Keller High-Pressure-Calibrator Operating Manual

HigH-Pressure-Calibrator
standard- and Full-Version
May 2006
KELLER
Included In delIvery: HIgH Pressure calIbrator standard-versIon
Included In delIvery: HIgH Pressure calIbrator Full-versIon
– 1 Calibrator
– 1 Test-Unit Adapter G 1/4” / G 1/2”
– 1 Test-Unit Adapter G 1/4” / G 1/8”
– 1 Spare Battery (3,6 V)
– 1 Allen Wrench
– 1 Manual
– 1TestCerticate
… in addition for version with
Logger-Option:
– K101-Cable (RS 232)
– 2 Program Discs:
• Logger Windows
• Logger MS-DOS
– 1 Calibrator
– 1 Test-Unit Adapter G 1/4” / G 1/2”
– 1 Test-Unit Adapter G 1/4” / G 3/8”
– 1 Pressure Switch Adapter
– 1 Adapter (4…20 mA / 0…20 mA)
– 1 Adapter (0…10 V / 0…20 V)
– 1 Mains Adapter
– 1 Spare Battery (3,6 V)
– 1 Allen Wrench
– 1 Manual
– 1TestCerticate
… in addition for version with
Logger-Option:
– K101-Cable (RS 232)
– 2 Program Discs:
• Logger Windows
• Logger MS-DOS
lIst oF contents:
Outline Drawing 1
General Description 2
Operating the Calibrator 3
– Turn-On
– Connecting Test-Units
– Executing Functions and Commands
– Pressure Generation
– Turn-Off
Description of the Functions 4
Measuring Mode 5
Mano-Mode
Temporarily Setting new Zero Reference 6
tara-Mode
Selection of Pressure Unit 6
unit-Mode
Preparing the Leak-Test 7
leak-Mode
Executing the Leak-Test 8
leak-Mode
Executing Storage of Measured Values 12
reC-Mode
Measuring the Linearity* 13*
lin-Mode*
Calculating the Linearity* 14*
lin-Mode*
Adjusting the Display-Resolution 15
reso-Mode
Permanently Setting new Zero Reference 15
Zero-Mode
Data Transfer to PC 16
Maintenance 17
– Priming the Pressure System
– Testing of Battery Capacity
– Battery Change
– Restarting the Calibrator
– External Supply of Test-Units
– Overow/OverpressureLimit
– Service
Operating Guide/Functions 18
Testing of Pressure Switches* 9*
Pst-Mode*
Preparing Storage of Measured Values 11
reC-Mode
Technical Data 19
Declaration of Conformity 20
* Full-Version only
outlIne drawIng
max. Oil-Level
KELLER
KELLER
10 12
13
14
11a
9815
7
6
5
4
32
1b
1a
11b
1a) Connector for RS 232 Output
(Standard Version)
1b)* Connector for RS 232 Output
(Full-Version)
2) Instrument Base (with battery compartment*)
3) * LED-Diode
4) HOLD-Key (H-Key)
5) Shut-off Valve
6) Display
7) Function-Key (F-Key)
8) Oil Chamber
 9) OilRellScrew
10) Oil Screw Compressor
High-Pressure-Calibrator Ranges: 200 bar, 350 bar, 700 bar (Sealed Gauge)
1
11a) Outlet Pressure Port (G 1/4”, female)
with Hexagon Plug (200 bar and 350 bar
Calibrator with Overpressure Valve)
11b) Outlet Pressure Port (G 1/4”, female)
with Hexagon Plug (700 bar Calibrator
without Overpressure Valve)
12) Priming Pump
13)* Input Plug for external Power Supply to
Transmitter/Switch under Test
14)* Electrical Connection for Test-Unit Adap-
ters (Transmitters or Pressure Switches)
15) Return Pipe
* Full-Version only
general descrIPtIon
The calibrator combines a number of functions
and performance features. Please take your
time reading this manual thoroughly prior to
setting the calibrator into operation. The safety
instructions on page 17 are meant to enhance
your safety and to maintain the functions as well
as the accuracy of the calibrator and its internal
sensor.
The instrument is battery powered, thus very
well suited for outdoor use.
