration due to gravity at its place of location ( only if the balance has not already been adjusted to the location in the factory). This adjustment process must be carried out during the initial start-up, after change in lo­cation and variation of surrounding temperature. It is also recommendable to adjust the balance periodically during weighing operation in order to obtain exact measured values.
6.6 Adjusting
Using a precision weight, the accuracy of the balance can be checked at any time and adjusted.
Adjustment procedure:
Check that the surrounding conditions are stable. A short warm-up time of about 5 minutes is recommended for stabilisation.
Operation Display
Turn balance on using ON/OFF button Hold down TARE button for 3 seconds The weight value for adjusting the scales will appear in the
display. Place the required adjustment weight on the weighing plat-
form (see section 1, “Technical data“) Shortly afterwards, this will appear: If adjustment has not worked correctly or if the wrong ad-
justment weight was used then this will be displayed: In this case switch the power off and on and repeat the adjustment process. Remove the weight from the weighing platform and the display will automatically return to normal weighing mode. Adjustment has been successfully completed.
0,00
10,00
F
0,00
E
ECE-BA-e-0711
12
7 Operation
7.1 Weighing
The balance is switched on by pressing the "ON/OFF" key.
The balance performs auto-diagnostics (for 2 s) and then displays "0.00".
If the weighing object is heavier than the weighing range the display indicates
"E" (=Overload).
The balance is switched off by pressing the "ON/OFF" key once.
7.2 Taring (TARE key)
Switch the balance with the “ON/OFF” key on and wait until display appears as "0.00". Position the tare load and press the “TARE” key.
The balance display becomes "0.00".
The item to be weighed can now be placed in the tare container and the
weight can be read off.
ECE-BA-e-0711
13
8 Maintenance, upkeep, disposal
8.1 Cleaning
Please disconnect the device from the operating voltage before cleaning. Only use a cloth dampened with mild suds and not aggressive cleaning agents (sol-
vents or similar). Please ensure that fluids are not able to get into the device and rub off using a clean, soft cloth. Loose sample residue/powder can be removed carefully using a brush or hand vac­uum cleaner.
Remove any spilt material to be weighed immediately.
8.2 Maintenance, upkeep
The device may only be opened by trained service engineers authorised by KERN. Disconnect from the mains supply before opening.
8.3 Disposal
The operating company shall dispose of the packaging and the device in compliance with the valid national or regional law of the operating location.
ECE-BA-e-0711
14
9 Troubleshooting
The balance should be switched off for a short time following an interruption in the programme sequence and disconnected from the mains supply. It is then necessary to repeat the weighing process from the beginning.
Help:
Interruption Possible cause
Weight display is not illuminated.
The balance is not switched on.
The mains supply connection has been
interrupted (mains cable not plugged in/faulty).
Power supply interrupted. .
The weight display changes continu­ally
Draught/air movement
Table/floor vibrations
The weighing plate is in contact with for-
eign matter.
Electromagnetic fields / static charging (choose different location/switch off inter­fering device if possible)
The weighing result is obviously in­correct
The balance display is not set to zero
Adjustment is no longer correct.
Great fluctuations in temperature.
Electromagnetic fields / static charging
(choose different location/switch off inter­fering device if possible)
Switch the balance off if other error messages should appear and then switch on a­gain. Contact the manufacturer if the error message does not disappear.