Kane KM940 User Manual

KM940
Hand-held Combustion
Analyser
Page No:
1. ANALYSER LAYOUT AND FEATURES............................................................ 4-7
1.1 Instrument features and keypad...........................................................................4
1.2 Instrument layout (Rear) .....................................................................................5
1.3 Standard Probe Configuration.............................................................................6
1.4 Analyser connections ..........................................................................................7
2. SAFETY WARNING ...................................................................................................8
3. FIRST TIME USE ........................................................................................................8
4. NORMAL START UP SEQUENCE...................................................................... 9-15
4.1 Every Time You Use The Analyser ....................................................................9
4.2 Automatic Calibration .........................................................................................9
4.3 Main Display Parameters ..................................................................................11
4.3.1 4 Page Mode.........................................................................................11
4.3.2 Line Scroll Mode..................................................................................12
4.3.3 8 Page Mode.........................................................................................12
4.4 Sampling the Flue Gas ......................................................................................13
4.5 Taking a Pressure Reading................................................................................13
4.6 Regular Checks During Sampling.....................................................................14
4.7 Normal Shutdown Sequence .............................................................................14
4.8 Electromagnetic Compatibility .........................................................................15
5. MOVING THROUGH THE MENUS.................................................................. 16-22
5.1 Basic Operation .................................................................................................16
5.2 Menu Options and Settings ...............................................................................17
5.2.1 Main Menu ............................................................................................17
5.2.2 Select Menu ..........................................................................................17
5.2.3 Units Menu............................................................................................19
5.2.4 Display Menu ........................................................................................20
5.2.5 Set-Up Menu .........................................................................................20
6. PRINTING INFORMATION.....................................................................................23
6.1 Printing a ‘Live’ Test..........................................................................................23
6.2 Standard Printout ................................................................................................23
7. STORING AND RETRIEVING DATA .............................................................. 24-25
7.1 Storing a ‘Live’ Test .........................................................................................24
7.2 Viewing and Printing a ‘Stored’ Test................................................................24
7.3 Deleting Data.....................................................................................................25
- 2 -
8. MAINTENANCE .......................................................................................................26
8.1 Emptying and Cleaning the in-line water trap ..................................................26
8.2 Changing the particle filter................................................................................26
9. PROBLEM SOLVING ...............................................................................................27
10. ANNUAL RE-CALIBRATION .................................................................................27
11. PRODUCT SPECIFICATION ............................................................................. 28-29
APPENDICES:
A. MAIN DISPLAY PARAMETERS ............................................................................30
B. COMBUSTION EFFICIENCY CALCULATION............................................... 32-33
C. CALCULATION OF FUEL DATA...........................................................................34
D. ELECTROMAGNETIC COMPATABILITY STATEMENT ...................................35
- 3 -
1. ANALYSER LAYOUT AND FEATURES
1.1 Instrument Features and Keypad
RS232 (8 pin din)
Charger socket
Store
Print
Pump
Enter
ON/OFF
MENU
Allows access to all menu functions
PUMP
Turns pump on and off
ENTER
Accepts a command ie enters a menu option
Infra-red emitter
Pressure Inlet
Inlet Port (side)
Menu
On/Off
Up
Down
Keypad
Flue Temperature Connector (underside)
UP
Scrolls up through options ie Fuel
DOWN
Scrolls down through options
STORE
Enters data storage menu
PRINT
Prints current data
- 4 -
1.2 Instrument Layout (Rear)
e
Options Label
Exhaust Port
Charger Socket
SENSORS FITTED
NOTE! - Cover Does Not Remov
*NOTE! Do not cover exhaust port as this will
severely affect analyser operation
- 5 -
Serial Number
1.3 Standard Probe Configuration
- 6 -
1.4 Analyser Connections
Gas Connection to Instrument
Particle Filter
In-line Water Trap
Neoprene Hose
Flue Temperature Connector
- 7 -
2. SAFETY WARNING
This analyser extracts combustion gases that may be toxic in relatively low concentrations. These gases are exhausted from the side of the instrument. This instrument must only be used in well ventilated locations. It must only be used by trained and competent persons after due consideration of all the potential hazards.
Protection Against Electric Shock (in accordance with EN 61010-1 : 1993)
This instrument is designated as Class III equipment and should only be connected to SELV circuits. The battery charger is designated as:
Class II equipment Installation category II Pollution degree 2 Indoor use only Altitude to 2000m Ambient temperature 0°C-40°C Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to 50%RH at 40°C Mains supply fluctuations not to exceed 10% of the nominal voltage.
3. FIRST TIME USE
Charge the battery for 12 hours, following this an overnight charge should be sufficient for an average 8 hour day. See Main Parameter displays for Battery Indicator.
The KANE940 has a rechargeable lead acid battery which uses a different charger than other Kane analysers. Ensure the correct charger is used or damage may occur to the instrument.