The calibrator itself is powered by a 3,6 V
battery. Test-units (transmitters or electronic
switches) can either be powered by the internal
9 V batteries or external power supply.*
The high pressure calibrator generates pressure
with both the integrated priming pump and the
oilscrewcompressor.Thepressurecanbene
adjusted.
The high technology integrated in this instru-
ment allows it to measure precisely both the
pressure generated and the output signal of a
test-unit connected.
Thus, test-units can be measured and their
characteristics can exactly be documented.
The calibrator has an internal memory for recor-
dingdata. It can either belledwith the linea
rity measurements (function lin)*, the switching
points of a pressure switch (function Pst)* or
with continuously measured recordings. This
data can later be transferred to a PC for evalua-
tion via RS 232 serial interface.
The calibrator is operated with the function key
(F-key), marked by a pointed triangle.
An additional key, the Hold-key, will freeze the
display at any time during operation.
* Full-Version only
2
oPeratIng tHe calIbrator
turn-on
The calibrator is turned on by pressing the func-
tion key once.
connectIng test-unIts*
Adapter:
Included in the delivery of the calibrator are
different test-unit adapters, one for each of the
following transmitter types:
– 4…20 mA / 0…20 mA transmitter
– 0…10 V / 0…20 V transmitter
– Pressure switch
Plug the test-unit adapter into socket 14.
electrIcal connectIon oF test-unIt*
Select the correct adapter. Wire the transmitter/
switch under test to the adapter. Check that
polarity is observed and that good connections
are made.
Pressure connectIon For tHe test-unIt**
Your test-unit can be connected with the
calibrator via the pressure connection 11a resp.
11b.
Pressure generatIon
Attention: To avoid an overpressure within the
oilchamber,theoilrellscrew(9)mustbeopen
when operating the calibrator!
Generate pressure with the priming pump (12)
up to approx. 10 bar.
Increase or reduce pressure with screw com-
pressor (10).
decreasIng tHe Pressure
1) To return pressure to zero, completely
screw back compressor (10).
2) Open shut-off valve (5).
Do not open (5) when system is under high
pressure! (Especially when using an external
pres
sure source, damage/injury may result).
If you are not able to reach the desired pressure,
read the chapter “Maintenance” on page 17 on
how to completely prime the pressure system.
executIng FunctIons and commands
Read the following chapter to get acquainted
with the functions.
turn-oFF
Press the function key until the command oFF
appears on the display. Release the key imme-
diately and the instrument turns off.
** To avoid leakage and damage of the check valve, the test-unit has to be cleaned prior to connecting it to the
calibrator.
3
descrIPtIon oF tHe FunctIons
For a better understanding, the operating ele-
ments are described again:
FunctIon-Key (F-Key)
The function key on the front of the instrument,
marked by a pointed triangle, serves to operate
the instrument.
Hold-Key (H-Key)
The Hold-Key, integrated in the back of the
instrument housing, freezes the values during
the Mano-Mode. It has additional functions in
other modes (see following pages).
command structure
When keeping the F-Key pressed, the functions
exe, oFF, Mano, tara, unit, leak, Pst*, reC, lin*,
Zero, reso and ascending numbers appear suc-
cessively on the display. Releasing the F-Key at
a displayed function activates and leads into the
function. This cycle (constantly pressing the F-
Key and letting it go at a desired function) is cal-
led activation (e.g. to activate Mano --> Press
F-Key until display shows Mano --> Release the
F-Key).
exe: In Mano-Mode: To reset peak- and
trough-pressure.
oFF: To turn off the instrument.
mano: To display the actual pressure
generated, the units, peak- and
trough pressure.
tara: To set a new, volatile reference for
the zero point (Tare).
unIt: To display the pressure in different
units (bar, Psi, kPa, MWC…).
leaK: To measure the pressure changes
over a programmable time.
Pst*: To test pressure switches.
rec: To record the measured values in
programmable intervals.
lIn*: To determine the linearity of a
test-unit.
Withinspecicfunctions,theFKeycanexecute
additional commands (exe, steP, reset).
Activating oFF turns off the instrument. When
starting the instrument again, the display will
automatically lead into the function from which
the instrument was previously turned off.
Activate Mano if normal operation is desired.
reso: To reduce the resolution by
factor 10 or to return to the
original resolution.
