Keithley Instruments 2400 Data Sheet

Series 2400
Family of products offers wide
dynamic range: 10pA to 10A, 1µV to 1100V, 20W to 1000W
4-quadrant operation
0.012% basic accuracy with
Tightly coupled precision sourcing and measurement
1
2-digit resolution
5
6-wire Ω measurement with
programmable I source and
SourceMeter®Line
issues associated with using multiple instruments. And, their compact half-rack size conserves pre­cious “real estate” in the test rack or bench.
The Power of Three Instruments in One
The tightly coupled nature of a SourceMeter instrument provides many advantages over separate instruments. For example, it provides faster test times by reducing GPIB traffic and simplifies the remote programming interface. It also protects the device under test from damage due to accidental overloads, thermal runaway, etc. Both the current and voltage source are programmable with read­back to help maximize device measurement integrity. If the readback reaches a programmed compli­ance limit, then the source is clamped at the limit, providing fault protection.
V clamp
1700 readings/second at
1
2 digits via GPIB
4
Built-in comparator for fast
pass/fail testing
Optional contact check function
Digital I/O for fast binning and
connection to component handlers
GPIB, RS-232, and Trigger Link
interfaces
TestPoint
and LabVIEW™drivers
TEST LEADS AND PROBES
1754 2-Wire Universal 10-Piece Test Lead Kit 5804 Kelvin (4-Wire) Universal 10-Piece Test Lead Kit 5805 Kelvin (4-Wire) Spring-Loaded Probes 5806 Kelvin (4-Wire) Clip Lead Set 8607 2-Wire, 1000VBanana Cables, 1m (3.3 ft) CA-18-1 Shielded Dual Banana Cable, 1.2m (4 ft)
SWITCHING HARDWARE
7001 Two-Slot Switch System 7002 Ten-Slot Switch System 7019-C 6-Wire Ohms Switch Card 7053 High-Current Switch Card
CABLES/ADAPTERS
7007-1 Shielded GPIB Cable, 1m (3.3 ft) 7007-2 Shielded GPIB Cable, 2m (6.6 ft) 7009-5 RS-232 Cable
COMMUNICATION INTERFACE
KPCI-488LP IEEE-488 Interface/Controller for the PCI Bus KPXI-488 IEEE-488 Interface Board for the PXI Bus KUSB-488A IEEE-488 USB-to-GPIB Interface Adapter
Keithley’s SourceMeter family is designed specifically for test applications that demand tightly coupled sourcing and measure­ment. All SourceMeter models provide precision voltage and cur­rent sourcing as well as measurement capabilities. Each SourceMeter instrument is both a highly stable DC power source and a true instrument-grade 5
1
2-digit multimeter. The power source characteristics include low noise, precision, and readback. The multimeter capabilities include high repeatability and low noise. The result is a compact, single-channel, DC parametric tester. In operation, these instruments can act as a voltage source, a current source, a voltage meter, a current meter, and an ohm­meter. Manufacturers of components and modules for the com­munications, semiconductor, computer, automotive, and medical industries will find the SourceMeter instruments invaluable for a wide range of characterization and production test applications.
Advantages of a Tightly Integrated Instrument
By linking source and measurement circuitry in a single unit, these instruments offer a variety of advantages over systems con­figured with separate source and measurement instruments. For example, they minimize the time required for test station devel­opment, setup, and maintenance, while lowering the overall cost of system ownership. They simplify the test process itself by elimi­nating many of the complex synchronization and connection
ACCESS ORIES AVAILABLE
TRIGGERING AND CONTROL
2499-DIGIO Digital I/O Expander Assembly 8501-1 Trigger Link Cable, DIN-to-DIN, 1m (3.3 ft) 8501-2 Trigger Link Cable, DIN-to-DIN, 2m (6.6 ft) 8502 Trigger Link to BNCBreakout Box 8503 Trigger Link Cable, DIN-to-Dual BNC, 1m (3.3 ft) 8505 Male to 2-Female Y-DINCable for Trigger Link
RACK MOUNT KITS
4288-1 Single Fixed Rack Mount Kit 4288-2 Dual Fixed Rack Mount Kit
SOFTWARE
LabTracer 2.0 Curve Tracing Software (downloadable)
OTHER
1050 Padded Carrying Case 2400-EW 1-Year Warranty Extension 2410-EW 1-Year Warranty Extension 2420-EW 1-Year Warranty Extension 2425-EW 1-Year Warranty Extension 2430-EW 1-Year Warranty Extension 2440-EW 1-Year Warranty Extension
1.888.KEITHLEY (U.S. only)
SOURCE AND MEASURE
www.keithley.com
A GREATER MEASURE OF CONFIDENCE
Series 2400
Ordering Information
2400 200V, 1A, 20W
2400-C 200V, 1A, 20W
2400-LV 20V, 1A, 20W
2410 1100V, 1A, 20W
2410-C 1100V, 1A, 20W
2420 60V, 3A, 60W
2420-C 60V, 3A, 60W
2425 100V, 3A, 100W
2425-C 100V, 3A, 100W
2430 100V, 10A, 1000W Pulse
2430-C 100V, 10A, 1000W Pulse
2440 40V, 5A, 50W
2440-C 40V, 5A, 50W
Extended warranty, service, and calibration contracts are available
Accessories Supplied
Test Leads LabVIEW Software Driver
LabTracer
TestPoint Software Driver
SourceMeter
SourceMeter with Contact Check
SourceMeter
SourceMeter
SourceMeter with Contact Check
SourceMeter
SourceMeter with Contact Check
SourceMeter
SourceMeter with Contact Check
Mode SourceMeter
Mode SourceMeter with Contact Check
SourceMeter
SourceMeter with Contact Check
(downloadable)
®
Software
(downloadable)
(downloadable)
SourceMeter Line
I-V Characteristics
All SourceMeter instruments provide four-quadrant operation. In the first and third quadrants they operate as a source, delivering power to a load. In the second and fourth quadrants they operate as a sink, dissipating power internally. Voltage, current, and resistance can be measured during source or sink operation.
