Kanomax S-6531-1E, S-6541-0E, S-6533-0E, S-6533-1E, S-6541-1E User Manual

...
actoolsupply.com
Kanomax Climomaster 6501-0E Hot-wire Anemometer Kanomax Climomaster 6501-AE Hot-wire Anemometer Kanomax Climomaster 6501-BE Hot-wire Anemometer Kanomax Climomaster 6501-CE Hot-wire Anemometer Kanomax Climomaster S-6531-0E Hot-wire Anemometer Kanomax Climomaster S-6531-1E Hot-wire Anemometer Kanomax Climomaster S-6533-0E Hot-wire Anemometer Kanomax Climomaster S-6533-1E Hot-wire Anemometer Kanomax Climomaster S-6541-0E Hot-wire Anemometer Kanomax Climomaster S-6541-1E Hot-wire Anemometer Kanomax Climomaster S-6542-0E Hot-wire Anemometer Kanomax Climomaster S-6542-1E Hot-wire Anemometer Kanomax Climomaster S-6543-0E Hot-wire Anemometer Kanomax Climomaster S-6543-1E Hot-wire Anemometer Kanomax Climomaster S-6551-0E Hot-wire Anemometer Kanomax Climomaster S-6551-1E Hot-wire Anemometer Kanomax Climomaster S-6552-0E Hot-wire Anemometer Kanomax Climomaster S-6552-1E Hot-wire Anemometer Kanomax Climomaster S-6561-0E Hot-wire Anemometer Kanomax Climomaster S-6561-1E Hot-wire Anemometer
actoolsupply.com
actoolsupply.com
i
List of Components
Standard
Items
MODEL
Qty
Functions
Main Body
6501-00
One of
these
Main Body with analog output
6501-A0
With analog output function
Main Body with pressure
sensor
6501-B0
With pressure measurement function
Main body with pressure
sensor and analog output
6501-C0
With pressure measurement and analog output functions
Probe
6531-21
One of
these
Air velocity, air temperature, humidity sensor (directivity)
6541-21
Air velocity, air temperature
sensor (directivity)
6542-21
Air velocity, air temperature sensor (needle-type)
6533-21
Air velocity, air temperature, humidity sensor (spherical)
6543-21
Air velocity, air temperature sensor (spherical)
6551-21
Air velocity sensor (mini-spherical, straight type)
6552-21
Air velocity sensor (mini-spherical, articulating type)
6561-21
Air velocity, air temperature sensor (directivity)
Probe Cable
6531-06
1
To connect Probe to Main Body
USB Cable
1
USB Communication Cable
Carrying Case
1
Hard case
Operation Manual
1
Manganese AA Batteries
6
Available Accessories
Items
MODEL
Functions
Extension Rod (Flexible)
6531-04
For measurements in high places
Extension Rod (Straight)
6531-05
For measurements in high places
AC Adaptor
6113-02
Power Supply
Printer (Recommended)
DPU-S245
To print out data, calculation results, etc.
Printer Cable
6000-31
To connect the Main Body to the Printer
Mobile ANEMOMASTER Measurement Software
6000-41
Data acquisition software
Hands-free Case
6000-61
For hands-free measurement
actoolsupply.com
actoolsupply.com
ii
Important Safety Information
The symbols for the warnings used in this manual are defined below:
Classifications
Danger: To Prevent Serious Injury or Death
Warnings in this classification indicate a danger that may result in serious injury or death if not observed.
Caution: To Prevent Damage to the Product
Warnings in this classification indicate a risk of damage to the product that may void the product warranty if not observed.
Description of Symbols
This symbol indicates a condition that requires caution (including
danger). The subject of each caution is illustrated inside the triangle (e.g. the high temperature caution symbol is shown on the left).
This symbol indicates a prohibition. Do not take the prohibited action shown inside or near this symbol (e.g. the disassembly prohibition symbol is shown on the left).
●This symbol indicates a mandatory action. A specific action is given
near the symbol.
actoolsupply.com
actoolsupply.com
iii
Danger
Never bring the probe close to a flammable gas atmosphere.
The heated sensor may cause a fire or explosion.
Never touch the sensor.
The sensor is heated during operation. Touching the heated sensor may cause burns, and may also damage the sensor itself.
Do not disassemble or heat the batteries, or put them into a fire.
This may cause burns and the batteries may burst.
If abnormal noises, smells or smoke occur, or if liquid enters the instrument, turn off the instrument immediately, and remove the batteries or pull out the plug.
There is possibility of malfunction, electric shock, and/or fire. Please contact your local distributor or our service center for repair.
Caution
Always unplug the instrument from the electrical outlet when the instrument is not in use.
Failure to do so may cause an electrical shock, fire or circuit damage.
Do not use the instrument in a water vapor atmosphere.
Condensed steam on the sensor will change the heat dissipation rate, resulting in inaccurate measurements. It may also cause damage to the sensor.
This instrument is designed to be used in an environment with a clean air stream without any dust or foreign materials.
Foreign materials may cause damage to the sensor. Also dust or foreign materials on the sensor will impede accurate measurements.
Do not apply force to the sensor.
If the sensor is deformed, the accuracy of the sensor may be affected.
When measuring, ensure that the direction mark is facing into the airflow.
Otherwise, the measurement may be inaccurate, as some sensors (uni-directional probes) have a specific directivity.
Forbidden
Hot surface
Don’t touch
Explosive
Use properly
Forbidden
Set up properly
Caution
Use properly
Forbidden
Forbidden
actoolsupply.com
actoolsupply.com
iv
Do not use or leave the instrument in a high temperature, high humidity or dusty environment. Do not leave this instrument under direct sunlight for a prolonged period.
The instrument may not function properly out of the specified operating conditions.
Do not subject the instrument or the probe to strong impacts.
Dropping the unit or placing heavy objects on it may cause damage or malfunction to the instrument.
Never disassemble, modify or repair the product.
Failure to observe the above may cause a short circuit and/or other failures that will affect the performance.
Do not pick up or carry the instrument by the probe cable.
It may cause a malfunction or the wire may break.
Remove the batteries from the battery compartment when storing the instrument for a long period. Do not leave exhausted batteries in the battery compartment. When inserting batteries be sure to insert them with the polarity facing the correct direction.
Failure to do so may cause battery leakage.
Do not wipe the instrument with a volatile solvent.
The body may deform or deteriorate. Use a soft dry cloth to remove stains. If stains persist, soak the cloth in a neutral detergent and wipe the instrument with the soft cloth. Never use volatile solvents such as thinner or benzene.
Discharge any built-up static electricity from your body before touching the instrument.
The built-up static electricity may influence the readings and cause damage to the circuit.
Regularly check the head of the probe for contamination. Impurities (such as dust) on the sensor may affect the accuracy of the measurements.
To get rid of dust, use a blow blush for cameras to blow it off, or you can rinse it with water and allow it to air-dry completely. *Be sure to turn the power off before you clean it. *Never dry the probe with heat. (Heat may cause permanent damage to the sensor.)
Do not remodel/
disassemble
Electrified
Forbidden
Forbidden
Forbidden
Forbidden
Forbidden
Forbidden
Set up properly
Use properly
actoolsupply.com
actoolsupply.com
v
Do not move the main unit and the probe from a cold place to a warm place quickly. It will cause condensation.
Even when used in an environment within the specified operating temperature and humidity, a sudden temperature change may cause condensation. Condensation generated on the sensor may cause inaccurate measurements. Condensation on metal parts may cause rusting and lead to a malfunction.
Do not touch the LCD screen with a sharp-pointed object or with excessive pressure.
It may cause distortion of the screen or a malfunction. Also a rapid temperature change may cause a malfunction of the screen.
When storing the instrument, put the instrument in the carrying case and keep it in a place with an ambient temperature of -10 to 50°C and no condensation.
Do not dispose of the instrument as household waste.
Please note that the disposal of the instrument and the batteries should be in line with your local or national legislation. For details, please contact your local distributor.
Use properly
Forbidden
Forbidden
Forbidden
actoolsupply.com
actoolsupply.com
vi
Table of Contents
§ 1 Part Names and Functions ....................................................................... 7
Main Body ............................................................................................................. 7
Touch Panel ........................................................................................................... 8
Probe .................................................................................................................... 9
Probe Cable ......................................................................................................... 10
§ 2 Getting Started ...................................................................................... 11
Installing Batteries ................................................................................................ 11
Connecting the Probe and the Probe Cable .............................................................. 12
Disconnecting the Probe/Probe Cable ...................................................................... 12
Turning ON/OFF the Power .................................................................................... 13
Precautions for Measurements ............................................................................... 14
Duct Shape / Size Input ........................................................................................ 16
§ 3 Normal Measurement NORMAL MODE ............................................... 18
§ 4 Measuring Max, Min, and Mean: CALCULATION Mode ............................. 22
§ 5 Measurement of Air Flow FLOW RATE Mode ......................................... 25
SINGLE FLOW RATE Mode ..................................................................................... 25
MULTI FLOW RATE Mode ....................................................................................... 29
§ 6 Store and Recall Measured Data ............................................................. 31
§ 7 Print Out ............................................................................................... 34
Preparation for Printing ......................................................................................... 34
Print Out in NORMAL Mode .................................................................................... 34
Print Out in CALCULATION Mode and FLOW RATE Mode ............................................ 35
Print Out Stored DataSingle Page ...................................................................... 36
Print Out Stored DataPlural Pages ..................................................................... 38
§ 8 Digital Output ........................................................................................ 39
Preparation .......................................................................................................... 39
To Access the CLIMOMASTER from Your PC ............................................................. 39
§ 9 Analog Output ....................................................................................... 43
§ 10 Other Settings ....................................................................................... 45
Date.................................................................................................................... 45
Units and Baud Rate ........................................................................................... 46
Settings for LCD Backlight .................................................................................. 47
To Delete Data ..................................................................................................... 48
To Delete All the Data ........................................................................................... 49
Contrast Adjustment ............................................................................................. 50
§ 11 How to Clean a Probe ............................................................................ 51
§ 12 Specification .......................................................................................... 52
§ 13 Principle of Measurement ...................................................................... 53
Principle of Hot-wire Anemometer .......................................................................... 53
What are Discomfort Index (DI) and Dew Point Temperature (DT)? ............................ 55
About Compensation ............................................................................................. 56
§ 14 Probe Directivity (Air Velocity) ............................................................... 57
§ 15 Trouble Shooting .................................................................................. 59
§ 16 Warranty and After Service .................................................................... 61
Kanomax Limited Warranty .................................................................................... 61
§ 17 Contact Information .............................................................................. 62
actoolsupply.com
actoolsupply.com
7
§ 1 Part Names and Functions
Main Body
Pressure Port
Probe Socket
Analog Output Terminal
Serial
Communication
Terminal
DC Input Terminal
Power Switch
Battery Compartment
LCD Display
Touch Panel
Pressure port is only
available on MODEL6501-B0/-C0.
Analog output is only
available on
MODEL6501-A0/-C0.
actoolsupply.com
actoolsupply.com
8 Part Names and Functions
Touch Panel
MENU Key
Enters the Main Menu to select a function. *Press the MENU Key to CANCEL out of the current measurement or
settings and return to the Main Menu.
START/HOLD Key
This key will start and stop the calculation and/or measurement. It also functions as a hold key during measurements.
MODE Key
You can select the measurement mode
SET Key
Press the key to execute the selected item.
MENUaa1.NORMAL
aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO
▲ ▼ Navigation Keys
1. In Normal Mode, this key allows you to select
the time constant (1, 5, or 10 sec) (TC1TC5TC10)
TC1: Displays the instantaneous value
at every 1 second
TC5: Displays the moving average for 5 sec
TC10: Displays the moving average for 10 sec
2. In the Main Menu, press these keys to scroll in the desired direction
*1 Air temperature is only available on
MODEL6531/6541/6542/6543/6533/6561.
*2 Humidity is only available on MODEL6531/6533.
*3 Pressure is only available on MODEL6501-B0/6501-C0
*1 *2
Air Flow
Air Temp.
Humidity
Air Velocity
Air Temp.
Humidity
Pressure
*3
……………………
……………
………
……………
………
…………
…………………
……
Normal Measurement Mode
To input duct shape and size Calculation Mode To measure flow rate To output stored data To delete stored data
To set the date, time and units
Zero point adjustment
Zero point adjustment is only available on
MODEL6501-B0/-C0.
actoolsupply.com
actoolsupply.com
9
Probe
There are 8 different types of probes available for CLIMOMASTER. The model number and specifications depends on the type of the probe you have. The compatibility feature of the probes allows you to swap out probes freely without having to recalibrate the main unit.
MODEL6542-21
MODEL6531-21/6541-21/6561-21
*Model6531 only
MODEL6543-21
MODEL6533-21
MODEL6551-21
MODEL6552-21
*Unit: mm
○1
○2
○3 ○
4
*
○6
○5
○6
○1
○2○3
○1
2
○3
○4
○6
○1
○2
○7
○6
○6
○7
○2
○1
○1Air Velocity Sensor ○2Temp. Compensation Sensor ○
3
Temperature Sensor 4Humidity Sensor
○5Wind Direction Mark ○6Probe Number2 Shielded Cable
○1 ○
2○3
○6
actoolsupply.com
actoolsupply.com
10 Part Names and Functions
Probe Cable
The probe can be extended by using a probe cable between the main unit and the probe.