Check that you have all the items you have ordered. Take time to read this manual fully. When using the analyser for the first time you will need to choose from:-
Language selection Calibration countdown time CO gas alarm NOx percentage for calculation Time and Date Printed header name and telephone number
The SET UP MENU (Section 5.2.5) gives details of how to change the above settings.
- 8 -
4. NORMAL START UP SEQUENCE
4.1 Every Time You Use The Analyser
BEFORE SWITCH-ON CHECK THAT:
the particle filter is not dirty
the water trap and probe line are empty of water
all hose connections, etc, are properly made
the probe is sampling CLEAN AMBIENT air
the water trap is correctly fitted and the instrument upright
the flue temperature is connected
Switch ON the instrument by pressing
4.2 Automatic Calibration
During this sequence the analyser pumps fresh air into the sensors to allow toxic sensors (if fitted) to be set to zero and the Oxygen sensor to be set to 20.9 %.
After switch-on the analyser will briefly display header information :-
Kane International
(44)-1707-375550
And then show the countdown screen :-
ZERO CAL Time : 180
FRESH AIR PURGE
The calibration time will count down in seconds to zero. Calibration time may be changed from 2 to 6 minutes. See Set-Up menu section 5.2.5.
Note ! Three minutes is recommended to allow the sensors to stabilise fully. Anything less
than this may result in drift of the toxic and oxygen sensors in clean ambient air.
To obtain the quoted specification an instrument should be calibrated with clean ambient air at standard temperature and pressure (STP).
- 9 -
Once the time has reached zero an audible beep will be heard and will show the selected fuel on the following display:-
NATURAL GAS
PRESS -MENU- KEY
Press
This zeros the toxic sensor and sets Oxygen to 20.9%. The next screen is the MAIN DISPLAY of the analyser:-
NETT C . . . . 0.0
O2 % . . . 20.9 CO ppm . . . 0000 EFF (G) % . . . 0.0
Use and to change the display.
CO2 % . . . . . . . 0.0
FLUE C. . . . . . 0.0
INLT . . . . NOT FITTED
AMBIENT C . . . . 21
All parameters are detailed in Appendix A - MAIN DISPLAY PARAMETERS.
- 10 -
4.3 Main Displays
The main display can be changed to show either 4 or 8 parameters at one time. Two options are available when 4 parameters are selected.
4 Page Mode displays 4 lines of data in set format, each page is predefined.
Line scroll mode allows you to customise the display to show the data you require.
8 Page Mode displays 8 parameters on 4 lines in set format, the bottom two can be
changed.
Changing between the different modes is detailed in Display Menu Section 5.2.4.
4.3.1 4 Page Mode
Use the and keys to change the information that is displayed on the screen. The following pages are available.
NATURAL GAS DATE . . . 07-08-96 TIME . . . 12:31:35 BATTERY % . . 54
NETT C . . . 0.0 O2 % . . . 20.9 CO ppm . . 0000 EFF (G) % . . . 0.0
CO2 % . . . 0.0 FLUE C . . . 0.0 INLT NOT FITTED AMBIENT C . . . 21
CO/CO2 R . . 0.0001 P INDEX % . . . 0.01 XAIR % . . . 0.0 Prs mbar 0.00
This screen only displayed on an analyser fitted with an NO sensor
NO ppm . . 0000 NOx ppm . . 0000 NOx calc% . . . 5 O2 ref % . . . 3.0
TIP - In 4 page mode only turns the backlight ON and OFF.
- 11 -
4.3.2 Line Scroll Mode
Line scroll mode allows you to customise the display.
Use the and keys to change the bottom line of the display. Once the correct line is displayed press to confirm and move the line up. Select the next parameter and repeat until all lines display the desired parameters.
Change bottom line using and
NETT C . . . 0.0 O2 % . . . 20.9 CO ppm . . 0000 CO2 % . . . 0.0
to select and move parameter up
O2 % . . . 20.9 CO ppm . . 0000 CO2 % . . . 0.0 CO2 % . . . 0.0
Select next parameter. Repeat above until display reads desired data
O2 % . . . 20.9 CO ppm . . 0000 CO2 % . . . 0.0 CO/CO2 R . . 0.0001
4.3.3 8 Page Mode
Displays 8 parameters on the screen at one time. Symbols used in this mode are different to those used in 4 page and line scroll modes and are detailed in Appendix A - MAIN DISPLAY PARAMETERS.
O2
CO PI
λ
The bottom line of the display can be changed to display other parameters.
Use the and keys to change this line.
- 12 -
:20.9 % : 0ppm : - ­: - -
CO2 Eff
T Tf
: - ­: - ­: 0C : 21C
4.4 Sampling the Flue Gas
Once the automatic calibration procedure has been completed and the specific fuel has been selected (See SELECT menu) the probe can be inserted into the desired sampling point.