Zero: To write a new zero point into the
non-volatile memory.
28, 29… Ascending numbers (have no
function).
* Full-Version only
4
mano-mode
The Mano-Mode is the standard mode of the ca-
librator. In this mode, the calibrator shows on
the upper display the actual pressure generated.
The smaller displays below indicate the peak-
and trough values. The measured values on the
smaller displays always appear with a reduced
resolution. Activating exe resets the peak- and
trough values, meaning that they are reset to the
actual pressure.
Full-Version only: If a transmitter is connected,
the display automatically indicates the value
of the transmitter in mA or V on the lower left
display. The lower right display shows the tem-
perature.
Pressing the Hold-Key (H-Key) freezes the
measured values. The actual pressure is now
shown on the lower right display. Pressing the
H-Key again leads back into the Mano-Mode
with the peak- and trough values reset to the
actual pressure (same as exe).
Actual Pressure
TroughPeak
Press H-Key
Hold-mode
Frozen Value
Actual PressurePeak
With H-Key back to MAno-Mode
mano-mode
disPlay WitH transMitter (Full-Version only)
Actual Pressure
mano-mode
disPlay WitHout transMitter
Actual Pressure
TroughPeak
Activate ExE
= Is being read as a “1” between 100 bar and 200 bar when displaying with increased resolution --> 133.41 bar
Signal of
Transmitter
Hold-mode
Frozen Value
of Transmitter
Press H-Key
Temperature of the Reference­Sensor
Frozen Value
Actual Pressure
5
tara-mode / unIt-mode
The tara-Mode serves to set a temporary zero
reference. Unlike to the Zero-Mode, this new
zero reference will not remain after the instru-
ment has been turned off.
Activate tara:
Value to be set to zero
Actual
Pressure
Activate ExE
mano-mode
The unit-Mode allows the selection of one of
four pressure units.
Activate unIt:
Activate ExE
Activate ExE
With the command exe the new zero is being
determined and stays as a reference until a new
tara is being executed or until the instrument is
turned off.
The program returns to the Mano-Mode.
Activate ExE
Activate Mano to return to the Mano-Mode with
the new selected unit.
Please note that, depending on pressure range,
the units can differ from the above.
6
leaK-mode
The leak-Mode serves to measure a pressure
change over a programmable time. The unit to
be tested has to be connected to the calibrator.
Setting the Test Time:
Activate leaK:
Actual
Pressure
Activate SEl
Flashing. exe changes
to hr (hours),
exe again to day.
Activate StEp
Test Time
Seconddigitisashing,
exe encreases digit… With steP to next digit.
When the test time is set, activate PreP. The
display is now back in PreP-Mode.
The test time is now programmed (35:00). The
start of the leak-test is described on the next
page.
Firstdigitisashing,exe
encreases digit. With steP to next digit.
Activate StEp
7
leaK-mode
Start of the leaK-Test:
After setting the test time, the display is in the
PreP-Mode:
Actual
Pressure
Press H-Key or activate Run
Test Time
During Test:
Pressure Change
Time counting down to zero
Pressure at
test start
After the test, the measured values are frozen in.
Activate PreP to return to the PreP-Mode from
the end display.
A new test can be started as follows (also when
measurements are running).
– Press the H-Key or
– Activate run with the F-Key
At any time, you may return to the Mano-Mode
by activating Mano.
Activating Mano interrupts a running leak-test.
After Test:
Pressure at
test start
Pressure change after the test
Selected test time
8
Pst-mode
(Full-Version only)
This mode allows the testing of mechanical and
electronic switches. The display always indi-
cates the switch status and the switch pressure
points. The actual status (oPen/Close) and the
switch pressure are being displayed with each
change of status in the pressure switch.
The actual pressure is always indicated on the
top display.
All switch points are being recorded and can be
written into a PC and displayed on the screen
(only with Logger-Option). Activating end followed
by the activation of another function quits the
Pst-Mode and leads into the new selected func-
tion.
Activate Pst:
The pressure switch closes. The status and the switch pressure are displayed.
The pressure switch opens. The status and the switch pressure are displayed.
Press H-Key
(recording of switch pressure points ends)
Connect pressure switch.
