Keithley Instruments 2302 Data Sheet

Page 1
2302 Battery Simulator
DC VOLTAGE OUTPUT (2 YEARS,23°C ± 5°C)
OUTPUT VOLTAGE: 0 to +15VDC.
OUTPUT ACCURACY: ±(0.05% + 3mV).
PROGRAMMING RESOLUTION: 1mV.
READBACK ACCURACY
1
READBACK RESOLUTION: 1mV.
OUTPUT VOLTAGE SETTLING TIME: 5ms to within stated accuracy.
LOAD REGULATION: 0.01% + 2mV.
LINE REGULATION: 0.5mV.
STABILITY
MEASUREMENT TIME CHOICES: 0.01 to 10 PLC
2
: 0.01% + 0.5mV.
7
, in 0.01PLC steps.
AVERAGE READINGS: 1 to 10.
READING TIME
1, 8,9
: 31ms, typical.
TRANSIENT RESPONSE: High Bandwidth Low Bandwidth
Transient Recovery Time
13
<40µs3or <60µs
Transient Voltage Drop <75mV3or <100mV4<250mV3or <400mV
4
<80µs3or <100µs
4
4
REMOTE SENSE: 1V max. drop in each lead. Add 2mV to the voltage load regulation
specification for each 1V change in the negative output lead due to load current change. Remote sense required. Integrity of connection continually monitored. If compromised, output will turn off automatically once settable window (±0 to ±8 volts) around normal voltage exceeded.
VARIABLE OUTPUT IMPEDANCE
RANGE: 0 to 1.00in 0.01steps. Value can be changed with output on.
DC CURRENT (2 Years, 23°C ± 5°C)
OUTPUT CURRENT:
0–4V: 5A max. >4V: I
= 60W/(V
MAX
SOURCE COMPLIANCE ACCURACY: ±(0.16% + 5mA)
PROGRAMMED SOURCE COMPLIANCE RESOLUTION: 1.25mA.
READBACK ACCURACY
READBACK RESOLUTION: 5A Range: 100µA.
CURRENT SINK CAPACITY:
0–5V: 3A max. 5V–15V: Derate 0.2A per volt above 5V. Compliance setting controls sinking.
LOAD REGULATION: 0.01% + 1mA.
LINE REGULATION: 0.5mA.
STABILITY
4
: 0.01% + 50µA.
MEASUREMENT TIME CHOICES: 0.01 to 10 PLC
AVERAGE READINGS: 1 to 10.
READING TIME
+6) (not intended to be operated in parallel).
OUT
1
: 5A Range: ±(0.2% + 200µA).
5mA Range: ±(0.2% + 1µA).
5mA Range: 0.1µA.
7
1, 8,9
: 31ms, typical.
5
.
, in 0.01PLC steps.
PULSE CURRENT MEASUREMENT OPERATION
TRIGGER LEVEL: 5A Range: 5mA to 5A, in 5mA steps.
TRIGGER DELAY: 0 to 100ms, in 10µs steps.
INTERNAL TRIGGER DELAY: 15µs.
HIGH/LOW/AVERAGE MODE:
Measurement Aperture Settings: 33.3µs to 833ms, in 33.3µs steps. Average Readings: 1 to 100.
PULSE CURRENT MEASUREMENT ACCURACY
100 µs – 200 µs 0.2% + 900 µA + 1.5mA 200 µs – 500 µs 0.2% + 900 µA + 1 mA
500 µs – <1 PLC 0.2% + 600 µA + 0.8mA
1A Range: 1mA to 1A, in 1mA steps. 100mA Range: 0.1mA to 100mA, in 100µA steps.
11
(2 Years,23°C ±5°C):
APERTURE ACCURACY ±(% reading + offset + rms noise10)
<100 µs 0.2% + 900 µA + 2 mA
12
1 PLC
0.2% + 400 µA + 0 mA
>1 PLC 0.2% + 400 µA + 100 µA
BURST MODE CURRENT MEASUREMENT
MEASUREMENT APERTURE: 33.3µs.
CONVERSION RATE: 3650/second, typical.
INTERNAL TRIGGER DELAY: 15µs.
NUMBER OF SAMPLES: 1 to 5000.
TRANSFER SAMPLES ACROSS IEEE BUS IN BINARY MODE: 4800 bytes/s, typical.
LONG INTEGRATION MODE CURRENT MEASUREMENT
MEASUREMENT TIME6: 850ms (840ms) to 60 seconds in 1ms steps.
DIGITAL VOLTMETER INPUT (2 Years,23°C ± 5°C)
INPUT VOLTAGE RANGE: –5 to +30VDC.
INPUT IMPEDANCE: 2Mtypical.
MAXIMUM VOLTAGE (either input terminal) WITH RESPECT TO OUTPUT LOW:
–5V, +30V.
READING ACCURACY
READING RESOLUTION: 1mV.
CONNECTOR: HI and LO input pair part of Output #1’s terminal block.
MEASUREMENT TIME CHOICES: 0.01 to 10 PLC
AVERAGE READINGS: 1 to 10.
READING TIME
1
1, 8,9
: 31ms, typical.
7
, in 0.01PLC steps.
Page 2
2302 Battery Simulator
GENERAL
ISOLATION (low–earth): 22VDC max. PROGRAMMING: IEEE-488.2 (SCPI). USER-DEFINABLE POWER-UP STATES: 5. REAR PANEL CONNECTORS: 8-position quick disconnect terminal block for output (4), sense
(2), and DVM (2).
TEMPERATURE COEFFICIENT(outside 23°C ±5°C): Derate accuracy specification by (0.1 × spec-
ification)/°C.
