Keithley SourceMeter 2611, SourceMeter 2602, SourceMeter 2612, SourceMeter 2601 User Manual

www.keithley.com
Series 2600 System SourceMeter
®
User’s Manual
2600S-900-01 Rev. A / May 2006
A GREATER MEASURE OF CONFIDENCE

WARRANTY

Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship for a period of 1 year from date of shipment.
Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables, rechargeable batteries, diskettes, and documentation.
During the warranty period, we will, at our option, either repair or replace any product that proves to be defective.
To exercise this warranty, write or call your local Keithley representative, or contact Keithley headquarters in Cleveland, Ohio. You will be given prompt assistance and return instructions. Send the product, transportation prepaid, to the indicated service facility. Repairs will be made and the product returned, transportation prepaid. Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days.
LIMITATION OF WARRANTY
This warranty does not apply to defects resulting from product modification without Keithley’s express written consent, or misuse of any product or part. This warranty also does not apply to fuses, software, non-recharge able batteries, damage from battery leakage, or problems arising from normal wear or failure to follow instruc­tions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUD­ING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE. THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OF ITS INSTRUMENTS AND SOFTWARE EVEN IF KEITHLEY INSTRUMENTS, INC., HAS BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COSTS OF REMOVAL AND INSTALLATION, LOSSES SUSTAINED AS THE RESULT OF INJURY TO ANY PERSON, OR DAMAGE TO PROPERTY.
-
A G R E A T E R M E A S U R E O F C O N F I D E N C E
Keithley Instruments, Inc.
Corporate Headquarters • 28775 Aurora Road • Cleveland, Ohio 44139
440-248-0400 • Fax: 440-248-6168 • 1-888-KEITHLEY (534-8453) • www.keithley.com
12/04
Series 2600 System
®
SourceMeter
Instruments
User’s Manual
©2006, Keithley Instruments, Inc.
All rights reserved.
Cleveland, Ohio, U.S.A.
Document Number: 2600S-900-01 Rev. A

Manual Print History

The print history shown below lists the printing dates of all Revisions and Addenda creat­ed for this manual. The Revision Level letter increases alphabetically as the manual under­goes subsequent updates. Addenda, which are released between Revisions, contain important change information that the user should incorporate immediately into the manual. Addenda are numbered sequentially. When a new Revision is created, all Addenda associ ated with the previous Revision of the manual are incorporated into the new Revision of the manual. Each new Revision includes a revised copy of this print history page.
Revision A (Document Number 2600S-900-01).................................................................. May 2006
-
All Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc. Other brand names are trademarks or registered trademarks of their respective holders.

Safety Precautions

The following safety precautions should be observed before using this product and any associated instrumentation. Although some instruments and accessories would normally be used with non-hazardous voltages, there are situations where hazardous conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before us­ing the product. Refer to the manual for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product may be impaired.
The types of product users are:
Responsible body is the individual or group responsible for the use and maintenance of equipment, for ensuring that the equipment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained.
Operators use the product for its intended function. They must be trained in electrical safety procedures and proper use of the instrument. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product to keep it operating properly, for example, setting the line voltage or replacing consumable materials. Maintenance procedures are described in the manual. The procedures explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel are trained to work on live circuits, and perform safe installations and repairs of products. Only properly trained service personnel may perform installation and service procedures.
Keithley products are designed for use with electrical signals that are rated Measurement Category I and Measurement Category II, as described in the International Electrotechnical Commission (IEC) Standard IEC 60664. Most measurement, control, and data I/O signals are Measurement Category I and must not be directly connected to mains voltage or to voltage sources with high transient over-voltages. Measurement Category II connections require protection for high transient over-voltages often associated with local AC mains connections. Assume all measurement, control, and data I/O connections are for connection to Category I sources unless otherwise marked or described in the Manual.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30V RMS,
42.4V peak, or 60VDC are present. A good safety practice is to expect that hazardous voltage is present in any unknown
circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000 volts, no conductive part of the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance limited sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective devices to limit fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connected to a properly grounded power receptacle. Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input power disconnect device must be provided, in close proximity to the equipment and within easy reach of the operator.
5/03
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being measured.
The instrument and accessories must be used in accordance with its specifications and operating instructions or the safety of the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating information, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with same type and rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground connections.
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use of a lid interlock.
If a screw is present, connect it to safety earth ground using the wire recommended in the user documentation.
The ! symbol on an instrument indicates that the user should refer to the operating instructions located in the manual.
The symbol on an instrument shows that it can source or measure 1000 volts or more, including the combined effect of normal and common mode voltages. Use standard safety precautions to avoid personal contact with these voltages.
The symbol indicates a connection terminal to the equipment frame.
The WARNING heading in a manual explains dangers that might result in personal injury or death. Always read the associated information very carefully before performing the indicated procedure.
The CAUTION heading in a manual explains hazards that could damage the instrument. Such damage may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
To maintain protection from electric shock and fire, replacement components in mains circuits, including the power transformer, test leads, and input jacks, must be purchased from Keithley Instruments. Standard fuses, with applicable national safety approvals, may be used if the rating and type are the same. Other components that are not safety related may be purchased from other suppliers as long as they are equivalent to the original component. (Note that selected parts should be purchased only through Keithley Instruments to maintain accuracy and functionality of the product.) If you are unsure about the applicability of a replacement component, call a Keithley Instruments office for information.
To clean an instrument, use a damp cloth or mild, water based cleaner. Clean the exterior of the instrument only. Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with no case or chassis (e.g., data acquisition board for installation into a computer) should never require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the factory for proper cleaning/servicing.

