Kane KM940 User Manual

KM940
Hand-held Combustion
Analyser
Page No:
1. ANALYSER LAYOUT AND FEATURES............................................................ 4-7
1.1 Instrument features and keypad...........................................................................4
1.2 Instrument layout (Rear) .....................................................................................5
1.3 Standard Probe Configuration.............................................................................6
1.4 Analyser connections ..........................................................................................7
2. SAFETY WARNING ...................................................................................................8
3. FIRST TIME USE ........................................................................................................8
4. NORMAL START UP SEQUENCE...................................................................... 9-15
4.1 Every Time You Use The Analyser ....................................................................9
4.2 Automatic Calibration .........................................................................................9
4.3 Main Display Parameters ..................................................................................11
4.3.1 4 Page Mode.........................................................................................11
4.3.2 Line Scroll Mode..................................................................................12
4.3.3 8 Page Mode.........................................................................................12
4.4 Sampling the Flue Gas ......................................................................................13
4.5 Taking a Pressure Reading................................................................................13
4.6 Regular Checks During Sampling.....................................................................14
4.7 Normal Shutdown Sequence .............................................................................14
4.8 Electromagnetic Compatibility .........................................................................15
5. MOVING THROUGH THE MENUS.................................................................. 16-22
5.1 Basic Operation .................................................................................................16
5.2 Menu Options and Settings ...............................................................................17
5.2.1 Main Menu ............................................................................................17
5.2.2 Select Menu ..........................................................................................17
5.2.3 Units Menu............................................................................................19
5.2.4 Display Menu ........................................................................................20
5.2.5 Set-Up Menu .........................................................................................20
6. PRINTING INFORMATION.....................................................................................23
6.1 Printing a ‘Live’ Test..........................................................................................23
6.2 Standard Printout ................................................................................................23
7. STORING AND RETRIEVING DATA .............................................................. 24-25
7.1 Storing a ‘Live’ Test .........................................................................................24
7.2 Viewing and Printing a ‘Stored’ Test................................................................24
7.3 Deleting Data.....................................................................................................25
- 2 -
8. MAINTENANCE .......................................................................................................26
8.1 Emptying and Cleaning the in-line water trap ..................................................26
8.2 Changing the particle filter................................................................................26
9. PROBLEM SOLVING ...............................................................................................27
10. ANNUAL RE-CALIBRATION .................................................................................27
11. PRODUCT SPECIFICATION ............................................................................. 28-29
APPENDICES:
A. MAIN DISPLAY PARAMETERS ............................................................................30
B. COMBUSTION EFFICIENCY CALCULATION............................................... 32-33
C. CALCULATION OF FUEL DATA...........................................................................34
D. ELECTROMAGNETIC COMPATABILITY STATEMENT ...................................35
- 3 -
1. ANALYSER LAYOUT AND FEATURES
1.1 Instrument Features and Keypad
RS232 (8 pin din)
Charger socket
Store
Print
Pump
Enter
ON/OFF
MENU
Allows access to all menu functions
PUMP
Turns pump on and off
ENTER
Accepts a command ie enters a menu option
Infra-red emitter
Pressure Inlet
Inlet Port (side)
Menu
On/Off
Up
Down
Keypad
Flue Temperature Connector (underside)
UP
Scrolls up through options ie Fuel
DOWN
Scrolls down through options
STORE
Enters data storage menu
PRINT
Prints current data
- 4 -
1.2 Instrument Layout (Rear)
e
Options Label
Exhaust Port
Charger Socket
SENSORS FITTED
NOTE! - Cover Does Not Remov
*NOTE! Do not cover exhaust port as this will
severely affect analyser operation
- 5 -
Serial Number
1.3 Standard Probe Configuration
- 6 -
1.4 Analyser Connections
Gas Connection to Instrument
Particle Filter
In-line Water Trap
Neoprene Hose
Flue Temperature Connector
- 7 -
2. SAFETY WARNING
This analyser extracts combustion gases that may be toxic in relatively low concentrations. These gases are exhausted from the side of the instrument. This instrument must only be used in well ventilated locations. It must only be used by trained and competent persons after due consideration of all the potential hazards.
