2001 7
2002 8
• True 71⁄2- (Model 2001) or 81⁄2-
digit (Model 2002) resolution
• Exceptional measurement
integrity with high speed
• Broad range of built-in
measurement functions
• Built-in 10 channel scanner
option
• IEEE-488.2 and SCPI compatible
• Model 2002 has HP 3458A
emulation mode
1
⁄2-Digit High Performance Multimeter
1
⁄2-Digit High Performance Multimeter
DMM users whose applications demand exceptional resolution, accuracy,
and sensitivity combined with high throughput now have two attractive
alternatives to high priced, high-end DMMs. Keithley’s 7
2001 and 8
1
⁄2-digit Model 2002 High Performance Digital Multimeters not
1
⁄2-digit Model
only deliver performance specifications usually associated only with instruments that cost thousands more, but they also offer a broad range of functions not typically available from DMMs.
True 71⁄2- (or 81⁄2-) Digit Resolution
While other DMMs may claim 71⁄2- or 81⁄2-digit resolution, those instruments must average multiple readings to extend their resolution. The resolution specifications of the 2001 and 2002 are based on a 28-bit A/D converter that provides the resolution needed to discern smaller changes. This
higher resolution also provides greater dynamic range, making it possible
to measure from 1µV to 20V on a single range, thus avoiding range-shift
errors and delays.
High Throughput, High Accuracy DCV and
Resistance Measurements
In applications where high throughput is critical, both the Model 2001 and
2002 provide more than 2000 readings per second at 4
1
At 7
⁄2 digits, the Model 2002 maintains full rated accuracy at reading rates
1
⁄2-digit resolution.
up to 44/second on DCV and ohms.
The Model 2002 uses a unique single-phase method for 4-wire ohms measurements. This makes it twice as fast for a given power line cycle rate. This
also eliminates errors due to changing lead resistances that can result from
fast test handlers. A built-in “open-lead” detection circuit also eliminates
many production test problems.
High Accuracy ACV Measurements
A patented circuit design makes the 2001 and
2002’s AC measurements several times more
accurate than competitive DMMs, thus maintaining very good accuracy (better than 0.1%) down
to 1Hz. The wide bandwidth of these DMMs
allows for accurate measurements of high frequency AC signals without the need for a special
AC meter. Both the 2001 and 2002 feature TRMS
AC, Average AC, Peak AC, AC+DC, and Crest
Factor measurement capability for a wide variety
of applications.
Built-In Scanner (Multiplexer)
Options
With the addition of a plug-in scanner card, it
ACCESSORIES AVAILABLE
TEST LEADS AND PROBES
5805 Kelvin Probes, 0.9m (3 ft)
8606 High Performance Modular Probe Kit
CABLES/ADAPTERS
7007-1 Shielded GPIB Cable, 1m (3.3 ft)
7007-2
7009-5 RS-232 Cable
RACK MOUNT KITS
4288-1 Single Fixed Rack Mount Kit
4288-4
THER
O
KPC-488.2 IEEE-488.2 Interface Card for the ISA Bus
KPCI-488 IEEE-488 Interface/Controller for the PCI Bus
KUSB-488 IEEE-488.2 USB-to-GPIB Interface Adapter
Shielded GPIB Cable, 2m (6.6 ft)
Side by Side Rack Mount Kit
becomes a complete scan and measure system
for applications involving up to ten measurement points. The additional resolution and
measurement ranges provided by the 2002 make
it an excellent choice for production test, design
verification, and metrology applications where
high accuracy is critical.
High resolution, high accuracy DMMs
1.888.KEITHLEY
www.keithley.com
(U.S. only)
A GR
TER M
EA
EASU
RE OF CONFIDENCE
DIGITAL MULTIMETERS & SYSTEMS
27

2001 7
2002 8
1
⁄2-Digit High Performance Multimeter
1
⁄2-Digit High Performance Multimeter
Ordering Information
2001 High Performance 71⁄2-Digit
2002 High Performance 8
2
2001-SCAN
2001-TCSCAN
These products are available
with an Extended Warranty.
Accessories Supplied
Model 8605 High Performance
Modular Test Leads, user’s manual, option slot cover, and full
calibration data.