Table of Contents

Section Title Page
1 Front Panel Operation
Safety symbols and terms .......................................................................... 1-2
Front and rear panel familiarization ............................................................ 1-2
Front panel summaries ........................................................................ 1-2
Rear panel summaries ......................................................................... 1-6
What are the source-measure capabilities? ............................................... 1-8
Models 2601 and 2602 ........................................................................ 1-8
Models 2611 and 2612 ........................................................................ 1-8
How do I power-up the instrument? ........................................................... 1-9
How do I make measurements? .............................................................. 1-10
How do I use the buffer? .......................................................................... 1-15
2 Remote Operation
How do I use the remote interface? ........................................................... 2-2
How do I use Test Script Builder? .............................................................. 2-4
How do I use TSB to make measurements? .............................................. 2-9
How do I use other programs? ................................................................. 2-12
Using LabVIEW ................................................................................. 2-12
Using Visual Basic ............................................................................. 2-14
3 Test Script Processor Interaction
What is a script? ......................................................................................... 3-2
Factory scripts ..................................................................................... 3-2
User scripts .......................................................................................... 3-2
How do I run a script from the front panel? ................................................ 3-2
How do I interact with scripts using Test Script Builder? ............................ 3-3
Running a factory script ....................................................................... 3-4
Modifying a factory script ..................................................................... 3-6
2600S-900-01 Rev. A / May 2006 i
Table of Contents Series 2600 System SourceMeters User’s Manual
Running the user script ...................................................................... 3-12
Deleting a user script and user tests ................................................ 3-13
How do I use other programs? ................................................................. 3-14
Using LabVIEW ................................................................................. 3-14
Using Visual Basic ............................................................................. 3-16
4 Controlling Multiple SourceMeters (TSP-Link)
How do I set up the TSP-Link system? ...................................................... 4-2
How do I use the expanded system? ......................................................... 4-4
Accessing resources of TSP-Link nodes ............................................. 4-4
Running scripts in a TSP-Link system ................................................. 4-5
Appendix Title Page
A Specifications
.................................................................................................................. A-2
B Frequently Asked Questions
How do I optimize performance? .............................................................. B-2
Setting speed ...................................................................................... B-2
Disabling auto zero to increase speed ............................................... B-2
How do I use the Digital I/O port? ............................................................. B-3
Digital I/O port terminals ..................................................................... B-3
Controlling the Digital I/O port ............................................................. B-3
How do I trigger other instruments? .......................................................... B-5
Triggering a scanner ........................................................................... B-5
Programming triggering ...................................................................... B-5
How do I generate a service request? ...................................................... B-6
Setting up a service request ............................................................... B-6
Service request programming example .............................................. B-6
Polling for SRQs ................................................................................. B-6
How do I store measurements in non-volatile memory? ........................... B-7
Front panel operation .......................................................................... B-7
Remote programming ......................................................................... B-7
How do I stack channels to output higher voltage? ................................... B-9
How do I parallel channels to output higher current? ............................... B-11
How do I make contact check measurements? ...................................... B-12
Contact check connections ............................................................... B-12
Contact check programming example .............................................. B-13
2 2600-900-01 Rev. A / May 2006