I
meter
I
source
V
/Compliance
meter
Local
Remote
Remote
Local
IN/OUT HI
SENSE HI
DUT
SENSE LO
IN/OUT LO
Source I, Measure V, I, or Ω configuration Source V, Measure I, V, or Ω configuration
+1A
+100mA
–200V
2400 only
–100mA
–1A
+20V–20V
+200V
2400
only
Duty cycle limited
Model 2400 and 2400-LV SourceMeter
+3A
+1A
+100mA
–60V
.
+20V–20V
–100mA
–1A
–3A
+60V
Duty cycle limited
Model 2420 3A SourceMeter
+10A
+3A
+1A
+100mA
+60
–100V
+20–20–60 +100V
–100mA
–1A
–3A
–10A
Duty cycle limited
Pulse mode only
Model 2430 1kW Pulse Mode SourceMeter
I
meter
V
source
/Compliance
–200V
Feedback to
Adjust V
source
Local
IN/OUT HI
Remote
SENSE HI
V
meter
Remote
SENSE LO
Local
IN/OUT LO
+1A
+200mA
+20mA
+20V–20V
–20mA
–100mA
–1A
+200V
+1100V–1100V
Duty cycle limited
Model 2410 High-Voltage SourceMeter
+3A +1A
+100mA
–100V
–20–60
+20
–100mA
–1A
–3A
+60
+100V
Duty cycle limited
Model 2425 100W SourceMeter
5A
3A 1A
–40V
100mA
–10V +10V
–100mA –1A
-–3A
-5A
40V
Duty cycle limited
Model 2440 5A SourceMeter
DUT
Tightly coupled precision sourcing and measurement
1.888.KEITHLEY (U.S. only)
www.keithley.com
SOURCE AND MEASURE
A GREATER MEASURE OF CONFIDENCE
Series 2400
Test 3
Test 2
V
I
Test 1
I
R
V
F2
V
F1
SourceMeter Line
Automation for Speed
A SourceMeter instrument streamlines production testing. It sources voltage or current while making measurements without needing to change connections. It is designed for reliable operation in non­stop production environments. To provide the throughput demanded by production applications, the SourceMeter instrument offers many built-in features that allow it to run complex test sequences without computer control or GPIB communications slowing things down.
Standard and Custom Sweeps
Sweep solutions greatly accelerate testing with
Stop
automation hooks. Three basic sweep waveforms are provided that can be programmed for single-
Start
event or continuous operation. They are ideal for I/V, I/R, V/I, and V/R characterization.
Bias
• Linear Staircase Sweep: Moves from the start level to the stop level in equal linear steps
Linear staircase sweep
• Logarithmic Staircase Sweep: Done on a log scale with a specified number of steps per
Stop
decade
User
defined
steps
Start
Start
• Custom Sweep: Allows constr uction of special sweeps by specifying the number of measure­ment points and the source level at each point
• Up to 1700 readings/second at 4
1
2 digits to the
GPIB bus
• 5000 5
1
2-digit readings can be stored in the
Bias
Logarithmic staircase sweep
non-volatile buffer memory
Built-In Test Sequencer (Source Memory List)
Tightly coupled precision sourcing and measurement
The Source Memory list provides faster and easi­er testing by allowing you to setup and execute up to 100 different tests that run without PC
Bias
Custom sweep
intervention.
• Stores up to 100 instr ument configurations, each containing source settings, measurement settings, pass/fail criteria, etc.
• Pass/fail limit test as fast as 500µs per point
• Onboard comparator eliminates the delay caused when sending data to the computer for analysis
• Built-in, user definable math functions to calculate derived parameters
Stop
Bias
Bias
Bias
TYPICAL APPLICATIONS Devices:
• Discrete semiconductor devices
• Passive devices
• Transient suppression devices
• ICs, RFICs, MMICs
• Laser diodes, laser diode modules, LEDs, photodetectors
• Circuit protection devices: TVS, MOV, Fuses, etc.
