Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship for a period of 3 years from
date of shipment.
Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables, rechargeable batteries,
diskettes, and documentation.
During the warranty period, we will, at our option, either repair or replace any product that proves to be defective.
To exercise this warranty, write or call your local Keithley representative, or contact Keithley headquarters in Cleveland, Ohio. You
will be given prompt assistance and return instructions. Send the product, transportation prepaid, to the indicated service facility.
Repairs will be made and the product returned, transportation prepaid. Repaired or replaced products are warranted for the balance
of the original warranty period, or at least 90 days.
LIMITATION OF WARRANTY
This warranty does not apply to defects resulting from product modification without Keithley’s express written consent, or misuse
of any product or part. This warranty also does not apply to fuses, software, non-rechargeable batteries, damage from battery leakage, or problems arising from normal wear or failure to follow instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED
WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE. THE REMEDIES PROVIDED HEREIN
ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR ANY DIRECT,
INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OF ITS INSTRUMENTS AND SOFTWARE EVEN IF KEITHLEY INSTRUMENTS, INC., HAS BEEN ADVISED IN ADVANCE OF THE
POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO:
COSTS OF REMOVAL AND INSTALLATION, LOSSES SUSTAINED AS THE RESULT OF INJURY TO ANY PERSON, OR
DAMAGE TO PROPERTY.
A G R E A T E R M E A S U R E O F C O N F I D E N C E
The print history shown below lists the printing dates of all Revisions and Addenda created for this manual. The Revision
Level letter increases alphabetically as the manual undergoes subsequent updates. Addenda, which are released between Revisions, contain important change information that the user should incorporate immediately into the manual. Addenda are numbered sequentially. When a new Revision is created, all Addenda associated with the previous Revision of the manual are
incorporated into the new Revision of the manual. Each new Revision includes a revised copy of this print history page.
Revision A (Document Number 2001-905-01) ....................................................................................... April 1992
Revision B (Document Number 2001-905-01) ........................................................................................ June 1992
Revision C (Document Number 2001-905-01) ........................................................................................ May 1993
Addendum C (Document Number 2001-905-02)..................................................................................... June 1993
Addendum C (Document Number 2001-905-03)............................................................................November 1993
Addendum C (Document Number 2001-905-04)................................................................................ January 1995
Revision D (Document Number 2001-905-01) .................................................................................... August 1995
Revision E (Document Number 2001-905-01) ........................................................................................ April 1996
Revision F (Document Number 2001-905-01)................................................................................November 2003
Revision G (Document Number 2001-905-01) ........................................................................................ May 2004
All Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc.
Other brand and product names are trademarks or registered trademarks of their respective holders.
Safety Precautions
The following safety precautions should be observed before using
this product and any associated instrumentation. Although some instruments and accessories would normally be used with non-hazardous voltages, there are situations where hazardous conditions
may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury. Read and follow all installation,
operation, and maintenance information carefully before using the
product. Refer to the manual for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product may be impaired.
The types of product users are:
Responsible body is the individual or group responsible for the use
and maintenance of equipment, for ensuring that the equipment is
operated within its specifications and operating limits, and for ensuring that operators are adequately trained.
Operators use the product for its intended function. They must be
trained in electrical safety procedures and proper use of the instrument. They must be protected from electric shock and contact with
hazardous live circuits.
Maintenance personnel perform routine procedures on the product
to keep it operating properly, for example, setting the line voltage
or replacing consumable materials. Maintenance procedures are described in the manual. The procedures explicitly state if the operator
may perform them. Otherwise, they should be performed only by
service personnel.
Service personnel are trained to work on live circuits, and perform
safe installations and repairs of products. Only properly trained service personnel may perform installation and service procedures.
Keithley products are designed for use with electrical signals that
are rated Measurement Category I and Measurement Category II, as
described in the International Electrotechnical Commission (IEC)
Standard IEC 60664. Most measurement, control, and data I/O signals are Measurement Category I and must not be directly connected to mains voltage or to voltage sources with high transient overvoltages. Measurement Category II connections require protection
for high transient over-voltages often associated with local AC
mains connections. Assume all measurement, control, and data I/O
connections are for connection to Category I sources unless otherwise marked or described in the Manual.
Exercise extreme caution when a shock hazard is present. Lethal
voltage may be present on cable connector jacks or test fixtures.
The American National Standards Institute (ANSI) states that a
shock hazard exists when voltage levels greater than 30V RMS,
42.4V peak, or 60VDC are present. A good safety practice is to ex-
pect that hazardous voltage is present in any unknown circuit
before measuring.
Operators of this product must be protected from electric shock at
all times. The responsible body must ensure that operators are prevented access and/or insulated from every connection point. In
some cases, connections must be exposed to potential human contact. Product operators in these circumstances must be trained to
protect themselves from the risk of electric shock. If the circuit is
capable of operating at or above 1000 volts, no conductive part of
the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits.
They are intended to be used with impedance limited sources.
NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective devices to limit
fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connected to a properly grounded power receptacle. Inspect the connecting
cables, test leads, and jumpers for possible wear, cracks, or breaks
before each use.
When installing equipment where access to the main power cord is
restricted, such as rack mounting, a separate main input power disconnect device must be provided, in close proximity to the equipment and within easy reach of the operator.
For maximum safety, do not touch the product, test cables, or any
other instruments while power is applied to the circuit under test.
ALWAYS remove power from the entire test system and discharge
any capacitors before: connecting or disconnecting cables or jumpers, installing or removing switching cards, or making internal
changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground. Always
make measurements with dry hands while standing on a dry, insulated
surface capable of withstanding the voltage being measured.
The instrument and accessories must be used in accordance with its
specifications and operating instructions or the safety of the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating information, and as shown on the instrument or test fixture panels, or
switching card.
When fuses are used in a product, replace with same type and rating
for continued protection against fire hazard.
Chassis connections must only be used as shield connections for
measuring circuits, NOT as safety earth ground connections.
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use of a
lid interlock.
5/03
If a screw is present, connect it to safety earth ground using the
wire recommended in the user documentation.
!
The symbol on an instrument indicates that the user should refer to the operating instructions located in the manual.
The symbol on an instrument shows that it can source or measure 1000 volts or more, including the combined effect of normal
and common mode voltages. Use standard safety precautions to
avoid personal contact with these voltages.
The symbol indicates a connection terminal to the equipment
frame.
The WA RN ING heading in a manual explains dangers that might
result in personal injury or death. Always read the associated information very carefully before performing the indicated procedure.
