Kanomax OMX-TDM User manual

Handheld Odor Meter
OMX-TDM
Operation Manual
Ver. 2
Find Quality Products Online at: sales@GlobalTestSupply.com
20160719
Table of Contents
1. Introduction............................................................................................................... 4
1-1. Supplied Components List ..................................................................................... 4
1-2. Warning ................................................................................................................... 5
1-3. Cautions .................................................................................................................. 6
1-4. Unit Dimension ...................................................................................................... 7
2. General Description .................................................................................................. 8
2-1. Features .................................................................................................................. 9
2-1-1. LCD Display ........................................................................................................ 9
2-1-2. Base Unit ........................................................................................................... 10
2-2. Accessories ............................................................................................................ 11
3. Measurement ........................................................................................................... 12
3-1. Real-time Sampling Mode .................................................................................... 12
3-1-1. Measurement Preparation ............................................................................... 12
3-1-2. Measurement .................................................................................................... 13
3-2. Memory Sampling Mode ...................................................................................... 13
3-2-1. Measurement Preparation ............................................................................... 13
3-2-2. Measurement .................................................................................................... 13
3-2-3. Memory Overflow .............................................................................................. 14
3-3. Peak-hold Function .............................................................................................. 15
3-4. Zero-point Function .............................................................................................. 15
4. Data Filing Software ................................................................................................ 16
4-1. How to Download Software.................................................................................. 16
4-2. How to Install Software ....................................................................................... 16
How to Install USD Drivers Software ......................................................................... 17
4-3. Preparation ........................................................................................................... 18
2
Find Quality Products Online at: sales@GlobalTestSupply.com
4-4. Open Software ...................................................................................................... 18
5. Specification ............................................................................................................. 19
6. Maintenance and Servicing ..................................................................................... 20
6-1. Replacement of “AA” Alkaline Batteries ............................................................ 20
6-2. Replacement of Button Battery ........................................................................... 20
6-3. Maintenance of Air Purifying Unit ..................................................................... 21
6-4. Replacement of Filter for Suction Nozzle ........................................................... 21
6-5. Troubleshooting .................................................................................................... 22
3
Find Quality Products Online at: sales@GlobalTestSupply.com
Name
Q’ty
OMX-TDM Base Unit
1
Suction Nozzle
1
AC Adaptor
1
‘AA’ Alkaline Battery
4
USB Connection Cable
1
Air Purifying Unit
1
Activated Carbon (50cc/bag)
1
Replacement Filter for Air Purifying Unit
4
Replacement Filter for Suction Nozzle
5
Operation Manual
1
1. Introduction 1-1. Supplied Components List
Thank you for purchasing of Shinyei Handheld Odor Meter.
Handheld Odor Meter is packaged carefully and includes all the components necessary for full
operation.
Upon receipt, check the contents and find you received the items as below in good condition.
The instrument kit contains:
4
Find Quality Products Online at: sales@GlobalTestSupply.com
1-2. Warning
Warning
Do not use any adaptors except supplied original AC adaptor. Do not use in explosive atmosphere. Do not use the device where combustible and/or explosive gasses exist.
The device is not designed for explosion proof.
Do not dismantle the base unit.
Find Quality Products Online at: sales@GlobalTestSupply.com
5
1-3. Cautions
Cautions
Do not install, use, or store the unit at a place where:
- Ambient temperature is not in the range of -10 to 40 degree in C. (14 to 104 degree in
F)
- Ambient humidity exceeds 90%R.H.
- Condensation occurs due to rapid temperature change.
- Corrosive and/or explosive gases exist.
- Direct vibration and/or stream occur.
- Mush dust, salty materials, and/or iron powders exist.
- Static electrons, magnetic, and/or electronic noise occur.
- Cool or warm air from conditioners influences directly.
- The unit is exposed to direct sunlight.
- Heating occurs due to heat radiation.
Do not use or store the unit at a place where the following gases exist:
- Extremely concentrated odor.
- Cigarette smoke and/or oil mist.
- Silicon compound gas included in hair spray, adhesive, and silicon rubber etc.
- High concentrated chloride, acidic substances, corrosive gas, and chlorofluorocarbon
etc.
Exchange new batteries if battery status displays “Low.” Do not use rechargeable batteries. Do not use liquid such as benzene and/or thinner to clean the unit surface. Do not block suction nozzle and/or exhaust nozzle.
Find Quality Products Online at: sales@GlobalTestSupply.com
6
1-4. Unit Dimension
Unit: mm
7
Find Quality Products Online at: sales@GlobalTestSupply.com
2. General Description
Handheld Odor Meter is the most popular simplified tool for odor analysis enables to show
TVOC in the value corresponding to toluene gas numerically by comparing odor gases and
purified air.
The reading shown on the display is calculated based on our original principle.
With the installation of data filing software, measured data can be transferred to PC in easy
steps.
The Value Corresponding to Toluene Gas
Simply turn on the unit and start measurement.
Real-time TVOC gas concentration corresponding to toluene gas can be measured continuously.
Peak-hold
Retains maximum gas concentration.
