Keithley Instruments 2001 Calibration Manual

Model 2001 Multimeter
Calibration Manual
A GREATER MEASURE OF CONFIDENCE
Model 2001 Multimeter
Calibration Manual
©1992, Keithley Instruments, Inc.
All rights reserved.
Cleveland, Ohio, U.S.A.
Document Number: 2001-905-01 Rev. G
WARRANTY
Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship for a period of 3 years from date of shipment.
Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables, rechargeable batteries, diskettes, and documentation.
During the warranty period, we will, at our option, either repair or replace any product that proves to be defective.
To exercise this warranty, write or call your local Keithley representative, or contact Keithley headquarters in Cleveland, Ohio. You will be given prompt assistance and return instructions. Send the product, transportation prepaid, to the indicated service facility. Repairs will be made and the product returned, transportation prepaid. Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days.
LIMITATION OF WARRANTY
This warranty does not apply to defects resulting from product modification without Keithley’s express written consent, or misuse of any product or part. This warranty also does not apply to fuses, software, non-rechargeable batteries, damage from battery leak­age, or problems arising from normal wear or failure to follow instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE. THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OF ITS INSTRU­MENTS AND SOFTWARE EVEN IF KEITHLEY INSTRUMENTS, INC., HAS BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COSTS OF REMOVAL AND INSTALLATION, LOSSES SUSTAINED AS THE RESULT OF INJURY TO ANY PERSON, OR DAMAGE TO PROPERTY.
A G R E A T E R M E A S U R E O F C O N F I D E N C E
Keithley Instruments, Inc.
Corporate Headquarters • 28775 Aurora Road • Cleveland, Ohio 44139 • 440-248-0400 • Fax: 440-248-6168 • 1-888-KEITHLEY (534-8453) • www.keithley.com
Belgium: Sint-Pieters-Leeuw • 02-363 00 40 • Fax: 02-363 00 64 • www.keithley.nl Italy: Milano • 02-48 39 16 01 • Fax: 02- 48 39 16 28 • www.keithley.it China: Beijing • 8610-82251886 • Fax: 8610-82251892 • www.keithley.com.cn Japan: Tokyo • 81-3-5733-7555 • Fax: 81-3-5733-7556 • www.keithley.jp Finland: Helsinki • 09-5306-6560 • Fax: 09-5306-6565 • www.keithley.com Korea: Seoul • 82-2-574-7778 • Fax: 82-2-574-7838 • www.keithley.com France: Saint-Aubin • 01-64 53 20 20 • Fax: 01-60 11 77 26 • www.keithley.fr Netherlands: Gorinchem • 0183-635333 • Fax: 0183-630821 • www.keithley.nl Germany: Germering • 089/84 93 07-40 • Fax: 089/84 93 07-34 • www.keithley.de Singapore: Singapore • 65-6747-9077 • Fax: 65-6747-2991 • www.keithley.com Great Britain: Theale • 0118 929 7500 • Fax: 0118 929 7519 • www.keithley.co.uk Sweden: Solna • 08-509 04 600 • Fax: 08-655 26 10 • www.keithley.com India: Bangalore • 91-80 2212 8027 • Fax: 91-80 2212 8005 • www.keithley.com Tai wa n : Hsinchu • 886-3-572-9077 • Fax: 886-3-572-9031 • www.keithley.com.tw
3/04
Manual Print History
The print history shown below lists the printing dates of all Revisions and Addenda created for this manual. The Revision Level letter increases alphabetically as the manual undergoes subsequent updates. Addenda, which are released between Revi­sions, contain important change information that the user should incorporate immediately into the manual. Addenda are num­bered sequentially. When a new Revision is created, all Addenda associated with the previous Revision of the manual are incorporated into the new Revision of the manual. Each new Revision includes a revised copy of this print history page.
Revision A (Document Number 2001-905-01) ....................................................................................... April 1992
Revision B (Document Number 2001-905-01) ........................................................................................ June 1992
Revision C (Document Number 2001-905-01) ........................................................................................ May 1993
Addendum C (Document Number 2001-905-02)..................................................................................... June 1993
Addendum C (Document Number 2001-905-03)............................................................................November 1993
Addendum C (Document Number 2001-905-04)................................................................................ January 1995
Revision D (Document Number 2001-905-01) .................................................................................... August 1995
Revision E (Document Number 2001-905-01) ........................................................................................ April 1996
Revision F (Document Number 2001-905-01)................................................................................November 2003
Revision G (Document Number 2001-905-01) ........................................................................................ May 2004
All Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc. Other brand and product names are trademarks or registered trademarks of their respective holders.

Safety Precautions

The following safety precautions should be observed before using this product and any associated instrumentation. Although some in­struments and accessories would normally be used with non-haz­ardous voltages, there are situations where hazardous conditions may be present.
This product is intended for use by qualified personnel who recog­nize shock hazards and are familiar with the safety precautions re­quired to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using the product. Refer to the manual for complete product specifications.
If the product is used in a manner not specified, the protection pro­vided by the product may be impaired.
The types of product users are: Responsible body is the individual or group responsible for the use
and maintenance of equipment, for ensuring that the equipment is operated within its specifications and operating limits, and for en­suring that operators are adequately trained.
Operators use the product for its intended function. They must be trained in electrical safety procedures and proper use of the instru­ment. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product to keep it operating properly, for example, setting the line voltage or replacing consumable materials. Maintenance procedures are de­scribed in the manual. The procedures explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel are trained to work on live circuits, and perform safe installations and repairs of products. Only properly trained ser­vice personnel may perform installation and service procedures.
Keithley products are designed for use with electrical signals that are rated Measurement Category I and Measurement Category II, as described in the International Electrotechnical Commission (IEC) Standard IEC 60664. Most measurement, control, and data I/O sig­nals are Measurement Category I and must not be directly connect­ed to mains voltage or to voltage sources with high transient over­voltages. Measurement Category II connections require protection for high transient over-voltages often associated with local AC mains connections. Assume all measurement, control, and data I/O connections are for connection to Category I sources unless other­wise marked or described in the Manual.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30V RMS,
42.4V peak, or 60VDC are present. A good safety practice is to ex-
pect that hazardous voltage is present in any unknown circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators are pre­vented access and/or insulated from every connection point. In some cases, connections must be exposed to potential human con­tact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000 volts, no conductive part of
the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance limited sources. NEVER connect switching cards directly to AC mains. When con­necting sources to switching cards, install protective devices to limit fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connect­ed to a properly grounded power receptacle. Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input power dis­connect device must be provided, in close proximity to the equip­ment and within easy reach of the operator.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting cables or jump­ers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the com­mon side of the circuit under test or power line (earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being measured.
The instrument and accessories must be used in accordance with its specifications and operating instructions or the safety of the equip­ment may be impaired.
Do not exceed the maximum signal levels of the instruments and ac­cessories, as defined in the specifications and operating informa­tion, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with same type and rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground connections.
If you are using a test fixture, keep the lid closed while power is ap­plied to the device under test. Safe operation requires the use of a lid interlock.
5/03
If a screw is present, connect it to safety earth ground using the wire recommended in the user documentation.
!
The symbol on an instrument indicates that the user should re­fer to the operating instructions located in the manual.
The symbol on an instrument shows that it can source or mea­sure 1000 volts or more, including the combined effect of normal and common mode voltages. Use standard safety precautions to avoid personal contact with these voltages.
The symbol indicates a connection terminal to the equipment frame.
The WA RN ING heading in a manual explains dangers that might result in personal injury or death. Always read the associated infor­mation very carefully before performing the indicated procedure.
The CAUTION heading in a manual explains hazards that could damage the instrument. Such damage may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans. Before performing any maintenance, disconnect the line cord and
all test cables.
To maintain protection from electric shock and fire, replacement components in mains circuits, including the power transformer, test leads, and input jacks, must be purchased from Keithley Instru­ments. Standard fuses, with applicable national safety approvals, may be used if the rating and type are the same. Other components that are not safety related may be purchased from other suppliers as long as they are equivalent to the original component. (Note that se­lected parts should be purchased only through Keithley Instruments to maintain accuracy and functionality of the product.) If you are unsure about the applicability of a replacement component, call a Keithley Instruments office for information.
To clean an instrument, use a damp cloth or mild, water based cleaner. Clean the exterior of the instrument only. Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with no case or chassis (e.g., data acquisition board for installation into a com­puter) should never require cleaning if handled according to instruc­tions. If the board becomes contaminated and operation is affected, the board should be returned to the factory for proper cleaning/ser­vicing.

