Keithley Instruments 6517B User Manual

Model 6517B Electrometer
User’s Manual
6517B-900-01 Rev. A / June 2008
www.keithley.com
A GREATER MEASURE OF CONFIDENCE

WARRANTY

Keithley Instruments, Inc. warrant s this product to be free from defect s in materia l and workmanship fo r a per iod of one (1) year from date of shipment.
Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables, software, rechargeable batteries, diskettes, and documentation.
During the warranty period, Keithley Instruments will, at its option, either repair or replace any product that proves to be defective.
To exercise this warranty, write or call your local Keithley Instruments representative, or contact Keithley Instruments headquarters in Cleveland, Ohio. You will be given prompt assistance and return instructions. Send the product, transportation prepaid, to the indicated service facility. Repairs will be made and the product returned, transportation prepaid. Repaired or replaced products are warranted for the balan ce of the original warranty period, or at least 90 days.
LIMITATION OF WARRANTY
This warranty does not apply to defects resulting from product modification without Keithley Instruments’ express written consent, or misuse of any product or part. This warranty also does not apply to fuses, software, non-rechargeable batteries, damage from batter y leakage, or problems arising from normal wear or failure to follow instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE. THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OF ITS INSTRUMENTS AND SOFTWARE, EVEN IF KEITHLEY INSTRUMENTS, INC. HAS BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COST OF REMOVAL AND INSTALLATION, LOSSES SUSTAINED AS THE RESULT OF INJURY TO ANY PERSON, OR DAMAGE TO PROPERTY.
A G R E A T E R M E A S U R E O F C O N F I D E N C E
Keithley Instruments, Inc.
440-248-0400 • Fax: 440-248-6168 • 1-888-KEITHLEY (1-888-534-8453) • www.keithley.com
3/07
Model 6517B
Any unauthorized reproduction, photocopy, or use the information herein, in whole or in part, without the prior written
approval of Keithley Instruments, Inc. is strictly prohibited.
TSP, TSP-Link, and TSP-Net are trademarks of Keithley Instruments, Inc.
All Keithley Instruments product names are trademarks or registered trademarks of Keithley Instruments, Inc.
Other brand names are trademarks or registered trademarks of their respective holders.
Electrometer
User’s Manual
©2008, Keithley Instruments, Inc.
All rights reserved.
Cleveland, Ohio, U.S.A.
Document Number:
6517B-900-01 Rev. A / Jun 2008
The following safety precautions should be observed before using this product an d any associated instrumentation. Although some instruments and accessories would normally be used with non-hazardous voltages, there are situations where hazardous conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using the product. Refer to the user documentation for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product warranty may be impaired. The types of product users are: Responsible body is the individual or group responsible for the use and maintenance of equipment, for ensuring that the equipment is
operated within its specifications and operating limits, and for ensuring that operators are adequately trained. Operators use the product for its intended function. They must be trained in electrical safety procedures and proper use of the instrument.
They must be protected from electric shock and contact with hazardous live circuits. Maintenance personnel perform routine procedures on the product to keep it operating properly, for example, setting the line voltage or
replacing consumable materials. Maintenance procedures are described in the user documentation. The procedures explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel.