Extension Rod (Optional)
Straight
Flexible
Do not use the Flexible Extension Rod in an environment with an air velocity of 20 m/s or greater; use the Straight Extension Rod for these higher velocities.
Probe Connector
Main Body Connector
*Unit: mm
*Unit: mm
*Unit: mm
Approx.Φ15
Approx.Φ16
Approx.230 to 870
Approx.50
Approx.130
Approx.Φ23
Cable length: Approx.2000 including the rod
Cable length: Approx.2000 including the rod
Approx.Φ15
Approx.Φ23
Approx.130
Approx.330 to 960
Approx.50
Approx.Φ16
Approx.2000
Approx.43
Approx.43
Approx. Φ15
Approx. Φ14
actoolsupply.com
actoolsupply.com
11
§ 2 Getting Started
Installing Batteries
Backside of the body
4. Insert the batteries according to the
indicated polarity chart. This instrument requires six (6) AA size batteries that must be of the same type. DO NOT mix different types of
batteries. Failure to observe this may cause battery leakage or damage the
instrument.
*Batteries CANNOT be recharged by
the AC adapter.
3. Lift the cover away from the body.
2. Slide the cover until it stops.
5. Put the cover back on by reversing
the above procedure.
1. Press down on the battery cover.
Manganese AA size batteries
Alkaline AA size batteries
Ni-Cd AA size batteries
Ni-MH AA size batteries
Types of batteries that can be used are:
actoolsupply.com
actoolsupply.com
12 Getting Started
Connecting the Probe and the Probe Cable
This section describes how to connect the probe or the probe cable to the main unit.
Make sure that the power is turned OFF when connecting or disconnecting the probe or the probe cable.
Disconnecting the Probe/Probe Cable
To remove the probe, follow the procedure below:
Make sure that the power is turned OFF when connecting or disconnecting the probe or the probe cable.
Probe Terminal to connect
the Probe/Probe Cable
1. Align the connector of the probe or
the probe cable with the probe terminal on the body.
*When connecting the probe, make sure
the probe number faces towards the front
side of the main unit.
2. Line up the connector of the probe
and the terminal of the body and gently push until it clicks.
*DO NOT FORCE in the probe/probe cable
connector to the main body and DO NOT
ROTATE the probe/probe cable once it is
connected. Doing so may cause serious
damage to the instrument.
1. Lightly lift the locking ring of the probe/probe cable (A, as shown left).
A
2. Pull the probe/probe cable out
vertically while holding the lock ring up.
*DO NOT rotate the connector of the probe/probe cable when it is connected to the main body. Doing so may cause serious damage to the instrument.
Probe Number
Probe No.
actoolsupply.com
actoolsupply.com
13
Turning ON/OFF the Power
The power switch to turn the power ON/OFF is located at the side of the instrument. When you turn the power on after connecting the probe/probe cable, a test screen with the Kanomax logo, model, and the version of the firmware will be displayed for a few seconds before it switches to the Normal Mode screen.
Battery Level Indicator
Check the Battery Level Indicator to confirm the remaining battery level. The battery consumption rate largely depends on the measured air velocity. When the batteries need to be replaced (or
recharged), the indicator will start blinking. The screen may freeze if a high velocity is measured when the battery level indicator is blinking. The indicator changes as indicated below according to the remaining battery level:
CONNECT PROBE AND RESTART!
NO PROBE!
'12/04/12a15:40:23 ~
TCC1
0.87f
25.3e
64.1g
Power Switch
|: Power ON
: Power OFF
Power ON
Connect the probe/probe
cable to the instrument
Normal Mode Screen (refer to P.18 for details)
1. Current date and time 4. Air temperature
*1
2. Time constant (P.20) 5. Humidity
*2
3. Air velocity 6. Battery level indicator
*1Air temperature is available only on MODEL6531/6541/6542/6543/6533/6561. *2Humidity is available only on MODEL6531/6533.
Batteries must be replaced.
Blinks
- Locked Status -
When is displayed, all functions of the
instrument will be locked. The current measurement
will be cancelled and data will NOT be saved. To release
the lock, turn off the power and replace the batteries.
Probe is not
connected
properly
Annual Calibration
A year after
calibration
Press any key or 10 sec later
When the probe is not connected properly, the “NO PROBE!” warning will be displayed as shown above. Turn the power off, and check the connection of the probe.
Normal
Flow
Last calibration
date
A year after the last calibration, a notice to
perform the annual calibration will be displayed on the screen. The screen will switch to normal
measurement mode 10 seconds later or after any key is pressed.
'12/04/12a15:40:23 ~
TCC1
0.87f
25.3e
64.1g
aaaaCLIMOMASTER aaaaVersiona4.00
aa*aNowaLoadinga!a*
ooCalibrationoDue ooooooLastoDone Ooooo2010/02/23 o<PushoANYoKey:Start>
1.
6.
2.
3.
4.
5.
actoolsupply.com
actoolsupply.com
14 Getting Started
Precautions for Measurements
Air Velocity
MODEL6531/6541/6561: The probe has its own
directivity characteristics. Make sure that the direction mark is facing into the airflow. If you are not sure of the airflow direction, slowly rotate the probe and select the point where you get the maximum velocity reading.
MODEL6533/6542/6543/6551/6552: This omnidirectional
probe has a horizontal, 360 degree dynamic characteristic
(Refer to “Directivity of the Probe” on P.57 for details).
The probe relies on both an air velocity sensor and temperature
compensation sensor to accurately measure air velocity. In order to do so, it is necessary that both sensors are evenly exposed to the airflow under the same temperature conditions (Refer to the figures of the probes on P.9 for the location of the sensors).
For measurements in an environment with rapid air temperature
changes, measure for at least 20 seconds and wait for the reading to stabilize before starting the actual measurement (i.e. the data will not be accurate until the probe has time to acclimate to the environment.)
Air Temperature
The faster the wind, the shorter the response time for temperature
measurements. The normal response time is approximately 30 seconds when the air velocity is 1 m/s (90% response). Wait for the reading to become stable before taking a measurement.
When no airflow is present, the air temperature reading may be
higher than the actual temperature due to the heat generated by the air velocity sensor. It is recommended that you SLOWLY wave the probe to create an environment with approx. 0.1m/s airflow to obtain accurate temperature readings.
Humidity
Humidity readings in an atmosphere with extremely high humidity
or rapid temperature changes may be higher than the actual humidity due to condensation generated on the surface of the sensor. If this occurs, allow the probe to dry in an atmosphere with 40%RH humidity or less for 24 hours.
―― Comparing with ASSMANN Psychrometer ――
The humidity measurement function of CLIMOMASTER is strictly calibrated, traceable to Japanese National Standards carried by JEMIC (Japan Electric Meters Inspection Corporation), and it assures you highly accurate performance. The electronic hygrometer, used in CLIMOMASTER, is known to be more stable and has a shorter response time compared to conventional ASSMAN Psychrometers. Also, the electronic hygrometer is not dependent upon the measurement conditions, while an ASSMANN Psychrometer can easily be influenced by many factors such as dust, condensation and the way the gauze is wrapped.
Direction Mark
Airflow
*For more
information on proper
handling of the ASSMAN Psychrometer, please refer to Japanese Industrial Standard (JIS-Z8806 Method of measuring Humidity), or its
counterpart standards that apply.
actoolsupply.com
actoolsupply.com
15
Pressure
Do not expose the pressure sensor to more than 75kPa of pressure.
Excess pressure may cause serious damage to the sensor.
The operating temperature must be between 5oC and 40oC or 41oF
and 104oF when measuring pressure. The instrument may not operate properly outside of this temperature range.
Make sure to perform a zero adjustment before measuring
pressure. When performing a zero adjustment, leave both pressure ports (+) and (-) open.
The zero adjustment range is – 0.5 to 0.5 kPa. If the reading is out
of this range, an error message will be displayed.
Zero Adjustment Procedure
Press to enter the Main Menu.
Use to select “8. PRESSURE ZERO”
and press .
Use to select “1. ZERO ADJUST”
and press .
Use to select “YES”.
Press to display 0.00 kPa.
Use to select “2. SAVE INFO”.
Press . This will complete the zero
adjustment and you are back in Main
Menu.
If you press before you save, you will go back to the Main Menu
without completing the zero adjustment.
Connecting the Pressure Tube
Connect the pressure tube to the (+) or (-) port as you see on the left.
Connect/insert the other end of the tube to the measuring hole of the duct. If the pressure that you are going to measure is positive, connect the tube to (+) port. If it is negative, connect the tube to (-) port.
aaaaPRESSUREaZERO
1.ZEROaADJUSTaaaYES
2.SAVEaINFO
00.00h
aaaaPRESSUREaZERO
1.ZEROaADJUSTaaaYES
2.SAVEaINFO
00.00h
aaaaPRESSUREaZERO
1.ZEROaADJUSTaaaYES
2.SAVEaINFO
-0.00h
MENUaa1.NORMAL
aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO
*In order to take
an accurate measurement,
make sure that the
tube is properly secured without any leakage, or
bends.
Pressure sensor is
only available on
MODEL6501-B0/-C0.
actoolsupply.com
actoolsupply.com
16 Getting Started
Duct Shape / Size Input
Before measuring the volumetric flow rate, the duct shape and size settings must be entered. Up to 25 different duct types can be registered in this instrument. Select a duct type from the registered duct types to measure the air flow rate.
Press to enter the Main Menu.
Use to select ”2. DUCT TYPE”.
Press .
<To Set Duct ENTRY No.>
You can choose a memory address from 1 to 25 to store the duct data in.
Use to select “1. ENTRY No.
and press .
Use to select ENTRY No. and
press .
<To Set Duct SHAPE (RECTANGLE/CIRCLE)>
Use to select “2. SHAPE” and press .
Use to select “RECTANGLE” or “CIRCLE” and press .
Rectangular Duct Circular Duct
aaaaaaaDUCTaTYPE
1.ENTRYaNo.aaaaaaaa1
2.SHAPEaaaaCIRCLE
3.DIAMETERaaaaaaa300
4.UNIT(mm/inch)aaamm
6.SAVEaINFO
aaaaaaaDUCTaTYPE
1.ENTRYaNo.aaaaaaaa1
2.SHAPEaaaaRECTANGLE
3.WaSIZEaaaaaaaaaaa-
4.HaSIZEaaaaaaaaaaa-
5.UNIT(mm/inch)aaamm
6.SAVEaINFO
aaaaaaaDUCTaTYPE
1.ENTRYaNo.aaaaaaaa1
2.SHAPEaaaaRECTANGLE
3.WaSIZEaaaaaaaaaaa-
4.HaSIZEaaaaaaaaaaa-
5.UNIT(mm/inch)aaamm
6.SAVEaINFO
MENUaa1.NORMAL
aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO
actoolsupply.com
actoolsupply.com
17
<To Set Duct Size>
Registering a Rectangular Duct
Use to select “3. WIDTH” and
press
.
Use to enter the
width and press
.
Use to select “4. HEIGHT and
press . Use to enter the
height and press .
The maximum is 9999mm or 999.9 inch.
Registering a Circular Duct
Use to select “3. DIAMETER”
and press . Use to enter
the diameter of the duct and press .
The maximum is 9999mm or 999.9inch.
<To Set Units (mm/inch)>
Use to select “5. UNIT(mm/inch)”
and press .
*For a circular duct, select
“4. UNIT (mm/inch)”.
Use to select mm or inch and
press .
<To Save the Setting>
Use to select “6. SAVE INFO” and
press .
This will complete the Duct Shape/Size
Input and you are back in Main Menu.
If you press before you save, you will go back to the Main Menu without completing the Duct Shape/Size Input.
aaaaaaaDUCTaTYPE
1.ENTRYaNo.aaaaaaaa1
2.SHAPEaaaaRECTANGLE
3.WIDTHoooooooooo800
4.HEIGHTooooooooo500
5.UNIT(mm/inch)aaamm
6.SAVEaINFO
aaaaaaaDUCTaTYPE
1.ENTRYaNo.aaaaaaaa1
2.SHAPEaaaaRECTANGLE
3.WaSIZEaaaaaaaaa800
4.HaSIZEaaaaaaaaa500
5.UNIT(mm/inch)aaamm
6.SAVEaINFO
aaaaaaaDUCTaTYPE
1.ENTRYaNo.aaaaaaaa1
2.SHAPEaaaaRECTANGLE
3.WaSIZEaaaaaaaaa800
4.HaSIZEaaaaaaaaa500
5.UNIT(mm/inch)aaamm
6.SAVEaINFO
aaaaaaaDUCTaTYPE
1.ENTRYaNo.aaaaaaaa1
2.SHAPEaaaaCIRCLE
3.DIAMETERaaaaaaa300
4.UNIT(mm/inch)aaamm
6.SAVEaINFO
actoolsupply.com
actoolsupply.com
18 Normal Measurement NORMAL MODE
§ 3 Normal Measurement NORMAL MODE
This is the mode that you will be in, when you first turn on the instrument. In this mode you cannot save any data. The display is updated every 1 second.