List of Illustrations

Section Figure Title Page
1 Front Panel Operation
Figure 1-1 Series 2600 front panels 1-3
Figure 1-2 Series 2600 rear panels ................................................................. 1-6
Figure 1-3 DUT connections to 10kΩ resistor ............................................... 1-12
Figure 1-4 Interlock circuit ............................................................................. 1-13
Figure 1-5 Display modes.............................................................................. 1-14
Figure 1-6 Buffer display format .................................................................... 1-17
2 Remote Operation
Figure 2-1 GPIB cable ..................................................................................... 2-2
Figure 2-2 RS-232 cable ................................................................................. 2-2
Figure 2-3 Test Script Builder initial startup screen ......................................... 2-6
Figure 2-4 Instrument console control icons.................................................... 2-7
Figure 2-5 Select Instrument Resource dialog box ......................................... 2-8
Figure 2-6 Source-measure command sequence in console window .......... 2-11
Figure 2-7 LabVIEW source-measure example block diagram ..................... 2-13
Figure 2-8 Visual Basic example user interface ............................................ 2-14
Figure 2-9 Example program test results....................................................... 2-17
2600-900-01 Rev. A / May 2006 3
LIst of Illustrations Series 2600 System SourceMeters User’s Manual
3 Test Script Processor Interaction
Figure 3-1 Pulse-measure cycle for the PulseVMeasureI function.................. 3-3
Figure 3-2 Importing a factory script project from the Series 2600.................. 3-7
Figure 3-3 KIGeneral project imported into the Test Script Builder ................. 3-9
Figure 3-4 Run configuration example - Main tab ......................................... 3-11
Figure 3-5 Run configuration example - Script Attributes tab ....................... 3-12
Figure 3-6 LabVIEW source step example.................................................... 3-15
Figure 3-7 GUI after loading the non-function script (GPIB).......................... 3-18
Figure 3-8 GUI after running the non-function script (GPIB) ......................... 3-19
Figure 3-9 GUI after loading and running the function script (GPIB)............. 3-21
Figure 3-10 GUI after calling the function (GPIB) ............................................ 3-23
4 Controlling Multiple SourceMeters (TSP-Link)
Figure 4-1 TSP-Link connections .................................................................... 4-2
Appendix Figure Title Page
B Frequently Asked Questions
Figure B-1 Digital I/O port terminals................................................................ B-3
Figure B-2 Triggering a scanner ..................................................................... B-5
Figure B-3 Stacking channels for higher voltage .......................................... B-10
Figure B-4 Connecting channels in parallel for higher current....................... B-11
Figure B-5 Contact check connections ......................................................... B-12
4 2600-900-01 Rev. A / May 2006

List of Tables

Section Table Title Page
1 Front Panel Operation
Table 1-1 Model 2601 and 2602 source-measure capabilities....................... 1-8
Table 1-2 Model 2611 and 2612 source-measure capabilities ....................... 1-8
3 Test Script Processor Interaction
4 Controlling Multiple SourceMeters (TSP-Link)
Table 4-1 Assigning a node number to an instrument from the front panel.... 4-3
Table 4-2 Resetting the TSP-Link from the front panel .................................. 4-3
Appendix Table Title Page
B Frequently Asked Questions
Table B-1 Commands for basic I/O port ........................................................ B-4
2600-900-01 Rev. A / May 2006 5
List of Tables Series 2600 System SourceMeters User’s Manual
This page left blank intentionally.
6 2600-900-01 Rev. A / May 2006
In this section:
Top ic
Safety symbols and terms 1-2
Front and rear panel familiarization 1-2
Front panel overview 1-2
Rear panel overview 1-6
What are the source-measure capabilities? 1-8
Models 2601 and 2602 1-8
Models 2611 and 2612 1-8
How do I power-up the instrument? 1-8
Connect to line power 1-9
Turn on power 1-9
Set line frequency 1-9
How do I make measurements? 1-10
Connect the DUT 1-10
Select source and set source level 1-10
Set compliance limit 1-11
Select measurement function and range 1-11
Turn output on 1-11
Make measurements 1-11
Turn output off 1-12
How do I use the buffer? 1-15
Connect the DUT 1-15
Section 1
Front Panel Operation
Page
Return to In this section: 2600S-900-01 Rev. A / May 2006
1-2 Front Panel Operation Series 2600 System SourceMeters User’s Manual
Set up source and measure functions 1-15
Configure the buffer 1-16
Turn on the output 1-16
Store readings 1-16
Turn off the output 1-16
Recall readings 1-16

Safety symbols and terms

The following symbols and terms may be found on the instrument or used in this manual:
!
The operating instructions located in the manual.
The symbol on the instrument shows that high voltage may be present on the terminal(s). Use standard safety precautions to avoid personal contact with these voltages.
symbol on an instrument indicates that the user should refer to the
The symbol on an instrument shows that the surface may be hot. Avoid personal contact to prevent burns.
The WARNING heading used in this manual explains dangers that might result in personal injury or death. Always read the associated information very carefully before performing the indicated procedure.
The CAUTION heading used in this manual explains hazards that could damage the instrument. Such damage may invalidate the warranty.

