NOTE: This manual contains important information concerning safety and daily use as
well as maintenance instructions and should be stored for future use.
This manual is valid for all cabinets that are delivered after 2001.
Manufactured by:
KANMED AB
January 2004
PART NO: GE-1002-070/3
USER and TECHNICAL MANUAL KANMED Warming Cabinet GE-2-1370 Page 2 of 16
11 CIRCUIT DIAGRAM FOR CABINETS DELIVERED UNTIL 2003 .................................... 12
12 CIRCUIT DIAGRAM FOR CABINETS DELIVERED FROM 2004.................................... 13
USER and TECHNICAL MANUAL KANMED Warming Cabinet GE-2-1370 Page 3 of 16
1 SAFETY INSTRUCTIONS
INTENDED USE
KANMED Warning Cabinet GE-1380-XX is primarily intended for warming of Gel Pads,
blankets, fluids, up to a temperature of 42°C
KANMED Warning Cabinet shall be used according to this user manual and in accordance
with normal hospital routines.
GENERAL ADVICE
Adjust the cabinet so that it is correctly positioned, vertically and horizontally.
Secure the cabinet to the wall using the built in brackets at the top.
Never pull out all shelves/baskets at the same time. The cabinet can tilt forward if it is not
secured to the wall.
Take extra caution if the cabinet is mounted on wheels, especially when the cabinet is moved
around.
Note! If the cabinet is not secured to a wall, don’t’ pull out more than one shelf/basket at a
time.
Do not overload the shelves/basket - maximum 8 kilos per shelf. Max 2 pcs. of gel pads
(45x50x1.5) Max load of the basket is 20kg.
Do not overfill the top shelf - there must be at least 5 cm's free space to ensure air
circulation.
Don't block the ventilation holes at the inside rear of the cabinet.
EXPLANATION OF SYMBOLS
Consult Users Manual
O OFF
I ON
~ AC current
UP button, rises temperature
DOWN button, decreases temperature
SET button, for displaying desired temperature
USER and TECHNICAL MANUAL KANMED Warming Cabinet GE-2-1370 Page 4 of 16
2 GENERAL DESCRIPTION
THE CABINET
The cabinet is made of stainless steel and insulated in order to reduce heat loss to ambience
and to reduce noise. For the same reason the door is double glassed.
THE HEATING COMPARTMENT
Heating element, fan, thermostat, temperature regulator/indicator and electrical connections
are all placed on a "shelf" at the top of the cabinet. A spare fan is also included and mounted
in the heating compartment. The heating compartment is kept in place by a small screw that,
when loosened, allows the whole heating compartment to be pulled completely out to make
service quick and easy.
THE SHELVES-BASKETS
The shelves/baskets run on wheels and can be pulled fully out until they are automatically
stopped. To completely remove the shelves/basket for cleaning or repositioning, lift the front
of about 5 cm upwards and remove it. The shelves can carry maximum 8 kilos, basket 20 kg.
EXTRA SHELVES- EXTRA BASKETS
Extra shelves/baskets and rails are available.
Mount the rails with the screws supplied using the prepared mounting holes and then push
the shelf/basket in place by holding the front higher than the rear.
3 DESCRIPTION OF FUNCTION
The warming cabinet and its contents are warmed by warm air heated by a 1000 Watt
warming element. The hot air is circulated by a fan and distributed evenly through the outlets
at the rear of the cabinet. The temperature is regulated, by the temperature regulator T1 to
the set temperature.
Thermostat T2 functions as an over temperature protection and will take over the
temperature control in case the air temperature exceeds 45°C. At the same time the red
lamp on the front panel will be lit to indicate that there is a malfunction. Inside the heating
element there is an additional over temperature protection that is self-resetting. It will be
activated in case the fan stops or goes too slow.
On old versions of the cabinets a power switch at the door interrupts the power when the
door is opened. When restarted again the temperature goes back to the set temp as it was
before interrupted.
On the new cabinets this power switch is removed
NOTE: If the red over temperature lamp is lit there is an error that requires a technician.
USER and TECHNICAL MANUAL KANMED Warming Cabinet GE-2-1370 Page 5 of 16
4 TECHNICAL DATA
ELECTRICAL DATA:
Voltage 220-240 Volt AC
Power max. 1200 Watt *
Frequency 50 -60 Hz
Fuses T 6,3A L250V
PHYSICAL DATA
Cabinet GE-1380
Height 175 cm
Width 66 cm
Dept 60 cm
GE-41500
Shelves inner measure
Height 2 cm (edge)
Width 52 cm
Dept 50 cm
Working temperature 35°C – 42°C, selectable in steps of 1°C
(Of circulating air)
Over temperature protection Visual alarm 45°C
Accuracy ± 1°C
Hysteresis (T2 overtemp.) 4°C max
WARMING CAPACITY
Warming of 10 gel pads GE-455015 3 hours
45 x 50 x 1.5 cm from 22°C to 40°.
MODE OF OPERATION Designed for continuous use
CE - MARKING According to 89/366/EEC and 93/68/EEC
(EN 60601-1-2, EN55011 (1991)
* The nominal effect of the heating element 1000 W. Due to the thermostat regulation the
average power consumption is about 400 W