1 Basic information _______________________________________________ 3 2 Introduction____________________________________________________ 4 3 General Information _____________________________________________ 4 4 Error messages_________________________________________________ 5 5 Explanation of display ___________________________________________ 6
5.1 Symbol indicator _________________________________________________ 7
6 Explanation of keyboard _________________________________________ 8 7 External calibration______________________________________________ 9
7.1 Calibration of non-approval models (CXB) ____________________________ 9
7.2 Back to the original value of calibration_______________________________ 9
7.3 Calibration of approval models (CXB-M) _____________________________ 10
8 Service mode access ___________________________________________ 11 9 Linearity calibration ____________________________________________ 12 10 Spec. calibration _______________________________________________ 15
10.1 Reference values for capacity setting _______________________________ 17
11 Temperature calibration_________________________________________ 18
11.1 Indoor air (normal) temperature calibration___________________________ 18
11.2 High temperature calibration_______________________________________ 19
11.3 Low temperature calibration _______________________________________ 20
11.4 Check / Delete temperature calibration data __________________________ 20
12 Spare parts ___________________________________________________ 21
2 CXB-SH-e-0913
1 Basic information
Grundlegende Hinweise
The device must be repaired only by trained specialist staff or personnel with professional formation (such as a repair-specialist accredited by law concerning verification).
The service manual is obligatory for repair work. After repair, original conditions of the device have to be restored. Only original spare parts should be used.
Instructions about conformity-evaluated scales: Repair must be carried only at 100% compliance with the type approval.
A violation of this specification will result in a loss of the type approval! After successful repair the balance will have to be reverified before it can be
used again in a statutorily regulated field.
Das Gerät darf nur von geschultem oder beruflich ausgebildetem Fachpersonal (z. B. eichrechtlich anerkannter Instandsetzer) repariert werden.
Die Serviceanleitung ist bindend für Reparaturen. Das Gerät muss nach erfolgter Reparatur wieder in den Originalzustand
zurückversetzt werden. Es dürfen nur Originalersatzteile verwendet werden.
Hinweis zu konformitätsbewerteten Waagen: Reparatur darf nur in 100% -iger Übereinstimmung mit der
Bauartzulassung erfolgen. Ein Verstoß gegen diese Vorgabe führt zum Erlöschen der Bauartzulassung!
Nach erfolgreicher Reparatur muss eine Nacheichung erfolgen, um die Waage wieder im gesetzlich geregelten Bereich verwenden zu können.
CXB-SH-e-0913 3
2 Introduction
1. There are 3 ways to rectify the linearity for setting the temperature calibration,
which can make the scale more accurate.
n Inside air temperature (normal temperature) o High temperature p Low temperature
It is adequate to only do the inside air temperature calibration. Upon the customer’s request, we also offer the high temperature and low temperature calibration as well.
2. Sampling speed is also depended on the user’s need (01~05).
3. Linearity calibration, capacity setting and temperature calibration can be set
separately.
4. Weighing unit: g or kg (depending on model)
5. Tare range is full range tare.
3 General Information
1. The newly manufactured scale without passing the calibration and setting process will automatically enter the linearity calibration mode. After the linearity calibration procedure is done, the program will automatically enter the Specification Setting Mode.
2. Zero Range
The internal value should be in between 120,000~250,000 with the platter on and without any load (check Æ see chapter 10 “Spec. calibration”)
3. Maximum Capacity
The internal value should be over 300,000 when the scale is fully loaded.
4. When rubbing the tetragon, the inaccuracy value should be smaller than 10 divisions.
5. Tare range is full range tare.
4 CXB-SH-e-0913
4 Error messages
Non-approval models: E1 ZERO over 300,000 internal value.
E2 ZERO below 10,000 internal value.
Approval models: E1 ZERO over the zero position in weight calibration + 10% of full scale.
E2 ZERO over the zero position in weight calibration - 10% of full scale.
All models: E4 ZERO is unstable. E5 Turning on machine core value is under 0. E6 Core value is over 250,000. E7 Core value is below 120,000.
- - OL - The weighing range is exceeded.
NOTES:
Non-approval models (CXB): After E1, E2 or E4 is displayed press
to continue.
Approval models (CXB-M): After E1, E2 or E4 appear, set the jumper SWA1
on main board to “ADJ” position and press to return the counting mode.
Then press
+ + at the same time followed by pressing
to access calibration (see chap. 7.3).