Table of Contents

1 Performance Verification
1.1 Introduction..........................................................................................................................................................1-1
1.2 Environmental conditions .................................................................................................................................... 1-1
1.3 Warm-up period ...................................................................................................................................................1-1
1.4 Line power ...........................................................................................................................................................1-1
1.5 Recommended test equipment .............................................................................................................................1-2
1.6 Verification limits ................................................................................................................................................1-2
1.7 Restoring default conditions ................................................................................................................................1-2
1.8 Verification procedures........................................................................................................................................1-4
1.8.1 DC volts verification.................................................................................................................................... 1-4
1.8.2 AC volts verification.................................................................................................................................... 1-5
1.8.3 DC current verification ..............................................................................................................................1-10
1.8.4 AC current verification ..............................................................................................................................1-11
1.8.5 Resistance verification ...............................................................................................................................1-13
1.8.6 Frequency accuracy verification................................................................................................................1-16
1.8.7 Temperature reading checks ...................................................................................................................... 1-17
2Calibration
2.1 Introduction..........................................................................................................................................................2-1
2.2 Environmental conditions .................................................................................................................................... 2-1
2.3 Warm-up period ...................................................................................................................................................2-2
2.4 Line power ...........................................................................................................................................................2-2
2.5 Calibration lock.................................................................................................................................................... 2-2
2.5.1 Comprehensive calibration lock...................................................................................................................2-2
2.5.2 Low-level calibration lock ...........................................................................................................................2-2
2.5.3 IEEE-488 bus calibration lock status ........................................................................................................... 2-2
2.6 IEEE-488 bus calibration commands and program .............................................................................................2-2
2.6.1 Calibration commands .................................................................................................................................2-2
2.6.2 Required order of command execution........................................................................................................2-4
2.6.3 Example calibration command program ......................................................................................................2-4
2.7 Calibration errors ................................................................................................................................................. 2-5
i
2.7.1 Front panel error message summary.............................................................................................................2-5
2.7.2 IEEE-488 bus error reporting.......................................................................................................................2-6
2.8 Comprehensive calibration...................................................................................................................................2-6
2.8.1 Recommended equipment for comprehensive calibration...........................................................................2-6
2.8.2 Front panel comprehensive calibration ........................................................................................................2-6
2.8.3 IEEE-488 bus comprehensive calibration....................................................................................................2-9
2.9 AC self-calibration .............................................................................................................................................2-12
2.9.1 Front panel AC calibration.........................................................................................................................2-12
2.9.2 IEEE-488 bus AC self-calibration..............................................................................................................2-12
2.10 Low-level calibration..........................................................................................................................................2-12
2.10.1 Recommended equipment for low-level calibration ..................................................................................2-13
2.10.2 Low-level calibration summary..................................................................................................................2-13
2.10.3 Front panel low-level calibration procedure...............................................................................................2-15
2.10.4 IEEE-488 bus low-level calibration procedure ..........................................................................................2-18
3 Calibration Command Reference
3.1 Introduction .........................................................................................................................................................3-1
3.2 Command summary..............................................................................................................................................3-1
3.3 :CALibration:PROTected subsystem...................................................................................................................3-3
3.3.1 :LOCK..........................................................................................................................................................3-3
3.3.2 :SWITch?......................................................................................................................................................3-3
3.3.3 :SAVE...........................................................................................................................................................3-4
3.3.4 :DATA?........................................................................................................................................................3-4
3.3.5 :DATE..........................................................................................................................................................3-5
3.3.6 :DATE?.........................................................................................................................................................3-5
3.3.7 :NDUE..........................................................................................................................................................3-5
3.3.8 :NDUE?........................................................................................................................................................3-6
3.3.9 :LLEVel........................................................................................................................................................3-6
3.3.10 :DC ...............................................................................................................................................................3-9
3.4 :CALibration:UNPRotected Subsystem.............................................................................................................3-12
3.4.1 :ACCompensation ......................................................................................................................................3-12
3.5 Bus error reporting .............................................................................................................................................3-13
3.5.1 Calibration error summary .........................................................................................................................3-13
3.5.2 Detecting calibration errors........................................................................................................................3-13
3.6 Detecting calibration step completion................................................................................................................3-13
3.6.1 Using the *OPC? query..............................................................................................................................3-13
3.6.2 Using the *OPC command.........................................................................................................................3-14
Appendices
A Model 2001 Specifications..................................................................................................................................A-1
B Calibration Programs...........................................................................................................................................B-1
C Calibration Messages...........................................................................................................................................C-1
D Alternate Calibration Sources .............................................................................................................................D-1
ii

List of Illustrations

1 Performance Verification
Figure 1-1 Connections for DC volts verification ...................................................................................................... 1-5
Figure 1-2 Connections for AC volts verification (all except 2MHz test).................................................................. 1-6
Figure 1-3 Connections for AC volts verification (2MHz frequency only) ............................................................... 1-7
Figure 1-4 Connections for DC current verification................................................................................................. 1-11
Figure 1-5 Connections for AC current verification................................................................................................. 1-12
Figure 1-6 Connections for resistance verification (20¾-200k¾ ranges)................................................................. 1-14
Figure 1-7 Connections for resistance verification (2M¾ - 200M¾ ranges) ........................................................... 1-15
Figure 1-8 1G¾ resistor test box construction.......................................................................................................... 1-15
Figure 1-9 Connections for frequency accuracy verification.................................................................................... 1-16
2 Calibration
Figure 2-1 Low-thermal short connections................................................................................................................. 2-7
Figure 2-2 Connections for comprehensive calibration.............................................................................................. 2-8
Figure 2-3 Calibration voltage connections.............................................................................................................. 2-16
Figure 2-4 Current calibration connections .............................................................................................................. 2-17
Figure 2-5 Synthesizer connections .......................................................................................................................... 2-18
Appendices
Figure B-1 Low-thermal short connections..................................................................................................................................................... B-3
Figure B-2 Calibration connection for comprehensive calibration............................................................................................................... B-4
Figure B-3 Calibration voltage connections..................................................................................................................................................... B-4
Figure B-4 Calibration current connections..................................................................................................................................................... B-5
Figure B-5 Synthesizer connections.................................................................................................................................................................. B-5
iii

