Keithley Instruments 2002 Data Sheet

48
DIGITAL MULTIMETERS
2001 and 2002 High Performance Multimeters
• True 71⁄
2- (Model 2001) or 8
(Model 2002) resolution
• Exceptional measurement integrity with high speed
1
2-digit
DMM users whose applications demand exceptional resolution, accuracy, and sensitivity combined with high throughput now have two attractive alternatives to high-priced, high-end DMMs. Keithley’s 7½- digit Model 2001 and 8½-digit Model 2002 High Performance Digital Multimeters not only deliver per­formance specifications usually associated only with instruments that cost thousands more, but they also offer a broad range of functions not typically available from DMMs. The 2002 is based on the same superior measurement technology as the 2001. The front panels of both instruments have the same look,
• High-speed function and range changing
• Broad range of built-in measurement functions
• Multiple measurement display
feel, and response, so the learning curve is reduced dramatically.
True 71⁄2- (or 81⁄2-) Digit Resolution
While other DMMs may claim 7½-or 8½-digit resolution, those instruments must average multiple readings to extend their resolution. The resolution specifications of the 2001 and 2002 are based on a 28-bit A/D converter that provides the
• Flexible option slot design
• Nanovolt Preamp option (Model
1801)
• IEEE-488.2 and SCPI compatible
• Model 2002 has HP 3458A emulation mode
resolution needed to discern smaller changes. This higher resolution also provides greater dynamic range, making it possible to measure from 1µV to 20V on a single range and to avoid range-shift errors.
Applications
The 2001 is well suited to general-purpose measurements that require high resolution and accuracy. With the addition of a plug-in scanner card, it becomes a complete scan and measure system for applications involving up to ten measurement points. The additional resolution and measurement ranges
Model 2001
1p
1n
–12
10
10
DC Voltage
AC Voltage
AC Voltage
Bandwidth
DC Current
DC In-Circuit
Current
AC Current
Frequency
Frequency Sensitivity
Period
Temperature–RTD
Temperature–TC
Crest Factor
Peak Spikes–Repetitive
Peak Spikes–Single
Resistance (2-Wire)
Resistance (4-Wire)
1nV
10nV
100nV
10pA
100pA
67ns
100n
100n
Logarithmic scale
Both the 2001 and 2002 provide exceptional measurement range. In addition, the 2002 offers extended DCV and resistance mea­surement capabilities.
Model 2002
1m
1µ
–3
–6
–9
10
10
1Hz
100µA
1Hz
60mV
0.001°
0.001°
0.1°
20Hz
500ns
2µs 16s
1µΩ
1µΩ
1M
6
10
2MHz
1MHz
210k
15MHz
2.1M
1G
9
10
1G
1k
1
3
0
10
10
1100V
775V rms
2.1A
12A
2.1A
775V rms
1s
630°C
1820°C
16s
ORDERING INFORMATION
All models include Model 8605 High Performance Modular Test Leads, user’s manual, option slot cover, and full calibra­tion data.
provided by the 2002 make it an excellent choice for production test, design verifi­cation, and metrology applications
2001 High Performance 71⁄
DMM with 8K Memory
2001/MEM1 High Performance 7
DMM with 32K Memory
2001/MEM2 High Performance 7
DMM with 128K Memory
2002 High Performance 8
DMM with 8K Memory
2002/MEM1 High Performance 8
DMM with 32K Memory
2002/MEM2 High Performance 8
DMM with 128K Memory
These products are available with an Extended Warranty. See page 635 for com­plete ordering information.
2
-Digit
1
2
-Digit
1
⁄2-Digit
1
⁄2-Digit
1
2
-Digit
1
2
-Digit
where high accuracy is critical.
High Accuracy ACV Measurements
A patented circuit design makes the 2001 and 2002’s AC measurements several times more accurate than competitive DMMs. In this circuit, the signal bypasses the prime error-contributing section of conventional rms converters. This not only increases the accuracy at almost any voltage level, but also increases sensitivity down to a guar­anteed 1% of the selected range, compared to 5–10% for most other DMMs. The result is highly accurate measurements over a broad range of input signals.
Applications involving vibration, servo, guidance, shock, and control systems often require accurate low-frequency ACV measurements.
