KEM SRZ Series, SRZ-10, SRZ-40, SRZ HR, SRZ-400 Operation Manual

...
Operation Manual
SRZ Helical Flow Meters
Küppers Elektromechanik
Operating Instructions for Helical Flow Meters Type SRZ
Contents
Liebigstr. 2 • 85757 Karlsfeld • Tel. +49 81 31/59 39 10 • Fax +49 81 31/9 26 04 ... Fehler! Textmarke nicht
definiert.
1 General Information .......................................................................................... 3
2 Getting started .................................................................................................. 5
3 Installation ....................................................................................................... 6
4 Operation......................................................................................................... 8
5 Service and Maintenance .................................................................................... 9
6 Listings ...........................................................................................................10
1.1 At a Glance................................................................................................................... 3
1.2 Safety........................................................................................................................... 3
1.2.1 General Safety........................................................................................................................................ 3
1.2.2 Special requirements for Ex installations .................................................................................................3
1.3 Ordering Codes and Accessories .................................................................................... 4
1.3.1 Ordering Code........................................................................................................................................4
1.3.2 Accessories ............................................................................................................................................4
2.1 Unpacking.................................................................................................................... 5
2.2 Operating Elements....................................................................................................... 5
2.3 Pin Assignments ........................................................................................................... 5
3.1 Mechanical................................................................................................................... 6
3.1.1 Installing the SRZ...................................................................................................................................6
3.2 Electrical ......................................................................................................................7
3.3 Ex installation .............................................................................................................. 7
4.1 Evaluation.................................................................................................................... 8
5.1 Maintenance................................................................................................................. 9
5.2 Trouble shooting .......................................................................................................... 9
5.3 Service .........................................................................................................................9
6.1 Warranty ....................................................................................................................10
6.2 Declaration of conformity.............................................................................................10
6.3 Technical Data.............................................................................................................10
6.3.1 General................................................................................................................................................ 10
6.3.2 Electrical Data...................................................................................................................................... 10
6.3.3 Ex Supply Data.................................................................................................................................... 11
6.4 Mechanical Drawings...................................................................................................11
6.5 WEEE 2002/96/EC........................................................................................................11
6.6 Address.......................................................................................................................12
Operating Instructions for Helical Flow Meters Type SRZ
1 General Information
1.1 At a Glance
The SRZ Helical flow meters are positive displacement meters with built-in hall sensors. They provide high resolution and reverse-flow detection.
The SRZ meters are perfect for media with high viscosity (30 to 10 x 10 accuracy and low pressure drop.
The measurement is pulsation free and insensitive to pulsating flows.
1.2 Safety
1.2.1 General Safety
All statements regarding safety of operation and technical data in this manual will only apply when the unit is operated correctly in accordance with this manual.
The data for Ingress Protection (IPxx) will only apply when all connectors are caped properly with the corresponding counterpart with the same or better IP rating. Cable glands must be populated with cables with the specified diameter and closed properly.
During operation all openings of the housing must be closed unless otherwise noted in this manual.
All connections to the load and to the supply must be made with shielded cables unless otherwise noted in this manual. This unit must be grounded.
1.2.2 Special requirements for Ex installations
When used in hazardous area a suitable separation amplifier or zener barriers must be used.
When using long cables make sure that the maximum inductances and capacitances for the respective voltage or gas group are not exceeded.
