Keithley Instruments 2001-TCSCAN Data Sheet

DIGITAL MULTIMETERS
2000 Series Scanners
Keithley has designed three plug-in scanner cards for use with Model 2000, 2001, 2002, and 2010 DMMs. Simply by inserting the appropriate card in the option slot on the DMM’s back panel, users can create a multipoint test and measurement system that’s well suited to applications involving up to ten measurement points. Combining scan and measurement capabilities in a single instrument reduces the need for extra equipment, eliminates complexities of triggering, timing, and processing issues, and helps reduce test time significantly. Also, since the scanners were designed specifically to work with these DMMs, measurement integrity is assured. For larger ATE applications, the DMMs can be easily combined with Keithley’s 7000 Series switch matrices and cards for high-speed, high­accuracy scanning and measurement.
Model 2000-SCAN
Model 2001-SCAN
Model 2000-SCAN Scanner Card
• Ten analog input channels (2-pole)
• Configurable as 4-pole, 5-channel
Model 2000, 2001, 2002, and 2010 DMM users can create an economical ten-channel data logger simply by insert­ing the Model 2000-SCAN scanner card into the DMM’s back panel option slot. Users can multiplex one of ten 2­pole or one of five 4-pole signals into the Model 2000 and/or any combination of 2- or 4-pole signals. Contact potential is typically less than 500nV per contact pair (1µV maximum) for high measurement integrity.
Model 2001-SCAN Scanner Card
• Ten analog input channels
• Two channels of 2-pole high-speed solid-state switching
• High-speed ratio and delta measurements
ORDERING INFORMATION
2000-SCAN 10-Channel Scanner Card
with Test Leads for Models 2000, 2001, 2002, and 2010 Multimeters
2001-SCAN 10-Channel Scanner Card
with Test Leads and two high-speed channels for Models 2000, 2001, 2002, and 2010 Multimeters
2001-TCSCAN 9-Channel Thermocouple/
General Purpose Scanner Card with Test Leads for Models 2000, 2001, 2002, and 2010 Multimeters
2000-SCAN-20 See page 46 for a descrip-
tion and specifications for this 20-channel scanner card, used with the 2000-20 DMM.
These products are available with an Extended Warranty. See page 635 for com­plete ordering information.
The 2001-SCAN is a high-speed multiplexing general­purpose scanner card for use with the 2000, 2001, 2002, or 2010 multimeter. Installing the 2001-SCAN card into the meter transforms it into a high-accuracy, high-speed ten-channel data logger for a wide variety of mixed-signal applications. The card’s 500nV–125V signal range makes it well suited for both low and high level signals. Two high-speed solid-state channels make it possible to calculate ratio and delta when used with the 2001, 2002, or 2010.
Model 2001-TCSCAN
ACCESSORIES AVAILABLE
7401 Type K Thermocouple Wire Kit (100 ft)
See page 235 for descriptions of all accessories.
QUESTIONS?
1-800-552-1115 (U.S. only)
Call toll free for technical assistance,
product support or ordering information, or
visit our website at www.keithley.com.
Model 2001-TCSCAN Thermocouple Scanner Card
• Nine analog input channels
• Scan up to 43 channels per second
• Built-in temperature reference for thermocouple cold-junction compensation
• Suitable for general-purpose measurements of voltage, resistance, and frequency
• 500nV–125V signal range
When installed in a 2000, 2001, 2002, or 2010 DMM, the Model 2001-TCSCAN provides up to nine channels of cold-junction compensated temperature measurements and/or voltage, resistance, and frequency measurements. When the card is installed in the 2001 or 2002, the DMM will linearize type J, K, E, R, S, B, and T thermocouples automatically. The Model 2000 DMM provides built-in support for J, K, and T-type thermocouples. The Model 2010 has built-in support for J, K, T, and N thermocouples. Readings are output directly in the user’s choice of Celsius, Fahrenheit, or Kelvin scales. Total system accuracy is 0.65°C (with Type J TCs). When used with the 2001, 2002, or 2010, the 2001-TCSCAN can also measure temperature directly using 2- or 4-wire RTDs.
www.keithley.com
www.keithley.com
DIGITAL MULTIMETERS
QUESTIONS?