List of Tables

1 Performance Verification
Table 1-1 Recommended equipment for performance verification........................................................................... 1-2
Table 1-2 Limits for DC volts verification................................................................................................................ 1-5
Table 1-3 Limits for normal mode AC voltage verification...................................................................................... 1-8
Table 1-4 Limits for low-frequency mode AC voltage verification.......................................................................... 1-9
Table 1-5 Limits for AC peak voltage verification ................................................................................................. 1-10
Table 1-6 Limits for DC current verification .......................................................................................................... 1-11
Table 1-7 Limits for AC current verification .......................................................................................................... 1-12
Table 1-8 Limits for resistance verification (20¾-200M¾ ranges)......................................................................... 1-13
Table 1-9 Limits for resistance verification (1G¾ range) ....................................................................................... 1-15
Table 1-10 Frequency verification limits .................................................................................................................. 1-16
Table 1-11 Thermocouple temperature reading checks............................................................................................. 1-17
Table 1-12 RTD probe temperature reading checks.................................................................................................. 1-18
2 Calibration
Table 2-1 IEEE-488 bus calibration command summary.......................................................................................... 2-3
Table 2-2 Calibration error messages........................................................................................................................ 2-5
Table 2-3 Recommended equipment for comprehensive calibration........................................................................ 2-6
Table 2-4 Front panel comprehensive calibration summary ..................................................................................... 2-6
Table 2-5 IEEE-488 bus comprehensive calibration summary ............................................................................... 2-10
Table 2-6 Recommended equipment for low-level calibration............................................................................... 2-13
Table 2-7 Low-level calibration summary .............................................................................................................. 2-14
3 Calibration Command Reference
Table 3-1 IEEE-488 bus calibration command summary.......................................................................................... 3-2
Table 3-2 Low-level calibration commands.............................................................................................................. 3-6
Table 3-3 Comprehensive calibration commands ..................................................................................................... 3-9
Table 3-4 Calibration error summary ...................................................................................................................... 3-13
v
Appendices
Table B-1 Recommended equipment for comprehensive calibration ....................................................................... B-2
Table B-2 Recommended equipment for low-level calibration................................................................................. B-2
Table C-1 Calibration errors...................................................................................................................................... C-2
Table D-1 Alternate calibration sources .................................................................................................................... D-1
vi
1

Performance Verification

1.1 Introduction

The procedures in this section are intended to verify that Model 2001 accuracy is within the limits stated in the instru­ment one-year specifications. These procedures can be per­formed when the instrument is first received to ensure that no damage or misadjustment has occurred during shipment. Verification may also be performed whenever there is a ques­tion of instrument accuracy, or following calibration, if de­sired.
NOTE
If the instrument is still under warranty, and its performance is outside specified limits, contact your Keithley representa­tive or the factory to determine the correct course of action.
This section includes the following:
1.2 Environmental conditions: Covers the temperature and humidity limits for verification.
1.3 Warm-up period: Describes the length of time the Model 2001 should be allowed to warm up before test­ing.
1.4 Line power: Covers power line voltage ranges during testing.
1.5 Recommended equipment: Summarizes recom­mended equipment and pertinent specifications.
1.6 Verification limits: Explains how reading limits were calculated.
1.7 Restoring factory default conditions: Gives step-by­step procedures for restoring default conditions before each test procedure.
1.8 Verification procedures: Details procedures to verify measurement accuracy of all Model 2001 measure­ment functions.

1.2 Environmental conditions

Verification measurements should be made at an ambient temperature of 18-28°C (65-82°F), and at a relative humidity of less than 80% unless otherwise noted.

1.3 Warm-up period

The Model 2001 must be allowed to warm up for at least one hour before performing the verification procedures. If the in­strument has been subjected to temperature extremes (out­side the range stated in paragraph 1.2), allow additional time for internal temperatures to stabilize. Typically, it takes one additional hour to stabilize a unit that is 10°C (18°F) outside the specified temperature range.
The calibration equipment should also be allowed to warm up for the minimum period specified by the manufacturer.
1-1
Performance Verification

1.4 Line power

The Model 2001 should be tested while operating from a line voltage in the range of 90-134V or 180-250V at a frequency of 50, 60, or 400Hz.

1.5 Recommended test equipment

Table 1-1 lists all test equipment required for verification. Alternate equipment may be used as long as that equipment has specifications at least as good as those listed in the table. See Appendix D for a list of alternate calibration sources.

1.6 Verification limits

The verification limits stated in this section have been calcu­lated using only Model 2001 one year specifications, and they do not include test equipment tolerance. If a particular measurement falls slightly outside the allowed range, recal­culate new limits based both on Model 2001 specifications and pertinent calibration equipment specifications.

1.7 Restoring default conditions

Before performing each performance verification procedure, restore instrument bench default conditions as follows:
1. From the normal display mode, press the MENU key. The instrument will display the following:
MAIN MENU
SAVESETUP GPIB CALIBRATION
2. Select SAVESETUP, and press ENTER. The following will be displayed:
SETUP MENU
SAVE RESTORE POWERON RESET
3. Select RESET, and press ENTER. The display will then appear as follows:
RESET ORIGINAL DFLTS
BENCH GPIB
4. Select BENCH, then press ENTER. The following will be displayed:
RESETTING INSTRUMENT
ENTER to confirm; EXIT to abort
5. Press ENTER again to confirm instrument reset.
Table 1-1
Recommended equipment for performance verification
Mfg. Model Description Specifications*
Fluke 5700A Calibrator ±5ppm basic uncertainty.
DC voltage:
190mV: ±11ppm
1.9V: ±5ppm 19V: ±5ppm 190V: ±7ppm 1000V: ±9ppm
AC voltage, 10Hz-1MHz (40Hz-20kHz specifications):
190mV: ±150ppm
1.9V: ±78ppm 19V: ±78ppm 190V: ±85ppm
DC current:
190µA: ±102ppm
1.9mA: ±55ppm 19mA: ±55ppm 190mA: ±65ppm
1.9A: ±96ppm
1-2
Performance Verification
Table 1-1 (cont.)
Recommended equipment for performance verification
Mfg. Model Description Specifications*
Fluke 5700A Calibrator AC current, 40Hz-10kHz (40Hz-1kHz specifications):
190µA: ±245ppm
1.9mA: ±160ppm 19mA: ±160ppm 190mA: ±170ppm
1.9A: ±670ppm
Resistance:
19¾: ±26ppm 190¾: ±17ppm
1.9k¾: ±12ppm 19k¾: ±11ppm 190k¾: ±13ppm
1.9M¾: ±19ppm 19M¾: ±47ppm
100M¾: ±120ppm Fluke 5725A Amplifier AC voltage, 1kHz-10kHz: 750V: ±85ppm Fluke 5700A-03 Wideband AC option 190mV, 1.9V @ 2MHz, ±0.1% Fluke 5440A-7002 Low thermal cable set Keithley CA-18-1 Low capacitance cable Low capacitance dual banana to dual banana shielded cable (for
ACV), 1.2m (4 ft.) in length.
Keithley R-289-1G 1G¾ resistor NOTE: Resistor should be characterized to within ±10,000 ppm and
mounted in shielded test box (see procedure).
Metal component box (for 1G¾ resistor)
Insulated banana plugs (2)
(for test box) Keithley 3940 Multifunction Synthesizer 1Hz-15MHz, ±5ppm General
Radio
1433-T Precision Decade Resis-
tance Box
10-400¾, ±0.02%
⎯⎯ Megohmmeter 1G¾, ±1%
* 90-day calibrator specifications shown include total uncertainty at specified output. The 1.9V output includes 0.5ppm transfer uncertainty. See Appendix D for recommendation on alternate calibration sources.
1-3
Performance Verification