CXB-SH-e-0913 5
5 Explanation of display
Weight Reference weight
Loading status display Pieces
6 CXB-SH-e-0913
5.1 Symbol indicator 3
Display weight
Here, the weight of your goods is displayed.
Overlay W indicates:
Æ0Å Zeroing display
PRE-
TARE
Tare in memory
Battery very low
Display reference weight
The reference weight of a sample is shown here. This value is either entered by user of calculated by balance.
Overlay W indicates:
Placed number of pieces insufficient for reference calculation Placed reference weight insufficient for reference calculation
Display quantity
Here, all the parts placed on balance are immediately displayed by number.
Overlay W indicates:
M+
Data in summation memory Stability display
Battery charge status display
red
Battery is almost discharged
green
Battery is completely discharged
CXB-SH-e-0913 7
6 Explanation of keyboard
Choice Function
-
Number keys
Delete key
Call counting with tolerance control
Store reference weights in memory
Call stored reference weights
Addition in total memory
Call up total memory
Delete summation memory
Enter reference weight through weighing
Display reference weight stored last
Enter target number of pieces
Numeric entry reference weight
Display reference weight stored last
Enter target weight
Zeroing key
Back to weighing mode
Taring key
Enter numerical tare
8 CXB-SH-e-0913
7 External calibration
7.1 Calibration of non-approval models (CXB)
Before calibrating, adjust the mini jumper SWA1 to “ADJ” position (see chap. 8) Observe stable environmental conditions. A warming up time of 30 minutes is
required for stabilization.
Operation
Ensure that there are no objects on the weighing plate.
0
W
a
PRE­TARE
Press the key and keep it pressed down whilst simultaneously operating the key
Weight value of calibration weight is flashing on display. However, you may enter a value of your choice via the number keys. *
3000
W
a
PRE­TARE
CAL
Carefully place calibration weight in the centre of the weighing plate. Calibration will be carried out automatically after dead stop control.
Remove calibration weight during count down to zero. The balance returns automatically into weighing mode. An error message will
appear on the display should a calibration error occur or should the calibration weight be incorrect. Turn balance off, then restart it and repeat the calibration process.
* The calibration should be made with the recommended calibration weight (see chap. 1 "Technical data" in the operating manual).
Note Use the key to exit calibration mode. The balance returns to weighing mode.
7.2 Back to the original value of calibration
(only non-approval models)
n Press and hold key followed by pressing key until the scale is counting
backward to zero, then release it.
o The scale will show the original value, which is the value setting at the factory.
CXB-SH-e-0913 9
7.3 Calibration of approval models (CXB-M)
Before calibrating, adjust the mini jumper SWA1 to “ADJ” position (see chap. 8) Observe stable environmental conditions. A warming up time of 30 minutes is
required for stabilization.
Operation
Ensure that there are no objects on the weighing plate.
Ö Press + 1 + 5 keys
Ö Press key
The display shows:
0.000
M+
0
(Example)
Ö Press key The display shows:
15.000
(Example)
M+
0
Weight value of calibration weight is flashing on display.
Ö Carefully place calibration weight in the centre of the weighing plate.
Ö Press key
Calibration will be carried out automatically after dead stop control. Ö Remove calibration weight during count down to zero. The balance returns automatically into weighing mode. An error message will
appear on the display should a calibration error occur or should the calibration weight be incorrect. Turn balance off, then restart it and repeat the calibration process.
194452
194452
* The calibration should be made with the recommended calibration weight (see chap. 1 "Technical data" in the operating manual).
10 CXB-SH-e-0913
8 Service mode access
Access to the service mode is controlled by jumper SWA1 on the main PCB. The
jumper has two positions LOCK, which prevents access to the service mode and ADJ, which allows access to the calibration and configuration routines.
SWA 1 jumper is located adjacent to the RF screening cover of the main PCB.
Place your balance upside down and remove seal/cap. For calibration and
configuration routines the jumper “SWA1” must be set to position “ADJ”.
Sealing mark / unlocking switch
Position of unlocking switch Status
To the right Unlock the balance for calibration process (ADJ) To the left Verification position - calibration locked (LOCK)
CXB-SH-e-0913 11
9 Linearity calibration
Before calibrating, please adjust the mini jumper SWA1 to “ADJ” position. After finish the calibration, please re-adjust the mini jumper SWA1 to
“LOCK” position.