To move to NORMAL Mode from
another measuring mode, press .
Select “1. NORMAL” and press .
Selecting the Measuring Parameters
The measurement parameters can be selected in Normal Mode as described below.
<Air Velocity Measurement Screen>
While in Normal Mode as shown left,
the measured item can be changed by
pressing . The item changes in the following sequence: PressureAir
RateAir Velocity
<Pressure Measurement Screen>
Pressure measurement is a feature for
MODEL 6501-B0/6501-C0.
This screen is displayed on MODEL 6501-B0/-C0, which has a built-in pressure sensor.
<Flow Rate Measurement Screen>
The selected duct type will be
displayed at the lower left of the
screen.
To select the duct type, use
while pressing .
If you have not selected a duct type or your selected
duct type (number) is not available, the instrument will
not give you the flow rate reading. Make sure the duct
type is properly registered and selected. See Getting
Started Duct Shape/Size Input on P.16 for details.
'12/04/12a16:24:37 ~
TCC1
DCC15 WCCC­HCCC-
------a
24.9e
63.7g
'12/04/12a16:24:37 ~
TCC1
DCCC3 C1050
6037a
24.9e
63.7g
'12/04/12a16:24:37 ~
0.31h
'12/04/12a15:40:23 ~
TCC1
0.87f
25.3e
64.1g
MENUaa1.NORMAL
aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO
Duct type
Humidity is only
available on
MODEL6531/6533.
Air temperature is
only available on MODEL6531/6541 /6542/6533/6543/ 6561
Pressure is only
available on MODEL6501-B0
/-C0.
actoolsupply.com
actoolsupply.com
19
Hold the Reading
While measuring (in NORMAL Mode),
press the key to hold the current
reading. This is applicable in both the
flow rate and pressure measurement
screens as well.
The ”HOLD” indicator appears on the display to indicate that the reading shown is held.
Press again to release.
Hold the Maximum Value
While measuring (NORMAL Mode),
press and hold .
The “HOLD” indicator appears on the
display. As long as you keep pressing
down the button you can hold the
maximum value of each parameter (air
velocity, air temperature, humidity and
pressure).
When you release , the reading
shown will be frozen. Press again
to release.
'12/04/12a15:40:23 ~
TCC1 HOLD
0.87f
25.3e
64.1g
'12/04/12a15:40:23 ~
TCC1
0.87f
25.3e
64.1g
'12/04/12a15:40:23 ~
TCC1 HOLD
0.87f
25.3e
64.1g
'12/04/12a15:40:23 ~
TCC1
0.87f
25.3e
64.1g
Humidity is only
available on
MODEL6531/6533.
Air temperature is
only available on MODEL6531/6541 /6542/6533/6543/
6561.
Pressure is only
available on MODEL6501-B0/-C0.
actoolsupply.com
actoolsupply.com
20 Normal Measurement NORMAL MODE
Setting the Time Constant
In NORMAL Mode, you can change the
Time Constant (TC) by pressing .
You can select the Time Constant from 1, 5 or 10seconds.
TC1: Displays the instantaneous value at every 1 second TC5: Displays moving average for 5 seconds TC10: Displays moving average for 10 seconds
*The Time Constant is only effective in Normal Mode. When you first turn
on the instrument, it is effective only for the AIR VELOCITY and the AIR FLOW measurements in Normal Mode. If you need to apply the time constant to the AIR TEMPERATURE, HUMIDITY and PRESSURE measurements in Normal Mode, refer to the next section “Changing
Time Constant Application”.
*** What is Time Constant? ***
Time Constant determines the time span of the moving average. When you set the Time Constant to a larger (longer) value, the readings will become stable. When you select a smaller (shorter) value, the readings will become more responsive and sensitive to changes in air velocity. This function is not available in Calculation or Flow Rate Mode.
MODE
How To Take In Measured Data
EXPLANATION
TC1
Displays the instantaneous value at every 1 sec.
TC5
Updates every second. The value displayed is the average of the past 5 seconds.
TC10
Updates every second. The value displayed is the average of the past 10 seconds.
'12/04/12a15:40:23 ~
TCC1
0.87f
25.3e
64.1g
Time Constant
Data for 1 sec
Average of 5 sec
Average of 10 sec
Humidity is only
available on
MODEL6531/6533.
Air temperature is
only available on MODEL6531/6541 /6542/6533/6543/
6561.
sec. (Measuring Time)
sec. (Measuring Time)
sec. (Measuring Time)
Pressure is only
available on MODEL6501-B0/-C0.
actoolsupply.com
actoolsupply.com
21
Changing the Time Constant Application
When you first turn on the instrument, the Time Constant is only effective for AIR VELOCITY and AIR FLOW in Normal Mode. If you want to make it effective for AIR TEMPERATURE, HUMIDITY and PRESSURE, the following change must be made.
Press to enter the Main Menu.
Use and select “7. UTILITY”.
Press .
Use and select “4. TC(T.H.P) SET”. Press .
*If your instrument does not have the optional pressure sensor, the screen will
display “4. TC(T.H)”.
Select “1. TC(T.H.P) SET” and press
. Use to select YES or NO.
Press .
YES: The time constant is effective for
all parameters in Normal Mode.
NO: The time constant is only effective
for the AIR VELOCITY and FLOW RATE.
<Save the Settings>
Use to select “2. SAVE INFO”.
Press to save the new settings
and go back to Main Menu.
If you press before saving the settings you will return back to
the Main Menu and any setting changes you made will not be saved.
TC(T.H.P) SET
1.TC(T.H.P) SET NO
2.SAVE INFO
TC(T.H.P) SET
1.TC(T.H.P) SET YES
2.SAVE INFO
aaaaaaaUTILITY
1.TIME AJUST
2.UNIT SELECT
3.ANALOG OUTPUT
4.TC(T.H.P) SET
MENUaa1.NORMAL
aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO
*P indicates Pressure (MODEL6501-B0/-C0).
*
T indicates Air Temperature (MODEL6531/6541/6542/6533/6543/6561).
*H indicates Humidity (MODEL6531/6533).
oooooooUTILITY
1.TIMEoADJUST
2.UNIToSELECT
3.ANALOGoOUTPUT
4.TC(T.H.P)oSET
5.BACKLIGHToON/OFF
Pressure is only
available on MODEL6501-B0
/-C0.
actoolsupply.com
actoolsupply.com
22 Measuring Max, Min, and Mean: CALCULATION Mode
§ 4 Measuring Max, Min, and Mean: CALCULATION Mode
CALCULATION Mode will automatically calculate the maximum, minimum and mean of the measured data.
Press to enter the Main Menu. Use
to select “3. CALCULATION
and press .
CALCULATION Mode Display Icons
1. CALCULATION Mode AVERAGE: Takes a measurement
every second and then displays the average for the set sampling time. INSTANT: Takes a single measurement at the end of the sampling time.
2. SAMPLING TIME (1 to 999 sec)
To set the length of the sampling time.
3. No. TRIAL (N) (1 to 9999 times)
To set the number of samples (trials) needed. Each trial will be for the duration specified in sampling time.
4. DATA STORAGE (YES or NO)
5. SET TO START
Save the settings and return to standby.
MENUaa1.NORMAL
aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO
Remaining Memory Available
Total Memory
The 1st trial The 2nd trial The Nth trial
TRIAL(1) TRIAL(2) …………… TRIAL(N)
Measuring Time……Sampling Time x N (Number of Trial)
Sampling Time
CALCULATE R20000/20000
1.MODE11111111AVERAGE
2.SAMPLING(TIME)0012s
3.SAMPLING(n)110050
4.DATA2STORAGE2?22YES
5.SET2TO2START
actoolsupply.com
actoolsupply.com
23
<Setting the CALCULATION MODE>
Select “1. MODE“ and press .
Use to select AVERAGE or
INSTANT and press .
<Setting the SAMPLING TIME>
Use to select
“2. SAMPLING (TIME) and press .
Use to select SAMPLING TIME
(1 to 999 sec) and press .
<Setting the No. TRIAL (N)>
Use to select “3. SAMPLING(n)
and press .
Use to select No. TRIAL
(1 to 9999 times) and press .
Data obtained by the number of trials
set here will be stored.
<DATA STORAGE>
Use to select
“4. DATA STORAGE?” and press .
Use to select YES or NO and
press .
There must be sufficient memory to store the calculation data. If the measured data would exceed the available number of empty data locations, Climomaster will automatically adjust the number of trials to match the amount of free remaining memory locations. (Ex.: if there is R0020/20000 remaining, you can only measure 20 times even if you set the No. TRIAL to more than 20.
<Save the Settings>
Use to select
“5. SET TO START” and press .
(In the example shown at the left, the
instrument will perform fifty (50)
measurements of one (1) second intervals.)
CALCULATE R20000/20000
1.MODE11111111AVERAGE
2.SAMPLING(TIME)0012s
3.SAMPLING(n)110050
4.DATA2STORAGE2?22YES
5.SET2TO2START
CALCULATE R20000/20000
1.MODE11111111AVERAGE
2.SAMPLING(TIME)0012s
3.SAMPLING(n)110050
4.DATA2STORAGE2?22YES
5.SET2TO2START
CALCULATE R20000/20000
1.MODE11111111AVERAGE
2.SAMPLING(TIME)0012s
3.SAMPLING(n)110050
4.DATA2STORAGE2?22YES
5.SET2TO2START
CALCULATE R20000/20000
1.MODE11111111AVERAGE
2.SAMPLING(TIME)0012s
3.SAMPLING(n)110050
4.DATA2STORAGE2?22YES
5.SET2TO2START
CALCULATE R20000/20000
1.MODE11111111AVERAGE
2.SAMPLING(TIME)0012s
3.SAMPLING(n)110050
4.DATA2STORAGE2?22YES
5.SET2TO2START
actoolsupply.com
actoolsupply.com
24 Measuring Max, Min, and Mean: CALCULATION Mode
<READY>
The instrument is ready for measuring.
Press to change the applied
parameters. (Air Velocity, Air Temp,
Humidity – Flow Rate, Air Temp,
Humidity – Pressure)
Press to start.
<While Measuring>
Press to stop.
If you have selected “YES” for “4. DATA STORAGE?”, the measured data will be
stored.
Pressing will also stop the
measurement but the data will not be
saved.
<RESULT>
After all the trials are finished, the calculated results will be displayed.
Press to check each parameter
in the following sequence: Air Velocity
(or Air Flow) Air Temp. Humidity
Dew Point Temperature (DT),
Discomfort Index (DI).
*If you measured Pressure, only Pressure will be displayed.
Press to return to the setup menu of the CALCULATION Mode.
Press to return to the Main Menu. If YES is selected for “DATA
STORAGE”, the results will be stored.
Related Functions:
If a printer is connected, press to print out the results.
To recall stored data P. 31
Print out P. 34
What is Dew Point Temperature (DT) and Discomfort Index (DI)?
P. 55
'12/04/12a15:40:23 ~
RDY.
N2221 /2250
0.87f
25.3e
64.1g
Humidity is only
available on
MODEL6531/6533.
Air temperature is
only available on MODEL6531/6541 /6542/6533/6543/
6561.
'12/04/12a15:42:57 ~
SMP .
N2223 /2250
2.11f
25.3e
64.2g
2222CALCULATION
MAX
AVG
MIN
2.76f
1.43f
0.81f
Pressure is only
available on MODEL6501-B0
/-C0.
actoolsupply.com
actoolsupply.com
25
§ 5 Measurement of Air Flow FLOW RATE Mode
CLIMOMASTER features an accurate Flow Rate Mode which corresponds to the industry measurement standards such as ASHRAE.
SINGLE FLOW RATE Mode
This mode is useful for performing a measurement of the airflow of duct (i.e. a duct traverse). Following the steps listed below, you can program Climomaster with the following parameters:
Sampling Time: The instrument will take a sample at the end
of this entered sampling time. Example: if you setup a sampling time of 10 seconds, Climomaster will wait 10 seconds and then take a measurement. If you are taking multiple samples, it will wait 10 seconds between each one before taking the next sample.
Number of Trials: this is the number of samples you wish to
collect from a single point in the traverse.
Number of Points: this is the total number of points in the
traverse.
In single mode, the average, maximum and minimum values from all points will be calculated. The instrument will record the average from each point as well as providing a final average of all the points in the traverse.
The value of each point is recorded as the average of TRIAL(1) to TRIAL(N). This value
(for each individual point) is stored in the memory.
To take an accurate measurement of the airflow of duct, you need to figure out the average velocity of air inside the duct and the area of the cross section of the duct.
TRIAL(1) (2) (3) ……(N) TRIAL(1) (2) (3) ……(N) TRIAL(1) (2) (3) (N)
Time needed to measure a point…Sampling Time X N (Number of Trial)
Value at point 1 Value at point 2 Value at point M
POINT(1) POINT(2) ……………… POINT(M)
POINT(1): avg(1)ΣTRIAL(N)N
POINT(2): avg(2)ΣTRIAL(N)N
・・・
POINT(M): avg(M)ΣTRIAL(N)N
<Result>
Average (AVG): AVG(avg(1)+ avg(2)+ ・・・・+avg(M))M
Maximum (MAX): MAX Maximum value from avg(1) to avg(M).