OPERATING TEMPERATURE: 0° to 50°C (Derate to 70%). 0° to 35°C (Full power). STORAGE TEMPERATURE: –20° to 70°C. HUMIDITY: <80% @ 35°C non-condensing. DISPLAY TYPE: 2-line × 16 character VFD. REMOTE DISPLAY/KEYPAD OPTION: Disables standard front panel. DIMENSIONS: 89mm high × 213mm wide × 411mm deep (3 NET WEIGHT: 3.2kg (7.1 lbs). SHIPPING WEIGHT: 5.4kg (12 lbs). INPUT POWER: 100–120VAC/220–240VAC, 50 or 60Hz (auto detected at power-up). POWER CONSUMPTION: 150VA max. WARRANTY: Two years parts and labor on materials and workmanship. EMC: Conforms with European Union Directive directive 89/336/EEC, EN 55011, EN 50082-1,
EN 61000-3-2 and 61000-3-3, FCC part 15 class B.
SAFETY: Conforms with European Union Directive 73/23/EEC, EN 61010-1. AC LINE LEAKAGE CURRENT: 450µA @ 110VAC, typ.; 600µA @ 220V, typ. RELAY CONTROL PORT: 4-channel, each capable of 100mA sink, 24V max. Total port sink
capacity (all 4 combined) is 250mA max. Accepts DB-9 male plug.
ACCESSORIES SUPPLIED: User and service manual, output connectors mating terminal (part
no. CS-846).
ACCESSORIES AVAILABLE:
Model 2304-DISP: Remote LCD Display/Keypad (4.6 in × 2.7 in × 1.5 in). Includes 2.7m (9 ft)
cable and rack mount kit.
1
⁄2in × 83⁄8 in × 163⁄16in).
1
PLC = 1.00.
2
Following 15 minute warm-up, the change in output over 8 hours under ambient temperature, constant load, and line operating conditions.
3
Remote sense, at output terminals, 0.5A to 5A typical.
4
Remote sense, with 4.5m (15 ft) of 16 gauge (1.31mm2) wire and 1Ω resist- ance in each lead to simulate typical test environment, 1.5A load change (0.15A to 1.65A).
5
Minimum current in constant current mode is 6mA.
6
60Hz (50Hz).
7
PLC = Power Line Cycle. 1PLC = 16.7ms for 60Hz operation, 20ms for 50Hz operation.
8
Display off.
9
Speed includes measurement and binary data transfer out of GPIB.
10
Typical values, peak-to-peak noise equals 6 times rms noise.
11
Based on settled signal: 100µs pulse trigger delay.
12
Also applies to other apertures that are integer multiples of 1PLC.
13
Recovery to within 20mV of previous level.
Specifications are subject to change without notice.
Page 3
2302-PJ Battery Simulator
DC VOLTAGE OUTPUT (2 YEARS,23°C ± 5°C)
OUTPUT VOLTAGE: 0 to +15VDC.
OUTPUT ACCURACY: ±(0.05% + 3mV).
PROGRAMMING RESOLUTION: 1mV.
READBACK ACCURACY
1
: ±(0.05% + 3mV).
READBACK RESOLUTION: 1mV.
OUTPUT VOLTAGE SETTLING TIME: 5ms to within stated accuracy.
LOAD REGULATION: 0.01% + 2mV.
LINE REGULATION: 0.5mV.
STABILITY
MEASUREMENT TIME CHOICES: 0.01 to 10 PLC
2
: 0.01% + 0.5mV.
7
, in 0.01PLC steps.
AVERAGE READINGS: 1 to 10.
READING TIME
1, 8,9
: 31ms, typical.
TRANSIENT RESPONSE: High Bandwidth Low Bandwidth
Transient Recovery Time
13
<40µs3or <60µs
Transient Voltage Drop <75mV3or <100mV4<250mV3or <400mV
4
<80µs3or <100µs
4
4
REMOTE SENSE 1V max. drop in each lead. Add 2mV to the voltage load regulation
specification for each 1V change in the negative output lead due to load current change. Remote sense required. Integrity of connection continually monitored. If compromised, output will turn off automatically once settable window (±0 to ±8 volts) around normal voltage exceeded.
VARIABLE OUTPUT IMPEDANCE
RANGE: 0 to 1.00in 0.01steps. Value can be changed with output on.
DC CURRENT (2 Years, 23°C ± 5°C)
CONTINUOUS AVERAGE OUTPUT CURRENT:
0–4V: 5A max.
>4V: I
= 60W / (V
MAX
Peak currents can be a maximum of 5A provided the average current is within the above limits.
CONTINUOUS AVERAGE SINK CURRENT:
0–5V: 3A max. 5–15V: Derate 0.2A per volt above 5V. Compliance setting controls sinking.
SOURCE COMPLIANCE ACCURACY: ±(0.16% + 5mA)
PROGRAMMED SOURCE COMPLIANCE RESOLUTION: 1.25mA.
READBACK ACCURACY
READBACK RESOLUTION: 5A Range: 100µA.
LOAD REGULATION: 0.01% + 1mA.
LINE REGULATION: 0.5mA.
STABILITY
4
: 0.01% + 50µA.
MEASUREMENT TIME CHOICES: 0.01 to 10 PLC
AVERAGE READINGS: 1 to 10.
READING TIME
1, 8,9
+ 6) (Not intended to be operated in parallel).
SET
5
.
1
: 5A Range: ±(0.2% + 200µA).
500mA Range: ±(0.2% + 20µA).
500mA Range: 10µA.
7
, in 0.01PLC steps.
: 31ms, typical.
PULSE CURRENT MEASUREMENT OPERATION
TRIGGER LEVEL:
5A Current Range
5A Range: 5mA to 5A, in 5mA steps. 1A Range: 1mA to 1A, in 1mA steps. 100mA Range: 0.1mA to 100mA, in 100µA steps.
500mA Current Range
500mA Range: 0.5mA to 500mA, in 0.5mA steps. 100mA Range: 0.1mA to 100mA, in 100µA steps. 10mA Range: 100µA to 10mA, in 100µA steps.
TRIGGER DELAY: 0 to 100ms, in 10µs steps.