Safety Precautions

Service personnel are trained to work on live circuits, perform safe installations, and repair products. Only properly trained service
personnel may perform installation and service procedures. Keithley Instruments products are designed for use with electrical signals that are rate d Measurement Category I and Measurement
Category II, as described in the International Electrotechnical Commission (IEC) Standard IEC 60664. Most measurement, control, and data I/O signals are Measurement Category I and must not be directly connected to mains voltage or to voltage sources with high transient over-voltages. Measurement Category II connections require protection for high transient over-voltages often associated with local AC mains connections. Assume all measurement, control, and data I/O connections are for connection to Category I sources unless otherwise marked or described in the user documentation.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30V RMS, 42.4V peak, or 60VDC are present. A good safety practice is to expect that hazardous voltage is present in any unknown circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000V, no conductive part of the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance-limited sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective devices to limit fault current and voltage to the card.
Before operating an instrument, ensure that the line cord is connected to a properly-grounded power receptacle. Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
11/07
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input power
!
disconnect device must be provided in close proximity to the equipment and within easy reach of the operator. For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under test.
AL W AYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being measured.
The instrument and accessories must be used in accordance with its specifications and operating instructions, or the safety of the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating information, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with the same type and rating for continued protection against fire hazard. Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground connections. If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use of a lid
interlock. If a screw is present, connect it to safety earth ground using the wire recommended in the user documentation. The symbol on an instrument indicates that the user should refer to the operating instructions located in the user documentation. The symbol on an instrument shows that it can source or measure 1000V or more, including the combined effect of normal and
common mode voltages. Use standard safety precautions to avoid personal contact with these voltages. The symbol on an instrument shows that the surface may be hot. Avoid personal contact to prevent burns. The symbol indicates a connection terminal to the equipment frame.
If this symbol is on a product, it indicates that mercury is present in the display lamp. Please note that the lamp must be properly disposed of according to federal, state, and local laws.
The WARNING heading in the user documentation explains dangers that might result in personal injury or death. Always read the associated information very carefully before performing the indicated procedure.
The CAUTION heading in the user documentation explains hazards that could damage the instrument. Such damage may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans. Before performing any maintenance, disconnect the line cord and all test cables. T o maintain protection from electric shock and fire, replacement components in mains circuits - including the power transformer, test leads,
and input jacks - must be purchased from Keithley Instruments. Standard fuses with applicable national safety approvals may be used if the rating and type are the same. Other components that are not safety-related may be purchased from other suppliers as long as they are equivalent to the original component (note that selected parts should be purchased only through Keithley Instruments to maintain accuracy and functionality of the product). If you are unsure about the applicability of a replacement component, call a Keithley Instruments office for information.
To clean an instrument, use a damp cloth or mild, water-based cleaner. Clean the exterior of the instrument only. Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with no case or chassis (e.g., a data acquisition board for installation into a computer) should never require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the factory for proper cleaning/servicing.