Minimum (MIN): MIN Minimum value from avg(1) to avg(M).
actoolsupply.com
actoolsupply.com
26 Measurement of Air Flow FLOW RATE Mode
Air Flow: Air Volume per Time Unit [m3/min, m3/h, ft3/min, ft3/h] Air Flow (Q) = Average Air Velocity (U) x Cross Sectional Area (A)
Press to enter the Main Menu.
Use to select
“4. FLOW RATE” and press .
Use to select
“1. SINGLE FLOW RATE and press
.
<Setting the SAMPLING TIME>
Use to select
“1. SAMPLING(TIME)” and press .
Use to select SAMPLING TIME
(1 to 999) and press .
SINGLE FLOW RATE Mode Display Icons
1. SAMPLING TIME (1 to 999 sec)
To set the interval time between each sampling. (INSTANT)
2. No. TRIAL (N) (1 to 9999 times)
To set the number of trials (data) to be sampled at each point.
3. MEAS. POINT (1 to 2000 points)
To set the number of measuring points.
4. DATA STORAGE (YES or NO)
5. DUCT ENTRY NO.
Select the duct entry number of the duct that is to be measured. Registration and changes to the duct information can also be made.
6. SET TO START
Save the settings and return to standby.
MENUaa1.NORMAL
aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO
SAMPLING MODE
1.SINGLEoFLOWoRATE
2.MULTIoFLOWoRATE
SINGLE (R20000/20000
1.SAMPLING(TIME)001os
2.SAMPLING(n)110050
3.MEAS.2POINT220010
4.DATA2STORAGE2?22YES
5.DUCT2ENTRY2NO.2221
6.SET2TO2START
Remaining Memory Available
Total Memory
r1=0.316R r2=0.548R r3=0.707R r4=0.837R r5=0.949R
Measurement
Point
Circular Duct
R
r1
r2
r3
r4
r5
Rectangular Duct
SINGLE (R20000/20000
1.SAMPLING(TIME)001os
2.SAMPLING(n)110050
3.MEAS.2POINT220010
4.DATA2STORAGE2?22YES
5.DUCT2ENTRY2NO.2221
6.SET2TO2START
actoolsupply.com
actoolsupply.com
27
<Setting the No. TRIAL>
Use to select
“2. SAMPLING(n)and press .
Use to select No. TRIAL (1 to
9999) and press .
<Setting MEAS. POINT>
Use to select
“3. MEAS. POINT” and press . Use to select the number of
MEAS. POINT (1 to 2000) and press
.
<Data Storage>
Use to select
“4. DATA STORAGE?” and press .
Use to select YES or NO and
press
.
<Selecting DUCT ENTRY No.>
Use to select
“5. DUCT ENTRY NO.” and press .
*If you select an empty ENTRY NO., you
cannot start the Flow Rate measurement.
The duct type settings screen will be displayed. Here you can input the dimensions of the duct.
*Refer to “Duct Shape/Size Input” on P.16 for
details.
<Save the Settings>
Use to select
“6. SET TO START” and press .
If you press before saving the settings you will return back to the Main Menu and any setting changes you made will not be saved.
aaaaaaaDUCTaTYPE
1.ENTRYaNo.aaaaaaaa1
2.SHAPEaaaaRECTANGLE
3.WIDTHoooooooooooo-
4.HEIGHTooooooooooo-
5.UNIT(mm/inch)aaamm
6.SAVEaRETURN
SINGLE (R20000/20000
1.SAMPLING(TIME)001os
2.SAMPLING(n)110050
3.MEAS.2POINT220010
4.DATA2STORAGE2?22YES
5.DUCT2ENTRY2NO.2221
6.SET2TO2START
SINGLE (R20000/20000
1.SAMPLING(TIME)001os
2.SAMPLING(n)110050
3.MEAS.2POINT220010
4.DATA2STORAGE2?22YES
5.DUCT2ENTRY2NO.2221
6.SET2TO2START
SINGLE (R20000/20000
1.SAMPLING(TIME)001os
2.SAMPLING(n)110050
3.MEAS.2POINT220010
4.DATA2STORAGE2?22YES
5.DUCT2ENTRY2NO.2221
6.SET2TO2START
SINGLE (R20000/20000
1.SAMPLING(TIME)001os
2.SAMPLING(n)110050
3.MEAS.2POINT220010
4.DATA2STORAGE2?22YES
5.DUCT2ENTRY2NO.2221
6.SET2TO2START
SINGLE (R20000/20000
1.SAMPLING(TIME)001os
2.SAMPLING(n)110050
3.MEAS.2POINT220010
4.DATA2STORAGE2?22YES
5.DUCT2ENTRY2NO.2221
6.SET2TO2START
actoolsupply.com
actoolsupply.com
28 Measurement of Air Flow FLOW RATE Mode
aaaaaa0001/0015aaa~
RDY.
P
2222
N
2221
D
2223
C
2300
---502a
27.1e
62.9g
<Ready>
The ready screen will be displayed.
Use while pressing to
change the entry no. of the duct type.
Press to start measuring.
Display Icons
1. RDY: Current Status (READY/SAMPLE)
2. P 1: Current Number of Points
3. N 1: Current Number of Trial
4. D 3: Selected Duct Entry Number
5. C300: Duct Shape and Size (In this case, the duct is circular and 300mm in diameter.) For the rectangular duct, W*** and H*** will be displayed.
<While Measuring>
Press to stop the measurement. To
continue measuring, press again.
Press to complete the
measurement. Then the results will be
displayed.
If Point 1 is not completed, no results will be displayed on the screen.
Press to complete the measurement without storing data.
<READY After Point 1>
After completing the measurement of Point 1, the instrument will be in STANDBY, ready to measure Point 2.
<RESULT>
After the completion of the measurement, the results will be displayed.
Press to select the parameter in the following sequence: Air Flow
Air TemperatureHumidity
Air Velocity
Press to return to the Main Menu.
Press to return to the setup screen of the FLOW RATE Mode.
*If “DATA STORAGE” is set to YES, the results will be stored.
'12/04/12a16:11:59 ~
RDY.
P
2221
N
2221
D
2223
C
2300
--6037a
25.1e
63.7g
1
2 3 4
5
Number of Points already measured
Number of Points
Humidity is only
available on
MODEL6531/6533.
Air temperature is
only available on MODEL6531/6541 /6542/6533/6543/
6561.
'12/04/12a16:24:37 ~
SMP .
P
2221
N
2224
D
2223
C
2300
--6037a
24.9e
63.7g
aNEXT112001/015222 ~
RDY.
P
2222
N
2221
D
2223
C
2300
--6502a
27.1e
62.9g
aNEXT1FLOW2RATE
MAX
AVG
MIN
106433a
085134a 066797a
actoolsupply.com
actoolsupply.com
29
MULTI FLOW RATE Mode
MULTI FLOW RATE is very similar to the SINGLE FLOW RATE MODE, except it adds an additional parameter for the number of locations. This allows you to take several air flow measurements (i.e. traverses) and get a total average, for example measuring flow rate from multiple ducts to get an overall average.
The value of each point is the average of TRIAL(1) to TRIAL(N). The value of each location is the average of POINT(1) to POINT(M). The value of each location will be stored in the memory.
Press to enter the Main Menu.
Use to select “4. FLOW RATE”
and press .
Use to select
2. MULTI FLOW RATE” and press
.
As the same as the SINGLE mode, set
the following: 1. Sampling Time, 2. No.
Trial, 4. Data Storage, and 5. Duct
Entry No.
Use to select
3. MEAS./LOC.” and press .
Use to select the number of
MEAS. points (1 to 999) and press .
Then use to select the
number MEAS. locations (1 to 999) and
press .
MULTI (R20000/20000
1.SAMPLING(TIME)001ps
2.SAMPLING(n)110050
3.MEAS./LOC.oo010/010
4.DATA2STORAGE2?22YES
5.DUCT2ENTRY2NO.2221
6.SET2TO2START
SAMPLING MODE
1.SINGLEoFLOWoRATE
2.MULTIoFLOWoRATE
LOCATION(1) :AVG(1)=ΣPOINT(i)M LOCATION(2) :AVG(2)=ΣPOINT(i)M
・・・
LOCATION(L) :AVG(L)=ΣPOINT(i)M
<Result>
Average(AVG): AVG(AVG(1)+ AVG(2)+ ・・・・+AVG(L))L Maximum(MAX): MAX
Maximum value from AVG(1) to AVG(L)
Minimum(MIN): MINMinimum value from AVG(1) to AVG(L)
POINT(1) avg(1)=ΣTRIAL(i)N
POINT(2) avg(2)=ΣTRIAL(i)N
・・・
POINT(M) :avg(M)=ΣTRIAL(i)/N
MENUaa1.NORMAL
aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO
actoolsupply.com
actoolsupply.com
30 Measurement of Air Flow FLOW RATE Mode
'12/04/12a16:11:59 ~
RDY.
P
2221
N
2221
L
2221
C
2300
---502a
27.1e
62.9g
<Save the Setting>
Use to select
“6. SET TO START” and press .
If you press before saving the settings you will return back to the Main Menu and any setting changes you made will not be saved.
<Ready>
The ready screen will be displayed.
Use , while pressing to
change the entry number of the duct
type. Press to start measuring.
Follow the same procedure as SINGLE
mode to measure.
Display Icons
1. RDY: Current Status (READY/SAMPLE)
2. P 1: Current Number of Points
3. N 1: Current Number of Trial
4. L 1: Current Number of Location
5. C300: Duct Shape and Size (In this case, the duct is circular and 300mm in diameter.)
For the rectangular duct, W*** and H*** will be displayed.
<Display the Results>
After all the measurements at all locations are completed, the calculation results will be displayed.
Press to display the results of
each item in the following sequence:
Flow rateAir TempHumidity
Air velocity
Press to print the results, when a printer is connected.
Press to return to the Main Menu.
Press to return to the setup screen of FLOW RATE Mode.
*When “DATA STORAGE” is set to YES, values of each location will
be stored.
MULTI (R20000/20000
1.SAMPLING(TIME)001ps
2.SAMPLING(n)110050
3.MEAS./LOC.oo010/010
4.DATA2STORAGE2?22YES
5.DUCT2ENTRY2NO.2221
6.SET2TO2START
① ② ③ ④ ⑤
aNEXT1FLOW2RATE
MAX
AVG
MIN
106433a
085134a 066797a
actoolsupply.com
actoolsupply.com
31
§ 6 Store and Recall Measured Data
Data which can be stored in the memory of the instrument is shown below.
What can be Stored
MODEL
Measuring
Mode
CALCULATION Mode
FLOW RATE Mode
6531 6533
Display
V T H
W
T
H
*PV T
H
W
T H
Stored
Parameters
V T H
V, W
T, H
*P
V, W
T, H
V, W
T, H
6541 6542 6543 6561
Display
V, T
W, T
*P
V, T
W, T
Stored
Parameters
V, T V W
T
*PV W
T
V
W
T
6551 6552
Display
V W *PV W
Stored
Parameters
V
V, W
*P
V, W
V, W
To Recall Stored Data
Press to enter the Main Menu.
Use to select “5. DATA OUTPUT”
and press .
Use to select “1. DISPLAY”
and press .
V: Air Velocity W: Flow Rate T: Air Temperature H: Humidity
*P: Pressure is only available on MODEL 6501-B0/6501-C0
DATA OUTPUT
1.DISPLAY
2.PRINTER(SINGLE)
3.PRINTER(MULTI)
MENUaa1.NORMAL
aaaaaa
2.DUCTaTYPE
aaaaaa
3.CALCULATION
aaaaaa
4.FLOWaRATE
aaaaaa
5.DATAaOUTPUT
aaaaaa
6.DATAaCLEAR
aaaaaa
7.UTILITY
aaaaaa8.PRESSUREaZERO
actoolsupply.com
actoolsupply.com
32 Store and Recall Measured Data
<Setting the PAGE>
Use to select the page that
you would like to recall and press .
<Displaying the Recalled Data>
The recalled data of the selected page will be displayed.
Use to scroll.
If you measured Flow Rate in CALCULATION Mode or if you
measured in FLOW RATE Mode, you can select Flow Rate or Air
Velocity to be displayed by pressing .
Only Pressure will be displayed if you measured Pressure.
Humidity is only
available on
MODEL6531/6533.
Air temperature is
only available on MODEL6531/6541 /6542/6533/6543/
6561.
……………………… Page Number
………… Measured Mode (CALCULATION (A): AVERAGE/(I): INSTANT ) ………………… Measured Date (Year/Month/Day) ………………… Measured Time (Hour : Minute : Second)
……… Number of Trial
……………………… Duct Shape/Size (For Air Flow Measurement only)
……………………… Page Number
………… Measured Mode ………………… Measured Date (Year/Month/Day) ………………… Measured Time (Hour : Minute : Second)
………
Number of Trial/Number of Measuring Point
……………………… Duct Shape/Size
CALCULATION
SINGLE FLOW RATE
……… Calculation Range
……
Data Number, Air Velocity, Air Temperature, Humidity
START:00100END:050
NUM.
