Keller LEX1 Specifications

LEX 1
Digital precision manometer
Features
• High precision of up to 0,01 %FS
Insulated piezoresistive pressure sensor encapsulated in an oil-lled metal housing
• RS485 bus interface for communication with up to 128 devices
• Battery-operated with up to 2000 hours of battery life
• Can be used with external power supply
• License-free KELLER software available to download
• Intrinsically safe, explosion-proof version LEX 1 Ei available
Functions
• Wide range of units of pressure to choose from
• Zero point calibration via button
• Automatic shutdown
• Min/max display
User-dened units of pressure can be congured
• Digital calibration (zero point, end point)
Typical applications
• Calibration
• Laboratory use
• Industrial applications
Accuracy
± 0,05 %FS
Total error band
± 0,05 %FS
Pressure ranges
from 0...30 mbar to 0...1000 bar
LEX-1 with piezoresistive measuring cell LEX-1 with capacitive measuring cell
KELLER AG für Druckmesstechnik
CH-8404 Winterthur
+41 52 235 25 25 info@keller-druck.com
KELLER Ges. für Druckmesstechnik mbH
DE-79798 Jestetten
+49 7745 9214 0 eurocenter@keller-druck.com
KELLER America
Newport News, Virginia
1 877 253 5537 sales@kelleramerica.com
Subject to alterations Companies approved to ISO 9001
www.keller-druck.com
Page 2 of 6
LEX 1 – Specications
Standard pressure ranges
LEX 1 piezoresistive
Gauge pressure, PR
Absolute pressure, PAA
Absolute pressure, PA
Overload resistance
Display resolution
LEX 1 capacitive
Gauge pressure, PR
Differential pressure, PD
Overload resistance
Negative overload resistance
Display resolution
Key
Performance
-1…1 -1…3 -1…6 -1…10 -1…16 -1…30 bar rel.
0…2 0…4 0…7 0…11 0…17 0…31 0…61 0…101 0…161 bar abs.
8 8 20 20 90 90 300 300 600 600 800 1100 1100 bar
0,1 0,1 1 1 1 1 1 10 10 10 20 50 100 mbar
0…30 0…100 0…300
300 1000 1500 mbar
30 100 300 mbar
0,01 0,01 0,1 mbar
PR Gauge pressure Zero at atmospheric pressure PAA Absolute pressure Zero at 0 mbar abs. (vacuum) PA Absolute pressure Zero at 1000 mbar abs. PD Differential
0…300 0…400 0…700 0…1000 bar
mbar
mbar diff.
LEX 1 piezoresistive
Accuracy @ RT (20…25 °C) ≤ ± 0,05 %FS
Nonlinearity (BFSL), pressure hysteresis, non-repeatability,
zero point, amplication
Total error band (0...50 °C) ≤ ± 0,05 %FS Max. deviation within the specied pressure and temperature range
Long-term stability
Degree of dependency on location ≤ ± 1,5 mbar
0,1 %FS > 1 bar, per year under reference conditions, yearly recalibration recommended
1,0 mbar ≤ 1 bar, per year under reference conditions, yearly recalibration recommended
Calibrated in vertical installation position with pressure connection facing downwards
Accuracy of temperature measurement ± 1 °C typ.
Pressure range reserve ± 10%
Vacuum endurance ≤ 0,2 bar abs. Of operation ≤ 0,2 bar abs. upon request
LEX 1 capacitive
Accuracy @ RT (20…25 °C) ≤ ± 0,1 %FS
Nonlinearity (BFSL), pressure hysteresis, non-repeatability,
zero point, amplication
Total error band (0...50 °C) ≤ ± 0,2 %FS Max. deviation within the specied pressure and temperature range
Long-term stability 0,1 %FS Per year under reference conditions, yearly recalibration recommended
Long-term stability 30 mbar range 0.1 mbar Per year under reference conditions, yearly recalibration recommended
Degree of dependency on location none
Line pressure dependency (PD versions) ≤ ± 0,005 %FS / bar
Accuracy of temperature measurement ± 1 °C typ.