It is recommended that the sampling point be located at least two flue diameters downstream of any bend and that the probe tip is in the centre of the flue. With balanced flues and other domestic units the probe should be positioned far enough into the flue so that no air can ‘back flush’ into the probe. This will be indicated by a low oxygen reading and/or a low ‘Poison Index’ reading.
The probe depth stop cone provided with the instrument allows the probe to be used in holes whose diameters range from 8 mm to 21 mm (5/16 to
The standard probe is rated at 650°C. Temperatures of up to 1100°C (2012°F) can be accommodated using an optional high temperature probe.
TIP To conserve battery power, switch off the pump when you are not taking a
measurement. Use the key to turn ON and OFF the pump.
13
/16 inch).
4.5 Taking a Pressure Reading
With the optional pressure module fitted a flue draught measurement can be made at any time.
Connect the standard probe to the pressure sensor inlet and the probe in the flue. The pressure reading will be displayed :-
CO/CO2 R . . 0.0001 P INDEX % . . . 0.01 XAIR % . . . 0.0 Prs mbar 0.00
To perform a combustion test and display draught pressure at the same time a special probe is
required. Contact Kane International or Authorised Distributor for details.
- 13 -
4.6 Regular Checks During Sampling
Care must be taken at all times not to exceed the analysers operating specifications, in particular ensure the following :-
Do not exceed the maximum temperature of the flue probe.
The analyser internal temperature does not exceed normal operating range, typically 0- 40°C.
DO NOT PLACE THE INSTRUMENT ON A HOT SURFACE.
The water trap is vertical at all times. Water condenses in the probe line and can quickly
fill the water trap when the probe is moved. Take care and watch the water trap closely.
The in-line particle filter is clean and does not become blocked.
4.7 Normal Shutdown Sequence
DO THIS EVERY TIME YOU USE THE ANALYSER
Remove the probe from the flue - TAKE CARE ! THE PROBE WILL BE HOT - and allow it to cool naturally. Do not immerse the probe in water as this will be drawn into the analyser and damage the pump and sensors.
Once the probe is removed from the flue press and the analyser will count down from 30 to switch off.
OFF 30
MENU TO ESCAPE
If you have not finished but press by mistake, you can press to return to normal operation and not switch OFF.
- 14 -
4.8 Electromagnetic Compatibility
The European Council Directive 89/336/EEC requires that electronic equipment does not generate electromagnetic disturbances that exceed defined levels and has an adequate level of immunity to enable it to be operated as intended. The specific standards applicable to this product are detailed in the appendices.
Since there are many electrical products in use that pre-date this Directive and may emit electromagnetic radiation in excess of the standards defined in the Directive there may be occasions where it would be appropriate to check the analyser prior to use.
The following procedure should be adopted:
Go through the normal start up sequence in the location where the equipment is to be used.
Switch on all localised electrical equipment that might be capable of causing interference.
Check that all readings are as expected. (A level of disturbance in the readings is acceptable). If not adjust the position of the instrument to minimise interference or switch off, if possible, the offending equipment for the duration of the test.
N.B. Maximum cable lengths must be less than 3 metres.
At the time of writing this manual (November 2005) Kane International Ltd is not aware of any field based situation where such interference has ever occurred and this advice is only given to satisfy the requirements of the Directive.
- 15 -
5. MOVING THROUGH THE MENUS
5.1 Basic Operation
From the MAIN DISPLAY
Press to access the MAIN MENU
Press and to move cursor up and down
Press to access selected Menu
NETT C . . . . 0.0
O2 % . . . 20.9 CO ppm . . . 0000 EFF (G) % . . . 0.0
MAIN MENU
1. SELECT 3. DISPLAY
2. UNITS 4. SETUP
MAIN MENU
1. SELECT 3. DISPLAY
2. UNITS 4. SETUP
MAIN MENU
1. SELECT 3. DISPLAY
2. UNITS 4. SETUP
Press to select parameter
FUEL
O2 Ref SMOKE RESET
Use and to change setting i.e. fuel selected
FUEL O2 Ref SMOKE RESET
Press to enter value and move to next parameter
FUEL
O2 Ref
SMOKE RESET
Press to save settings and return to the MAIN MENU
1. SELECT 3. DISPLAY
2. UNITS 4. SETUP
Press to return to the MAIN DISPLAY
: LIGHT OIL : OFF : OFF : NO
: NATURAL GAS : OFF : OFF : NO
: LIGHT OIL : OFF : OFF : NO
MAIN MENU
- 16 -
5.2 Menu Options and Settings
5.2.1 Main Menu
The MAIN MENU consists of 4 sub menus which are shown below and detailed on the following pages.
MAIN MENU
1. SELECT 3. DISPLAY
2. UNITS 4. SETUP
All sub-menus are accessed using and exited using
The and keys move the cursor within a menu and allow
parameters to be changed.
TIP Holding down one of these keys scrolls through the data quicker.