The CAUTION heading in a manual explains hazards that could
damage the instrument. Such damage may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and
all test cables.
To maintain protection from electric shock and fire, replacement
components in mains circuits, including the power transformer, test
leads, and input jacks, must be purchased from Keithley Instruments. Standard fuses, with applicable national safety approvals,
may be used if the rating and type are the same. Other components
that are not safety related may be purchased from other suppliers as
long as they are equivalent to the original component. (Note that selected parts should be purchased only through Keithley Instruments
to maintain accuracy and functionality of the product.) If you are
unsure about the applicability of a replacement component, call a
Keithley Instruments office for information.
To clean an instrument, use a damp cloth or mild, water based
cleaner. Clean the exterior of the instrument only. Do not apply
cleaner directly to the instrument or allow liquids to enter or spill on
the instrument. Products that consist of a circuit board with no case
or chassis (e.g., data acquisition board for installation into a computer) should never require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected,
the board should be returned to the factory for proper cleaning/servicing.
1.3Warm-up period ...................................................................................................................................................1-1
1.4Line power ...........................................................................................................................................................1-1
1.5Recommended test equipment .............................................................................................................................1-2
1.8.3DC current verification ..............................................................................................................................1-10
1.8.4AC current verification ..............................................................................................................................1-11
2.3Warm-up period ...................................................................................................................................................2-2
2.4Line power ...........................................................................................................................................................2-2
2.5.3IEEE-488 bus calibration lock status ........................................................................................................... 2-2
2.6IEEE-488 bus calibration commands and program .............................................................................................2-2
2.7.2IEEE-488 bus error reporting.......................................................................................................................2-6
2.9.1Front panel AC calibration.........................................................................................................................2-12
2.9.2IEEE-488 bus AC self-calibration..............................................................................................................2-12
3.6.1Using the *OPC? query..............................................................................................................................3-13
3.6.2Using the *OPC command.........................................................................................................................3-14
Figure 1-1 Connections for DC volts verification ...................................................................................................... 1-5
Figure 1-2 Connections for AC volts verification (all except 2MHz test).................................................................. 1-6
Figure 1-3 Connections for AC volts verification (2MHz frequency only) ............................................................... 1-7
Figure 1-4 Connections for DC current verification................................................................................................. 1-11
Figure 1-5 Connections for AC current verification................................................................................................. 1-12
Figure 1-6 Connections for resistance verification (20¾-200k¾ ranges)................................................................. 1-14
Table 1-1 Recommended equipment for performance verification........................................................................... 1-2
Table 1-2 Limits for DC volts verification................................................................................................................ 1-5
Table 1-3 Limits for normal mode AC voltage verification...................................................................................... 1-8
Table 1-4 Limits for low-frequency mode AC voltage verification.......................................................................... 1-9
Table 1-5 Limits for AC peak voltage verification ................................................................................................. 1-10
Table 1-6 Limits for DC current verification .......................................................................................................... 1-11
Table 1-7 Limits for AC current verification .......................................................................................................... 1-12
Table 1-8 Limits for resistance verification (20¾-200M¾ ranges)......................................................................... 1-13
Table 1-9 Limits for resistance verification (1G¾ range) ....................................................................................... 1-15
Table 1-10 Frequency verification limits .................................................................................................................. 1-16
Table 1-11 Thermocouple temperature reading checks............................................................................................. 1-17
Table 1-12 RTD probe temperature reading checks.................................................................................................. 1-18
2Calibration
Table 2-1 IEEE-488 bus calibration command summary.......................................................................................... 2-3
The procedures in this section are intended to verify that
Model 2001 accuracy is within the limits stated in the instrument one-year specifications. These procedures can be performed when the instrument is first received to ensure that no
damage or misadjustment has occurred during shipment.
Verification may also be performed whenever there is a question of instrument accuracy, or following calibration, if desired.
NOTE
If the instrument is still under warranty,
and its performance is outside specified
limits, contact your Keithley representative or the factory to determine the correct
course of action.
This section includes the following:
1.2Environmental conditions: Covers the temperature
and humidity limits for verification.
1.3Warm-up period: Describes the length of time the
Model 2001 should be allowed to warm up before testing.
1.4Line power: Covers power line voltage ranges during
testing.
1.5Recommended equipment: Summarizes recommended equipment and pertinent specifications.
1.6Verification limits: Explains how reading limits were
calculated.
1.7Restoring factory default conditions: Gives step-bystep procedures for restoring default conditions before
each test procedure.
1.8Verification procedures: Details procedures to verify
measurement accuracy of all Model 2001 measurement functions.
1.2Environmental conditions
Verification measurements should be made at an ambient
temperature of 18-28°C (65-82°F), and at a relative humidity
of less than 80% unless otherwise noted.
1.3Warm-up period
The Model 2001 must be allowed to warm up for at least one
hour before performing the verification procedures. If the instrument has been subjected to temperature extremes (outside the range stated in paragraph 1.2), allow additional time
for internal temperatures to stabilize. Typically, it takes one
additional hour to stabilize a unit that is 10°C (18°F) outside
the specified temperature range.
The calibration equipment should also be allowed to warm
up for the minimum period specified by the manufacturer.
1-1
Performance Verification
1.4Line power
The Model 2001 should be tested while operating from a line
voltage in the range of 90-134V or 180-250V at a frequency
of 50, 60, or 400Hz.
1.5Recommended test equipment
Table 1-1 lists all test equipment required for verification.
Alternate equipment may be used as long as that equipment
has specifications at least as good as those listed in the table.
See Appendix D for a list of alternate calibration sources.
1.6Verification limits
The verification limits stated in this section have been calculated using only Model 2001 one year specifications, and
they do not include test equipment tolerance. If a particular
measurement falls slightly outside the allowed range, recalculate new limits based both on Model 2001 specifications
and pertinent calibration equipment specifications.
1.7Restoring default conditions
Before performing each performance verification procedure,
restore instrument bench default conditions as follows:
1. From the normal display mode, press the MENU key.
The instrument will display the following:
MAIN MENU
SAVESETUP GPIB CALIBRATION
2. Select SAVESETUP, and press ENTER. The following
will be displayed:
SETUP MENU
SAVE RESTORE POWERON RESET
3. Select RESET, and press ENTER. The display will then
appear as follows:
RESET ORIGINAL DFLTS
BENCH GPIB
4. Select BENCH, then press ENTER. The following will
be displayed:
RESETTING INSTRUMENT
ENTER to confirm; EXIT to abort
5. Press ENTER again to confirm instrument reset.
Table 1-1
Recommended equipment for performance verification
Mfg.ModelDescriptionSpecifications*
Fluke5700ACalibrator±5ppm basic uncertainty.
