A. MAIN DISPLAY PARAMETERS ............................................................................30
B. COMBUSTION EFFICIENCY CALCULATION............................................... 32-33
C. CALCULATION OF FUEL DATA...........................................................................34
D. ELECTROMAGNETIC COMPATABILITY STATEMENT ...................................35
- 3 -
1. ANALYSER LAYOUT AND FEATURES
1.1 Instrument Features and Keypad
RS232 (8 pin din)
Charger socket
Store
Print
Pump
Enter
ON/OFF
MENU
Allows access to all menu functions
PUMP
Turns pump on and off
ENTER
Accepts a command ie enters a
menu option
Infra-red emitter
Pressure Inlet
Inlet Port (side)
Menu
On/Off
Up
Down
Keypad
Flue Temperature
Connector (underside)
UP
Scrolls up through options ie Fuel
DOWN
Scrolls down through options
STORE
Enters data storage menu
PRINT
Prints current data
- 4 -
1.2 Instrument Layout (Rear)
e
Options Label
Exhaust Port
Charger Socket
SENSORS FITTED
NOTE! - Cover
Does Not Remov
*NOTE! Do not cover exhaust port as this will
severely affect analyser operation
- 5 -
Serial Number
1.3 Standard Probe Configuration
- 6 -
1.4 Analyser Connections
Gas Connection
to Instrument
Particle Filter
In-line Water Trap
Neoprene Hose
Flue Temperature
Connector
- 7 -
2. SAFETY WARNING
This analyser extracts combustion gases that may be toxic in relatively low concentrations.
These gases are exhausted from the side of the instrument. This instrument must only be used
in well ventilated locations. It must only be used by trained and competent persons after due
consideration of all the potential hazards.
Protection Against Electric Shock (in accordance with EN 61010-1 : 1993)
This instrument is designated as Class III equipment and should only be connected to SELV
circuits. The battery charger is designated as:
Class II equipment
Installation category II
Pollution degree 2
Indoor use only
Altitude to 2000m
Ambient temperature 0°C-40°C
Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to 50%RH
at 40°C
Mains supply fluctuations not to exceed 10% of the nominal voltage.
3. FIRST TIME USE
Charge the battery for 12 hours, following this an overnight charge should be sufficient for an
average 8 hour day. See Main Parameter displays for Battery Indicator.
The KANE940 has a rechargeable lead acid battery which uses a different charger than other
Kane analysers. Ensure the correct charger is used or damage may occur to the instrument.
Check that you have all the items you have ordered.
Take time to read this manual fully.
When using the analyser for the first time you will need to choose from:-
Language selection
Calibration countdown time
CO gas alarm
NOx percentage for calculation
Time and Date
Printed header name and telephone number
The SET UP MENU (Section 5.2.5) gives details of how to change the above settings.
- 8 -
4. NORMAL START UP SEQUENCE
4.1 Every Time You Use The Analyser
BEFORE SWITCH-ON CHECK THAT:
the particle filter is not dirty
the water trap and probe line are empty of water
all hose connections, etc, are properly made
the probe is sampling CLEAN AMBIENT air
the water trap is correctly fitted and the instrument upright
the flue temperature is connected
Switch ON the instrument by pressing
4.2 Automatic Calibration
During this sequence the analyser pumps fresh air into the sensors to allow toxic sensors (if
fitted) to be set to zero and the Oxygen sensor to be set to 20.9 %.
After switch-on the analyser will briefly display header information :-
Kane International
(44)-1707-375550
And then show the countdown screen :-
ZERO CAL
Time : 180
FRESH AIR PURGE
The calibration time will count down in seconds to zero. Calibration time may be changed
from 2 to 6 minutes. See Set-Up menu section 5.2.5.
Note ! Three minutes is recommended to allow the sensors to stabilise fully. Anything less
than this may result in drift of the toxic and oxygen sensors in clean ambient air.
To obtain the quoted specification an instrument should be calibrated with clean ambient air
at standard temperature and pressure (STP).
- 9 -
Once the time has reached zero an audible beep will be heard and will show the selected fuel
on the following display:-
NATURAL GAS
∗PRESS -MENU- KEY∗
Press
This zeros the toxic sensor and sets Oxygen to 20.9%. The next screen is the MAIN DISPLAY of the analyser:-