KERN CXB 6K0.5, CXB 15K5M, CXB 30K10M, CXP 30K2, CXB 15K1 Operating Instruction

...
KERN & Sohn GmbH
Ziegelei 1 D-72336 Balingen E-Mail: info@kern-sohn.com
Tel: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com
Operating instruction Counting scales
KERN CXB/CXP
Version 1.6 11/2010 GB
CXB/CXP-BA-e-1016
CXB/CXP-BA-e-1016
2
GB
KERN CXB/CXP
Version 1.6 11/2010
Operating instruction Counting scales
Table of Contents
1 Technical data................................................................................................ 4
2 Declaration of conformity ............................................................................. 7
3 Appliance overview ....................................................................................... 8
3.1 Overview of display................................................................................................................ 10
3.1.1 Display weight .......................................................................................................................... 11
3.1.2 Display reference weight .......................................................................................................... 11
3.1.3 Display quantity ........................................................................................................................ 11
3.1.4 Battery charge status display ................................................................................................... 11
3.2 Keyboard overview................................................................................................................. 12
4 Basic Information (General)........................................................................ 14
4.1 Proper use............................................................................................................................... 14
4.2 Improper Use .......................................................................................................................... 14
4.3 Warranty .................................................................................................................................. 14
4.4 Monitoring of Test Resources............................................................................................... 15
5 Basic Safety Precautions ............................................................................ 15
5.1 Pay attention to the instructions in the Operation Manual ................................................ 15
5.2 Personnel training.................................................................................................................. 15
6 Transport and storage................................................................................. 15
6.1 Testing upon acceptance ...................................................................................................... 15
6.2 Packaging................................................................................................................................ 15
7 Unpacking, Setup and Commissioning ..................................................... 16
7.1 Installation Site, Location of Use.......................................................................................... 16
7.2 Unpacking ............................................................................................................................... 16
7.2.1 Setup ........................................................................................................................................ 17
7.2.2 Scope of delivery...................................................................................................................... 18
7.3 Mains connection ................................................................................................................... 18
7.4 Battery power supply............................................................................................................. 18
7.5 Initial Commissioning ............................................................................................................ 18
7.5.1 Start-up..................................................................................................................................... 19
7.5.2 Turn off – CXB models ............................................................................................................. 19
7.5.3 Turn off/standby mode – CXP models ..................................................................................... 19
7.5.4 Balance zero display ................................................................................................................ 19
7.5.5 Stability display......................................................................................................................... 19
7.6 Adjustment with external weight .......................................................................................... 20
7.6.1 Adjustment of models CXP ...................................................................................................... 20
7.6.2 Adjustment of models CXB ...................................................................................................... 21
7.6.3 Adjustment of models CXB_M ................................................................................................. 23
8 Verification ................................................................................................... 24
8.1 Adjustment key and seals ..................................................................................................... 25
9 Parts counting.............................................................................................. 26
9.1 Determination of the reference weight by weighing .......................................................... 26
9.2 Numeric entering of the reference weight ........................................................................... 27
CXB/CXP-BA-e-1016
3
9.3 Automatic reference optimization ........................................................................................ 27
9.4 Store/call reference weight ................................................................................................... 28
9.4.1 Save.......................................................................................................................................... 28