DC voltage:
190mV: ±11ppm
1.9V: ±5ppm
19V: ±5ppm
190V: ±7ppm
1000V: ±9ppm
AC voltage, 10Hz-1MHz (40Hz-20kHz specifications):
190mV: ±150ppm
1.9V: ±78ppm
19V: ±78ppm
190V: ±85ppm
DC current:
190µA: ±102ppm
1.9mA: ±55ppm
19mA: ±55ppm
190mA: ±65ppm
1.9A: ±96ppm
1-2
Performance Verification
Table 1-1 (cont.)
Recommended equipment for performance verification
Keithley R-289-1G1G¾ resistorNOTE: Resistor should be characterized to within ±10,000 ppm and
mounted in shielded test box (see procedure).
Metal component box (for
1G¾ resistor)
Insulated banana plugs (2)
(for test box)
Keithley 3940Multifunction Synthesizer1Hz-15MHz, ±5ppm
General
Radio
1433-TPrecision Decade Resis-
tance Box
10-400¾, ±0.02%
⎯⎯Megohmmeter1G¾, ±1%
* 90-day calibrator specifications shown include total uncertainty at specified output. The 1.9V output includes 0.5ppm transfer uncertainty. See Appendix
D for recommendation on alternate calibration sources.
1-3
Performance Verification
1.8Verification procedures
The following paragraphs contain procedures for verifying
instrument accuracy specifications for the following measuring functions:
•DC volts
• AC volts
• DC current
• AC current
• Resistance
•Frequency
• Temperature
If the Model 2001 is out of specifications and not under warranty, refer to the calibration procedures in Section 2.
WARNING
The maximum common-mode voltage
(voltage between INPUT LO and chassis ground) is 500V peak. Exceeding this
value may cause a breakdown in insulation, creating a shock hazard. Some of
the procedures in this section may expose you to dangerous voltages. Use
standard safety precautions when such
dangerous voltages are encountered to
avoid personal injury caused by electric
shock.
NOTE
Do not connect test equipment to the Model 2001 through a scanner.
1.8.1DC volts verification
DC voltage accuracy is verified by applying accurate DC
voltages from a calibrator to the Model 2001 input and verifying that the displayed readings fall within specified ranges.
Follow the steps below to verify DCV measurement accuracy.
CAUTION
Do not exceed 1100V peak between INPUT HI and INPUT LO, or instrument
damage may occur.
1. Turn on the Model 2001 and the calibrator, and allow a
one-hour warm-up period before making measurements.
NOTE
Use shielded, low-thermal connections
when testing the 200mV range to avoid errors caused by noise or thermal offsets.
Connect the shield to calibrator output
LO. (See Table 1-1.)
2. Connect the Model 2001 to the calibrator, as shown in
Figure 1-1. Be sure to connect calibrator HI to Model
2001 INPUT HI and calibrator LO to Model 2001 INPUT LO as shown.
3. Restore Model 2001 factory default conditions, as explained in paragraph 1.7.
4. Set digital filter averaging as follows:
A. From normal display, press CONFIG then DCV.
B. Select FILTER, then press ENTER.
C. Select AVERAGING, then press ENTER.
D. Using the cursor and range keys, set the averaging
parameter to 10 readings, then press ENTER.
E. Press EXIT as necessary to return to normal display.
F. If the FILT annunciator is off, press FILTER to en-
able the filter.
5. Select the Model 2001 200mV DC range.
NOTE
Do not use auto-ranging for any of the verification tests because auto-range hysteresis may cause the Model 2001 to be on an
incorrect range.
6. Set the calibrator output to 0.000000mVDC, and allow
the reading to settle.
7. Enable the Model 2001 REL mode. Leave REL enabled
for the remainder of the DC volts verification test.
8. Set the calibrator output to +190.0000mVDC, and allow
the reading to settle.
9. Verify that the Model 2001 reading is within the limits
summarized in Table 1-2.
10. Repeat steps 8 and 9 for the remaining ranges and voltages listed in Table 1-2.
11. Repeat the procedure for each of the ranges with negative voltages of the same magnitude as those listed in Table 1-2.
1-4
Performance Verification
5700A Calibrator (Output DC Voltage)
Input HI
INPUT
HI
1100V
PEAK
LO
500V
PEAK
R
2A 250V
AMPS
Input
LO
+1. 900000 VDC
PREV
DCV ACV DCI ACI Ω2 Ω4
DISPLAY
NEXT
REL TRIG STORE
POWER
LOCAL CHAN SCAN CONFIG MENU EXIT ENTER
INFO
Model 2001
2001 MULTIMETER
RECALL
FREQ TEMP
FILTER MATH
SENSE
Ω 4 WIRE
350V
PEAK
INPUTS
F
RANGE
FRONT/REAR
AUTO
RANGE
CAL
Note: Use shielded, low-thermal cables
when testing 200mV range. Use
internal Guard (EX GRD LED is off).
Figure 1-1
Connections for DC volts verification
Table 1-2
Limits for DC volts verification
2001
DCV
range
200mV
200V
1000V
Notes:
1. Repeat procedure for negative voltages.
2. Reading limits shown do not include calibrator uncertainty.
2V
20V
Applied DC
voltage
190.0000mV
1.900000V
19.00000V
190.0000V
1000.000V
Reading limits
(18° to 28°C, 1 year)
189.9918mV to 190.0082mV
1.899949V to 1.900052V
18.99946V to 19.00054V
189.9922V to 190.0078V
999.953V to 1000.047V
Output HI
Output
LO
Ground link installed.
Normal mode
1. Turn on the Model 2001, calibrator, and amplifier, and
allow a one-hour warm-up period before making measurements.
2. Connect the Model 2001 to the calibrator, as shown in
Figure 1-2. Be sure to connect the amplifier HI to Model
2001 INPUT HI, and amplifier LO to Model 2001 INPUT LO as shown. Connect the power amplifier to the
calibrator using the appropriate connector on the rear of
the calibrator.
3. Restore Model 2001 factory default conditions, as explained in paragraph 1.7.
4. Select the ACV function and the 200mV range on the
Model 2001, and make sure that REL is disabled.