INTERNAL TRIGGER DELAY: 15µs.
HIGH/LOW/AVERAGE MODE:
Measurement Aperture Settings: 33.3µs to 833ms, in 33.3µs steps. Average Readings: 1 to 100.
PULSE CURRENT MEASUREMENT ACCURACY
11
(2 Years,23°C ±5°C):
ACCURACY ±(% reading + offset + rms noise10)
APERTURE 5A RANGE 500mA RANGE
<100 µs 0.2% + 900 µA + 2 mA 0.2% + 90 µA + 2 mA 100 µs – 200 µs 0.2% + 900 µA + 1.5 mA 0.2% + 90 µA + 1.5 mA 200 µs – 500 µs 0.2% + 900 µA + 1 mA 0.2% + 90 µA + 1 mA
500 µs – <1 PLC 0.2% + 600 µA + 0.8 mA 0.2% + 60 µA + 0.8 mA
12
1 PLC
0.2% + 400 µA + 0 mA 0.2% + 40 µA + 0 mA
>1 PLC 0.2% + 400 µA + 100 µA 0.2% + 40 µA + 100 µA
BURST MODE CURRENT MEASUREMENT
MEASUREMENT APERTURE: 33.3µs.
CONVERSION RATE: 3650/second, typical.
INTERNAL TRIGGER DELAY: 15µs.
NUMBER OF SAMPLES: 1 to 5000.
TRANSFER SAMPLES ACROSS IEEE BUS IN BINARY MODE: 4800 bytes/s, typical.
LONG INTEGRATION MODE CURRENT MEASUREMENT
Available on both 5A and 500mA current ranges.
MEASUREMENT TIME
6
: 850ms (840ms) to 60 seconds in 1ms steps.
DIGITAL VOLTMETER INPUT (2 Years,23°C ± 5°C)
INPUT VOLTAGE RANGE: –5 to +30VDC.
INPUT IMPEDANCE: 2Mtypical.
MAXIMUM VOLTAGE (either input terminal) WITH RESPECT TO OUTPUT LOW:
–5V, +30V.
READING ACCURACY
READING RESOLUTION: 1mV.
CONNECTOR: HI and LO input pair part of Output #1’s terminal block.
MEASUREMENT TIME CHOICES: 0.01 to 10 PLC
AVERAGE READINGS: 1 to 10.
READING TIME
1
: ±(0.05% + 3mV).
1, 8,9
: 31ms, typical.
7
, in 0.01PLC steps.
Page 4
2302-PJ Battery Simulator
GENERAL
ISOLATION (low–earth): 22VDC max. Do not exceed 60VDC between any two ter-
minals of either connector.
PROGRAMMING: IEEE-488.2 (SCPI). USER-DEFINABLE POWER-UP STATES: 4. REAR PANEL CONNECTORS: 8-position quick disconnect terminal block for out-
put (4), sense (2), and DVM (2).
TEMPERATURE COEFFICIENT (outside 23°C ±5°C): Derate accuracy specification
by (0.1 × specification)/°C.
OPERATING TEMPERATURE: 0° to 50°C (Derate to 70%). 0° to 35°C (Full power). STORAGE TEMPERATURE: –20° to 70°C. HUMIDITY: <80% @ 35°C non-condensing. DISPLAY TYPE: 2-line × 16 character VFD. REMOTE DISPLAY/KEYPAD OPTION: Disables standard front panel. DIMENSIONS: 89mm high × 213mm wide × 411mm deep (3 NET WEIGHT: 3.2kg (7.1 lbs). SHIPPING WEIGHT: 5.4kg (12 lbs). INPUT POWER: 100–120VAC/220–240VAC, 50 or 60Hz (auto detected at power-up). POWER CONSUMPTION: 150VA max. WARRANTY: Two years parts and labor on materials and workmanship. EMC: Conforms with European Union Directive directive 89/336/EEC. SAFETY: Conforms with European Union Directive 73/23/EEC. AC LINE LEAKAGE CURRENT: 450µA @ 110VAC, typ.; 600µA @ 220V, typ. RELAY CONTROL PORT: 4-channel, each capable of 100mA sink, 24V max. Total
port sink capacity (all 4 combined) is 250mA max. Accepts DB-9 male plug.
ACCESSORIES SUPPLIED: User and service manual, output connectors mating
terminal (part no. CS-846).
ACCESSORIES AVAILABLE:
Model 2304-DISP: Remote LCD Display/Keypad (4.6 in × 2.7 in × 1.5 in). Includes
2.7m (9 ft) cable and rack mount kit.
1
⁄2in × 83⁄8 in × 163⁄16in).
1
2
3
4
5
6
7
8
9
10
11
12
13
Specifications are subject to change without notice.
PLC = 1.00. Following 15 minute warm-up, the change in output over 8 hours under ambient temper-
ature, constant load, and line operating conditions. Remote sense, at output terminals, 0.5A to 5A typical. Remote sense, with 4.5m (15 ft) of 16 gauge (1.31mm2) wire and 1resistance in each lead
to simulate typical test environment, 1.5A load change (0.15A to 1.65A). Minimum current in constant current mode is 6mA. 60Hz (50Hz). PLC = Power Line Cycle. 1PLC = 16.7ms for 60Hz operation, 20ms for 50Hz operation. Display off. Speed includes measurement and binary data transfer out of GPIB. Typical values, peak-to-peak noise equals 6 times rms noise. Based on settled signal: 100µs pulse trigger delay. Also applies to other apertures that are integer multiples of 1PLC. Recovery to within 20mV of previous level.
Page 5
2306 Dual Channel Battery/Charger Simulator
OUTPUT #1 (BATTERY)
DC VOLTAGE OUTPUT (2 YEARS,23°C ± 5°C)
OUTPUT VOLTAGE: 0 to +15VDC.
OUTPUT ACCURACY: ±(0.05% + 3mV).
PROGRAMMING RESOLUTION: 1mV.
