KERN ECB, KERN ECE, ECB 10K10, ECB 20K20, ECB 50K50 Service Manual

...
KERN & Sohn GmbH
Ziegelei 1 D-72336 Balingen E-Mail: info@kern-sohn.com
Tel: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com
GB
Service Manual
Electronic Platform Scales
Page 2
KERN ECB / ECE
Version 2.0 4/2008
ECB/ECE-SH-e-0820
GB
KERN ECB / ECE
Version 2.0 4/2008
Service Manual Electronic Platform Scales
Table of Contents
1 Features.......................................................................................................................3
2 Calibration Procedure (CAL) .....................................................................................3
3 Internal Calibration Procedure (Linearity Adjustment) ...........................................4
4 Display Segment Test .............................................................................................. 5
5 Functional Block Diagram / Description...................................................................6
5.1 Functional Block Diagram.....................................................................................................6
5.2 Function Description.............................................................................................................7
6 Trouble Shooting........................................................................................................ 8
7 To Replace PCB.......................................................................................................... 9
8 Schematics................................................................................................................ 10
9 Components Layout................................................................................................. 12
2 ECB/ECE-SH-e-0820
ECB/ECE-SH-e-0820 3
1 Features
Zero weight function
Negative value indication
Low battery indicator
Auto zero
Display segments test function
Auto off function in battery mode - enable/disable (only ECB)
Overload protection
Overload indication
Digital span calibration
Digital internal calibration
Operated by 6 x AA size alkaline batteries
Rechargeable battery as option
Zero tracking - enable/disable (only ECB)
Backlight - 3 modes (only ECB)
2 Calibration Procedure (CAL)
1. Turn scale on and allow the unit to acclimatize and stabilize for 10 minutes.
2. Press and hold the [Tare] key for approx. 3 seconds. Display will show [CAL] and then the required calibration weight.
3. Gently place the correct calibration weight on the center on the weighing pan.
4. Wait until the display shows [ F] and then return to normal weighing mode automatically.
5. Remove the calibration weight and press the [TARE] key to reset the zero point.
In case [ E] is displayed instead of [ F], this indicates a calibration procedure error. Turn the scale off and then on and repeat the procedure.
3 Internal Calibration Procedure (Linearity Adjustment)
1. Remove the bottom cover under the PCB of the scale.
2. Connect J1 on the main board by soldering the pads together.
4 ECB/ECE-SH-e-0820
3. Place scale on a hard level surface. Turn scale on. Display shall show the internal counts.
4. The internal counts shall fall in the range from 3 000 to 6 000. In case out of this range, remove the four screws from the main board and lift it to see the rear side of the main board.
5. Connect J9 left side or right side to increase or reduce the reading. Connecting or disconnecting pads on J5, J6 or J7 can fine tune the reading.