1.8.2AC volts verification
AC voltage accuracy is checked by applying accurate AC
voltages at specific frequencies from an AC calibration
source and then verifying that each Model 2001 AC voltage
reading falls within the specified range. The three ACV verification procedures that follow include:
• Normal mode
• Low-frequency mode
• Peak ACV
CAUTION
Do not exceed 1100V peak or 2 ×
7
V•Hz between INPUT HI and IN-
10
PUT LO, or instrument damage may occur.
NOTE
Do not use REL to null offsets when performing AC volts tests.
5. Set the calibrator output to 190.000mVAC at a frequency of 20Hz, and allow the reading to settle.
6. Verify that the Model 2001 reading is within the limits
summarized in Table 1-3.
7. Repeat steps 5 and 6 for 190mVAC at the remaining frequencies listed in Table 1-3 (except 2MHz).Verify that
instrument readings fall within the required limits listed
in the table.
8. Repeat steps 5 through 7 for the 2V, 20V, 200V, and
750VAC ranges, using the input voltages and limits stated in Table 1-3.
9. Connect the Model 2001 to the wideband calibrator output (Figure 1-3).
1-5
Performance Verification
10. Set the calibrator output to 190.0000mV at a frequency
of 2MHz.
11. Verify that the reading is within limits stated in Table 1-
3.
12. Repeat steps 10 and 11 for 1.900V input on the 2V
range.
Low-frequency mode
1. Turn on the Model 2001, calibrator, and amplifier, and
allow a one-hour warm-up period before making measurements.
2. Connect the Model 2001 to the calibrator, as shown in
Figure 1-2. Be sure to connect the amplifier HI to Model
2001 INPUT HI, and amplifier LO to Model 2001 INPUT LO as shown. Connect the power amplifier to the
calibrator using the appropriate connector on the rear of
the calibrator.
3. Restore Model 2001 factory default conditions, as explained in paragraph 1.7.
4. Select the ACV function and the 200mV range on the
Model 2001, and make sure that REL is disabled.
NOTE
Do not use REL to null offsets when performing AC volts tests. Also, do not enable the filter.
5. Select the low-frequency mode as follows:
A. Press CONFIG ACV, select AC-TYPE, then press
ENTER.
B. Select LOW-FREQ-RMS, then press ENTER.
C. Press EXIT as required to return to normal display.
6. Set the calibrator output to 190.000mVAC at a frequency of 10Hz, and allow the reading to settle.
7. Verify that the Model 2001 reading is within the limits
summarized in Table 1-4.
8. Repeat steps 6 and 7 for 190mVAC at the remaining frequencies listed in the table.
9. Repeat steps 6 through 8 for the 2V, 20V, 200V, and
750VAC ranges, using the input voltages and limits stated in Table 1-4.
CAUTION
Do not apply more than 400V at 50kHz,
80V at 250kHz, 40V at 500kHz, or 20V
at 1MHz, or instrument damage may
occur.
Input HI
INPUT
HI
1100V
PEAK
LO
500V
PEAK
R
2A 250V
AMPS
Input
LO
CA-18-1 Lowcapacitance cable
Output HI
Output
LO
1. 900000 VAC RMS
PREV
DCV ACV DCI ACI Ω2 Ω4
DISPLAY
NEXT
REL TRIG STORE RECALL
POWER
INFO LOCAL CHAN SCAN CONFIG MENU EXIT ENTER
Model 2001
2001 MULTIMETER
FREQ TEMP
FILTER MATH
SENSE
Ω 4 WIRE
350V
PEAK
INPUTS
F
RANGE
FRONT/REAR
AUTO
RANGE
CAL
Figure 1-2
Connections for AC volts verification (all except 2MHz test)
5725 Amplifier (Connect to calibrator)
5700A Calibrator (Output AC Voltage)
Ground link installed.
Note: Use internal Guard (EX GRD LED is off).
1-6
1. 900000 VAC RMS
PREV
DCV ACV DCI ACI Ω2 Ω4
DISPLAY
NEXT
REL TRIG STORE RECALL
POWER
INFO LOCAL CHAN SCAN CONFIG MENU EXIT ENTER
Model 2001
2001 MULTIMETER
FREQ TEMP
FILTER MATH
RANGE
AUTO
RANGE
BNC to dual
banana
SENSE
INPUT
Ω 4 WIRE
HI
350V
PEAK
LO
INPUTS
F
R
FRONT/REAR
2A 250V
AMPS
CAL
1100V
PEAK
500V
PEAK
Performance Verification
5725 Amplifier (Connect to calibrator)
50Ω
terminator
50Ω Coax
Wideband
output
Note: Use internal Guard (EX GRD LED is off).
Figure 1-3
Connections for AC volts verification (2MHz frequency only)
**Use wideband option and connections when performing 2MHz tests.
NOTE: Limits shown do not include calibrator uncertainty. Reading limits do include the adder for AC Coupling of the input.
Table 1-4
Limits for low-frequency mode AC voltage verification
Allowable readings
2001 ACV
range
Applied
voltage
10Hz50Hz100Hz
(1 year, 18° to 28°C)
Performance Verification
200mV
190mV
189.837mV
190.163mV
2V
20V
200V
750V
NOTE: Specifications above 100Hz are the same as normal mode. Limits shown do not include
calibrator uncertainty.
1.9V
19V
190V
750V
AC peak mode
1. Turn on the Model 2001, calibrator, and amplifier, and
allow a one-hour warm-up period before making measurements.
2. Connect the Model 2001 to the calibrator, as shown in
Figure 1-2. Be sure to connect the amplifier HI to Model
2001 INPUT HI, and amplifier LO to Model 2001 INPUT LO as shown. Connect the power amplifier to the
calibrator using the appropriate connector on the rear of
the calibrator.
3. Restore Model 2001 factory default conditions, as explained in paragraph 1.7.
4. Select the ACV function and the 200mV range on the
Model 2001, and make sure that REL is disabled.
NOTE
to
1.89837V
to
1.90163V
18.9818V
to
19.0182V
189.811V
to
190.189V
—
189.875mV
to
190.125mV
1.89875V
to
1.90125V
18.9856V
to
19.0144V
189.849V
to
190.151V
748.72V
to
751.28V
189.875mV
to
190.125mV
1.89875V
to
1.90125V
18.9856V
to
19.0144V
189.849V
to
190.151V
748.72V
to
751.28V
C. Select FILTER, then press ENTER.
D. Select AVERAGING, then press ENTER.
E. Using the cursor and range keys, set the averaging
parameter to 10 readings, then press ENTER.