READBACK ACCURACY
1
READBACK RESOLUTION: 1mV.
OUTPUT VOLTAGE SETTLING TIME: 5ms to within stated accuracy.
LOAD REGULATION: 0.01% + 2mV.
LINE REGULATION: 0.5mV.
STABILITY
MEASUREMENT TIME CHOICES: 0.01 to 10 PLC
2
: 0.01% + 0.5mV.
7
, in 0.01PLC steps.
AVERAGE READINGS: 1 to 10.
READING TIME
1, 8,9
: 31ms, typical.
TRANSIENT RESPONSE: High Bandwidth Low Bandwidth
Transient Recovery Time
13
<40µs3or <60µs
Transient Voltage Drop <75mV3or <100mV4<250mV3or <400mV
4
<80µs3or <100µs
4
4
REMOTE SENSE 1V max. drop in each lead. Add 2mV to the voltage load regulation
specification for each 1V change in the negative output lead due to load current change. Remote sense required. Integrity of connection continually monitored. If compromised, output will turn off automatically once settable window (±0 to ±8 volts) around normal voltage exceeded.
VARIABLE OUTPUT IMPEDANCE
RANGE: 0 to 1.00in 0.01steps. Value can be changed with output on.
DC CURRENT (2 Years, 23°C ± 5°C)
CONTINUOUS AVERAGE OUTPUT CURRENT:
Channel #2 (Charger) OFF:
I = 50W/(Vset channel 1 + 6V ); 5A max.
Channel #2 (Charger) ON:
I = (50W – Power consumed by channel #2)/(Vset channel 1 + 6V); 5A max.
The power consumed by channel #2 is calculated as:
Channel #2 sourcing current:
Power consumed = (Vset channel 2 + 6V) × (current supplied)
Channel #2 sinking current:
Power consumed = 5 × (sink current)
Peak currents can be a maximum of 5A provided the average current is within the above limits.
CONTINUOUS AVERAGE SINK CURRENT:
Channel #2 (Charger) OFF:
0–5V: 3A max. 5–15V: Derate 0.2A per volt above 5V. Compliance setting controls sinking.
Channel #2 (Charger) ON:
Available current = (50W – Power consumed by channel #2)/5; 3A max. (0–5V). Derate 0.2A per volt above 5V.
SOURCE COMPLIANCE ACCURACY: ±(0.16% + 5mA)
PROGRAMMED SOURCE COMPLIANCE RESOLUTION: 1.25mA.
READBACK ACCURACY
1
: 5A Range: ±(0.2% + 200µA).
5mA Range: ±(0.2% + 1µA).
READBACK RESOLUTION: 5A Range: 100µA.
5mA Range: 0.1µA.
LOAD REGULATION: 0.01% + 1mA.
LINE REGULATION: 0.5mA.
STABILITY
4
: 0.01% + 50µA.
MEASUREMENT TIME CHOICES: 0.01 to 10 PLC
AVERAGE READINGS: 1 to 10.
READING TIME
1, 8,9
: 31ms, typical.
5
.
7
, in 0.01PLC steps.
PULSE CURRENT MEASUREMENT OPERATION
TRIGGER LEVEL: 5A Range: 5mA to 5A, in 5mA steps.
TRIGGER DELAY: 0 to 100ms, in 10µs steps.
INTERNAL TRIGGER DELAY: 15µs.
HIGH/LOW/AVERAGE MODE:
Measurement Aperture Settings: 33.3µs to 833ms, in 33.3µs steps. Average Readings: 1 to 100.
PULSE CURRENT MEASUREMENT ACCURACY
100 µs – 200 µs 0.2% + 900 µA + 1.5mA 200 µs – 500 µs 0.2% + 900 µA + 1 mA
500 µs – <1 PLC 0.2% + 600 µA + 0.8mA
1A Range: 1mA to 1A, in 1mA steps. 100mA Range: 0.1mA to 100mA, in 100µA steps.
11
(2 Years,23°C ±5°C):
APERTURE ACCURACY ±(% reading + offset + rms noise10)
<100 µs 0.2% + 900 µA + 2 mA
12
1 PLC
0.2% + 400 µA + 0 mA
>1 PLC 0.2% + 400 µA + 100 µA
BURST MODE CURRENT MEASUREMENT
MEASUREMENT APERTURE: 33.3µs.
CONVERSION RATE: 3650/second, typical.
INTERNAL TRIGGER DELAY: 15µs.
NUMBER OF SAMPLES: 1 to 5000.
TRANSFER SAMPLES ACROSS IEEE BUS IN BINARY MODE: 4800 bytes/s, typical.
LONG INTEGRATION MODE CURRENT MEASUREMENT
MEASUREMENT TIME6: 850ms (840ms) to 60 seconds in 1ms steps.
DIGITAL VOLTMETER INPUT (2 Years,23°C ± 5°C)
INPUT VOLTAGE RANGE: –5 to +30VDC.
INPUT IMPEDANCE: 2Mtypical.
MAXIMUM VOLTAGE (either input terminal) WITH RESPECT TO OUTPUT LOW:
–5V, +30V.
READING ACCURACY
READING RESOLUTION: 1mV.
CONNECTOR: HI and LO input pair part of Output #1’s terminal block.
MEASUREMENT TIME CHOICES: 0.01 to 10 PLC
AVERAGE READINGS: 1 to 10.
READING TIME
1
1, 8,9
: 31ms, typical.
7
, in 0.01PLC steps.
Page 6
2306 Dual Channel Battery/Charger Simulator
OUTPUT #2 (CHARGER)
DC VOLTAGE OUTPUT (2 YEARS,23°C ± 5°C)
OUTPUT VOLTAGE: 0 to +15VDC.
OUTPUT ACCURACY: ±(0.05% + 10mV).
PROGRAMMING RESOLUTION: 10mV.
READBACK ACCURACY
1
READBACK RESOLUTION: 1mV.
