Keithley Instruments 7154 Data Sheet

7154 High Voltage Scanner Card
10-Channel
CONTACT LIFE: >10
(at maximum signal level).
CONTACT RESISTANCE: <200minitial, 2to rated life. CONTACT POTENTIAL: <35µV per contact pair with copper leads. ACTUATION TIME: <2ms exclusive of mainframe. CHANNEL ISOLATION: 10 INPUT ISOLATION: Differential: >10
Common Mode: >10 COMMON MODE VOLTAGE: 1100V peak. ENVIRONMENT: Operating: 0° to 50° up to 35°C at 70% R.H.
Storage: –5° to +65°C.
Ch. 1
Ch. 2 - 9
Ch. 10
1100 volts peak
2-pole switching
High and low fused
Ordering Information
7154 High Voltage
The Model 7154 switches voltages to 1100V peak or currents to 0.5A. The current carry ca­pacity of each relay contact is 1A. Two-pole relays switch both circuit High and Low for full floating measurements and each input line is fuse protected against current overload. A Guard input common to all channels is provided for shielding or as a Guard driven from a single in­strument. Guards may be isolated by removing jumpers installed at each input. Multiple switched guard circuits can be achieved by re­moving the jumper and connecting circuit Guard to the Low input terminal.
CHANNELS PER CARD: 10. CONTACT CONFIGURATION: 2-pole Form A with user-selectable
shield or driven Guard. Each pole is fused using #38AWG mag­net wire.
CONNECTOR TYPE: Screw terminals, #16AWG maximum wire
size.
RELAY DRIVE CURRENT: 57mA per relay typical. MAXIMUM SIGNAL LEVEL: 1100V peak, 0.5A DC or rms
switched, 1A DC or rms carry, 10W.
8
closures (cold switching); >5×106closures
10
, <10pF.
9
9
, <150pF.
H
L
G
H
L
G
J1
J10
, <10pF.
H
L
G
7158 Low Current Scanner Card
Use with 7001 and 7002 Switch Mainframes
Sub-picoamp offset current
Maintains current path for
unselected channel
BNC connectors
Ordering Information
7158 Low Current
Accessories Supplied
4801 Low Noise Male to Male
BNC Input Cable
10-Channel
The Model 7158 provides quality low-current switching at an affordable price. The offset cur­rent error generated is specified <1pA, with typi­cal performance at <30fA. When used with a voltage source and electrometer or picoammeter, this card can easily automate insulation resist­ance tests, reverse leakage tests on semiconduc­tor junctions, or gate leakage tests on FETs.
The Model 7158 is designed to maintain the cur­rent path even when the channel is de-selected. Input connectors are BNC for shielding of the sensitive measurements and for compatibility with low noise coaxial cables such as Keithley accessory cables Models 4801 and 4803. Two outputs are provided to allow for chaining sever­al scanner cards to one measurement instru­ment, and an isolation relay in the output HI minimizes interaction between cards.
CHANNELS PER CARD: 10. CONTACT CONFIGURATION: Single pole, simultaneous break
and make for signal HI input. Signal LO is common for all 10 channels and output. When a channel is off, signal HI is con­nected to signal LO.
CONNECTOR TYPE: BNC. RELAY DRIVE CURRENT: 100mA per card typical (regardless of
channel closures selected).
MAXIMUM SIGNAL LEVEL: 30V, 100mA peak (resistive load). CONTACT LIFE: >10
sures at low signal levels.
CONTACT RESISTANCE: <1. CONTACT POTENTIAL: <200µV. OFFSET CURRENT: <1pA (<30fA typical). 3dB BANDWIDTH: 1MHz typical. ACTUATION TIME: <1ms, exclusive of mainframe. CHANNEL ISOLATION: >10 INPUT ISOLATION: Differential: >10
Common Mode: >10
COMMON MODE VOLTAGE: <30V maximum.
6
closures at maximum signal levels; >107clo-
14
.
9
9
, <150pF.
, <50pF.
ACCESSORIES AVAILABLE
4801 Low Noise Male to Male BNC Input Cable 4802-10 Low Noise BNC to Unterminated Cable, 10 ft. 4803 Low Noise BNC Cable Kit for 7158
Ch. 1
Ch. 2 – 9
Ch. 10
Output
1
Output
2
1.888.KEITHLEY (U.S. only)
SWITCHING & CONTROL
96
www.keithley.com
A GREATER MEASURE OF CONFIDENCE
Loading...