F. Press EXIT as necessary to return to normal display.
G. If the FLT annunciator is off, press FILTER to en-
able the filter.
6. Set the calibrator output to 100.000mVAC at a frequency of 5kHz, and allow the reading to settle.
7. Verify that the Model 2001 reading is within the limits
summarized in Table 1-5.
8. Repeat steps 6 and 7 for 100mVAC at the remaining frequencies listed in the table.
9. Repeat steps 6 through 8 for the 2V, 20V, 200V, and
750VAC ranges, using the input voltages and limits stated in Table 1-5.
Do not use REL to null offsets when performing AC volts tests. Use AC coupling
for 5kHz-1MHz tests. Use AC+DC coupling for 20Hz tests. (Use CONFIG-ACV
to set coupling.)
5. Select the AC peak and filter modes as follows:
A. Press CONFIG then ACV, select AC-TYPE, then
press ENTER.
B. Select PEAK, then press ENTER.
CAUTION
Do not apply more than 400V at 50kHz,
80V at 250kHz, 40V at 500kHz, or 20V
at 1MHz, or instrument damage may
occur.
10. Set input coupling to AC+DC, then repeat the procedure
for a 20Hz input signal.
1-9
Performance Verification
Table 1-5
Limits for AC peak voltage verification
2001
ACV
range
Applied
voltage*20Hz†
Allowable Readings (1 year, 18° to 28°C)
5kHz25kHz50kHz100kHz250kHz500kHz750kHz1MHz
200mV100mV139.9mV
to
142.9mV
2V1V1.407V
to
1.421V
20V10V13.99V
to
14.29V
200V190V267.8V
to
269.6V
750V750V
*Calibrator voltage is given as an RMS value. Model 2001 reading limits are peak AC values.
**CAUTION: Do not apply more than 2 × 10
†Use AC+DC input coupling for 20Hz tests only. (Use CONFIG-ACV to set coupling.)
NOTE: Limits shown do not include uncertainty calibrator.
⎯
1.8.3DC current verification
DC current accuracy is checked by applying accurate DC
currents from a calibrator to the instrument AMPS input and
then verifying that the current readings fall within appropriate limits.
Follow the steps below to verify DCI measurement accuracy.
139.9mV
to
142.9mV
1.407V
to
1.421V
13.98V
to
14.30V
267.8V
to
269.6V
1054V
to
1067V
7
V•Hz
139.9mV
to
142.9mV
1.407V
to
1.421V
13.98V
to
14.30V
267.7V
to
269.7V
1053V
to
1068V
139.8mV
to
143.0mV
1.406V
to
1.422V
13.97V
to
14.31V
267.6V
to
269.8V
************
139.6mV
143.2mV
1.404V
1.424V
13.96V
14.32V
267.4V
270.0V
4. Set digital filter averaging as follows:
A. From normal display, press CONFIG then DCI.
B. Select FILTER, then press ENTER.
C. Select AVERAGING, then press ENTER.
D. Using the cursor and range keys, set the averaging
E. Press EXIT as necessary to return to normal display.
F. If the FILT annunciator is off, press FILTER to en-
CAUTION
Do not apply more than 2A, 250V to the
AMPS input, or the amps protection
fuse will blow.
1. Turn on the Model 2001 and the calibrator, and allow a
one-hour warm-up period before making measurements.
2. Connect the Model 2001 to the calibrator, as shown in
Figure 1-4. Be sure to connect calibrator HI to the
AMPS input, and connect calibrator LO to INPUT LO
as shown.
5. Select the DC current function (DCI) and the 200µA
range on the Model 2001.
6. Set the calibrator output to +190.0000µADC, and allow
the reading to settle.
7. Verify that the Model 2001 reading is within the limits
summarized in Table 1-6.
8. Repeat steps 6 and 7 for the remaining currents listed in
Table 1-6.
9. Repeat the procedure for each of the ranges with negative currents of the same magnitude as those listed in Table 1-6.
3. Restore Model 2001 factory default conditions, as explained in paragraph 1.7.
138.6mV
to
to
to
to
to
144.2mV
1.394V
to
1.434V
13.86V
to
14.42V
********
136.5mV
to
146.3mV
1.373V
to
1.455V
13.65V
to
14.63V
132.2mV
to
150.6mV
1.330V
to
1.498V
13.22V
to
15.06V
parameter to 10 readings, then press ENTER.
able the filter.
127.3mV
to
155.5mV
1.281V
to
1.547V
12.73V
to
15.55V
1-10
Performance Verification
5700A Calibrator (Output DC Current)
19.00000 mADC
PREV
DCV ACV DCI ACI Ω2 Ω4
DISPLAY
NEXT
REL TRIG STORE RECALL
POWER
INFO LOCAL CHAN SCAN CONFIG MENU EXIT ENTER
Model 2001
2001 MULTIMETER
FREQ TEMP
FILTER MATH
SENSE
Ω 4 WIRE
350V
PEAK
INPUTS
F
RANGE
FRONT/REAR
AUTO
RANGE
CAL
Input
INPUT
LO
HI
1100V
PEAK
LO
500V
PEAK
R
2A 250V
AMPS
Amps
Output HI
Output
LO
Note: Use internal Guard (EX GRD LED is off).
Figure 1-4
Connections for DC current verification
Table 1-6
Limits for DC current verification
2001 DCI
range
200µA
Applied DC
current
190.0000µA
Reading limits
(1 year, 18° to 28°C)
189.9000µA
to
190.1000µA
2mA
1.900000mA
1.899200mA
to
1.900800mA
20mA
19.00000mA
18.99200mA
to
19.00800mA
200mA
190.0000mA
189.9010mA
to
190.0990mA
2A
1.900000A
1.898200A
to
1.901800A
NOTES:
1. Repeat procedure for negative currents.
2. Reading limits shown do not include calibrator uncertainty.
1.8.4AC current verification
AC current verification is performed by applying accurate
AC currents at specific frequencies and then verifying that
Model 2001 readings fall within specified limits.
Ground link installed.
Follow the steps below to verify ACI measurement accuracy.
CAUTION
Do not apply more than 2A, 250V to the
AMPS input, or the current protection
fuse will blow.
1. Turn on the Model 2001 and the calibrator, and allow a
one-hour warm-up period before making measurements.
2. Connect the Model 2001 to the calibrator, as shown in
Figure 1-5. Be sure to connect calibrator HI to the
AMPS input, and connect calibrator LO to INPUT LO
as shown.