OUTPUT VOLTAGE SETTLING TIME: 5ms to within stated accuracy.
LOAD REGULATION: 0.01% + 2mV.
LINE REGULATION: 0.5mV.
STABILITY
MEASUREMENT TIME CHOICES: 0.01 to 10 PLC
2
: 0.01% + 0.5mV.
7
, in 0.01PLC steps.
AVERAGE READINGS: 1 to 10.
READING TIME
1, 8,9
: 31ms, typical.
TRANSIENT RESPONSE: High Bandwidth Low Bandwidth
Transient Recovery Time
13
<50µs3or <80µs
Transient Voltage Drop <120mV3or <150mV4<160mV3or <200mV
4
<60µs3or <100µs
4
4
REMOTE SENSE: 1V max. drop in each lead. Add 2mV to the voltage load regulation
specification for each 1V change in the negative output lead due to load current change. Remote sense required. Integrity of connection continually monitored. If compromised, output will turn off automatically once settable window (±0 to ±8 volts) around normal voltage exceeded.
DC CURRENT (2 Years, 23°C ± 5°C)
CONTINUOUS AVERAGE OUTPUT CURRENT:
Channel #1 (Battery) OFF:
I = 50W/(Vset channel 2 + 6V ); 5A max.
Channel #1 (Battery) ON:
I = (50W – Power consumed by channel #1)/(Vset channel 2 + 6V); 5A max.
The power consumed by channel #1 is calculated as:
Channel #1 sourcing current:
Power consumed = (Vset channel 1 + 6V) × (current supplied)
Channel #1 sinking current:
Power consumed = 5 × (sink current)
Peak currents can be a maximum of 5A provided the average current is within the above limits.
CONTINUOUS AVERAGE SINK CURRENT:
Channel #1 (Battery) OFF:
0–5V: 3A max. 5–15V: Derate 0.2A per volt above 5V. Compliance setting controls sinking.
Channel #1 (Battery) ON:
Available current = (50W – Power consumed by channel #1)/5; 3A max. (0–5V). Derate 0.2A per volt above 5V.
SOURCE COMPLIANCE ACCURACY: ±(0.16% + 5mA)
PROGRAMMED SOURCE COMPLIANCE RESOLUTION: 1.25mA.
READBACK ACCURACY
1
: 5A Range: ±(0.2% + 200µA).
5mA Range: ±(0.2% + 1µA).
READBACK RESOLUTION: 5A Range: 100µA.
5mA Range: 0.1µA.
LOAD REGULATION: 0.01% + 1mA.
LINE REGULATION: 0.5mA.
STABILITY
4
: 0.01% + 50µA.
MEASUREMENT TIME CHOICES: 0.01 to 10 PLC
AVERAGE READINGS: 1 to 10.
READING TIME
1, 8,9
: 31ms, typical.
5
.
7
, in 0.01PLC steps.
PULSE CURRENT MEASUREMENT OPERATION
TRIGGER LEVEL: 5mA to 5A, in 5mA steps.
TRIGGER DELAY: 0 to 100ms, in 10µs steps.
INTERNAL TRIGGER DELAY: 15µs.
HIGH/LOW/AVERAGE MODE:
Measurement Aperture Settings: 33.3µs to 833ms, in 33.3µs steps. Average Readings: 1 to 100.
PULSE CURRENT MEASUREMENT ACCURACY
11
(2 Years,23°C ±5°C):
APERTURE ACCURACY ±(% reading + offset + rms noise10)
<100 µs 0.2% + 900 µA + 2 mA 100 µs – 200 µs 0.2% + 900 µA + 1.5mA 200 µs – 500 µs 0.2% + 900 µA + 1 mA
500 µs – <1 PLC 0.2% + 600 µA + 0.8mA
12
1 PLC
0.2% + 400 µA + 0 mA
>1 PLC 0.2% + 400 µA + 100 µA
BURST MODE CURRENT MEASUREMENT
MEASUREMENT APERTURE: 33.3µs.
CONVERSION RATE: 2040/second, typical.
INTERNAL TRIGGER DELAY: 15µs.
NUMBER OF SAMPLES: 1 to 5000.
TRANSFER SAMPLES ACROSS IEEE BUS IN BINARY MODE: 4800 bytes/s, typical.
LONG INTEGRATION MODE CURRENT MEASUREMENT
MEASUREMENT TIME6: 850ms (840ms) to 60 seconds in 1ms steps.
DIGITAL VOLTMETER INPUT (2 Years,23°C ± 5°C)
INPUT VOLTAGE RANGE: –5 to +30VDC.
INPUT IMPEDANCE: 2Mtypical.
MAXIMUM VOLTAGE (either input terminal) WITH RESPECT TO OUTPUT LOW:
–5V, +30V.
READING ACCURACY
READING RESOLUTION: 1mV.
CONNECTOR: HI and LO input pair part of Output #2’s terminal block.
MEASUREMENT TIME CHOICES: 0.01 to 10 PLC
AVERAGE READINGS: 1 to 10.
READING TIME
1
1, 8,9
: 31ms, typical.
7
, in 0.01PLC steps.
Page 7
2306 Dual Channel Battery/Charger Simulator
1
GENERAL
ISOLATION (low–earth): 22VDC max. Do not exceed 60VDC between any two terminals of
either connector.
PROGRAMMING: IEEE-488.2 (SCPI). USER-DEFINABLE POWER-UP STATES: 5. REAR PANEL CONNECTORS: Two 8-position quick disconnect terminal block for output (4),
sense (2), and DVM (2).
TEMPERATURE COEFFICIENT(outside 23°C ±5°C): Derate accuracy specification by (0.1 × spec-
ification)/°C.
