Autonomous DAtA ColleCtor
ca. 208 mm
smAllest DiAmeter – Ø 16 mm
The DCX-16 is an autonomous, battery-powered data collector in a stainless steel housing with a very small diameter of only 16 mm. The housing and the pressure sensor element are comple­tely welded in, so sealing rings are eliminated at this point. In applications where a small probe diameter is an advantage, the logger can record the water level (pressure) and the tempe-rature over long periods.
As well as the battery compartment with its double seal, the small-diameter (16 mm) submersible sensor includes electronic circuitry featuring the latest microprocessor technology. It records the pressure and temperature of the medium with high measurement accuracy and resolu-tion, and it uses a mathematical model to correct any linearity or temperature errors made by the pressure sensor. High data reliability is guaranteed thanks to the use of a nonvolatile data memory.
The various conguration options allow the data logger to be adapted to the measuring point so that only specied events will be detected and stored. Event-controlled recording and log-inter-
val recording can be set independently of one another. In addition, installation data and com­ments on the measuring point can be stored in the logger. Installation is fast and simple with
xing disks of various sizes which can be tted as options: these match locking units (caps)
from different manufacturers for levels of 1” or more (2” or more with light plummet aperture), so measuring points can be implemented at a fraction of the previous cost. Three versions of the
data collector are available:
DCX-16
The sensor, electronics and battery are accommodated in one housing. To extract data, the
data collector must be removed from the measuring point, and the watertight screw cap that allows access to the read-out plug/interface must be unscrewed. The DCX-16 operates with an
absolute pressure sensor. In shallow water, when the inuence of air pressure uctuations has
to be taken into account, a second logger (barometric logger, obtainable separately) positioned on the surface of the water must be used to record the air pressure progression. The com-puter software then calculates the differential pressure or the water level by subtracting the two me­asured data.
DCX-16 SG/VG
Instead of a watertight screwed closure, these versions have a cable output. The interface plug
is xed on the sounding tube above the surface of the water with the help of a screw-on xing
disk. This means that there is no need to remove the DCX-16 SG/VG from the sounding tube in order to read the data. In the VG version (reference pressure measurement), the reference pressure-compensating ca­pillary is routed in the same cable into the read-out plug housing; this also contains the reference
aperture (protected by a Gore-Tex® diaphragm) which produces the pressure equalization.
DCX-16 (sG/VG)
Version DCX-16
Version DCX-16 VG
M20x1,5
Thread for screw-on
xing disks
KELLER AG für Druckmesstechnik CH-8404 Winterthur
+41 52 235 25 25 info@keller-druck.com
DCX-16 SG/VG DCX-16
±1
82
Venting Holes for VG-Version
±1
192,5
KELLER Ges. für Druckmesstechnik mbH
D-79798 Jestetten
+49 7745 9214 0 eurocenter@keller-druck.com
PE cable reference
Ø 5,8 / 5-pole
Mark of the diaphragm position on the level sensor
27 mm
Ø16
Screw cap/ interface plug
KELLER America
Newport News, Virginia
1 877 253 5537 sales@kelleramerica.com
Electronics- resp.
battery housing
ø 16 mm
Mark of the diaphragm position on the level sensor
Subject to alterations Companies approved to ISO 9001
www.keller-druck.com
Page 2
Specications
Pressure Ranges 10 mWC 20 mWC 50 mWC 100 mWC
DCX-16 PAA 0,8…2 0,8…3 0,8…6 0,8…11 bar abs. DCX-16 SG PAA 0,8…2 0,8…3 0,8…6 0,8…11 bar abs. DCX-16 VG PR 1 2 5 10 bar
Overpressure 2 x Pressure Range
System length* 10 m 20 m 50 m 100 m
PAA: Absolute. Zero at vacuum PR: Vented Gauge. Zero at atmospheric pressure (other ranges on request) *The system length can be selected (optional); for lengths of 100 m or more, use an anchor clamp. Tolerance for system length: ≤ 10 m: ± 2 cm; > 10 m: ± 1% of system length
Supply Lithium battery 3,6 V (Type AAA) Battery Life * 4 years @ 1 measurement/hour Output RS 485 digital
Electrical Connection Fischer DEE 103A054
Pressure Sensor Specications
Comp. Temperature Range -10…40 °C
Accuracy typ. 0,05 %FS Error Band *** (-10…40 °C) 0,1 %FS Resolution max. 0,0025 %FS
Long Term Stability typ. 1 mbar
Temperature Compensation -10…40 °C (others on request)
Temperature Measurement Accuracy typ. ± 1 °C Shortest Measuring Range 1x per second Memory 2 MBit: 57’000 measuring values @
storage measuring
ed time) Material Stainless steel AISI 316L
O-Ring: Viton
interval 15 s, otherwise 28’000
values (always with attribut-
®
Cable PE cable
Probe Weight 150 g (without cable) Options Other pressure connections, larger data memory material
, different accuracy, other
: e.g. Hastelloy or Titanium
* exterior inuences could reduce battery life ** Linearity + Temperature Error *** Includes Linearity + Repeatability + Hysteresis
LOGGER 5
The Logger 5 software makes it possible to congure and read autonomous KELLER data loggers. This software assists users during measure­ments in the eld, with processing the data and also with forwarding it to partners or end customers.
Measurement data can be displayed in graphic form, exported, compensated for air pressure or converted into different units. The online function
displays the current device values.
The software is included in the scope of delivery for the interface converter cables, or it can be downloaded free of charge at www.keller-druck.com.
• Supports Windows operating systems
Overview of functions: Logger 5
- Pressure and temperature channels, selectable
- Adjustable measurement interval (1s…99 days)
- Averaging with selectable number of measurements
- Recording modes:
• continuous interval measurement
• event-controlled recording:
· recording starts when value is exceeded
· recording starts when value is undercut
· recording starts when value changes → combination of continuous and event-controlled recording is possible
- Adjustment of pressure zero point
- Start measurements immediately or at a set time
- Data storage: linear or ring-type memory
- Battery status display
- Online display of measuring channels
- Management of notications and images for stations
Processing and forwarding measurement data
- Graphic display of measurement data
- Simple export of measurement data and graphics (supports Microsoft Ofce and these le formats:
CSV-1, CSV-2, XML, Hydras, TNO, Wiski, BNA)
- Generation of measurement reports
- Station information stored in SQ Lite database
KELLER AG für Druckmesstechnik CH-8404 Winterthur
+41 52 235 25 25 info@keller-druck.com
KELLER Ges. für Druckmesstechnik mbH
D-79798 Jestetten
+49 7745 9214 0 eurocenter@keller-druck.com
KELLER America
Newport News, Virginia
1 877 253 5537 sales@kelleramerica.com
Subject to alterations Companies approved to ISO 9001
www.keller-druck.com
Loading...