• Airbags
• Connectors, switches, relays
Tests:
• Leakage
• Low voltage/resistances
• LIV
• IDDQ
• I-V characterization
• Isolation and trace resistance
• Temperature coefficient
• Forward voltage, reverse breakdown, leakage current
• DC parametric test
• DC power source
•HIPOT
• Dielectric withstanding
Example Test Sequence
1.888.KEITHLEY (U.S. only)
SOURCE AND MEASURE
www.keithley.com
Test Pass/Fail Test If Passes Test If Fails Test
Test 1 Check VF1at
100mA against pass/fail limits
Test 2 Check VF2at 1A
against pass/fail limits
Test 3 Check leakage
current at –500V and test against pass/fail limits
Go to Test 2
Go to Test 3
1. Bin par t to good bin
2. Transmit readings to com­puter while handler is placing new part
3. Return to Test 1
1. Bin par t to bad bin
2. Transmit data to computer while handler is placing new part
3. Return to Test 1
A GREATER MEASURE OF CONFIDENCE
Series 2400
SourceMeter Line
Digital I/O Interface
The digital I/O interface can link the SourceMeter instrument to many pop­ular component handlers, including Aetrium, Aeco, and Robotronics. Other capabilities of the interface include:
• Tight systems integration for applications such as binning and sorting
• Built-in component handler interface
• Start of test and end of test signals
• 5V, 300mA power supply
• Optional expander accessory (Model 2499-DIGIO) adds 16 digital I/O lines
Trigger Link Interface
All SourceMeter instruments include Keithley’s unique Trigger Link inter­face which provides high-speed, seamless communications with many of Keithley’s other instruments. For example, use the Trigger Link interface to connect a SourceMeter instrument with a Series 7000 Switching System for a complete multi-point test solution. With Trigger Link, the 7000 Series Switching Systems can be controlled by a SourceMeter during a high-speed test sequence independent of a computer and GPIB.
Optional Contact Check Function
The Contact Check function makes it simple to verify good connections quickly and easily before an automated test sequence begins. This elimi­nates measurement errors and false product failures associated with con­tact fatigue, breakage, contamination, loose or broken connection, relay failures, etc. Some capabilities of this function are:
• 350µs verification and notification process time
• The output of the SourceMeter instr ument is automatically shut off after a fault and is not re-activated until good contact is verified, protecting the device under test from damage and the operator from potential safety hazards
• 3 pass/fail threshold values: 2Ω, 15Ω, and 50Ω
• No energy passes through the device under test during the operation
• Enabled either from the front panel or remotely over the GPIB
• 3 fault notification methods
Unique 6-Wire Ohms Technique
SourceMeter instruments can make standard 4-wire, split Kelvin, and 6­wire, guarded ohms measurements and can be configured for either the constant current or constant voltage method. The 6-wire ohms technique:
• Uses guard and guard sense leads in addition to the 4-wire sense and source leads
• Locks out parallel current paths when measuring resistor networks or hybrid circuits to isolate the component under test
• Allows users to configure and plot data easily from Series 2400 SourceMeter instruments, making characterization of two, three, and four terminal devices a snap
+
GUARD
-
I
meter
I
source
6-Wire Ohms Circuit. All test current flows through R1 because the high current guard drives the voltage across R2 to 0V.
V
meter
GUARD SENSE
IN/OUT HI
SENSE HI
R1
SENSE LO
IN/OUT LO
DUT
R2
R3
Tightly coupled precision sourcing and measurement
+
-
I
meter
V or I
V
Source
Contact check option for 4-wire or 6-wire applications
meter
350μs
Contact Check
(optional)
GUARD
GUARD SENSE IN/OUT HI
SENSE HI
SENSE LO IN/OUT LO
1.888.KEITHLEY (U.S. only)
www.keithley.com
Pass
Fail
Pass
Free LabTracer 2.0 device characterization software (downloadable)
SOURCE AND MEASURE
A GREATER MEASURE OF CONFIDENCE
Series 2400
SourceMeter Line
Voltage Accuracy (Local or Remote Sense)
SOURCE
MODEL RANGE
2400, 2400-C
2400-LV
2410, 2410-C
2420, 2420-C
2425, 2425-C
2430, 2430-C
Series 2400 condensed specifications
2440, 2440-C
TEMPERATURE COEFFICIENT (0°–18°C & 28°–50°C): ±(0.15 × accuracy specification)/°C. VOLTAGE REGULATION: Line: 0.01% of range. Load: 0.01% of range + 100µV. OVER VOLTAGE PROTECTION: User selectable values, 5% tolerance. Factory default = none. CURRENT LIMIT:Bipolar current limit (compliance) set with single value. Min. 0.1% of range. OVERSHOOT:<0.1% typical (full scale step, resistive load, 10mA range).