3. Restore Model 2001 factory default conditions, as explained in paragraph 1.7.
4. Select the AC current function and the 200µA range on
the Model 2001.
5. Set the calibrator output to 190.000µA AC at a frequency of 40Hz, and allow the reading to settle.
6. Verify that the Model 2001 reading is within the limits
for the present current and frequency summarized in Table 1-7.
7. Repeat steps 4 and 5 for each frequency listed in Table
1-7.
8. Repeat steps 4 through 7 for the remaining ranges and
frequencies listed in Table 1-7.
1-11
Performance Verification
5700A Calibrator (Output AC Current)
190.0000 μAAC RMS
PREV
DCV ACV DCI ACI Ω2 Ω4
DISPLAY
NEXT
REL TRIG STORE RECALL
POWER
INFO LOCAL CHAN SCAN CONFIG MENU EXIT ENTER
Model 2001
2001 MULTIMETER
FREQ TEMP
FILTER MATH
SENSE
Ω 4 WIRE
350V
PEAK
INPUTS
F
RANGE
FRONT/REAR
AUTO
RANGE
Figure 1-5
Connections for AC current verification
Table 1-7
Limits for AC current verification
2001 ACI
range
Applied AC
current
200µA190.000µA188.260µA
2mA1.90000mA1.88355mA
20mA19.0000mA18.8355mA
200mA190.000mA188.355mA
2A1.90000A1.88250A
Note: Reading limits shown do not include calibrator uncertainty.
Input
INPUT
LO
HI
1100V
PEAK
LO
500V
PEAK
R
2A 250V
AMPS
CAL
Amps
Output HI
Output
LO
Ground link installed.
Note: Use internal Guard (EX GRD LED is off).
Reading limits (1 year, 18° to 28°C)
40Hz100Hz1kHz10kHz
to
191.740µA
to
1.91645mA
to
19.1645mA
to
191.645mA
to
1.91750A
189.560µA
to
190.440µA
1.89657mA
to
1.90344mA
18.9657mA
to
19.0344mA
189.657mA
to
190.344mA
1.89556A
to
1.90444A
189.210µA
to
190.790µA
1.89742mA
to
1.90258mA
18.9742mA
to
19.0258mA
189.742mA
to
190.258mA
1.89390A
to
1.90610A
189.020µA
to
190.980µA
1.89742mA
to
1.90258mA
18.9742mA
to
19.0258mA
189.685mA
to
190.315mA
1.89105A
to
1.90895A
1-12
Performance Verification
1.8.5Resistance verification
Resistance verification is performed by connecting accurate
resistance values to the instrument and verifying that Model
2001 resistance readings are within stated limits.
Follow the steps below to verify resistance measurement accuracy.
CAUTION
Do not apply more than 1100V peak between INPUT HI and LO or more than
350V peak between SENSE HI and LO,
or instrument damage may occur.
20¾ - 200k¾ range verification
1. Turn on the Model 2001 and the calibrator, and allow a
one-hour warm-up period before making measurements.
2. Set the calibrator for 4-wire resistance (external sense
on).
3. Using shielded 4-wire connections, connect the Model
2001 to the calibrator, as shown in Figure 1-6. Be sure
to connect calibrator HI and LO terminals to the Model
2001 HI and LO terminals (including SENSE HI and
LO) as shown.
4. Restore Model 2001 factory default conditions, as explained in paragraph 1.7.
5. Set operating modes as follows:
A. From normal display, press CONFIG then ¾4.
B. Select FILTER, then press ENTER.
C. Select AVERAGING, then press ENTER.
D. Using the cursor and range keys, set the averaging
parameter to 10 readings, then press ENTER.
E. Select OFFSETCOMP, then press ENTER.
F. Select ON, then press ENTER.
G. Press EXIT to return to normal display.
6. Set the calibrator to output 19.000¾, and allow the reading to settle. Verify that the reading is within the limits
stated in Table 1-8.
NOTE
Resistance values available in the Model
5700A calibrator may be slightly different
than the stated nominal resistance values.
Calculated limits stated in Table 1-8
should be recalculated based on actual calibrator resistance values.
7. Set the calibrator output to 190.000¾, and allow the
reading to settle.
8. Verify that the reading is within the limits stated in Table
1-8. (NOTE: Recalculate limits if calibrator resistance is
not exactly as listed.)
9. Repeat steps 11 and 12 for the 2k¾ through 200k¾ ranges using the values listed in Table 1-8. NOTE: Turn offset compensation off when testing the 200k¾ range (see
step 5).
Table 1-8
Limits for resistance verification (20¾-200M¾ ranges)
2001 ¾ range
20¾
Applied
resistance
19.0000¾
Reading limits
(1 year, 18° to 28°C)
18.99849¾
to
19.00151¾
200¾
190.000¾
189.9880¾
to
190.0120¾
2k¾
1.90000k¾
1.899897k¾
to
1.900103k¾
20k¾
19.0000k¾
18.99897k¾
to
19.00103k¾
200k¾
190.000k¾
189.9820k¾
to
190.0180k¾
2M¾
1.90000M¾
1.899687M¾
to
1.900313M¾
20M¾
19.0000M¾
18.98281M¾
to
19.01719M¾
200M¾
100.000M¾
97.9800M¾
to
102.0200M¾
NOTES:
1. Limits shown do not include calibrator uncertainty and are based on
absolute calibration values shown. Recalculate limits using Model
2001 specifications if calibrator resistance values differ from nominal values shown.
2. Use 4-wire connections and function for 20¾-200k¾ ranges. Use 2wire connections and function for 2M¾-200M¾ ranges.
1-13
Performance Verification
5700A Calibrator (Output 2-Wire Resistance)
Input HI
INPUT
HI
1100V
PEAK
LO
500V
PEAK
R
2A 250V
AMPS
Input
LO
Output HI
Output
LO
+1. 900000 kΩ
PREV
DCV ACV DCI ACI Ω2 Ω4
DISPLAY
NEXT
REL TRIG STORE RECALL
POWER
INFO LOCAL CHAN SCAN CONFIG MENU EXIT ENTER
Model 2001
2001 MULTIMETER
FREQ TEMP
FILTER MATH
SENSE
Ω 4 WIRE
350V
PEAK
INPUTS
F
RANGE
FRONT/REAR
AUTO
RANGE
CAL
Note: Use shielded cable to minimize noise.
Disable calibrator external sense mode.
Use internal Guard (EX GRD LED is off).