Pressure range reserve ± 10%
Line pressure ≤ 2 bar
KELLER AG für Druckmesstechnik CH-8404 Winterthur
+41 52 235 25 25 info@keller-druck.com
KELLER Ges. für Druckmesstechnik mbH
DE-79798 Jestetten
+49 7745 9214 0 eurocenter@keller-druck.com
KELLER America
Newport News, Virginia
1 877 253 5537 sales@kelleramerica.com
Subject to alterations Companies approved to ISO 9001
www.keller-druck.com
Page 3 of 6
LEX 1 – Specications
Performance
Optional for LEX 1 piezoresistive
Precision @ RT (20…25 °C)
Accuracy @ RT ≤ ± 0,01 %FS
For PA/PAA ≥ 10 bar, except for ush versions.
Accuracy ± 0,05 %FS, with KELLER test report ex works (standard)
Keller uses pressure sources to calibrate its products that are at least four times more accurate than the product to be tested. This enables us to produce products in our factory with an absolute accuracy of up to ± 0,05 %FS.
Precision ± 0,01 %FS / 0,025 %FS, with KELLER test report ex works
Additional measurement work and selection of a specic pressure transducer means that optimum repeatability
is guaranteed for selected pressure transmitters and digital manometers. Owing to the residual measurement
uncertainty of the pressure sources used at its factory, KELLER cannot provide any verication of measurement
accuracy at scales below ± 0,05 %FS for these ultra-precise devices. KELLER therefore uses the term "preci­sion" to denote the ability of a pressure transmitter or manometer to repeat measured values within a tolerance of 0,01 %FS based on the pressure sources used at the factory.
Accuracy ± 0,01 %FS with DakkS (German accreditation body) certicate,
issued by an external accredited calibration laboratory
By calibrating the zero point and performing amplication via the digital interface, an accredited calibration
laboratory (ilac.org) can adapt ultra-precise KELLER products to their more accurate pressure sources and record the results. External calibration to an accuracy of up to ± 0,01 %FS is performed in accordance with the guidelines set out by the German Calibration Service (DKD) and is conducted under reference conditions without any consideration of long-term effects.
≤ ± 0,01 %FS
≤ ± 0.025 %FS
Nonlinearity (BFSL), pressure hysteresis, non-repeatability, with KELLER test report ex works
With DakkS (German accreditation body) certicate issued by external calibration laboratory
Temperature ranges
Compensated temperature range 0…50 °C
Media temperature 0…50 °C
Ambient temperature 0…50 °C
Storage temperature -20…70 °C
Electrical data
Battery 3 V, type CR2430
Battery life approx. 2000 hours when used continuously
External power supply 8...28 VDC
Overvoltage protection and reverse polarity of external power supply
RS485 voltage insulation -7...+12 VDC
GND case insulation > 10 MΩ @ 50 VDC
External interface RS485 half-duplex
Interface measuring rate 30 measurements per second
Electrical connection Fischer D 103 A054-130
CE conformity as per 2014/30/EU (EMC)
± 32 VDC
EN 61000-6-1 to -6-4 EN 61326-1 / EN 61326-2-3
LEX1 Ei only permitted for use in haz­ardous atmospheres when used with the Renata CR2430
