197A 5
1
2-Digit Portable Multimeter
1µV, 1m, 1nA sensitivity
TRMS AC measurements to
100kHz
100-point data logger, relative
reference, dB
Backlit display
ACCESSORIES AVAILABLE
1681 Clip-On Test Lead Set 1754 Safety Univ. Test Lead Kit 5806 Kelvin Clip Leads 1978 Rechargeable Battery Pack
Battery option
DC VOLTS
Ordering Information
Measures voltage, current, and resistance economically
197A Autoranging Microvolt
DMM
197A/1978
Autoranging Microvolt DMM with Recharge­able Battery Pack
RANGE LUTION RESISTANCE 18°–28°C
200 mV 1 µV >1 G 0.016 + 3
200 V 1 mV 10 M 0.015 + 2
1000 V 10 mV 10 M 0.015 + 2
RESO- INPUT 1 Year,
2 V 10 µV >1 G 0.011 + 2
20 V 100 µV 11M 0.015 + 3
The Model 197A is a lightweight, portable DMM with a battery pack option. It combines 5
1
2-digit resolution (on a 220,000-count display)
with 1µV, 1nA, 1msensitivity. Basic DCV accuracy is 0.011%.
Extended Performance
Measure voltage over nine decades from 1µV to 1000V. Input resistance on the two lowest voltage ranges exceeds 1G. TRMS AC measure­ments can be made to 100kHz. Make 2- or 4-terminal ohms measure­ments over 11 decades from 1mto 220M. Six current ranges meas­ure from 1nA to 10A.
Battery Option
The Model 1978 battery pack option enables data to be collected in locations where AC power is not available or where isolation from line voltage is required for safety.
GENERAL
DISPLAY: Backlit ±220,000-count LCD, 0.45 in height; polari-
ty, function, range, and status indication.
RELATIVE: Pushbutton allows zeroing of on range readings.
Allows readings to be made with respect to baseline value. Front panel annunciator indicates REL mode.
DATA LOGGER and MIN/MAX: 100 reading storage capacity; records
data at one of six selectable rates from 3 readings/second to 1 read­ing/hour or by manual triggering. Also detects and stores maximum and minimum readings continuously while in data logger mode.
CONVERSION RATE: 3 readings/second. OVERRANGE INDICATION: “OL” displayed.
ACCURACY
±(%rdg + counts)
ENVIRONMENT: Operating: 0°–50°C; <80% relative
humidity up to 35°C; linearly derate 3% RH/°C, 35°–50°C.
Storage: –40° to +70°C. WARM–UP: 1 hour to rated accuracy. POWER: 105–125V or 210–250V (external switch selected),
90–110V available; 50–60Hz, 12VA. Removable power cord.
Optional 5 hour battery pack, Model 1978. DIMENSIONS, WEIGHT: 89mm high × 235mm wide ×
275mm deep (3
(3 lb, 14 oz).
1
2 in × 91⁄4 in × 103⁄4 in). Net weight 1.8kg
Extended warranty, service, and calibration contracts are available
Accessories Supplied
Instruction Manual and Model 1751 Safety Test Leads
1.888.KEITHLEY (U.S. only)
SWITCHING SYSTEMS
DMMS, DATA ACQUISITION/
www.keithley.com
TRMS AC VOLTS
.
2V–200 V 1.00 + 100 0.35 + 100 0.6 + 200 1.5 + 250 3 + 400
OHMS
ACCURACY (1 Year)
RANGE 20Hz–50Hz 50Hz–10kHz 10kHz–20kHz 20kHz–50kHz 50kHz–100kHz
750 V 1.25 + 100 0.75 + 100 1.0 + 200 1.8 + 250 3 + 400
RANGE RESOLUTION I–SHORT UNKNOWN 22°–24°C 18°–28°C
200 1m 2 mA 0.5 V 0.01 + 2 0.018 + 3
2k 10 mΩ 2 mA 4.0 V 0.01 + 2 0.018 + 2
20 k 100 m 400 µA 4.0 V 0.014 + 2 0.026 + 2
200 k 1 40 µA 4.0 V 0.014 + 2 0.026 + 2
2M 10 Ω 4 µA 4.0 V 0.02 + 2 0.035 + 2
20 M 100 400 nA 4.0 V 0.10 + 2 0.12 + 2
200 M 10 k 400 nA 5.0 V 2.00 + 1 2.00 + 1
A GREATER MEASURE OF CONFIDENCE
NOMINAL ACROSS 24 Hr., 1 Year,
18°–28°C ±(%rdg + counts)
OUTPUT ACCURACY
MAX. V ±(%rdg+counts)