The 2001 and 2002 maintain very good accuracy (better than 0.1%) down to 1Hz. The wide bandwidth of these DMMs allows for accurate measurements of high­frequency AC signals without the need for a special AC meter. Both the 2001 and 2002 feature TRMS AC, Average AC, Peak AC, and AC+DC measurement capability for a wide variety of applications.
High Speed for High Throughput
I
n applications where high throughput is critical, both the 2001 and 2002 provide more than 2000 readings per second at 4½-digit resolution. At 7½ digits, the 2002 maintains full rated accuracy at reading rates up to 44/second on DCV and ohms.
QQZQ QZQQ
QZZQ
48
www.keithley.com
DIGITAL MULTIMETERS
2001 and 2002 High Performance Multimeters
49
High Speed,High Precision Resistance Measurement
The Model 2002 uses a unique single-phase method for 4-wire ohms measurements. This makes it twice as fast for a given power line cycle rate. This also eliminates errors due to changing lead resist­ances that can result from fast test handlers. A built-in “open-lead” detection circuit also eliminates many production test problems.
Fast, Flexible Triggering
Trigger latency—the delay between trigger and measurement—is often a barrier to higher through­put. Also, variability in latency can complicate predicting measurement timing. The 2001 and 2002 trigger in less than 2µs ±1µs, which is 100–1000 times faster than other DMMs.
The unique Trigger-Link feature included in the Model 2001 and 2002 and a growing number of other Keithley test and measurement products can be used to coordinate the operation of two or more instruments. Trigger-Link combines six independent software selectable trigger lines on a sin­gle connector for simple, direct control over all instruments in a system.
Spot Trends with the Bar-Graph Display
The ability to track reading trends around a target value easily can be just as important as the abso­lute readings. A unique bar-graph display function in the 2001 and 2002 indicates data as a percent­age of the selected range from ±0.01% to ±100%. Whether adjusting about zero or any other desired value, this display can replace a nulling differential voltmeter.
Capture Spikes Down to 1µs
Both the 2001 and 2002 have internal peak detectors that can catch 1µs spikes such as power sup­ply spikes and transients, AC line power surges, and short-duration dropouts on components. These peak detectors operate up to 1MHz for repetitive signals or down to 1µs for single spikes, so there’s no need for a separate scope. The DMMs will automatically display and store the highest value or
QQZQ
display the maximum and minimum values of spikes.
Built-in Features and Capabilities
The 2001 and 2002 offer many built-in measurements that are typically unavailable in instruments of this type, including in-circuit current, temperature with thermocouples or RTDs, and peak spikes.
QZQQ
QZZQ
Four separate outputs linked to limits simplify configuring the DMMs for use in binning operations. Built-in AC crest factor measurement helps ensure the accuracy of AC measurements. Other DMMs
typically perform AC measurements for signals without excessive crest factor—the ratio of peak val­ue to rms values. However, when crest factor rises, measurements may not meet specs. With a 2001 or 2002, there’s no need for an oscilloscope to determine if the crest factor is acceptable—the DMM measures it directly.
While some DMMs will calculate average AC from the rms value, these calculations apply only to sine wave inputs. The 2001 and 2002 allow measuring peak value, average and true rms directly to obtain a complete characterization of the signal. This capability makes these DMMs ideal for AC cir­cuit design or test applications and for verifying test voltages specified only in averages.
When measuring AC or digital signals, frequency is critical. The 2001 and 2002 accurately measure frequency up to 15MHz. Accurate triggering on the signal is critical to measure frequency reliably. The frequency counters in the 2001 and 2002 have a fully adjustable trigger level for good measure­ments of noisy signals.
Multiple Measurement Display
The 2001 and 2002 can simultaneously display DC and AC volts and the AC frequency from a single measurement connection . Several other multiple-measurement displays are available, including crest factor and bar graph. By measuring sequentially and displaying simultaneously, the 2001/2002 operate as if three different meters are working together.
Option Slot Extends DMM Performance
An option slot in the back of the 2001 and 2002 opens the door to a wide range of measurement capabilities. Choose a 10-channel general-purpose scanner card or a 9-channel thermocouple scanner card to make measurements on multiple test points or devices. This can eliminate the need for a separate scanner and sig­nificantly reduce programming and setup time. The Model 1801 Nanovolt Preamp extends the performance of either DMM by extending the low-level measurement ranges and amplifies the voltage sensitivity dramatically.
