KEPCO JQE 6-10, JQE 100-1, JQE 55-2, JQE 75-1.5, JQE 6-10M Operator's Manual

...
O P E R AT O R ’ S M A N U A L
JQE 1/4 RACK
POWER SUPPLY
KEPCO INC.
An ISO 9001 Company.
IMPORTANT NOTES:
1) This manual is valid for the following Model and associated serial numbers:
MODEL SERIAL NO. REV. NO.
INCLUDES MODEL SUFFIX
MODEL
100W
JQE 6-10, JQE 15-6, JQE 25-4, JQE 36-3,
JQE 55-2, JQE 75-1.5,
JQE 100-1
M, DM
2) A Change Page may be included at the end of the manual. All applicable changes and revision number changes are documented with reference to the equipment serial num­bers. Before using this Instruction Manual, check your equipment serial number to identify your model. If in doubt, contact your nearest Kepco Representative, or the Kepco Docu­mentation Office in New York, (718) 461-7000, requesting the correct revision for your par­ticular model and serial number.
3) The contents of this manual are protected by copyright. Reproduction of any part can be made only with the specific written permission of Kepco, Inc.
Data subject to change without notice.
©2019, KEPCO, INC.
P/N 228-1730-R2
KEPCO, INC. z 131-38 SANFORD AVENUE z FLUSHING, NY. 11355 U.S.A. z TEL (718) 461-7000 z FAX (718) 767-1102
email: hq@kepcopower.com z World Wide Web: http://www.kepcopower.com
KEPCO®
THE POWER SUPPLIER™
KEPCO®
THE POWER SUPPLIER™
OPERATOR MANUAL CORRECTION JQE MODELS
1. Digital Meters JQE models with the suffix DM indicates that the unit has digital meters in place of the analog meters.
(NOTE: analog meters are no longer available.) V o ltag e meter accuracy is 1%; current meter accuracy is
1.5%. Meter range selector switches are no longer needed with the digital meters and have been removed on DM models. Refer to Figures 2, 3 and 4 for front panel views of the qu arter rack, half rack, and full rack DM models.
2, PAR. 1-9 ADD: b) Side Support “L” Bracket P/N 128-1775 mounts to vertica l rails to provide additional sid e support
for rack-mounted models.19.6 in. long x 2.3 in. high x 3.5 in. wide. Two (2) required per unit.
FIGURE 1. SIDE SUPPORT “L” BRACKET P/N 128-1775
3. SECTION III Correction The instructions throughout Section III are based on a value of the front panel VOLTAGE control (R102)
that requires an Ratio, of 1000 Ohms/Volt.
To allow for the fact that the value of R102 may now be different, use the following procedure to deter­mine the actual
1. Turn the unit off and disconnect unit from source power.
2. Remove the link between positions 5 and 6 of TB501 (for JQE quarter rack models: remove link between positions 6 and 7 of TB1).
3. Set the Voltage potentiometer on the front panel to full clockwise.
4. Use a precision digital ohmmeter to measure the resistance between positions 4 and 5 of TB 501 (for JQE quarter rack models: measure resistance between positions 7 and 8 of TB1). This is the actual resistance of the front panel VOLTAGE control (R102), referred to as R
5. Use the measured value of R102 to calculate Ib as follows: I
b
where: E R I
b
E
I
of 1mA and a corresponding CONTROL RATIO (Kc), also referred to as Programming
b
I
and the corresponding CONTROL RATIO (1/Ib )
b
VC Internal.
= (EO + (EO))/ R
is the nominal output voltage (Volts).
O
VC Internal
is the actual control current.
O
is the value of R102 measured in step 4 (Kilohms).
functions as voltage headroom to allow full scale from the front p anel control.
VC Internal
JQE-r2/030519
KEPCO, INC. 131-38 SANFORD AVENUE  FLUSHING, NY. 11355 U.S.A.  TEL (718) 461-7000  FAX (718) 767-1102
email: hq@kepcopower.com World Wide Web: www.kepcopower.com
1
1
6. Calculate the CONTROL RATIO as 1/I
131−1418
(Kilohms/Volt) using the value o f Ib calculated in step 5.
b
7. Use the calculated values of I
and CONTROL RATIO for all procedures of Section III.
b
As an example, referring to the Procedure: Precision Programming Ratio Adjustment, PAR. 16, step c), instead of R
= 5K Ohms, use RVC = 5 (1/Ib) (in Kilohms) in order to obtain an output
VC
voltage of 5V.