5.2.2 Select Menu
FUEL
O2 Ref SMOKE RESET
This menu allows selections to be made for the parameters detailed below.
FUEL : Select the fuel being used by the boiler from either a standard fuel stored in the
analyser or by entering the user fuel. Once the correct fuel has been selected press to view the fuel constants.
: NATURAL GAS : OFF : OFF : NO
NATURAL GAS
K1g K_2 K_4
: 0.350 : 11.89 : 32
K1n K_3 O2r
: 0.390 : 9.83 : 3.0
- 17 -
Calculation of fuel constants are detailed in the Appendix. Fuel constants will
have to be calculated before a user fuel can be entered.
To enter the user fuel select ‘User Fuel’ and Press
USER FUEL
K1g K_2 K_4
Use and to select the correct value.
: 0.000
: 0.00 : 00
K1n K_3 O2r
: 0.000 : 0.00 : 00
USER FUEL
K1g K_2 K_4
: 0.350
: 0.00 : 0
K1n K_3 O2r
: 0.000 : 0.00 : 00
Use to move to the next parameter, repeat above until all parameters
are correct. Press to return to SELECT menu.
O2 Ref : Toxic gas measurements can be referenced to defined oxygen levels. Reference
values can be set from 1-20%, to AUTO or more normally to the default value
- OFF. Setting to AUTO uses the figure in the FUEL constants data.
Oxygen referencing is required by some regulations such as TA-LUFT. If a
reference value is selected then toxic gas measurements will be displayed with the symbol (n) attached to the reading. i.e. CO(n)
What does Oxygen reference mean ?
If 3 % O
gas values will be recalculated as if 3 % were measured. The equation for referencing is detailed in the Appendix.
Oxygen referencing prevents false readings being submitted, e.g. allowing
more air into the boiler will increase the oxygen level in the flue and hence dilute any toxic gas reading. Oxygen referencing gives readings as if they were undiluted.
SMOKE: Allows the user to enter a smoke test number from 0-9. This value will be
printed on the standard printout. Default value is OFF.
RESET : Allows the user set the Oxygen to 20.9% and zero the toxic sensors without
turning the analyser off.
- 18 -
reference is selected and 5 % O2 is measured in the flue then toxic
2
Selecting YES and will display the following screen.
RESET SENSORS
O2 % : 20.9 CO & NO = 0
PRESS ENTER
MENU TO ESCAPE
After pressing the analyser will count down for 5 seconds and then return to the main display.
WARNING : The sensors must only be reset if you are sure they have been
sampling fresh air for at least 3 minutes. Errors in measurement will occur if the sensors are reset during or just after sampling.
5.2.3 Units Menu
TEMP
GAS PRESS. EFF.
Allows all displayed units to be changed.
TEMP : Choose selections from Centigrade °C or Fahrenheit °F.
GAS : Changes the toxic gas measurement units. Select from volumetric readings,
parts per million (ppm) or mass flow reading milligrams per cubic meter (mg/m3) .
PRESS. : Flue draught can be displayed in millibar (mbar), hectaPascals (hPa),
millimeters water gauge (mmWG) or inches water gauge (in WG).
EFF. : Efficiency can be selected for Gross or Net values. Gross efficiency assumes
latent heat of vapourisation is lost in the boiler and hence will be lower than Net efficiency. For Natural Gas the difference will be approximately 11%.
: C : ppm : mbar : GROSS
- 19 -
5.2.4 Display Menu
LIGHT
MODE CONTRAST
Allows the configuration of the display to be changed.
LIGHT : Choose from ON or OFF.
MODE : Select 4 or 8 Page Mode or Line Scroll Mode as detailed in section 4.3 Main
Displays.
CONTRAST :The contrast is set to a DEFAULT value or can be adjusted ↑ LIGHTER or ↓ DARKER. Use the and keys to
adjust.
: OFF : 8-PAGE : DEFAULT
5.2.5. Set-Up Menu
The set up menu allows the following parameters to be set / altered.
Language.
Automatic calibration time
CO gas alarm
NOx percentage for calculation
Date and time
Printout Header
LANG
CO AL
CALEN
: ENGLISH ARM : 400 DAR
ZERO : 3 NOx% : 5 HEADER
Parameter Description Settings
LANG :
ZERO :
CO ALARM :
Changes the analysers displayed and printed language.
Allows setting of the Autocalibration time in minutes. Care must be taken when changing this parameter as sensors may drift from zero if too short a time is used. Kane International advise 3 minute countdown.
Allows an alarm level to be set on for the CO reading. This is set as a default at 1000 ppm.
ENGLISH SPANISH DUTCH FRENCH ITALIAN
2-6 minutes
OFF 0-4000 ppm
- 20 -
Once an alarm has been exceeded the display will flash every two minutes warning the user of an alarm state and display the gas concentration. A similar display will be shown during a RECHARGE BATTERY and PUMP OFF alarms.