Protection Against Electric Shock (in accordance with EN 61010-1 : 1993)
This instrument is designated as Class III equipment and should only be connected to SELV circuits. The battery charger is designated as:
Class II equipment Installation category II Pollution degree 2 Indoor use only Altitude to 2000m Ambient temperature 0°C-40°C Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to 50%RH at 40°C Mains supply fluctuations not to exceed 10% of the nominal voltage.
3. FIRST TIME USE
Charge the battery for 12 hours, following this an overnight charge should be sufficient for an average 8 hour day. See Main Parameter displays for Battery Indicator.
The KANE940 has a rechargeable lead acid battery which uses a different charger than other Kane analysers. Ensure the correct charger is used or damage may occur to the instrument.
Check that you have all the items you have ordered. Take time to read this manual fully. When using the analyser for the first time you will need to choose from:-
Language selection Calibration countdown time CO gas alarm NOx percentage for calculation Time and Date Printed header name and telephone number
The SET UP MENU (Section 5.2.5) gives details of how to change the above settings.
- 8 -
4. NORMAL START UP SEQUENCE
4.1 Every Time You Use The Analyser
BEFORE SWITCH-ON CHECK THAT:
the particle filter is not dirty
the water trap and probe line are empty of water
all hose connections, etc, are properly made
the probe is sampling CLEAN AMBIENT air
the water trap is correctly fitted and the instrument upright
the flue temperature is connected
Switch ON the instrument by pressing
4.2 Automatic Calibration
During this sequence the analyser pumps fresh air into the sensors to allow toxic sensors (if fitted) to be set to zero and the Oxygen sensor to be set to 20.9 %.
After switch-on the analyser will briefly display header information :-
Kane International
(44)-1707-375550
And then show the countdown screen :-
ZERO CAL Time : 180
FRESH AIR PURGE
The calibration time will count down in seconds to zero. Calibration time may be changed from 2 to 6 minutes. See Set-Up menu section 5.2.5.
Note ! Three minutes is recommended to allow the sensors to stabilise fully. Anything less
than this may result in drift of the toxic and oxygen sensors in clean ambient air.
To obtain the quoted specification an instrument should be calibrated with clean ambient air at standard temperature and pressure (STP).
- 9 -
Once the time has reached zero an audible beep will be heard and will show the selected fuel on the following display:-
NATURAL GAS
PRESS -MENU- KEY
Press
This zeros the toxic sensor and sets Oxygen to 20.9%. The next screen is the MAIN DISPLAY of the analyser:-
NETT C . . . . 0.0
O2 % . . . 20.9 CO ppm . . . 0000 EFF (G) % . . . 0.0
Use and to change the display.
CO2 % . . . . . . . 0.0
FLUE C. . . . . . 0.0
INLT . . . . NOT FITTED
AMBIENT C . . . . 21
All parameters are detailed in Appendix A - MAIN DISPLAY PARAMETERS.
- 10 -
4.3 Main Displays
The main display can be changed to show either 4 or 8 parameters at one time. Two options are available when 4 parameters are selected.
4 Page Mode displays 4 lines of data in set format, each page is predefined.
Line scroll mode allows you to customise the display to show the data you require.
8 Page Mode displays 8 parameters on 4 lines in set format, the bottom two can be
changed.
Changing between the different modes is detailed in Display Menu Section 5.2.4.
4.3.1 4 Page Mode
Use the and keys to change the information that is displayed on the screen. The following pages are available.
NATURAL GAS DATE . . . 07-08-96 TIME . . . 12:31:35 BATTERY % . . 54
NETT C . . . 0.0 O2 % . . . 20.9 CO ppm . . 0000 EFF (G) % . . . 0.0
CO2 % . . . 0.0 FLUE C . . . 0.0 INLT NOT FITTED AMBIENT C . . . 21
CO/CO2 R . . 0.0001 P INDEX % . . . 0.01 XAIR % . . . 0.0 Prs mbar 0.00
This screen only displayed on an analyser fitted with an NO sensor
NO ppm . . 0000 NOx ppm . . 0000 NOx calc% . . . 5 O2 ref % . . . 3.0
TIP - In 4 page mode only turns the backlight ON and OFF.
- 11 -
Loading...
+ 24 hidden pages