Table of Contents

Section Topic Page
1 Introduction............................................................................................. 1-1
Overview..................................................................................................... 1-2
Capabilities and features overview....................................................... 1-2
Options and accessories...................................................................... 1-3
Specifications....................................................................................... 1-5
Unpacking and inspection........................................................................... 1-5
Inspection for damage.......................................................................... 1-5
Handling precautions ........................................................................... 1-5
Shipment contents ............................................................................... 1-5
Documentation..................................................................................... 1-5
Repacking for shipment........................................................................ 1-6
2 Getting Started....................................................................................... 2-1
Introduction................................................................................................. 2-2
Front and rear panel familiarization............................................................ 2-2
Front panel summary........................................................................... 2-2
Rear panel summary............................................................................ 2-4
Power-up .................................................................................................... 2-6
Line power connection......................................................................... 2-6
Power-up procedure............................................................................. 2-6
Display........................................................................................................ 2-6
Primary display line.............................................................................. 2-6
Secondary display line ......................................................................... 2-7
Bench defaults............................................................................................ 2-7
Restoring bench defaults...................................................................... 2-7
Function default settings...................................................................... 2-7
3 Basic Measurements ............................................................................ 3-1
Introduction................................................................................................. 3-2
Changing functions..................................................................................... 3-2
Voltage measurements............................................................................... 3-2
Guarding .............................................................................................. 3-3
Current measurements............................................................................... 3-4
Resistance measurements......................................................................... 3-5
Auto V-Source...................................................................................... 3-6
Amps Rel.............................................................................................. 3-6
Resistivity measurements ........................................................................... 3-7
Charge measurements ............................................................................... 3-9
Auto discharge ................................................................................... 3-10
Temperature and humidity measurements ............................................... 3-10
Connections....................................................................................... 3-10
Measurement control ......................................................................... 3-11
Temperature units............................................................................... 3-11
Display reading options...................................................................... 3-12
4 Measurement Options.......................................................................... 4-1
Introduction................................................................................................. 4-3
Multiple displays ......................................................................................... 4-3
Menus......................................................................................................... 4-5
Navigating menus ................................................................................ 4-6
Table of Contents Model 6517B Electrometer User’s Manual
Section Topic Page
Voltage source ............................................................................................ 4-7
Connections.......................................................................................... 4-7
Basic operation..................................................................................... 4-9
Configuring V-Source........................................................................... 4-9
Relative....................................................................................................... 4-9
Configuring rel .................................................................................... 4-10
Zero check ................................................................................................ 4-10
Zero correct............................................................................................... 4-10
Triggering .................................................................................................. 4-10
Idle...................................................................................................... 4-11
Control sources .................................................................................. 4-12
Device action...................................................................................... 4-12
Output trigger...................................................................................... 4-12
Counters............................................................................................. 4-12
Bench default setup............................................................................ 4-12
Speed ........................................................................................................ 4-13
Configuring speed .............................................................................. 4-13
Resolution................................................................................................. 4-13
Configuring resolution......................................................................... 4-13
Filter.......................................................................................................... 4-13
Filter types.......................................................................................... 4-14
Filter modes........................................................................................ 4-14
Median filter........................................................................................ 4-14
Configuring filters................................................................................ 4-15
Buffer (data store) ..................................................................................... 4-15
Storing readings.................................................................................. 4-15
Math.......................................................................................................... 4-16
Selecting and configuring math.......................................................... 4-17
Enabling math..................................................................................... 4-18
Additional math operations................................................................. 4-18
Test sequences ......................................................................................... 4-19
Configuring test sequences................................................................ 4-19
Connections........................................................................................ 4-20
Running a test .................................................................................... 4-21
Internal scanning ....................................................................................... 4-21
Configure internal scanner ................................................................. 4-21
Perform the scan................................................................................ 4-21
Close/open channels.......................................................................... 4-22
External scanning...................................................................................... 4-22
Trigger connections............................................................................ 4-23
Configure external channels............................................................... 4-23
Perform the scan................................................................................ 4-23
5 Remote Operation.................................................................................. 5-1
Overview..................................................................................................... 5-2
Software support......................................................................................... 5-2
LabView driver...................................................................................... 5-2
Remote interfaces....................................................................................... 5-2
IEEE-488 bus ....................................................................................... 5-2
RS-232 interface................................................................................... 5-3
Common commands................................................................................... 5-5
SCPI commands......................................................................................... 5-5
SCPI command syntax......................................................................... 5-6
SCPI signal-oriented commands.......................................................... 5-8
SCPI subsystem commands ................................................................ 5-8
Program examples.................................................................................... 5-13
Changing function and range ............................................................. 5-13
One-shot triggering............................................................................. 5-14
Continuous triggering #1 .................................................................... 5-15
Continuous triggering #2 .................................................................... 5-15
ii 6517B-900-01 Rev. A / Jun 2008
Model 6517B Electrometer User’s Manual Table of Contents
Section Topic Page
Generating SRQ on buffer full ............................................................ 5-16
Storing readings in the buffer.............................................................. 5-17
Taking readings with the scanner card ............................................... 5-18
Test sequence: Staircase sweep........................................................ 5-19
Index ............................................................................................................................ I-1
6517B-900-01 Rev. A / Jun 2008 iii
Table of Contents Model 6517B Electrometer User’s Manual
This page left blank intentionally.
iv 6517B-900-01 Rev. A / Jun 2008

List of Figures

Section Figure Title Page
2 Figure 2-1 Model 6517B front panel................................................................ 2-2
2 Figure 2-2 Model 6517B rear panel ................................................................ 2-4
3 Figure 3-1 Voltage measurements.................................................................. 3-3
3 Figure 3-2 Current measurements.................................................................. 3-4
3 Figure 3-3 Resistance measurements ............................................................ 3-6
3 Figure 3-4 Resistivity measurements.............................................................. 3-9
3 Figure 3-5 Charge measurements................................................................ 3-10
3 Figure 3-6 External temperature and humidity measurements..................... 3-11
4 Figure 4-1 Main menu display......................................................................... 4-5
4 Figure 4-2 Independent V-Source connections............................................... 4-8
4 Figure 4-3 FVMI connections.......................................................................... 4-8
4 Figure 4-4 Basic trigger model (simplified).................................................... 4-11
4 Figure 4-5 Advanced trigger model (simplified) ............................................ 4-11
4 Figure 4-6 Filter response/noise window ...................................................... 4-14
4 Figure 4-7 Combining math calculations....................................................... 4-18
4 Figure 4-8 Typical FVMI connections............................................................ 4-20
4 Figure 4-9 Trigger connections using trigger link.......................................... 4-23
5 Figure 5-1 IEEE-488 connections ................................................................... 5-3
5 Figure 5-2 Hierarchy of the example SENSe1 commands ............................. 5-6
List of Figures Model 6517B Electrometer User’s Manual
This page left blank intentionally.
vi 6517B-900-01 Rev. A / Jun 2008

