RANGE CHANGE3: 50/s.
FUNCTION CHANGE
AUTORANGE TIME
ASCII READINGS TO RS-232 (19.2K BAUD): 55/s.
MAX. INTERNAL TRIGGER RATE: 2000/s.
MAX. EXTERNAL TRIGGER RATE: 500/s.
DC GENERAL
LINEARITY OF 10VDC RANGE: ±(2ppm of reading + 1ppm of range).
DCV, Ω, TEMPERATURE, CONTINUITY, DIODE TEST INPUT
PROTECTION: 1000V, all ranges.
MAXIMUM 4WΩ LEAD RESISTANCE: 10% of range per lead for 100Ω
and 1kΩ ranges; 1kΩ per lead for all other ranges.
DC CURRENT INPUT PROTECTION: 3A, 250V fuse.
SHUNT RESISTOR: 0.1Ω for 3A, 1A and 100mA ranges. 10Ω for 10mA
range.
CONTINUITY THRESHOLD: Adjustable 1Ω to 1000Ω.
AUTOZERO OFF ERROR: Add ±(2ppm of range error + 5µV) for <10 min-
utes and ±1°C change.
OVERRANGE: 120% of range except on 1000V, 3A and Diode.
3
: 45/s.
3,10
: <30 ms.
3, 4
2
3, 7
1
⁄
2
3, 5
2
3, 5
1
5
⁄
2
5
51⁄
2
5
1
5
⁄
2
5
1
4
⁄
2
2, 6
2
8
510
301
501
2700.1
5000.1
10000.04
20000.01
SPEED AND NOISE REJECTION
RATEREADINGS/S DIGITS10V RANGENMRR12CMRR
10 PLC561⁄
1 PLC506
0.1 PLC50051⁄
0.01 PLC200041⁄
DC Notes
1
Add the following to “ppm of range” uncertainty:1V and 100V, 2ppm; 100mV, 15ppm; 100Ω, 15ppm; <1MΩ,
2ppm; 10mA and 1A, 10ppm; 100mA, 40ppm.
2
Speeds are for 60 Hz operation using factory default operating conditions (*RST). Autorange off, Display off,
Trigger delay = 0.
3
Speeds include measurement and binary data transfer out the GPIB.
4
Auto zero off.
5
Sample count = 1024, auto zero off.
6
Auto zero off, NPLC = 0.01.
7
Ohms = 24 readings/second.
8
1 PLC = 16.67ms @ 60Hz, 20ms @ 50Hz/400Hz. The frequency is automatically determined at power up.
9
For signal levels >500V, add 0.02ppm/V uncertainty for the portion exceeding 500V.
10
Add 120ms for ohms.
11
Must have 10% matching of lead resistance in Input HI and LO.
12
For line frequency ±0.1%.
13
For 1kΩ unbalance in LO lead.
14
Relative to calibration accuracy.
15
Specifications are for 4-wire ohms. For 2-wire ohms, add 1Ω additional uncertainty.
16
For rear inputs, add the following to Temperature Coefficient “ppm of reading” uncertainty: 10MΩ 70ppm,
100MΩ 385ppm. Operating environment specified for 0° to 50°C and 50% RH at 35°C.
RMS NOISE
2
< 1.5 µV60 dB140 dB
1
⁄
2
2
2
< 4 µV60 dB140 dB
< 22 µV—80 dB
< 150 µV—80 dB
13
Rev. F
HW 8/22/02
Page 2
2000 61⁄2-Digit Multimeter
TRUE RMS AC VOLTAGE AND CURRENT CHARACTERISTICS
ACCURACY1: ±(% of reading + % of range), 23°C ±5 °C
FUNCTION/RANGE CHANGE6: 4/s.
AUTORANGE TIME: <3 s.
ASCII READINGS TO RS-232 (19.2k BAUD)
MAX. INTERNAL TRIGGER RATE
MAX. EXTERNAL TRIGGER RATE
4
: 300/s.
4
: 300/s.
4
: 50/s.
AC GENERAL
INPUT IMPEDANCE: 1MΩ ±2% paralleled by <100pF.
ACV INPUT PROTECTION: 1000Vp.
MAXIMUM DCV: 400V on any ACV range.
ACI INPUT PROTECTION: 3A, 250V fuse.
BURDEN VOLTAGE: 1A Range: <0.3V rms. 3A Range: <1V rms.
SHUNT RESISTOR: 0.1Ω on all ACI ranges.
AC CMRR: >70dB with 1kΩ in LO lead.
MAXIMUM CREST FACTOR: 5 at full scale.
VOLT HERTZ PRODUCT: ≤8 × 10
OVERRANGE: 120% of range except on 750V and 3A ranges.
AC Notes
1
Specifications are for SLOW rate and sinewave inputs >5% of range.
2
Speeds are for 60 Hz operation using factory default operating conditions (*RST).
Auto zero off, Auto range off, Display off, includes measurement and binary data
transfer out the GPIB.
3
0.01% of step settling error. Trigger delay = 400ms.
4
Trigger delay = 0.
5
DETector:BANDwidth 300, NPLC = 0.01.
6
Maximum useful limit with trigger delay = 175ms.
7
Applies to non-sinewaves >5Hz and <500Hz. (Guaranteed by design for Crest
Factors >4.3)
8
Applies to 0°–18°C and 28°–50°C.
9
For signal levels > 2.2A, add additional 0.4% to “of reading” uncertainty.
7
V·Hz.
Rev. F
HW 8/22/02
Page 3
2000 61⁄2-Digit Multimeter
FREQUENCY AND PERIOD CHARACTERISTICS
ACVFREQUENCY PERIODGATE±(ppm of90 DAY/1 YEAR
RANGERANGERANGETIMEreading)±(% of reading)
100 mV3 Hz333 ms1 s
tototo(SLOW)0.30.01
750 V500 kHz2 µs
Frequency Notes
1
Specifications are for squarewave inputs >10% of ACV range, except 100mV range. On 100mV range frequency
must be >10Hz if voltage is <20mV.