9.4.2 Call-up ...................................................................................................................................... 29
9.5 Count with tolerance control - Fill to target ........................................................................ 29
9.5.1 Set tolerance value for target number of pieces ...................................................................... 29
9.5.2 Set tolerance value for target weight........................................................................................ 30
10 Taring ............................................................................................................ 31
10.1 Determination of the tare weight by weighing .................................................................... 31
10.2 Numerical input of tare (PRE-TARE) .................................................................................... 32
10.3 Numerical input of tare (PRE-TARE) .................................................................................... 35
11 Add up .......................................................................................................... 37
11.1 Add up “number parts".......................................................................................................... 37
11.2 Add up “weight" ..................................................................................................................... 38
11.3 Delete stored values .............................................................................................................. 39
12 Application menu......................................................................................... 39
12.1 Navigation in the menu.......................................................................................................... 39
12.2 Menu overview [USER] ........................................................................................................... 40
13 Configurations menu ................................................................................... 41
13.1.1 Display background illumination .......................................................................................... 41
13.1.2 Setting the display speed..................................................................................................... 42
14 Data output ................................................................................................... 43
14.1 RS 232C interface................................................................................................................... 43
14.2 Description of interface ......................................................................................................... 44
14.2.1 Example: Setting an output format....................................................................................... 48
15 Service, maintenance, disposal.................................................................. 49
15.1 Cleaning .................................................................................................................................. 49
15.2 Service, maintenance............................................................................................................. 49
15.3 Disposal................................................................................................................................... 49
16 Instant help................................................................................................... 50
17 Appendix ASCII CODE table ....................................................................... 51
CXB/CXP-BA-e-1016
4
1 Technical data
CXB models:
KERN CXB 3K0.2 CXB 6K0.5 CXB 15K1 CXB 30 K2
Readability (d) 0.2 g 0.5 g 1 g 2 g Weighing range (max) 3 kg 6 kg 15 kg 30 kg Reproducibility 0.2 g 0.5 g 1 g 2 g Linearity ± 0.4 g ± 1.0 g ± 2 g ± 4 g Stabilization time 2 sec. 2 sec. 2 sec. 2 sec. Recommended adjustment
weight, not added (class)
3 kg (M1) 5 kg (M1) 15 kg (M1) 30 kg (M1)
Weighing unit g g g g Minimum piece weight 0.1 g 0.2 g 0.5 g 1 g Warming up time (operat-
ing temperature)
30 min
Reference quantity freely selectable Net weight (kg) 4 kg Permissible ambient
condition
-10° C to 40° C
Humidity of air 15% - 85% (non-condensing) Weighing plate, stainless
steel
300 x 225 mm
Dimensions of the housing (B x D x H)
300 x 330 x 110 mm
Mains connection Mains adapter 230 V, 50/60 Hz ; 9 V DC balance, 800 mA
Without backlit display:
Service life c. 200 hours / loading time ca. 8 hrs.
Rechargeable battery
With backlit display:
Operating time ca. 60h. / loading time ca. 8 hrs.
CXB/CXP-BA-e-1016
5
KERN CXB
3K1M
CXB
6K2M
CXB
15K5M
CXB
30K10M
Readability (d) 1 g 2 g 5 g 10 g Weighing range (Max) 3 kg 6 kg 15 kg 30 kg Reproducibility 1 g 2 g 5 g 10 g Linearity 1 g 4 g 10 g 20 g Verification class III III III III Stabilization time 2 sec. 2 sec. 2 sec. 2 sec. Recommended adjustment
weight, not added (class)
3 kg (M1) 5 kg (M1) 15 kg (M1) 30 kg (M1)
Weighing unit g g kg kg Minimum piece weight 100 mg 200 mg 500 mg 1g Warming up time (operat-
ing temperature)
10 min
Reference quantity freely selectable Net weight (kg) 4 kg Permissible ambient
condition
-10° C bis 40° C
Allowable air humidity 15% - 85% (non-condensing) Weighing plate, stainless
steel
300 x 225 mm
Dimensions of the housing plastic material (B x D x H)
300 x 330 x 110
Voltage 220-240 V, 50/60 Hz;
Without backlit display:
Service life ca. 200Std. / loading time ca. 8 Std.
Rechargeable battery
with back light:
Operating time ca. 60Std. / loading time ca. 8 Std.
CXB/CXP-BA-e-1016
6
CXP models:
KERN CXP 30K2 CXP 75K5 CXP 150K10
Readability (d) 2 g 5 g 10 g Weighing range (max) 30 kg 75 kg 150 kg Reproducibility 2 g 5 g 10 g Linearity ± 4 g ± 10 g ± 20 g Stabilization time 2 sec. 2 sec. 2 sec. Recommended adjustment
weight, not added (class)
20 kg (M1) 50 kg (M1) 100 kg (M1)
Weighing unit g g g Minimum piece weight 0.5 g 1 g 2.5 g Warming up time (operat-
ing temperature)
30 min
Reference quantity freely selectable Net weight (kg) 8.9 kg Permissible ambient
condition
-10° C to 40° C
Allowable air humidity 15 % - 85 % (non-condensing) Weighing plate, stainless
steel
400 x 300 mm
400 x 300 x 100 mm (platform) Dimensions of the housing
plastic material (B x D x H)
290 x 140 mm (terminal)
Voltage 230 V (AC)
Without backlit display:
Service life c. 200 hours / loading time ca. 16 hrs.