ACCESSORIES AVAILABLE
ACCESSORIES AND COMPANION PROD­UCTS
2000-SCAN
2001-SCAN
2001-TCSCAN
1801 Nanovolt Preamp
TEST LEADS AND PROBES
5805 Kelvin Probes, 0.9m (3 ft) 5805-12 Kelvin Probes, 3.6m (12 ft) 5806 Kelvin Clip Leads 8502 Micro-DIN to 6 BNCs Adapter Box
8530 Centronics Adapter 8605 High Performance 2-Wire Modular
8606 High Performance Modular
8610 Low Thermal Shorting Plug 8680 RTD Probe Adapter 8681 Low Cost RTD 8693 General Purpose RTD Probe 8695 Surface RTD Probe 8696 Air/Gas RTD Probe
CABLES/ADAPTERS
7007-1 Shielded GPIB Cable, 1m (3.3 ft) 7001-2 Shielded GPIB Cable, 2m (6.6 ft) 8501-1 Trigger Link Cable, 1m (3.3 ft) 8501-2 Trigger Link Cable, 2m (6.6 ft) 8502 Trigger Link Adapter Box 7009-5 RS-232 Cable
RACK MOUNT KITS
4288-1 Single Fixed Rack Mount Kit 4288-4 Side by Side Rack Mount Kit
OTHER
KPCI-488 IEEE-488 Interface/Controller
KPC-488.2AT IEEE-488 Interface Card for IBM
KPC-TM Trigger Master Interface TestPoint Test Development Software 1050 Padded Carrying Case 2001-EW 1 Year Extended Warranty 2002-EW 1 Year Extended Warranty
See page 235 for descriptions of all accessories.
For more information, request the Model 2001 and 2002 Technical Specifications books.
10-Channel Scanner Card
10-Channel Scanner Card with two high-speed channels
9-Channel Thermocouple Scanner Card
with 8501-1 Cable
Test Leads
Probe Kit
for the PCI Bus
PC/AT (full slot)
www.keithley.com
49
50
DIGITAL MULTIMETERS
2001 and 2002 High Performance Multimeters
2001 Condensed Specifications
DC VOLTS 10PLC, D
RANGE FULL SCALE RESOLUTION RESOLUTION RESISTANCE 5 Minutes 24 Hours 90 Days 1 Year 2 Years
200 mV ±210.00000 10 nV 100 nV >10 G 3 + 3 10 + 6 25 + 6 37 + 6 50 + 6
2 V ±2.1000000 100 nV 1 µV >10 G 2 + 1.5 7 + 2 18 + 2 25 + 2 32 + 2
20 V ±21.000000 1 µV 10 µV >10 G 2 + 1.5 7 + 4 18 + 4 24 + 4 32 + 4
200 V ±210.00000 10 µV 100 µV 10 M±1% 2 + 1.5 13 + 3 27 + 3 38 + 3 52 + 3
1000 V ±1100.0000 100 µV 1 mV 10 M±1% 10 + 1.5 17 + 6 31 + 6 41 + 6 55 + 6
10 ACCURACY
FILT
DEFAULT INPUT ± (ppm of reading + ppm of range)
AC VOLTS ACCURACY
RANGE 20–50Hz 50–100Hz 0.1–2kHz 2–10kHz 10–30kHz 30–50kHz 50–100kHz 100–200kHz 0.2–1MHz 1–2MHz
200 mV 0.25 + 0.015 0.07 + 0.015 0.03 + 0.015 0.03 + 0.015 0.035 + 0.015 0.05 + 0.015 0.17+ 0.015 0.5 + 0.025 2 + 0.1 5 + 0.2
2 V 0.25 + 0.015 0.07 + 0.015 0.03 + 0.015 0.03 + 0.015 0.035 + 0.015 0.05 + 0.015 0.17 + 0.015 0.5 + 0.025 2 + 0.1 5 + 0.2
20 V 0.25 + 0.015 0.07 + 0.015 0.04 + 0.015 0.06 + 0.015 0.08 + 0.015 0.1 + 0.015 0.17 + 0.015 0.5 + 0.025 4 + 0.2 7 + 0.2 200 V 0.25 + 0.015 0.07 + 0.015 0.04 + 0.015 0.06 + 0.015 0.08 + 0.015 0.1 + 0.015 0.17 + 0.015 0.5 + 0.025 4 + 0.2 750 V 0.25 + 0.015 0.1 + 0.015 0.08 + 0.015 0.09 + 0.015 0.12 + 0.015 0.15 + 0.015 0.5 + 0.015
(Normal Mode–TRMS), Low Freq.mode extends down to 1Hz; Peak AC, dB, AC+DC also available.