Front and rear panel familiarization

Front panel overview
The front panels of the Series 2600 are shown in Figure 1-1 and Figure 1-2. Sum- maries of the front panel controls follow these figures.
2600S-900-01 Rev. A / May 2006 Return to In this section:
Series 2600 System SourceMeters User’s Manual Front Panel Operation 1-3
Figure 1-1
Series 2600 front panels
Model 2601 and Model 2611
SYSTEM SourceMeter
KEITHLEY SourceMeter
89
7
DISPLAY
CONFIG
POWER
SRC LIMITMEAS
5
4
SPEED
DIGITS
1
2
LOAD RUN STORE RECALL
6
REL
3
1 2
SYSTEM SourceMeter
KEITHLEY SourceMeter
CHANNEL A CHANNEL B
89
7
DISPLAY
CONFIG
POWER
SRC LIMITMEAS
5
4
SPEED
DIGITS
1
2
LOAD RUN STORE RECALL
6
REL
3
®
+ / -
MODE
0
FILTER
0000
TRIG MENU
LOCAL
EXIT ENTER
Model 2602 and Model 2612
®
+ / -
MODE
0
FILTER
0000
SRC LIMITMEAS
SPEED
DIGITS
TRIG MENU
MODE
REL
FILTER
LOCAL
EXIT ENTER
RANGE
AUTO
3
RANGE
AUT O
RANGE
P
R
CURSOR
U
S
H
T
O
E
D
I
T
/
E
N
T
E
R
E
T
N
E
/
T
I
D
E
O
T
H
S
U
P
OUTPUT
ON/OFF
4
P
R
CURSOR
OUTPUT
U
S
H
T
T
/
E
N
T
E
R
CHAN B
O
E
D
I
E
T
N
E
/
T
I
D
E
O
T
H
S
U
P
CHAN A
ON/OFF ON/OFF
5
1 2
NOTE The Models 2601 and 2611 have one SourceMeter channel
(Channel A) and the Models 2602 and 2612 have two SourceMeter channels (Channel A and Channel B).
3
4
5
Return to In this section: 2600S-900-01 Rev. A / May 2006
1-4 Front Panel Operation Series 2600 System SourceMeters User’s Manual
1. Special keys and power switch:
DISPLAY Toggles between the various source-measure displays and the user
message mode. Selects Model 2602/2612 single or dual-channel
display. CONFIG Used to configure a function or operation. POWER Power switch – In position turns SourceMeter on (I), out position
turns SourceMeter off (O). Number Keys The Number Keys (0-9, +/-, 0000) allow direct numeric entry in the
EDIT mode.
2. Source measure setup, performance control and special operation:
Top Row – Source measure setup Models 2601/2611 and Models 2602/2612:
SRC Channel A – Selects the source function (V or A) and places cursor
in the source field for editing. MEAS Channel A – Cycles through measure functions (V, A, Ω or W). LIMIT Channel A – Places the cursor in the compliance limit field for editing. MODE Channel A – Directly chooses the measurement function (V, A, Ω or W).
Top Row – Source measure setup Models 2602/2612 only:
SRC Channel B – Selects the source function (V or A) and places cursor
in the source field. MEAS Channel B – Cycles through measure functions (V,A, Ω or W). LIMIT Channel B – Places the cursor in the compliance limit field for editing. MODE Channel B – Directly chooses the measurement function (V, A, Ω or W).
Middle Row – Source measure setup Model 2601/2611 and Models 2602/2612:
DIGITS Channel A – Changes resolution display to 4-1/2, 5-1/2, or 6-1/2
digits. SPEED Channel A – Sets the measurement speed by controlling the A/D
converter measurement aperture. REL Channel A – Controls relative, which allows a baseline value to be
subtracted from a reading. FILTER Channel A – Controls the digital filter, which can be used to reduce
reading noise.
Middle Row – Source measure setup Models 2602/2612 only:
DIGITS Channel B – Changes resolution display to 4-1/2, 5-1/2, or 6-1/2 digits. SPEED Channel B – Sets the measurement speed by controlling the A/D
converter measurement aperture. REL Channel B – Controls relative, which allows a baseline value to be
subtracted from a reading. FILTER Channel B – Controls the digital filter, which can be used to reduce
reading noise.
2600S-900-01 Rev. A / May 2006 Return to In this section:
Series 2600 System SourceMeters User’s Manual Front Panel Operation 1-5
Bottom Row – Source measure setup Models 2601/2611 2602/2612:
LOAD Loads factory or user-defined scripts for execution. RUN Runs last selected factory or user-defined scripts. STORE Stores readings, source values, and timestamp values in one of two
internal buffers for later recall.
RECALL Recalls stored readings, source values, and timestamp values from
either of the two buffers. TRIG Triggers readings. MENU Accesses the Main Menu for saving and recalling setups, selecting