-- -- -- -- -- -- -- -- -- -- -- -- -- -­CO ALARM
1010 ppm
-- -- -- -- -- -- -- -- -- -- -- -- -- --
NO REF:
CALENDAR:
Displayed on the Nitric Oxide unit only. Allows the percentage P in the following calculation to be set. The default value set is 5%. Note the percentage allows for NO2 in a typical boiler.
NOx = NO + P% NO
Allows the user to change the date and time. (24 hour clock). The following screen will be shown once the parameter is entered:
OFF 1-9 %
TIME : FORMAT : DATE :
FORMAT :
To change the time position the cursor on Time and press . The cursor will now be to the left of the 13 as shown below:
Changes the date format for display and printing.
TIME : FORMAT : DATE :
Using and scroll through the setting options i.e. 0-23.
Once the correct hour is set press to move to the next parameter, the cursor will move to the left of minutes (53). Move to each parameter until the correct time is set. Pressing after setting the seconds will return the cursor to the left of the screen.
Format and Date are set in a similar manner.
hh
: mm : ss
13
: 53 : 26
dd
: mm : yy
03
: 01 : 96
dd : mm : yy yy : mm : dd mm : dd : yy
hh
: mm : ss
13
: 53 : 26
dd
: mm : yy
03
: 01 : 96
- 21 -
Header :
Allows two lines of 20 characters to be programmed into the analyser. The header appears on the top of the standard printout. This can be used to print your company name and/or phone number.
Name/Phone
K
ane International
(44)-1707-375550
‘LEFT’ USE STORE KEY
The screen above shows the standard header setting with the cursor now shown underlining the K in Kane. By using and any letter or number can be chosen.
Once the correct character is displayed, use to move right to the next. Move along until all characters spell the desired name or phone number. If you need to go back and change a character use to move left.
Press to return to the SET UP menu.
- 22 -
6. PRINTING INFORMATION
Supplied as accessories for the KANE940 are an infra-red thermal printer or a dot matrix serial printer. Read the manual supplied with each printer prior to operation. Connections to the KANE940 are detailed below :
Infra-red thermal printer - this does not require a cable to transmit the data but uses an
infra-red (IR) link similar to a TV remote control. The IR emitter is positioned on the top of the KANE940 and the bottom of the printer. Ensure they are pointing at each other and within 300 mm, with no obstructions in the way. Data may be lost if transmission is interrupted. Keep the KANE940 pointing at the printer until the printout has finished.
Dot matrix serial printer - requires the supplied serial cable to transmit data. Connect the
cable to the 8 pin DIN socket on the top of the KANE940 and the 25 pin D-connector on the printer.
Data can either be printed from a ‘live’ test or from stored data. Printing of stored data is detailed in STORING AND RETREIVING DATA.
6.1 Printing a ‘Live’ Test
During a combustion test the KANE940 will print data on request. With the analyser showing the MAIN DISPLAY press and current data will be sent to the printer.
The display will show the following until data transmission is complete.
6.2 Standard Printout
The standard printout is shown below :-
∗∗∗∗∗ Printing ∗∗∗∗∗
- 23 -
7. STORING AND RETRIEVING DATA
The KANE940 can store up to 100 combustion tests. Once stored, the data can be viewed on the display or downloaded to a PC or printer.
7.1 Storing a ‘Live’ Test
While performing a test and viewing the data on the MAIN display access the STORE menu as follows :-
Press STORE MENU to access the STORE MENU MODE
LOCATION
Mode: Select from the following :-
STORE - Allows data to be stored in memory.
VIEW / PRINT - Stored data can be viewed or printed.
DELETE - Clears all data in memory.
Location: Automatically allocates a location in the memory of the instrument for the next
test. On the display shown above the next location will be 3.
To store a test set MODE to STORE and press . The current readings will be stored in the analysers memory.
Tip : Make a note of the location number for your particular test as it may be useful
when downloading or printing.
PRESS ‘STORE’ TO LOG
: STORE : 3
7.2 Viewing and Printing a ‘Stored’ Test
Multiple tests can be printed easily with the KANE940.
Select PRINT under MODE in the STORE menu. This feature is in addition to the VIEW/PRINT, STORE and DELETE options.
Press STORE MENU to access the STORE MENU MODE
LOCATION
PRESS ‘PRINT’
: PRINT : 1 TO 10
The cursor will move to the first number, use the and to select the location and start printing.
Press to move the cursor to the second number, select the last location to print.
- 24 -
To print the data press . In the screen shown above locations 1 to 10 will be printed.
During printing the following will be shown.