Figure 1-6
Connections for resistance verification (20¾-200k¾ ranges)
2M¾ – 200M¾ range verification
1. Connect the DC calibrator and Model 2001 using the 2wire connections shown in Figure 1-7.
2. Set the calibrator to the 2-wire mode (external sense
off).
3. Set operating modes as follows:
A. From normal display, press CONFIG then ¾2.
B. Select FILTER, then press ENTER.
C. Select AVERAGING, then press ENTER.
D. Using the cursor and range keys, set the averaging
parameter to 10 readings, then press ENTER.
E. Press EXIT to return to normal display.
F. If the FILT annunciator is off, press FILTER to en-
able the filter.
4. Select the Model 2001 ¾2 function, and change to the
2M¾ range.
5. Set the calibrator to output 1.90000M¾, and allow the
reading to settle.
6. Verify that the reading is within the limits for the 2M¾
range stated in Table 1-8. (NOTE: Recalculate limits if
actual calibrator resistance differs from value shown.)
7. Repeat steps 4 through 6 for the 20M¾ (output
19.0000M¾) and 200M¾ (output 100.000M¾) ranges.
1G¾ range verification
Ground link installed.
2. Characterize the 1G¾ resistor to within ±10,000ppm or
better using an accurate megohmmeter (see Table 1-1).
Record the characterized value where indicated in Table
1-9. Also, compute the limits based on the value of R using the formula at the bottom of the table.
NOTE
The value of the 1G¾ resistor should not
exceed 1.05G¾.
3. Set operating modes as follows:
A. From normal display, press CONFIG then ¾2.
B. Select FILTER, then press ENTER.
C. Select AVERAGING, then press ENTER.
D. Using the cursor and range keys, set the averaging
parameter to 10 readings, then press ENTER.
E. Press EXIT to return to normal display.
F. If the FILT annunciator is off, press FILTER to en-
able the filter.
4. Select the 2-wire ohms function (¾2) and the 1G¾
range on the Model 2001.
5. Connect the 1G¾ resistor test box (from steps 1 and 2)
to the INPUT HI and LO terminals of the Model 2001.
Allow the reading to settle.
6. Verify that the Model 2001 reading is within the limits
you calculated and recorded in Table 1-9.
1. Mount the 1G¾ resistor and the banana plugs to the test
box, as shown in Figure 1-8. Be sure to mount the banana plugs with the correct spacing. The resistor should
be completely enclosed in and shielded by the metal test
box. The resistor LO lead should be electrically connected to the test box to provide adequate shielding.
1-14
190.0000 kΩ
PREV
DCV ACV DCI ACI Ω2 Ω4
DISPLAY
NEXT
REL TRIG STORE RECALL
POWER
INFO LOCAL CHAN SCAN CONFIG MENU EXIT ENTER
Model 2001
2001 MULTIMETER
FREQ TEMP
FILTER MATH
Sense LO
Performance Verification
Sense LO
Sense HI
Sense HI
SENSE
INPUT
Ω 4 WIRE
HI
350V
PEAK
F
RANGE
FRONT/REAR
AUTO
RANGE
Input HI
1100V
PEAK
LO
500V
PEAK
INPUTS
R
2A 250V
AMPS
CAL
Input
LO
Output HI
Output
LO
5700A Calibrator (Output 4-wire Resistance)
Note: Use shielded cables to minimize noise.
Enable calibrator external sense mode.
Use internal Guard (EX GRD LED is off).
Figure 1-7
Connections for resistance verification (2M¾ - 200M¾ ranges)
Insulated
Plug
HI
0.75"
LO
Banana
Plugs
Non-insulated Plug
Ground link installed.
1GΩ Resistor (Keithley
part # R-289-1G)
Metal
Test Box
Figure 1-8
1G¾ resistor test box construction
Table 1-9
Limits for resistance verification (1G¾ range)
Characterized
resistor (R)Reading limit (1 year, 18° to 28°C)*
____________ G¾_________G¾ to _________G¾
*1 Year limits = R ± (0.04R + 100,000)
Where R = characterized value of 1G¾ resistor.
Note: Resistor must be accurately characterized
before use (see text).
1-15
Performance Verification
1.8.6Frequency accuracy verification
Frequency accuracy verification is performed by connecting
an accurate frequency source to the Model 2001 inputs, and
then verifying that the frequency readings are within stated
limits.
Use the procedure below to verify the frequency measurement accuracy of the Model 2001.
1. Connect the frequency synthesizer to the Model 2001
INPUT terminals, as shown in Figure 1-9.
2. Turn on both instruments, and allow a one-hour warmup period before measurement.
3. Set the synthesizer operating modes as follows:
FREQ: 1Hz
AMPTD: 5V p-p
OFFSET: 0V
MODE: CONT
FCTN: sine wave
4. Restore Model 2001 factory defaults, as explained in
paragraph 1.7.
5. Press FREQ to place the Model 2001 in the frequency
measurement mode.
6. Set maximum signal level to 10V as follows:
A. Press CONFIG then FREQ.
B. Select MAX-SIGNAL-LEVEL, then press ENTER.
C. Select VOLTAGE, then press ENTER.
D. Select 10V, then press ENTER.
E. Press EXIT to return to normal display.
7. Verify that the Model 2001 frequency reading is within
the limits shown in the first line of Table 1-10.
8. Set the synthesizer to each of the frequencies listed in
Table 1-10, and verify that the Model 2001 frequency
reading is within the required limits.
Table 1-10
Frequency verification limits
Synthesizer
frequency
1Hz
10Hz
100Hz
1kHz
10kHz
100kHz
1MHz
10MHz
15MHz
Reading limits
(1 year, 18° to 28°C)
0.9997Hz to 1.0003Hz
9.997Hz to 10.003Hz
99.97Hz to 100.03Hz
0.9997kHz to 1.0003kHz
9.997kHz to 10.003kHz
99.97kHz to 100.03kHz
0.9997MHz to 1.0003MHz
9.997MHz to 10.003MHz
14.995MHz to 15.005MHz
Model 2001
SENSE
Ω 4 WIRE
1.0000 MHz
PREV
DCV ACV DCI ACI Ω2 Ω4 FREQ TEMP
DISPLAY
NEXT
REL TRIG STORE RECALL
POWER
INFO LOCAL CHAN SCAN CONFIG MENU EXIT ENTER
2001 MULTIMETER
FILTER MATH
Figure 1-9
Connections for frequency accuracy verification
350V
PEAK
LO
INPUTS
F
FRONT/REAR
R
CAL
RANGE
AUTO
RANGE
BNC-to-Dual
Banana Plug
Model 3940 Synthesizer
3940 MULTIFUNCTION SYNTHESIZER
Adapter
INPUT
HI
1100V
PEAK
500V
PEAK
2A 250V
AMPS
Main
Function
Output
50Ω BNC Coaxial Cable
1-16
Performance Verification
1.8.7Temperature reading checks
When using thermocouples, the Model 2001 displays temperature by measuring the DC thermocouple voltage, and
then calculating the corresponding temperature. Similarly,
the instrument computes RTD temperature readings by measuring the resistance of the RTD probe and calculating temperature from the resistance value.