Press H-Key or activate Run
(recording of switch pressure points starts)
Actual Pressure
9
Pst-mode
(Full-Version only)
Note:
The test measurements can be started in each
position of the connected pressure switch.
In the stoP mode, a new switch can be connected
withoutinuenceonstatus
or recording.
Press H-Key
New start of measurement
Back to the Mano-Mode:
Activate end -->
The rst point is automatically written into the
small right display (Low) if the measurement
started at high pressure and the pressure switch
was switching at a lower pressure.
An increase of pressure from this status writes
the new switch point automatically into the
small left display (High). The pressure decrea-
sing again results in writing the new switch point
into the Low again.
This allows various switch points to be regis-
tered from the same pressure switch. They are
being recorded in the memory in pairs.
Activating run can start a new measurement at
any point of the function.
Activate MAno
10
rec-mode
In the reC- (or Record) Mode, the measured
values can be stored. The number of measure-
ments and the interval (in minutes and seconds)
are programmable.
Subsequently, one storage cycle over the num-
ber of measurements is called a “recording”.
Full-Version only:
If no transmitter is connected, only the actual
pressure will be stored when recording. When
recording with a connected transmitter, three
values will be stored per recording: Actual pres-
sure, signal of transmitter and temperature.
Please note that when recording with three
values per recording, the storage requirements
are tripling (1000 measurements with three
values each lead to 3000 measuring values).
Firstdigitisashing;
exe increases digit.
With steP to next digit.
Activate StEp
Seconddigitisashing;
exe increases digit.
With steP to next digit…
Adjustment of the Storage Parameters:
Activate rec:
Recording
Number of recordings
Activate SEl
interval
11
After setting the number of recordings, set the
recording interval in min:sec in the same way.
The adjustment of the storage parameters can
be terminated by activating PreP.
The display is now in the PreP-Mode. The newly
set storage parameters are shown in the dis-
play:
Recording
Number of recordings
interval
– The start of the recording is described on
the next page.
rec-mode
Start of the Recordings:
After the adjustment of the storage parameters,
the display is in the PreP-Mode:
Recording
Number of recordings
Activate Run
interval
During Recording:
Actual Pressure
Time remaining until next recording
Remaining recordings
(descending)
(descending)
The display end indicates the end of the recor-
ding. Activate PreP to return to the PreP-Mode
from the end display.
At any time, a record can prematurely be termi-
nated by activating res.
In the reC-Mode you may also display the tem-
perature.
Pressing the H-key at PreP display indicates the
temperature on ther lower right display.
To quit the reC-Mode, activate PreP, followed by
the activation of any function.
After Recording:
Number of recordings
Recording interval
12
lIn-mode
(Full-Version only)
The linearity mode allows the determination of
the linearity error of a test-unit. The linearity is
calculated as terminal linearity, that means, the
lineissetbetweentherstandthelastmeasu-
ring point.
The linearity error is indicated as a percen-
tageofthespanbetweentherstandthelast
measuring value. Maximal 10 linearity values are
allowed.
After having activated lin, activating the com-
mand saVe writes the actual values of the
reference pressure and of the test-unit into the
memory. This cycle can be quit by activating
saVe.
The Linearity Test:
Activate lIn:
Output signal
of test-unit
Actual Pressure
Number of measuring point
Press H-Key or activate SAvE
Values are written into the memory
New number indicates: Values are registered
Pressure increase >>>
Press H-Key or activate SAvE
13
Second value is registered
lIn-mode
(Full-Version only)
Calculating the Linearity:
After registration of the measuring points, the
calculation of the linearity is initiated with the
command CalC.
After the calculation, the pressure and measu-
ring point with the highest linearity error are
shown:
Actual Pressure
Number of last measuring
Value of
test-unit
Activate cAlc
Max. linearity
error in %
point
Pressure at max. linearity error
Number of measuring point, where error is maximal
From this position, the linearity of all other
measuring points can be called upon:
Pressure at point 3
Activate ExE
Pressure at point 4
Linearity error at point 4
Activate ExE to go to points 5, 1, 2…
Activate pREp for new linearity test
14
reso-mode / Zero-mode
reso-Mode:
The activation of reso reduces the resolution by
factor 10 or sets the display back to the original
resolution.