~ Divide the capacity into 5 sections before linearity calibration.
E.g. CXB 30K2 (capacity = 30kg)
Divide it into 5 phases: 3kg, 6kg, 12kg, 12kg, 3kg (The total calibrated weight should be between 1.1~1.25 times of the capacity weight. E.g.: 3kg + 6kg + 12kg + 12kg + 3kg = 36kg are 1.2 times of 30kg)
n Turn on the machine, after the scale resets back to zero,
then press key + 1 key + 6 key (at the same time) to enter the linearity calibration mode. (Please press 9 key to pass over this linearity calibration mode.)
Weight Piece Weight Pcs
0 0 L. XXXXXX
key in *n value core value display the number of times of calibration (zero internal value)
*n = the calibrated value of phase X / the calibrated value of 1st phase
[*n = multiplier (2,3,4…) of 1st calibration weight / 1st phase]
*n ⇒ must be an integral
o Press . key
Weight Piece Weight Pcs
0 1 1 XXXXXX
internal value of the phase
12 CXB-SH-e-0913
p Put the 1st weight mass (e.g. weight mass=3kg, *n=1), then press . key
Weight Piece Weight Pcs
0 2 1 XXXXXX
input *n value internal value of the phase
q Add the 2
nd
weight mass (e.g. 6kg, in sum 9kg) and input *n value (e.g. *n=2),
then press . key
Weight Piece Weight Pcs
0 3 1 XXXXXX
input *n value, e.g. n=2 internal value of the phase
r Add the 3rd weight mass (e.g. 12kg, in sum 21kg) and input *n value
(e.g. *n=4), then press . key
Weight Piece Weight Pcs
0 4 1 XXXXXX
input *n value, e.g. n=4 internal value of the phase
s Add the 4
th
weight mass (e.g. 12kg, in sum 33kg) and input *n value
(e.g. *n=4), then press
. key
Weight Piece Weight Pcs
0 5 1 XXXXXX
input *n value, e.g. n=4 internal value of the phase
CXB-SH-e-0913 13
t Add the 5
th
weight mass (e.g. 3kg, in sum 36kg) and input *n value (e.g. *n=1),
then press . key
Weight Piece Weight Pcs
0 6 1 XXXXXX
input *n value, e.g. n=1 internal value of the phase
u Press key (“CONFIRM” key), the buzzer beeps once
Weight Piece Weight Pcs
S Y S U P
v Remove all the weight mass, the scale will reset back to zero automatically,
then the procedures finished.
14 CXB-SH-e-0913
10 Spec. calibration
Before calibrating, please adjust the mini jumper SWA1 to “ADJ” position. After finish the calibration, please re-adjust the mini jumper SWA1 to “LOCK”
position. n Turn on the machine, after the scale resets back to zero,
then press key + 1 key + 7 key (at the same time) to enter the spec. calibration mode.
Weight Piece Weight Pcs
0 3 0 0 1 0 2 1 X X 2 0 2 0 5 3
program version keyboard default value core value
(to check keyboard function) (internal value of zero)
see chapter 3
“General Information“
o Press key
Weight Piece Weight Pcs
Ñ0Ò0 0 0 0 0 0 0 0 0 0 0 0 0 0 1
full capacity setting preset calibrated value refer to “Approx. 1” key in “ full scale+9d ” suggest key in “full scale” (overload function max.+ 9d)
Approx. 1: 0 0 0 0 0 1
a b c d e
a input the division 1 = e.g. 1g / 10g 2 = e.g. 0,2g / 2g 5 = e.g. 0,5g / 5g b input the decimal point 0 = e.g. 1g 1 = e.g. 0,2g
2 = e.g. 0,01kg 3 = e.g. 0,001kg
c input the unit of the unit weight 0, 1, 2, 8, 9 (full capacity are 10n of division)
d define the keyboard function of ; kg/lb ; PST CE keys (use always 1 )
0 = PST CE, unit weight in kg 1 = 10 preset, unit weight in kg (OIML approval model) 2 = kg/lb (1st unit = kg, 2nd unit = lb) 3 = 10 preset, kg/lb (1st unit kg, 2nd unit lb) 4 = PST CE, lb unit only 5 = 10 preset, lb unit only 6 = lb/kg (1
st
unit lb, 2nd unit kg) 7 = 10 preset, lb/kg (1st unit lb, 2nd unit kg)
0 = 10
0
(x 1) » CXB 15K1 / CXB 30K2
1 = 10
1
(x 10)
2 = 10
2
(x 100)
8 = 10
-2
(x 0.01)
9 = 10
-1
(x 0.1) » CXB 3K0.2 / CXB 6K0.5
Example: CXB 3K0.2 a= 2. value of a x 10n = Readout 2 x 10
-1
= 0,2g c= 9
CXB-SH-e-0913 15
e different versions of models setting
0 = Non-approval model (External calibration is applicable) 1 = OIML approval model (The display resolution should be set at 1/3000,
External calibration is NOT applicable) 4 = Non-approval model (External calibration is NOT applicable)
p After input spec., press
key (not necessary to put the weight mass on the platter, the calibrated weight value is the reference for the temperature calibration procedure).