200.000 mV 5µV 0.02% + 600 µV 1µV 0.012% + 300 µV
2.00000 V 50 µV 0.02% + 600 µV 10 µV 0.012% + 300 µV
20.0000 V 500 µV 0.02% + 2.4 mV 100 µV 0.015% + 1.5 mV 0.08 V/µs
200.000 V 5mV 0.02% + 24 mV 1 mV 0.015% + 10 mV 0.5 V/µs
200.000 mV 5µV 0.02% + 600 µV 1µV 0.012% + 300 µV
2.00000 V 50 µV 0.02% + 600 µV 10 µV 0.012% + 300 µV
20.0000 V 500 µV 0.02% + 2.4 mV 100 µV 0.015% + 1.5 mV 0.08 V/µs
200.000 mV 5µV 0.02% + 600 µV 1µV 0.012% + 300 µV
2.00000 V 50 µV 0.02% + 600 µV 10 µV 0.012% + 300 µV
20.0000 V 500 µV 0.02% + 2.4 mV 100 µV 0.015% + 1 mV 0.15 V/µs
1000.00 V 50 mV 0.02% + 100 mV 10 mV 0.015% + 50 mV 0.5 V/µs
200.000 mV 5µV 0.02% + 600 µV 1µV 0.012% + 300 µV
2.00000 V 50 µV 0.02% + 600 µV 10 µV 0.012% + 300 µV
20.0000 V 500 µV 0.02% + 2.4 mV 100 µV 0.015% + 1 mV 0.08 V/µs
60.0000 V 1.5 mV 0.02% + 7.2 mV 1 mV 0.015% + 3 mV 0.14 V/µs
200.000 mV 5µV 0.02% + 600 µV 1µV 0.012% + 300 µV
2.00000 V 50 µV 0.02% + 600 µV 10 µV 0.012% + 300 µV
20.0000 V 500 µV 0.02% + 2.4 mV 100 µV 0.015% + 1 mV 0.08 V/µs
100.0000 V 2.5 mV 0.02% + 12 mV 1 mV 0.015% + 5 mV 0.25 V/µs
200.000 mV 5µV 0.02% + 600 µV 1µV 0.012% + 300 µV
2.00000 V 50 µV 0.02% + 600 µV 10 µV 0.012% + 300 µV
20.0000 V 500 µV 0.02% + 2.4 mV 100 µV 0.015% + 1 mV 0.08 V/µs
100.0000 V 2.5 mV 0.02% + 12 mV 1 mV 0.015% + 5 mV 0.25 V/µs
200.000 mV 5µV 0.02% + 600 µV 1µV 0.012% + 300 µV
2.00000 V 50 µV 0.02% + 600 µV 10 µV 0.012% + 300 µV
10.0000 V 500 µV 0.02% + 1.2 mV 100 µV 0.015% + 750 µV 0.08 V/µs
40.0000 V 5mV 0.02% + 4.8 mV 1 mV 0.015% + 3 mV 0.25 V/µs
PROGRAMMING
RESOLUTION
ADDITIONAL SOURCE SPECIFICATIONS (All Models)
TRANSIENT RESPONSE TIME: 30µs minimum for the output to recover to its spec. following a step
change in load.
COMMAND PROCESSING TIME: Maximum time required for the output to begin to change follow-
ing the receipt of :SOURce:VOLTage|CURRent <nrf> command. Autorange On: 10ms. Autorange
Off: 7ms.
OUTPUT SETTLING TIME: Time required to reach 0.1% of final value after command is processed.
100µs typical. Resistive load. 10µA to 100mA range.
DC FLOATING VOLTAGE: Output can be floated up to ±250VDC (Model 2440 ±40VDC) from chas-
sis ground.
REMOTE SENSE: Up to 1V drop per load lead. COMPLIANCE ACCURACY:Add 0.3% of range and ±0.02% of reading to base specification. OVER TEMPERATURE PROTECTION: Internally sensed temperature overload puts unit in standby
mode.
RANGE CHANGE OVERSHOOT:Overshoot into a fully resistive 100kΩ load, 10Hz to 1MHz BW, adja-
cent ranges: 100mV typical, except 20V/200V (20V/60V on Model 2420), 20V/100V on Model 2425 and 2430, range boundary, and Model 2440.
MINIMUM COMPLIANCE VALUE:0.1% of range.
ACCURACY (1 Year)
23°C ±5°C
±(% rdg. + volts)
1
DEFAULT
MEASUREMENT
RESOLUTION
MEASUREMENT
ACCURACY (1 Year)
23°C ±5°C
±(% rdg. + volts)
2, 3, 4
OUTPUT SLEW
RATE (±30%)
SOURCE/SINK
LIMIT
±21 V @ ±1.05 A
±210 V @ ±105 mA
±21 V @ ±1.05 A
±21 V @ ±1.05 A
±1100 V @ ±21 mA
±21 V @ ±3.15 A ±63 V @ ±1.05 A
±21 V @ ±3.15 A
±105 V @ ±1.05 A
±105 V @ ±1.05 A
±105 V @ ±10.5 A
(pulse mode only)
±10.5 V @ ±5.25 A
±42 V @ ±1.05 A
ADDITIONAL PULSE MODE SOURCE SPECIFICATIONS (2430 and 2430-C only)
MAXIMUM DUTY CYCLE:8%, hardware limited, 10A range only. All other ranges 100%. MAXIMUM PULSE WIDTH: 5ms from 90% rising to 90% falling edge, 2.5ms 10A range. MINIMUM PULSE WIDTH: 150µs. MINIMUM PULSE RESOLUTION: 50µs typical, 70µs max., limited by system jitter. SOURCE ACCURACY:Determined by settling time and source range specifications. OUTPUT SETTLING TIME 0.1%:
800µs typ., source I = 10A into 10Ω, limited by voltage slew rate. 500µs typ., source I = 10A into 1Ω, limited by voltage slew rate.