90 Days, ±2°C from last AC self-cal for 1% to 100% of range ±(% of reading + % of range)
RESISTANCE ACCURACY
RANGE FULL SCALE RESOLUTION SOURCE 24 Hours 90 Days 1 Year 2 Years
20 21.000000 1 µ 9.2 mA 29 + 7 52 + 7 72 + 7 110 + 7
200 210.00000 10 µ 0.98 mA 24 + 7 36 + 7 56 + 7 90 + 7
2k 2100.0000 100 µΩ 0.98 mA 22 + 4 33 + 4 50 + 4 80 + 4.5
20 k 21.000000 1 m 89 µA 19 + 4 32 + 4 50 + 4 80 + 4.5
200 k 210.00000 10 m 7 µA 20 + 4.5 72 + 4.5 90 + 4.5 130 + 5
2M 2.1000000 100 mΩ 770 nA 50 + 4.5 110 + 4.5 160 + 4.5 230 + 5
20 M 21.000000 1 70 nA 160 + 4.5 560 + 4.5 900 + 4.5 1100 + 5
200 M 210.00000 10 4.4 nA 3000 + 100 10000 + 100 20000 + 100 30000 + 100
1G 1.0500000 100 Ω 4.4 nA 9000 + 100 20000 + 100 40000 + 100 60000 + 100
DC AMPS
RANGE FULL SCALE RESOLUTION VOLTAGE 24 Hours 90 Days 1 Year 2 Years
200 µA 210.00000 10 pA 0.25 V 63 + 25 300 + 25 500 + 25 1350 + 25
2 mA 2.1000000 100 pA 0.31 V 64 + 20 300 + 20 400 + 20 750 + 20
20 mA 21.000000 1 nA 0.4 V 65 + 20 300 + 20 400 + 20 750 + 20
200 mA 210.00000 10 nA 0.5 V 96 + 20 300 + 20 500 + 20 750 + 20
2 A 2.1000000 100 nA 1.5 V 500 + 20 600 + 20 900 + 20 1350 + 20
CURRENT ± (ppm of reading + ppm of range)
MAXIMUM BURDEN ± (ppm of reading + ppm of range)
ACCURACY
AC AMPS ACCURACY 90 Days, 1 Year or 2 Years, TCAL ±5°C, for 5% to 100% of range,±(% of reading + % of range)
RANGE 20Hz – 50Hz 50Hz – 200Hz 200Hz – 1kHz 1kHz – 10kHz 10kHz – 30kHz 30kHz – 50kHz 50kHz – 100kHz
200 µA 0.35 + 0.015 0.2 + 0.015 0.4 + 0.015 0.5 + 0.015
2 mA 0.3 + 0.015 0.15 + 0.015 0.12 + 0.015 0.12 + 0.015 0.25 + 0.015 0.3 + 0.015 0.5 + 0.015
20 mA 0.2 + 0.015 0.15 + 0.015 0.12 + 0.015 0.12 + 0.015 0.25 + 0.015 0.3 + 0.015 0.5 + 0.015
200 mA 0.3 + 0.015 0.15 + 0.015 0.12 + 0.015 0.15 + 0.015 0.5 + 0.015 1 + 0.015 3 + 0.015
2 A 0.35 + 0.015 0.2 + 0.015 0.3 + 0.015 0.45 + 0.015 1.5 + 0.015 4 + 0.015
QQZQ QZQQ
QZZQ
50
FREQUENCY COUNTER
RANGE: 1Hz–15MHz. ACCURACY: ±(0.03% of reading).
DC IN-CIRCUIT CURRENT
RANGE: 100µA to 12A. ACCURACY: ±(5% + 2 counts) over 2 years. TRACE RESISTANCE: 1mto 10typical.
TEMPERATURE
Built-in linearization for J, K, N, T, E, R, S, B thermocouple types to ITS-90
and 100platinum RTDs DIN 43760, IPTS-68, or ITS-90.
For complete specifications, refer to the 2001 Technical Data Book.
GENERAL/STANDARDS COMPLIANCE
POWER
Voltage: 90–134V and 180–264V, universal self-selecting. Frequency: 50Hz, 60Hz, or 400Hz self-identifying. Consumption: <55 VA.