0m/s0000°C000%RH
001
000.810025.400
64.7
002
000.950025.400
64.7
003
000.980025.600
64.9
004
001.050025.700
65.1
005
001.210025.700
65.0
006000.990025.90065.2
……………………… Page Number
………… Measured Mode ………………… Measured Date (Year/Month/Day) ………………… Measured Time (Hour : Minute : Second)
………
Number of Trial/Number of Measuring Location
……………………… Duct Shape/Size
22222DAPAGEOSET
PAGE:0001 MODE:CALCULATION(A) DATE:2012/04/12 TIME:2213:40:53 MEAS./LOC.:2220/2ooo­SHAPE:C0ooo-
22222DAPAGEOSET
PAGE:0002 MODE:SINGLEoFLOW2RATE DATE:2012/04/12 TIME:2215:42:53 DATA/DIV.:oooo3/oo210 SHAPE:C0900
22222DAPAGEOSET
PAGE:0003 MODE:MULTIoFLOW2RATE DATE:2012/04/12 TIME:2215:45:53 MEAS./LOC.:222o3/210 SHAPE:C0900
MULTI FLOW RATE
Pressure is only
available on MODEL6501-B0
/-C0.
actoolsupply.com
actoolsupply.com
33
<Setting the CALCULATION RANGE>
You can select the range of calculation.
If the data range is correct, press to calculate the range of data
displayed on the screen.
If the range is not correct, press to
display a cursor on “START”.
Use to select the first number of
the data (START) that is to be calculated
and press .
The cursor will move to “END”. Use to select the number of
the last data (END) that is to be calculated and press .
Press to calculate for the displayed
data range.
*You CANNOT set more than one range.
<In FLOW RATE Mode>
Press to shift the calculation results in the following
sequence: Flow RateAir TemperatureHumidityAir Velocity.
<In CALCULATION Mode>
Press to shift the calculation results in the following sequence:
Air Velocity (Flow Rate)Air TemperatureHumidityDew Point Temperature, Discomfort Index.
Press to return to PAGE SET.
Press to return to the MAIN MENU.
START:00100END:050
NUM.
0m/s0000°C000%RH
001
000.810025.400
64.7
002
000.950025.400
64.7
003
000.980025.600
64.9
004
001.050025.700
65.1
005
001.210025.700
65.0
006000.990025.90065.2
aNEXT1FLOW2RATE
MAX
AVG
MIN
106433a
085134a 066797a
actoolsupply.com
actoolsupply.com
34 Print Out
§ 7 Print Out
Preparation for Printing
Connect the printer cable to the Serial Communication terminal located on the side of the instrument to print out measurement data.
<Requirements>
Printer (optional)……Recommended: Seiko Instruments Model
DPU-S245
Printer Cable (optional)
<Check the BAUD RATE>
You need to enter the same baud rate and data transmission conditions on both the Main Body and the printer. The factory setting of the Main Body is as follows:
<Connecting the Printer>
1. Connect the printer to the Main Body (Serial Communication Terminal) using the printer cable.
2. Turn ON the CLIMOMASTER first, and then turn the printer ON.
3. Make sure that the CLIMOMASTER is displaying NORMAL Mode.
Print Out in NORMAL Mode
When the screen is displaying in
NORMAL Mode, press to HOLD the
display. Press to print out the data
on the screen.
* If the printer is not connected properly, ”PERR” will be displayed in
the lower left of the display.
Examples of Print Out
Air Velocity Mode Pressure Mode
Data Bit Length
8 bits
Parity
None
Baud Rate
Set Value
Stop Bit
1
'12/04/12a15:40:23 ~
TC
C1
HOLD
0.87f
25.3e
64.1g
*To change the BAUD RATE, refer to
“Units and Baud Rate” on P.
46.
*
To configure the settings of the
printer, refer to the printer operation manual.
2001/06/19 13:42:09 Velocity 0.12 m/s Temperature 25.9 C Humidity 62.7 %RH
Air VelocityAir Temperature
2001/06/19 13:42:28 Pressure 0.23 kPa
Serial
Communication
Terminal
Humidity is only
available on
MODEL6531/6533.
Air temperature is
only available on MODEL6531/6541 /6542/6533/6543/
6561.
Pressure is only
available on MODEL6501-B0
/-C0.
actoolsupply.com
actoolsupply.com
35
Print Out in CALCULATION Mode and FLOW RATE Mode
<Measurement and Calculation Mode (CALCULATION)>
Press after the measurement and
calculation has been performed to print
out the result.
<Airflow Measurement Mode (FLOW RATE)>
Press after the measurement and
calculation has been performed to print
out the result.
Examples of Print Out
Stored Location
Measurement Mode
Measured Date
Measured Time
No. of Data Records
No. of Points Measured
Sampling Time
Calculation Range
Duct Type
Air Velocity
Flow Rate
Air Temperature
Humidity
Dew Point Temperature
Discomfort Index
Air Velocity
PAGE:004 MODE:CALCULATION(A)
DATE:2001/06/19 TIME:17:24:33 DATA:005 SAMPLING TIME:001 START:001 END:005 MAX 1.26 m/s AVG 0.90 m/s MIN 0.55 m/s MAX 25.6 C AVG 25.6 C MIN 25.5 C
MAX 64.6 %RH
AVG 64.5 %RH MIN 64.4 %RH DT 18.4 C
DI 74.1
PAGE:002
MODE:FLOW RATE DATE:2001/06/19 TIME:13:35:23 DATA:003 DIV.:002 SAMPLING TIME:001 START:001 END:002 SHAPE D 900mm MAX 361554 m3/h AVG 220582 m3/h MIN 79609 m3/h MAX 25.5 C
AVG 25.4 ゚ C
MIN 25.4 C MAX 63.7 %RH AVG 62.6 %RH MIN 61.4 %RH MAX 1.58 m/s AVG 0.96 m/s
MIN 0.35 m/s
aNEXCALCULATION
MAX
AVG
MIN
102.76f
081.43f
060.81f
aNEXT1FLOW2RATE
MAX
AVG
MIN
106433a
085134a 066797a
actoolsupply.com
actoolsupply.com
36 Print Out
Print Out Stored DataSingle Page
Press to enter the Main Menu.
Use to select “5. DATA OUTPUT”
and press .
Use to select
“2. PRINTER(SINGLE)” and press .
Use to select the page that you
would to print out and press .
The data you have selected will be displayed.
Press to scroll.
If you measured Air Flow in CALCULATION mode or in FLOW RATE
mode, you can select Air Flow or Air Velocity to be displayed by
pressing .
*If you measured Pressure, only Pressure will be displayed.
<Setting the range of calculation.>
You can select the range of calculation. If you are not going to change the data range, press to
calculate all the data.
Press . Cursor will appear on
“START”.
Use to select starting point.
Press . Cursor will move to “END”.
You CANNOT set more than one range.
START:00100END:050
NUM.
0m/s0000°C000%RH
001
000.810025.400
64.7
002
000.950025.400
64.7
003
000.980025.600
64.9
004
001.050025.700
65.1
005
001.210025.700
65.0
006000.990025.90065.2
START:00100END:050
NUM.
0m/s0000°C000%RH
001
000.810025.400
64.7
002
000.950025.400
64.7
003
000.980025.600
64.9
004
001.050025.700
65.1
005
001.210025.700
65.0
006000.990025.90065.2
MENUaa1.NORMAL
aaaaaa
2.DUCTaTYPE
aaaaaa
3.CALCULATION
aaaaaa
4.FLOWaRATE
aaaaaa
5.DATAaOUTPUT
aaaaaa
6.DATAaCLEAR
aaaaaa
7.UTILITY
aaaaaa8.PRESSUREaZERO
DATA OUTPUT
1.DISPLAY
2.PRINTER(SINGLE)
3.PRINTER(MULTI)
……………………… Page Number
……… Measurement Mode (CALCULATION or FLOW RATE)
……………… Measurement Date (Year/Month/Day)
…………… Measured Time (Hour : Minute : Day)
……… Number of Trial/Number of Measuring Point
……………………… Duct Shape/Size (for Air Flow Measurement only)
………… Calculation Range
……… Data Number, Air Velocity, Air Temperature, Humidity
Humidity is only
available on
MODEL6531/6533.
Air temperature is
only available on MODEL6531/6541 /6542/6533/6543/
6561.
Humidity is only
available on
MODEL6531/6533.
Air temperature is
only available on MODEL6531/6541 /6542/6533/6543/
6561.
Pressure is only
available on MODEL6501-B0
/-C0.
22222DAPAGEOSET
PAGE:0002 MODE:SINGLEoFLOW2RATE DATE:2012/04/12 TIME:2215:42:53 DATA/DIV.:oooo3/oo210 SHAPE:C0900
actoolsupply.com
actoolsupply.com
37
Press to select the contents of the
Print Out.
Use to select from items 1 to 3
and press to print out.
Please refer to the printing example shown below.
1. RESULT… Conditions and Calculation Results
2. DATA…… Conditions and Stored Data
3. ALL……… Conditions, Calculation Results and Stored Data
Example of Print Out
Measurement
Data
PAGE SET
PAGE:001 MODE:CALCULATION(A) DATE:2001/05/19 TIME:11:32:26 DATA:010 SAMPLING TIME:001 START:001 END:010
MAX 0.75 m/s
AVG 0.40 m/s MIN 0.05 m/s MAX 25.8 C AVG 25.6 C MIN 25.4 C MAX 59.1 %RH AVG 58.2 %RH MIN 57.5 %RH DT 16.8 C DI 73.4 NUM. m/s C %RH
001 0.05 25.8 57.5
002 0.06 25.8 57.6 003 0.48 25.6 57.7 004 0.48 25.6 57.8 005 0.56 25.6 58.0 006 0.48 25.6 58.3 007 0.39 25.5 58.5 008 0.75 25.5 58.7 009 0.27 25.4 58.8
010 0.46 25.4 59.1
PAGE SET
PAGE:011 MODE:FLOW RATE DATE:2001/06/21 TIME:16:23:08 DATA:003 DIV.:005 SAMPLING TIME:001
START:001 END:005
SHAPE W 200 H 300mm MAX 32194 m3/h AVG 16871 m3/h MIN 5446 m3/h MAX 25.9 C AVG 25.9 C MIN 25.8 C MAX 72.2 %RH AVG 71.2 %RH MIN 69.7 %RH MAX 1.49 m/s
AVG 0.78 m/s
MIN 0.25 m/s NUM. m3/h C %RH
001 20596 25.8 69.7 002 19906 25.9 72.2 003 6214 25.9 71.3 004 5446 25.9 71.4
005 32194 25.9 71.4
Conditions
(Always
printed out)
Calculation
Results
(RESULT)
Humidity is only available on MODEL6531/6533.
Air temperature is only available on MODEL6531/6541/6542/6533/6543/6561.
Pressure is only available on MODE 6051-B0/6051-C0
AaaaaPRINT OUTPUT
1
.RESULT
2.DATA
3.ALL
actoolsupply.com
actoolsupply.com
38 Print Out
Print Out Stored DataPlural Pages
Press to enter the Main Menu.
Use to select “5. DATA OUTPUT”
and press .
Use to select
3. PRINTER(MULTI) and press .
Use to select
1.SELECT and press .
Use to select the starting page.
Press .
Use to select the ending page.
Press .
Use to select “YES” for
selecting the content of the Print Out.
Press .
Use to select from items 1 to 3
and press to print out.
Use to select
2.ALL DATA and press , and use
to select “YES” for printing out
all pages. Press .
ooDATAoOUTPUT(MULTI)
1.SELECToSTART:o0001
oooooooooENDoo:o0003 oooooooooSEToo:oooNO
2.ALLoDATAooooooooNO
oooooooooR00020/20000
DATA OUTPUT
1.DISPLAY
2.PRINTER(SINGLE)
3.PRINTER(MULTI)
ooDATAoOUTPUT(MULTI)
1.SELECToSTART:o0001 oooooooooENDoo:o0003 oooooooooSEToo:oooNO
2.ALLoDATAooooooooNO
oooooooooR00020/20000
MENUaa1.NORMAL
aaaaaa
2.DUCTaTYPE
aaaaaa
3.CALCULATION
aaaaaa
4.FLOWaRATE
aaaaaa
5.DATAaOUTPUT
aaaaaa
6.DATAaCLEAR
aaaaaa
7.UTILITY
aaaaaa8.PRESSUREaZERO
ooDATAoOUTPUT(MULTI)
1.SELECToSTART:o0001
oooooooooENDoo:o0003 oooooooooSEToo:oooNO
2.ALLoDATAooooooooNO
oooooooooR00020/20000
ooDATAoOUTPUT(MULTI)
1.SELECToSTART:o0001
oooooooooENDoo:o0003 oooooooooSEToo:oooNO
2.ALLoDATAoooooooYES
oooooooooR00020/20000
ooDATAoOUTPUT(MULTI)
1.SELECToSTART:o0001
oooooooooENDoo:o0003 oooooooooSEToo:ooYES
2.ALLoDATAooooooooNO
oooooooooR00020/20000
AaaaaPRINT OUTPUT
1
.RESULT
2.DATA
3.ALL
actoolsupply.com
actoolsupply.com
39
§ 8 Digital Output
Preparation
You can download the data stored in the CLIMOMASTER to your PC, by connecting the CLIMOMASTER to your PC with the USB cable.
