Keller Acculevel Data sheet

HigH AccurAcy SubmerSible level TrAnSmiTTer
THermAlly-compenSATed, cuSTom rAngeS, duAl ouTpuT SubmerSible
Acculevel
mitters, including standard ±0.25% FS or optional ±0.1% Total Error Band (TEB)3 accuracy. This level transmitter
is approved to NSF 61 and NSF 372
standards for water quality and includes Keller America’s guaranteed
7
lightning protection, making the Acculevel an outstanding value for liquid level measurement.
The ability of the Acculevel to provide this level of sustained performance over a wide range of operating
conditions makes it ideally suited to an equally wide range of applications, including drinking water, stream and
reservoir level, environmental monitoring, tank level, wastewater, and for installations in areas prone to chronic
damage due to transients caused by lightning.
For more information on the Acculevel, or any other Keller product, please contact Keller America, or view the
entire Keller catalog at http://www.kelleramerica.com/datasheets.html.
FEATURES
NSF 61 / NSF 372 approved construction for use in drinking water applications
7
4...20mA models include guaranteed lightning protection at no additional cost.
16-bit internal digital error correction for cost-effective low Total Error Band (TEB)
3
316L stainless construction standard - Optional Titanium for severe applications.
2-year warranty covers defects in materials and workmanship.
User-rangeable analog output ensures compatibility as requirements change.
RS485 modied-MODBUS compatible interface allows up to 128 transmitters on a single bus.
Standard dual (analog & RS485) outputs simplify interface to controls, data collection, and telemetry systems.
Built in the U.S.A. ARRA Section 1605 Compliant.
Standard 3-day lead time
KELLER America Inc.
Newport News, Virginia
www.kelleramerica.com sales@kelleramerica.com
Output White Black Red Blue Yellow
2-wire
(mA)
3-wire
(VDC)
OUT /
GND
+Vcc N/A RS485A RS485B
GND +VCC +OUT RS485A RS485B
Braided shield wire connected to transmitter housing
Min. pipe bend radius 1” IPS 1.5” IPS 2” IPS
Schedule 40 22” 4” 3”
Schedule 80 22” 6” 4”
Edition 07/2019
Subject to alterations
877-253-5537
757-596-6659
Pressure Ranges
1,2
Relative Innite between 0...3 thru 0...900 ft W.C.
Absolute Innite between 0...2 thru 0...11 bar
1.The Acculevel can be provided with custom calibration at no extra cost. For uids other than water,
the specic gravity must be given at the time the order is placed.
2. Intermediate ranges are realized by deranging the analog output from the next highest basic range:
1, 3, 10, and 30 bar (relative) 2, 4, and 11 bar (absolute). Level range may be specied in units of lb/
in2(psi), inches WC or feet WC. Keller America uses the International Standard conversion of 2.3067 feet WC/psi.
Accuracy
3
Static Standard ±0.1% FS, Optional ±0.05% FS
Total Error Band Standard ±0.25% BR, Optional ±0.1% BR
3. Static accuracy includes the combined effects of non-linearity, hysteresis, and non-repeatability at room temperature (25°C). Total Error Band (TEB) includes the combined effects of non-linearity, hysteresis, and non-repeatability as well as thermal dependencies, over the compensated temperature range, expressed as a percentage of the basic range (BR).
The calculation for maximum TEB on intermediate ranges (IR) is:
TEBIR = (BR/IR) X TEB
BR
Output
Current 4...20mA + RS485
Voltage
Resolution
4. Other voltage output options available on request.
5. Resolution applies to digital output only. Analog resolution is continuous and limited by the process meter and not the instrument.
4
5
0...5, 0-10VDC + RS485
0.002%
Certications
CE EN50081-1, EN50082-2
NSF / ANSI
7
61, 372
Electrical
6
Supply (4-20mA) 11-28 VDC
Supply (0-5VDC) 8...28 VDC
Supply (0-10VDC) 13...28 VDC
Load Resistance (mA) <(Supply-11V)/0.022A
Load Resistance (VDC) >4k ohm
6. Nominal values may be higher depending upon cable length. Internal lightning protection increases
the minimum-required supply voltage from 8VDC to 11VDC, due to internal resistance of the surge protectors. In addition, cable resistance (~70Ω / 1000ft) adds to the supply requirement. In order to
insure proper system operation, calculate the minimum required supply voltage (at the source) as follows:
For two-part (internal+external) system (recommended):
MINIMUM SUPPLY VOLTAGE = 11.6 + 0.022 (CABLE LENGTH x 0.07) VDC
For internal only protector (standard with 4-20mA output):
MINIMUM SUPPLY VOLTAGE = 11 + 0.022 (CABLE LENGTH x 0.07) VDC
Environmental
Protection Rating IP68
Operating Temp. -10...60° C
Compensated Temp. -10...80° C
Wetted Materials 316 L Stainless Steel
Titanum Optional
Polyamide
Cable & Sealing PE & EPDM for water / wastewater
7
Hytrel & Viton for hydrocarbons
Tefzel & Viton or EPDM as required for
chemical interaction
7. NSF 61 and NSF 372 approval applies to both 316L stainless steel & titanium construction with PE
& EPDM cable sealing option, which is standard on this instrument unless otherwise specied.
Optional Accessories
1/2” NPT Conduit Fitting Drying Tube Assembly Bellows Assembly Cable Hanger Termination Enclosure Pressure Test Adapter
Stabilizing Weight Interface Converter Process Meter Open-faced Nose Cap Signal Line Surge Protector
KELLER America Inc.
Newport News, Virginia
www.kelleramerica.com sales@kelleramerica.com
Edition 07/2019
Subject to alterations
877-253-5537
757-596-6659