OUTPUT SLEW RATE:
Voltage (10Ω load): 0.25V/µs ±30% on 100V range. 0.08V/µs ±30% on 20V range, 10A range. Current (0Ω load): 0.25A/µs ±30% on 100V range. 0.08A/µs ±30% on 20V range, 10Arange.
NOTES
1. 2400, 2410 Only: Specifications valid for continuous output currents below 105mA. For operation above 105mA con­tinuous for >1 minute, derate accuracy 10%/35mA above 105mA.
2. Speed = Normal (1 PLC). For 0.1 PLC, add 0.005% of range to offset specifications, except 200mV, 1A, 10A ranges, add 0.05%. For 0.01 PLC, add 0.05% of range to offset specifications, except 200mV, 1A, 10A ranges, add 0.5%.
3. Accuracies apply to 2- or 4-wire mode when properly zeroed.
4. In pulse mode, limited to 0.1 PLC measurement.
1.888.KEITHLEY (U.S. only)
SOURCE AND MEASURE
www.keithley.com
A GREATER MEASURE OF CONFIDENCE
Series 2400
SourceMeter Line
Current Accuracy (Local or Remote Sense)
1, 3
SOURCE ACCURACY
MODEL RANGE
2400, 2400-C, 2400-LV
2410, 2410-C
2420, 2420-C
2425, 2425-C
2430, 2430-C
2440, 2440-C
TEMPERATURE COEFFICIENT (0°–18°C & 28°–50°C): ±(0.15 × accuracy specification)/°C. CURRENT REGULATION: Line: 0.01% of range. Load: 0.01% of range (except Model 2440 5A range
0.05%) + 100pA.
VOLTAGE LIMIT:Bipolar voltage limit (compliance) set with single value. Min. 0.1% of range. OVERSHOOT:<0.1% typical (1mA step, RL = 10kΩ, 20V range for Model 2400, 2410, 2420, 2425,
2430), (10V range for Model 2440).
1.00000 µA 50 pA 0.035% + 600 pA 10 pA 0.029% + 300 pA
10.0000 µA 500 pA 0.033% + 2 nA 100 pA 0.027% + 700 pA
100.000 µA 5 nA 0.031% + 20 nA 1 nA 0.025% + 6 nA
1.00000 mA 50 nA 0.034% + 200 nA 10 nA 0.027% + 60 nA
10.0000 mA 500 nA 0.045% + 2 µA 100 nA 0.035% + 600 nA
100.000 mA 5 µA 0.066% + 20 µA 1 µ A 0.055% + 6 µA
2
1.00000 A
1.00000 µA 50 pA 0.035% + 600 pA 10 pA 0.029% + 300 pA
10.0000 µA 500 pA 0.033% + 2 nA 100 pA 0.027% + 700 pA
100.000 µA 5 nA 0.031% + 20 nA 1 nA 0.025% + 6 nA
1.00000 mA 50 nA 0.034% + 200 nA 10 nA 0.027% + 60 nA
20.0000 mA 500 nA 0.045% + 4 µA 100 nA 0.035% + 1.2 µA
100.000 mA 5 µA 0.066% + 20 µA 1 µ A 0.055% + 6 µA
2
1.00000 A
10.0000 µA 500 pA 0.033% + 2 nA 100 pA 0.027% + 700 pA
100.000 µA 5 nA 0.031% + 20 nA 1 nA 0.025% + 6 nA
1.00000 mA 50 nA 0.034% + 200 nA 10 nA 0.027% + 60 nA
10.0000 mA 500 nA 0.045% + 2 µA 100 nA 0.035% + 600 nA
100.000 mA 5 µA 0.066% + 20 µA 1 µ A 0.055% + 6 µA
2
1.00000 A
2
3.00000 A
10.0000 µA 500 pA 0.033% + 2 nA 100 pA 0.027% + 700 pA
100.000 µA 5 nA 0.031% + 20 nA 1 nA 0.025% + 6 nA
1.00000 mA 50 nA 0.034% + 200 nA 10 nA 0.027% + 60 nA
10.0000 mA 500 nA 0.045% + 2 µA 100 nA 0.035% + 600 nA
100.000 mA 5 µA 0.066% + 20 µA 1 µ A 0.055% + 6 µA
2
1.00000 A
2
3.00000 A
10.0000 µA 500 pA 0.033% + 2 nA 100 pA 0.027% + 700 pA
100.000 µA 5 nA 0.031% + 20 nA 1 nA 0.025% + 6 nA
1.00000 mA 50 nA 0.034% + 200 nA 10 nA 0.027% + 60 nA
10.0000 mA 500 nA 0.045% + 2 µA 100 nA 0.035% + 600 nA
100.000 mA 5 µA 0.066% + 20 µA 1 µ A 0.055% + 6 µA
1.00000 A 50 µA 0.067% + 900 µA 10 µA 0.060% + 570 µA
2
3.00000 A
4
10.00000 A
10.0000 µA 500 pA 0.033% + 2 nA 100 pA 0.027% + 700 pA
100.000 µA 5 nA 0.031% + 20 nA 1 nA 0.025% + 6 nA
1.00000 mA 50 nA 0.034% + 200 nA 10 nA 0.027% + 60 nA
10.0000 mA 500 nA 0.045% + 2 µA 100 nA 0.035% + 600 nA