OPERATING TEMPERATURE: 0° to 50°C (Derate to 70%). 0° to 35°C (Full power). STORAGE TEMPERATURE: –20° to 70°C. HUMIDITY: <80% @ 35°C non-condensing. DISPLAY TYPE: 2-line × 16 character VFD. REMOTE DISPLAY/KEYPAD OPTION: Disables standard front panel. DIMENSIONS: 89mm high × 213mm wide × 411mm deep (3 NET WEIGHT: 3.2kg (7.1 lbs). SHIPPING WEIGHT: 5.4kg (12 lbs). INPUT POWER: 100–120VAC/220–240VAC, 50 or 60Hz (auto detected at power-up). POWER CONSUMPTION: 150VA max. WARRANTY: Two years parts and labor on materials and workmanship. EMC: Conforms with European Union Directive directive 89/336/EEC, EN 55011, EN 50082-1,
EN 61000-3-2 and 61000-3-3, FCC part 15 class B.
SAFETY: Conforms with European Union Directive 73/23/EEC, EN 61010-1. AC LINE LEAKAGE CURRENT: 450µA @ 110VAC, typ.; 600µA @ 220V, typ. RELAY CONTROL PORT: 4-channel, each capable of 100mA sink, 24V max. Total port sink
capacity (all 4 combined) is 250mA max. Accepts DB-9 male plug.
ACCESSORIES SUPPLIED: User and service manual, output connectors mating terminal (part
no. CS-846).
ACCESSORIES AVAILABLE:
Model 2304-DISP: Remote LCD Display/Keypad (4.6 in × 2.7 in × 1.5 in). Includes 2.7m (9 ft)
cable and rack mount kit.
1
⁄2in × 83⁄8 in × 163⁄16in).
PLC = 1.00.
2
Following 15 minute warm-up, the change in output over 8 hours under ambient temperature, constant load, and line operating conditions.
3
Remote sense, at output terminals, 0.5A to 5A typical.
4
Remote sense, with 4.5m (15 ft) of 16 gauge (1.31mm2) wire and 1Ω resist- ance in each lead to simulate typical test environment, 1.5A load change (0.15A to 1.65A).
5
Minimum current in constant current mode is 6mA.
6
60Hz (50Hz).
7
PLC = Power Line Cycle. 1PLC = 16.7ms for 60Hz operation, 20ms for 50Hz operation.
8
Display off.
9
Speed includes measurement and binary data transfer out of GPIB.
10
Typical values, peak-to-peak noise equals 6 times rms noise.
11
Based on settled signal: 100µs pulse trigger delay.
12
Also applies to other apertures that are integer multiples of 1PLC.
13
Recovery to within 20mV of previous level.
Specifications are subject to change without notice.
Page 8
2306-PJ Dual Channel Battery/Charger Simulator
OUTPUT #1 (BATTERY)
DC VOLTAGE OUTPUT (2 YEARS,23°C ± 5°C)
OUTPUT VOLTAGE: 0 to +15VDC.
OUTPUT ACCURACY: ±(0.05% + 3mV).
PROGRAMMING RESOLUTION: 1mV.
READBACK ACCURACY
1
: ±(0.05% + 3mV).
READBACK RESOLUTION: 1mV.
OUTPUT VOLTAGE SETTLING TIME: 5ms to within stated accuracy.
LOAD REGULATION: 0.01% + 2mV.
LINE REGULATION: 0.5mV.
STABILITY
MEASUREMENT TIME CHOICES: 0.01 to 10 PLC
2
: 0.01% + 0.5mV.
7
, in 0.01PLC steps.
AVERAGE READINGS: 1 to 10.
READING TIME
1, 8,9
: 31ms, typical.
TRANSIENT RESPONSE: High Bandwidth Low Bandwidth
Transient Recovery Time
13
<40µs3or <60µs
Transient Voltage Drop <75mV3or <100mV4<250mV3or <400mV
4
<80µs3or <100µs
4
4
REMOTE SENSE 1V max. drop in each lead. Add 2mV to the voltage load regulation
specification for each 1V change in the negative output lead due to load current change. Remote sense required. Integrity of connection continually monitored. If compromised, output will turn off automatically once settable window (±0 to ±8 volts) around normal voltage exceeded.
VARIABLE OUTPUT IMPEDANCE
RANGE: 0 to 1.00in 0.01steps. Value can be changed with output on.
DC CURRENT (2 Years, 23°C ± 5°C)
CONTINUOUS AVERAGE OUTPUT CURRENT:
Channel #2 (Charger) OFF:
I = 50W/(Vset channel 1 + 6V ); 5A max.
Channel #2 (Charger) ON:
I = (50W – Power consumed by channel #2)/(Vset channel 1 + 6V); 5A max.
The power consumed by channel #2 is calculated as:
Channel #2 sourcing current:
Power consumed = (Vset channel 2 + 6V) × (current supplied)
Channel #2 sinking current:
Power consumed = 5 × (sink current)
Peak currents can be a maximum of 5A provided the average current is within the above limits.
CONTINUOUS AVERAGE SINK CURRENT:
Channel #2 (Charger) OFF:
0–5V: 3A max. 5–15V: Derate 0.2A per volt above 5V. Compliance setting controls sinking.
Channel #2 (Charger) ON:
Available current = (50W – Power consumed by channel #2)/5; 3A max. (0–5V). Derate 0.2A per volt above 5V.
SOURCE COMPLIANCE ACCURACY: ±(0.16% + 5mA)
PROGRAMMED SOURCE COMPLIANCE RESOLUTION: 1.25mA.
READBACK ACCURACY
1
: 5A Range: ±(0.2% + 200µA).
500mA Range: ±(0.2% + 20µA).
READBACK RESOLUTION: 5A Range: 100µA.
500mA Range: 10µA.
LOAD REGULATION: 0.01% + 1mA.
LINE REGULATION: 0.5mA.
STABILITY
4
: 0.01% + 50µA.
MEASUREMENT TIME CHOICES: 0.01 to 10 PLC
AVERAGE READINGS: 1 to 10.
READING TIME
1, 8,9
: 31ms, typical.
5
.
7
, in 0.01PLC steps.