remote interface, line frequency, self-tests, serial number and beeper
control. EXIT Cancels selection, and backs out of menu structure. Used as a
LOCAL key to take the unit out of remote. ENTER Accepts selection, moves to next choice or exits menu.
3. Range keys:
and Selects the next higher or lower source or measure range.
AUTO Enables or disables source or measure auto range.
4. Output control and LED status indicator:
OUTPUT ON/OFFTurns source output on or off. LED indicator Lights up when output is on.
5. Rotary Knob and CURSOR keys:
When in source edit, use CURSOR keys for cursor control and then turn the Rotary Knob to change a source or compliance value. The Rotary Knob can also be used to enable or disable the source edit mode.
When in a menu, use the CURSOR keys or Rotary Knob for menu item cursor control. When displaying a menu value, use the CURSOR keys for cursor control and turn the Rotary Knob to change the value. Pressing the Rotary Knob opens a menu item, or selects a menu option or value.
6. Display annunciators (not shown):
EDIT Unit is in the source editing mode. ERR Questionable reading or invalid cal step. REM Unit in remote mode. TALK Unit addressed to talk. LSTN Unit addressed to listen. SRQ Service request. REL Relative mode enabled. FILT Analog filter or Averaging filter is enabled. AUTO Auto source or measure range selected. ARM Unit armed and ready to run. TRIG External triggering selected. * (asterisk) Readings being stored in buffer.
Return to In this section: 2600S-900-01 Rev. A / May 2006
1-6 Front Panel Operation Series 2600 System SourceMeters User’s Manual
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
Rear panel overview
The rear panels of the Series 2600 are is shown in Figure 1-2. Summaries of the rear panel components follow Figure 1-2.
Figure 1-2
Series 2600 rear panel
s
Model 2601/2611
U
L
C
MADE IN
U.S.A.
DIGITAL I/O
2
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
LISTED
SourceMeter
4ZA4
US
!
RS-232
TSP-Link
LO LO HIHI GGGG
R
1
CHANNEL A
CAT I
LINE FUSE
SLOWBLOW
3.15A, 250V
!
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
SS
!
3
45 7 8
1
S
LO
G
G
G
HI
G
S
HI
CAT I CHANNEL B
DIGITAL I/O
2
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
LO
!
6
Model 2602/2612
U
L
C
US
LISTED
SourceMeter
4ZA4
!
RS-232
TSP-Link
MADE IN
U.S.A.
9
1
CHANNEL A
CAT I
LINE FUSE
SLOWBLOW
3.15A, 250V
!
SS
LO LO HIHI GGGG
!
R
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
3
45 7 8
2600S-900-01 Rev. A / May 2006 Return to In this section:
6
9
Series 2600 System SourceMeters User’s Manual Front Panel Operation 1-7
1. CHANNEL A and CHANNEL B (Channel B on Models 2602/2612 only)
Input/output connections for source, sense, and guard.
2. DIGITAL I/O
Female DB-25 connector. Fourteen pins for digital input or output, one pin for Output
Enable. Use a cable equipped with a male DB-25 connector (Keithley Instruments part number CA-126-1CA).
3. IEEE-488
Connector for IEEE-488 (GPIB) operation. Use a shielded cable, such as the Model
7007-1 or Model 7007-2.
4. Cooling exhaust vent
Exhaust vent for internal cooling fan. Keep vent free of obstructions to prevent
overheating.
5. Chassis ground
Ground screw for connections to chassis ground.
6. Low noise chassis ground
Ground jack for connecting Output HI or LO to chassis.
7. RS-232
Female DB-9 connector. For RS-232 operation, use a straight-through (not null
modem) DB-9 shielded cable for connection to the PC (Keithley Instruments Model 7009-
5).
8. TSP-Link
Expansion interface that allows a Series 2600 and other TSP-enabled instruments to
trigger and communicate with each other. Use a category 5e or higher LAN crossover cable (Keithley Instruments part number CA-180-3A).
9. Power module
Contains the AC line receptacle and power line fuse. The instrument can operate on
line voltages of 100V to 240VAC at line frequencies of 50Hz or 60Hz. Refer to Section 17 of the Series 2600 Reference Manual for line fuse replacement instructions.
Return to In this section: 2600S-900-01 Rev. A / May 2006
1-8 Front Panel Operation Series 2600 System SourceMeters User’s Manual

What are the source-measure capabilities?