Memory Sampling Measurement Mode
Up to 511 measuring data can be saved upon selecting sampling rate.
Data Filing Software
By using original data filing software, measuring data from the unit can be transferred to PC.
8
Find Quality Products Online at: sales@GlobalTestSupply.com
2-1. Features
No.
Name
Function
Ready
Blinks for about 10min for initializing when the unit is turned on.
Start measuring when it stops blinking.
TVOC gas concentration
corresponding to toluene gas
Displays TVOC gas concentration corresponding to toluene gas
from 0 to 9999 in µg/m³.
Calendar Function
Upper: Year-Month-Date
Lower: Hour-Time
Peak Hold
Retains maximum gas concentration.
Battery Status
■■■■ Full
■■■□ Good – Measureable for 3 to 4 hours approx.
■■□□ Moderate – Measureable for 1 to 2 hours approx.
■□□□ Low – Change battery immediately.
□□□□ No battery.
USB
USB Cable is connected.
Data No.
Shows number of saved data.
Memory
Displays remaining memory capacity.
Sampling Rate
Select sampling rate.
NOTE “RT” refers to Real-time Sampling that measures gas every 1
second continuously.
2-1-1. LCD Display
Find Quality Products Online at: sales@GlobalTestSupply.com
9
2-1-2. Base Unit
No.
Name
Function
Hold 3 seconds for “peak-hold function”. (See 3-3.) Hold and turn on the device at the same time to set calendar.
Select sampling rate. (See 3-2.) Set calendar.
Use for “zero-point function". (See 3-4.) Set calendar.
Confirm sampling rate and/or “zero-point” function.
Power Switch
Turn on/off the unit.
AC Adaptor Port
Connect AC adaptor if necessary.
Suction Nozzle
Takes sampling air and/or object gases.
Exhaust Nozzle
Exhaust sampling air and/or object gases.
USB Port
Connect attached USB cable for PC connection.
Battery Box
Open/Close to replace batteries.
Hook Pin
Use for straps.
Calendar Settings
1. Hold and turn on the device at the same time to set calendar.
2. When “Year” is blinking, change to the arbitrary number.
3. Press to confirm the settings.
4. Repeat 2. And 3. To set month, date, house, and minute.
10
Find Quality Products Online at: sales@GlobalTestSupply.com
2-2. Accessories
Name
Function
AC Adaptor
Connect to the unit during the measurement if necessary.
Air Purifying Unit
Attach to the unit for “zero-point function”. (See 3-4.)
Activated Carbon
Exchange activated carbon when deodorization using air
purifying unit is not effective.
Replacement Filter for Air
Purifying Unit
Replace filter when deodorization using air purifying unit is not
effective.
Replacement Filter for Suction
Nozzle
Replace filter when gets dirt.
USB Connection Cable
Use to transfer measuring data from the unit to PC.
11
Find Quality Products Online at: sales@GlobalTestSupply.com
3. Measurement 3-1. Real-time Sampling Mode
3-1-1. Measurement Preparation
1. Place filter to suction nozzle.
2. Attach suction nozzle to the unit before measurement.
3. Use “AA” alkaline batteries or AC adaptor.
or
4. Turn on the unit at clean ambient.
It takes about 10 min for the unit initializing and start measuring.
NOTE The unit sets “zero” point when the initializing is done.
NOTE To get the zero point stable, initializing at clean environment is recommended.
NOTE To get “Zero” point forcibly, use air purifying unit. (See 3-4.)
12
Find Quality Products Online at: sales@GlobalTestSupply.com
3-1-2. Measurement
1. Turn on the unit and wait for about 10 minutes until “Ready” stop blinking and the
initializing is done.
2. Start measuring.
3. Use data filing software if necessary.
4. Turn off the unit when odor strength shows close to “0”.
3-2. Memory Sampling Mode
3-2-1. Measurement Preparation
Refer to Page. 11.
3-2-2. Measurement
1. Turn on the unit and wait for about 10 minutes until “Ready” stop blinking and the
initializing is done.
2. Press until sampling rate on LCD display starts blinking.
3. Select sampling rate by pressing.
4. Press to finish selecting sampling rate.
5. Press (one time) and (twice) to save data.
13
Find Quality Products Online at: sales@GlobalTestSupply.com
3-2-3. Memory Overflow
Memory overflow occurs when the internal memory is full.
If there is no space remained, calendar on LCD display starts blinking and data number becomes
“0”.
To delete saved memory, press and at the
same time while the unit is turned off.
Keep pressing and and turn on the device.
NOTE Saved memory in the unit will not be deleted automatically when it is transferred to PC.
14
Find Quality Products Online at: sales@GlobalTestSupply.com
3-3. Peak-hold Function
Press for over 3 seconds during Real-time sampling.
By using this function, the highest value of odor strength keeps being displayed until the unit
measures higher value.
NOTE Press for over 3 seconds again to cancel the function.
NOTE The function is not displayed on the saved data.
3-4. Zero-point Function
By using this function, the unit sets the baseline accurately when it is turned on.
The unit sets the baseline at the place where the initialization is done.