With backlit display:
Operating time ca. 60h. / loading time ca. 16 hrs.
Rechargeable battery
with back light + RS 232:
Operating time ca. 56h. / loading time ca. 16 hrs.
Data interface RS 232C
CXB/CXP-BA-e-1016
7
2 Declaration of conformity
KERN & Sohn GmbH
D-72322 Balingen-Frommern Postfach 4052 E-Mail: info@kern-sohn.de
Tel: 0049-[0]7433- 9933-0 Fax: 0049-[0]7433-9933-149 Internet: www.kern-sohn.de
Declaration of conformity
EC-Konformitätserklärung EC-Declaration of -Conformity EC-
Déclaration de conformité EC-Declaración de Conformidad
EC-Dichiarazione di conformità EC-Conformiteitverklaring EC- Declaração de conformidade EC- Prohlášení o shode EC-Deklaracja zgodności EC-Заявление о соответствии
D
Konformitäts­erklärung
Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den nachstehenden Normen übereinstimmt.
GB
Declaration of conformity
We hereby declare that the product to which this declaration refers conforms with the following standards.
CZ
Prohlášení o shode
Tímto prohlašujeme, že výrobek, kterého se toto prohlášení týká, je v souladu s níže uvedenými normami.
E
Declaración de conformidad
Manifestamos en la presente que el producto al que se refiere esta declaración está de acuerdo con las normas siguientes
F
Déclaration de conformité
Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente déclaration, est conforme aux normes citées ci-après.
I
Dichiarazione di conformitá
Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferi­sce è conforme alle norme di seguito citate.
NL
Conformiteit­verklaring
Wij verklaren hiermede dat het product, waarop deze verklaring betrekking heeft, met de hierna vermelde normen overeenstemt.
P
Declaração de conformidade
Declaramos por meio da presente que o produto no qual se refere esta declaração, corresponde às normas seguintes.
PL
Deklaracja zgodności
Niniejszym oświadczamy, że produkt, którego niniejsze oświadczenie dotyczy, jest zgodny z poniższymi normami.
RUS
Заявление о соответствии
Мы заявляем, что продукт, к которому относится данная декларация, соответствует перечисленным ниже нормам.
Electronic Balance: KERN CXB/RXB/CXP
EU Directive Standards
2004/108/EC EN 61326-1:2006
EN 61000-3-2:2006 EN 61000-3-3:1995/A2:2005
2006/95/EC EN 61010-1:2001
Date: 12.07.2010 Signature:
KERN & Sohn GmbH Management
KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0 Fax +49-[0]7433/9933-149, E-Mail: info@kern-sohn.com, Internet: www.kern-sohn.com
CXB/CXP-BA-e-1016
8
3 Appliance overview
CXB models:
1. Bubble level
2. Battery compartment
3. Mains cable connection
4. ON/OFF switch
1
3
4
2
CXB/CXP-BA-e-1016
9
3
2
1
CXP models:
1. ON/OFF switch
2. Mains cable connection
3. RS 232 interface
4. Bubble level
CXB/CXP-BA-e-1016
10
Loading status display
3.1 Overview of display
CXB models:
Weight Reference weight
Loading status display Pieces
CXP models:
Weight Reference weight Pieces
CXB/CXP-BA-e-1016
11
3.1.1 Display weight
Here, the weight of your goods is displayed.
Overlay W indicates:
a
Zeroing display
PRE-
TARE
Tare in memory
Battery very low
3.1.2 Display reference weight
The reference weight of a sample is shown here. This value is either entered by user of calculated by balance.
Overlay W indicates:
Placed number of pieces insufficient for reference calculation
Placed reference weight insufficient for reference calculation
3.1.3 Display quantity
Here, all the parts placed on balance are immediately displayed by number.
Overlay W indicates:
M+
Data in summation memory
Stability display
3.1.4 Battery charge status display
red
Battery is almost discharged
green
Battery is completely discharged
CXB/CXP-BA-e-1016
12
3.2 Keyboard overview
CXB models:
CXP models:
CXB/CXP-BA-e-1016
13
Choice Function
-
Number keys
Delete key
Call counting with tolerance control
Store reference weights in memory
Call stored reference weights
Addition in total memory
Call up total memory
Delete summation memory
Enter reference weight through weighing
Display reference weight stored last
Enter target number of pieces
Numeric entry reference weight
Display reference weight stored last
Enter target weight
Zeroing key
Back to weighing mode
CXB models
Taring key
Enter numerical tare
CXP models
ON/OFF standby
Taring key
Enter numerical tare
CXB/CXP-BA-e-1016
14
4 Basic Information (General)
4.1 Proper use
The balance you purchased is intended to determine the weighing value of material to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the mate­rial to be weighed is manually and carefully placed in the centre of the weighing plate. As soon as a stable weighing value is reached the weighing value can be read.