PULSE CURRENT MEASUREMENT OPERATION
TRIGGER LEVEL:
5A Current Range
5A Range: 5mA to 5A, in 5mA steps. 1A Range: 1mA to 1A, in 1mA steps. 100mA Range: 0.1mA to 100mA, in 100µA steps.
500mA Current Range
500mA Range: 0.5mA to 500mA, in 0.5mA steps. 100mA Range: 0.1mA to 100mA, in 100µA steps. 10mA Range: 100µA to 10mA, in 100µA steps.
TRIGGER DELAY: 0 to 100ms, in 10µs steps.
INTERNAL TRIGGER DELAY: 15µs.
HIGH/LOW/AVERAGE MODE:
Measurement Aperture Settings: 33.3µs to 833ms, in 33.3µs steps. Average Readings: 1 to 100.
PULSE CURRENT MEASUREMENT ACCURACY
11
(2 Years,23°C ±5°C):
ACCURACY ±(% reading + offset + rms noise10)
APERTURE 5A RANGE 500mA RANGE
<100 µs 0.2% + 900 µA + 2 mA 0.2% + 90 µA + 2 mA 100 µs – 200 µs 0.2% + 900 µA + 1.5 mA 0.2% + 90 µA + 1.5 mA 200 µs – 500 µs 0.2% + 900 µA + 1 mA 0.2% + 90 µA + 1 mA
500 µs – <1 PLC 0.2% + 600 µA + 0.8 mA 0.2% + 60 µA + 0.8 mA
12
1 PLC
0.2% + 400 µA + 0 mA 0.2% + 40 µA + 0 mA
>1 PLC 0.2% + 400 µA + 100 µA 0.2% + 40 µA + 100 µA
BURST MODE CURRENT MEASUREMENT
MEASUREMENT APERTURE: 33.3µs.
CONVERSION RATE: 3650/second, typical.
INTERNAL TRIGGER DELAY: 15µs.
NUMBER OF SAMPLES: 1 to 5000.
TRANSFER SAMPLES ACROSS IEEE BUS IN BINARY MODE: 4800 bytes/s, typical.
LONG INTEGRATION MODE CURRENT MEASUREMENT
Available on both 5A and 500mA current ranges.
MEASUREMENT TIME
6
: 850ms (840ms) to 60 seconds in 1ms steps.
DIGITAL VOLTMETER INPUT (2 Years,23°C ± 5°C)
INPUT VOLTAGE RANGE: –5 to +30VDC.
INPUT IMPEDANCE: 2Mtypical.
MAXIMUM VOLTAGE (either input terminal) WITH RESPECT TO OUTPUT LOW:
–5V, +30V.
READING ACCURACY
READING RESOLUTION: 1mV.
CONNECTOR: HI and LO input pair part of Output #1’s terminal block.
MEASUREMENT TIME CHOICES: 0.01 to 10 PLC
AVERAGE READINGS: 1 to 10.
READING TIME
1
: ±(0.05% + 3mV).
1, 8,9
: 31ms, typical.
7
, in 0.01PLC steps.
Page 9
2306-PJ Dual Channel Battery/Charger Simulator
OUTPUT #2 (CHARGER)
DC VOLTAGE OUTPUT (2 YEARS,23°C ± 5°C)
OUTPUT VOLTAGE: 0 to +15VDC.
OUTPUT ACCURACY: ±(0.05% + 10mV).
PROGRAMMING RESOLUTION: 10mV.
READBACK ACCURACY
1
: ±(0.05% + 3mV).
READBACK RESOLUTION: 1mV.
OUTPUT VOLTAGE SETTLING TIME: 5ms to within stated accuracy.
LOAD REGULATION: 0.01% + 2mV.
LINE REGULATION: 0.5mV.
STABILITY
MEASUREMENT TIME CHOICES: 0.01 to 10 PLC
2
: 0.01% + 0.5mV.
7
, in 0.01PLC steps.
AVERAGE READINGS: 1 to 10.
READING TIME
1, 8,9
: 31ms, typical.
TRANSIENT RESPONSE: High Bandwidth Low Bandwidth
Transient Recovery Time
13
<50µs3or <80µs
Transient Voltage Drop <120mV3or <150mV4<160mV3or <200mV
4
<60µs3or <100µs
4
4
REMOTE SENSE: 1V max. drop in each lead. Add 2mV to the voltage load regulation
specification for each 1V change in the negative output lead due to load current change. Remote sense required. Integrity of connection continually monitored. If compromised, output will turn off automatically once settable window (±0 to ±8 volts) around normal voltage exceeded.
DC CURRENT (2 Years, 23°C ± 5°C)
CONTINUOUS AVERAGE OUTPUT CURRENT:
Channel #1 (Battery) OFF:
I = 50W/(Vset channel 2 + 6V ); 5A max.
Channel #1 (Battery) ON:
I = (50W – Power consumed by channel #1)/(Vset channel 2 + 6V); 5A max.
The power consumed by channel #1 is calculated as:
Channel #1 sourcing current:
Power consumed = (Vset channel 1 + 6V) × (current supplied)
Channel #1 sinking current:
Power consumed = 5 × (sink current)
Peak currents can be a maximum of 5A provided the average current is within the above limits.
CONTINUOUS AVERAGE SINK CURRENT:
Channel #1 (Battery) OFF:
0–5V: 3A max. 5–15V: Derate 0.2A per volt above 5V. Compliance setting controls sinking.
Channel #1 (Battery) ON:
Available current = (50W – Power consumed by channel #1)/5; 3A max. (0–5V). Derate 0.2A per volt above 5V.
SOURCE COMPLIANCE ACCURACY: ±(0.16% + 5mA)
PROGRAMMED SOURCE COMPLIANCE RESOLUTION: 1.25mA.
READBACK ACCURACY
1
: 5A Range: ±(0.2% + 200µA).
5mA Range: ±(0.2% + 1µA).