PRINT TESTS
1 to 10
PRINTING TEST 1
NOTE While the display above is shown (i.e. the instrument is printing a test) the keypad is disabled. To exit from printing wait until the current test has finished and the display below is shown:
Press to exit the print routine. The instrument will return to main display
PRINT TESTS
1 to 10
PLEASE WAIT
MENU TO ESCAPE
7.3 Deleting Data
To delete the data in stored memory press to obtain the STORE MENU (as above) :-
Press STORE MENU to access the STORE MENU MODE
LOCATION
PRESS ‘ENTER’ TO DELETE
: DELETE : 3
Press
ENTER to ERASE DATA
to access delete data screen
MENU to ESCAPE
Press to delete data in memory, press to exit delete data screen.
- 25 -
8. MAINTENANCE
8.1 Emptying and Cleaning the In-line Water Trap
The in-line water trap should be checked and emptied on a regular basis. Water vapour will condense and gather in the probe line. This may move suddenly to the trap when the probe is moved. Care should be taken at all times.
Emptying of the water trap is detailed below :-
Carefully remove the end cap from the in-line housing. Dispose of the condensate in a suitable drain, care must be taken as it could be acidic. If condensate spills onto the skin or clothing, clean off immediately using fresh water, seek medical advice if problems occur.
8.2 Changing the Particle Filter
This is a very important part of the analyser and should be changed regularly. It prevents dust and dirty particles entering the pump and sensors and hence causing damage. The filter MUST be changed when it is discoloured.
Remove the end cap from the in-line filter housing. Carefully remove the paper filter element and dispose of it. Clean the inside of the filter housing with a suitable soft cloth. Insert a new filter element onto the spigot in the filter housing and carefully replace the end cap.
- 26 -
9. PROBLEM SOLVING
The following is a list of problems that may occur on the instrument through its operating life. If the cause of the fault is not easy to identify then we advise you contact Kane International Service Department or an International Distributor for expert advice.
Fault symptom Causes
Oxygen too high
CO2 too low
Oxygen Error (FAULT)
Toxic sensor Error (FAULT)
Analyser not holding charge
Analyser not charging
Analyser does not respond to flue
gas
Flue temperature readings erratic Temperature plug reversed in socket.
Analyser automatically switches
off in operation.
Display shows dark lines and no
reponse from ON/OFF key.
Air leaking into probe, tubing, water trap,
connectors or internal to instrument.
Oxygen cell needs replacing.
Calibration time set too short and
instrument not allowed to stabilise
Instrument has been stored in a cold
environment and is not at normal working temperature.
Oxygen cell or toxic sensors needs
replacing.
Battery exhausted.
AC charger not giving correct output.
Fuse blown in charger plug.
Particle filter blocked.
Probe or tubing blocked.
Pump not working or damaged with
contaminents.
Probe connected to pressure connector.
Faulty connection or break in cable or plug.
Battery below alarm level.
Ambient temperature above 50°C.
Battery quickly discharging and is faulty.
Fault has occured on the instrument
electronics and requires resetting. Contact Kane International or Distibutor.
10. ANNUAL RE-CALIBRATION
Whilst the sensors have an expected life of more than two years in normal use it is recommended that the analyser is re-calibrated at least annually. This is so that long term drift on the sensors and electronics can be eliminated. Local regulation may require more frequent re-calibration and users should check with appropriate authorities to ensure they comply with relevant guidelines.
- 27 -
11. PRODUCT SPECIFICATION
Parameter Resolution Accuracy Range
Temp Measurement
Flue Temperature with probe
1.0ºC/F +2.0ºC +0.3% reading 0-600ºC 32-1112ºF
Inlet Temperature 0.1ºC/F +1ºC +0.3% reading 0-50ºC/32-122ºF
Pressure
Gas Measurement *1
0.01 mbar +2% of full scale +150mbar to -150mbar
Oxygen 0.1% +0.2% 0-21%
Carbon Monoxide
(standard: H compensated)
1ppm +20ppm <400ppm
+5%<5000ppm
0-10,000ppm
+10%>5000ppm
Carbon Monoxide
(high range)
Nitric Oxide
(standard)
Nitric Oxide
(low range)
0.01% +5% reading from 0.1%
to 10%
1ppm +5ppm<100ppm
+5%>100ppm
1 ppm +2ppm<30ppm
+5ppm>30ppm
0 - 10%
0-5000ppm
0 -100ppm
Nitrogen Dioxide
1 ppm +5ppm<100ppm
+10ppm<500ppm
0 - 1000ppm
+5%>500ppm
Sulphur Dioxide 1ppm +5ppm<100ppm
0-5000ppm
+5%>100ppm
Calculations
Carbon Dioxide*2 0.1% +0.3% reading 0-99.9%
Losses 0.1% +1.0% reading 0-99.9%
Efficiency 0.1% +1.0% reading 0-99.9%
Excess Air 0.1% +0.2% 0-2885.0%
Temp (Nett) 1.0ºC/F +2ºC +0.3% reading 0-600ºC/32-1112ºF
CO/CO2 ratio 0.0001 +0.0001 0-0.9999
Poison Index 0.01% +0.01 0-99.99
Pre-programmed Fuels
Natural gas, Town gas, Gascor, Light Oil, Heavy Oil, Propane, Butane, Anthracite, Coke, Coal, Kinsale Gas
Dimensions
Weight Handset Probe
1kg 220mm x 55mm x120mm L 420mm x Dia 8mm with stainless steel shaft, type K thermocouple and 3m hose
Ambient operating
0oC to 45oC/ 10% to 90% RH non condensing
range
Power supply
(battery charger)
Battery Life
Input 110Vac / 220Vac nominal Output: 12Vac off load
>6 hours from full charge
*1
Using dry gases at STP
*2
Calculated
- 29 -
APPENDICES
A - Main Display Parameters
The parameters and their meanings are detailed as follows : -
DATE : Analyser date. See Set-Up menu section 5.2.5 to change.