1.8 Verification procedures

The following paragraphs contain procedures for verifying instrument accuracy specifications for the following measur­ing functions:
•DC volts
• AC volts
• DC current
• AC current
• Resistance
•Frequency
• Temperature
If the Model 2001 is out of specifications and not under war­ranty, refer to the calibration procedures in Section 2.
WARNING
The maximum common-mode voltage (voltage between INPUT LO and chas­sis ground) is 500V peak. Exceeding this value may cause a breakdown in insula­tion, creating a shock hazard. Some of the procedures in this section may ex­pose you to dangerous voltages. Use standard safety precautions when such dangerous voltages are encountered to avoid personal injury caused by electric shock.
NOTE
Do not connect test equipment to the Mod­el 2001 through a scanner.
1.8.1 DC volts verification
DC voltage accuracy is verified by applying accurate DC voltages from a calibrator to the Model 2001 input and veri­fying that the displayed readings fall within specified ranges.
Follow the steps below to verify DCV measurement accura­cy.
CAUTION
Do not exceed 1100V peak between IN­PUT HI and INPUT LO, or instrument damage may occur.
1. Turn on the Model 2001 and the calibrator, and allow a one-hour warm-up period before making measure­ments.
NOTE
Use shielded, low-thermal connections when testing the 200mV range to avoid er­rors caused by noise or thermal offsets. Connect the shield to calibrator output LO. (See Table 1-1.)
2. Connect the Model 2001 to the calibrator, as shown in Figure 1-1. Be sure to connect calibrator HI to Model 2001 INPUT HI and calibrator LO to Model 2001 IN­PUT LO as shown.
3. Restore Model 2001 factory default conditions, as ex­plained in paragraph 1.7.
4. Set digital filter averaging as follows: A. From normal display, press CONFIG then DCV. B. Select FILTER, then press ENTER. C. Select AVERAGING, then press ENTER. D. Using the cursor and range keys, set the averaging
parameter to 10 readings, then press ENTER. E. Press EXIT as necessary to return to normal display. F. If the FILT annunciator is off, press FILTER to en-
able the filter.
5. Select the Model 2001 200mV DC range.
NOTE
Do not use auto-ranging for any of the ver­ification tests because auto-range hystere­sis may cause the Model 2001 to be on an incorrect range.
6. Set the calibrator output to 0.000000mVDC, and allow the reading to settle.
7. Enable the Model 2001 REL mode. Leave REL enabled for the remainder of the DC volts verification test.
8. Set the calibrator output to +190.0000mVDC, and allow the reading to settle.
9. Verify that the Model 2001 reading is within the limits summarized in Table 1-2.
10. Repeat steps 8 and 9 for the remaining ranges and volt­ages listed in Table 1-2.
11. Repeat the procedure for each of the ranges with nega­tive voltages of the same magnitude as those listed in Ta­ble 1-2.
1-4
Performance Verification
5700A Calibrator (Output DC Voltage)
Input HI
INPUT
HI
1100V PEAK
LO
500V PEAK
R
2A 250V
AMPS
Input LO
+1. 900000 VDC
PREV
DCV ACV DCI ACI Ω2 Ω4
DISPLAY
NEXT
REL TRIG STORE
POWER
LOCAL CHAN SCAN CONFIG MENU EXIT ENTER
INFO
Model 2001
2001 MULTIMETER
RECALL
FREQ TEMP
FILTER MATH
SENSE
Ω 4 WIRE
350V PEAK
INPUTS
F
RANGE
FRONT/REAR
AUTO
RANGE
CAL
Note: Use shielded, low-thermal cables when testing 200mV range. Use internal Guard (EX GRD LED is off).
Figure 1-1
Connections for DC volts verification
Table 1-2
Limits for DC volts verification
2001
DCV
range
200mV
200V
1000V
Notes:
1. Repeat procedure for negative voltages.
2. Reading limits shown do not include calibrator uncertainty.
2V
20V
Applied DC
voltage
190.0000mV
1.900000V
19.00000V
190.0000V
1000.000V
Reading limits
(18° to 28°C, 1 year)
189.9918mV to 190.0082mV
1.899949V to 1.900052V
18.99946V to 19.00054V
189.9922V to 190.0078V
999.953V to 1000.047V
Output HI
Output LO
Ground link installed.
Normal mode
1. Turn on the Model 2001, calibrator, and amplifier, and allow a one-hour warm-up period before making mea­surements.
2. Connect the Model 2001 to the calibrator, as shown in Figure 1-2. Be sure to connect the amplifier HI to Model 2001 INPUT HI, and amplifier LO to Model 2001 IN­PUT LO as shown. Connect the power amplifier to the calibrator using the appropriate connector on the rear of the calibrator.
3. Restore Model 2001 factory default conditions, as ex­plained in paragraph 1.7.
4. Select the ACV function and the 200mV range on the Model 2001, and make sure that REL is disabled.
1.8.2 AC volts verification
AC voltage accuracy is checked by applying accurate AC voltages at specific frequencies from an AC calibration source and then verifying that each Model 2001 AC voltage reading falls within the specified range. The three ACV ver­ification procedures that follow include:
• Normal mode
• Low-frequency mode
• Peak ACV
CAUTION
Do not exceed 1100V peak or 2 ×
7
V•Hz between INPUT HI and IN-
10 PUT LO, or instrument damage may oc­cur.
NOTE
Do not use REL to null offsets when per­forming AC volts tests.
5. Set the calibrator output to 190.000mVAC at a frequen­cy of 20Hz, and allow the reading to settle.
6. Verify that the Model 2001 reading is within the limits summarized in Table 1-3.
7. Repeat steps 5 and 6 for 190mVAC at the remaining fre­quencies listed in Table 1-3 (except 2MHz).Verify that instrument readings fall within the required limits listed in the table.
8. Repeat steps 5 through 7 for the 2V, 20V, 200V, and 750VAC ranges, using the input voltages and limits stat­ed in Table 1-3.
9. Connect the Model 2001 to the wideband calibrator out­put (Figure 1-3).
1-5
Performance Verification
10. Set the calibrator output to 190.0000mV at a frequency of 2MHz.
11. Verify that the reading is within limits stated in Table 1-
3.
12. Repeat steps 10 and 11 for 1.900V input on the 2V range.
Low-frequency mode
1. Turn on the Model 2001, calibrator, and amplifier, and allow a one-hour warm-up period before making mea­surements.
2. Connect the Model 2001 to the calibrator, as shown in Figure 1-2. Be sure to connect the amplifier HI to Model 2001 INPUT HI, and amplifier LO to Model 2001 IN­PUT LO as shown. Connect the power amplifier to the calibrator using the appropriate connector on the rear of the calibrator.
3. Restore Model 2001 factory default conditions, as ex­plained in paragraph 1.7.
4. Select the ACV function and the 200mV range on the Model 2001, and make sure that REL is disabled.
NOTE
Do not use REL to null offsets when per­forming AC volts tests. Also, do not en­able the filter.
5. Select the low-frequency mode as follows: A. Press CONFIG ACV, select AC-TYPE, then press
ENTER. B. Select LOW-FREQ-RMS, then press ENTER. C. Press EXIT as required to return to normal display.
6. Set the calibrator output to 190.000mVAC at a frequen­cy of 10Hz, and allow the reading to settle.
7. Verify that the Model 2001 reading is within the limits summarized in Table 1-4.
8. Repeat steps 6 and 7 for 190mVAC at the remaining fre­quencies listed in the table.
9. Repeat steps 6 through 8 for the 2V, 20V, 200V, and 750VAC ranges, using the input voltages and limits stat­ed in Table 1-4.
CAUTION
Do not apply more than 400V at 50kHz, 80V at 250kHz, 40V at 500kHz, or 20V at 1MHz, or instrument damage may occur.
Input HI
INPUT
HI
1100V PEAK
LO
500V PEAK
R
2A 250V
AMPS
Input LO
CA-18-1 Low­capacitance cable
Output HI
Output LO
1. 900000 VAC RMS
PREV
DCV ACV DCI ACI Ω2 Ω4
DISPLAY
NEXT
REL TRIG STORE RECALL
POWER
INFO LOCAL CHAN SCAN CONFIG MENU EXIT ENTER
Model 2001
2001 MULTIMETER
FREQ TEMP
FILTER MATH
SENSE
Ω 4 WIRE
350V PEAK
INPUTS
F
RANGE
FRONT/REAR
AUTO
RANGE
CAL
Figure 1-2
Connections for AC volts verification (all except 2MHz test)
5725 Amplifier (Connect to calibrator)
5700A Calibrator (Output AC Voltage)
Ground link installed.