ENVIRONMENT
Operating Temperature: 0° to 50°C. Storage Temperature: –40° to +70°C. Humidity: 80% R.H., 0° to 35°C. Altitude: 4,500m (15,000 ft) operating; 12,000m (40,000 ft.) non-
operating.
PHYSICAL
Case Dimensions: 90mm high × 214mm wide × 369mm deep
(31⁄2in × 81⁄2in × 141⁄2in).
Unit Weight: 4.2kg (9.2 lbs).
STANDARDS
EMI/RFI: Conforms to VDE 0871B (per Vfg 1046/1984), IEC 801-2,
FCC part 15 Class B, CISPR-22 (EN55022).
Safety: Conforms to IEC348, CAN/CSC22.2 No. 231, MIL-T-
28800E1.
www.keithley.com
DIGITAL MULTIMETERS
2001 and 2002 High Performance Multimeters
2002 Condensed Specifications
DC VOLTS 10PLC, D
RANGE FULL SCALE RESOLUTION RESISTANCE Transfer 24 Hours 90 Days 1 Year 2 Years
200 mV ±210.000000 1 nV >100 G 0.4 + 1.5 3.5 + 3 15 + 8 19 + 9 23 + 10
2 V ±2.10000000 10 nV >100 G 0.2 + 0.15 1.2 + 0.3 6 + 0.8 10 + 0.9 14 + 1
20 V ±21.0000000 100 nV >100 G 0.1 + 0.05 1.2 + 0.1 6 + 0.15 10 + 0.15 14 + 0.15
200 V ±210.000000 1 µV 10 M±1% 0.5 + 0.08 5 + 0.4 14 + 2 22 + 2 30 + 2
1000 V ±1100.00000 10 µV 10 M±1% 1 + 0.05 5 + 0.08 14 + 0.4 22 + 0.4 30 + 0.4
10 ACCURACY
FILT
INPUT ± (ppm of reading + ppm of range)
51
AC VOLTS ACCURACY
RANGE 20–50Hz 50–100Hz 0.1–2kHz 2–10kHz 10–30kHz 30–50kHz 50–100kHz 100–200kHz 0.2–1MHz 1–2MHz
200 mV 0.25 + 0.015 0.07 + 0.015 0.02 + 0.01 0.02 + 0.01 0.025 + 0.01 0.05 + 0.01 0.3 + 0.015 0.75 + 0.025 2 + 0.1 5 + 0.2
20 V 0.25 + 0.015 0.07 + 0.015 0.03 + 0.015 0.04 + 0.015 0.05 + 0.015 0.07 + 0.015 0.3 + 0.015 0.75 + 0.025 4 + 0.2 7 + 0.2 200 V 0.25 + 0.015 0.07 + 0.015 0.03 + 0.015 0.04 + 0.015 0.05 + 0.015 0.07 + 0.015 0.3 + 0.015 0.75 + 0.025 4 + 0.2 750 V 0.25 + 0.015 0.1 + 0.015 0.05 + 0.015 0.06 + 0.015 0.08 + 0.015 0.1 + 0.015 0.5 + 0.015
RESISTANCE 10PLC,Offset comp. on, D
200 210.000000 1 µ 960 µA 2.5 + 2 5 + 3 15 + 4 17 + 4 20 + 4
200 k 210.000000 1 m 9.6 µA 2.5 + 0.4 5.5 + 0.5 29 + 0.8 35 + 0.9 40 + 1
QQZQ QZQQ
QZZQ
200 M 210.000000 1 1.4 µA 50 + 0.5 150 + 1 500 + 3 550 + 3 600 + 3
DC AMPS
200 µA 210.00000 10 pA 0.25 V 50 + 6 275 + 25 350 + 25 500 + 25
200 mA 210.00000 10 nA 0.35 V 75 + 5 300 + 20 375 + 20 525 + 20
AC AMPS ACCURACY 90 Days,1 Year or 2 Years, TCAL ±5°C, for 5% to 100% of range,±(% of reading + % of range)
200 µA 0.35 + 0.015 0.2 + 0.015 0.4 + 0.015 0.5 + 0.015
20 mA 0.3 + 0.015 0.15 + 0.015 0.12 + 0.015 0.12 + 0.015 0.25 + 0.015 0.3 + 0.015 0.5 + 0.015
200 mA 0.3 + 0.015 0.15 + 0.015 0.12 + 0.015 0.15 + 0.015 0.5 + 0.015 1 + 0.015 3 + 0.015
(Normal Mode–TRMS), Low Freq.mode extends down to 1Hz; Peak AC, dB, AC+DC also available.