6
mm²/s) featuring high
KEM 3
Operating Instructions for Helical Flow Meters Type SRZ
1.3 Ordering Codes and Accessories
1.3.1 Ordering Code
SRZ ** –**–* *–*
O-ring material
V = Viton T = Teflon
Pickup
A, C, R= integral dual pickup HR = integral dual high-resolution pickup
Material, connections
ST = standard stainless steel, G 3/4"
Size, measuring Range
40 (HR) = 0.4 up to 8 ltr./min 40 (A, C. R) = 0.4 up to 40 ltr./min 100 (A, C. R) = 1.0 up to 100 ltr./min 400 (A, C. R) = 4.0 up to 400 ltr./min
1.3.2 Accessories
Ordering Code Short description
Connector M12; 5 pin, male; 713 cable, 5-core, black, 5 m with connector M12
EWS* Intrinsically safe power supply and separation amplifier
Operating Instructions for Helical Flow Meters Type SRZ
2 Getting started
2.1 Unpacking
Verify that you have received the following items:
SRZ Helical flow meter
Operational Manual
2.2 Operating Elements
1 = electrical Connector 2 = flow input 3 = flow output
2.3 Pin Assignments
Type Pin 1 Pin 2 Pin 3 Pin 4 Pin 5
SRZ-A + UB fA GND 2 * f f
B
SRZ-C + UB GND 2 * f PE SRZ-R + UB Direction GND 2 * f n. c. SRZ-HR + UB f
B
GND f
A
n. c.
+UB = 12 … 30V DC
SRZ-A is not available as Ex.
KEM 5
Operating Instructions for Helical Flow Meters Type SRZ
3 Installation
Note
All installations murst be carried out by qualified service personnel.
3.1 Mechanical
Make sure the measuring range of the meter cannot be exceeded by more than 20%. Purge your pipe system before installing the SRZ. Particles must not enter the SRZ as they could block the screw spindles. This is particularly important for the smallest size SRZ 10. As a precaution the use of the following filters for the measuring medium is recommended:
SRZ 10: .......... 120 micron
SRZ 40: .......... 200 micron
SRZ 100.......... 300 micron
SRZ 400 .......... 300 micron
3.1.1 Installing the SRZ
Mount the SRZ in a way that the main flow is in direction fo the arrow on the SRZ type plate. The SRZ works independent of the mechanical orientation. However, we recommend a vertical orientation with upwards going flow for low flow rates and low viscosities.
Connect the SRZ with the corresponding fittings (see chapter 6.4) and use the required sealants.
Note
Never use fibrous sealants such as hamp of teflon strip for sealing.
Operating Instructions for Helical Flow Meters Type SRZ
3.2 Electrical
Make sure that the flow meter is grounded.
Note
The housing of the SRZ* is connected to the shield of the cable. In bigger systems a separate grounding of the meter might be required for avoiding high currents in the cable shield. In that case the shield might be left open or grounded via a capacitance of e. g. 100nF to avoid ground loops.
Note
Inproper grounding and shielding may lead to bad EMC behaviour or danger to your health!
Connect the SRZ* via a shielded cable to the control unit. The SRZ is specified for a supply voltage of 15 V or 24 V and will operate at supply voltages from 12 V up to 30 V,
Note
Make sure that all cable and wires are connected and fixed properly before applying power to the SRZ*!
Figure 1: Output stage of the VTE*
The output driver for the frequency output is an asymmetrical push pull output stage with the protective resistors of R1 = 720 Ω and R2 = 470 Ω as shown in Note .
Note
The output can withstand a short circuit to ground for an infinite period whereas a short circuit to the positive supply for longer than a few seconds might destroy the output stage!
3.3 Ex installation
The SRZ*-Ex is intrinsically save when operated with an EWS or zener barriers as described in the technical data (see chapter 6.3)
Mount the SRZ* as described in chapter 3.1.
Connect the SRZ to the intrinsically safe power supply type EWS or to suitable the zener barriers.
Note
When used in hazardous areas the SRZ* must not be connected to any circuitry except our intrinsically safe power supply type EWS or suitable zener barriers.
KEM 7
Operating Instructions for Helical Flow Meters Type SRZ
K
4 Operation
The SRZ* is ready to use and does not require any special settings. After applying the supply voltage the unit immediately provides an output signal proporional to the flow.
4.1 Evaluation
The K-factor is individually determined for each SRZ, this factor specifies the exact pulse rate per litre. The K-factor can be found in the calibration record, which is supplied with every SRZ. The calibration record contains specifications to adjust the evaluation equipment.