List of Tables

Section Table Title Page
4 Table 4-1 Multiple displays by function.......................................................... 4-3
4 Table 4-2 Main menu summary..................................................................... 4-5
4 Table 4-3 Configuration settings for each measurement function................. 4-6
4 Table 4-4 Configuration settings for instrument operations........................... 4-6
4 Table 4-5 Menu summary.............................................................................. 4-7
List of Tables Model 6517B Electrometer User’s Manual
This page left blank intentionally.
viii 6517B-900-01 Rev. A / Jun 2008
In this section:
Topic Page
Overview ................................................................................................. 1-2
Unpacking and inspection....................................................................... 1-5
Section 1
Introduction
Capabilities and features overview ................................................. 1-2
Options and accessories................................................................. 1-3
Specifications.................................................................................. 1-5
Inspection for damage .................................................................... 1-5
Handling precautions ...................................................................... 1-5
Shipment contents.......................................................................... 1-5
Repacking for shipment .................................................................. 1-6
Section 1: Introduction Model 6517B Electrometer User’s Manual

Overview

This section contains general information about the Keithley Instruments Model 6517B Electrometer.
If you have any questions after reviewing this information, please cont act your local Keithley Instruments representative or call one of our applications engineers at 1-888-KEITHLEY (1-888-534-8453) within the U.S. and Canada. You can also contact us through our website at
www.keithley.com.
Capabilities and features overview
The Model 6517B is a 6½-digit electrometer/high-resistance test and measurement system with the following measurement capabilities:
DC voltage measurements from 1μV to 210V.
DC current measurements from 10aA to 21mA.
Charge measurements from 10fC to 2.1μC.
Resistance measurements from 10Ω to 210PΩ.
Surface resistivity measurements.
Volume resistivity measurements.
External temperature measurements from -25°C to 150°C using the supplied Model 6517-TP thermocouple.
Relative humidity measurements (0 to 100%) using the optional Model 6517-RH probe.
Some additional capabilities of the Model 6517B include:
Built-in V-Source. The 100V range provides up to ±100V at 10mA, while the 1000V range provides up to ±1000V at 1mA.
Data storage (50,000 points).
Single button zeroing (REL).
Built-in math functions.
Filtering, averaging, and median.
Built-in test sequences.
Remote operation using the IEEE-488 (GPIB) bus or the RS-232 interface.
Scan (measure) channels of an external scanner.
Scan (measure) channels of an internal scanner card (for example, Keithley Instruments Model 6521 or Model 6522) installed in the option slot.
1-2 Return to Section Topics 6517B-900-01 Rev. A / Jun 2008
Model 6517B Electrometer User’s Manual Section 1: Introduction
Options and accessories
The following options and accessories are available from Keithley Instruments for use with the Model
6517B Electrometer:
NOTE Check the Keithley Instruments website (www.keithley.com) for additional options and
accessories that may have been added to the Keithley Instruments product line for use with the Model
Cables and adapters
Model 237-ALG-2 Triax Cable: This is a 2-meter (6.6ft) low-noise triax cable terminated with a
3-slot male triax connector on one end and 3 alligator clips on the other. Model 237-BNC-TRX Adapter: This is a male BNC to 3-lug female triax adapter (guard
disconnected). It is used to terminate a triax cable with a BNC plug. Suit able for use with the Model 6517B V-Source in high voltage applications.
Model 237-TRX-T Adapter: This is a 3-slot male to dual 3-lug fema le triax tee adapter for use with 7078-TRX triax cables. Suitable for use with the Model 6517B V-Source in high voltage applications.
6517B Electrometer.
Model 237-TRX-TBC Connector: This is a 3-lug female triax bulkhead connector with cap for assembly of custom panels and interface connections. Suitable for use with the Model 6517B V-Source in high voltage applications.
