Minimizing System and Instrument Noise. Even if a characterization system is
properly shielded and grounded, it’s still possible for noise to corrupt measurement results.
Typically, instruments contribute very little to the total noise error in the measurements. For
example, a good characterization system has a noise specification of about 0.2% of range,
meaning the p-p noise on the lowest current range is just a few femtoamps. Noise can be
further reduced with proper signal averaging (through filtering and/or increasing the
measurement integration period, i.e., integrating over a larger number of power line cycles).
The most likely sources of noise are other test system components, such as long
cables or switching hardware inappropriate for the application. Therefore, it is advisable to
use the best switch matrix available, designed specifically for ultra-low current
measurements. Then, keep all connecting cables as short as possible.
Generally, system noise has the greatest impact on measurement integrity when the
DUT signal is very small (i.e., low signal-to-noise ratio). This leads to the classic problem of
amplifying noise along with the signal. Clearly then, increasing the signal-to-noise ratio is
key to low level measurement accuracy.
Some characterization systems offer a low noise pre-amplifier option that allows
measurements down to the sub-femptoamp level. To get that level of sensitivity, it is best to
mount the pre-amps remotely on a probe station platen. With this arrangement, the signal
must travel only a very short distance (just the length of the probe needle) before it is
amplified. Then, the amplified signal is routed through the cables and switch matrix into the
measurement hardware.
Test System Settling Time. Fast, accurate low current measurements depend a great
deal on the way system elements work together. Measuring instruments must be properly
synchronized with the prober and switching matrix, if one is used. Improper synchronization
and source-measure delay may lead to collection of signals unrelated to the real device
parameters.
A step voltage test is typically used to characterize system settling time. A 10V step is
applied across two open-circuit probe tips, and then current is monitored continuously for a
period of time. The resulting current vs. time (I-t) curve (Figure 7) reveals several important
system characteristics.