90 Days, 1 Year or 2 Years, ±2°C from last AC self-cal for 1% to 100% of range, ±(% of reading + % of range)
2 V 0.25 + 0.015 0.07 + 0.015 0.02 + 0.01 0.02 + 0.01 0.025 + 0.01 0.05 + 0.01 0.3 + 0.015 0.75 + 0.025 2 + 0.1 5 + 0.2
10 ACCURACY
FILT
RANGE FULL SCALE RESOLUTION SOURCE Transfer 24 Hours 90 Days 1 Year 2 Years
20 21.0000000 100 n 7.2 mA 2.5 + 3 5 + 4.5 15 + 6 17 + 6 20 + 6
2k 2100.00000 10 µΩ 960 µA 1.3 + 0.2 2.5 + 0.3 7 + 0.4 9 + 0.4 11 + 0.4
20 k 21.0000000 100 µ 96 µA 1.3 + 0.2 2.5 + 0.3 7 + 0.4 9 + 0.4 11 + 0.4
2M 2.10000000 10 mΩ 1.9 µA 5 + 0.2 12 + 0.3 53 + 0.5 65 + 0.5 75 + 0.5
20 M 21.0000000 100 m 1.4 µA 15 + 0.1 50 + 0.2 175 + 0.6 250 + 0.6 300 + 0.6
1G 1.05000000 10 Ω 1.4 µA 250 + 2.5 750 + 5 2000 + 15 2050 + 15 2100 + 15
RANGE FULL SCALE RESOLUTION VOLTAGE 24 Hours 90 Days 1 Year 2 Years
2 mA 2.1000000 100 pA 0.3 V 50 + 5 275 + 20 350 + 20 500 + 20
20 mA 21.000000 1 nA 0.35 V 50 + 5 275 + 20 350 + 20 500 + 20
2 A 2.1000000 100 nA 1.1 V 350 + 5 600 + 20 750 + 20 1000 + 20
RANGE 20Hz – 50Hz 50Hz – 200Hz 200Hz – 1kHz 1kHz – 10kHz 10kHz – 30kHz 30kHz – 50kHz 50kHz – 100kHz
2 mA 0.3 + 0.015 0.15 + 0.015 0.12 + 0.015 0.12 + 0.015 0.25 + 0.015 0.3 + 0.015 0.5 + 0.015
2 A 0.35 + 0.015 0.2 + 0.015 0.3 + 0.015 0.45 + 0.015 1.5 + 0.015 4 +0.015
CURRENT ± (ppm of reading + ppm of range)
MAXIMUM BURDEN ± (ppm of reading + ppm of range)
ACCURACY
GENERAL/STANDARDS COMPLIANCE
POWER
Voltage: 90–134V and 180–264V, universal self-selecting. Frequency: 50Hz, 60Hz, or 400Hz self-identifying. Consumption: <55 VA.
ENVIRONMENT
Operating Temperature: 0° to 50°C. Storage Temperature: –40° to +70°C. Humidity: 80% R.H., 0° to 35°C.
PHYSICAL
Case Dimensions: 90mm high × 214mm wide × 369mm deep
1
⁄2in × 81⁄2in × 141⁄2in).
(3
Unit Weight: <4.2kg (<9.2 lbs).
STANDARDS
EMI/RFI: Conforms to VDE 0871B (per Vfg 1046/1984), IEC 801-2,
FCC part 15 Class B, CISPR-22 (EN55022).
Safety: Conforms to IEC348, CAN/CSC22.2 No. 231, MIL-T-
28800E1.
www.keithley.com
FREQUENCY COUNTER
RANGE: 1Hz–15MHz. ACCURACY: ±(0.03% of reading)
DC IN-CIRCUIT CURRENT
RANGE: 100µA to 12A. ACCURACY: ±(5% + 500µA), 1 year or 2 years. TRACE RESISTANCE: 1mto 10Ω.
TEMPERATURE
Built-in linearization for J, K, T, E, R, S, N, B thermocouple types to ITS-90
and 100platinum RTDs DIN43760, IPTS-68, ITS-90, and SPRTs to ITS-90.
For complete specifications, refer to the 2002 Technical Data Book.