Reference Refer to the specifications in Appendix A of this manual as well as
Sections 4 and 8 of the Series 2600 Reference Manual for more detailed information.
Models 2601 and 2602
Basic DC source-measure capabilities for the Models 2601 and 2602 are listed in
Table 1-1.
Table 1-1
Models 2601 and 2602 source-measure capabilities
Function Capabilities
Source ±DC voltage Source ±DC current
Measure ±DC voltage Measure ±DC current
Models 2611 and 2612
Basic DC source-measure capabilities for the Models 2611 and 2612 are listed in
Table 1-2.
Table 1-2
Models 2611 and 2612 source-measure capabilities
Function Capabilities
Source ±DC voltage Source ±DC current
Measure ±DC voltage Measure ±DC current
1μV to 40.4V 1pA to 3.03A
1μV to 40.8V 1pA to 3.06A
1μV to 202V 1pA to 1.515A
1μV to 204V 1pA to 1.53
1
1
1. 10A maximum available in pulse mode only.
2600S-900-01 Rev. A / May 2006 Return to In this section:
Series 2600 System SourceMeters User’s Manual Front Panel Operation 1-9

How do I power-up the instrument?

Reference Refer to “Power-up” in Section 1 of the Series 2600 Reference Man-
ual for more information on line power.
Step 1: Connect to line power
CAUTION Operating the instrument on an incorrect line voltage may
cause damage to the instrument, possibly voiding the warranty.
1. Before plugging in the power cord, make sure that the front panel power switch is in the off (O) position.
2. Connect the female end of the supplied power cord to the AC receptacle on the rear panel. Connect the other end of the power cord to a grounded AC outlet.
WARNING Failure to use a grounded outlet may result in personal injury
or death due to electric shock.
Step 2: Turn on power
Turn on the instrument by pressing the front panel power switch to the on (I) position. Assuming no errors occur, the Series 2600 will power-up as follows:
After a few seconds with the OUTPUT indicators and display pixels on, the instrument model number and firmware revision levels are briefly dis played.
The node and the GPIB address are displayed briefly as follows:
KEITHLEY MODEL 26xx NODE = 1 GPIB = 26
The TSP-Link node and serial port parameters are displayed briefly:
KEITHLEY MODEL 26xx NODE = 1 SERIAL = 9600,8,N,1,NONE
1
1
Step 3: Set line frequency
The Series 2600 will operate at line frequencies of either 50Hz or 60Hz. For best measurement noise performance, the unit should be configured to match the actual line frequency used by selecting MENU > LINE-FREQ > 50Hz, 60Hz, or AUTO (When set to AUTO, the SourceMeter will automatically detect the power line frequency at power-up.).
1. Instrument will display actual model number: 2601, 2602, 2611, or 2612.
Return to In this section: 2600S-900-01 Rev. A / May 2006
-
1-10 Front Panel Operation Series 2600 System SourceMeters User’s Manual

How do I make measurements?