Model 2001 and 2002 Specifications
MM with 8K Memory
D
DMM with 8K Memory
000-SCAN
10-Channel Scanner Card
10-Channel Scanner Card
with two high-speed
channels
9-Channel Thermocouple
Scanner Card
1
⁄2-Digit
2001 Condensed Specifications
C VOLTS
D
ESO- INPUT + ppm of range)
RANGE LUTION RESISTANCE 90 Days
200 mV 10 nV >10 GΩ 25 + 6
20 V 1 µV >10 GΩ 18 + 4
00 V 10 µV 10 M
2
000 V 100 µV 10 M
1
RESISTANCE
RANGE LUTION SOURCE 90 Days
20 Ω 1µΩ 9.2 mA 52 + 7
00
2
20 k
00 k
2
20MΩ 1 Ω 7
200M
R
2 V 100 nV >10 G
CCURACY
A
RESO- CURRENT + ppm of range)
Ω 10µΩ 0
2kΩ 100 µΩ 0.98 mA 33 + 4
Ω 1mΩ 89 µA 32 + 4
Ω 10mΩ 7
M
2
1GΩ 100 Ω 4.4 nA 20000 + 100
00m
Ω 1
Ω 10 Ω 4.4 nA 10000 + 100
.98 mA 36 + 7
70 nA 110 + 4.5
Ω 7
0 nA 560 + 4.5
DC AMPS
MAXIMUM ±(ppm of reading
RANGE LUTION VOLTAGE 90 Days
200 µA 10 pA 0.25 V 300 + 25
20 mA 1 nA 0.4 V 300 + 20
200 mA 10 nA 0.5 V 300 + 20
RESO- BURDEN + ppm of range)
2 mA 100 pA 0.31 V 300 + 20
2 A 100 nA 1.5 V 600 + 20
ACCURACY
(ppm of reading
±
Ω 18 + 2
1% 27 + 3
Ω ±
1% 31 + 6
Ω ±
(ppm of reading
±
µA 72 + 4.5
ACCURACY
2002 Condensed Specifications
C VOLTS
D
ESO- INPUT + ppm of range)
RANGE LUTION RESISTANCE 90 Days
200 mV 1 nV >100 GΩ 15 + 8
20 V 100 nV >100 GΩ 6 + 0.15
00 V 1 µV 10 M
2
000 V 10 µV 10 M
1
RESISTANCE
RANGE LUTION SOURCE 90 Days
20 Ω 100 nΩ 7.2 mA 15 + 6
00
2
20 k
00 k
2
20MΩ 1
200M
R
2 V 10 nV >100 G
CCURACY
A
RESO- CURRENT + ppm of range)
Ω 1µΩ 9
2kΩ 10 µΩ 960 µA 7 + 0.4
Ω 100 µΩ 96 µA 7 + 0.4
Ω 1mΩ 9
M
2
Ω 10mΩ 1
00m
Ω 1 Ω 1.4 µA 500 + 3
1GΩ 10 Ω 1.4 µA 2000 + 15
60 µA 15 + 4
.6 µA 29 + 0.8
.9 µA 53 + 0.5
.4 µA 175 + 0.6
Ω 1
DC AMPS
MAXIMUM ±(ppm of reading
RANGE LUTION VOLTAGE 90 Days
200 µA 10 pA 0.25 V 275 + 25
20 mA 1 nA 0.35 V 275 + 20
200 mA 10 nA 0.35 V 300 + 20
RESO- BURDEN + ppm of range)
2 mA 100 pA 0.3 V 275 + 20
2 A 100 nA 1.1 V 600 + 20
ACCURACY
(ppm of reading
±
Ω 6 + 0.8
1% 14 + 2
Ω ±
1% 14 + 0.4
Ω ±
(ppm of reading
±
ACCURACY
28
For more information, request the
Model 2001 and 2002 Technical
Specifications books.
1.888.KEITHLEY
DIGITAL MULTIMETERS & SYSTEMS
www.keithley.com
(U.S. only)
GENERAL
POWER
Voltage:
90–134V and 180–264V, universal self-selecting.
Frequency: 50Hz, 60Hz, or 400Hz self-identifying.
Consumption: <55 VA.
ENVIRONMENT
Operating Temperature:
Storage Temperature: –40°
Humidity: 80% R.H., 0° to 35°C.
Altitude: 4,500m (15,000 ft) operating; 12,000m (40,000
ft.) non-operating.
PHYSICAL
Case Dimensions:
1
deep (3
⁄2 in × 81⁄2 in × 141⁄2 in).
Unit Weight: 4.2kg (9.2 lbs).
STANDARDS
EMI/RFI:
Conforms to VDE 0871B (per Vfg 1046/1984),
IEC 801-2, FCC part 15 Class B, CISPR-22 (EN55022).
Safety: Conforms to IEC348, CAN/CSC22.2 No. 231,
-28800E1.
T
MIL-
0° to 50°C.
C.
°
to +70
90mm high × 214mm wide × 369mm
A GREATER MEASURE OF CONFIDENCE
FREQUENCY COUNTER
RANGE: 1Hz–15MHz.
ACCURACY: ±
DC I
RANGE: 100µA to 12A.
ACCUR
TR
(0.03% of reading).
N-CIRCUIT CURRENT
:
ACY
(5% + 2 counts) over 2 years.
±
ACE RESISTANCE:
1mΩ to 10Ω typical.
TEMPERATURE
, S, B thermocouple types
Built-in linearization for J, K, N, T
to ITS-90 and 100
For complete specifications, refer to the 2001 or 2002
echnical Data Book.
T
Ω platinum RTDs.
, E, R