Activate RESo
Activate RESo
Zero-Mode:
Ambient pressure changes, dependency of posi-
tionorthe inuence of temperature can result in
zero shifts.
These shifts can be permanently corrected as
follows.
Activate Zero:
Actual
Zero
Activate ExE
15
The new zero is set and permanently stored.
This zero will be maintained when turning off the
instrument.
The programme automatically returns to the
Mano-Mode after setting the new zero.
data transFer to Pc
data transFer to Pc
(only with Logger-Option)
In order to transfer the data from the calibrator
into a PC, a special cable (K101) and a special
software is required. The transfer of the data
is accomplished over a RS 232 serial interface
connected to the calibrator via socket 1a or 1b.
logger-soFtware
This software serves to transfer the data to a
PC, offering many further processing possibili-
ties of the read-out data.
The LOGGER-Software meets all the require-
ments of modern Windows-Software:
– Graphical Display
– Tabulated Display
– Export into other Windows-Programs
– Storage into Files
tIme allocatIon
After each recording (lin*, Pst*, reC), the calib-
rator writes the status of the calibrator’s time
meter into the memory. When transferring, the
absolute time-axis is calculated from the time
set in the PC, the time meter of the calibrator
and the recorded time. The calibrator should
therefore not be turned off between the recor-
ding and the transfer.
During the recording process, ve additional
recording locations for the storage of time are
used for each complete recording cycle.
memory
Each of the functions lin*, Pst* and reC ll the
memory. New data is always stored after the last
recordedvalue,continuouslyllingthememory.
Since these instruments feature a ring-memory,
new data will only overwrite the oldest data. This
ensures that the maximal amount of the most
recent data is always at your disposal when the
memory is full.
The transfer of the data always starts at the last
recording and can be extended to read out the
entire memory. The memory may be read out as
many times as needed.
total storage locatIons
Standard: 900
Extended: 8000
(only with memory option)
* Full-Version only
16
maIntenance
PrImIng tHe Pressure system
Completely release pressure. Open valve (5) and
screw (9). If the pressure is not zero, activate
rsttara and then exe.
Purge air from the system by repeatedly pressing
the priming pump (12). The system is free of air
when air bubbles cease coming out of the return
pipe (15). Close shut-off valve (5).
batterIes
The calibrator houses a 3,6 V battery behind the
display for the calibrator functions, and two 9 V
batteries* for the test-unit supply.
3,6 v battery
The calibrator will not indicate a battery low. If
the display starts to fade, it is an indication that
the battery charge is weakening. At this point,
we recommend changing the battery. The bat-
teries can be acquired from KELLER.
3,6 v battery cHange
Turn the display ring beyond the limit stop. It will
detach from the main housing. Disconnect the
battery connections and remove the old battery.
Insert the new battery. Set the display ring back
in place.
2 x 9 v battery cHange*
The LED diode is also an indicator for the test-
unit supply. The batteries have to be changed
if the LED does not illuminate when connecting
your test-unit. Open the battery compartment of
the instrument base and change the batteries.
restartIng tHe calIbrator
If it should occur that the programme is locked,
meaning that the instrument does not react
anymore when operating the F-Key, in most
cases an interruption of the power supply can
activate the instrument again.
Just follow the procedure described for the 3,6 V
battery change and disconnect the instrument
from the battery for at least 20 seconds. After
reconnection, start the instrument again.
external test-unIt suPPly*
An external supply can be plugged to connec-
tion (13), e.g. 13 V for KELLER transmitters.
The adapter plug, included in the delivery, auto-
matically interrupts the 9 V battery supply.
When testing mechanical switches, the mains
adapter should be used.
overFlow/overPressure lImIt
The calibrator displays the pressure until 5 to
10% above the indicated pressure range. If this
range is being exceeded, the display indicates
oVFl” (overow). Do not increase the applied
pressure any more!
If the pressure exceeds the indicated pressure
range by more than 20%, the sensor may be
destroyed.
servIce
KELLER pressure calibrators are maintenance-
free. The cycle for recalibration depends on the
application conditions. Recommended recali-
bration cycle: 1 year.
17
* Full-Version only
oPeratIng guIde / FunctIons
turn-on
The calibrator is turned on by pressing the func-
tion key (7) once.