Note: Use internal Guard (EX GRD LED is off).
1-6
1. 900000 VAC RMS
PREV
DCV ACV DCI ACI Ω2 Ω4
DISPLAY
NEXT
REL TRIG STORE RECALL
POWER
INFO LOCAL CHAN SCAN CONFIG MENU EXIT ENTER
Model 2001
2001 MULTIMETER
FREQ TEMP
FILTER MATH
RANGE
AUTO
RANGE
BNC to dual
banana
SENSE
INPUT
Ω 4 WIRE
HI
350V PEAK
LO
INPUTS
F
R
FRONT/REAR
2A 250V
AMPS
CAL
1100V PEAK
500V PEAK
Performance Verification
5725 Amplifier (Connect to calibrator)
50Ω
terminator
50Ω Coax
Wideband
output
Note: Use internal Guard (EX GRD LED is off).
Figure 1-3
Connections for AC volts verification (2MHz frequency only)
5700A Calibrator (Output AC Voltage)
Ground link installed.
1-7
Performance Verification
180.100mVto199.900mV
1.80100Vto1.99900V
to
186.000mV
1.86000Vto1.94000V
194.000mV
*
18.2000Vto19.8000V
to
189.000mV
189.647mVto190.353mV
1.89000Vto1.91000V
191.000mV
1.89647Vto1.90353V
to
189.856mV
189.875mVto190.125mV
189.875mVto190.125mV
1.89856Vto1.90144V
190.144mV
1.89875Vto1.90125V
1.89875Vto1.90125V
to
189.875mV
1.89875Vto1.90125V
190.125mV
***
18.9000Vto19.1000V
18.9647Vto19.0353V
18.9723Vto19.0277V
18.9742Vto19.0258V
18.9809Vto19.0192V
18.9856Vto19.0144V
189.640Vto190.360V
189.716Vto190.284V
189.735Vto190.265V
189.802Vto190.198V
189.849Vto190.151V
*****
748.12Vto751.88V
748.49Vto751.51V
748.72Vto751.28V
1-8
Table 1-3
to
to
19.1284V
189.678Vto190.322V
to
200V 190V 188.709V
189.685mV
1.89685Vto1.90315V
190.315mV
to
Allowable readings (1 year, 18° to 28°C)
Applied
voltage 20Hz 50Hz 1kHz 5kHz 25kHz 50kHz 100kHz 200kHz 1MHz 2MHz**
191.284mV
2V 1.9V 1.88716V
ACV
200mV 190mV 188.716mV
range
Limits for normal mode AC voltage verification
2001
18.9685Vto19.0315V
to
1.91284V
20V 19V 18.8716V
191.291V
to
748.12V
750V 750V
751.88V
7
V•Hz input.
*CAUTION: Do not exceed 2 × 10
**Use wideband option and connections when performing 2MHz tests.
NOTE: Limits shown do not include calibrator uncertainty. Reading limits do include the adder for AC Coupling of the input.
Table 1-4
Limits for low-frequency mode AC voltage verification
Allowable readings
2001 ACV
range
Applied
voltage
10Hz 50Hz 100Hz
(1 year, 18° to 28°C)
Performance Verification
200mV
190mV
189.837mV
190.163mV
2V
20V
200V
750V
NOTE: Specifications above 100Hz are the same as normal mode. Limits shown do not include calibrator uncertainty.
1.9V
19V
190V
750V
AC peak mode
1. Turn on the Model 2001, calibrator, and amplifier, and allow a one-hour warm-up period before making mea­surements.
2. Connect the Model 2001 to the calibrator, as shown in Figure 1-2. Be sure to connect the amplifier HI to Model 2001 INPUT HI, and amplifier LO to Model 2001 IN­PUT LO as shown. Connect the power amplifier to the calibrator using the appropriate connector on the rear of the calibrator.
3. Restore Model 2001 factory default conditions, as ex­plained in paragraph 1.7.
4. Select the ACV function and the 200mV range on the Model 2001, and make sure that REL is disabled.
NOTE
to
1.89837V to
1.90163V
18.9818V to
19.0182V
189.811V to
190.189V
189.875mV to
190.125mV
1.89875V to
1.90125V
18.9856V to
19.0144V
189.849V to
190.151V
748.72V to
751.28V
189.875mV to
190.125mV
1.89875V to
1.90125V
18.9856V to
19.0144V
189.849V to
190.151V
748.72V to
751.28V
C. Select FILTER, then press ENTER. D. Select AVERAGING, then press ENTER. E. Using the cursor and range keys, set the averaging
parameter to 10 readings, then press ENTER. F. Press EXIT as necessary to return to normal display. G. If the FLT annunciator is off, press FILTER to en-
able the filter.
6. Set the calibrator output to 100.000mVAC at a frequen­cy of 5kHz, and allow the reading to settle.
7. Verify that the Model 2001 reading is within the limits summarized in Table 1-5.
8. Repeat steps 6 and 7 for 100mVAC at the remaining fre­quencies listed in the table.
9. Repeat steps 6 through 8 for the 2V, 20V, 200V, and 750VAC ranges, using the input voltages and limits stat­ed in Table 1-5.
Do not use REL to null offsets when per­forming AC volts tests. Use AC coupling for 5kHz-1MHz tests. Use AC+DC cou­pling for 20Hz tests. (Use CONFIG-ACV to set coupling.)
5. Select the AC peak and filter modes as follows: A. Press CONFIG then ACV, select AC-TYPE, then
press ENTER.
B. Select PEAK, then press ENTER.
CAUTION
Do not apply more than 400V at 50kHz, 80V at 250kHz, 40V at 500kHz, or 20V at 1MHz, or instrument damage may occur.
10. Set input coupling to AC+DC, then repeat the procedure for a 20Hz input signal.
1-9
Performance Verification
Table 1-5
Limits for AC peak voltage verification
2001
ACV
range
Applied
voltage* 20Hz†
Allowable Readings (1 year, 18° to 28°C)
5kHz 25kHz 50kHz 100kHz 250kHz 500kHz 750kHz 1MHz
200mV 100mV 139.9mV
to
142.9mV
2V 1V 1.407V
to
1.421V
20V 10V 13.99V
to
14.29V
200V 190V 267.8V
to
269.6V
750V 750V
*Calibrator voltage is given as an RMS value. Model 2001 reading limits are peak AC values. **CAUTION: Do not apply more than 2 × 10 †Use AC+DC input coupling for 20Hz tests only. (Use CONFIG-ACV to set coupling.) NOTE: Limits shown do not include uncertainty calibrator.
1.8.3 DC current verification
DC current accuracy is checked by applying accurate DC currents from a calibrator to the instrument AMPS input and then verifying that the current readings fall within appropri­ate limits.
Follow the steps below to verify DCI measurement accuracy.
139.9mV to
142.9mV
1.407V to
1.421V
13.98V to
14.30V
267.8V to
269.6V
1054V
to
1067V
7
V•Hz
139.9mV to
142.9mV
1.407V to
1.421V
13.98V to
14.30V
267.7V to
269.7V
1053V
to
1068V
139.8mV to
143.0mV
1.406V to
1.422V
13.97V to
14.31V
267.6V to
269.8V ** ** ** ** ** **
139.6mV
143.2mV
1.404V
1.424V
13.96V
14.32V
267.4V
270.0V
4. Set digital filter averaging as follows: A. From normal display, press CONFIG then DCI. B. Select FILTER, then press ENTER. C. Select AVERAGING, then press ENTER. D. Using the cursor and range keys, set the averaging
E. Press EXIT as necessary to return to normal display. F. If the FILT annunciator is off, press FILTER to en-
CAUTION
Do not apply more than 2A, 250V to the AMPS input, or the amps protection fuse will blow.
1. Turn on the Model 2001 and the calibrator, and allow a one-hour warm-up period before making measure­ments.
2. Connect the Model 2001 to the calibrator, as shown in Figure 1-4. Be sure to connect calibrator HI to the AMPS input, and connect calibrator LO to INPUT LO as shown.
5. Select the DC current function (DCI) and the 200µA range on the Model 2001.
6. Set the calibrator output to +190.0000µADC, and allow the reading to settle.
7. Verify that the Model 2001 reading is within the limits summarized in Table 1-6.
8. Repeat steps 6 and 7 for the remaining currents listed in Table 1-6.
9. Repeat the procedure for each of the ranges with nega­tive currents of the same magnitude as those listed in Ta­ble 1-6.
3. Restore Model 2001 factory default conditions, as ex­plained in paragraph 1.7.
138.6mV
to
to
to
to
to
144.2mV
1.394V to
1.434V
13.86V to
14.42V ** ** ** **
136.5mV to
146.3mV
1.373V to
1.455V
13.65V to
14.63V
132.2mV to
150.6mV
1.330V to
1.498V
13.22V to
15.06V
parameter to 10 readings, then press ENTER.
able the filter.
127.3mV to
155.5mV
1.281V to
1.547V
12.73V to
15.55V
1-10
Performance Verification
5700A Calibrator (Output DC Current)
19.00000 mADC
PREV
DCV ACV DCI ACI Ω2 Ω4
DISPLAY
NEXT
REL TRIG STORE RECALL
POWER
INFO LOCAL CHAN SCAN CONFIG MENU EXIT ENTER
Model 2001
2001 MULTIMETER
FREQ TEMP
FILTER MATH
SENSE
Ω 4 WIRE
350V PEAK
INPUTS
F
RANGE
FRONT/REAR
AUTO
RANGE
CAL
Input
INPUT
LO
HI
1100V
PEAK
LO
500V PEAK
R
2A 250V
AMPS
Amps
Output HI
Output LO
Note: Use internal Guard (EX GRD LED is off).
Figure 1-4
Connections for DC current verification
Table 1-6
Limits for DC current verification
2001 DCI
range
200µA
Applied DC
current
190.0000µA
Reading limits
(1 year, 18° to 28°C)
189.9000µA to
190.1000µA