LEX 1 Ei devices must not be supplied with power from an external source and the RS485 interface must not be used in the Ex zone. See operating instructions for further information
KELLER AG für Druckmesstechnik CH-8404 Winterthur
+41 52 235 25 25 info@keller-druck.com
KELLER Ges. für Druckmesstechnik mbH
DE-79798 Jestetten
+49 7745 9214 0 eurocenter@keller-druck.com
KELLER America
Newport News, Virginia
1 877 253 5537 sales@kelleramerica.com
Subject to alterations Companies approved to ISO 9001
www.keller-druck.com
Page 4 of 6
LEX 1 – Specications
Display
Dimensions/appearance
Number of digits on LC display 2 rows with 5 digits each
Display mode Pressure + min/max or pressure + temperature
Measuring rate 2 measurements per second
Congurable units of pressure
Additional units of pressure 5 user-dened units can be congured
Width x height: 27,8 x 30,0 mm " see Dimensions & options
[bar], [mbar], [hPa], [kPa], [MPa], [PSI], [mH2O], [cmH2O], [inH2O], [ftH2O], [mmHg], [inHg], [kp/cm2]
Mechanical data
Materials in contact with media
Component LEX 1 piezoresistive LEX 1 capacitive
Pressure connection Stainless steel AISI 316L Stainless steel AISI 316L
Pressure transducer separating diaphragm Stainless steel AISI 316L Aluminium oxide 96%, gold-plated
Pressure transducer seal (internal) FKM (Viton® type A) Nitrile
Pressure connection seal (external) FKM (Viton® type A) FKM (Viton® type A)
Other materials
Component LEX 1 piezoresistive LEX 1 capacitive
Display housing Faradex AS-1003 Faradex AS-1003
Oil lling pressure transducer Silicone oil None
Further details
Component LEX 1 piezoresistive LEX 1 capacitive
Pressure connection G1/4, see Dimensions & options
Diameter x height x depth (approx.) 76 x 118 x 55 mm 76 x 148 x 55 mm
Weight (approx.) 300 g 335 g
Protection IP65
Explosion protection LEX 1 Ei
Intrinsically safe version in accordance with 2014/34/EU and IECEx
KELLER AG für Druckmesstechnik CH-8404 Winterthur
+41 52 235 25 25 info@keller-druck.com
PTB 05 ATEX 2012 X IECEx PTB 13.0028 X Zone 1: Ex II 2 G Ex ia IIC T6 Gb
KELLER Ges. für Druckmesstechnik mbH
DE-79798 Jestetten
+49 7745 9214 0 eurocenter@keller-druck.com
Permitted max. ambient temperature range -20…65 °C
KELLER America
Newport News, Virginia
1 877 253 5537 sales@kelleramerica.com
Subject to alterations Companies approved to ISO 9001
www.keller-druck.com
Page 5 of 6
LEX 1 – Dimensions & options
J
E
2 : 1
G
2 : 1
J
G
2 : 1
J
Fischer Flanschdose
DEE 103 A054-130
Binder Flanschstecker
Serie 723
Stiftplatte mPm-393
1
2
3
4
5
2
0
5
G1/2
1
4
1/4-18 NPT
1
5
2
0
2
0
5
1/4-18 NPT
1
5
2
0
2
0
5
2
0
2
0
5
6
1
5
3
7
°
43,5
1
4
1
6
G1/2
2
0
5
6
1
2
G1/4
G1/2
1
4
1/4-18 NPT
1
5
1/2-14 NPT
2
0
1
3
3
5
1
5
3
7
°
50,5
43,5
1
6
G1/2
2
0
5
6
G1/2
1
4
1/4-18 NPT
1
5
1/2-14 NPT
2
0
1
3
3
5
50,5
43,5
G1/2
2
0
5
6
1/4-18 NPT
1
5
1/2-14 NPT
2
0
1
3
3
5
G1/4
G1/2
2
0
5
6
1/2-14 NPT
2
0
1
3
3
5
G1/4
G1/2
2
0
5
6
1
3
3
5
G1/4
Display
Placement Display information
Width x height: 27,8 x 30,0 mm
Digit height: 8,4 pt. large 6,3 pt. small
External connection
Placement Connection Pin assignment
Fischer connector D 103 A054-130 Red Reference point
1 GND
1
5
2
4
3
2 n.c.