The scale will be self-tested and reset back to zero.
Press
key1 key5 key, then press key to enter the temperature
calibration mode (see chapter 11).
NOTES:
(1) The last number of full capacity value must be variable number, can be fixed “0”. (2) The digit and location of preset calibrated value, refer to the location of capacity digit. (3) Input capacity + 9d in order to run 9 times overload function. (4) If max. number of digit is “0”, please input “0”. (5) Reference values for capacity setting see chapter 10.1.
16 CXB-SH-e-0913
10.1 Reference values for capacity setting
Weighing unitg –
Weight Piece Weight Pcs
Ñ3Ò0 0 1 8 3 0 0 0 0 2 1 9 1 0 (1)
Weight Piece Weight Pcs
Ñ6Ò0 0 4 5 6 0 0 0 0 5 1 9 1 0 (1)
Weight Piece Weight Pcs
Ñ1Ò5 0 0 9 1 5 0 0 0 1 0 0 1 0 (1)
Weight Piece Weight Pcs
Ñ3Ò0 0 1 8 3 0 0 0 0 2 0 0 1 0 (1)
CXB 3K0.2
CXB 6K0.5
CXB 15K1
CXB 30K2
Weight Piece Weight Pcs
Ñ3Ò0 0 1 8 3 0 0 0 0 1 0 0 1 1 (1)
Weight Piece Weight Pcs
Ñ6Ò0 0 4 5 6 0 0 0 0 2 0 0 1 1 (1)
CXB 3K1M
CXB 6K2M
Weighing unitkg –
Weight Piece Weight Pcs
Ñ1Ò5 0 0 9 1 5 0 0 0 1 3 0 1 0 (1)
Weight Piece Weight Pcs
Ñ3Ò0 0 1 8 3 0 0 0 0 2 3 0 1 0 (1)
CXB 15K1
CXB 30K2
Weight Piece Weight
Pcs
Ñ0Ò3 0 0 9 0 3 0 0 0 1 3 0 1 1 (1)
Weight Piece Weight Pcs
Ñ0Ò6 0 1 8 0 6 0 0 0 2 3 0 1 1 (1)
Weight Piece Weight Pcs
Ñ1Ò5 0 4 5 1 5 0 0 0 5 3 0 1 1 (1)
Weight Piece Weight Pcs
Ñ0Ò3 0 0 9 0 3 0 0 0 1 2 0 1 1 (1)
CXB 3K1M
CXB 6K2M
CXB 15K5M
CXB 30K10M
CXB-SH-e-0913 17
11 Temperature calibration
Before calibrating, please adjust the mini jumper SWA1 to “ADJ” position. After finish the calibration, please re-adjust the mini jumper SWA1 to “LOCK”
position.
11.1 Indoor air (normal) temperature calibration
n * The scale will be self-tested and reset back to zero.
Press key1 key+ 5 key, then press key to enter the temperature calibration mode.
Weight Piece Weight Pcs
0. 0 0 0 XX XXXXXX
indoor temperature(°C) zero internal value
o No weight on the platter, press . key to calibrate zero.
Weight Piece Weight Pcs
3 0. 0 0 0 XX XXXXXX
full capacity value indoor temperature(°C) internal value of full capacity
p Put the full capacity weight on the platter, press . key to calibrate full capacity
value.
q After scale is stable, it will reset back to zero automatically (remove the weight
mass from the platter before it resets back to zero), then the procedures finished. Put the weight mass on the platter again to check if the display shows the accurate weight. If it’s correct, turn off the power, then re-adjust the mini jumper SWA1 to “LOCK” position.