2mA
1.900000mA
1.899200mA to
1.900800mA
20mA
19.00000mA
18.99200mA to
19.00800mA
200mA
190.0000mA
189.9010mA to
190.0990mA
2A
1.900000A
1.898200A to
1.901800A
NOTES:
1. Repeat procedure for negative currents.
2. Reading limits shown do not include calibrator uncertainty.
1.8.4 AC current verification
AC current verification is performed by applying accurate AC currents at specific frequencies and then verifying that Model 2001 readings fall within specified limits.
Ground link installed.
Follow the steps below to verify ACI measurement accuracy.
CAUTION
Do not apply more than 2A, 250V to the AMPS input, or the current protection fuse will blow.
1. Turn on the Model 2001 and the calibrator, and allow a one-hour warm-up period before making measure­ments.
2. Connect the Model 2001 to the calibrator, as shown in Figure 1-5. Be sure to connect calibrator HI to the AMPS input, and connect calibrator LO to INPUT LO as shown.
3. Restore Model 2001 factory default conditions, as ex­plained in paragraph 1.7.
4. Select the AC current function and the 200µA range on the Model 2001.
5. Set the calibrator output to 190.000µA AC at a frequen­cy of 40Hz, and allow the reading to settle.
6. Verify that the Model 2001 reading is within the limits for the present current and frequency summarized in Ta­ble 1-7.
7. Repeat steps 4 and 5 for each frequency listed in Table 1-7.
8. Repeat steps 4 through 7 for the remaining ranges and frequencies listed in Table 1-7.
1-11
Performance Verification
5700A Calibrator (Output AC Current)
190.0000 μAAC RMS
PREV
DCV ACV DCI ACI Ω2 Ω4
DISPLAY
NEXT
REL TRIG STORE RECALL
POWER
INFO LOCAL CHAN SCAN CONFIG MENU EXIT ENTER
Model 2001
2001 MULTIMETER
FREQ TEMP
FILTER MATH
SENSE
Ω 4 WIRE
350V PEAK
INPUTS
F
RANGE
FRONT/REAR
AUTO
RANGE
Figure 1-5
Connections for AC current verification
Table 1-7
Limits for AC current verification
2001 ACI
range
Applied AC
current
200µA 190.000µA 188.260µA
2mA 1.90000mA 1.88355mA
20mA 19.0000mA 18.8355mA
200mA 190.000mA 188.355mA
2A 1.90000A 1.88250A
Note: Reading limits shown do not include calibrator uncertainty.
Input
INPUT
LO
HI
1100V PEAK
LO
500V PEAK
R
2A 250V
AMPS
CAL
Amps
Output HI
Output LO
Ground link installed.
Note: Use internal Guard (EX GRD LED is off).
Reading limits (1 year, 18° to 28°C)
40Hz 100Hz 1kHz 10kHz
to
191.740µA
to
1.91645mA
to
19.1645mA
to
191.645mA
to
1.91750A
189.560µA to
190.440µA
1.89657mA to
1.90344mA
18.9657mA to
19.0344mA
189.657mA to
190.344mA
1.89556A to
1.90444A
189.210µA to
190.790µA
1.89742mA to
1.90258mA
18.9742mA to
19.0258mA
189.742mA to
190.258mA
1.89390A to
1.90610A
189.020µA to
190.980µA
1.89742mA to
1.90258mA
18.9742mA to
19.0258mA
189.685mA to
190.315mA
1.89105A to
1.90895A
1-12
Performance Verification
1.8.5 Resistance verification
Resistance verification is performed by connecting accurate resistance values to the instrument and verifying that Model 2001 resistance readings are within stated limits.
Follow the steps below to verify resistance measurement ac­curacy.
CAUTION
Do not apply more than 1100V peak be­tween INPUT HI and LO or more than 350V peak between SENSE HI and LO, or instrument damage may occur.
20¾ - 200k¾ range verification
1. Turn on the Model 2001 and the calibrator, and allow a one-hour warm-up period before making measure­ments.
2. Set the calibrator for 4-wire resistance (external sense on).
3. Using shielded 4-wire connections, connect the Model 2001 to the calibrator, as shown in Figure 1-6. Be sure to connect calibrator HI and LO terminals to the Model 2001 HI and LO terminals (including SENSE HI and LO) as shown.
4. Restore Model 2001 factory default conditions, as ex­plained in paragraph 1.7.
5. Set operating modes as follows: A. From normal display, press CONFIG then ¾4. B. Select FILTER, then press ENTER. C. Select AVERAGING, then press ENTER. D. Using the cursor and range keys, set the averaging
parameter to 10 readings, then press ENTER. E. Select OFFSETCOMP, then press ENTER. F. Select ON, then press ENTER. G. Press EXIT to return to normal display.
6. Set the calibrator to output 19.000¾, and allow the read­ing to settle. Verify that the reading is within the limits stated in Table 1-8.
NOTE
Resistance values available in the Model 5700A calibrator may be slightly different than the stated nominal resistance values. Calculated limits stated in Table 1-8 should be recalculated based on actual cal­ibrator resistance values.
7. Set the calibrator output to 190.000¾, and allow the reading to settle.
8. Verify that the reading is within the limits stated in Table 1-8. (NOTE: Recalculate limits if calibrator resistance is not exactly as listed.)
9. Repeat steps 11 and 12 for the 2k¾ through 200k¾ rang­es using the values listed in Table 1-8. NOTE: Turn off­set compensation off when testing the 200k¾ range (see step 5).
Table 1-8
Limits for resistance verification (20¾-200M¾ ranges)
2001 ¾ range
20¾
Applied
resistance
19.0000¾
Reading limits
(1 year, 18° to 28°C)
18.99849¾ to
19.00151¾
200¾
190.000¾
189.9880¾ to
190.0120¾
2k¾
1.90000k¾
1.899897k¾ to
1.900103k¾
20k¾
19.0000k¾
18.99897k¾ to
19.00103k¾
200k¾
190.000k¾
189.9820k¾ to
190.0180k¾
2M¾
1.90000M¾
1.899687M¾ to
1.900313M¾
20M¾
19.0000M¾
18.98281M¾ to
19.01719M¾
200M¾
100.000M¾
97.9800M¾ to
102.0200M¾
NOTES:
1. Limits shown do not include calibrator uncertainty and are based on absolute calibration values shown. Recalculate limits using Model 2001 specifications if calibrator resistance values differ from nomi­nal values shown.
2. Use 4-wire connections and function for 20¾-200k¾ ranges. Use 2­wire connections and function for 2M¾-200M¾ ranges.
1-13
Performance Verification
5700A Calibrator (Output 2-Wire Resistance)
Input HI
INPUT
HI
1100V
PEAK
LO
500V PEAK
R
2A 250V
AMPS
Input LO
Output HI
Output LO
+1. 900000 kΩ
PREV
DCV ACV DCI ACI Ω2 Ω4
DISPLAY
NEXT
REL TRIG STORE RECALL
POWER
INFO LOCAL CHAN SCAN CONFIG MENU EXIT ENTER
Model 2001
2001 MULTIMETER
FREQ TEMP
FILTER MATH
SENSE
Ω 4 WIRE
350V PEAK
INPUTS
F
RANGE
FRONT/REAR
AUTO
RANGE
CAL
Note: Use shielded cable to minimize noise.
Disable calibrator external sense mode.
Use internal Guard (EX GRD LED is off).
Figure 1-6
Connections for resistance verification (20¾-200k¾ ranges)
2M¾ – 200M¾ range verification
1. Connect the DC calibrator and Model 2001 using the 2­wire connections shown in Figure 1-7.
2. Set the calibrator to the 2-wire mode (external sense off).
3. Set operating modes as follows: A. From normal display, press CONFIG then ¾2. B. Select FILTER, then press ENTER. C. Select AVERAGING, then press ENTER. D. Using the cursor and range keys, set the averaging
parameter to 10 readings, then press ENTER. E. Press EXIT to return to normal display. F. If the FILT annunciator is off, press FILTER to en-
able the filter.
4. Select the Model 2001 ¾2 function, and change to the 2M¾ range.
5. Set the calibrator to output 1.90000M¾, and allow the reading to settle.
6. Verify that the reading is within the limits for the 2M¾ range stated in Table 1-8. (NOTE: Recalculate limits if actual calibrator resistance differs from value shown.)
7. Repeat steps 4 through 6 for the 20M¾ (output
19.0000M¾) and 200M¾ (output 100.000M¾) ranges.
1G¾ range verification
Ground link installed.
2. Characterize the 1G¾ resistor to within ±10,000ppm or better using an accurate megohmmeter (see Table 1-1). Record the characterized value where indicated in Table 1-9. Also, compute the limits based on the value of R us­ing the formula at the bottom of the table.
NOTE
The value of the 1G¾ resistor should not exceed 1.05G¾.
3. Set operating modes as follows: A. From normal display, press CONFIG then ¾2. B. Select FILTER, then press ENTER. C. Select AVERAGING, then press ENTER. D. Using the cursor and range keys, set the averaging