TIME : Analyser time. Use Set-Up menu section 5.2.5 to change.
BATTERY : Displays the battery level from 0-100%. The analyser will flash (BAT) RECHARGE BATTERY at less than 10 % of charge. With the charger
connected the display shows AC ON.
NETT : Nett temperature calculated by deducting the internal AMBIENT
temperature
(T) from the measured FLUE temperature. Displays in either °C (C) or °F (F)
and will display NOT FITTED (N/F) if flue probe is not connected.
If an external INLET probe is used then INLET is deducted from FLUE.
O2 : Oxygen reading in percentage %.
CO : Carbon Monoxide reading indicated in ppm or mg/m3. If the figures are
referenced to oxygen then the display will show CO(n). See SELECT menu
5.2.2 for oxygen reference. The display will read
‘O2 > 20%’ if referenced values selected and instrument is in clean ambient
air.
EFF (G) : Combustion Efficiency calculation displayed in percentage. Gross G or Net
N can be set see SELECT menu 5.2.3. The calculation is determined by fuel type see Appendix B for calculation. The efficiency is displayed during a combustion test, ‘- -’ is displayed while in fresh air.
CO2 : Carbon Dioxide calculation determined by the type of fuel. This only shows
a reading when a combustion test is being carried out. ‘- -’ is displayed while in fresh air.
FLUE : Temperature measured by flue gas probe in Centigrade or (Tf) Fahrenheit. Will show ambient temperature after fresh air calibration and NOT FITTED (N/F) or FAULT (FLT) if probe disconnected.
INLET : Temperature measured by the optional inlet air probe. This probe is (Ti) plugged into the instrument through the RS232 socket. This figure is
used to calculate the NET temperature instead of AMBIENT when fitted.
AMBIENT : Temperature measured by the internal sensor, used in the NET
- 30 -
(Ta) temperature calculation if an INLET probe is not fitted.
CO/CO2 R : The CO/CO2 ratio, is the ratio of measured CO divided by calculated CO2.
It gives an indication of the following :-
How good a gas sample the instrument is reading.
How clean the boiler is running.
For example : A new or clean domestic boiler will display a ratio of less
than 0.004, a unit in need of cleaning 0.004-0.008 and a unit in need of major overhaul will show greater than 0.008.
This only shows a reading when a combustion test is being carried out. ‘- -’
is displayed while in clean ambient air.
P INDEX : The CO/CO2 ratio expressed as a percentage %, called the ‘Poison (PI) Index’ i.e. P INDEX % = 100 x CO/CO2. ‘- -’ is displayed while in clean ambient air.
XAIR % : Excess air calculated from the measured oxygen and type of fuel (λ) used. During a combustion test ‘O2 > 20%’ will be displayed while in
clean ambient air.
Prs : Flue draught pressure reading. Displayed when pressure sensor fitted. See
UNITS menu 5.2.3. for scales.
NO: Nitric Oxide reading in ppm or mg/m3. Displayed when Nitric Oxide sensor
fitted. Also displayed as NO (n) when referenced to oxygen. The display will read ‘O2 > 20%’ if referenced values selected and instrument is in clean ambient air.
NOx : Calculated total Nitric oxides displayed in ppm or mg/m3. Where NOx =
NO + P%NO, note P can be set from 0-9%, default = 5%. See SELECT menu 5.2.2. Also displayed as NOx (n) referenced to oxygen. The display will read ‘O2 > 20%’ if referenced values are selected and instrument is sampling clean ambient air.
SO2 : Sulphur Dioxide reading in ppm or mg/m3. Displayed when Sulphur
Dioxide sensor fitted. Also displayed as SO2 (n) referenced to oxygen. The display will read ‘O2 > 20%’ if referenced values selected and instrument is in clean ambient air.
O2 ref % : Toxic gas measurements can be referenced to defined oxygen levels. (O2r) See SELECT menu 5.2.2 for details.
- 31 -
B. COMBUSTION EFFICIENCY CALCULATION
The efficiency calculation is based upon British Standard BS845.