After 10 minutes initializing, the unit will display Odor Strength “0” (baseline) automatically
regardless of the measurement environment.
If you need to set the baseline accurately, connect Air Purifying Unit to purify the air forcibly.
1. Press until calendar on LCD display “0” and starts blinking.
2. Press to proceed, or press to cancel.
15
Find Quality Products Online at: sales@GlobalTestSupply.com
TDM)
4. Data Filing Software
Data filing software is used to transfer data from the unit to PC by using USB connection cable.
4-1. How to Download Software
NOTE Register personal information in order to download the file.
4-2. How to Install Software
Open “SetupTDM_en.msi” and follow the instruction on the window.
NOTE Install software under Windows administrative privileges.
16
Find Quality Products Online at: sales@GlobalTestSupply.com
How to Install USD Drivers Software
Install USB Drivers software before connecting USB cable to PC.
1. Download the latest software from the link as below.
2. Install the software.
NOTE If PC CPU is 32bit, select
“CP210xVCPInstaler_x86.exe”.
If CPU is 64bit, select
“CP210xVCPInstaler_x64.exe”.
NOTE Warning message appears if
incorrect file is selected.
3. Confirm Device Manager displays “Silicon Labs CP210x USB to UART Bridge (COM)”
after the installation.
4. Confirm assigned COM number.
17
Find Quality Products Online at: sales@GlobalTestSupply.com
4-3. Preparation
Connect the unit to PC by USB connection cable.
4-4. Open Software
1. Double-click shortcut icon on desktop, or open the file from program files.
2. Confirm main screen appears.
3. Read “Handheld Odor Meter Data Filing Software Manual” for details.
18
Find Quality Products Online at: sales@GlobalTestSupply.com
5. Specification
Detection Method
Semiconductor gas sensors
Object Gas
TVOC, etc.
Sampling Method
Continuous sampling by built-in pump
Detection Range
TVOC gas concentration corresponding to toluene gas from 0 to
9999 in µg/m³.
Display
Digital LCD display
0 to 9999 in µg/m³
Response Time
Toluene (500µg/m³): Reaches 90% responsiveness within 40 seconds
Toluene (1500µg/m³): Reaches 90% responsiveness within 60 seconds
Toluene (6000µg/m³): Reaches 90% responsiveness within 60 seconds
Power Source
2 “AA” alkaline batteries or AC adaptor
Consumed Power
1,100mW approx. with “AA” alkaline batteries
Storage Temp. Range
0 to 40 degree in C
Dimensions
74(W) x 167.50(L) x 35(D)
Memory
32,732 data or 511 files
19
Find Quality Products Online at: sales@GlobalTestSupply.com
6. Maintenance and Servicing
6-1. Replacement of “AA” Alkaline Batteries
Replace “AA” alkaline batteries when battery status on LCD display becomes “Low”.
6-2. Replacement of Button Battery
The unit uses a button battery (model: CR2032) for memory back up.
When to replace:
- Pasts 5 years approx. after purchasing.
- Confirms when calendar or time on LCD display is incorrect.
1. Turn of the unit and remove “AA” alkaline batteries.
NOTE Do not connect AC adaptor.
2. Push the round part of button battery firmly until the left part of button battery moves up.
3. Take out the battery.
NOTE Do not use metal tools or easily broken stools like
screwdrivers or pencils.
4. Insert the new battery.
20
Find Quality Products Online at: sales@GlobalTestSupply.com
Top
Top Wire Gauze
Top Filter
Arrow Label
Activated Carbon
Bottom Filter
Bottom Wire Gauze
Bottom
Filter
Suction
Nozzle
Do not exceed the line.
6-3. Maintenance of Air Purifying Unit
Replace activated carbon and filters in Air Purifying Unit to
prevent gases absorbed in the unit.
1. Twist the Air Purifying Unit to open.
2. Take out top wire gauze, top filter, activated carbon, and bottom filter.
NOTE Do not take out bottom wire gauze.
3. Replace bottom filter and fill in activated carbon up to the line.
NOTE Do not exceed the line when replace activated carbon.
4. Replace top filter and top wire gauze.
6-4. Replacement of Filter for Suction Nozzle
Twist Suction Nozzle and replace filter if necessary.
21
Find Quality Products Online at: sales@GlobalTestSupply.com
6-5. Troubleshooting
If you find:
Check:
The unit does not operate. Check battery status or connect
AC adaptor.
Check battery polarity.
Odor Strength does not change.
Close the unit to object.
Connect Air Purifying Unit to
clean the absorbed gas in the
unit.
Odor Strength on LCD display keeps blinking.
Odor concentration is too high and
overflowed.
Change display range or try to lower
the gas concentration.
Time on LCD display keeps blinking.
Memory is overflowed.
Delete data.
Sampling rate cannot be selected.
Memory is overflowed.
Delete data.
Data # becomes “0”.
Memory is overflowed.
Delete data.
“SErr” is shown on LCD display.
Sensor malfunction.
Contact us of the message keeps
being displayed.
“PErr” is shown on LCD display.
Built-in pump malfunction.
Contact us of the message keeps
being displayed.