parameter to 10 readings, then press ENTER. E. Press EXIT to return to normal display. F. If the FILT annunciator is off, press FILTER to en-
able the filter.
4. Select the 2-wire ohms function (¾2) and the 1G¾ range on the Model 2001.
5. Connect the 1G¾ resistor test box (from steps 1 and 2) to the INPUT HI and LO terminals of the Model 2001. Allow the reading to settle.
6. Verify that the Model 2001 reading is within the limits you calculated and recorded in Table 1-9.
1. Mount the 1G¾ resistor and the banana plugs to the test box, as shown in Figure 1-8. Be sure to mount the ba­nana plugs with the correct spacing. The resistor should be completely enclosed in and shielded by the metal test box. The resistor LO lead should be electrically con­nected to the test box to provide adequate shielding.
1-14
190.0000 kΩ
PREV
DCV ACV DCI ACI Ω2 Ω4
DISPLAY
NEXT
REL TRIG STORE RECALL
POWER
INFO LOCAL CHAN SCAN CONFIG MENU EXIT ENTER
Model 2001
2001 MULTIMETER
FREQ TEMP
FILTER MATH
Sense LO
Performance Verification
Sense LO
Sense HI
Sense HI
SENSE
INPUT
Ω 4 WIRE
HI
350V PEAK
F
RANGE
FRONT/REAR
AUTO
RANGE
Input HI
1100V PEAK
LO
500V PEAK
INPUTS
R
2A 250V
AMPS
CAL
Input LO
Output HI
Output LO
5700A Calibrator (Output 4-wire Resistance)
Note: Use shielded cables to minimize noise. Enable calibrator external sense mode. Use internal Guard (EX GRD LED is off).
Figure 1-7
Connections for resistance verification (2M¾ - 200M¾ ranges)
Insulated
Plug
HI
0.75"
LO
Banana
Plugs
Non-insulated Plug
Ground link installed.
1GΩ Resistor (Keithley part # R-289-1G)
Metal
Test Box
Figure 1-8
1G¾ resistor test box construction
Table 1-9
Limits for resistance verification (1G¾ range)
Characterized
resistor (R) Reading limit (1 year, 18° to 28°C)*
____________ G¾ _________G¾ to _________G¾
*1 Year limits = R ± (0.04R + 100,000) Where R = characterized value of 1G¾ resistor.
Note: Resistor must be accurately characterized before use (see text).
1-15
Performance Verification
1.8.6 Frequency accuracy verification
Frequency accuracy verification is performed by connecting an accurate frequency source to the Model 2001 inputs, and then verifying that the frequency readings are within stated limits.
Use the procedure below to verify the frequency measure­ment accuracy of the Model 2001.
1. Connect the frequency synthesizer to the Model 2001 INPUT terminals, as shown in Figure 1-9.
2. Turn on both instruments, and allow a one-hour warm­up period before measurement.
3. Set the synthesizer operating modes as follows: FREQ: 1Hz
AMPTD: 5V p-p OFFSET: 0V MODE: CONT FCTN: sine wave
4. Restore Model 2001 factory defaults, as explained in paragraph 1.7.
5. Press FREQ to place the Model 2001 in the frequency measurement mode.
6. Set maximum signal level to 10V as follows: A. Press CONFIG then FREQ. B. Select MAX-SIGNAL-LEVEL, then press ENTER. C. Select VOLTAGE, then press ENTER. D. Select 10V, then press ENTER. E. Press EXIT to return to normal display.
7. Verify that the Model 2001 frequency reading is within the limits shown in the first line of Table 1-10.
8. Set the synthesizer to each of the frequencies listed in Table 1-10, and verify that the Model 2001 frequency reading is within the required limits.
Table 1-10
Frequency verification limits
Synthesizer
frequency
1Hz
10Hz
100Hz
1kHz
10kHz
100kHz
1MHz 10MHz 15MHz
Reading limits
(1 year, 18° to 28°C)
0.9997Hz to 1.0003Hz
9.997Hz to 10.003Hz
99.97Hz to 100.03Hz
0.9997kHz to 1.0003kHz
9.997kHz to 10.003kHz
99.97kHz to 100.03kHz
0.9997MHz to 1.0003MHz
9.997MHz to 10.003MHz
14.995MHz to 15.005MHz
Model 2001
SENSE Ω 4 WIRE
1.0000 MHz
PREV
DCV ACV DCI ACI Ω2 Ω4 FREQ TEMP
DISPLAY
NEXT
REL TRIG STORE RECALL
POWER
INFO LOCAL CHAN SCAN CONFIG MENU EXIT ENTER
2001 MULTIMETER
FILTER MATH
Figure 1-9
Connections for frequency accuracy verification
350V PEAK
LO
INPUTS
F
FRONT/REAR
R
CAL
RANGE
AUTO
RANGE
BNC-to-Dual
Banana Plug
Model 3940 Synthesizer
3940 MULTIFUNCTION SYNTHESIZER
Adapter
INPUT
HI
1100V PEAK
500V PEAK
2A 250V
AMPS
Main
Function
Output
50Ω BNC Coaxial Cable
1-16
Performance Verification
1.8.7 Temperature reading checks
When using thermocouples, the Model 2001 displays tem­perature by measuring the DC thermocouple voltage, and then calculating the corresponding temperature. Similarly, the instrument computes RTD temperature readings by mea­suring the resistance of the RTD probe and calculating tem­perature from the resistance value.
Since the instrument computes temperature from DCV and resistance measurements, verifying the accuracy of those DCV and resistance measurement functions guarantees the accuracy of corresponding temperature measurements. Thus, it is not necessary to perform a comprehensive temper­ature verification procedure if DCV and resistance verifica­tion procedures show the instrument meets its specifications in those areas. However, those who wish to verify that the Model 2001 does in fact properly display temperature can use the following procedure to do so.
Selecting the temperature sensor
Follow the steps below to select the type of temperature sen­sor:
1. From normal display, press CONFIG then TEMP.
2. Select SENSOR, then press ENTER.
3. Select 4-WIRE RTD or THERMOCOUPLE as desired, then press ENTER.
4. Select the type of RTD probe or thermocouple you wish to test, then return to the CONFIG TEMPERATURE menu.
5. Select UNITS, then press ENTER.
6. Select DEG-C, then press ENTER.
7. Press EXIT as necessary to return to normal display.
8. Press the TEMP key to place the Model 2001 in the tem­perature display mode. Refer to further information be­low on how to check thermocouple and RTD probe readings.
Thermocouple temperature reading checks
To check thermocouple readings, simply apply the appropri­ate DC voltage listed in Table 1-11 to the Model 2001 IN­PUT jacks using a precision DC voltage source (such as the one used to verify DC voltage accuracy in paragraph 1.8.1), and check the displayed temperature reading. Be sure to use low-thermal cables for connections between the DC calibra­tor and the Model 2001 when making these tests.
NOTE
The voltages shown are based on a 0°C reference junction temperature. Use CON­FIG TEMP to set the default reference junction temperature to 0°C.
Table 1-11
Thermocouple temperature reading checks
Thermo-
couple type
J -4.215mV
K -3.242mV
T -3.089mV
E -4.777mV
R0.054mV
S0.055mV
B0.632mV
*Voltages shown are based on 0°C reference junction temperature. Use CONFIG-TEMP menu to set default reference junction to 0°C.
Applied DC
voltage*
0mV
1.277mV
5.268mV
42.283mV
0mV
1.000mV
4.095mV
54.125mV
0mV
0.992mV
4.277mV
20.252mV
0mV
1.495mV
6.317mV
75.608mV
0.647mV
4.471mV
20.878mV
0.645mV
4.234mV
18.504mV
1.241mV
4.833mV
13.585mV
Displayed
temperature (°C)
-90.5 to -89.5
-0.5 to +0.5
24.5 to 25.5
99.5 to 100.5
749.5 to 750.5
-90.5 to -89.5
-0.5 to +0.5
24.5 to 25.5
99.5 to 100.5
1349.5 to 1350.5
-90.5 to -89.5
-0.5 to +0.5
24.5 to 25.5
99.5 to 100.5
389.5 to 390.5
-90.6 to -89.4
-0.6 to +0.6
24.4 to 25.6
99.4 to 100.6
989.4 to 990.6 7 to 13
97 to 103
497 to 503
1747 to 1753
7 to 13
97 to 103
497 to 503
1747 to 1753
355 to 365 495 to 505
995 to 1005
1795 to 1805
1-17
Performance Verification
RTD Temperature reading checks
Use a precision decade resistance box (see Table 1-1) to sim­ulate probe resistances at various temperatures (Table 1-12). Be sure to use 4-wire connections between the decade resis­tance box and the Model 2001.
Table 1-12
RTD probe temperature reading checks
RTD probe
type
PT385
(=0.00385055)
PT3916
(=0.00392)
Applied
resistance
64.30¾ 100¾
109.73¾
138.51¾
313.71¾
63.68¾ 100¾
109.90¾
139.16¾
266.94¾
Displayed
temperature (°C)
-90.08 to -89.92
-0.08 to +0.08
24.92 to 25.08
99.92 to 100.08
599.86 to 600.14
-90.08 to -89.92
-0.08 to +0.08
24.92 to 25.08
99.92 to 100.08
449.86 to 450.14
1-18
2