This identifies three sources of loss associated with fuel burning:
Losses due to flue gasses: Dry Flue gas loss, Moisture and hydrogen Sensible heat of water vapour Unburned gas
Losses due to refuse: Combustible in ash Combustible in riddlings Combustible in dust
Other losses: radiation convection conduction other unmeasured losses
Net efficiency calculations assume that the energy contained in the water vapour (formed as a product of combustion and from wet fuel) is recovered and the wet loss term is zero. Gross efficiency calculations assume that the energy contained in the water vapour is not recovered.
Since the fuel air mixture is never consistent there is the possibility of unburned/partially unburned fuel passing through the flue. This is represented by the unburned carbon loss.
Losses due to combustible matter in ashes, riddlings, dust and grit, radiation, convection and conduction are not included.
Efficiency Calculation:
Known Data - Fuel: Qgr = Gross Calorific Value (kJ/kg) Qnet = Net Calorific Value (kJ/kg) K1 = Constant based on Gross or Net Calorific Value: K1g = ( 255 x %Carbon in fuel )/Qgr K1n = ( 255 x %Carbon in fuel )/Q
net
K2 = % max theoretical CO2 (dry basis) K3 = % Wet Loss H H
= % Hydrogen
2
O = % Water
2
Measured Data: Tf = Flue Temperature Ti = Inlet Temperature O O
m = % Oxygen in flue gas
2
r = Oxygen reference %
2
- 32 -
Calculated data: Tnet = Net Temperature % CO2 content in flue gas % Dry Flue Gas losses % Wet losses % Unburned carbon loss % Efficiency
Tnet = Flue Temperature - Inlet Temperature
Dry flue gas loss % = 20.9 x K1 x (Tnet) / K2 x (20.9 - O2m)
Wet loss % = 9 x H2 + H2O / Qgr x [2488 + 2.1Tf - 4.2 Ti]
simplified = [(9 x H2 + H2O) / Qgr] x 2425 x [1 + 0.001 Tnet]
Wet loss % = K3(1+0.001xTnet)
Where K3 = [(9 x H2 + H2O) / Qgr] x 2425
Net Efficiency % = 100 - dry flue gas losses
= 100 - 20.9 x K1n x (Tnet) / K2 x (20.9 - O2m)
Gross Efficiency % = 100 - {dry flue gas losses + wet losses}
= 100 – {[20.9 x K1g x (Tnet) / K2 x (20.9 - O2m)]+ [K3 x (1 + 0.001 x Tnett)]}
Excess Air = [(20.9% / (20.9% - 02m%)) – 1] x 100%
CO2% = [(20.9 - O2m) x K2 / 20.9]
Unburned fuel Loss % = K4 x CO / ( CO + CO
) Note: CO scaled in %
2
Where K4 = 70 for coke = 65 for anthracite = 63 for Bituminous coal = 62 for coal tar fuel = 48 for liquid petroleum fuel = 32 for natural gas
The formula for K4 is based on the gross calorific value Qgr. To obtain the loss based on net calorific value multiply by Qgr/Qnet. Since this loss is usually small this conversion has been ignored. This loss is subtracted from the efficiency.
Oxygen Reference CO(n) = CO x
(20.9 - O
r)
2
(20.9 - O2m)
- 33 -
C. CALCULATION OF FUEL DATA
For any fuel not specified by Kane International the net calorific value, gross calorific value and composition should be obtained from the fuel supplier.
The following fuel data has been calculated with reference to the efficiency calculation.
Example 1:
Chemical composition: C 25% H H Q Q Max CO
K1n = (255 x % carbon in fuel) / Q = (255 x 25) / 8350 = 0.763
K1g = (255 x % carbon in fuel) / Qg (kJ/Kg) = (255 x 25) / 9300 = 0.685
K2 = Max % CO2 = 20.40
K3 = Wet Loss = [(9 x %H2 + %H2O) / 9300] x 2425
= [(9 x 3 + 50) / 9300] x 2425 = (77 / 9300) x 2425 = 20.08
K4 = 65 (an approximation for wood) *
The fuel values to program into the Analyser are as follows:
3%
2
O 50%
2
8.35 MJ/kg
net
9.3 MJ/kg *
g
20.4%
2
(kJ/Kg)
net
NATURAL GAS
K1g K_2 K_4
: 0.763 : 20.4 : 65
* Assumed values in the absence of supplied data. See previous appendix for other fuels.
- 34 -
K1n K_3 O2r
: 0.685 : 20.08 : 8.0
D. ELECTROMAGNETIC COMPATABILITY (CE) STATEMENT
This product has been tested for compliance with the following generic standards:
EN 61000-6-3 EN 61000-6-1
and is certified to be compliant
Specification EC/EMC/KI/KANE940 details the specific test configuration, performance and conditions of use.
Please Note: Batteries used in this instrument should be disposed of in
accordance with current legislation and local guidelines.
- 35 -
Loading...