MDDEL 150A
C.
GENERAL PRECAUTIONS
1.
Source Resistance - In SECTION II - SPECIFICATIONS under INRJT RESISTANCE, the maximum source resistance for use with
each voltage range is specified. Reasonable operation is possible
with source resistance up to ten times greater than those specified;
however, the measurement will suffer from a considerable decrease in
speed of response, and measuring accuracy. If the instrument is
left completely open-circuited, the meter will generally drift off
scale on any voltage range. On current ranges this does not happen
because of the input shunting resistors.
2.
Shielding - Since the instrument operates with a modulator fre-
quency of 120 cps, it is no% generally sensitive to 60 cps pickup
unless it is large enough to overload the amplifier.
The pickup
may be a source of difficulty when using the amplifier with high
impedances on the more sensitive voltage ranges and on the two or
three most sensitive current ranges. In these cases it is desirable
to shield the leads and the sources as completely as possible. In
some cases a simple low-pass filter at the input to eliminate frequencies of about 1 cps and above will be helpful. No use is made
of an input filter in this instrument, since any input series impedance due to the filter will increase the input noise and the thermal
drift.
When operating above ground, the case of the instrument must
be grounded.
3.
Thermal E&iF - Extreme precautions have been taken in the input
circuit to minimize thermal EMF's so that the residual EMF is less
than 0.6 microvolt.
'Ihe material used in the input circuit is pure
copper. Any other metal will generate a thermocouple potential.
Lead solder is particularly troublesome.
Where thermal EMF's are
a problem,
soldering should be done with the cadmium-tin solder
supplied with the instrument.
Ir.
Input Noise:
The noise at the input is a function of input
resistance and is approximately given by
E q 1.29 x lo-lo (R+2000)$
where E is the rms noise,
and R is the source resistance.
It is
assumed that the bandwidth of the instrument is about 1 cps and the
temperature is SOoF.
If noise is observed,
calculate the theoretica
noise and compare results.
Also bear in mind that only wire-wound
resistors approach the ideal resistor.
However, if Evanohm or
manganin resistors are used,a considerable thermal EMF of the re-
sistor material against copper will be observed.
5.
Checking the Zero Point - At low levels, spurious E&W's may be
generated simply by contact between the input leads and the terminals
under test.
If possible,
always leave the instrument connected and
adjust the zero after establishing a zero reference in the apparatus
under test.
For example,
in bridge measurements, disconnect the
bridge exciting voltage; or with a phototube, shield the tube from
light.
3-2
0763~