Model 6517B-ILC-3 Interlock Cable: This is a 3 meter (6.6 ft) cable terminated with a 4 pin Phoenix connector on one end and a 4 pin Switchcraft connector on the other end.
Model 7078-TRX-BNC Adapter: This is a 3-slot male triax to female BNC adapter. This adapter lets you connect a BNC cable to the triax input of the Model 6517B. Suitable for use with the Model 6517B in high voltage applications.
Models 7078-TRX-3, 7078-TRX-10 and 7078-TRX-20 Triax Cables: These are low noise triax cables terminated at both ends with 3-slot male triax connectors. The Model 7078-TRX-3 is 3ftft (0.9m) in length, the Model 7078-TRX-10 is 10f t (3m) in length, and the Model 7078-TRX-20 is 20f t (6m) in length.
Model 7078-TRX-TBC Connector: This is a 3-lug female triax bulkhead connector with cap for assembly of custom panels and interface connections. Suitable for use with the Model 6517B V-Source in high voltage applications.
Models 7007-1 and 7007-2 Shielded IEEE-488 Cables: Connect the Model 6517B to the IEEE-488 bus using shielded cables and connectors to reduce electromagnetic interference (EMI). The Model 7007-1 is one meter long; the Model 7007-2 is two meters long.
Models 8501-1 and 8501-2 Trigger Link Cables: Connect the Model 6517B to other instruments with Trigger Link connectors (for example, Model 7001 Switch System). The Model 8501-1 is one meter long; the Model 8501-2 is two meters long.
Model 8502 Trigger Link Adapter: Allows you to connect the Trigger Link of the Model 6517B to instruments that use the standard BNC (In/Out) external trig g eri ng tech n iqu e.
Model 8606 High Performance Probe Tip Kit : Consists of two spade lugs, two alligator clips, and two spring hook test probes (the spade lugs and alligator clips are rated at 30V RMS, 42.4V peak; the test probes are rated at 1000V). These components are designed to be used with high performance test leads terminated with banana plugs, such as the Model 8607 High Performance Banana Cables.
6517B-900-01 Rev. A / Jun 2008 Return to Section T o pics 1-3
Section 1: Introduction Model 6517B Electrometer User’s Manual
Model 8607 High Performance Banana Cables: Consists of two high voltage (1000V) banana cables. The cables are terminated with banana plugs that have retractable sheaths.
CS-751 Barrel Adapter: This is a barrel adapter that allows you to connect two triax cables together. Both ends of the adapter are terminated with 3-lug female triax connectors.
Case and rack mount kits
Model 1050 Padded Carrying Case: A carrying case for a Model 6517B. Includes handles and
shoulder strap. Model 4299-1 Single Fixed Rack Mount Kit: Moun ts a sing le Model 6517B in a sta ndard 19-in ch
rack. Model 4299-2 Side-by-side Rack Mount Kit: Mounts two instruments side-by-side in a standard
19-inch rack. Model 4288-4 Side-by-side Rack Mount Kit: Mount s a Mo del 651 7B an d a 5.25-inch in str ument
side-by-side in a standard 19-inch rack.
Probes
Model 6517-RH Humidity Probe with Cable: This sensor allows the Model 6517B to make
relative humidity measurements (0 to 100%). Also included is an extension cable (part number CA-129-1).
Model 6517-TP Thermocouple with Leads: This type K thermocouple sensor allows the Model 6517B to make external temperature measurements from –190°C to 1350°C.
Scanner cards
Model 6521 Low Current Scanner Card: This 10-channel low current scanner card is terminated
with BNC connectors and plugs into the option slot of the Model 6517B. Model 6522 Low Current/Low Voltage Scanner Card: This 10-channel low current/low voltage
scanner card is terminated with triax connectors and plugs into the option slot of the Model 6517B.
T est fixture
Model 8009 Resistivity Test Fixture: This is a guarded test fixture for measuring volume and
surface resistivities. It can accommodate sheet samples 64 to 102mm (2-1/2 to 4 in.) in diameter and up to 3.175mm (1/8 in.) thick.
Software
Model 6524 Hi-R Software Package: Desig ned to aid in ma king more r epeat able high r esista nce/
resistivity measurements. Four windows-driven programs increase measurement precision, ease download and analysis of Hi-R data, and allow cross-correlation of environmental factors.
1-4 Return to Section Topics 6517B-900-01 Rev. A / Jun 2008
Model 6517B Electrometer User’s Manual Section 1: Introduction
Specifications
Full Model 6517B specifications are included in the Model 6 517B Refere nce Manual, Ap pen dix A. Check the Keithley Instruments website at www.keithley.com for the latest updates to the specifications.