3 +Vcc
4 RS485 A
5 RS485 B
Available pressure connections
G1/4 (standard) G1/2 1/4 NPT 1/2 NPT G1/2 EN 837
0
2 1
G1/4
4 1
G1/2
5 1
1/4-18 NPT
0 2
1/2-14 NPT
G1/2 ush G3/4 ush 7/16-20 UNF Clamp DIN 32676 ush G1/4 EN 837
5 1
43,5
50,5
G1/2
G3/4
7/16-20 UNF-2A
3
7
°
Other pressure connections available upon request.
Other customer-specic options
• Other pressure and temperature ranges
• Parts that come into contact with media made from Hastelloy, Inconel or titanium
Customer-specic front covers
Integration of application-specic calculations
Customer-specic rmware
• Detached pressure transducer
• Axial pressure connection
• Other sealing materials
Other oil llings pressure transducer
2
5
G1/2
6
KELLER AG für Druckmesstechnik CH-8404 Winterthur
+41 52 235 25 25 info@keller-druck.com
KELLER Ges. für Druckmesstechnik mbH
DE-79798 Jestetten
+49 7745 9214 0 eurocenter@keller-druck.com
KELLER America
Newport News, Virginia
1 877 253 5537 sales@kelleramerica.com
Subject to alterations Companies approved to ISO 9001
www.keller-druck.com
Page 6 of 6
LEX 1 – Software & accessories
Swiss Made
«CCS30» software
Recording measured values
• Live visualisation
• Adjustable measuring and storage interval
• Export function
Conguration
• Call up of information (pressure and temperature
range, rmware version, serial number etc.)
«ManoCong» software
The ManoCong program is compatible with all
types of KELLER manometers and allows end cus-
tomers to congure the devices.
Range of functions
• Display of online measured values
Conguring the wait period before automatic shutdown
• Selecting standard pressure units
Interface
The LEX 1 manometer has a digital interface (RS485 half-duplex) which supports the KELLER bus proto­col. The pressure measurement rate via the inter­face is up to 30 x per second. Details of the commu­nication protocols can be found at www.keller-druck. com. Documentation, a Dynamic Link Library (DLL) and various programming examples are available to integrate the communication protocol into your own software.
• Activating/deactivating pressure units
User-dened pressure units can be programmed
• Restoring to factory settings
• Calibrating the manometer
Interface converters
The connection to a computer is established via an RS485-USB interface converter. Suitable converters
Software download
are available as accessories. To ensure smooth op­eration, we recommend the K-114 A converter with
www.keller-druck.com/software
the corresponding USB connector.
Scope of delivery
Plastic case CR2430 battery KELLER test report Operating instructions D/E/F
LEX 1
Aktueller Druckwert Actual Pressure Value Valeur de pression actuelle
Min.-/Max.-Druckwert Min.-/Max. Pressure Value Valeur de pression Min./Max.
ENTER Min./Max.
SELECT
Druckanschluss Pressure Connection Raccord pression
Hochgenaues digitales Manometer
Highly Precise Digital Manometer
Manomètre Numérique de Haute Précision
Accessories
Interface converters Calibration certicate Rubber cover Carry case
K-114 A
• With Fischer plug (5-pin)
• Various adapter cables available
K-114 BT A
• with Bluetooth interface and integrated recharge­able battery
• Wireless connection via
Serial Port Prole (SPP)
• 15 V measuring device
supply from the converter’s
internal battery
Issued by the external calibration laboratory of the German accreditation body DAkkS or the Swiss accredita­tion body SAS
For additional protection in harsh environments
With belt loop
KELLER AG für Druckmesstechnik CH-8404 Winterthur
+41 52 235 25 25 info@keller-druck.com
KELLER Ges. für Druckmesstechnik mbH
DE-79798 Jestetten
+49 7745 9214 0 eurocenter@keller-druck.com
KELLER America
Newport News, Virginia
1 877 253 5537 sales@kelleramerica.com
Subject to alterations Companies approved to ISO 9001
www.keller-druck.com
Loading...