1-800-552-1115 (U.S. only)
Call toll free for technical assistance,
product support or ordering information, or
visit our website at www.keithley.com.
2000 Series Scanners
SCANNER OPTION 2000-SCAN
GENERAL: 10 channels of 2-pole relay input. All
channels configurable to 4-pole.
CAPABILITIES: Multiplex one of ten 2-pole or
one of five 4-pole signals into DMM.
INPUTS
Maximum Signal Level:
DC Signals: 110V DC, 1A switched, 30VA
maximum (resistive load).
AC Signals: 125V AC rms or 175V AC peak,
100kHz maximum, 1A switched, 62.5VA maximum (resistive load).
Contact Life: >10
5
operations at maximum
signal level; >10
8
operations cold switch-
ing.
Contact Resistance: <1at end of contact
life.
Actuation Time: 2.5ms maximum on/off.
Contact Potential:
<±500nV typical per contact, 1µV max. <±500nV typical per contact pair, 1µV max.
Connector Type: Screw terminal, #22 AWG
wire size.
Isolation Between Any Two Terminals:
>10
9
, <75pF.
Isolation Between Any Terminal and Earth:
>109Ω, <150pF.
Common Mode Voltage: 350V peak between
any terminal and earth.
Maximum Voltage Between Any Two
Terminals: 200V peak.
Maximum Voltage Between Any Terminal
and Model 2001 Input LO: 200V peak.
ENVIRONMENTAL: Meets all Model 2000 envi-
ronmental specifications.
DIMENSIONS, WEIGHT: 21mm high × 72mm
wide × 221mm deep (0.83 in. × 2.83 in. × 8.7 in.). Adds 0.4kg (10 oz.).
SCANNER OPTION 2001-SCAN
GENERAL: 10 Channels:
8 channels of 2-pole relay input. 2 channels of 2-pole solid-state input. All channels configurable to 4-pole.
CAPABILITIES:
a. Multiplex one of ten 2-pole or one of five 4-
pole signals into DMM and/or any combin­ation of 2- or 4-pole signals, or
b. Two-channel, high-speed multiplexing into
the DMM, or
c. High-speed ratio or delta measurements.
FUNCTIONS (can be mixed from channel to channel)
D C V, A C V, 4-Wire Ω,
4-Wire RTD, 2-Wire
Frequency DCI, ACI 2-Wire RTD
Relay Channels: Yes Yes* Yes
Solid State Channels: Yes
* On 2 channels, using user-supplied 1⁄
4
W shunt
resistor.
RELAY INPUTS Maximum Signal Level:
DC Signals: 110V DC, 1A switched, 30VA max-
imum (resistive load).
AC Signals: 125V AC rms or 175V AC peak,
100kHz maximum, 1A switched, 62.5VA maximum (resistive load).
Contact Life: >10
5
operations at maximum
signal level; >10
8
operations cold switching.
Contact Resistance: <1at end of contact life. Actuation Time: 2.5ms maximum on/off.
SOLID STATE INPUTS
Resistance: <275(including 200series
resistor).
Maximum Signal Level: 110V DC, 175V peak
AC, 50mA, 100kHz maximum.
Actuation Time: 150µs maximum on, 100µs
maximum off.
ALL INPUTS
Contact Potential:
<±500nV typical per contact, 1µV max. <±500nV typical per contact pair, 1µV max.
Operating Speed: See Operating Speed sec-
tion for complete speed specifications.
Connector Type: Screw terminal, #22 AWG
wire size.
Isolation Between Any Two Terminals: >10
9
,
<75pF.
Isolation Between Any Terminal and Earth:
>10
9
, <150pF.
Common Mode Voltage: 350V peak between
any terminal and earth.
Maximum Voltage Between Any Two Termin-
als: 200V peak.
Maximum Voltage Between Any Terminal
and Model 2001 Input LO: 200V peak.