Unpacking and inspection

Inspection for damage
The Model 6517B was carefully inspected electrically and mechanically before shipmen t. After unpacking all items from the shipping carton, check for a ny obvious si gns of physical d amage th at may have occurred during transit (there may be a protective film over the display lens, which can be removed). Report any damage to the shipping agent immediately. Save the original packing carton for possible future shipment. Before removing the Model 6517B from the anti-static bag, observe the following handling precautions.
Handling precautions
Always grasp the Model 6517B by the covers.
After removing the Model 6517B from its anti-static bag, inspect it for any obvious signs of physical damage; report any such damage to the shipping agent immediately.
When the Model 6517B is not installed and connected, keep the unit in its anti-static bag and store it in the original packing carton.
Shipment contents
The following items are included with every Model 6517B order:
Model 6517B Electrometer with line cord
Protective Triax Shield/Cap (CAP-28-1)
237-ALG-2 Triax Cable terminated with alligator clips on one end
Accessories as ordered
Certificate of calibration
CD containing complete documentation and required software
Documentation
A CD-ROM containing this User’s Manual, a Re ference Manual, and r equired sof tware is included with each Model 6517B. Check the Keithley Instruments website at www.keithley.com for the latest versions of the manuals. The latest manuals can be downloaded (in PDF format) from the website.
6517B-900-01 Rev. A / Jun 2008 Return to Section T o pics 1-5
Section 1: Introduction Model 6517B Electrometer User’s Manual
Repacking for shipment
Should it become necessary to return the Model 6517B for repair, carefully pack the unit in its original packing carton or the equivalent, and follow these instructions:
Call the repair department toll-free at 1-888-KEITHLEY (1-888-534-8453), within the U.S. and Canada, to obtain a Return Material Authorization (RMA) number.
Advise as to the warranty status of the Model 6517B.
Write ATTENTION REPAIR DEPARTMENT and the RMA number on the shipping label.
Complete and include the Service Form located at the back of this manual.
1-6 Return to Section Topics 6517B-900-01 Rev. A / Jun 2008
In this section:
Topic Page
Introduction .............................................................................. ... ............ 2-2
Front and rear panel familiarization......................................................... 2-2
Power-up ............................... ... ....................................... ... .................... 2-6
Display .................................................................................................... 2-6
Section 2
Getting Started
Front panel summary .................................... ........... ........ ........... .... 2-2
Rear panel summary .............................................................. .. ...... 2-4
Line power connection ................................................................... 2-6
Power-up procedure ....................................................................... 2-6
Primary display line ........................................................................ 2-6
Secondary display line ................................................................... 2-7
Bench defaults ....................................... ....................................... ... ....... 2-7
Restoring bench defaults ................................ ................................ 2-7
Function default settings ................................................................ 2-7
Section 2: Getting Started Model 6517B Electrometer User’s Manual
VOLTAGE
SOURCE
OPERATE
6517B ELECTROMETER/HIGH RESISTANCE METER
EDIT ERR REM TALK LSTN SRQ REAR REL FILT MATH 4W AUTO ARM TRIG SMPL
1 2 3 4 5 6 7 8 9 10 1 2 3 4 5 6 7 8 9 10
PREV
NEXT
DISPLAY
Z-CHK REL
FILTER
MATH TRIG SEQ
INFO
LOCAL STORE
RECALL
CONFIG
MENU
EXIT
ENTER
VI R
Q
CARD OPER
OPTION
VOLTAGE SOURCE
POWER
AUTO
RANGE
RANGE
5
19
273864
Secondary display line
Primary display line

Introduction

This section contains identification and descriptions of controls and components of the Keithley Instruments Model 6517B Electrometer and det ailed informa tion for powering up the Mode l 6517B.