FIGURE 2. FRONT PANEL VIEW OF JQE QUARTER RACK DM (DIGITAL METER) MODELS
FIGURE 3. FRONT PANEL VIEW OF JQE HALF RACK DM (DIGITAL METER) MODELS
2
FIGURE 4. FRONT PANEL VIEW OF JQE FULL RACK DM (DIGITAL METER) MODELS
3/(4 Blank)
Conditions of Conformance
When this product is used in applications governed by the requirements of the EEC, the following restric­tions and conditions apply:
1. For European applications, requiring compliance to the Low V olt age Directive, 2006/95/EC, this powe r supply is considered a component product, designed for "built in“ applications. Because it is incom­plete in construction, the end product enclosure must provide for compliance to any remaining electri­cal safety requirements and act as a fire enclosure. (EN61010-1:2010, Cl. 6, Cl. 7, Cl.8, and Cl. 9)
2. This power supply is designed for stationary installation, with mains power applied via a detachable power supply cord or via direct wiring to the source power terminal block.
3. This power supply is considered a Class 1 (earthed) product, and as such depends upon proper con­nection to protective earth for sa fety from electric shock. (EN61010-1 Cl. 6.5.5)
4. This power supply is intended for use as part of equipment meant for test, measurement and labora­tory use, and is designed to operate from single phase, three wire power systems. This equipment must be installed within a suitably wired equipment rack, utilizing a three wire (grounded) mains con­nection. See wiring section of this manual for complete electrical wiring instructions. (EN61010-1 Cl.
6.5.5 and Cl.6.10.1)
5. For European applications requiring co mp lia nc e wit h the RoHS 2 directive, 2011/65/EU, this power supply is designed for scientific research and technological development, and is excluded from the scope of the directive [2011, 65/EU, Article 2 (4) (j)].
6. This power supply has secondary output circuits that are considered hazardous, and which exceed 240 VA at a potential of 2V or more.
7. The output wiring terminals of this power supply has not been evalu ated for field wiring and, th erefore , must be properly configured by the end product manufacturer prior to use.
8. This power supply employs a supplementary circuit protector in the form of a circuit breaker mounted on the front panel. This circuit breaker protects the power supply itself from damage in the event of a fault condition. For complete circuit protection of the end product, as well as the building wiring, it is required that a primary circuit protection device be fitted to the branch circuit wiring. (EN61010­1:2010, Cl. 9.6)
9. Hazardous voltages are present within this power supply during normal operation. All operator adjust­ments to the product are made via externally accessible switches, controls and signal lines as speci­fied within the product operating instructions. There are no user or operator serviceable parts within the product enclosure. Refer all servicing to qualified and trained Kepco service technicians.
228-1351 COND/CONFORM 030519 A
SAFETY INSTRUCTIONS
!
!
!
!
!
1. Installation, Operation and Service Precautions
This product is designed for use in accordance with EN 61010-1 and UL 3101 for Installation Category 2, Pollution Degree 2. Hazardous voltages are present within this product during normal oper ation. The product should never be operated with the cover removed un less equivalent protection of the operator from accidental contact with hazardous internal voltages is provided:
There are no operator serviceable part s or adjustments within the prod uct enclosure. Refer all servicing to trained service technician.
Source power must be removed from the product prior to performing any servicing.
This product is factory-wired for the nominal a-c mains voltage indicated on the rat­ing nameplate located adjacent to the source power conne ction on the product's rear panel. To reconfigure the product input for other nominal mains voltages as listed herein, the product must be modified by a trained service technician .
2. Grounding
This product is a Class 1 device which utilizes protective earthing to ensure operator safety.
The PROTECTIVE EARTHING CONDUCTOR TERMINAL must be properly con­nected prior to application of source power to the product (see instr uctions on instal­lation herein) in order to ensure safety from electric shock.
PROTECTIVE EARTHING CONDUCTOR TERMINAL - This symbol indicates the point on the product to which the protective earthing conductor must be attached.
EARTH (GROUND) TERMINAL - This symbol is used to indicate a point which is connected to the PROTECTIVE EARTHING TERMINAL. The component installer/ assembler must ensure that this point is connected to the PROTECTIVE EARTH­ING TERMINAL.
CHASSIS TERMINAL -This symbol indicates frame (chassis) connection, which is supplied as a point of convenience for performance purposes (see instructions on grounding herein). This is not to be confused with the protective earthing poin t, and may not be used in place of it.
3. Electric Shock Hazards
This product outputs hazardous voltage and energy levels as a function of normal operation. Operators must be trained in its use and exercise caution as well a s common sense during use to preven t a ccidental shock.
This symbol appears adjacent to any external terminals a t which ha zardous vo ltage levels as high as 500V d-c may exist in the course of normal or single fault condi­tions.
This symbol appears adjacent to any external terminals a t which ha zardous vo ltage levels in excess of 500V d-c may exist in the course of normal or single fault condi­tions.
B 228-1352 SAFETY - (COVER REMOVAL) 030519