FunctIons and commands
The functions and commands are activated
with the F-key (activation: release F-key when
desired command is shown on display).
Pressure generatIon
Generate pressure with the priming pump (12)
up to approx. 10 bar. Increase or reduce pressure
with screw compressor (10).
decreasIng tHe Pressure
Completely screw back compressor (10).
Open shut-off valve (5).
Do not open when system is under high pres-
sure! (Especially when using an external pres-
sure source, damage/injury may result)
connectIng transmItters*
Plug in transmitter adapter. Internal supply is
OK, when LED (3) lights. Connect the transmit-
ter. The lower left display now indicates the out-
put of the transmitter in mA or V.
connectIng Pressure swItcHes*
Plug in the pressure switch adapter. Internal
supply is OK, when LED (3) lights.
Connect the pressure switch to adapter.
FUNCTIONS
mano:
Display of pressure, max. and min. pressure. exe
resets peak- and trough values.
With test-unit connected*: Display of actual
pressure, transmitter signal and temperature.
Hold:
Activated with the HKey; display of pressure
and of test-unit signal are frozen and the actual
pressure is indicated on lower right display.
Pressing H-Key again leads back to Mano-mode
with max./min. values reset.
tara:
Sets the pressure display to zero (Tare).
Activate tara, execute with exe (page 6).
unIt:
Setting the pressure units (page 6).
leaK:
Pressure changes over time (page 7/8).
Pst*:
Testing of pressure switches (page 9/10).
rec:
Recording of measured values (page 11/12).
lIn*:
Determining test-unit linearity (page 13/14).
reso:
Reducing/increasing resolution (page 15).
Zero:
Setting a new permanent zero (page 15).
* Full-Version only
18
tecHnIcal data
Total Accuracy of displayed Pressure
(1)
Total Accuracy of displayed Test Unit Signal
Resolution of Display
200 bar Calibrator
350 bar Calibrator
700 bar Calibrator
Overpressure
Selectable Pressure Units
Measurement Cycle
Displayed Temperature
Storage Temperature
Operating Temperature
Compensated Temperature Range
Air Humidity
Memory
Power Supply of Calibrator-Electronics
Lithium-Battery-Life
Power Supply of Transmitter
Total Oil Chamber Volume
Oil Type
Dead Volume
Total Weight, including Batteries
Dimensions (L x W x H)
±0,1% FS (Full Scale), from 0…50 °C
±0,1% FS ±1 Digit
10 mbar
100 mbar
100 mbar
FS + 10%
BAR / PSI / kPA / mWC (others on request)
2 Measurements per Second
Temperature of Reference Sensor in °C
-20…60 °C
0…50 °C
0…50 °C
5…95% Relative Humidity
900 Storage Places (8000 optional)
Lithium-Battery 3,6 V (not rechargeable)
> 200 Days at continuous operation
2 x 9 V Batteries (not rechargeable)
57 ml
Type HLP 22 BP
7,7 ml
≈3,9kg
315 (337) x 153 x 205 mm
Protection (only Standard)
(1)
Includes linearity, repeatability, hysteresis, temperature error and resolution of the display
IP 65
19
declaratIon oF conFormIty
Herewith we declare, that the following products or product range
HIGH PRESSURE CALIBRATORS
Standard- and Full-Version
according to drawing 80139.3
meet the basic requirements for the electromagnetic compatibility, which are established in the guide-
lines of the European Community (89 / 336 / EWG).
This declaration is valid for High Pressure Calibrators, Standard- or Full-Version, of which a drawing,
marked with the CE sign, exists, and which are produced according to this drawing. This explanation
has no validity without such a drawing.
As criteria for the electromagnetic compatibility, the following norms are applied:
EN - 61000 - 6 - 1
EN - 61000 - 6 - 3
This declaration is given for the manufacturer
Keller AG für Druckmesstechnik
St. Gallerstrasse 119
CH - 8404 Winterthur
in full responsibility by
Keller GmbH
Schwarzwaldstrasse 17
D - 79798 Jestetten
Jestetten / May 22, 2006
H.W. Keller
Geschäftsführer
with legally effective signature
20
www.keller-druck.com
Loading...