Reference Refer to the following information in Section 4 of the Series 2600
Reference Manual:
• “Basic-source-measure procedure”
• “Ohms measurements”
• “Power measurements”
The measurement procedure below demonstrates how to set up the Series 2600 to source 10V and measure current through a 10kΩ resistor.
WARNING Hazardous voltages may be present on the output and guard
terminals. To prevent electrical shock that could cause injury or death, NEVER make or break connections to the Series 2600 while the output is on. Power off the equipment from the front panel or disconnect the main power cord from the rear of the SourceMeter before handling cables connected to the outputs. Putting the equipment into standby does not guarantee the outputs are not powered if a hardware or software fault occurs.
Step 1: Connect the DUT
Connect a 10kΩ resistor to the SourceMeter Channel A HI and LO terminals as shown in then plug the terminal block into the Channel A connector. Refer to Section 3 of the Series 2600 Reference Manual for complete information.
Figure 1-3. Connect the resistor to the supplied screw terminal block,
Step 2: Select source and set source level
Perform the following steps to select the voltage source and set the source value to 10V:
1. Press Channel A SRC to select the V-Source as indicated by the “V” units in the source field on the display. The flashing cursor indicates which value is presently selected for editing.
2. Press the up RANGE key, and select the 20V source range (Models 2611/
2612) or 40V source range (Models 2601/2602). Use the lowest possible source range for best accuracy.
3. Use the CURSOR keys to move the cursor to the 10s digit, then press the Rotary Knob to enter the EDIT mode (EDIT annunciator on).
4. Using either the numeric keys or the Rotary Knob, set the source value to
10.0000V, then press ENTER or the Rotary Knob.
2600S-900-01 Rev. A / May 2006 Return to In this section:
Series 2600 System SourceMeters User’s Manual Front Panel Operation 1-11
Step 3: Set compliance limit
Perform the following steps to set the current compliance limit value to 10mA:
1. Put the Model 2602/2612 in the single-channel display mode with the DISPLAY key.
2. Press the LIMIT key, then press ENTER or the Rotary Knob.
3. Move the cursor to the 10s digit, then press the Rotary Knob to enter the EDIT mode.
4. Using the numeric keys or the Rotary Knob, set the limit value to
10.000mA, then press ENTER or the Rotary Knob.
Step 4: Select measurement function and range
1. Put the Model 2602/2612 in the single-channel display mode with the DISPLAY key, then select the current measurement function by pressing MEAS or MODE.
2. Select the measurement range with the RANGE keys. For the purposes of this example, press AUTO RANGE, and note the AUTO annunciator turns on. The instrument will automatically select the best range based on the measured value. You can also use manual ranging by pressing the up or down RANGE key, but be sure to use the lowest possible range for best accuracy.
Step 5: Turn output on
Turn the output on by pressing the appropriate OUTPUT ON/OFF key. The ON/OFF indicator LED will light when the channel output is on.
NOTE The Models 2611 and 2612 are equipped with a safety
interlock circuit that prevents operation on the 200V range if the circuit is not energized. This circuit is illustrated in
Figure 1-4. Refer to Section 10 of the Series 2600 Reference
manual for more information.
Step 6: Make measurements
1. Observe the readings on the display (Press TRIG if necessary to trigger the unit to begin taking readings.). For the single-channel display mode, the readings will appear on the top line, while source and limit values are on the bottom line. For the 10kΩ resistor under test, typical display values are:
1.00000mA
SrcA: +10.0000 V LimA:010.0000mA
Return to In this section: 2600S-900-01 Rev. A / May 2006
1-12 Front Panel Operation Series 2600 System SourceMeters User’s Manual
2. Use the DISPLAY key to cycle through the various display modes shown in
Figure 1-5 (The User State display messages are defined with specific
display commands; refer to Section 14 of the Series 2600 Reference manual.).
3. Press the MEAS key several times to display measured voltage, resis­tance, and power. Typical values for the 10kΩ resistor are: 10.0000V,
10.0000kΩ, and 10.0000mW.
Step 7: Turn output off
When finished making measurements, turn the output off by pressing the OUTPUT ON/OFF key. The OUTPUT indicator light will turn off.
Figure 1-3
DUT connectio
KEITHLEY SERIES 2600
LO GLO HI G G G HI
ns to 10kΩ resistor
CHANNEL A
SS
HI
DUT 10kW Resistor
LO
2600S-900-01 Rev. A / May 2006 Return to In this section:
Series 2600 System SourceMeters User’s Manual Front Panel Operation 1-13
Figure 1-4
Interlock circuit
Read by firmware
INTERLOCK pin
(on DIGITAL I/O connector)
10kW
Chassis ground
Rear panel
Coil resistance 145W +/- 10%
+220V supply
-220V supply
To output stage
Return to In this section: 2600S-900-01 Rev. A / May 2006
1-14 Front Panel Operation Series 2600 System SourceMeters User’s Manual
Figure 1-5
Display modes
1.00000mA .    V
SrcA:+10.0000V SrcB:+000.000mV
Press DISPLAY key
1.00000mA
SrcA:+10.0000V LimA:10.0000mA
Press DISPLAY key
.    V
SrcB:+000.000mV LimB:100.000mA
Press DISPLAY key
User State
Press DISPLAY key
Source-Measure display for SMU A and SMU B:
Top line displays the measure function (V, A, W or W). Bottom line displays the source function (V or A)
and level.
Source-Measure and Compliance Limit display for SMU A:
Top line displays the measure function (V, A, W or W). Bottom line displays the source function (V or A) and level,
and the compliance limit (A or V).
Source-Measure and Compliance Limit display for SMU B:
Top line displays the measure function (V, A, W or W). Bottom line displays the source function (V or A) and level,
and the compliance limit (A or V).
Display for user-defined messages and prompts.
     Indicates that a measured reading has not been
triggered.
2600S-900-01 Rev. A / May 2006 Return to In this section:
Series 2600 System SourceMeters User’s Manual Front Panel Operation 1-15

How do I use the buffer?