Calibration

2.1 Introduction

This section gives detailed procedures for calibrating the Model 2001. There are three types of calibration procedures:
• Comprehensive calibration
• AC self-calibration
• Low-level calibration
Comprehensive calibration requires accurate calibration equipment to supply precise DC voltages and resistance val­ues. AC self-calibration requires no external equipment and can be performed at any time by the operator. Low-level cal­ibration is normally performed only at the factory where the instrument is manufactured and is not usually required in the field.
NOTE
Low-level calibration is required in the field only if the Model 2001 has been re­paired, or if the other calibration proce­dures cannot bring the instrument within stated specifications.
2.5 Calibration lock: Explains how to unlock calibration with the CAL switch.
2.6 IEEE-488 bus calibration commands and program:
Summarizes bus commands used for calibration, lists a simple calibration program, and also discusses other important aspects of calibrating the instrument over the bus.
2.7 Calibration errors: Details front panel error messages that might occur during calibration and also explains how to check for errors over the bus.
2.8 Comprehensive calibration: Covers comprehensive (user) calibration from the front panel and over the IEEE-488 bus.
2.9 AC self-calibration: Discusses the AC user calibra­tion process, both from the front panel and over the IEEE-488 bus.
2.10 Low-level calibration: Explains how to perform the low-level calibration procedure, which is normally re­quired only at the factory.
Section 2 includes the following information:
2.2 Environmental conditions: States the temperature and humidity limits for calibration.
2.3 Warm-up period: Discusses the length of time the Model 2001 should be allowed to warm up before cal­ibration.
2.4 Line power: States the power line voltage limits when calibrating the unit.

2.2 Environmental conditions

Calibration procedures should be performed at an ambient temperature of 23°± 1°C, and at a relative humidity of less than 80% unless otherwise noted.
2-1
Loading...
+ 77 hidden pages