100.000 mA 5 µA 0.066% + 20 µA 1 µA 0.055% + 6 µA
1.00000 A 50 µA 0.067% + 900 µA 10 µA 0.060% + 570 µ A
5.00000 A 50 µA 0.10 % + 5.4 mA 10 µA 0.10 % + 3.42 mA
PROGRAMMING
RESOLUTION
50 µA 0.27 % + 900 µA 10 µA 0.22 % + 570 µA
50 µA 0.27 % + 900 µA 10 µA 0.22 % + 570 µA
50 µA 0.067% + 900 µA 10 µA 0.066% + 570 µ A 50 µA 0.059% + 2.7 mA 10 µA 0.052% + 1.71 mA
50 µA 0.067% + 900 µA 10 µA 0.060% + 570 µ A 50 µA 0.059% + 2.8 mA 10 µA 0.052% + 1.71 mA
500 µA 500 µA
CONTACT CHECK SPECIFICATIONS (requires -C version)
SPEED: 350µs for verification and notification. CONTACT CHECK: 2 Ω 15 Ω 50 Ω
No contact check failure <1.00 Ω <13.5 Ω <47.5 Ω Always contact check failure >3.00 Ω >16.5 Ω >52.5 Ω
(1 Year)
23°C ±5°C
±(% rdg. + amps)
0.059% + 2.8 mA
0.089% + 5.9 mA
3
DEFAULT
MEASUREMENT
RESOLUTION
10 µA 10 µA
MEASUREMENT
ACCURACY (1 Year)
23°C ±5°C
±(% rdg. + amps)
0.052% + 1.71 mA
0.082% + 1.71 mA
5, 6, 7
SOURCE/SINK LIMIT
±1.05A @ ±21 V
±105 mA @ ±210 V
±1.05A @ ±21 V
±21 mA @ ±1100 V
±3.15A @ ±21 V
±1.05 A @ ±63 V
±3.15A @ ±21 V
±1.05 A @ ±105 V
±1.05A @ ±105 V
±10.5 A @ ±105 V
(pulse mode only)
±5.25A @ ±10.5 V
±1.05 A @ ±42 V
NOTES
1. 2400, 2410 Only: Specifications valid for continuous output currents below 105mA. For operation above 105mA con-
tinuous for >1 minute, derate accuracy 10%/35mA above 105mA.
2. Full operation (1A) regardless of load to 30°C (50°C for Model 2420 and 2440). Above 30°C (50°C for Model 2420
and 2440) ambient, derate 35mA/°C and prorate 35mA/Ω load. 4-wire mode. For current sink operation on 1A, 3A, or 5A ranges, maximum continuous power is limited to approximately 1/2 rated power or less, depending on cur­rent, up to 30°C ambient. See power equations in the User’s Manual to calculate allowable duty cycle for specific conditions.
3. For sink mode, 1µA to 100mA range, accuracy is:
Model 2400: ±(0.15% + offset*4). Models 2410, 2420, 2425, 2430, 2440: ±(0.5% + offset*3). For 1Arange, accuracy is:
Model 2400: ±(1.5% + offset*8). Models 2410, 2420, 2425, 2430, 2440: ±(1.5% + offset*3).
4. 10A range only in pulse mode. Limited to 2.5ms pulse width maximum. 10% duty cycle maximum.
5. Speed = Normal (1 PLC). For 0.1 PLC, add 0.005% of range to offset specifications, except 200mV, 1A, 10A ranges,
add 0.05%. For 0.01 PLC, add 0.05% of range to offset specifications, except 200mV, 1A, 10A ranges, add 0.5%.
6. Accuracies apply to 2- or 4-wire mode when properly zeroed.
7. In pulse mode, limited to 0.1 PLC measurement.
8. Model 2400 and 2400-C only.
8
Series 2400 condensed specifications
1.888.KEITHLEY (U.S. only)
www.keithley.com
SOURCE AND MEASURE
A GREATER MEASURE OF CONFIDENCE
Series 2400
SourceMeter Line
RESISTANCE MEASUREMENT Accuracy (Local or Remote Sense)
RANGE
<0.20000 Ω
2.00000 Ω
RESOLUTION
3 3
10 µΩ
20.0000 Ω 100 µΩ
200.000 Ω 1mΩ
2.00000 kΩ 10 mΩ
20.0000 kΩ 100 mΩ
200.000 kΩ 1 Ω
6
DEFAULT
2.00000 MΩ
20.0000 MΩ
200.000 MΩ
>200.000 MΩ
10 Ω
7
100 Ω
3 3
1kΩ
CURRENT
2400, 2410
1A
100 mA 100 mA
10 mA 10 mA
1mA 1mA
100 µA 100 µA
10 µA 10 µA
1µA 1µA 1µA 1µA
100 nA
TEMPERATURE COEFFICIENT (0°–18°C & 28°–50°C): ±(0.15 × accuracy specifica-
tion)/°C.
SOURCE I MODE, MANUAL OHMS: Total uncertainty = I source accuracy + V measure
accuracy (4-wire remote sense).