Front and rear panel familiarization

Front panel summary
The front panel of the Model 6517B is shown in Figure 2-1. The descriptions of the front panel controls follow Figure 2-1.
Figure 2-1
Model 6517B front panel
1 ANNUNCIATORS
EDIT Editing voltage source values ERR Questionable reading REM In remote TALK Addressed to talk LSTN Addressed to listen SRQ Service Request REL Relative reading displayed FILT Digital filter enabled MATH Math calculation enabled AUTO Autoranging enabled ARM Trigger armed; not in idle *(asterisk) Reading being stored
2-2 Return to Section Topics 6517B-900-01 Rev. A / Jun 2008
Model 6517B Electrometer User’s Manual Section 2: Getting Started
2 F UNCTION KEYS
Select measurement function; volts (V), amps (I), resistivity (R) or coulombs (Q).
3 RAN GE KEYS
Moves to higher range; increments digit. Moves to lower range; decrements digit.
AUTO Enables/disables autorange.
4 HANDLE
Pull out and rotate to desired position.
5 DISPLAY KEYS
PREV/NEXT Scroll through multiple displays of a function.
6 OPERATION KEYS
Z-CHK Enables/disables zero check: enable zero check before changing functions. REL Enables/disables relative reading. FILTER Displays digital filter status for present function and toggles filter on/off. MATH Displays math calculation and toggles math on/off if configured. TRIG Triggers unit. SEQ Performs selected test sequence.
and Moves cursor among data entry digits, menu selections, and information displays.
INFO Shows context-sensitive information about the present display. LOCAL Cancels remote operation. STORE Enables data storage. RECALL Displays reading data (reading, number, time). Use PREV/NEXT DISPLAY for
maximum, minimum average, standard deviation. CONFIG Configures functions and operations. MENU Saves/restores instrument conditions; sets up communications; performs calibration
and self-tests; defines limits, digital output, and other miscellaneous operations. EXIT Cancels selection, moves back within menu structure. ENTER Holds reading, enters selection, moves down within menu structure.
7 VOLTAGE SOURCE KEYS
OPER Toggles V-source between operate and standby.
and Adjusts V-source value.
8 VO LTA GE SOU RCE OPERATE
Indicator light is on when in operation, off in standby.
9 OPTION CARD KEY
Use to program and operate an optional feature. Also use to view external scanner channels.
6517B-900-01 Rev. A / Jun 2008 Return to Section T o pics 2-3
Section 2: Getting Started Model 6517B Electrometer User’s Manual
8
9101112131415
Input high
Input low Chassis ground
Unguarded (guard off)
Input high
Guard Chassis ground
Guarded (guard off)
OPTION SLOT
250VDC MAX
MADE IN
U.S.A
LINE RATING
50-60 Hz
100 VA MAX.
LINE
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS, SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD, REPLACE FUSE WITH SAME TYPE AND RATING.
1250Vpk
PREAMP
OUT
1000VDC
MAX
FUSE
630mAT
315mAT
100V 120V 220V 240V
COMMON 2V OUT
LO HI
V SOURCE
TRIGGER LINK RS-232
DIGITAL I/O
HUMIDITY
TEMP
TYPE K
INTERLOCK
IEEE-488
2VDC MAX
1234567
Rear panel summary
The rear panel of the Model 6517B is shown in Figure 2-2. The descriptions of the rear panel components follow Figure 2-2.
Figure 2-2
Model 6517B rear panel
1 INPUT CONNECTOR
Unguarded Configuration Disable GUARD for amps, ohms, coulombs
and unguarded voltage measurements.
Guarded Configuration Enable GUARD for guarded voltage measurements.
2PREAMP OUT
Follows the signal amplitude applied to the INPUT terminal. With GUARD on, PREAMP OUT is connected to the inner shell of the INPUT triax connector to configure the input for guarded voltage measurements. Referenced to COMMON. See the Model 6517B Reference Manual for more information.
2-4 Return to Section Topics 6517B-900-01 Rev. A / Jun 2008
Model 6517B Electrometer User’s Manual Section 2: Getting Started
3COMMON
Connector that is internally connected to INPUT low.
4 CHASSIS GROUND
Attached cable that connects the chassis to ground through the power line cord. COMMON can be connected to chassis ground by plugging the cable into COMMON. For floating measurements, make sure the ground cable connection between COMMON and Chassis Ground is open.
5 2V ANALOG OUTPUT
Connector provides a scaled 0 to 2V output that is referenced to COMMON. It is typically connected to a measuring device such as a chart recorder. See the Model 6517B Reference Manual for more information.
6 HUMIDITY
Connect the optional Keithley Instruments Model 6517-RH probe for relative humidity measurements.
7 TEMP TYPE K
Connect the Keithley Instruments Model 6517-TP type K thermocouple for temperature measurements.
8 POWER LINE INPUT MODULE
Provides connections for power line input and contains the line fuse. If the fuse needs to be replaced, refer to the procedure in Section 2 of the Model 6517B Reference Manual.
9 IEEE-488 CONNECTOR
Connects the Model 6517B to the IEEE-488 (GPIB) bus. Use shielded IEEE-488 cables.
10 INTERLOCK
Connects the safety interlock to a test fixture using an appropriate cable. Interlock is automatically enabled when the appropriate interlock cable is connected to the 6517B. Rated at 50-60 Hz, 140 VA max. The interlock’s CS-1305 connector includes four pins (left to right as viewed from rear of the 6517B):
Pin 1: Interlock safe
Pin 2: Ground
Pin 3: +5 VDC output
Pin 4: Surface/volume select (low = volume, high = surface)
11 RS-232
DB-9 connector for the RS-232 interface. Use a standard RS-232 cable.
12 DIGITAL I/O
A male DB-9 connector for the four TTL-compatible digital output lines.
13 TRIGGER LINK
An 8-pin micro DIN connector for sending and receiving trigger pulses to and from other instruments.
14 V-SOURCE HI and LO
Safety banana jacks for the voltage source.
15 OPTION SLOT
An option card, such as the Keithley Instruments Model 6521 or Model 6522 scanner card, can be installed in this slot.
6517B-900-01 Rev. A / Jun 2008 Return to Section T o pics 2-5
Section 2: Getting Started Model 6517B Electrometer User’s Manual