Requirements
Personal Computer
USB Cable (supplied)
Communication Software
(such as Hyper Terminal for Windows)
Connecting PC
1. Connect the PC to the Climomaster using the USB cable.
2. Turn ON the CLIMOMASTER.
3. Make sure that the CLIMOMASTER is displaying NORMAL Mode.
Measurement Software for Windows is available.Optional
To Access CLIMOMASTER from Your PC
<Icons and their Meaning>
Space
Press Enter
* :A Number
*All inputs need to be capital letters.
Command
Function
****
Number of Downloading Data
Cancel
Output the Measuring Condition (of On-Time Data)
Output the Measuring Units
Output the Duct Shape/Size
Output the Page Number
****
Output the Stored Data
****
Output the Measuring Condition (of Stored Data)
Output the Measuring Condition of All Pages
Serial Communication
Terminal
actoolsupply.com
actoolsupply.com
40 Digital Output
Transmission of On-Time Data (Measured at every 1sec)
<Enter the Number of Data Needed>
1. Type **** ”.
*Type in the number of data(4 digits) after D”.
2. After “AD”, the data will be displayed.
3. Each data represents 1 second of measurement. If you ask for 20 data, it takes approximately 20 seconds to display.
4. The maximum number that can be set is 9999. If you need more, re-send the command
Output Content
Air Velocity Mode: Velocity; Temperature; Humidity
Humidity Flow Rate Mode: Flow Rate; Temperature; Humidity
Pressure Mode: 0000000; 0000000; Pressure
<To Cancel>
Type N ”.
<To Download the Measuring Conditions>
1. Type “S ”
2. After “AS”, the data will be displayed as described below.
Output Content
Measuring Parameter (WTH, VTH or PRS); Time Constant (P. 20); Duct Size (Width, Height, Diameter); Units of the Duct Size
WTH: Flow Rate, Temperature, Humidity
VTH: Air Velocity, Temperature, Humidity
PRS: Pressure
<To Download the Measuring Units>
1. Type U
2. After “AU”, the data will be displayed.
Output Content
Velocity Unit; Temperature Unit; Humidity Unit; Pressure Unit; Flow Rate Unit
”0000000” will be
displayed for humidity on MODEL6541/
6542/6543/6561.
”0000000” will be
displayed for temperature and humidity on MODEL6551 and 6552
Example: (Air Velocity Mode)
typed 0004
Example
Example
Example
Pressure is only
available on
MODEL6501-B0/-C0.
actoolsupply.com
actoolsupply.com
41
Transmission of Stored Memory
<To Download the Duct Setting>
1. Type K
2. After “AK”, the data will be displayed.
3. All stored duct settings
(1 through 25) will be downloaded.
Output Content
Entry No. (Data Location); Width; Height; Diameter; Units of Duct Size
<To Download the Page Number>
1. Type P”.
2. After “AP”, the data will be displayed.
3. Number of data stored pages will be downloaded.
<To Download the Stored Data>
1. Type T**** ”.
* Type in the desired page number(4 digits) after “T”.
2. After “AT”, the data will be displayed.
Output Content
Air Velocity Mode: Number of Data; Velocity; Temperature; Humidity
Flow Rate Mode: Number of Data; Flow Rate; Temperature; Humidity
Pressure Mode: Number of Data; 0000000; 0000000; Pressure
*The data will be displayed in the currently selected units.
NOT in the units the data was saved with.
*Calculated data will not be downloaded.
The format of the date
output is fixed to
Year/Month/Day. This is not affected by date
setting of the main body.
Example
Example
AP P0011
Example
”0000000” will be
displayed for humidity on MODEL6541, 6542, 6543 and
6561.
”0000000” will be
displayed for temperature and humidity on MODEL6551 and
6552
Pressure is only
available on
MODEL6501-B0/-C0.
actoolsupply.com
actoolsupply.com
42 Digital Output
<To Download the Measuring Condition of Stored Data>
1. Press “M**** ”.
*Type in the desired page number(4 digits) after “M”.
2. After “AM”, the data will be displayed.
Output Content
A. Measuring Parameter
WTH: Flow Rate; Temperature; Humidity
WT: Flow Rate; Temperature VTH: Velocity; Temperature; Humidity VT: Velocity; Temperature PRS: Pressure
B. Measuring Mode
000: Calculation Mode 001: Flow Rate Mode
C. Sampling Time D. Number of Data E. Calculation Mode
AVG: Average
INS: Instantaneous In the case of Flow Rate Mode, the number of points is indicated.
F. Width G. Height
H. Diameter I. Units of Duct Size
<To Download the Measuring Conditions of All Pages>
1. Press “B ”
2. After “AB”, the data will be displayed.
Output Content
*Same as shown above.
<Error Message>
Re-type the command.
A B C D E F G H I
Example
Example
Example
Pressure is only
available on MODEL6501-B0
/-C0.
actoolsupply.com
actoolsupply.com
43
§ 9 Analog Output
1. Data Update……every 0.1 sec
(Except Humidity; updated every 1sec)
2. Load Impedance……Above 5KΩ
3. Output Voltage……DC 0 to 1V
For analog output, you must select one setting from the table below. (There is no analog output for Flow Rate)
Parameter
Output Range
Conversion Formula
(V: voltage)
Velocity
(U)
0 to 5 m/s
U= 5× m/s
0 to 10 m/s
U= 10×m/s
0 to 30 m/s
U= 30×m/s
0 to 50 m/s
U= 50x V m/s
0 to 1000 FPM
U= 1000×FPM
0 to 2000 FPM
U= 2000×FPM
0 to 6000 FPM
U= 6000× V FPM
0 to 10000 FPM
U= 10000×FPM
Humidity
(H)
0 to 50 %RH
H= 50× %RH
0 to 100 %RH
H= 100× %RH
Temperature
(T)
-20 to 30 °C
T= 50×V-20 °C
0 to 50 °C
T= 50× °C
0 to 100 °C
T= 100× °C
-4 to 86 oF
T= 90× - 4
o
F
32 to 122 oF
T= 90×V+32
o
F
32 to 212 oF
T= 180×V+32
o
F
Pressure
(P)
-2 to +2 kPa
P= 4×- 2 kPa
-5 to +5 kPa
P= 10×- 5 kPa
For the output range, the low end will be set at 0V and the high end will be set at 1V. The voltage is linear. The Time Constant will be in effect.
Mode
Ways To Take In Measured Data (Analog Output
Explanation
TC1
Take 10 measurements each second and indicate the average as an instantaneous value at every 1 sec.
TC5
Output the average measured value for 5 seconds at every
0.1 second. Data shifts by 0.1 second.
TC10
Output the average measured value for 10 seconds at every
0.1 second. Data shifts by 0.1 second.
Analog Output
Terminal
0 5 10 15 20sec
Measuring time
0 0.5 1.0 1.5 2.0sec 0 2.5 5.0 7.5 10sec
Average of 10sec
Average of 5sec
Average of 1 sec
Analog output is only
available on
MODEL6501-A0/-C0.
To change the setting
of Time Constant, refer to To change Time Constant on
P. 20
.
actoolsupply.com
actoolsupply.com
44 Analog Output
Press to enter the Main Menu.
Use to select “7. UTILITY”.
Press .
Use to select
3. ANALOG OUTPUT”.
Press
<To Select Data>
Select “1. OUTPUT SELECT” and press
.
Use to select “VEL.” (Velocity), “PRS.” (*Pressure),
“HUM.” (*Humidity), or “TMP.” (*Temperature).
Press .
<To Set the Output Range>
Use to select “2.” (selected
parameter) and press
.
Use to select the range and
press
.
<To Save the Settings>
Use to select “3. SAVE INFO”.
Press .
If you press before you save, you will return to the Main Menu
and the settings will not be saved.
MENUaa1.NORMAL
aaaaaa
2.DUCTaTYPE
aaaaaa
3.CALCULATION
aaaaaa
4.FLOWaRATE
aaaaaa
5.DATAaOUTPUT
aaaaaa
6.DATAaCLEAR
aaaaaa
7.UTILITY
aaaaaa8.PRESSUREaZERO
AaaANALOG OUTPUT
1.OUTPUT
2SELECT22V
EL.
2.VEROCITY
22222222222
0-1022m/s
3.SAVE
2INFO
AaaANALOG OUTPUT
1.OUTPUT
2SELECT22
VEL.
2.VEROCITY
22222222222
0-1022m/s
3
.SAVE2INFO
aaaaaaaUTILITY
1.TIME AJUST
2.UNIT SELECT
3
.ANALOG OUTPUT
4.TC(T.H.P) SET
oooooooUTILITY
1.TIMEoADJUST
2.UNIToSELECT
3.ANALOGoOUTPUT
4.TC(T.H.P)oSET
5.BACKLIGHToON/OFF
AaaANALOG OUTPUT
1.OUTPUT
2SELECT22
VEL.
2.VEROCITY
22222222222
0-3022m/s
3.SAVE
2INFO
Pressure is only
available on MODEL6501-B0 /-C0.
Humidity is only
available on
MODEL6531/6533
Air temperature is
only available on MODEL6531/6541 /6542/6533/6543/
6561.
actoolsupply.com
actoolsupply.com
45
§ 10 Other Settings
Date
Press .
Use to select “7. UTILITY” and press .
Use to select “1. TIME ADJUST” and press .
Use to select "1. STYLE" or
"2. DATE" and press
.
1. STYLE: Select JP, US or EU
Japanese style (JP): YYYY/MM/DD
US style (US): MM/DD/YYYY EU style (EU): DD/MM/YYYY
2. DATE: Date
3. TIME: hour/minute/sec
Use to select the desired item.
Use to change the setting.
Press to select the setting and
move the cursor back to “1” or “2”.
To Save the Setting
Use to select “4. SAVE INFO” and press to save the change. The screen returns to the Main Menu.
*When the date is changed, the dates of existing stored
measurements will change as well.
If you press before you save, you will return to the Main Menu and the settings will not be saved.
AaaaaTIME2ADJUST
1.STYLE:
222JP
2.DATE
2:22012/04/12
3.TIME
2:22214:37:27
4.SAVE2INFO
AaaaaTIME2ADJUST
1.STYLE:
222JP
2.DATE
2:22012/04/12
3.TIME
2:22214:37:27
4.SAVE2INFO
oooooooUTILITY
1.TIMEoADJUST
2.UNIToSELECT
3.ANALOGoOUTPUT
4.TC(T.H.P)oSET
5.BACKLIGHToON/OFF
AaaaaTIME2ADJUST
1
.STYLE:222JP
2.DATE
2:22012/04/12
3.TIME
2:22214:37:27
4.SAVE2INFO
MENUaa1.NORMAL
aaaaaa
2.DUCTaTYPE
aaaaaa
3.CALCULATION
aaaaaa
4.FLOWaRATE
aaaaaa
5.DATAaOUTPUT
aaaaaa
6.DATAaCLEAR
aaaaaa
7.UTILITY
aaaaaa8.PRESSUREaZERO
*The style of date
displayed on the
screen or for printer
output depends on this setting, however, the style of the output to digital port is fixed to Japanese style.
actoolsupply.com
actoolsupply.com
46 Other Settings
22222DUNITASELECT
1
.VELOCITY2222222m/s
2.TEMPERATURE
22222°C
3.FLOW
2RATE222m³/h
4.PRESSURE
2222222kPa
5.BAUD RATE
222219200
6.SAVE INFO
22222DUNITASELECT
1.VELOCITY
2222222m/s
2.TEMPERATURE
22222°C
3.FLOW
2RATE222m³/h
4.PRESSURE
2222222kPa
5.BAUD RATE
222219200
6.SAVE INFO
22222DUNITASELECT
1.VELOCITY
2222222m/s
2.TEMPERATURE
22222°C
3.FLOW
2RATE222m³/h
4.PRESSURE
2222222kPa
5.BAUD RATE
222219200
6.SAVE INFO
Units and Baud Rate
Press to enter the Main Menu.
Use to select “7. UTILITY”.
Press .
Use to select “2. UNIT SELECT”
and press .
<Units and Baud Rate>
Use to select the desired item (1 through 5). Press .
Use to change the setting.
Press .
<To Save the Setting>
Use to select “6. SAVE INFO”
and press . The change will be saved
and the screen will return to the Main
Menu.
If you press before you save, you will return to the Main Menu
and the settings will not be saved.
oooooooUTILITY
1.TIMEoADJUST
2.UNIToSELECT
3.ANALOGoOUTPUT
4.TC(T.H.P)oSET
5.BACKLIGHToON/OFF
MENUaa1.NORMAL
aaaaaa
2.DUCTaTYPE
aaaaaa
3.CALCULATION
aaaaaa
4.FLOWaRATE
aaaaaa
5.DATAaOUTPUT
aaaaaa
6.DATAaCLEAR
aaaaaa
7.UTILITY
aaaaaa8.PRESSUREaZERO
<Units Conversion Table>
Velocity 1m/s = 196 FPM Temperature T(oF) = 1.8×T(°C+32
Flow Rate 1m3/h = 35.32ft3/h
………m/s or FPM for velocity
………°C or oF for temperature
………m3/h, m3/min, ft3/min, or ft3/h for flow rate
………kPa or Pa for pressure
………4800, 9600, 19200, or 38400 baud rate
Air temperature is
only available on MODEL6531/6541 /6542/6533/6543/
6561.