NOTES:
(1) The procedures are for the scale without the temperature IC or for the scale in
the normal temperature environment.
(2) To achieve the accuracy of the scale, please run the high temperature and low
temperature calibration procedures.
(3) The procedures for high temperature and low temperature are totally reversed.
18 CXB-SH-e-0913
11.2 High temperature calibration
scales must be in the temperature lab chamber
n Press key+ 1 key+ 5 key, then press key to enter the temperature
calibration mode
Weight Piece Weight Pcs
0. 0 0 0 XX XXXXXX
high temperature(°C) zero internal value
o No weight on the platter, press . key to calibrate zero
Weight Piece Weight Pcs
3 0. 0 0 0 XX XXXXXX
full capacity value high temperature(°C) internal value of full capacity
p Put the full capacity weight on the platter, press . key to calibrate full capacity
value.
q After scale is stable, it will reset back to zero automatically (remove the weight
mass from the platter before it resets back to zero), then the procedures finished. Put the weight mass on the platter again to check if the display showing the accurate weight. If it’s correct, turn off the power, then re-adjust the mini jumper SWA1 to “LOCK” position.
CXB-SH-e-0913 19
11.3 Low temperature calibration scales must be in the control room of low temperature
n Press key 1 key 5 key, then press key to enter the temperature
calibration mode
Weight Piece Weight Pcs
0. 0 0 0 XX XXXXXX
low temperature(°C) zero internal value
o No weight on the platter, press . key to calibrate zero
Weight Piece Weight Pcs
3 0. 0 0 0 XX XXXXXX
full capacity value low temperature(°C) internal value of full capacity
p Put the full capacity weight on the platter, press . key to calibrate full capacity
value.
q After scale is stable, it will reset back to zero automatically (remove the weight
mass from the platter before it resets back to zero), then the procedures finished. Put the weight mass on the platter again to check if the display showing the accurate weight. If it’s correct, turn off the power, then re-adjust the mini jumper SWA1 to “LOCK” position.
11.4 Check / Delete temperature calibration data
n Press key1 key+ 5 key
Weight Piece Weight Pcs
Sho t 0 0 XXXXXX
show temperature temperature (°C) internal value
o Press 2 or 8 key to check the temperature calibration data.
p Press 2 or 8 key to select the desirable data, and double-press . key to
delete it.
20 CXB-SH-e-0913
12 Spare parts
CXB-SH-e-0913 21
No. Description Q’ty
1
Switch on/off (2 pin) 1
2
Power socket 1
3
Level 1
4
Buckle plug (HP-13) 1
5
Terminal Cover (male) 5
6
Round head screw M3*6 2
7
Round head screw M4*6 2
8
Round head screw TP2 3*8 1
9
Flat head screw M3*10 1
10
Round head screwTP2 4*16 4
11
Hexagon screw M4*12 1
12
Round head screw M4*25 W/10 1
13
Hexagon screw M4*16 4
14
Washer 6*13 4
15
1/4 spring washer 4
16
Nut 3*6 3
17
Battery Cap Foam 1
18
Power nameplate 230V AC 1
19
Power sticker 1
20
Dust cover 1
21
FD aluminum support 1
22
FD aluminum bracket 1
23
Battery 6V/4AH 1
24
X2 capacitor 0.1µF/275V(P=15) 1
25
Wire Fixer 1
26
Front panel 1
27
Rear panel 1
28
Keypad 1
29
Keypad circuit 1
30
Load cell 1
31
9501 battery fixer 1
32
AE washer 2.0mm 2
33
FD Pad for Cockroach Filter Cap 4
34
9910 Upper noise Filter Cap t=0.8mm 1
35
9910 Lower Noise Filter Cap t=0.8mm 1
36
FD Anti-cockroach plastic 4
37
Waterproof rubber 27*18*1mm 1
38
Main board 1
39
FD plastic pan 1
40
FD stainless steel pan 1
41
Adjust feet 4
42
Battery cover (blue) 1
22 CXB-SH-e-0913
CXB-SH-e-0913 23
43
Upper housing 1
44
Lower housing 1
45
AE Sealing 1
46
Round head screw M4*8 6
47
Multi wire 150mm 1
48
Alloy hexagon screw M6*16 4
49
Power cable 1
50
Capacity sticker 1
51
Spring 1
52
Transformer 115/230V-10V 1