Reference Refer to Section 7 of the Series 2600 Reference Manual for more
detailed information on using the buffer.
The SourceMeter has two buffers per channel that can store from 1 to more than 100,000 readings. Each buffer reading is numbered and can also include the source value and a timestamp.
The following example shows how to store 100 readings, source values, and timestamps in Channel A, Buffer 1 and recall them from the front panel.
Step 1: Connect the DUT
Connect a 10kΩ resistor to the Channel A HI and LO terminals (Refer to
Figure 1-3).
Step 2: Set up source and measure functions
Using the procedure described in “How do I make measurements?” earlier in this section, set-up source and measure functions:
Source function: volts
Source range: 20V (2611/2612) or 40V (2601/2602)
Source value: 10V
Measure function: current
Measure range: auto
Step 3: Configure the buffer
1. Press the CONFIG key followed by the STORE key.
2. Choose COUNT, then set the number of readings to store to 100 using the Rotary Knob, and press ENTER or the Rotary Knob.
3. Select CHANA_BUFF, then press ENTER or the Rotary Knob to select Channel A.
4. Choose DEST, then CHANA_BUFFER1, and press ENTER or the Rotary Knob.
5. Choose BUFFER1, then press ENTER or the Rotary Knob to set up Buffer 1.
6. Select CLEAR, then YES to clear the buffer.
7. Choose ELEMENTS, then enable (ON) both SRC-VAL (source value) and TSTAMP (timestamp) storage.
8. Press EXIT several times to return to normal display.
Return to In this section: 2600S-900-01 Rev. A / May 2006
1-16 Front Panel Operation Series 2600 System SourceMeters User’s Manual
Step 4: Turn on the output
Press the CHAN A (Models 2602/2612) OUTPUT ON/OFF key to turn on the source output.
Step 5: Store readings
1. Press the STORE key to store readings. The asterisk (*) annunciator turns on to indicate data storage operation is enabled.
2. Press EXIT to stop data storage before it finishes.
Step 6: Turn off the output
Press the OUTPUT ON/OFF key to turn off the output after storage has completed.
Step 7: Recall readings
1. Press the RECALL key to access buffer readings (repeatedly pressing RECALL will cycle through Buffer 1 then Buffer 2 for Channel A and then Channel B). A message will be displayed if a buffer is empty.
2. Note the buffer display data (Refer to Figure 1-6):
The reading is on the top display at the left.
The buffer location number is on the right. For example, location #000001 indicates that the displayed reading is stored at the first memory location.
The source value is positioned at the lower left side of the display.
The timestamp is positioned at the lower right side. The first source­measure reading stored in the buffer (#000001) is timestamped at
00000000.001 seconds. Subsequent readings are timestamped relative to when the time storage was started. The interval between readings will depend on the reading rate.
3. To display the other readings stored in the buffer, choose the desired memory location number:
Use the Rotary Knob to increment and decrement the selected digit of the location number.
Set the cursor position with the Rotary Knob or CURSOR keys.
4. To exit from the data store recall mode, press EXIT.
2600S-900-01 Rev. A / May 2006 Return to In this section:
Series 2600 System SourceMeters User’s Manual Front Panel Operation 1-17
Figure 1-6
Buffer display format
Reading
Source Value
(SrcA1= Channel A, Buffer 1)
1.00000 mA #0000001
SrcA1:+10.0000 V @00000000.001s
Use Rotary Knob
1.00000 mA #0000002
SrcA1:+10.0000 V @00000000.002s
Use Rotary Knob
1.00000 mA #0000003
SrcA1:+10.0000 V @00000000.003s
Use Rotary Knob
1.00000 mA #0000004
SrcA1:+10.0000 V @00000000.004s
Buffer Location Number
Timestamp (seconds)
Note: Source values and timestamps will be displayed only if enabled when buffer is configured.
Return to In this section: 2600S-900-01 Rev. A / May 2006
1-18 Front Panel Operation Series 2600 System SourceMeters User’s Manual
This page left blank intentionally.
2600S-900-01 Rev. A / May 2006 Return to In this section:
Loading...
+ 78 hidden pages