Pressure is only
available on MODEL6501-B0
/-C0.
actoolsupply.com
actoolsupply.com
47
Settings for LCD Backlight
Press .
Use to select “7. UTILITY”.
Press .
Use to select
“5. BACKLIGHT ON/OFF”.
Press .
Use to select “1. BACKLIGHT”.
Press .
Use to select
BACKLIGHT ON/OFF. Press .
BACKIGHT OFFBacklight is always off. BACKLIGHT ONIf you press any button, the backlight will turn on
according to the following settings:
<Settings for Backlight Lit-Up Time>
Use and set the amount of
time which the backlight will stay on.
ALWAYS: The backlight is always on. FOR ***: The backlight is on for the specified amount of time (***).
e.g.) FOR 2min: It’s on for 2 minutes
<Saving the Change>
Use and select “6.SAVE INFO”
Press .
If you press before you save, you will return to the Main Menu
and the settings will not be saved.
ooooLCDoBACKLIGHT
1.BACKLIGHToON
ooooFORoooooooo01omin
2.SAVEoINFO
ooooLCDoBACKLIGHT
1.BACKLIGHToON
ooooFORoooooooo01omin
2.SAVEoINFO
ooooLCDoBACKLIGHT
1.BACKLIGHToON
ooooFORoooooooo01omin
2.SAVEoINFO
oooooooUTILITY
1.TIMEoADJUST
2.UNIToSELECT
3.ANALOGoOUTPUT
4.TC(T.H.P)oSET
5.BACKLIGHToON/OFF
MENUaa1.NORMAL
aaaaaa
2.DUCTaTYPE
aaaaaa
3.CALCULATION
aaaaaa
4.FLOWaRATE
aaaaaa
5.DATAaOUTPUT
aaaaaa
6.DATAaCLEAR
aaaaaa
7.UTILITY
aaaaaa8.PRESSUREaZERO
actoolsupply.com
actoolsupply.com
48 Other Settings
To Delete Data
Press to enter the Main Menu.
Use to select “6. DATA CLEAR”
and press .
Use to select “1. CLEAR” (delete
the selected data only) and press .
Use to select the starting page.
Press .
Use to select the ending page.
Press .
Use to select “YES” for partial
delete.
Press .
The selected page will be deleted and the remaining data will shift up. (See diagram below)
MENUaa1.NORMAL
aaaaaa
2.DUCTaTYPE
aaaaaa
3.CALCULATION
aaaaaa
4.FLOWaRATE
aaaaaa
5.DATAaOUTPUT
aaaaaa
6.DATAaCLEAR
aaaaaa
7.UTILITY
aaaaaa8.PRESSUREaZERO
AaaaaDATARCLEAR
1
.CLEARLSTART:000001
00000000
END00:000001
00000000
SET00:002YES
2.ALL
2CLEAR0000000NO
000000000R19950/20000
AaaaaDATARCLEAR
1.CLEAR
LSTART:000001
00000000
END00:000001
00000000
SET00:002YES
2.ALL
2CLEAR0000000NO
000000000R19950/20000
AaaaaDATARCLEAR
1.CLEAR
LSTART:000001
00000000
END00:000006
00000000
SET00:002YES
2.ALL
2CLEAR0000000NO
000000000R19950/20000
AaaaaDATARCLEAR
1.CLEAR
LSTART:000001
00000000
END00:000006
00000000
SET00:002YES
2.ALL
2CLEAR0000000NO
000000000R19950/20000
Page 1
Page 2
Page 4
Page 1
Page 2
Page 3
……Starting Page Number
…… Ending Page Number …… Partial Delete (YES or NO) …… All Delete YES or NO
…… Remaining Memory / Total Memory
(The number of sampling data that can be stored)
Page Number will change
automatically
Page 1
Page 2
Page 3
Page 4
actoolsupply.com
actoolsupply.com
49
To Delete All the Data
Press to enter the Main Menu.
Use t to select “6. DATA CLEAR”.
Press .
Use to select “2. ALL CLEAR”. Press .
Use to select “YES” to delete.
Press .
All the data will be deleted and the Remaining Memory will be 20000.
AaaaaDATARCLEAR
1.CLEAR
LSTART:000001
00000000
END00:000001
00000000
SET00:0020NO
2
.ALL2CLEAR0000000NO
000000000R19950/20000
AaaaaDATARCLEAR
1.CLEAR
LSTART:000001
00000000
END00:000001
00000000
SET00:0020NO
2.ALL
2CLEAR000000YES
000000000R19950/20000
AaaaaDATARCLEAR
1.CLEAR
LSTART:000001
00000000
END00:000001
00000000
SET00:0020NO
2
.ALL2CLEAR000000YES
000000000R20000/20000
MENUaa1.NORMAL
aaaaaa
2.DUCTaTYPE
aaaaaa
3.CALCULATION
aaaaaa
4.FLOWaRATE
aaaaaa
5.DATAaOUTPUT
aaaaaa
6.DATAaCLEAR
aaaaaa
7.UTILITY
aaaaaa8.PRESSUREaZERO
……All Delete (YES or NO)
……Remaining Memory / Total Memory
(The number of sampling data that can be stored)
actoolsupply.com
actoolsupply.com
50 Other Settings
Contrast Adjustment
If the LCD display of the CLIMOMASTER is too dark or too light, there is an adjustment dial at the back, bottom of the unit, inside the battery cover.
You can adjust the contrast by using a precision screwdriver (-) 0.9 to 1.5mm. Turn the dial clockwise to darken the LCD and counterclockwise to lighten it.
Contrast Adjusting Dial Lightens Darkens
actoolsupply.com
actoolsupply.com
51
§ 11 How to Clean a Probe
Dust or particles attached to the velocity sensor would alter the amount of heat diffusion, which leads to a less precise reading. Also, deformation or clogging of the protective mesh around the sensor of the probe would also affect the accuracy of the instrument. Users are encouraged to clean the probe regularly to maintain the accuracy of measurements.
Clean the sensor by soaking it in water in an ultrasonic cleaner for 10 to 20 seconds. Do not soak it for too long, since there is an increased risk of damaging the coating. Please use only water for cleaning the probe.
CAUTION
!) When cleaning, make sure that the power is turned OFF. !) Make sure that the sensor is dry before turning it ON. !) DO NOT let the humidity sensor come into contact with water
(MODEL6531/6533). When wet, let it air dry by placing it in an environment with less than 40%RH for more than 24 hours.
!) NEVER USE alcohol or any other organic fluid.
Alcohol will cause permanent damage to the humidity sensor. Once the sensor is damaged, it will need to be replaced even if it seems to be functioning.
Procedure
DO NOT place the sensor in an
upward position until it is completely dry as residual water
could damage the sensor.
Ultrasonic Cleaner for 10 to 20 seconds
Dry with Cool Air
actoolsupply.com
actoolsupply.com
52 Specification
§ 12 Specification
Product
CLIMOMASTER Air Velocity Meter
Model No. of the Main Body
6501
Model No. of the Probe
6531-21
6541
-21
6561
-21
6542
-21
6551
-21
6552
-21
6533-21
6543
-21
Measuring Object
Clean air flow
Air Velocity
[m/s]
Measuring Range
0.01 to 30.0, 0.01 to 50.0 (6561-21 only)
0.01 to 5.00
Resolution
0 to 9.99: 0.01 10 to 30: 0.1 30 to 50: 0.1 (6561-21 only)
0.01
Accuracy
±2% of reading or ±0.015 (±3FPM),
whichever is greater
0.01 to 0.99:±0.02
0.99 to 5.00: ±2% of reading or ±0.015 (±3FPM), whichever is greater
Response Time
[sec.]
Approx. 1
Approx. 4
Approx. 7
Approx. 7
(at 1m/s, 90% response time)
Air
Temperature
[oC]
Measuring Range
-20.0 to 70.0
-20.0 to 70.0
Resolution
0.1
0.1
Accuracy
±0.5
±0.5
Response Time
[sec.]
Approx. 30
(at 1m/s, 90% response time)
Approx. 30
(at 1m/s, 90%
response time)
Humidity
*1
[%RH]
Measuring Range
2.0 to 98.0
2.0 to 98.0
Resolution
0.1 - 0.1
Accuracy
2 to 80:
±2.0
80 to 98:
±3.0
2 to 80:
±2.0
80 to 98:
±3.0
Response Time
[sec.]
Approx.
15
Approx.
15
Pressure
※2
[kPa]
Measuring Range
-5.00+5.00
Resolution
0.01
Accuracy
±3% of reading or ±0.01
Response Time
[sec.]
Approx. 1
Temperature Compensation
[m/s]
±5% of reading or ±0.015 (±3FPM), whichever is greater (in the temperature range of 5 to 60.0oC)
Functions
Readings Hold, Max Hold, Time Constant (1, 5, 10 sec.), Battery Level Indicator (5 Levels), Unit selection (Air Velocity: m/s or FPM, Flow Rate: m3/min, m3/h, ft3/min, or ft3/h, Temperature*1: oC or oF, Humidity*2: %RH, Static Pressure*3: kPa or Pa) Max., Min., and Average Calculation, Sampling Time: 1 to 999 sec. No. of trails: 1 to 9999, Store up to 25 Duct Sizes (Square/Round, Size Range: 1 to 9999mm or 0.1 to 999.9inch(diameter, height, width)) Max.Data Storage: 20000
Output
Digital Output: USB (when connected to a printer, automatically switch to RS-232C, Baud Rate 4800, 9600, 19200 and 38400 bps) Analog Output*3: DC 0 to 1V (select one from Air Velocity, Temperature*1, Humidity*2 and Pressure*3)
Power
6 × AA Batteries , AC Adaptor*4: AC 100 to 240V (50/60Hz)
Battery Life
Approximately 10 continuous hours (at air velocity 5m/s, 20oC, with alkaline batteries)
Operating Environment
Main Body: 5 to 40oC (41 to 104oF) with no visible condensation
Probe: -20 to 70oC (-4 to 158 oF) with no visible condensation
Storage Environment
-10 to 50oC (14 to 122o F) with no visible condensation
Weight
Approximately 400g (batteries included)
Accessories
Carrying case, Operation manual, 6×AA Batteries, Probe cable (2m), USB cable
Options
Spare probe, Analog output, Pressure sensor, Extension rod, Printer, Printer cable, Software (for Windows), AC adaptor, Hands-free Case
*1: Air temperature is only available on MODEL6531/6541/6542/6533/6543/6561
*2: Humidity is only available on MODEL6531/6533 *3: Pressure is only available on MODEL6501-B0/-C0 , Analog Output is only available on MODEL6501-A0/-C0
*4: Optional.
actoolsupply.com
actoolsupply.com
53
§ 13 Principle of Measurement
Principle of Hot-wire Anemometer
The principle of the thermal Probe is based on a heated element from which heat is extracted by the colder impact flow. The temperature is kept constant via a regulating switch. The controlling current is directly proportional to the velocity. When thermal velocity Probes are used in turbulent flows, the measured result is influenced by the flows impacting the
heated body from all directions. In turbulent flows, a thermal velocity sensor indicates higher measured values than a vane Probe. This can be observed during measurements in ducts. Depending on the design of the duct, turbulent flows can occur even at low velocities. The amount of heat that is extracted by the colder impact flow from the sensor can be expressed by:
Where H: Heat diffusion quantity T: Temperature of the sensor Ta: Air temperature U: Air velocity a, b: Constant
Also, heat diffusion quantity can be expressed by the formula:
Where R is resistance and I is current (R is kept constant regardless of air velocity since the temperature in constant).
Therefore,
The Temperature Compensation
The air velocity sonsor is heated to an elevated temperature relative to the surrounding air by means of control electronics. The temperature compensation sensor senses the ambient, or surrounding air temperature and forces the velocity sensor to stay at a constant overheat above the ambient. The circuit forces the voltage to be equal by means of an operational amplifier. Air flowing past the sensor tends to cool the sensor, thus driving down its resistance. The amplifier responds by immediately delivering more power to the circuit to maintain voltage equilibrium. Delivered power is converted into an electrical signal to display.
Cooling
Current
Sensor Platinum Coil
Velocity m/s
Cu rrent
2
RIH
UbaRI
2
fomulasKingTaTUbaH    ・・・・・・ '))((
Current (i)
Velocity [U]
Ta1 < Ta2
Heat Diffusion
Quantity [H]
Ta
1
Ta2
Compensation
Heat Diffusion
Quantity [H]
Velocity [U]
))(( TaTUba
Wind
Air Velocity
Sensor
actoolsupply.com
actoolsupply.com
54 Principle of Measurement
Measurement of Wind-Temperature
(MODEL 6531/6541/6542/6533/6543/6561) The coefficient of resistance of the temperature sensor has a direct proportional relationship with the temperature. We can measure the wind temperature by adjusting the sensor’s temperature to the wind-temperature, and measuring its coefficient of resistance.