Power-up

Line power connection
Follow the procedure below to connect the Model 6517B to line power and turn on the instrument.
CAUTION Operating the instrument on an incorrect line voltage may cause damage to
the instrument, possibly voiding the warranty.
1. Before plugging in the power cord, make sure that the front panel power switch is in the off (O) position. Be sure that line input voltage is set to the correct range for the power supply being used to power the Model 6517B.
NOTE Line voltage is pre-set at the factory, but may be reset in the field by adjusting the voltage
selector behind the left ear. To access the voltage selector, first remove the handle then remove the left mounting ear. The current voltage setting is the marking closest to the small circle.
2. Connect the female end of the supplied power cord to the AC receptacle on the rear pan el. Connect the other end of the power cord to a grounded AC outlet.
WARNING The power cord supplied with the Model 6517B contain s a separate ground
wire for use with grounded outlets. When proper connections are made, instrument chassis is connected to power line ground through the gr ound wire in the power cord. Failure to use a grounded outlet may r esult in personal injury or death due to electric shock.
Power-up procedure
1. Turn on the instrument by pressing the front panel power switch to the on (I) position.
2. During power-up, the Model 6517B performs self-tests on its memory elements. If a failure occurs, the instrument momentarily displays an error message and turns on the ERR annunciator. If the instrument passes the self-tests, the firmware revision level and bus interface information are displayed. For complete information on the power-up sequence, see Section 2 in the Model 6517B Reference Manual.

Display

As shown in the Model 6517B front panel illustration (see Figure 2-1), the front panel has three lines of display information: the primary display line, the secondary display line, and annunciators.
Primary display line
The top line displays readings along with units. It can also display measu rement type, “hold,” math operating type, channel number, menu headings, and messages. Status and error messages are listed in Section 2 of the Reference Manual.
2-6 Return to Section Topics 6517B-900-01 Rev. A / Jun 2008
Loading...
+ 63 hidden pages