ENVIRONMENTAL: Meets all Model 2001 envi-
ronmental specifications.
DIMENSIONS, WEIGHT: 21mm high × 72mm
wide × 221mm deep (0.83 in. × 2.83 in. × 8.7 in.). Adds 0.4kg (10 oz.).
SCANNER CONFIGURATION: Same as 2000-
SCAN.
SCANNER OPTION 2001-TCSCAN
Thermocouple Accuracy
Total Absolute
Error
1
, 1 Year
Thermocouple Default 0°-18°C &
Type Range Resolution 18°-28°C 28°-50°C
J –100° to 760°C 0.1°C ±0.65°C ±1.08°C K –100° to 1372°C 0.1°C ±0.70°C ±1.32°C T –100° to 400°C 0.1°C ±0.68°C ±1.22°C E –100° to 1000°C 0.1°C ±0.67°C ±1.11°C R 0° to 1768°C 1.0°C ±1.31°C ±3.06°C S 0° to 1768°C 1.0°C ±1.30°C ±3.02°C B 350° to 1820°C 1.0°C ±1.65°C ±4.14°C
1
When used with Model 2001 Multimeter.
Specifications apply to channels 2–6. Add 0.06°C for each adjacent channel
away from channel 6.
Extended range (Types J, K, T, E): –200. 0°C to –100. 1°C add ±0.1°C.
Excluding thermocouple error.
GENERAL: 9 channels of 2-pole analog input, 1 cold junction sensor. FUNCTIONS: DCV, ACV, 4-wire , Thermocouple, 2-wire , 4-wire RTD, 2-
wire RTD, Frequency (can be mixed from channel to channel).
THERMOCOUPLE SCAN SPEED: 25 channels/second @ 0.1°C resolution;
43 channels/second @ 1°C resolution.
CAPABILITIES: Multiplex one of nine 2-pole or one of four 4-pole analog
signals into Model 2001 DMM and/or any combination of 2- or 4-pole analog signals.
REFERENCE OUTPUT:+200µV/°C
(+54.63mV at 0°C).
ALL INPUTS
Maximum Signal Levels
2
: DC: 110V DC, <1A switched, 30VA maximum (resistive load). AC: 125V AC rms or 175V AC peak, 1A switched, 62.5VA maximum (resis-
tive load).
Contact Life: >10
5
operations at maximum signal level; >108operations
cold switching.
Contact Resistance
2
: <1at end of contact life. Actuation Time: 2.5ms maximum on/off. Contact Potential
3
: <±500nV typical per contact, 1µV max. Connector Type: Screw terminal, #22 AWG wire size (0.062 O.D.). Isolation Between Any Two Terminals: >10
9
, <75pF.
Common Mode Voltage: 350V peak between any terminal and earth. Maximum Voltage Between Any Two Terminals: 110V DC, 125V AC rms.
DIMENSIONS,WEIGHT: 21mm high × 72mm wide × 221mm deep (0.83 in
× 2.83 in × 8.7 in). Net weight 283g (10 oz).
2
Channels 5 and 10 maximum power = 0.25 VA maximum (factory installed
120, 5%,
1
⁄4W resistors. User may replace with jumper. See note in man-
ual for complete instructions.)
3
Channels 5 and 10 contact potential = ±1µV typical, 2µV max.
HI
LO
HI
LO
HI
LO
HI
LO
HI
LO
4-pole 2-pole
Out A
Out B
Ch. 2–4
Ch. 1 (Reference
(Junction)
Ch. 5
Ch. 7–9
Ch. 6
Ch. 10
Scanner Configuration for Models 2000-SCAN and 2001-SCAN
Scanner Configuration for Model 2001-TCSCAN
HI
Ch. 1
LO
Ch. 2–4
HI
Ch. 5
LO
*Solid state switches on 2001-SCAN only
HI
Ch. 6
LO
Ch. 7–9
HI
Ch. 10
LO
*
4-pole 2-pole
*
HI
Out A
LO
HI
Out B
LO
Loading...