LOW RESISTANCE MEASUREMENTS
580 Micro-ohmmeter
• 10µsensitivity
• 20mV voltage clamp
• Selectable test current wave­forms
• 3 sets of test leads included
• Optional IEEE-488 interface
• Optional battery
ORDERING INFORMATION
580 Micro-ohmmeter with 5804,
580-1 Micro-ohmmeter without Test
580/5802 Micro-ohmmeter with Isolated
580/1978 Micro-ohmmeter with
580/02/78 Micro-ohmmeter with Isolated
This product is available with an
Warranty. See section C for complete order-
ing information.
5805, 5806 Test Leads and 5801 Test Lead Pouch
Leads and Pouch
Analog Output/IEEE-488 Interface
Rechargeable Battery Pack
Analog Output/IEEE-488 Interface and Rechargeable Battery Pack
Extended
ACCESSORIES AVAILABLE
TEST LEADS
5804 Test Lead Set 5805 Kelvin Probes, 0.9m (3 ft) 5805-12 Kelvin Probes, 3.6m (12 ft) 5806 Kelvin Clip Leads 5807-7 Helical Spring Point
Test Leads
RACK MOUNT KITS
1010 Single Rack Mounting Kit 1017 Dual Rack Mounting Kit
OTHER
1978 Rechargeable Battery
Pack 5801 Test Lead Pouch 5802 Isolated Analog Output/IEEE-488
Interface
See page A-231 for descriptions of all acces­sories.
The 580 Micro-ohmmeter combines high accuracy, resolution, and sensitivity with three special capabilities that make micro-ohm measurements easier and more versatile than ever:
1. When measuring contact and connector resistances, it is important not to puncture oxides and films that may have formed. The 580 ensures this dry circuit condition by clamping the open circuit test voltage to 20mV on the 200mΩ, 2Ω, and 20Ωranges when the 20mV MAX button is pressed.
2. The 580 can test more devices by enabling both test current polarity and waveform (DC or pulsed) to be selected. When using pulsed test current, the 580 automatically compensates for thermoelec­tric EMFs. Tests on inductive devices are best performed using DC test current, as this avoids the effects of time constants on pulsed current resistance measurements. For temperature-sensitive components, a single trigger mode of operation minimizes power delivered to the device. Single trigger operation is possible via the front panel.
3. The optional IEEE-488 bus interface permits operation as a stand-alone unit as well as in a PC-based system.
A microprocessor-based design gives the 580 a number of advanced capabil­ities. Resistance measurements can be made from 10µto 200kon seven ranges, with 4
1
⁄2-digit resolution and accuracy to within 0.04% of reading. Ranging can be performed either man­ually or automatically. Settling time is less than one second to within 10 counts on a given range.
The 580 is supplied with three sets of test leads—standard leads, Kelvin probes, and Kelvin clip leads—in a rug­ged strap-on pouch. For applications requiring portability or line power iso­lation, an optional battery pack is also available. Digital calibration is possible through front panel controls or over the optional IEEE-488 bus interface.
QUESTIONS?
1-800-552-1115 (U.S. only)
Call toll free for technical assistance,
product support or ordering information.
The 580 is supplied with three sets of test leads—standard leads, Kelvin probes, and Kelvin clip leads—in a rugged strap-on pouch.
LOW RESISTANCE MEASUREMENTS
T.O.C.
ALPHA
NUM.
INST.
D ATA
ACQ.
SERVICE
GLOS­SARY
SALES
REPS
LIT.
REQ.
PRINT
MAIN
MENU
580 Micro-ohmmeter
A-57
NON DRY CIRCUIT TEST DRY CIRCUIT TEST
MAXIMUM 1 Year, 18°–28°C POWER 1 Year, 18°–28°C
RANGE RESOLUTION CURRENT PULSED DC IN SAMPLE PULSED
200 m 10 µ 100 mA 0.04 + 2 0.04 + 3 500 µW 0.05 + 2
2 100 µ 10 mA 0.04 + 2 0.04 + 3 50 µW 0.05 + 2
20 1m 1 mA 0.04 + 2 0.04 + 3 5 µW 0.05 + 2
200 10 m 1 mA 0.04 + 2 0.04 + 2
2k 100 mΩ 1 mA 0.04 + 2 0.04 + 2
20 k 1 Ω 10 µA 0.05 + 2 0.05 + 2
200 k 10 10 µA 0.075 + 2 0.075 + 2
CONFIGURATION: 4-wire (two sense, two source). MAXIMUM SOURCE VOLTAGE: 20mV in Dry Circuit Test, 1V
otherwise.