4.2 Improper Use
Do not use balance for dynamic weighings. In the event that small quantities are re­moved or added to the material to be weighed, incorrect weighing results can be dis­played due to the “stability compensation“ in the balance. (Example: Slowly draining fluids from a container on the balance.) Do not leave permanent load on the weighing plate. This may damage the measuring system. Impacts and overloading exceeding the stated maximum load (max) of the balance, minus a possibly existing tare load, must be strictly avoided. Balance may be dam­age by this. Never operate balance in explosive environment. The serial version is not explosion protected. The structure of the balance may not be modified. This may lead to incorrect weigh­ing results, safety-related faults and destruction of the balance. The balance may only be used according to the described conditions. Other areas of use must be released by KERN in writing.
4.3 Warranty
Warranty claims shall be voided in case
Our conditions in the operation manual are ignored
The appliance is used outside the described uses
The appliance is modified or opened
Mechanical damage or damage by media, liquids, natural wear and tear
The appliance is improperly set up or incorrectly electrically connected
The measuring system is overloaded
CXB/CXP-BA-e-1016
15
4.4 Monitoring of Test Resources
In the framework of quality assurance the measuring-related properties of the bal­ance and, if applicable, the testing weight, must be checked regularly. The responsi­ble user must define a suitable interval as well as type and scope of this test. Infor­mation is available on KERN’s home page (www.kern-sohn.com with regard to the monitoring of balance test substances and the test weights required for this. In KERN’s accredited DKD calibration laboratory test weights and balances may be calibrated (return to the national standard) fast and at moderate cost.
5 Basic Safety Precautions
5.1 Pay attention to the instructions in the Operation Manual
Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances.
5.2 Personnel training
The appliance may only be operated and maintained by trained personnel.
6 Transport and storage
6.1 Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the appli­ance itself when unpacking for possible visible damage.
6.2 Packaging
Keep all parts of the original packaging in case you need to return the appliance. Only use original packaging for returning. Before sending, disconnect all connected cables and loose/movable parts. Attach possibly existing transport safeguards. Secure all parts, e.g. weighing plate, power unit etc., to prevent slipping and damage.
CXB/CXP-BA-e-1016
16
7 Unpacking, Setup and Commissioning
7.1 Installation Site, Location of Use
The balances are designed in a way that reliable weighing results are achieved in common conditions of use. You will work accurately and fast, if you select the right location for your balance.
Therefore, observe the following for the installation site:
Place the balance on a firm, level surface;
Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
Protect the balance against direct draughts due to open windows and doors;
Avoid jarring during weighing;
Protect the balance against high humidity, vapours and dust;
Do not expose the device to extreme dampness for longer periods of time.
Non-permitted condensation (condensation of air humidity on the appliance) may occur if a cold appliance is taken to a considerably warmer environment. In this case, acclimatize the disconnected appliance for ca. 2 hours at room temperature.
Avoid static charge of goods to be weighed or weighing container.
Major display deviations (incorrect weighing results) may be experienced should electromagnetic fields (e.g. due to mobile phones or radio equipment), static electric­ity accumulations or instable power supply occur. Change location or remove source of interference.
7.2 Unpacking
Carefully remove the balance from the packaging, remove plastic cover and setup balance at the intended workstation.
CXB/CXP-BA-e-1016
17
7.2.1 Setup
Level balance with foot screws until the air bubble of the water balance is in the pre­scribed circle.
Models CXP:
1 Weighing plate 6 Platform 2 Cantilever 7 Display part 3 Screws (x 2) 8 Support 4 Spring 9 Screws (x 4) 5 Substructure 10 Allen keys
A) Insert screw (3) in eye of spring (4) and screw it down.
Screw down second screw (3) as well.
B) Push the holder (8) into the guide rail of the display part (7). Attach the display part (7) to the weighing balance, using the four screws (9). Use an
Allen key (10) to tighten the screws.
Loading...