Since the instrument computes temperature from DCV and
resistance measurements, verifying the accuracy of those
DCV and resistance measurement functions guarantees the
accuracy of corresponding temperature measurements.
Thus, it is not necessary to perform a comprehensive temperature verification procedure if DCV and resistance verification procedures show the instrument meets its specifications
in those areas. However, those who wish to verify that the
Model 2001 does in fact properly display temperature can
use the following procedure to do so.
Selecting the temperature sensor
Follow the steps below to select the type of temperature sensor:
1. From normal display, press CONFIG then TEMP.
2. Select SENSOR, then press ENTER.
3. Select 4-WIRE RTD or THERMOCOUPLE as desired,
then press ENTER.
4. Select the type of RTD probe or thermocouple you wish
to test, then return to the CONFIG TEMPERATURE
menu.
5. Select UNITS, then press ENTER.
6. Select DEG-C, then press ENTER.
7. Press EXIT as necessary to return to normal display.
8. Press the TEMP key to place the Model 2001 in the temperature display mode. Refer to further information below on how to check thermocouple and RTD probe
readings.
Thermocouple temperature reading checks
To check thermocouple readings, simply apply the appropriate DC voltage listed in Table 1-11 to the Model 2001 INPUT jacks using a precision DC voltage source (such as the
one used to verify DC voltage accuracy in paragraph 1.8.1),
and check the displayed temperature reading. Be sure to use
low-thermal cables for connections between the DC calibrator and the Model 2001 when making these tests.
NOTE
The voltages shown are based on a 0°C
reference junction temperature. Use CONFIG TEMP to set the default reference
junction temperature to 0°C.
Table 1-11
Thermocouple temperature reading checks
Thermo-
couple type
J-4.215mV
K-3.242mV
T-3.089mV
E-4.777mV
R0.054mV
S0.055mV
B0.632mV
*Voltages shown are based on 0°C reference junction temperature. Use
CONFIG-TEMP menu to set default reference junction to 0°C.
Applied DC
voltage*
0mV
1.277mV
5.268mV
42.283mV
0mV
1.000mV
4.095mV
54.125mV
0mV
0.992mV
4.277mV
20.252mV
0mV
1.495mV
6.317mV
75.608mV
0.647mV
4.471mV
20.878mV
0.645mV
4.234mV
18.504mV
1.241mV
4.833mV
13.585mV
Displayed
temperature (°C)
-90.5 to -89.5
-0.5 to +0.5
24.5 to 25.5
99.5 to 100.5
749.5 to 750.5
-90.5 to -89.5
-0.5 to +0.5
24.5 to 25.5
99.5 to 100.5
1349.5 to 1350.5
-90.5 to -89.5
-0.5 to +0.5
24.5 to 25.5
99.5 to 100.5
389.5 to 390.5
-90.6 to -89.4
-0.6 to +0.6
24.4 to 25.6
99.4 to 100.6
989.4 to 990.6
7 to 13
97 to 103
497 to 503
1747 to 1753
7 to 13
97 to 103
497 to 503
1747 to 1753
355 to 365
495 to 505
995 to 1005
1795 to 1805
1-17
Performance Verification
RTD Temperature reading checks
Use a precision decade resistance box (see Table 1-1) to simulate probe resistances at various temperatures (Table 1-12).
Be sure to use 4-wire connections between the decade resistance box and the Model 2001.
Table 1-12
RTD probe temperature reading checks
RTD probe
type
PT385
(∝=0.00385055)
PT3916
(∝=0.00392)
Applied
resistance
64.30¾
100¾
109.73¾
138.51¾
313.71¾
63.68¾
100¾
109.90¾
139.16¾
266.94¾
Displayed
temperature (°C)
-90.08 to -89.92
-0.08 to +0.08
24.92 to 25.08
99.92 to 100.08
599.86 to 600.14
-90.08 to -89.92
-0.08 to +0.08
24.92 to 25.08
99.92 to 100.08
449.86 to 450.14
1-18
2
Calibration
2.1Introduction
This section gives detailed procedures for calibrating the
Model 2001. There are three types of calibration procedures:
• Comprehensive calibration
• AC self-calibration
• Low-level calibration
Comprehensive calibration requires accurate calibration
equipment to supply precise DC voltages and resistance values. AC self-calibration requires no external equipment and
can be performed at any time by the operator. Low-level calibration is normally performed only at the factory where the
instrument is manufactured and is not usually required in the
field.
NOTE
Low-level calibration is required in the
field only if the Model 2001 has been repaired, or if the other calibration procedures cannot bring the instrument within
stated specifications.
2.5Calibration lock: Explains how to unlock calibration
with the CAL switch.
2.6IEEE-488 bus calibration commands and program:
Summarizes bus commands used for calibration, lists
a simple calibration program, and also discusses other
important aspects of calibrating the instrument over
the bus.
2.7Calibration errors: Details front panel error messages
that might occur during calibration and also explains
how to check for errors over the bus.
2.8Comprehensive calibration: Covers comprehensive
(user) calibration from the front panel and over the
IEEE-488 bus.
2.9AC self-calibration: Discusses the AC user calibration process, both from the front panel and over the
IEEE-488 bus.
2.10 Low-level calibration: Explains how to perform the
low-level calibration procedure, which is normally required only at the factory.
Section 2 includes the following information:
2.2Environmental conditions: States the temperature
and humidity limits for calibration.
2.3Warm-up period: Discusses the length of time the
Model 2001 should be allowed to warm up before calibration.
2.4Line power: States the power line voltage limits when
calibrating the unit.
2.2Environmental conditions
Calibration procedures should be performed at an ambient
temperature of 23°± 1°C, and at a relative humidity of less
than 80% unless otherwise noted.
2-1
Loading...
+ 77 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.