READBACK RESOLUTION: 5A Range: 100µA.
5mA Range: 0.1µA.
LOAD REGULATION: 0.01% + 1mA.
LINE REGULATION: 0.5mA.
STABILITY
4
: 0.01% + 50µA.
MEASUREMENT TIME CHOICES: 0.01 to 10 PLC
AVERAGE READINGS: 1 to 10.
READING TIME
1, 8,9
: 31ms, typical.
5
.
7
, in 0.01PLC steps.
PULSE CURRENT MEASUREMENT OPERATION
TRIGGER LEVEL: 5mA to 5A, in 5mA steps.
TRIGGER DELAY: 0 to 100ms, in 10µs steps.
INTERNAL TRIGGER DELAY: 15µs.
HIGH/LOW/AVERAGE MODE:
Measurement Aperture Settings: 33.3µs to 833ms, in 33.3µs steps. Average Readings: 1 to 100.
PULSE CURRENT MEASUREMENT ACCURACY
11
(2 Years,23°C ±5°C):
APERTURE ACCURACY ±(% reading + offset + rms noise10)
<100 µs 0.2% + 900 µA + 2 mA 100 µs – 200 µs 0.2% + 900 µA + 1.5mA 200 µs – 500 µs 0.2% + 900 µA + 1 mA
500 µs – <1 PLC 0.2% + 600 µA + 0.8mA
12
1 PLC
0.2% + 400 µA + 0 mA
>1 PLC 0.2% + 400 µA + 100 µA
BURST MODE CURRENT MEASUREMENT
MEASUREMENT APERTURE: 33.3µs.
CONVERSION RATE: 2040/second, typical.
INTERNAL TRIGGER DELAY: 15µs.
NUMBER OF SAMPLES: 1 to 5000.
TRANSFER SAMPLES ACROSS IEEE BUS IN BINARY MODE: 4800 bytes/s, typical.
LONG INTEGRATION MODE CURRENT MEASUREMENT
MEASUREMENT TIME6: 850ms (840ms) to 60 seconds in 1ms steps.
DIGITAL VOLTMETER INPUT (2 Years,23°C ± 5°C)
INPUT VOLTAGE RANGE: –5 to +30VDC.
INPUT IMPEDANCE: 2Mtypical.
MAXIMUM VOLTAGE (either input terminal) WITH RESPECT TO OUTPUT LOW:
–5V, +30V.
READING ACCURACY
READING RESOLUTION: 1mV.
CONNECTOR: HI and LO input pair part of Output #2’s terminal block.
MEASUREMENT TIME CHOICES: 0.01 to 10 PLC
AVERAGE READINGS: 1 to 10.
READING TIME
1
: ±(0.05% + 3mV).
1, 8,9
: 31ms, typical.
7
, in 0.01PLC steps.
Page 10
2306-PJ Dual Channel Battery/Charger Simulator
1
GENERAL
ISOLATION (low–earth): 22VDC max. Do not exceed 60VDC between any two terminals of
either connector.
PROGRAMMING: IEEE-488.2 (SCPI). USER-DEFINABLE POWER-UP STATES: 4. REAR PANEL CONNECTORS: Two 8-position quick disconnect terminal block for output (4),
sense (2), and DVM (2).
TEMPERATURE COEFFICIENT (outside 23°C ±5°C): Derate accuracy specification by (0.1 ×
specification)/°C.
OPERATING TEMPERATURE: 0° to 50°C (Derate to 70%). 0° to 35°C (Full power). STORAGE TEMPERATURE: –20° to 70°C. HUMIDITY: <80% @ 35°C non-condensing. DISPLAY TYPE: 2-line × 16 character VFD. REMOTE DISPLAY/KEYPAD OPTION: Disables standard front panel. DIMENSIONS: 89mm high × 213mm wide × 411mm deep (3 NET WEIGHT: 3.2kg (7.1 lbs). SHIPPING WEIGHT: 5.4kg (12 lbs). INPUT POWER: 100–120VAC/220–240VAC, 50 or 60Hz (auto detected at power-up). POWER CONSUMPTION: 150VA max. WARRANTY: Two years parts and labor on materials and workmanship. EMC: Conforms with European Union Directive directive 89/336/EEC. SAFETY: Conforms with European Union Directive 73/23/EEC. AC LINE LEAKAGE CURRENT: 450µA @ 110VAC, typ.; 600µA @ 220V, typ. RELAY CONTROL PORT: 4-channel, each capable of 100mA sink, 24V max. Total port sink
capacity (all 4 combined) is 250mA max. Accepts DB-9 male plug.
ACCESSORIES SUPPLIED: User and service manual, output connectors mating terminal (part
no. CS-846).
ACCESSORIES AVAILABLE:
Model 2304-DISP: Remote LCD Display/Keypad (4.6 in × 2.7 in × 1.5 in). Includes 2.7m (9 ft)
cable and rack mount kit.
1
⁄2in × 83⁄8 in × 163⁄16in).
PLC = 1.00.
2
Following 15 minute warm-up, the change in output over 8 hours under ambient temperature, constant load, and line operating conditions.
3
Remote sense, at output terminals, 0.5A to 5A typical.
4
Remote sense, with 4.5m (15 ft) of 16 gauge (1.31mm2) wire and 1resistance in each lead to simulate typical test environment, 1.5A load change (0.15A to
1.65A).
5
Minimum current in constant current mode is 6mA.
6
60Hz (50Hz).
7
PLC = Power Line Cycle. 1PLC = 16.7ms for 60Hz operation, 20ms for 50Hz operation.
8
Display off.
9
Speed includes measurement and binary data transfer out of GPIB.
10
Typical values, peak-to-peak noise equals 6 times rms noise.
11
Based on settled signal: 100µs pulse trigger delay.
12
Also applies to other apertures that are integer multiples of 1PLC.
13
Recovery to within 20mV of previous level.
Specifications are subject to change without notice.