KEPCO Firmware Upgrader Utiliy User Manual

INSTRUCTION SHEET
An ISO 9001 Company.
FIRMWARE UPGRADE UTILITY
I — INTRODUCTION
SCOPE OF MANUAL. This Instruction Sheet covers the
installation and use of Kepco’s Firmware Upgrade Utility.
DESCRIPTION. The Kepco Firmware Upgrade Utility allows selected Kepco products to be upgraded to the lat­est Firmware without replacing any parts. The Firmware Upgrade Utility is available from the Kepco web site at
www.kepcopower.com/utilities/util.htm
NOTE: Check the web page referenced above to verify
that your Kepco product and/or Firmware ver­sion is 1) eligible to use this method to upgrade the firmware and 2) note any requirements for the port used to perform the upgrade.
Firmware
Upgrade
III — OPERATION
After the Firmware Upgrade Utility has been installed, pro­ceed as follows:
1. At your PC go to Start - Programs - Kepco - Utilities ­Firmware Upgraded to run the Firmware Upgrade Utility.
2. At the top menu bar, click Bus Selection (see Figure 1) and choose RS 232, IEEE 488 (GPIB) or LAN as the interface bus to use during the upgrade process.
3. Depending on the bus selected, enter the following
a. RS 232: Select applicable COM Port and Baud Rate
(see Figure 1).
II — INSTALLATION
1. Download and open the zip file from
www.kepcopower.com/utilities/util.htm
2. Run the file called setup and follow the instructions to install the program on your PC. The upgrade utility is self-contained and does not require VISA to be installed on your PC.
3. Installing the Firmware Upgrade utility accomplishes:
• Start Menu updated with KEPCO folder and UTLI­TIES subfolder.
• Program Files Folder updated with KEPCO sub­folder and UTILITIES sub-subfolder containing all files required for upgrade, including a runtime exe­cutable for CVI that performs the upgrade without affecting prior installations of CVI.
4. Verify communication between the power supply and PC using either the RS 232, GPIB (IEEE 488) or LAN bus.
FIGURE 1. OPENING SCREEN, RS 232 SELECTED
KEPCO, INC. 131-38 SANFORD AVENUE FLUSHING, NY. 11355 U.S.A. TEL (718) 461-7000 FAX (718) 767-1102
http://www.kepcopower.com email: hq@kepcopower.com
©2010, KEPCO, INC 1 Data subject to change without notice 228-1686 REV 1
b. IEEE 488: Select applicable GPIB board and GPIB
address of the power supply (Figure 2).
FIGURE 2. GPIB (IEEE 488) SELECTED
c. LAN: Enter applicable IP Address of the power sup-
ply (see Figure 3).
FIGURE 3. LAN SELECTED
4. Enter the power supply password in the Unit Pass­word field.
5. Click Select Filename to locate and select the firm­ware code file (e.g., klpupgrade V9_00.hex).
6. Click Start. The Model, Serial No. and Firmware Ver- sion Before Update are displayed. The Messages box displays Erasing to indicate the upgrade has started. If an error message is displayed (e.g., wrong password, device not found, etc.) correct the problem and click Start again. If unable to correct the problem, contact Kepco.
7. As the upgrade progresses, the Transmit indicator lights periodically when data is being transmitted from the PC to the power supply, and % progress is indi­cated both numerically and graphically.
9. When Firmware Version After Update shows the new firmware version and the % Progress indicators show 100%, the update is complete.
FIGURE 4. UPGRADE COMPLETE
10.Click EXIT! at the top menu bar to exit the upgrade util­ity.
11.Download documentation applicable to the new firm­ware revision from the Kepco website at www.kepcopower.com/support/opmals.htm.
12.For a brief list of differences between the old firmware and the upgraded firmware, refer to the readme.txt file included with the Firmware Upgrade utility.
8. When the upgrade is near completion, the message Powering Up will be displayed.
KEPCO, INC. 131-38 SANFORD AVENUE FLUSHING, NY. 11355 U.S.A. TEL (718) 461-7000 FAX (718) 767-1102
http://www.kepcopower.com email: hq@kepcopower.com
2 228-1686 REV 1 033010