SOURCE V MODE, MANUAL OHMS: Total uncertainty = V source accuracy + I measure
accuracy (4-wire remote sense).
6-WIRE OHMS MODE:Available using active ohms guard and guard sense. Max.
Guard Output Current: 50mA (except 1A range). Accuracy is load dependent. Refer to White Paper no. 2033 for calculation formula.
Series 2400 condensed specifications
GUARD OUTPUT IMPEDANCE:<0.1Ω in ohms mode.
DEFAULT TEST
DEFAULT TEST
CURRENT
2420, 2425 ,
2430, 2440
2400 2410 2420, 2425, 2430, 2440 2400
Source I
Source I
ACC
ACC
+ Meas. V
+ Meas V
ACC ACC
0.10% + 0.003 Ω 0.11% + 0.006Ω 0.10% + 0.003 Ω 0.07% + 0.001 Ω
0.08% + 0.03 Ω 0.09% + 0.1 Ω 0.08% + 0.03 Ω 0.05% + 0.01 Ω
0.07% + 0.3 Ω 0.08% + 0.6 Ω 0.07% + 0.3 Ω 0.05% + 0.1 Ω
0.06% + 3 Ω 0.07% + 6 Ω 0.06% + 3 Ω 0.04% + 1 Ω
0.07% + 30 Ω 0.07% + 60 Ω 0.07% + 30 Ω 0.05% + 10 Ω
0.11% + 300 Ω 0.12% + 600 Ω 0.11% + 300 Ω 0.05% + 100 Ω
0.11% + 1 kΩ 0.12% + 2.4 kΩ 0.11% + 1 kΩ 0.05% + 500 Ω
0.66% + 10 kΩ 0.66% + 24 kΩ
Source I
ACC
+ Meas. V
ACC
NOTES:
1. Speed = Normal (1 PLC). For 0.1 PLC, add 0.005% of range to offset specifications, except 200mV, 1A, 10A ranges, add 0.05%. For
0.01 PLC, add 0.05% of range to offset specifications, except 200mV, 1A, 10A ranges, add 0.5%.
2. Accuracies apply to 2- or 4-wire mode when properly zeroed.
3. Manual ohms only – except 2420, 2425, 2430, 2440 for 2Ω range and 2410 or 2400 for 200MΩ range.
4. Source readback enabled, offset compensation ON. Also available on 2410, 2420, 2425, 2430, and 2440 with similar accuracy enhancement.
5. In pulse mode, limited to 0.1 PLC measurement.
6. Except 2440; default test current is 5µA.
7. Except 2440; default test current is 0.5µA.
1, 2, 5
NORMAL ACCURACY (23°C ±5°C)
1 Year, ±(% rdg. + ohms)
Source I Source I
ACC ACC
+ Meas. V + Meas. V
Source I
ACC ACC
Source I
Source I
ACC
+ Meas. V
Source I
ACC
+ Meas. V
ACC
0.17% + 0.0003Ω
+ Meas. V
ACC
+ Meas. V
ACC
ENHANCED ACCURACY
(23°C ±5°C)
1 Year, ±(% rdg. + ohms)
ACC
Source I Source I
ACC ACC
0.35% + 5 kΩ
Source I
ACC ACC
ACC
4
+ Meas. V + Meas. V
+ Meas. V
ACC ACC
ACC
1.888.KEITHLEY (U.S. only)
SOURCE AND MEASURE
www.keithley.com
A GREATER MEASURE OF CONFIDENCE
Series 2400
SourceMeter Line
System Speeds
MEASUREMENT
MAXIMUM RANGE CHANGE RATE: 75/second. MAXIMUM MEASURE AUTORANGE TIME: 40ms (fixed source).