The actual flow is calculated as follows:
f60
Q
with Q = flow rate in litres per minute f = output frequency in Hz K = K-factor of the SRZ in pulses per litre
=
Operating Instructions for Helical Flow Meters Type SRZ
5 Service and Maintenance
5.1 Maintenance
The SRZ* does not require regular maintenance.
For best performance, however, we recommend a recalibration every year if operated 24 hours a day or every 2 years if the unit is running less than 24 hours a day.
When used with aggressive or abrasive media the SRZ* should be checked and recalibrated even more often.
If there is no flow over a longer period, the SRZ* should be emptied and cleaned. This is especially important if curing media like paints are measured.
For recalibration and test, please contact your nearest dealer or directly KEM. For the addresses see chapter 6.6
5.2 Trouble shooting
In case the VTE* does not work properly, first check the following items:
No output signal All cables properly connected? Î connect the missing cables VTE* properly mounted? Î Check mounting position and, if necessary, mount the VTE* properly
Output frequency too low All cables properly connected? Îreconnect the loose cables Meter working properly? Î Check and, if necessary, replace the meter
Output frequency too high or unstable Most probably EMC problems Shield and ground properly connected? Î Connect shield properly. If necessary, try additional means of grounding and shielding
5.3 Service
The SRZ* does not contain any user serviceable parts.
In case of malfunction, please contact your nearest dealer or directly KEM. For the addresses see chapter 6.6
KEM 9
Operating Instructions for Helical Flow Meters Type SRZ
6 Listings
6.1 Warranty
KEM warrants material and production for a period of 18 months after installation and start up, max. 24 months from delivery date.
6.2 Declaration of conformity
See separate sheet
6.3 Technical Data
6.3.1 General
Type
size
SRZ 40 HR SRZ 40 SRZ 100 SRZ 400
Measuring Range
ltr./min
0,40 0,40 1,00 4,00
up to up to up to up to
100 400
40
K-Factor
pulses/ltr.
8
33,000
3,500
850 214
Frequency Range
in Hz
20
up to
20
up to
8
up to
14
up to
Measuring media: fluids free of ferro-magnetic particles Linearity: ±0.5% of actual flow at 30 mm²/s or higher ±0.25% of actual flow at 100 mm²/s or higher Repeatability: ±0.1% Operating temperature: –20 up to +70°C Materials: SS 303 or SS316Ti/SS316L Ingress protection: IP67
Measuring ranges and accuracy are valid for viscosities of 30 mm²/s of higher. K-factors and frequencies are average values – individual values are included in calibration records.
6.3.2 Electrical Data
Nominal supply voltage: 15 VDC and 24 VDC Supply voltage range: 12 up to 30 VDC Supply current: typically < 20 mA at no load Output type: Push pull Max output current: > 10 mA max. (typically 20 mA) Output short circuit to GND: infinite to +UB < 1 s
5,000 1,740 1,750 1,800
Operating Instructions for Helical Flow Meters Type SRZ
6.3.3 Ex Supply Data
Supply circuit (pin 1 and 3) Voltage Ui = DC 30V Current Ii = 120mA Power Pi = 850mW effective internal capacitance Ci = 8 nF
Signal current ciruit push/pull (pin 4 and 3) Voltage Ui = 30 V Current Ii = 24.6 mA Power Pi = 185 mW effective internal capacitance Ci = 8 nF
6.4 Mechanical Drawings
Type G L D PN
SRZ 40 SRZ 100 SRZ 400
¾" 1" 1½"
155 mm 221 mm 318 mm
85 mm 110 mm 134 mm
400 bar 400 bar 400 bar
6.5 WEEE and RoHS
The unit described herein is not subject to the WEEE directive and the corresponding national laws. At the end of life forward the unit to a specialised recycling company and do not dispoes it of f as domestic waste.
The unit described herein fully complies with the RoHS directive.
KEM 11
Operating Instructions for Helical Flow Meters Type SRZ
6.6 Address
Main office
KEM Küppers Elektromechanik GmbH Liebigstr. 2 DE-85757 Karlsfeld Germany
Phone +49 81 31/59 39 10
info@kem-kueppers.com www.kem-kueppers.com