1.10 Filter Control *
The filter control is used to select the required filter.
(*Professional/Practitioner/Standard only.)
1.11 Filter Applications
Red Free (Green filter)
Is used to examine the blood vessels in fine detail. The green filter blocks red
rays showing blood vessels as black against a dark green background. This filter
is particularly useful for diabetic retinopathy.
Cobalt Blue *
Is used in conjunction with flourescein dye for the detection and examination of
corneal scars and abrasions.
(*Practitioner and Specialist only).
1.12 Pupillometer *
Hold the pupillometer adjacent to the patients eye to estimate pupil size. 1=1mm. The
range is 1mm to 8mm.
(*Applies to Specialist only.)
1.13 Precautions when using ophthalmoscopes
The intensity of light directed into the patients eye should be limited to the minimum
level necessary for diagnosis.
It is well established that exposure to the eye to intense light sources for extended
periods of time poses a risk of retinal photic injury. Many Ophthalmic instruments
illuminate the eye with intense light. The decision about the intensity of light level used
in any procedure must be made on a case by case basis. In each case, the clinician must
take a risk benefit judgement about the intensity of light to be used. Use of insufficient
intensity may result in inadequate visualization and in adverse effects more serious than
a retinal photic injury. Further, despite all efforts taken to minimize the risk of retinal
damage, damage may still occur. Retinal photic injury is a possible complication of the
need to use bright light to clearly visualize ocular structures during delicate ophthalmic
surgical procedures.
While no visible retinal photic lesions have been identified for ophthalmic instruments, it
is recommended that illumination levels be set to the minimum level necessary to perform
the diagnostic function. Young children and persons with diseased eyes may be at higher
risk. The risk may also be increased if the person being examined has had any exposure
with the same instrument or any other ophthalmic instrument using an intense visible
light source during the previous 24 hours. This will apply particularly if the eye has been
exposed to retinal photography.
The time to reach a potential optical radiation hazard for this device is 3 minutes when
the instrument is being operated at maximum intensity and maximum aperture. This
time is for a cumulative exposure in a day. It should be noted that there is a safety factor