Sweep Operation3Reading Rates (rdg./second) for 60Hz (50Hz):
SPEED NPLC/TRIGGER ORIGIN TO MEM. TO GPI B TO MEM. TO GPIB TO MEM. TO GPIB TO MEM. TO GPIB
Fast 0.01 / internal 2081 (2030) 1754 1551 (1515) 1369 902 (900) 981 165 (162) 165 IEEE-488.1 Mode 0.01 / external 1239 (1200) 1254 1018 (990) 1035 830 (830) 886 163 (160) 163 Fast 0.01 / internal 2081 (2030) 1198 (1210) 1551 (1515) 1000 (900) 902 (900) 809 (840) 165 (162) 164 (162) IEEE-488.2 Mode 0.01 / external 1239 (1200) 1079 (1050) 1018 (990) 916 (835) 830 (830) 756 (780) 163 (160) 162 (160) Medium 0.10 / internal 510 (433) 509 (433) 470 (405) 470 (410) 389 (343) 388 (343) 133 (126) 132 (126)
IEEE-488.2 Mode 0.10 / external 438 (380) 438 (380) 409 (360) 409 (365) 374 (333) 374 (333) 131 (125) 131 (125) Normal 1.00 / internal 59 (49) 59 (49) 58 (48) 58 (48) 56 (47) 56 (47) 44 (38) 44 (38)
IEEE-488.2 Mode 1.00 / external 57 (48) 57 (48) 57 (48) 57 (47) 56 (47) 56 (47) 44 (38) 44 (38)
1
2
MEASURE SOURCE-MEASURE
5
SOURCE-MEASURE PASS/FAIL TEST
4, 5
SOURCE-MEMORY
4
Single Reading Operation Reading Rates (rdg./second) for 60Hz (50Hz):
SPEED NPLC/TRIGGER ORIGIN TO GPIB TO GPIB TO GPIB
MEASURE SOURCE-MEASURE
Fast (488.1) 0.01 / internal 537 140 135 Fast (488.2) 0.01 / internal 256 (256) 79 (83) 79 (83) Medium (488.2) 0.10 / internal 167 (166) 72 (70) 69 (70) Normal (488.2) 1.00 / internal 49 (42) 34 (31) 35 (30)
Component for 60Hz (50Hz):
4, 6
MEASURE SOURCE SOURCE-MEASURE PASS/FAIL TEST
SPEED NPLC/TRIGGER ORIGIN TO GPIB PASS/FAIL TEST TO GPIB
Fast 0.01 / external 1.04 ms (1.08 ms) 0.5 ms (0.5 ms) 4.82 ms (5.3 ms) Medium 0.10 / external 2.55 ms (2.9 ms) 0.5 ms (0.5 ms) 6.27 ms (7.1 ms) Normal 1.00 / external 17.53 ms (20.9 ms) 0.5 ms (0.5 ms) 21.31 ms (25.0 ms)
1
Reading rates applicable for voltage or current measurements. Auto zero off, autorange off, filter off, display off, trig­ger delay = 0, and binary reading format.
2
Purely resistive lead. 1µA and 10µA ranges <65ms.
3
1000 point sweep was characterized with the source on a fixed range.
4
Pass/Fail test performed using one high limit and one low math limit.
5
Includes time to re-program source to a new level before making measurement.
6
Time from falling edge of START OF TEST signal to falling edge of END OF TESTsignal.
7
Command processing time of :SOURce:VOLTage|CURRent:TRIGgered <nrf> command not included.
5
SOURCE-MEASURE PASS/FAI L TEST
GENERAL
Noise Rejection:
NPLC NMRR CMRR
Fast 0.01 80 dB Medium 0.1 80 dB Slow 1 60 dB 100 dB
1
Except lowest 2 current ranges = 90dB.
LOAD IMPEDANCE: Stable into 20,000pF typical. COMMON MODE VOLTAGE: 250V DC (40V DC for Model 2440).
9
COMMON MODE ISOLATION: >10
Ω, <1000pF.
OVERRANGE: 105% of range, source and measure. MAX. VOLTAGE DROP BETWEEN INPUT/OUTPUT AND SENSE TERMINALS: 5V. MAX. SENSE LEAD RESISTANCE:1MΩ for rated accuracy.
10
SENSE INPUT IMPEDANCE: >10
Ω.
GUARD OFFSET VOLTAGE:<150µV, typical (300µV for Models 2430, 2440). SOURCE OUTPUT MODES:
Pulse (Model 2430 only) Fixed DC level Memory List (mixed function) Stair (linear and log)
MEMORY BUFFER: 5,000 readings @ 5 digits (two 2,500 point buffers). Includes selected measured
value(s) and time stamp. Lithium battery backup (3 yr+ battery life).
1
SOURCE MEMORY LIST: 100 points max. PROGRAMMABILITY: IEEE-488 (SCPI-1995.0), RS-232, 5 user-definable power-up states plus fac-
tory default and *RST.
DIGITAL INTERFACE:
Interlock: Active low input. Handler Interface: Start of test, end of test, 3 category bits. +5V@ 300mA supply. Digital I/O: 1 trigger input, 4 TTL/Relay Drive outputs (33V @ 500mA, diode clamped).
POWER SUPPLY: 100V to 240V rms, 50–60Hz (automatically detected at power up). Model
2400: 190VA. Model 2410:210VA. Model 2420: 220VA. Model 2425, 2430: 250VA. Model
2440: 240VA. COOLING: (Model 2410, 2420, 2425, 2430, 2440): Forced air, variable speed. WARRANTY: 1 year. EMC: Conforms to European Union Directive 89/336/EEC, EN 61326-1. SAFETY: Conforms to European Union Directive 73/23/EEC, EN61010-1. VIBRATION: MIL-PRF-28800F Class 3 Random. WARM-UP:1 hour to rated accuracies.
1
DIMENSIONS: 89mm high × 213mm wide × 370mm deep (3
Configuration (with handle & feet):104mm high × 238mm wide × 370mm deep (4
3
8 in × 149⁄16 in).
9
2 in × 83⁄8 in × 149⁄16 in). Bench
WEIGHT:3.21kg (7.08 lbs) (Model 2425, 2430, 2440: 4.1kg, 9.0 lbs). ENVIRONMENT: Operating:0°–50°C, 70%R.H. up to 35°C. Derate 3% R.H./°C, 35°–50°C.
Storage: –25°C to 65°C.
1
8 in ×
4,5
5, 7
Series 2400 condensed specifications
1.888.KEITHLEY (U.S. only)
www.keithley.com
SOURCE AND MEASURE