Measurement of Humidity (Hygrometer)
(MODEL6531/6533) The sensor is a pair of condensers and its electro-capacity is dependent upon the amount of water contained in its hydrophilic macromolecule (humidity sensitive) membrane. The components of the sensors are a glass foundation; a lower electrode; a macromolecule membrane and an upper electrode. The upper electrode will capture the change in permittivity caused by condensation or evaporation of water vapors to the macromolecule
membrane. (Electrostatic Capacity Type) There is another type of hygrometer that uses a resistor instead of a condenser. Typically Electrostatic Capacity types are good for low humidity measurements and the resistor types are good for high humidity measurements.
Theory of detecting static pressure (MODEL6501-B0/6501-C0) To measure static pressure, we use a diffusion-type semiconductor pressure sensor. The diffusion-type semiconductor pressure sensor is incorporated under the principle that the resonant frequency of a piezo-crystal decreases linearly with the pressure applied. On a thin diaphragm of silicon, there are four diffusion resistors (sensor chips) placed in a fixed distant apart. (Pic.1) When the pressure is applied from above, the diaphragm will deflect downward. When it is deflected, the sensors near the center (R3 and R4) will have compressing stress and the sensors near the perimeter (R1 and R2) will have tensile stress instead. (Pic.2) The diffusion coefficient of the diffusion resistor changes according to these two stresses.
If we create a bridge among these diffusion resistors (Pic.3), it is possible to detect voltage that is proportional to the amount of pressure applied. These sensors can be affected by the temperature, and therefore, a temperature compensation circuit must be added.
Pic.2
Upper Electrode
Lower El ectrode
Glass Foundation
Macromolecule Membrane
Diaphragm
Pressure Pic.1 Pressure Sensor
R1
R
R4
R1R4:Diffusion Resistant
Pic.3 Detecting Circuit
R1
R2
R4
R3
Current i
Output
Temp Sensor (Platinum Resistor)
Wind
actoolsupply.com
actoolsupply.com
55
What are Discomfort Index (DI) and Dew Point Temperature (DT)?
**What is DI and DT?**
Discomfort Index (DI)
Discomfort Index is an index to show how uncomfortable it is in the summer time. CLIMOMASTER use the formula, shown below, which is used by Meteorological Agencies.
(℃)     RH HumidityH:RelativeureT:Temperat
Where
THTDI
%
3.46)3.1499.0(01.081.0
DI Scale
Comfortable
Little
Uncomfortable
Uncomfortable
Very
Uncomfortable
Cannot
Endure
68 70 75 80 86
There are many formulas to calculate Dew Point Temperature,
but for A531, we have adopted the most general “Antnione” formula,
which is used to calculate saturated vapor pressure.
235)]ln(5614.23/[183.4030
::
:%:
100
322.133)]235/(183.40306686.18exp[
eDT
PaureVaporPressePareaporPressuSaturatedVE
eTemperaturTRHmidityRelativeHuH
Where
EH
e
TE
)    
(℃))     (
actoolsupply.com
actoolsupply.com
56 Principle of Measurement
About Compensation
This instrument has been calibrated at normal temperature and pressure. Therefore, if you are going to measure velocity at a different temperature and pressure, the indicating value will be affected.
Influence of Measuring Temperature
This instrument has been calibrated at normal temperature. Because of theoretical fundamentals of hot-wire anemometers, it can easily be effected by the ambient temperature. To prevent such influence, temperature compensation is needed. The temperature compensation sensor senses the ambient, or surrounding air temperature and forces the velocity sensor to stay at a constant overheat above the ambient temperature. By adopting temperature compensation sensors, you can measure air velocity accurately within a range of 560
o
C (41140
o
F)
Influence of Pressure at Measuring Point
The instrument is calibrated under atmospheric pressure of 1013 hPa. Since change in the atmospheric pressure will influence the heat dissipation amount, compensation of the atmospheric pressure is required. Compensation can be provided by using the following formula.
Where: Um: Actual Velocity[m/s]
Uc: Indicating Value Pm: Atmospheric Pressure at the Measuring Point [hPa]
Measuring Gas Components
This instrument has been calibrated using normal air (i.e. 79% nitrogen concentration & 21% oxygen concentration). If you wish to use the instrument in an environment other than normal air, you will need to compensate for the characteristics of the gas mixture
you’ll be using the instrument in. Please contact your local
Kanomax representative for details.
Uc
Pm
Um
1013
actoolsupply.com
actoolsupply.com
57
§ 14 Probe Directivity (Air Velocity) MODEL 6531-21 / 6541-21 / 6561-21
0°
180°
90°
90°
90°
90°
Horizontal
Wind Direction Indicator
At 5m/s
Vertical
Wind Direction Indicator
At 5m/s
actoolsupply.com
actoolsupply.com
58 Probe Directivity (Air Velocity)
Probe Directivity (Air Velocity) MODEL 6542-21
Probe Directivity (Air Velocity) M O D E L 6 5 3 3 - 2 1 / 6 5 4 3 - 2 1 / 6 5 5 1 - 2 1 / 6 5 5 2 - 21
Horizontal
At 1m/s
Vertical
Horizontal
Vertical
0°
180°
90°
90°
180°
0°
90°
90°
0°
180°
90°
90°
180°
0°
90° 90°
At 1m/s
actoolsupply.com
actoolsupply.com
59
§ 15 Trouble Shooting
Check battery
Problem
Possible cause(s) / Solution(s)
Reference
The instrument will not turn on
Battery is inserted in wrong polarity. Turn off the power and insert the battery correctly.
Page 11 &
13
The display is not visible
Display contrast is not set up properly. Adjust the contrast volume switch.
Page 50
The battery level indicator
flashes on the display
The battery is low. Turn off the instrument and replace the batteries.
Page 11 &
13
Initial operational check
Problem
Possible cause(s) / Solution(s)
Reference
The display is not visible
The display contrast is not set up properly. Adjust the contrast dial.
Page 50
[NO PROBE] is displayed
The probe is not connected. Turn off the instrument and connect a probe.
Page 12
Can I change the measurement unit?
Yes, you can change measurement units by following the steps on page 46.
Page 46
Normal operational check
Problem
Possible cause(s) / Solution(s)
Reference
[**.*] is displayed as measured value.
It indicates the measurement is outside the measuring range. The instrument must be used in the range stated in the specifications.
Page 52
[- - - - ] is displayed as air velocity measurement.
The probe or the probe cable is not connected properly. Check the connection between the probe and the probe cable.
Page 12
The probe or the probe cable may be damaged. Contact your local Kanomax representative for repair or replacement.
Page 62
The instrument is not reading the correct air speed.
The probe might be not facing into the wind. Make sure the wind direction indicator is facing into the wind.
Page 14
The instrument reads a higher temperature with no wind.
CLIMOMASTER can’t measure the correct
temperature with no wind (less than 0.1m/s of wind). *Model 6533 and 6543 can measure less than 0.1 m/s.
Page 14
The response time is slow
Set the correct time constant (TC).
Page 20
[ - - - - ] is displayed as the air volumetric flow measurement.
The duct information (shape and size) is not stored. Register duct information.
Page 16
actoolsupply.com
actoolsupply.com
60 Trouble Shooting
Output check (1) - Print out
Problem
Possible cause(s) / Solution(s)
Reference
The printer doesn’t
operate.
The printer is not connected properly. Confirm the printer connection.
Page 34
The Baud Rate might not be set properly. Check both the instrument and printer settings.
Page 46
Make sure to use the optional printer (Model DPU-S245) and printer cable (Model 6000-31).
Page 34
Confirm the connection procedure. Connect the printer to the printer port, then turn on the instrument and printer.
Page 34
Cannot printout the display
Display is not frozen. 1. Press to hold the display.
2. Press to print out.
Page 34
Want to cancel printing
You can’t cancel print while it is operating.
Page 34
Output check (2) - Digital output
Problem
Possible cause(s) / Solution(s)
Reference
PC doesn’t load the
measuring data.
Confirm the USB connection, which is different from printer connection.
Page 39
The Baud Rate might not be set properly. Check both the instrument and PC settings.
Page 34
Confirm the communication command (ex. Hyper Terminal command).
Page 39
Output check (3) - Analog output
Problem
Possible cause(s) / Solution(s)
Reference
No data is generated. Check the connection and confirm the polarity.
Page 43
The HOLD function might be active. Press to cancel the HOLD function.
Page 43
Data is tiered.
Data is outputted every one second.
Page 43
The output data is not correct.
Confirm the setting for analog output.
Page 43
Confirm the output range.
Page 43
The load impedance might not be optimum. Set the load impedance larger than 5k Ω.
Page 43
actoolsupply.com
actoolsupply.com
61
§ 16 Warranty and After Service
Kanomax Limited Warranty
The limited warranty set forth below is given by KANOMAX JAPAN, Inc. (hereafter referred to as “KJI”) with respect to the KANOMAX brand anemometer, and its attachment parts including probe and other accessories (hereafter referred to as “PRODUCT”) purchased directly from KJI or from and authorized distributor.
Your PRODUCT, when delivered to you in new condition in its original container, is warranted against defects in materials or workmanship as follows: for a period of two (2) years from the date of original purchase, defective parts or a defective PRODUCT returned to KJI, as applicable, and proven to be defective upon inspection, will be exchanged for a new or comparable rebuilt parts, or a refurbished PRODUCT as determined by KJI. Warranty for such replacements shall not extend the original warranty period of the defective PRODUCT.
This limited warranty covers all defects encountered in normal use of the PRODUCT, and does not apply in the following cases: (1) Use of parts or supplies other than the PRODUCT sold by KJI, which cause
damage to the PRODUCT or cause abnormally frequent service calls or
service problems. (2) If any PRODUCT has its serial number or date altered or removed. (3) Loss of damage to the PRODUCT due to abuse, mishandling, alternation,
improper packaging by the owner, accident, natural disaster, electrical
current fluctuations, failure to follow operation, maintenance or
environmental instructions prescribed in the PRODUCT’s operation manual
provided by KJI, or service performed by other than KJI.
NO IMPLIED WARRANTY, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE, APPLIES TO THE PRODUCT AFTER THE APPLICABLE PERIOD OF THE EXPRESS LIMITED WARRANTY STATED ABOVE, AND NO OTHER EXPRESS WARRANTY OR GUARANTY, EXCEPT AS MENTIONED ABOVE, GIVEN BY ANY PERSON OR ENTITY WITH RESPECT TO THE PRODUCT SHALL BIND KJI. KJI SHALL NOT BE LIABLE FOR LOSS OF STORAGE CHARGES, LOSS OR CORRUPTION OF DATA OR ANY OTHER SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES CAUSED BY THE USE OR MISUSE OF, OR INABILITY TO USE OR CONSEQUENTIAL DAMAGES CAUSED BY THE USE OR MISUSE OF, OR INABILITY TO USE, THE PRODUCT, REGARDLESS OF THE LEGAL THEORY ON WHICH THE CLAIMS IS BASED, AND EVEN IF KJI HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH DAMAGES. IN NO EVENT SHALL RECOVERY OF ANY KIND AGAINST KJI BE GREATER IN AMOUNT THAN THE PURCHASE PRICE OF THE PRODUCT SOLD BY KJI AND CAUSING THE ALLEGED DAMAGE. WITHOUT LIMITING THE FOREGOING, THE OWNER ASSUMES ALL RISK AND LIABILITY FOR LOSS, DAMAGE OF, OR INJURY TO THE OWNER AND THE OWNER'S PROPERTY AND TO OTHERS AND THEIR PROPERTY ARISING OUT OF USE OR MISUSE OF, OR INABILITY TO USE, THE PRODUCT NOT CAUSED DIRECTLY BY THE NEGLIGENCE OF KJI. THIS LIMITED WARRANTY SHALL NOT EXTEND TO ANYONE OTHER THAN THE ORIGINAL PURCHASER OF THE PRODUCT, OR THE PERSON FOR WHOM IT WAS PURCHASED AS A GIFT, AND STATES THE PURCHASER'S EXCLUSIVE REMEDY.
After Service
When you have a problem with your instrument, please check out
“Troubleshooting” first.
If that does not solve the problem, please contact your local distributor or call our
service center. (See last page for contact information.)
During the warranty period, we will repair at no charge a product that proves to be
defective due to material or workmanship under normal use. (Kanomax Limited Warranty.)
Repair after warranty expiration:
Upon request, we will repair the instrument at the customer’s expense, if the instrument’s performance is found to be recoverable by providing the repair.
Replacement parts are available for minimum period of five (5) years after
termination of production. This storage period of replacement parts is considered as the period during which we can provide repair service. For further information, please contact your local distributor or our service center.
actoolsupply.com
actoolsupply.com
62 Contact Information
§ 17 Contact Information
No copying, distribution, publication, modification, or incorporation of this document, in whole or part, is permitted for commercial purposes without the
express written permission of Kanomax.
The contents of this document may be changed without prior notice.
01002/12.04