MAXIMUM TEST LEAD RESISTANCE 200mand 2Ranges:
10in each SENSE lead with Non Dry Circuit Test; up to the selected full range resistance in each SOURCE lead and 10in each SENSE lead with Dry Circuit Test.
20through 200kRanges: Up to half of the selected range
in each test lead.
Up to 5in each SOURCE lead and
TEST ±(%rdg + counts) DISSIPATION ±(%rdg + counts)
ACCURACY MAXIMUM ACCURACY
CONVERSION RATE: 3 readings/second typical. RANGING: Auto or manual. AUTORANGING TIME: 200 ms per range change, average. SETTLING TIME: <1 second to within 10 counts on range. MAXIMUM INPUT OVERLOAD: 10V limited to 10A. MAXIMUM COMMON MODE VOLTAGE: 30V rms at DC, 50
or 60Hz.
TEMPERATURE COEFFICIENT (0°–18°C & 28°–50°C): ±(0.1
× applicable accuracy specification)/°C.
ISOLATED ANALOG OUTPUT/IEEE-488 INTERFACE (Model 5802 Option)
ANALOG OUTPUT LEVEL:
1V = 10,000 counts on ×1 gain.
1V = 100 counts on Maximum output voltage = ±4V.
ACCURACY: ±(0.25% of displayed reading + 2mV). In ×100 , 2mV output
= 0.2 displayed counts.
RESPONSE TIME: Follows display conversion rate. OUTPUT RESISTANCE: 1000Ω. ISOLATION: ANALOG OUTPUT LO is connected to IEEE COMMON.
Maximum common mode voltage from IEEE COMMON to earth ground is 30V rms at DC, 50 or 60Hz.
×100 gain.
GENERAL
DISPLAY: ±20,000-count LCD, range and
status information displayed.
OVERRANGE INDICATION:“OL” displayed. CONNECTORS: Measurement and rear panel
EXTERNAL TRIGGER inputs: Banana jacks.
RELATIVE: Allows zeroing of on-range read-
ings. Allows readings to be made with re­spect to baseline value. Display annunciator indicates REL.
DRIVE: Selects either pulsed or DC SOURCE
current. Pulsed drive provides automatic cancellation of thermal offsets, using 50% duty cycle pulse. Display annunciator in­dicates drive selected.
POLARITY: Selects either positive or nega-
tive SOURCE current in either drive. Dis­play annunciator indicates polarity selected.
TRIGGER: Allows single pulsed measurements. OPERATING ENVIRONMENT: 0°–50°C, <80%
relative humidity up to 35°C; linearly derate 3% R.H./°C from 35° to 50°C.
STORAGE ENVIRONMENT: –25° to 60°C. POWER: 105–125V or 210–250V (switch select-
ed), 90–110V available. 50–60Hz, 12VA. Optional 6-hour battery pack, Model 1978.
DIMENSIONS,WEIGHT: 89mm high ×
241mm wide × 300mm deep (31⁄2in × 91⁄ in × 113⁄4in). Net weight 3.2kg (7 lb). Test lead pouch adds 76mm (3 in) in height.
ACCESSORIES SUPPLIED: Models 5801,
5804, 5805, and 5806 test leads; Operator’s and Service Manuals.
2
IEEE-488 BUS
IMPLEMENTATION
MULTILINE COMMANDS: DCL, SDC, GET,
GTL UNT, UNL, SPE, SPD, LLO.
UNILINE COMMANDS: IFC, REN, EOI, SRQ,
ATN .
INTERFACE FUNCTIONS: SH1, AH1, T5,
TE0, L4, LE0, SR1, RL0, PP0, DC1, DT1, C0, E1.
PROGRAMMABLE PARAMETERS: Range,
DRY CIRCUIT TEST, Operate, RELative, POLARITY, DRIVE, TRIGger, Calibration, EOI, SRQ, Status, Data Format, Terminator.